Table 3.

Diagnostic Performance of the Sleepimage Ring with polysomnography (N = 39)

Variable Values (95% CI)
AHI ≥ 5 (N = 37) AHI ≥ 15 (N = 23) AHI ≥ 30 (N = 17)
Sensitivity 1.00 (1.00, 1.00) 0.87 (0.73, 1.00) 0.88 (0.73, 1.00)
Specificity 1.00 (1.00, 1.00) 0.69 (0.46, 0.91) 0.91 (0.79, 1.00)
False-positive rate 0.00 (0.00, 0.00) 0.31 (0.09, 0.54) 0.09 (0.00, 0.21)
False-negative rate 0.00 (0.00, 0.00) 0.13 (0.00, 0.27) 0.12 (0.00, 0.27)
Positive likelihood ratio NA 2.78 (1.32, 5.85) 9.71 (2.56, 36.80)
Negative likelihood ratio NA 0.19 (0.06, 0.57) 0.13 (0.03, 0.48)
Positive predictive value 1.00 (1.00, 1.00) 0.80 (0.64, 0.96) 0.88 (0.73, 1.00)
Negative predictive value 1.00 (1.00, 1.00) 0.79 (0.57, 1.00) 0.91 (0.79, 1.00)
Accuracy 1.00 (1.00, 1.00) 0.79 (0.67, 0.92) 0.90 (0.80, 0.99)
Youden index 1.00 (1.00, 1.00) 0.56 (0.29, 0.82) 0.79 (0.60, 0.99)
Optimal cut-off value 4 16 17

NPV, Negative predictive value; PPV, Positive predictive value; LR+, Positive likelihood ratio; LR-, Negative likelihood ratio