Table 5:
Problem | Possible Cause | Solution |
---|---|---|
High luminescent reading for all wells. | Bacterial contamination of CellTiter Glo. | Use a fresh aliquot stored at −20°C. |
Luminescence signal of MNV infected and cells-only controls are similar. | MNV infection of cells has not progressed enough. | Extend the timecourse of infection. |