Skip to main content
. 2023 Jul 25;77(Suppl 2):S125–S133. doi: 10.1093/cid/ciad326

Table 2.

Tests Used and Their Reported Performance

Pathogen Type of Test Test Name and Manufacturer Test Performance by Manufacturer and Independent Studies
Streptococcus pyogenes (Group A streptococci) Lateral flow RDT: detects Group A streptococcal antigen from throat swabs OSOM Strep A, Sekisui Diagnostics Sensitivity 96%; specificity 98%
(vs culture; from Manufacturer) [18]
Sensitivity 98%; specificity 99% (vs culture) [19]
Sensitivity 86% (83.3–87.6); specificity 95% (94.5–96.2) [20]
Streptococcus pneumoniae Lateral flow RDT: detects S. pneumoniae antigen in the urine of patients with pneumococcal pneumonia BinaxNOW  Streptococcus pneumoniae antigen card, Abbott/Alere Sensitivity 86%; specificity 94% (from Manufacturer; urine test) [21]
Sensitivity 74%; specificity 97% (from retrospective data) [22]
Influenza virus Lateral flow RDT: detects influenza virus type A, type B and A (H1N1) pandemic antigens directly from nasal/throat/nasopharyngeal swab or nasal/nasopharyngeal aspirate SD Bioline influenza Ag A/B/A(H1N1) PANDEMIC, Abbott/Alere Sensitivity/specificity not available not available from Manufacturer [23]
Sensitivity 56%; specificity 100% (vs RT-PCR) [24]
RSV Lateral flow RDT: detects respiratory syncytial virus fusion protein antigen in nasal wash and nasopharyngeal swab specimens Alere BinaxNOW® RSV Card, Abbott Sensitivity 93%; specificity 93% (from Manufacturer; prospective NP swab); sensitivity 89%; specificity 100% (retrospective nasal wash) [25]
Sensitivity 81%; specificity 93% (vs culture) [26]
Salmonella enterica serovar Typhi Lateral flow RDT: detects Salmonella typhi-specific IgM from serum or whole blood Test-it™ Typhoid IgM lateral flow assay, Life Assay Diagnostics Pty Ltd Sensitivity/specificity by Manufacturer (not available) [27]
Sensitivity 69% (95% CI: 59%–78%) for age >1 y old; specificity 90% (95% CI: 78%–93%) [28]
Denguea Immunochromatographic assay SD Bioline Dengue duo (NS1 Ag/IgG/IgM) Sensitivity NS1 92%, IgM/IgG 94% (vs RT-PCR); specificity NS1 98%, IgG/IgM 96% (vs ELISA; From Manufacturer) [29]
Sensitivity 89%; specificity 100% (vs0 ELISA) [30]
Plasmodium sp. (malaria) Lateral flow RDT as per national guidelines SD Bioline Malaria Ag P.F/PAN, Abbott Sensitivity 100% (Pf), 95%(non-Pf); specificity 99% (from Manufacturer) [31]
Sensitivity 99% (Pf), 93% (non-Pf); specificity, 98% (Pf), and 100%, non-Pf) [32]
CRP Immunoassay for the quantitative measurement of CRP level in human serum, plasma and whole blood Standard F CRP, SD Biosensor Coefficient of variation 7.6%–8.1% (CRP; from Manufacturer) [33]
White blood cell differentiation Five-part differentiation of white blood cell lines HemoCue® WBC DIFF System, Hemocue Measuring range: 0.3–30.0 × 109/L
(From Manufacturer) [34, 35]
Reliable comparability in the range of 0.4–30.0 × 09/L
(vs calibrated reference blood cell analyzer) [36]
Urine tests Urine dipstick test for urinary blood, bilirubin, ketones, pH, urobilinogen, protein, nitrites, leucocyte esterase and specific gravity Multistix SG-10, Siemens
SD UrocolorTM 10, Abbott
Multistix SG-10, Siemens
Sensitivity 88% (visual reading) 97% (instrument reading); specificity 93% (vs culture; from Manufacturer) [37]
Sensitivity nitrite, leukocyte esterase, blood and protein 97%
Specificity for nitrite, leukocyte esterase and blood 97% (94.2–98.6) (vs culture) [38]
SD Urocolor TM 10, Abbott
Sensitivity/specificity by Manufacturer (not available)

Abbreviations: CI, confidence interval; CRP, C-reactive protein; ELISA, enzyme-linked immunosorbent assay; IgG, immunoglobulin G; IgM, immunoglobulin M; NP, nasopharyngeal; Pf, P. falciparum; RDT, rapid diagnostic test; RSV, respiratory syncytial virus; RT-PCR, reverse transcription polymerase chain reaction; WBC, white blood cell.

Dengue tests were not performed in Uganda and Ghana.