TABLE 4.
Comparisons of the commonly used RT-LAMP-based methods for detection of HIV-1.
| Results detection technique | Advantages | Disadvantages | References |
| Agarose gel electrophoresis | High specificity and sensitivity | Require special equipment for agarose gel electrophoresis; time-consuming; need open the reaction tube for results identification |
Curtis et al., 2012 Current study |
| Turbidity | High specificity; without opening the reaction tubes for results identification | Require special equipment for turbidity measurement |
Hosaka et al., 2009 Current study |
| Fluorescence probe | High specificity and sensitivity; without opening the reaction tubes for results identification | Require expensive equipment for fluorescence signal detection | Ocwieja et al., 2015; Zhang et al., 2022 |
| Visual reagent | High specificity; without opening the reaction tubes for results identification | The results are ambiguous when the LAMP product concentrations are low |
Khan et al., 2023 Current study |
| AuNPs-LFA | High sensitivity and specificity; accurate and visual interpretation; easy to operate and cost-saving | Need open the reaction tube for results identification | Current study |