Table 5.

System suitability and validation parameters for the proposed HPLC–UV method

Parameters HPLC USP Reference values
Linearity (µg/mL) 1–200
Correlation coefficient (r) 0.9999
Slope 2.291
Intercept 0.7938
SD of residuals from line 0.2201
LOD (µg/mL) 0.317
LOQ (µg/mL) 0.961
Accuracy (recovery % ± SD) 101.34 ± 0.88
Precision (intraday) 99.65 ± 1.65
Precision (inter-day) 101.45 ± 1.98
Extraction recovery % 7.554
Matrix effect 8.423
tR, min 9.18 ± 0.05
Tailing factor (T) 1.04 T ≤ 2, T = 1 for symmetric peak
Capacity factor (K′) 2.11 K' = 1–10 acceptable
Plates number (N) 5884 N > 2000
Height equivalent to theoretical plate (HETP) 0.06 The smaller the value, the higher the column efficiency
Experimental resolution (Rs)** 2.08 Rs ≥ 2

**Experimental resolution between VA and internal standard