Skip to main content
. 2023 Jun 23;14(7):3597–3609. doi: 10.1364/BOE.494289

Table 1. Calculation of theoretical and experimental SNR for Raman data of biological samples.

Signal Metric Milk Tissue Units
Total signal intensity 14900, 25330 24140, 41038 ADU, e-
Raman intensity (1440 cm−1) 4559, 7750 4333, 7366 ADU, e-
Signal shot noise (1440 cm−1) 159 203 e-
Signal-free noise sum 7 7 e-

Signal standard deviation  Silent region, 1800-1900cm−1 45 58 ADU
Uncorrected, 1440 cm−1 201 304 ADU
MSC-corrected, 1440 cm−1 95 110 ADU

SNR1440 Theoretical 49 36 a.u.
Experimental (silent region) 101 75 a.u.
Experimental (uncorrected) 23 14 a.u.
Experimental (MSC-corrected) 48 39 a.u.