. 2023 May 20;4(7):417–426. doi: 10.1016/j.hroo.2023.05.002

Table 2.

Presentation and management of electrical lead abnormality by lead manufacturer

Variable All (n = 135) Abbott (n = 124) Medtronic (n = 10) Other (n = 1)
Presentation
 Electrical noise only 114 (84.4) 108 6 0
 Increased threshold only 2 (1.5) 1 1 0
 Electrical noise and impedance change 4 (3.0) 1 2 1
 Electrical noise and increased threshold 6 (4.5) 6 0
 Electrical noise and decreased sensing 7 (5.3) 7 0
 Impedance change and increased threshold 1 (0.7) 1 0
 Electrical noise, impedance change and increased threshold 1 (0.7) 0 1
Management
 Observation 77 (57) 75 (60) 2 (20) 0
 Reprogram 20 (15) 18 (14) 2 (20) 0
 Revision 38 (28) 31 (25) 6 (60) 1 (100)