Skip to main content
. 2023 Jul 12;14:1230533. doi: 10.3389/fmicb.2023.1230533

TABLE 3.

Comparing HIV-1 levels in clinical samples using RT-qPCR and our HIV-1-RT-LAMP-AuNPs-LFA methods.

HIV-1-LAMP-AuNPs-LFA HIV-1 RT-qRCR (reference method) Sensitivity Specificity
Positive Negative Total Value 95% CI Value 95% CI
Positive 38 (CRF01_AE, 16; CRF07_BC, 14; CRF08_BC, 4; Subtype B, 4) 0 38 100% 90.75–100.00% 100% 87.23–100.00%
Negative 0 27 27
Total 38 27 65