Skip to main content
Arquivos Brasileiros de Cardiologia logoLink to Arquivos Brasileiros de Cardiologia
. 2023 Jan 16;120(1):e20220892. [Article in Portuguese] doi: 10.36660/abc.20220892
View full-text in English

Diretriz Brasileira de Dispositivos Cardíacos Eletrônicos Implantáveis – 2023

Ricardo Alkmim Teixeira 1, Alexsandro Alves Fagundes 2, José Mário Baggio Junior 3, Júlio César de Oliveira 4, Paulo de Tarso Jorge Medeiros 5, Bruno Pereira Valdigem 5, Luiz Antônio Castilho Teno 6, Rodrigo Tavares Silva 7,8, Celso Salgado de Melo 9, Jorge Elias Neto 10, Antonio Vitor Moraes Júnior 11,12, Anisio Alexandre Andrade Pedrosa 13, Fernando Mello Porto 14, Hélio Lima de Brito Júnior 15, Thiago Gonçalves Schroder e Souza 16, José Carlos Pachón Mateos 5, Luis Gustavo Belo de Moraes 17, Alexander Romeno Janner Dal Forno 18, Andre Luiz Buchele D’Avila 18, Diogo Alberto de Magalhães Cavaco 19, Ricardo Ryoshim Kuniyoshi 20,21, Mauricio Pimentel 22, Luiz Eduardo Montenegro Camanho 23, Eduardo Benchimol Saad 23,24, Leandro Ioschpe Zimerman 25, Eduardo Bartholomay Oliveira 26, Mauricio Ibrahim Scanavacca 13, Martino Martinelli Filho 13, Carlos Eduardo Batista de Lima 27,28, Giselle de Lima Peixoto 29, Francisco Carlos da Costa Darrieux 13, Jussara de Oliveira Pinheiro Duarte 30, Silas dos Santos Galvão Filho 31, Eduardo Rodrigues Bento Costa 32, Enrique Indalécio Pachón Mateo 33, Sissy Lara De Melo 13, Thiago da Rocha Rodrigues 34, Eduardo Arrais Rocha 35, Denise Tessariol Hachul 13, Adalberto Menezes Lorga Filho 36, Silvana Angelina D’Orio Nishioka 13, Eduardo Barreto Gadelha 37, Roberto Costa 38, Veridiana Silva de Andrade 39, Gustavo Gomes Torres 40, Nestor Rodrigues de Oliveira Neto 40, Fernando Antonio Lucchese 41, Henrique Murad 42, José Wanderley Neto 43, Paulo Roberto Slud Brofman 44, Rui M S Almeida 45, João Carlos Ferreira Leal 46
PMCID: PMC10389103  PMID: 36700596

Diretriz Brasileira de Dispositivos Cardíacos Eletrônicos Implantáveis – 2023
O relatório abaixo lista as declarações de interesse conforme relatadas à SBC pelos especialistas durante o período de desenvolvimento deste posicionamento, 2019/2022.
Especialista Tipo de relacionamento com a indústria
Adalberto Menezes Lorga Filho
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Vatis / Libbs; Lixiana / Daiichi Sankyo; Eliquis / Pfizer: Fibrilação atrial.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Daiichi Sankyo: Lixiana

Alexander Romeno Janner Dal Forno
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biotronik: Marca-passo, Abbott: Fibrilação atrial.

  • B - Financiamento de pesquisas sob sua responsabilidade direta/pessoal (direcionado ao departamento ou instituição) provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Proadi.

Alexsandro Alves Fagundes
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer: Xarelto, Medtronic, Sankyo, Libbs.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Boston.

Andre Luiz Buchele D’Avila Nada a ser declarado
Anisio Alexandre Andrade Pedrosa
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Medtronic / Biomedical / Biocath / Imagem: Proctor

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Medtronic, Biocath, Biomedical.

Antonio Vitor Moraes Júnior
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biotronik.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Xarelto / Bayer: Fibrilação atrial.

Bruno Pereira Valdigem Nada a ser declarado
Carlos Eduardo Batista de Lima
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Daiichi Sankyo / Lixiana: Fibrilação atrial

  • C - Financiamento de pesquisa (pessoal), cujas receitas tenham sido provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - RIVER AF trial, Thrombosis Research Institute / Brazilian Clinical Research.

Celso Salgado de Melo Nada a ser declarado
Denise Tessariol Hachul Nada a ser declarado
Diogo Alberto de Magalhães Cavaco
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Boston Scientific/S-ICD.

Eduardo Arrais Rocha
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - ABBOTT: aula sobre monitoramento remoto.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biotronik / Bayer: Congresso.

  • Participação societária de qualquer natureza e qualquer valor economicamente apreciável de empresas na área de saúde, de ensino ou em empresas concorrentes ou fornecedoras da SBC:

  • - Área de Saúde.

Eduardo Barreto Gadelha Nada a ser declarado
Eduardo Bartholomay Oliveira
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biotronik.

Eduardo Benchimol Saad
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biosense Webster: Ablação por cateter.

Eduardo Rodrigues Bento Costa Nada a ser declarado
Enrique Indalécio Pachón Mateo Nada a ser declarado
Fernando Antônio Lucchese Nada a ser declarado
Fernando Mello Porto Nada a ser declarado
Francisco Carlos da Costa Darrieux
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer / Boehringer Ingelheim / Pfizer / Libbs / Biolab: Anticoagulação e fibrilação atrial.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer / Pfizer: Congressos e Cursos Virtuais.

Giselle de Lima Peixoto Nada a ser declarado
Gustavo Gomes Torres
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biotronik.

Hélio Lima de Brito Júnior
  • Outros relacionamentos

  • Participação societária de qualquer natureza e qualquer valor economicamente apreciável de empresas na área de saúde, de ensino ou em empresas concorrentes ou fornecedoras da SBC:

  • - Área de Saúde.

Henrique Murad Nada a ser declarado
Jorge Elias Neto Nada a ser declarado
João Carlos Ferreira Leal
  • Declaração financeira

  • B - Financiamento de pesquisas sob sua responsabilidade direta/pessoal (direcionado ao departamento ou instituição) provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Braile Biomédica: Inovare® Alpha - Prótese Valvular Biológica / Artivion-Neomex: Evita open plus, prótese híbrida.

José Carlos Pachón Mateos Nada a ser declarado
José Mário Baggio Junior
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Abbott: Palestra em Webinar; Biotronik: Simpósio satélite em congresso; Medtronic: Simpósio satélite em congresso.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Medtronic: Treinamento sobre "leadless pacing" e estimulação fisiológica; Biotronik: Congresso; Abbott: Treinamento de produtos.

José Wanderley Neto Nada a ser declarado
Júlio César de Oliveira
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Infinity.

Jussara de Oliveira Pinheiro Duarte Nada a ser declarado
Leandro Ioschpe Zimerman
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer/Xarelto; Daiichi Sankyo/Lixiana; Libbs/Propafenona e Amiodarona; Pfizer/Eliquis.

Luis Gustavo Belo de Moraes
  • Outros relacionamentos Participação societária de qualquer natureza e qualquer valor economicamente apreciável de empresas na área de saúde, de ensino ou em empresas concorrentes ou fornecedoras da SBC:

  • - Sócio cotista da Cardioritmo Serviços Médicos Ltda e Luis Belo Serviços Médicos.

Luiz Antônio Castilho Teno Nada a ser declarado
Luiz Eduardo Montenegro Camanho Nada a ser declarado
Martino Martinelli Filho Nada a ser declarado
Mauricio Ibrahim Scanavacca
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Johnson & Johnson.

  • B - Financiamento de pesquisas sob sua responsabilidade direta/pessoal (direcionado ao departamento ou instituição) provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Milestone Pharmaceuticals: etripamil NS for the treatment of Paroxysmal.

  • C - Financiamento de pesquisa (pessoal), cujas receitas tenham sido provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Medtronic, J&J and Abbott.

Mauricio Pimentel
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer: Anticoagulação.

Nestor Rodrigues de Oliveira Neto Nada a ser declarado
Paulo Roberto Slud Brofman Nada a ser declarado
Paulo de Tarso Jorge Medeiros Nada a ser declarado
Ricardo Alkmim Teixeira
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Boehringer-Ingelheim: Pradaxa, Jardiance; Daiichi-Sankyo: Lixiana; Abbott: Dispositivos Cardíacos Eletrônicos Implantáveis; Biotronik: Dispositivos Cardíacos Eletrônicos Implantáveis; Medtronic: Dispositivos Cardíacos Eletrônicos Implantáveis; Biomedical: Extração de Cabos-Eletrodos de Dispositivos Cardíacos Eletrônicos Implantáveis.

  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Biomedical: Extração de Cabos-Eletrodos de Dispositivos Cardíacos Eletrônicos Implantáveis.

Ricardo Ryoshim Kuniyoshi Nada a ser declarado
Roberto Costa
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Boston Scientific: Palestrante e consultor.

  • Outros relacionamentos

  • Participação em comitês de compras de materiais ou fármacos em instituições de saúde ou funções assemelhadas:

  • - Parecerista dos pregões do Hospital das Clínicas da FMUSP.

Rodrigo Tavares Silva Nada a ser declarado
Rui M. S. Almeida
  • Outros relacionamentos Participação societária de qualquer natureza e qualquer valor economicamente apreciável de empresas na área de saúde, de ensino ou em empresas concorrentes ou fornecedoras da SBC:

  • - Área da Saúde e Educação.

Silas dos Santos Galvão Filho Nada a ser declarado
Silvana Angelina D’Orio Nishioka
  • Outros relacionamentos

  • Financiamento de atividades de educação médica continuada, incluindo viagens, hospedagens e inscrições para congressos e cursos, provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Abbott.

  • Atuação no último ano como auditor médico para empresa operadora de planos de saúde ou assemelhada:

  • - Somente para aulas e treinamento médico.

Sissy Lara de Melo Nada a ser declarado
Thiago da Rocha Rodrigues
  • Declaração financeira

  • A - Pagamento de qualquer espécie e desde que economicamente apreciáveis, feitos a (i) você, (ii) ao seu cônjuge/ companheiro ou a qualquer outro membro que resida com você, (iii) a qualquer pessoa jurídica em que qualquer destes seja controlador, sócio, acionista ou participante, de forma direta ou indireta, recebimento por palestras, aulas, atuação como proctor de treinamentos, remunerações, honorários pagos por participações em conselhos consultivos, de investigadores, ou outros comitês, etc. Provenientes da indústria farmacêutica, de órteses, próteses, equipamentos e implantes, brasileiras ou estrangeiras:

  • - Bayer / Pfizer: eliquis; Daichi-Sanchio: lixiana; Abbott: rivacrist; Bhoeringer-Ingelheim: jardiance; Libbs: Vatis.

Thiago Gonçalves Schroder e Souza Nada a ser declarado
Veridiana Silva de Andrade Nada a ser declarado

Sumário

1. Recomendações Gerais 8

1.1. Sala Cirúrgica de DCEI 8

1.1.1. Recursos Humanos 8

1.1.2. Recursos Materiais 9

1.1.2.1. Radioscopia 9

1.1.2.2. Monitorização 9

1.1.2.3. Materiais Cirúrgicos 9

1.2. Clínica de Avaliação e Programação Eletrônica de DCEI 9

1.3. Avaliação Clínica antes do Implante de DCEI 9

1.4. Procedimento Cirúrgico e Tipos de DCEI 10

2. Recomendações para Implante de Marca-passo Convencional 10

2.1. Doença do Nó Sinusal 10

2.2. Bloqueios Atrioventriculares e Bloqueios Intraventriculares 12

2.2.1. Bloqueios Atrioventriculares 12

2.2.2. Bloqueios Intraventriculares (BIV) com Condução Atrioventricular 1:1 14

2.3. Síndrome da Hipersensibilidade do Seio Carotídeo 14

2.4. Síncope Vasovagal 16

2.5. Cardiomiopatia Hipertrófica 17

2.6. Doenças Neuromusculares 18

2.7. Síndrome da Apneia Obstrutiva do Sono 18

2.8. Síndrome do QT Longo Congênito (SQTLc) 19

2.9. Coração Transplantado 20

2.10. Escolha do Tipo de Marca-passo e do Modo de Estimulação 20

2.11. Estimulação Direta do Sistema Excito-condutor Cardíaco (Feixe de His, Ramo Esquerdo) 21

2.12. Estimulação sem Cabo-eletrodo (Leadless Pacemaker) 22

3. Recomendações para Implante de Marca-passo Multissítio/Terapia de Ressincronização Cardíaca (TRC) 24

3.1. Paciente em Ritmo Sinusal 24

3.2. Paciente com Fibrilação Atrial 25

3.3. Upgrade de Marca-passo Convencional 27

3.4. Na Indicação de Maca-passo Antibradicardia (1º Implante) 27

3.5. Na Indicação de Cardio-desfibrilador Implantável (TRC-D) 28

3.6. Estimulação Direta do Sistema Excito-condutor Cardíaco 28

3.6.1. Modulação de Contratilidade Cardíaca 29

4. Recomendações para Indicação de Cardioversor-desfibrilador Implantável 30

4.1. Prevenção Primária de Morte Súbita 30

4.1.1. Miocardiopatia Isquêmica 30

4.1.2. Miocardiopatia Não Isquêmica 31

4.1.3. Cardiomiopatia Hipertrófica 32

4.1.4. Cardiomiopatia Chagásica 34

4.1.5. Cardiomiopatia Arritmogênica do Ventrículo Direito 36

4.1.6. Miocardiopatia Não Compactada 38

4.1.7. Síndrome do QT Longo e Síndrome do QT Curto Congênito 39

4.1.8. Síndrome de Brugada 39

4.1.9. Taquicardia Ventricular Polimórfica Catecolaminérgica (TVPC) 40

4.1.10. Taquicardia Ventricular Idiopática 40

4.2. Prevenção Secundária de Morte Súbita 41

4.2.1. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada 41

4.2.1.1. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada na Presença de Cardiopatia Estrutural 41

4.2.1.2. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada na Ausência de Cardiopatia Estrutural 42

4.2.2. Síncope e Taquicardia/Fibrilação Ventricular no Estudo Eletrofisiológico 42

4.3. Crianças, Adolescentes e Cardiopatia Congênita 43

4.4. Escolha do Tipo de CDI e Modo de Estimulação 46

4.4.1. Técnica de Implante 46

4.4.2. Modo de Estimulação 46

4.5. Custo-efetividade do CDI na Prevenção Primária e Secundária de Morte Súbita 46

4.5.1. Prevenção Primária 47

4.5.2. Prevenção Secundária 47

5. Recomendações para Monitor de Eventos (Loop Recorder) Implantável 48

6. Recomendações para Avaliação e Programação Eletrônica dos DCEI 49

6.1. Marca-passo Convencional 49

6.1.1. Doença do Nó Sinusal 49

6.1.2. Bloqueio Atrioventricular 50

6.1.3. Fibrilação Atrial 50

6.1.4. Síncope Neuromediada e Síndrome do Seio Carotídeo 50

6.2. Terapia de Ressincronização Cardíaca 50

6.3. Cardioversor-desfibrilador Implantável 52

6.4. Monitor de Eventos Implantável ( Loop Recorder ) 53

6.5. Monitoramento Remoto ( Via Web ) 53

7. Recomendações para Prevenção e Tratamento de Infecções e Explante de DCEI 54

7.1. Prevenção e Tratamento de Infecções 54

7.2. Remoção de Cabos-eletrodos de Dispositivos Cardíacos Eletrônicos Implantáveis 59

8. Recomendações para prevenção de interferências eletromagnéticas 64

8.1. Cirurgia com Uso de Eletrocautério 64

8.2. Ressonância Magnética 64

8.3. Radioterapia 66

9. Conclusão 66

Referências 68

1. Recomendações Gerais

Apesar da normatização e simplificação das técnicas de implante de dispositivos cardíacos eletrônicos implantáveis (DCEI), além do conhecimento médico e da experiência cirúrgica, são necessários local e materiais adequados. O conhecimento de eletrocardiografia, principalmente das arritmias cardíacas e dos princípios de eletrofisiologia cardíaca, é fundamental.

1.1. Sala Cirúrgica de DCEI

Os procedimentos cirúrgicos de estimulação cardíaca artificial são realizados por médicos cirurgiões cardiovasculares ou cardiologistas com área de atuação em estimulação cardíaca eletrônica implantável (SOBRAC/SBC, ABEC/SBCCV, AMB). Os procedimentos devem ser realizados em centro cirúrgico, laboratório de hemodinâmica ou de eletrofisiologia. A sala operatória deve ter dimensões, iluminação e ventilação adequadas, lavatório para a antissepsia e sistema de eletricidade bivolt (com aterramento que elimine interferências eletromagnéticas e proteção dos equipamentos).

1.1.1. Recursos Humanos

Os profissionais envolvidos na realização de procedimentos cirúrgicos de DCEI são:

  1. Médicos com formação em estimulação cardíaca eletrônica implantável (ECEI), responsável e auxiliar

  2. Médico anestesiologista

  3. Instrumentador cirúrgico, preferencialmente com treinamento na área de ECEI

  4. Profissional de enfermagem, preferencialmente com treinamento na área de ECEI

  5. Técnico em ECEI

  6. Técnico de radiologia

1.1.2. Recursos Materiais

1.1.2.1. Radioscopia

Requisito ainda fundamental, a radioscopia (intensificador de imagem) pode ser fixa, como nos laboratórios de hemodinâmica, ou portátil (arco cirúrgico). A qualidade da imagem e os recursos de gravação e espelhamento de imagens facilitam bastante o procedimento, especialmente para terapia de ressincronização cardíaca (TRC). O intensificador de imagem deve possibilitar a visibilização de fios-guia de pequeno calibre e movimentos em diferentes projeções (oblíquas).

1.1.2.2. Monitorização

A monitorização eletrocardiográfica deve ser contínua, com a possibilidade de armazenamento dos traçados. As derivações disponíveis devem permitir adequada avaliação intraoperatória de TRC e de estimulação fisiológica do sistema de condução cardíaco (feixe de His, ramo esquerdo, septal profunda); ademais, nesses casos, é fundamental a análise de eletrogramas intracavitários (polígrafo).

A monitorização não invasiva da pressão arterial e o oxímetro de pulso devem estar disponíveis.

1.1.2.3. Materiais Cirúrgicos
  1. Caixa de instrumental cirúrgico apropriado

  2. Eletrocautério

  3. Cardioversor-desfibrilador externo

  4. Marca-passo (MP) externo temporário

  5. Estrutura de suporte avançado de vida

  6. Materiais e medicamentos para anestesia e estabilidade cardiovascular (analgésicos, anestésicos, antiarrítmicos, fármacos vasoativos, antibióticos etc.).

  7. Gerador de pulsos, cabos-eletrodos, introdutores, bainhas para cateterização de seio coronário e sistema de condução

  8. Programador e analisador específico do DCEI em uso ou a ser implantado.

  9. Ultrassonografia para acesso venoso pode ser útil para redução de complicações relacionadas à punção venosa profunda (p. ex., punção arterial acidental, pneumotórax)

  10. Ecocardiograma transesofágico: útil durante extração percutânea de cabos-eletrodos para diagnóstico precoce de tamponamento cardíaco

1.2. Clínica de Avaliação e Programação Eletrônica de DCEI

O médico responsável pela clínica de seguimento eletrônico de DCEI deve ter área de atuação em estimulação cardíaca eletrônica implantável (SOBRAC/SBC, ABEC/SBCCV, AMB). A clínica deve ter, em sua estrutura os seguintes recursos:

  1. Equipamento de eletrocardiograma (ECG)

  2. Programadores dos diversos fabricantes de DCEI

  3. Equipamento nobreak

  4. Cardioversor-desfibrilador externo com MP transcutâneo

  5. Imã

  6. Ecocardiograma transtorácico

  7. Acesso a exames complementares como teste ergométrico, Holter -24h, e exames de imagem (raio X, ressonância magnética, cintilografia miocárdica, tomografia computadorizada). O exame de teste de inclinação ( Tilt Table Test ) deve estar disponível na própria clínica de avaliação de DCEI ou em outra estrutura referenciada.

  8. Acesso a engenheiro especializado em ECEI

1.3. Avaliação Clínica antes do Implante de DCEI

A avaliação clínica inicial do paciente, antes da cirurgia de DCEI, deve incluir:

a) Anamnese e exame físico

A anamnese deve incluir investigação de sinais e sintomas relacionados às arritmias cardíacas, como síncope, pré-síncope, tonturas, palpitações, sinais e sintomas de insuficiência cardíaca (IC). História de morte súbita familiar, principalmente precoce e em parentes de primeiro grau, tem grande relevância.

O exame físico deve incluir inspeção, palpação dos pulsos periféricos, aferição da pressão arterial, ausculta cardíaca e de carótidas, frequência cardíaca e perfusão periférica.

Anticoagulantes orais e antiagregantes plaquetários devem ser suspensos temporariamente antes do procedimento cirúrgico, se possível. 1 Outros fármacos, geralmente, não precisam ser suspensos preventivamente. Pacientes com sinais de infecção ativa não devem receber implantes de dispositivos até a resolução do quadro.

b) Exames complementares pré-operatórios

  1. ECG em repouso

  2. Radiografia de tórax (PA + perfil esquerdo)

  3. Exames laboratoriais: todos os pacientes devem realizar hemograma completo e de coagulação. Para procedimentos em que se usa contraste endovenoso, como na TRC ou obstrução venosa, é fundamental avaliação da função renal com dosagem de eletrólitos. Nos pacientes diabéticos, a glicemia de jejum deve ser avaliada. Os exames de urina tipo I e urocultura devem ser indicados aos pacientes com queixas urinárias.

  4. Outros exames como ecocardiograma, Holter -24h, estudo eletrofisiológico e ultrassom ou flebografia de membros superiores estão indicados somente quando a condição clínica justificar

Os pacientes devem permanecer em jejum absoluto por pelo menos 6 a 8 horas antes da cirurgia, dependendo da complexidade da cirurgia e do tipo de anestesia. Tricotomia, antissepsia local e antibioticoterapia profilática 2 devem seguir protocolo institucional.

1.4. Procedimento Cirúrgico e Tipos de DCEI

a) Procedimentos cirúrgicos

Antes da cirurgia, as equipes médica e de enfermagem devem seguir os protocolos de cirurgia segura: confirmar nome do paciente, data de nascimento, número de registro hospitalar e lateralidade. Também devem confirmar a indicação do procedimento e checar os exames pré-operatórios.

O procedimento cirúrgico de DCEI deve ser realizado em centro cirúrgico ou laboratório de hemodinâmica/eletrofisiologia sob visão fluoroscópica, com monitorização eletrocardiográfica contínua, oximetria de pulso e registros intermitentes ou contínuos da pressão arterial. Dispositivos como os cardioversores-desfibriladores implantáveis (CDI) subcutâneos e os monitores de eventos implantáveis dispensam o uso do intensificador de imagem durante o implante.

A anestesia pode ser local, preferencialmente associada à sedação, ou geral. A escolha do tipo de anestesia depende da complexidade do procedimento, da via de acesso e das condições clínicas do paciente.

Para escolha do acesso cirúrgico, deve-se levar em consideração o local de implante do gerador de pulsos, o acesso ao coração para implante dos cabos-eletrodos (via transvenosa ou epicárdica) e a possibilidade de implante de sistemas sem cabo-eletrodo transvenoso ( leadless pacemaker, CDI subcutâneo). Outras variáveis que podem influenciar na escolha da estratégia cirúrgica são: uso de MP temporário e cateteres venosos, cirurgias torácicas prévias, necessidade de radioterapia (RT), características anatômicas, infecções de pele, membro superior dominante.

A região da bolsa do gerador de pulso geralmente é peitoral, podendo ser abdominal em situações específicas, em posição subcutânea ou submuscular. O acesso venoso é realizado por dissecção da veia cefálica ou punção de veia axilar, subclávia, jugular ou femoral. O número de cabos-eletrodos varia de acordo com o sistema implantado (geralmente de 1 a 3), sendo preferidos atualmente os cabos-eletrodos de fixação ativa ( screw-in ) em relação aos de fixação passiva, de acordo com a experiência profissional.

Durante o procedimento cirúrgico, é fundamental aferir os limiares de estimulação, de sensibilidade e as impedâncias dos cabos-eletrodos, além do eletrograma endocavitário ou epicárdico. Em caso de CDI, deve ser incluída medida de impedância de choque; a avaliação de limiar de desfibrilação é opcional para sistemas de CDI transvenosos (geralmente desnecessário), mas recomendável em implantes à direita, abdominal e CDI subcutâneo. 3 - 5 Nos sistemas multissítios, com implante do cabo-eletrodo para estimulação do ventrículo esquerdo (VE) pelo seio coronário, são necessários acessórios específicos para essa abordagem como bainhas, cateter de eletrofisiologia e cateter para venografia para escolha da melhor veia tributária para implante do cabo-eletrodo.

O relatório de cirurgia de DCEI deve incluir identificação do paciente, descrição do ato operatório, dados técnicos do sistema e a ocorrência ou não de complicações (p. ex., pneumotórax, hemotórax, falha de captura e/ou sensibilidade, má conexão eletrodo-gerador, deslocamento do cabo-eletrodo, perfuração do ventrículo direito (VD), estimulação diafragmática, hematoma na bolsa do gerador, contaminação e arritmias). O registro brasileiro de marca-passos e cardioversores-desfibriladores (RBM) deve ser preenchido.

Avaliação no pós-operatório e período de internação

Após a cirurgia de DCEI, é necessário realizar avaliação clínica do paciente, avaliação eletrônica do sistema implantado e exames de ECG e radiografia de tórax para confirmar o funcionamento adequado do DCEI, a posição dos cabos-eletrodos e diagnosticar possíveis disfunções e complicações.

O paciente geralmente permanece internado em ambiente hospitalar por 12 a 24h. Pacientes submetidos a procedimentos sem necessidade de acesso intravascular (troca de gerador de pulsos ou implante de dispositivo subcutâneo) habitualmente permanecem em observação pós-operatória pelo período de 6 a 12 horas (hospital-dia).

b) Tipos de DCEI

O Quadro 1 resume os principais DCEI e suas características.

Quadro 1. – Principais tipos e características dos DCEI.

Dispositivos Características
Marca-passo (MP)
  • ✓ Dispositivo cardíaco eletrônico multiprogramável, capaz de monitorar e promover a estimulação elétrica do coração, restaurar o sincronismo atrioventricular e a variabilidade da frequência cardíaca, detectar e registrar arritmias e tratar arritmias atriais usando mecanismos de overpace atrial

  • ✓ Geralmente são compostos por unidade de gerador de pulsos; um circuito eletrônico e por cabos-eletrodos que interligam o gerador à interface do endocárdio; podem ser uni, bicamerais

  • ✓ Recentemente introduzidos, os MP sem cabos-eletrodos ( leadless pacemaker ) são sistemas de estimulação elétrica, no qual todas as unidades do dispositivo são contidas em um único componente de implante intracardíaco realizado por sistema de bainhas transvenosas.

  • ✓ As bradiarritmias são o foco da terapia com MP

Terapia de ressincroização cardíaca (TRC)
  • ✓ Dispositivo cardíaco com função antibradicardia como os MP, mas que permitem a estimulação cardíaca multissítio, ou seja, estimulação do ventrículo direito e esquerdo (biventricular) para promover a correção da dissincronia intraventricular

  • ✓ O tratamento adjuvante da insuficiência cardíaca avançada é o foco primordial da TRC

Cardioversor-desfibrilador implantável (CDI)
  • ✓ Dispositivo cardíaco eletrônico que possui as funções do MP convencional, a capacidade de detectar, registrar e tratar automaticamente taquiarritmias ventriculares potencialmente fatais, por meio de terapia de choque (cardioversão ou desfibrilação) ou com estimulação ventricular programada ( overdrive – terapia antitaquicardia [do inglês, ATP])

  • ✓ As taquiarritmias ventriculares potencialmente letais associadas ou não às bradiarritmias podem ser tratadas com CDI

TRC + CDI ✓ São dispositivos que agregam em um único aparelho as funções da TRC e do CDI
Monitor de eventos implantável ✓ Dispositivo cardíaco de implante subcutâneo utilizado para monitoramento cardíaco prolongado (3 a 4 anos), com objetivo de detectar arritmias cardíacas intermitentes e esporádicas em casos de síncopes inexplicadas ou acidente vascular cerebral (AVC) criptogênico, por meio do registro contínuo da atividade elétrica do coração.

2. Recomendações para Implante de Marca-passo Convencional

2.1. Doença do Nó Sinusal

As disfunções do nó sinusal, quando resultam em sintomas, são denominadas doença do nó sinusal (DNS) e costumam ser a mais comum indicação para estimulação cardíaca artificial em âmbito global, correspondendo a aproximadamente metade dos implantes de MP definitivo. 6

A DNS é caracterizada, do ponto de vista eletrocardiográfico, por uma ou mais das seguintes manifestações: bradicardia sinusal, pausa ou parada sinusal, bloqueio sinoatrial, taquiarritmias atriais (principalmente flutter ou fibrilação atrial [FA]) associadas a bradiarritmias (pausas sinusais): síndrome bradi-taqui e incompetência cronotrópica (resposta inadequada da frequência cardíaca ao exercício ou estresse). 7 - 9

Os sintomas atribuíveis à DNS ocorrem devido à baixa frequência cardíaca ou de acordo com a duração da pausa dos batimentos cardíacos; os sintomas mais comuns são: palpitação, cansaço e dispneia, tontura, pré-síncope ou síncope. A síncope é um sintoma clínico comum, podendo estar presente em aproximadamente 50% dos pacientes encaminhados para implante de MP em decorrência de DNS. 10

Embora possa ocorrer em qualquer idade, a DNS aumenta com a idade, afetando 1 em cada 600 pacientes com idade superior a 65 anos, não havendo preferência por sexo. 11 - 13

A fisiopatologia da DNS é variada e geralmente envolve remodelamento eletrofisiológico e estrutural complexo. 14 , 15 Sua etiologia pode ser dividida em causas intrínsecas e extrínsecas.

As causas intrínsecas de DNS incluem processos inflamatórios, infecciosos e imunológicos, fibrose degenerativa, disfunção de canais iônicos e remodelamento do nó sinusal. A fibrose degenerativa idiopática, relacionada com a idade, é a causa intrínseca mais comum; no entanto, pesquisas recentes têm demonstrado que uma disfunção dos canais iônicos herdadas geneticamente também pode participar na gênese da disfunção sinusal resultante do envelhecimento. 16 - 19 Além disso, a resposta barorreflexa e a variabilidade da FC encontram-se diminuídas na população idosa. 20 , 21

Outros mecanismos intrínsecos de DNS incluem: doenças infiltrativas (p. ex., hemocromatose, amiloidose), inflamatórias (p. ex., sarcoidose), IC e FA (mudanças estruturais e anatômicas ao longo da crista terminalis ) e coronariopatia crônica (acometimento da artéria do nó sinusal). Análises genômicas já identificaram locus nas proteínas que interagem com canais iônicos e canais relacionados às frequências cardíacas normais e anormais em repouso, fornecendo informações sobre os mecanismos que controlam a frequência cardíaca. 22 - 24

Dentre as causas extrínsecas, agentes farmacológicos, distúrbios metabólicos e disfunção autonômica são os principais envolvidos na ocorrência de DNS. Os fármacos mais associados a DNS são os betabloqueadores, bloqueadores dos canais de cálcio, digitálicos, antiarrítmicos e medicações simpaticolíticas. 24 O lítio também pode estar associado à disfunção sinusal, muitas vezes definitiva. 25

A disfunção autonômica, com manifestação cadioinibitória, pode mimetizar ou intensificar a DNS na síncope vasovagal e na hipersensibilidade do seio carotídeo. 26

Outra causa rara de bradicardia recorrente e síncope é a epilepsia assistólica do lobo temporal. Nessa entidade, embora o implante de MP provisório possa ser fundamental na fase aguda, muitas vezes, o manejo específico (cirúrgico ou farmacológico) da epilepsia pode levar a regularização da disfunção sinusal e/ou bloqueio atrioventricular (BAV) sem a necessidade do implante de MP definitivo. As mesmas considerações são válidas para uma outra condição rara, a neuralgia do glossofaríngeo associada à assistolia e síncope. 27 , 28

Anormalidades metabólicas, tais como acidose sistêmica grave, hipercalemia, hipocalemia ou hipocalcemia, podem causar bradicardia sinusal, incomum em situações agudas. Outros fatores extrínsecos possíveis são hipotireoidismo, hipóxia, hipotermia e toxinas. Também tem sido descrita a associação entre a síndrome de Brugada, um tipo de canalopatia, e a ocorrência de DNS. 29

Os pacientes com DNS são geralmente assintomáticos nas fases iniciais no curso da doença, com os sintomas ocorrendo após vários anos de evolução. É também importante identificar quadros de bradicardia funcional assintomática, como ocorre durante o sono, em jovens saudáveis e em atletas, que não representam risco e não devem ser tratadas de forma geral.

A documentação da correlação das alterações eletrocardiográficas com as manifestações clínicas é essencial para o diagnóstico da DNS, seja por meio de ECG convencional de 12 derivações ou outros métodos, como Holter (24/48 horas, 7 dias), monitor de eventos externo ou monitor de eventos implantável. 30 O estudo eletrofisiológico invasivo (EEF) não deve ser utilizado regularmente na prática clínica devido à ausência de dados conclusivos sobre a real indicação de MP definitivo nos pacientes que apresentem um tempo de recuperação do nó sinusal (TRNS) ou tempo de condução sinoatrial (TCSA) alterado.

Para definição de conduta e eventual indicação de implante de MP em pacientes com DNS, é fundamental a correlação entre os sintomas clínicos e a bradicardia, além da identificação de causas reversíveis.

Na ausência de causas reversíveis e na presença de sintomas, o implante de MP definitivo representa o tratamento de escolha para a DNS, embora não haja evidências de que a estimulação cardíaca artificial tenha impacto quanto à sobrevida ou ao risco de morte súbita cardíaca dos pacientes com DNS em relação à população em geral. 30 Contudo, o implante de MP melhora significativamente a qualidade de vida, pode reduzir o risco de FA e tromboembolismo sistêmico, permite o uso de fármacos antiarrítmicos que podem causar bradicardia e possibilita monitoramento contínuo do ritmo cardíaco. 31 As recomendações para implante de MP definitivo na DNS encontram-se na Tabela 1 .

Tabela 1. – Recomendações para implante de marca-passo (MP) definitivo na doença do nó sinusal (DNS).

  Classe de recomendação Nível de evidência
Doença do nó sinusal (bradicardia sinusal, pausa/parada sinusal, bloqueio sinoatrial ou síndrome bradi-taquicardia) espontânea, sem causa passível de tratamento ou induzida por fármacos necessários e insubstituíveis, com documentação de sintomas (síncope, pré-síncope, tontura ou cansaço/fadiga) relacionados à bradicardia I C
  • Doença do nó sinusal (bradicardia sinusal, pausa/parada sinusal, bloqueio sinoatrial ou síndrome bradi-taquicardia) espontânea, sem causa passível de tratamento ou induzida por fármacos necessários e insubstituíveis, com sintomas (síncope/pré-síncope/tontura ou cansaço/fadiga) não documentados após investigação preconizada (diagnóstico presuntivo)

  • Síncope inexplicada com evidência de DNS no EEF

IIa C
Síncope inexplicada, quando houver documentação de pausa sinusal >6s assintomática IIb C
Paciente assintomático, com sintoma documentado e claramente não relacionado à bradicardia; paciente sintomático com bradicardia decorrente de causas reversíveis, incluindo-se fármacos não essenciais III C

DNS: doença do nó sinusal; EEF: estudo eletrofisiológico.

Quanto ao modo de estimulação, os grandes estudos randomizados não evidenciaram melhora da sobrevida com a estimulação atrial (AAI) ou atrioventricular (DDD) em relação à ventricular (VVI); entretanto, houve benefícios, como redução de taxas de FA, incidência de síncopes e síndrome do MP. 32 - 35 Em uma revisão sistemática dos grandes estudos randomizados, houve significativa redução de acidente vascular cerebral (AVC) ( hazard ratio [HR]:0,81) e FA (HR:0,80) com a estimulação AAI e/ou DDD comparada com a estimulação VVI. 36 O impacto do modo de estimulação na prevenção de IC ou AVC e a melhora na qualidade de vida são menos evidentes.

Os resultados do estudo DANPACE, 10 que analisou a estimulação AAI versus DDD em pacientes com DNS, mostraram que a estimulação AAIR associou-se a maior incidência de FA paroxística (HR:1,24) e o dobro de reoperações quando comparada com a estimulação DDDR. As reoperações ocorreram principalmente pela necessidade de upgrade de AAIR para DDDR, em decorrência do desenvolvimento de BAV durante o acompanhamento. Outro aspecto relevante observado nesse estudo foi que o benefício da estimulação AAI pode ser atenuado em pacientes com longos intervalos PR que podem resultar em insuficiência mitral diastólica.

A estimulação preferencial atrial (AAIR ou DDDR) é o modo predileto em pacientes com DNS ( Tabela 2 ).

Tabela 2. – Recomendações para escolha do modo de estimulação na doença do nó sinusal (DNS).

  Classe de recomendação Nível de evidência
  1. 1. AAI(R); na presença de condução AV normal

  2. 2. DDD(R), na presença de BAV avançado

I A
1. AAI(R) com reversão automática para DDD(R), na presença de BAV avançado intermitente I B
1. DDD com algoritmo para minimizar a estimulação ventricular em pacientes com condução AV preservada IIa B
1. VVI(R) em idosos, na ausência de condução retrógrada VA IIb B
1. VVI(R) em pacientes com possibilidade de demandar apenas curtos períodos de estimulação ventricular ou portadores de comorbidades significativas com impacto na sobrevida ou eventos clínicos IIb C
  1. 1. VVI(R) na presença de condução retrógrada VA; VDD(R)

  2. 2. AAI(R); na presença de BAV avançado

III C

AV: atrioventricular; BAV: bloqueio atrioventricular.

A maioria dos pacientes com DNS apresenta condução atrioventricular preservada. Por outro lado, sabe-se que a estimulação do VD tem sido associada a consequências fisiológicas negativas como resultado da dissincronia ventricular: remodelamento e redução da fração de ejeção do VE (FEVE) e insuficiência mitral funcional. 37 Em decorrência disso, em pacientes sem BAV associado, utiliza-se a programação de algoritmos que buscam diminuir a estimulação ventricular desnecessária, como a histerese atrioventricular e a mudança automática do modo DDD para AAI. 38

Algoritmos para suprimir a ocorrência de FA, como estimulação atrial contínua ( overpace atrial) ou estimulação atrial desencadeada pela sensibilidade de atividade atrial intrinseca, isoladamente ou em combinação, não têm seus beneficios comprovados. Da mesma forma, sítios alternativos de estimulação – tais como estimulação do fascículo de Bachmann, estimulação biatrial ou multissítio atrial – falharam em mostrar efeitos consistentes. 39

Os dispositivos atuais são dotados de um ou mais mecanismos de sensores de resposta de frequência, geralmente com base em movimento corporal (cristais piezoelétricos ou acelerômetros) ou ventilação/volume minuto. O objetivo principal dos sensores é aumentar a frequência cardíaca de maneira fisiológica e não necessariamente alterar os resultados clínicos. Embora o dispositivo não consiga avaliar com precisão a resposta cronotrópica atrial, ele pode fornecer indicadores da progressão da doença atrial através dos histogramas de frequência e arritmia, porcentagem de estimulação atrial e atividade diária do paciente. Tais dados podem ser úteis para a programação do sensor. Não existem evidências de que a utilização de uma combinação de sensores (p. ex., acelerômetro e ventilação minuto) possa proporcionar melhora da qualidade de vida.

2.2. Bloqueios Atrioventriculares e Bloqueios Intraventriculares

2.2.1. Bloqueios Atrioventriculares

O estímulo elétrico originado no nó sinusal é propagado pelo miocárdio pelo sistema de condução especializado. O retardo ou a falha na propagação do estímulo entre os átrios e os ventrículos caracterizam os BAV. Essa alteração da propagação do estímulo pode corresponder a uma alteração patológica ou ser um fenômeno funcional decorrente da refratariedade fisiológica (propriedade intrínseca das células do sistema de condução). 40

Do ponto de vista eletrocardiográfico, os BAV são classificados em 1º grau, 2º grau (Mobitz I, Mobitz II, 2:1, avançado) e 3º grau.

O BAV de 1º grau corresponde ao retardo na condução do estímulo do átrio ao ventrículo com um intervalo PR > 200ms.

No BAV de 2º grau tipo I ( Mobitz I), o bloqueio ocorre após prolongamento gradual dos intervalos PR (fenômeno de Wenckebach ), com uma onda P bloqueada ao final. No BAV de 2º grau tipo II ( Mobitz II), a onda P é subitamente bloqueada, ou seja, sem o alargamento progressivo do intervalo PR. Quando a condução AV ocorre com um padrão 2:1, o bloqueio de segundo grau habitualmente não pode ser classificado inequivocamente como tipo I ou tipo II sem o auxílio de manobras autonômicas, fármacos ou mesmo o estudo eletrofisiológico invasivo. O BAV avançado refere-se ao bloqueio de duas ou mais ondas P consecutivas com alguns batimentos conduzidos, o que indica alguma preservação da condução AV. Na presença de FA com pausas significativas (> 5s), deve ser considerada a possibilidade de BAV de alto grau.

Por fim, o BAV de 3º grau (bloqueio atrioventricular total [BAVT]) é definido como ausência de condução AV (completa dissociação entre ondas P e QRS, com frequência atrial superior à frequência ventricular). 41

Existem inúmeras patologias, congênitas ou mais frequentemente adquiridas, que podem afetar a condução AV. As causas degenerativas são as mais comuns na prática clínica e estão associadas ao aumento da idade, hipertensão arterial sistêmica e diabetes melito. Entre as causas infecciosas, em nosso meio, destacam-se a miocardite crônica da doença de Chagas e, em menor proporção, as miocardites agudas virais, que podem ocasionar bloqueios intermitentes agudos e definitivos.

O BAV atribuível à isquemia da parede inferior ou ao infarto agudo do miocárdio pode ser reversível, assim como os bloqueios mediados pelo sistema nervoso autônomo.

Causas iatrogênicas, principalmente por ação farmacológica, também devem ser lembradas na dependência da situação clínica.

O BAV pode ser classificado anatomicamente, de acordo com o sítio do bloqueio, em nodal AV, intra-His e infra-His. O BAV nodal está associado a progressão mais lenta, escape juncional mais rápido e confiável e melhor resposta à manipulação autonômica como administração de atropina, isoproterenol e epinefrina. Em contraste, os BAV intra-His ou infra-His progridem mais rapidamente e estão associados a escape ventricular mais lento e imprevisível, QRS mais largos e que respondem mal à atividade adrenérgica ou bloqueio vagal. Os bloqueios de alto grau (avançados ou 3º grau) apresentam maior risco de baixo débito e assistolias graves, o que implica necessidade de terapêutica urgente. Na presença de FA, identifica-se BAVT quando a resposta ventricular está baixa (< 50bpm) com intervalo RR regular.

O BAV de 1º grau geralmente é assintomático, mas pode resultar em fadiga ou intolerância ao esforço se o intervalo PR for longo o suficiente para permitir a perda do sincronismo AV. Essa alteração, “pseudossíndrome do MP”, pode ocorrer com intervalo PR > 300ms. 42 Da mesma forma, o BAV de 2º grau tipo I é frequentemente assintomático e observado em pacientes ativos e saudáveis com ou sem histórico de doença cardíaca, principalmente durante atividade parassimpática. No entanto, se ocorrer com frequência ou durante o exercício, pode causar sintomas de intolerância ao esforço ou tontura. 43

Em 61% dos pacientes com síncopes e bloqueio de ramo subjacente ou bloqueio bifascicular, podem estar presentes anormalidades da condução no sistema His-Purkinje significativas e clinicamente relevantes, identificados no EEF. 44

Em pacientes com BAV, a avaliação clínica pode ajudar a definir causas transitórias ou reversíveis, e o tratamento ou resolução pode tornar desnecessária a estimulação artificial permanente.

No BAVT congênito, a indicação de MP definitivo é mandatória na presença de sintomas ou quando a criança apresenta frequência cardíaca de repouso < 55 bpm ou < 70 bpm, se associada à doença cardíaca estrutural. 45 Nos casos assintomáticos, o acompanhamento regular com exames complementares é necessário para avaliar a FC média, o intervalo QT, pausas, arritmia ventricular, distúrbio de condução intraventricular (DCIV), presença ou surgimento de doença cardíaca estrutural, baixo desenvolvimento cognitivo-pondero-estatura e intolerância ao exercício físico, a fim de verificar a necessidade de MP definitivo. 46 Também recomenda-se implante de MP profilático, ou CDI, em alguns pacientes assintomáticos com disfunções neuromusculares e outras doenças genéticas.

Em pacientes com bradicardia indicativa de implante de MP e disfunção ventricular esquerda, o implante de CDI deve ser considerado (ver item 4).

O implante de MP não é indicado em pacientes assintomáticos com FA permanente com frequência cardíaca baixa em repouso, que apresentam resposta cronotrópica apropriada na vigília, independentemente da ocorrência e duração das pausas. Já nos casos com pausas significativas (> 3s) sintomáticas ou atribuíveis a bloqueio infranodal, indica-se MP. 30

O implante de MP para suporte terapêutico, especialmente em pacientes com IC ou doença coronária, pode ser necessário, especialmente frente à necessidade de uso crônico de betabloqueadores. 47

Para se determinar o melhor tipo de dispositivo e modo de estimulação artificial para pacientes com BAV, duas variáveis clínicas devem ser consideradas: porcentagem de estimulação ventricular esperada e função sistólica do VE (FEVE).

Estudos que compararam a estimulação com preservação da ativação atrioventricular sequencial com a estimulação unicameral ventricular em pacientes com BAV (PASE, MOST, CTOPP) não demonstraram redução significativa da mortalidade ou da taxa de AVC. 34 , 35 , 48

Por outro lado, em uma revisão sistemática, os autores identificaram que a estimulação bicameral seria a mais recomendada por diminuir a incidência de FA e reduzir a prevalência de síndrome do MP quando comparado à estimulação unicameral ventricular (VVI). 49 Já no estudo UKPace (pacientes ≥ 75 anos), não ficou demonstrado benefício da estimulação AV em termos de mortalidade, incidência de FA ou IC quando comparada à estimulação ventricular. Também observou taxas similares de AVC e maior taxa de complicações relacionadas ao procedimento cirúrgico nos pacientes submetidos ao implante de MP bicameral (7,8% x 3,5%; p < 0,001). 50 Assim, é razoável indicar implante de MP unicameral (VVI) em pacientes com fragilidade ou comorbidades significativas, idade avançada, estilo de vida muito sedentário ou pouca necessidade diária de estimulação nessa população de pacientes.

Em pacientes que apresentam condução retrógrada VA, a estimulação ventricular pode ocasionar sintomas da “síndrome do MP”. Nesses casos, deve-se preferir estimulação bicameral para evitar a dissincronia atrioventricular. 51

Os efeitos deletérios da estimulação crônica do VD foram demonstrados em vários estudos, embora apenas uma minoria (5% a 9%) de indivíduos com estimulação crônica do VD desenvolva disfunção ventricular grave com sintomas de IC. 52 - 54 Nesse sentido, pacientes com disfunção de VE e indicação de MP devido a BAV foram avaliados nos estudos COMBAT 55 (FEVE < 35%) e BLOCK-HF 56 (FEVE ≤ 50%). Nesses estudos, que compararam a TRC versus a estimulação convencional do VD, demostrou-se melhora clínica (NYHA) e remodelamento reverso do VE (com aumento da FEVE) com a TRC, com redução significativa de desfechos primários.

Em pacientes com FA e disfunção de VE, submetidos à ablação do nó AV para controle de FC, a estimulação com TRC ou do sistema excito-condutor cardíaco (feixe de His ou ramo esquerdo) parece estar associada a melhores resultados em comparação com a estimulação convencional do VD. 30

As recomendações para indicação de MP nos BAV estão sumarizadas na Tabela 3 .

Tabela 3. – Recomendações para implante de marca-passo (MP) definitivo no bloqueio atrioventricular (BAV).
  Classe de recomendação Nível de evidência
  • BAV adquirido, de 2° grau Mobitz II, grau avançado ou de 3° grau não atribuíveis à causa reversível ou fisiológica, independentemente da ocorrência de sintomas

  • BAV de 2º grau Mobitz II, avançado, de 3º grau ou bloqueio de ramo alternante, mesmo assintomático, e persistente após pelo menos 72 horas de IAM

  • BAV avançado ou bloqueio de ramo alternante após TAVI, persistente por pelo menos 24 a 48 horas

  • BAV avançado após IAM, persistente por pelo menos 5 dias

  • BAV avançado sintomático, persistente após pelo menos 5 dias de cirurgia cardíaca valvar, revascularização ou cirurgia de fibrilação atrial

  • BAV de 3º grau congênito sintomático

  • BAV de 3º grau congênito assintomático associado a fatores de risco (pausa superior a 3 vezes o ciclo RR basal, QRS largo, QTc prolongado, arritmia ventricular complexa, FC média <50bpm, disfunção ventricular)

I B
  • BAV adquirido de 2º grau Mobitz I sintomático não atribuível à causa reversível ou fármaco não essencial

  • BAV de 2º grau Mobitz II, avançado ou de 3º grau, assintomático e persistente após pelo menos 5 dias de cirurgia cardíaca valvar, revascularização ou cirurgia de fibrilação atrial

  • FA permanente com baixa resposta ventricular com sintomas atribuídos à bradicardia

I C
BAV de 3º grau congênito em adultos (>18 anos) assintomáticos IIa B
Em pacientes sintomáticos, claramente em decorrência de BAV de 1º grau significativo (pseudossíndrome do MP) IIa C

BAV: bloqueio atrioventricular; FA: fibrilação atrial; IAM: infarto agudo do miocárdio; MP: marca-passo; QTc: intervalo QT corrigido; TAVI: implante percutâneo transcateter de válvula aórtica.

2.2.2. Bloqueios Intraventriculares (BIV) com Condução Atrioventricular 1:1

As anormalidades do complexo QRS, representadas pelos bloqueios fasciculares ou bloqueios de ramo, são causadas por atraso da condução ou bloqueio de um ou mais ramos do sistema His-Purkinje.

O atraso de condução ou bloqueio do ramo direito associado a boqueio de um dos fascículos do ramo esquerdo é denominado bloqueio bifascicular (a mesma terminologia cabe no caso do bloqueio de ramo esquerdo [BRE]). Condições clínicas que podem ocasionar BIV incluem: genético-hereditárias, inflamatórias, infecciosas, infiltrativas, metabólicas, isquêmicas e degenerativas.

A presença isolada de BIV são raramente associados a sintomas, mas pode ser marcador de doença cardíaca estrutural; a presença ou surgimento de BRE pode causar dissincronia cardíaca e disfunção progressiva do VE. Alguns estudos demonstram correlação do BRE com doença coronariana e IC. 57 , 58 A progressão do BRE ou bloqueio bifascicular para BAV avançado é baixa, cerca de 1%/ano. 59 A presença de BRE costuma estar associada a maior mortalidade que os demais distúrbios da condução intraventricular. 60 O implante de MP é recomendado em pacientes com BIV em algumas doenças neuromusculares devido à alta incidência de BAVT e morte súbita cardíaca.

O EEF pode identificar distúrbios da condução de alto risco; no entanto, é um procedimento com sensibilidade variável e não é isento de riscos. Em pacientes com síncope, a presença de BIV é preditor de anormalidades ao EEF. 61

O bloqueio de ramo alternante, independentemente de sintomas, quando a morfologia do complexo QRS alterna espontaneamente entre BRE e BRD, também é indicativo de implante de MP, pois denota doença do sistema de condução em nível infranodal, com alto risco de BAVT grave.

As recomendações para indicação de MP nos BIV estão sumarizadas na Tabela 4 .

Tabela 4. – Recomendações para implante de marca-passo (MP) definitivo no bloqueio intraventricular (BIV).
  Classe de recomendação Nível de evidência
Em pacientes com síncope e bloqueio de ramo com registro de intervalo HV ≥ 70ms ou bloqueio infranodal no EEF, sem registro de TV hemodinamicamente instável, está recomendado o implante de MP definitivo Em pacientes com bloqueio de ramo alternante, independentemente de sintomas, está recomendado o implante de MP definitivo BRE novo persistente com QRS > 150ms por mais de 48 horas após TAVI, na presença de PRi > 240ms I C
Em pacientes com bloqueio bifascicular e síncope inexplicada, sem a realização de estudo eletrofisiológico (idosos, pacientes frágeis e síncope recorrente) IIb B
BRE novo persistente com QRS > 150ms por mais de 48 horas após TAVI, na presença de PRi normal IIb C
Pacientes assintomáticos com distúrbio da condução intraventricular isolado e condução AV 1:1 na ausência de outras indicações para implante de MP III B

BIV: bloqueio intraventricular; BRE: bloqueio de ramo esquerdo; EEF: estudo eletrofisiológico; MP: marca-passo; PRi: intervalo PR; TAVI: implante percutâneo transcateter de válvula aórtica.

2.3. Síndrome da Hipersensibilidade do Seio Carotídeo

A síndrome da hipersensibilidade do seio carotídeo (SHSC) é caracterizada pela história de síncope associada à resposta reflexa exacerbada decorrente da estimulação mecânica do seio carotídeo, espontânea ou por massagem (MSC). 62 A síncope é consequente à bradicardia e/ou hipotensão arterial significativas, deflagradas por movimentos da cabeça ou situações que ocasionam compressão involuntária do pescoço e seio carotídeo, embora esta correlação não seja clinicamente evidente em muitos pacientes. 63

O diagnóstico de SHSC é feito quando, na ausência de fármacos depressores do sistema excito-condutor, ocorre pausa > 3s (parada sinusal ou BAV) e/ou queda da pressão arterial sistólica (PAS) ≥ 50mmHg, com reprodução da síncope, durante manobra de compressão sequencial dos seios carotídeos direito e esquerdo, por 5 a 10s, realizada na posição supina e inclinada (teste de inclinação), em pacientes com mais de 40 anos de idade. 64 A reprodução da síncope durante a MSC aumenta a especificidade do teste diagnóstico, e a compressão carotídea na posição inclinada aumenta sua sensibilidade. 63

As respostas reflexas na SHSC podem ser classificadas, quanto ao perfil hemodinâmico, em três tipos: cardioinibidora (pausa ventricular > 3s), mista (pausa ventricular > 3s associada à queda da PA sistólica ≥ 50mmHg) ou vasodepressora (queda isolada da PAS ≥ 50mmHg). A incidência de SHSC aumenta com a idade (< 50 anos: ~0%, 50 a 59 anos: 2,4%, 60 a 69 anos: 9,1%, 70 a 79 anos: 20% e > 80 anos: 40%). 63

As evidências que suportam o implante de MP definitivo na SHSC são baseadas em pequenos estudos controlados e estudos observacionais retrospectivos. 65 - 68

Em uma revisão de 12 estudos, com 601 pacientes tratados com MP e 305 controles, apesar da heterogeneidade quanto à seleção dos pacientes, posição durante a MSC (horizontal ou inclinada), tempo de seguimento e modo de estimulação, observaram-se taxas menores de recorrência de síncope nos pacientes tratados (0% a 20%) quando comparados aos controles (20% a 60%). 68 A metanálise de três estudos controlados, com seguimento médio de 3,3 anos, demonstrou redução significativa (76%) na taxa de recorrência de síncope nos pacientes tratados com MP versus o grupo controle (9% e 38%, respectivamente; RR 0,24; IC 95% 0,12-0,48). 68

O estudo SAFE PACE, 69 que avaliou 175 pacientes idosos com quedas recorrentes inexplicadas, aparentemente sem perda de consciência e resposta cardioinibitória durante compressão do seio carotídeo, sugere que o diagnóstico de SHSC deve ser considerado nesses casos. No grupo randomizado para implante de MP definitivo, observou-se redução significativa na taxa de eventos (síncopes 53%, quedas 70% e traumatismos físicos 70%) durante o seguimento.

Em um estudo no qual o diagnóstico de SHSC foi complementado com o registro de pausas espontâneas por meio de monitores de eventos implantados, ocorreu redução de 98% na carga de síncope após o implante de MP (1,68 episódio por paciente/ano, IC 95% 1,66-1,70; vs. 0,04 episódio por paciente/ano, IC 95% 0,038-0,042). 70 Na investigação de síncopes inexplicadas em pacientes com mais de 40 anos, os monitores de eventos implantáveis podem ser úteis no diagnóstico da SHSC, assim como na síncope vasovagal, através do registro de pausas espontâneas, quando a investigação inicial através de MSC e teste de inclinação é negativa. 71 As recomendações para implante de MP na SHSC estão listadas na Tabela 5 .

Tabela 5. – Recomendações para implante de marca-passo (MP) definitivo na síndrome da hipersensibilidade do seio carotídeo (SHSC) e síncope vasovagal.

  Classe de recomendação Nível de evidência
Síncope recorrente, > 40 anos de idade e documentação de pausa sintomática espontânea >3s (pausa sinusal e/ou BAV) ou pausa >6s assintomática I A
Síncope recorrente, > 40 anos de idade e manobra de massagem do seio carotídeo com resposta cardioinibidora (pausa >3s, pausa sinusal e/ou BAV) ou mista (pausa > 3s + hipotensão) na ausência de fármaco depressor do sistema excito-condutor IIa B
Síncope recorrente, > 40 anos de idade e pausa sintomática > 3s (pausa sinusal e/ou BAV) induzida no teste de inclinação Queda recorrente, inexplicada, sem pródromos, > 40 anos de idade e manobra de massagem do seio carotídeo com resposta cardioinibidora (pausa > 3s, pausa sinusal e/ou BAV) IIb B
Paciente com síncope e ausência de resposta cardioinibidora documentada III B
Paciente assintomático(a) e massagem do seio carotídeo com resposta cardioinibidora III C

BAV: bloqueio atrioventricular; SHSC: síndrome da hipersensibilidade do seio carotídeo.

É importante salientar que pacientes idosos assintomáticos podem apresentar hipersensibilidade de seio carotídeo, porém sem caracterização da SHSC e sem indicação para implante de MP. Por outro lado, esses pacientes que apresentam quedas recorrentes inexplicadas e resposta cardioinibidora durante a compressão do seio carotídeo (hipersensibilidade de seio carotídeo), podem se beneficiar do MP definitivo. 69

Com relação ao modo de estimulação na SHSC, em um estudo que avaliou as alterações hemodinâmicas agudas em pacientes com SHSC submetidos à MSC, evidenciou-se que o modo VVI (câmara-única) esteve associado a maior queda da PAS e maior taxa de persistência dos sintomas do que a estimulação em modo DVI (dupla-câmara). 72 Estudos subsequentes comparando o modo de estimulação unicameral com o bicameral no longo prazo evidenciaram tendência de menores taxas de recorrência de síncope e pré-síncope em pacientes com estimulação dupla-câmara 73 - 76 ( Tabela 6 ).

Tabela 6. – Recomendações para modo de estimulação na síndrome da hipersensibilidade do seio carotídeo (SHSC) e síncope vasovagal.

  Classe de recomendação Nível de evidência
MP bicameral (DDD,O/R) na SHSC e na síncope vasovagal IIa B

MP: marca-passo; SHSC: síndrome da hipersensibilidade do seio carotídeo.

2.4. Síncope Vasovagal

A síncope vasovagal é caracterizada pela história de perda da consciência associada a reflexo neuromediado exacerbado que cursa com redução súbita do fluxo sanguíneo cerebral secundária à vasodilatação e/ou redução da frequência cardíaca. Na maioria dos casos, a síncope é secundária à queda súbita e significativa da PA, acompanhada de graus variáveis de bradicardia e geralmente precedida de manifestações prodrômicas, tais como: mal-estar, sudorese, sensação de calor, palidez e tontura, e seguida de fadiga. A síncope vasovagal, frequentemente, é desencadeada por um gatilho, como um estresse emocional importante, medo ou dor, e representa a principal causa de síncope, principalmente em indivíduos jovens. São fatores predisponentes: ortostatismo prolongado, ambientes fechados ou quentes, punção venosa, traumatismo físico e outros.

De acordo com alterações observadas na PA e na FC, classifica-se a resposta vasovagal em três tipos: tipo 1 ou resposta mista (queda significativa da PA acompanhada de diminuição discreta da FC); tipo 2 ou resposta cardioinibidora (diminuição importante da FC < 40 bpm ou assistolia > 3s); e tipo 3 ou resposta vasodepressora (queda significativa da PA sem diminuição significativa da FC). 77

Apesar de, eventualmente, estar relacionada a traumatismos físicos e inaptidão para realizar atividades de risco pessoal ou coletivo, a síncope vasovagal apresenta prognóstico benigno a longo prazo, e o tratamento, na maioria das vezes, é não farmacológico, por meio de orientações e mudanças de hábitos de vida. Entretanto, cerca de 14% dos pacientes apresentam formas severas de síncope vasovagal e necessitam de tratamento adicional (síncopes recorrentes muito frequentes, com comprometimento da qualidade de vida, com pródromos curtos ou ausentes e maior risco de traumatismos, ou associadas a atividades de alto risco (p. ex., motorista profissional, operador de máquinas, piloto de avião, esportista competitivo etc.). A idade do paciente é o fator mais importante na escolha da terapia mais apropriada.

Existem poucas opções terapêuticas baseadas em evidências na síncope vasovagal. A estimulação cardíaca artificial pode ser efetiva em pacientes com síncope vasovagal e reflexo cardioinibidor dominante; portanto, o foco da investigação clínica deve ser a documentação da correlação entre a síncope e a bradicardia.

A efetividade do MP definitivo foi avaliada em alguns estudos randomizados. O estudo VPS-I, randomizado aberto, avaliou pacientes com síncopes recorrentes (seis ou mais) e teste de inclinação positivo (bradicardia < 60bpm ou pausa > 1s) randomizados para tratamento com MP (DDD rate-drop response ) ou tratamento clínico. 78 A recorrência de síncope após 12 meses de seguimento foi 22% (6/27) no grupo MP e 70% (19/27) no grupo de tratamento clínico (RR 85,4%; 95% IC 59,7%-94,7%; 2p = 0,00002). O estudo SYDIT, randomizado aberto, incluiu pacientes com síncopes recorrentes (três ou mais) e teste de inclinação positivo (bradicardia) para tratamento com MP (DDD rate-drop response ) ou tratamento clínico (atenolol 100mg/dia). 79 A recorrência de síncope no grupo MP foi 4,3% (2/46), e no grupo atenolol foi 25,5% (12/47), após seguimento médio de 135 dias (OR 0,133; 95% IC 0,028-0,632; p = 0,004). No estudo VASIS, randomizado aberto, os pacientes foram randomizados para MP (DDI histerese de FC) ou nenhum tratamento. 80 A recorrência de síncope, após seguimento médio de 3,7 anos, foi 5% (1/19) no grupo MP e 61% (14/23) no grupo controle (p = 0,0006). No estudo VPS II, randomizado cego, foram selecionados pacientes com síncopes recorrentes (seis ou mais), porém bradicardia significativa no teste de inclinação não foi critério de inclusão. 81 Todos os pacientes foram submetidos a implante de MP, programados em “ativado/ ON ” ou “desativado/ OFF ”. A taxa de recorrência de síncope foi 33% (16/48) no grupo MP bicameral ( rate-drop response ) ativado e 42% (22/52) no grupo MP desativado (modo ODO), sem redução significativa no risco de síncope (RR 30%; 95% IC -33% a 63%; p = 0,14). No estudo randomizado duplo-cego mais importante publicado, o ISSUE-3, pacientes com mais de 40 anos de idade e documentação da síncope espontânea por meio de monitor de eventos implantável associada à assistolia >3s, ou assistolia > 6s na ausência de síncope, foram randomizados para MP bicameral ( rate drop response ) “ativado/ ON ” ou “desativado/ OFF ”. 82 Durante seguimento médio de 2 anos, ocorreu redução significativa (57%) na taxa de recorrência de síncope (25% no grupo “ ON ” e 57% no grupo “ OFF ”, p = 0,039). Nos principais estudos, utilizou-se MP bicameral com o modo rate drop response .

Em um pequeno estudo retrospectivo, o MP bicameral com o modo closed-loop stimulation (CLS) ativado, foi comparado com o modo desativado, com demonstração de menores taxas de recorrência de síncope. 83

As recomendações para implante de MP definitivo na síncope vasovagal estão resumidas na Tabela 5 , e as recomendações para escolha do modo de estimulação, na Tabela 6 .

2.5. Cardiomiopatia Hipertrófica

A cardiomiopatia hipertrófica (CMH) é uma doença cardiovascular genética comum, caracterizada por hipertrofia ventricular esquerda, na ausência de outras alterações cardíacas ou doenças sistêmicas capazes de produzir a magnitude de hipertrofia ventricular encontrada em seus portadores. 84 Em sua forma obstrutiva, existe um gradiente de pressão no trato de saída do VE, sendo que gradientes maiores estão associados a sintomas mais graves e aumento de mortalidade. 85

Em paciente com sintomas causados por obstrução da via de saída do VE (VSVE), as opções terapêuticas incluem fármacos com efeitos inotrópicos negativos, miectomia septal cirúrgica, ablação alcoólica septal e transplante cardíaco. 86

Em pacientes com obstrução da VSVE, a estimulação da ponta do VD promove alterações no padrão de contração ventricular e cria uma dissincronia regional, que tem como resultado a ativação tardia da parte basal do septo e a redução da contratilidade do VE, que, por sua vez, acarreta na redução do movimento sistólico anterior da valva mitral e reduz o gradiente pressórico na VSVE. 87 , 88 Entretanto, a dissincronia ventricular ocasionada pela ativação com QRS largo, por si só, reduz a contratilidade do VE e pode levar à redução do gradiente no trato de saída. Assim, o benefício, neste caso, estaria relacionado a um efeito colateral do MP.

A redução nos gradientes da VSVE com estimulação ventricular foi demonstrada em três pequenos estudos randomizados e controlados e em diversos estudos observacionais; contudo, a melhora dos sintomas e a qualidade de vida apresentaram resultados variáveis. 89 - 93 Em um estudo, a análise de subgrupos sugere que pacientes com mais de 65 anos de idade apresentam maior chance de benefício. 94

Em uma revisão da base de dados da Cochrane, os autores concluíram que os dados de benefício da estimulação ventricular na CMH são baseados em medidas de gradientes, sem evidências em relação a desfechos clínicos relevantes. 95 Além disso, em geral, a magnitude de redução dos gradientes é menor quando comparada com a miectomia ou a ablação septal. Desse modo, a indicação de MP dupla-câmara unicamente para reduzir o gradiente na VSVE está restrita a condições muito específicas: pacientes com mais de 65 anos de idade, com hipertrofia moderada, com sintomas definidos devido à obstrução da VSVE e que não tenham indicação de CDI. 93

Em geral, pacientes com CMH obstrutiva, com sintomas refratários ao tratamento farmacológico, devem ser considerados para miectomia ou ablação septal como primeira escolha. Casos muito graves poderão precisar de transplante cardíaco.

Nos pacientes submetidos a implante de MP para redução do gradiente da VSVE, a programação de intervalo AV curto é crucial (intervalo AV em VAT = 100 ± 30ms), para obter pré-excitação máxima do VD sem comprometer o enchimento ventricular diastólico. 96 Adicionalmente, deve-se programar a frequência máxima de seguimento atrial maior que a frequência máxima apresentada pelo paciente durante teste de esforço. Geralmente, pacientes com CMH, que toleram muito mal frequência cardíaca elevada, utilizam betabloqueadores – o que resulta em frequência máxima menor durante o esforço; por outro lado, são pacientes suscetíveis ao desenvolvimento de FA. Assim, deve-se programar a reversão automática de modo ( automatic mode switch [mas]) para DDI(R), evitando-se estimulação ventricular com frequência alta no caso de FA. Caso o eletrodo atrial seja pouco eficiente na detecção de FA, a frequência máxima de seguimento deve ser programada em valor reduzido.

Por fim, um significante número de pacientes com CMH recebe CDI para prevenção de morte súbita. Para esses pacientes, um dispositivo bicameral programado em DDD com intervalo AV curto pode reduzir o gradiente na VSVE e prevenir ou retardar a necessidade de intervenções complementares.

As recomendações para o implante de MP definitivo em pacientes com CMH estão elencadas na Tabela 7 .

Tabela 7. – Recomendações para implante de marca-passo (MP) definitivo na cardiomiopatia hipertrófica (CMH).

  Classe de recomendação Nível de evidência
Pacientes em ritmo sinusal com BAV de 2º grau Mobitz tipo II, grau avançado ou 3º grau espontâneo ou após ablação septal ou miectomia cirúrgica (recomendada estimulação bicameral) I B
Pacientes com CMH obstrutiva, com indicação de CDI, um dispositivo bicameral deverá ser considerado IIa C
Em pacientes com obstrução do trato de saída do VE (gradiente ≥ 50mmHg), a estimulação bicameral com intervalo AV curto pode ser considerada em casos refratários ao tratamento clínico, que não tenham indicação de CDI e que tenham contraindicação ou não aceitem ablação septal, miectomia cirúrgica ou transplante cardíaco IIb C

CDI: cardioversor-desfibrilador implantável; CHM: cardiomiopatia hipertrófica; MP: marca-passo; VE: ventrículo esquerdo.

2.6. Doenças Neuromusculares

Certas doenças neuromusculares podem provocar progressiva e insidiosa doença do sistema excito-condutor cardíaco. Entre elas, estão a distrofia muscular de Duchene, a distrofia muscular fascioescápuloumeral, a distrofia ligada ao cromossomo X, a miastenia gravis , a distrofia miotônica e a ataxia de Friedreich. 97

As principais manifestações identificadas estão relacionadas a distúrbios da condução infranodal, resultando em bloqueios fasciculares e BAV de 3º grau. Tais achados são particularmente observados na síndrome de Kearns-Sayre (oftalmoplegia externa progressiva, com retinopatia pigmentar), na síndrome de Guillain-Barré, na distrofia muscular miotônica, na distrofia muscular de Becker e na distrofia muscular fascioescápuloumeral.

A distrofia muscular miotônica e a síndrome de Kearns-Sayre são, ambas, associadas com alta incidência de doença do sistema de condução, que frequentemente progride rapidamente e não pode ser prevista por registros eletrocardiográficos ou intracavitários. A doença acomete quase sempre o sistema His-Purkinje e pode culminar ataques de Stokes-Adams ou morte súbita, exceto quando antecipados pelo implante de MP.

Em um estudo com 49 pacientes com distrofia miotônica (média de 46 anos de idade, intervalo HV ≥ 70ms), BAV de alto grau foi registrado em 47% dos pacientes após o implante do MP, mesmo sem evidência de bradicardia no início do estudo. 98 Os autores concluíram que o implante de MP definitivo deve ser considerado em pacientes com distrofia miotônica com intervalo HV aumentado (≥ 70ms), mesmo quando assintomáticos ou com bradicardia ao ECG.

Nos pacientes com doenças neuromusculares, a espera pela documentação de BAVT pode resultar em significante risco de morte súbita ou síncope. Por isso, o implante de MP definitivo deve ser considerado precocemente no curso da doença neuromuscular, assim que houver qualquer anormalidade da condução, mesmo em assintomáticos ( Tabela 8 ).

Tabela 8. – Recomendações para implante de marca-passo (MP) definitivo nas doenças neuromusculares.

  Classe de recomendação Nível de evidência
Pacientes com BAV de 2º grau (tipo II ou avançado) ou 3º grau, com ou sem sintomas. I B
Pacientes com BAV de 1º (PR > 240ms) ou QRS alargado (QRS > 120ms) assintomáticos (na distrofia muscular miotônica). IIb C

BAV: bloqueio atrioventricular; MP: marca-passo.

Anormalidades eletrocardiográficas como ritmo não sinusal, QRS > 120ms, PRi > 240ms, BAV de 2º ou 3º graus e taquiarritmias atriais foram preditores independentes de morte súbita em pacientes com distrofia muscular miotônica tipo 1. 99

2.7. Síndrome da Apneia Obstrutiva do Sono

Bradiarritmias, como bradicardia sinusal, pausas sinusais, BAV de 2º grau tipo I ou de grau avançado e ritmo de escape juncional são frequentes durante o sono, principalmente em jovens saudáveis com bom condicionamento físico. Dados diretos e indiretos têm mostrado relação com hipertonia vagal. Na maioria absoluta dos casos, esses achados são fisiológicos, sem indicação de tratamento específico. Entretanto, têm sido observadas arritmias cardíacas na síndrome da apneia obstrutiva do sono (SAOS). 100

Os casos mais importantes estão relacionados com síndrome de Pickwick, obesidade, hipertensão arterial sistêmica, síndrome metabólica, obstrução anatômica e/ou funcional das vias aéreas (p. ex., constitucional, macroglossia, hipertrofia do palato mole, hipertrofia tonsilar e de adenoides), doenças pulmonares crônicas, doenças neurológicas e outras.

Quando há obstrução de vias aéreas, durante a apneia, ocorre dessaturação de oxigênio que pode resultar em hipoxemia grave com consequente surgimento de bradi e taquiarritmias atriais e ventriculares. 101

O tratamento principal é dirigido à correção da apneia e perda de peso. O uso de aparelhos de pressão positiva em vias aéreas superiores (CPAP) para suporte respiratório durante o sono pode ser de grande importância, tendo sido observado, inclusive, desaparecimento das bradiarritmias em boa parte dos casos. 102 Assim, não há indicação primária de MP para as bradiarritmias relacionadas à SAOS.

Na prática clínica, é frequente que esta condição faça parte de um cenário caracterizado por hipertonia vagal, bradiarritmia noturna e FA comumente deflagrada pela bradicardia (síndrome braditaquicardia). 102 Quando a correção da obstrução das vias aéreas não é suficiente para melhorar o quadro, a ablação por cateter de radiofrequência pode ser indicada (cardioneuroablação e/ou ablação da FA). Em casos excepcionais, o implante de MP pode facilitar o tratamento da FA (betabloqueadores ou outros antiarrítmicos).

A hipertonia vagal excessiva pode ser bem caracterizada pela análise temporal e espectral da variabilidade RR ao Holter -24h, comparando-se os períodos de vigília e sono. A variabilidade RR e o número de pausas maiores que 2,5s pré e 1 ano pós-cardioneuroablação foram comparados em um estudo que incluiu 18 pacientes com história de SAOS e braditaquicardia e/ou hipertonia vagal. Nesse estudo, os autores demonstraram importante redução da variabilidade RR (SDNN pré-CNA 131,2 ± 38ms versus 91,9 ± 37ms 11 meses pós-CNA, p = 0,0001). Ademais, o número de pausas reduziu significativamente, de 6,5 ± 9,4 pré-CNA para 1,1 ± 3 após 11 meses da CNA, p = 0,03. Nenhum paciente recebeu implante de MP. 103 Dessa maneira, habitualmente, o implante de MP está reservado aos casos em que se identifica comprometimento do sistema de condução ( Tabela 9 ).

Tabela 9. – Recomendações para implante de marca-passo (MP) definitivo na síndrome da apneia obstrutiva do sono (SAOS).

  Classe de recomendação Nível de evidência
Pacientes com bradiarritmias noturnas, sem cardiopatia significativa, assintomáticos no período de vigília, com SAOS nos quais não foi realizado o tratamento específico III C

SAOS: síndrome da apneia obstrutiva do sono.

2.8. Síndrome do QT Longo Congênito (SQTLc)

A síndrome do QT longo congênito (SQTLc) é uma canalopatia causada por uma anormalidade da repolarização cardíaca e é caracterizada pela presença de intervalo QT prolongado, arritmias ventriculares (extrassistolia ventricular, taquicardia ventricular polimórfica, torsade de pointes ) e história de síncope e/ou de morte súbita. Em geral, existe história familiar em parentes próximos e pode ser autossômica recessiva (muito rara), acompanhada de surdez (síndrome de Jervell-Lange Nielsen) ou autossômica dominante, mais frequente (síndrome de Romano Ward). Esses dois tipos perfazem 90% dos casos de SQTLc; porém, atualmente, são conhecidos pelo menos 14 tipos diferentes da forma congênita dessa síndrome. 104

Existem portadores da anomalia genética que não apresentam nenhuma manifestação clínica espontânea, podendo moastrar, diante de certas condições, como uso de certos fármacos, estresse físico ou alterações eletrolíticas.

A medida do intervalo QT (QTi) pode ser realizada em qualquer derivação, porém, mais frequentemente, são utilizadas as derivações D2 e V5. 105 A medida é feita do início do complexo QRS até o final da onda T, excluindo-se a onda U. O QTi varia em condições fisiológicas, inversamente proporcional à frequência cardíaca. Desse modo, o valor normal deve ser corrigido pela frequência, sendo a fórmula de Bazett a mais utilizada. 106 Nesta, o QTi medido deve ser dividido pela raiz quadrada do intervalo RR precedente, sendo as unidades medidas em segundos. O resultado é o QT corrigido (QTc), cujo limite normal é 450ms e 460ms, respectivamente, para os gêneros masculino e feminino. Intervalo QTc > 480ms 4 minutos após teste de esforço é altamente sugestivo desta síndrome. Aproximadamente 20% dos casos com genótipo positivo apresentam QT normal.

Nos diversos tipos de síndrome de SQTLc, seja por redução na função dos canais de potássio ou por aumento na função do canal de sódio (retardo na inativação dos canais), ocorre retardo na repolarização celular que se manifesta por aumento do QTi. Aparentemente, as anormalidades eletrofisiológicas são heterogêneas e se tornam muito mais acentuadas diante de algumas condições como estimulação autonômica, estresse físico e mental, alterações eletrolíticas, ação de fármacos, isquemia etc., resultando em instabilidade elétrica, extrassistolia, taquicardia polimórfica, torsade de pointes , fibrilação ventricular e morte súbita.

É fundamental, sempre que possível, definir o tipo de SQTLc (especificamente, os tipos 1, 2 e 3) de acordo com a manifestação clínica e eletrocardiográfica, tendo em vista o tratamento recomendado para cada tipo.

Os tipos 1 e 2 são mais frequentes, ocasionados por diminuição da função de canais de potássio; o tipo 3 ocorre por aumento da função dos canais de sódio. 107 Tipicamente, as manifestações da SQTLc tendem a surgir na infância ou adolescência. Geralmente são mais precoces no sexo masculino (adolescência) do que no feminino (idade adulta). A síncope é a manifestação mais frequente, ocorrendo geralmente entre 5 e 15 anos de idade. História familiar de morte súbita é forte preditor de mortalidade. De modo geral, quanto maior o QTc, maior o risco de morte súbita.

A SQTLc tipo 1 geralmente tem as arritmias deflagradas durante esforço físico, notadamente a natação. Já o tipo 2, mais comumente, apresenta arritmias induzidas por estresse mental ocasionado, por exemplo, por fortes ruídos, principalmente durante descanso ou sono. O tipo 3 tipicamente apresenta as arritmias em repouso ou durante o sono, sem relação clara com uma condição de estresse.

É absolutamente fundamental que esses pacientes evitem distúrbios eletrolíticos como a hipopotassemia e estejam atentos para evitar o uso de fármacos que podem desencadear arritmias fatais. Estão disponíveis na internet diversos sites que listam fármacos com potencial risco de prolongamento do QTc (http://www.crediblemeds.org), e devem ser sempre consultados antes do uso de algum medicamento.

Todos os pacientes, sintomáticos, assintomáticos e portadores “silenciosos” devem reduzir acentuadamente a atividade física. Os esportes competitivos estão contraindicados. Certos deflagradores relacionados ao tipo, tais como natação extenuante no SQTL1 e ruídos muito altos na SQTL2, devem ser evitados. 108 Certa liberdade para esportes recreacionais não competitivos pode ser considerada, com cautela, para os pacientes com SQTL3, condicionado ao fácil acesso de um desfibrilador externo automático (DEA) no ambiente. Estão absolutamente contraindicados os medicamentos e as substâncias que prolongam a repolarização, tais como os bloqueadores dos canais de potássio, que podem induzir torsade de pointes mesmo em casos assintomáticos. Da mesma forma, também devem ser evitados os simpaticomiméticos. É recomendável que esses pacientes sejam portadores de um documento com a lista dos medicamentos proibidos.

O tratamento farmacológico baseia-se no uso de betabloqueadores, sendo os mais eficazes o propranolol e o nadolol, essencialmente na SQTL tipos 1 e 2. O metoprolol parece ser menos eficaz e não deve ser preferido. 109

Estudos retrospectivos têm demonstrado benefício inquestionável de betabloqueadores ou denervação cirúrgica (retirada do gânglio estrelado esquerdo), com mortalidade de 9% no grupo tratado versus 60% no grupo sem tratamento.

O QTc pode ser reduzido experimentalmente com agentes potencializadores da bomba de potássio, como o nicorandil na SQTL1, ou espironolactona combinada com potássio oral na SQTL2. A SQTL3 pode ser beneficiada com bloqueadores de canal de sódio, como mexiletine ou flecainida, que podem encurtar o QTc, mas este último pode induzir um fenótipo de Brugada. Existem relatos de tratamento com sucesso de tempestade elétrica na SQTL3 com mexiletine, que é recomendada por alguns profissionais quando há um QTc muito longo.

Não há, atualmente, indicação primária de implante de MP para prevenção de morte súbita na SQTLc, visto que o CDI confere segurança mais ampla. 110 Entretanto, o MP pode ser indicado em casos em que se identifica BAV ou quando existem arritmias ventriculares deflagradas ou agravadas por bradicardia ou pausas, desde que não haja história de morte súbita recuperada e que estejam ausentes os sinais de alto risco: surdez congênita, síncope, arritmias ventriculares complexas documentadas, história familiar de morte súbita, sexo feminino, QTc > 0,60s. Eventualmente, o MP pode ter indicação em conjunto com a terapia betabloqueadora para evitar bradicardia resultante da própria ação farmacológica. A estimulação cardíaca em uma frequência acima da frequência sinusal espontânea pode, reflexamente, inibir a ação simpática e ser bastante útil para controle de tempestades arrítmicas. O MP deve estimular somente átrio (evitar dissincronia ventricular) e pode ser programado com frequência mais elevada (p. ex., 80ppm), que pode reduzir a duração do QTc. Ademais, a ativação do sensor de resposta de frequência pode garantir adaptação cronotrópica, prejudicada devido ao uso de betabloqueadores ( Tabela 10 ).

Tabela 10. – Recomendações para implante de marca-passo (MP) definitivo na SQTLc.

  Classe de recomendação Nível de evidência
Pacientes de baixo risco (ausência das condições de alto risco*), principalmente na SQTL3, com bradiarritmia (sinusal ou BAV) agravada ou não pelo uso de betabloqueadores IIb C

*Alto risco na SQTLc: surdez congênita, síncope, arritmias ventriculares complexas documentadas, história familiar de morte súbita, sexo feminino, QTc > 0,60s. SQTLc: síndrome do QT Longo Congênito.

A denervação cirúrgica pode ser alternativa nos casos de síncopes recorrentes, apesar do uso regular em dose plena de betabloqueadores ou quando há impedimento para a terapia farmacológica otimizada (TFO) (p. ex., asma).

2.9. Coração Transplantado

A taxa de implante de MP definitivo após o transplante cardíaco varia entre 2% e 24% e tem apresentado queda significativa com emprego da técnica de anastomose bicaval, em relação às técnicas de anastomose biatrial. 111 - 113

A maioria dos estudos reporta a DNS como a anormalidade mais comum encontrada. As causas de DNS são variadas e incluem: traumatismo cirúrgico, danos à artéria do nó sinusal por traumatismo e isquemia, tempos prolongados de isquemia cardíaca, denervação cardíaca e características basais do coração doado. 114

Cerca de 10% dos pacientes que necessitam de MP apresentam distúrbios da condução AV, principalmente BAV de 2º e 3º graus, sendo postulado que essas alterações provavelmente estejam relacionadas com preservação inadequada do enxerto. 115

A bradicardia é comum no período precoce após transplante cardíaco, ocorrendo em cerca de dois terços dos pacientes; porém, frequentemente, tende a resolver espontaneamente. Se a bradicardia perdurar por algumas semanas e cursar com sintomas, o implante de MP pode ser necessário.

Estudos demonstram que vários pacientes com DNS e bradicardia que foram submetidos a implante de MP não são dependentes do dispositivo no final de 3 meses. Entretanto, os pacientes com BAV precoce, frequentemente, requerem estimulação cardíaca a longo prazo 116 , 117 ( Tabela 11 ).

Tabela 11. – Recomendações para implante de marca-passo (MP) definitivo após transplante cardíaco.

  Classe de recomendação Nível de evidência
Pacientes sintomáticos com bradiarritmias ou incompetência cronotrópica em que a resolução espontânea não é esperada ou que, mesmo que transitória, pode persistir por meses, o implante de marca-passo deve ser considerado IIa C

2.10. Escolha do Tipo de Marca-passo e do Modo de Estimulação

Estudos clínicos randomizados não evidenciaram impacto na sobrevida com a estimulação atrial ou atrioventricular (AAI/DDD) em relação à estimulação ventricular exclusiva (VVI); entretanto, demonstraram benefícios em relação à redução de ocorrência de FA, incidência de síncopes e síndrome do MP. 32 - 35 Em uma revisão sistemática, os autores demonstraram significativa redução das taxas de AVC ( hazard ratio [HR]:0,81) e FA (HR:0,80) com a estimulação atrioventricular sequencial em comparação com a estimulação ventricular. 36

Em pacientes com bradicardia persistente, a estimulação bicameral é o modo de escolha. Os resultados do estudo DANPACE mostraram que a estimulação AAIR associou-se com maior incidência de FA paroxística (HR:1,24) e o dobro de reoperações quando comparada com a estimulação DDDR. 118 A estimulação ventricular deve ser evitada em pacientes em ritmo sinusal, uma vez que pode causar FA e piora de IC 119 ( Figura 1 ). Ressalta-se que a programação de intervalo AV excessivamente longo, visando evitar estimulação ventricular, em pacientes com condução AV espontânea, pode ser prejudicial do ponto de vista hemodinâmico em pacientes com BAV de 1º grau amplo, podendo causar regurgitação mitral diastólica, pseudossíndrome do MP e FA. 120

Figura 1. – Algoritmo para preservação da condução atrioventricular intrínseca.

Figura 1

A ativação do sensor de variação de frequência pode ser benéfica em pacientes com incompetência cronotrópica. O estudo ADEPT comparou os modos DDDR versus DDD em termos de melhoria de qualidade de vida em 872 pacientes com incompetência cronotrópica. Aos 6 meses de seguimento, os pacientes randomizados para o modo DDDR tiveram maior pico de frequência cardíaca em comparação com aqueles em modo DDD (113,3ppm x 101,1ppm; p < 0,0001). 121 Contudo, com 1 ano, não ocorreram diferenças significativas entre os dois grupos com respeito à escala de atividade ou desfechos secundários de qualidade de vida.

2.11. Estimulação Direta do Sistema Excito-condutor Cardíaco (Feixe de His, Ramo Esquerdo)

O remodelamento e a consequente disfunção ventricular esquerda promovidos pela dissincronia associada à estimulação artificial do VD são desfechos que justificam a busca por sítios alternativos para estimulação em pacientes com bradiarritmias que necessitam de estimulação ventricular artificial.

O estudo MOST demonstrou que, em pacientes com disfunção sinusal, a estimulação ventricular na ponta do VD determinou aumento significativo de episódios de FA e internação por IC. 35 Mais importante ainda: esses eventos adversos estiveram diretamente relacionados ao percentual cumulativo de estimulação ventricular.

Dessa maneira, a busca por uma forma de estimulação artificial que mantenha o sincronismo intra e interventricular, além da correção da bradiarritmia, é uma necessidade de relevância clínica que vem sendo remetida à chamada estimulação fisiológica. 122 , 123

A estimulação direta do sistema de condução é a maneira mais fisiológica de estimulação ventricular artificial porque mantém a ativação elétrica natural do coração, uma vez que o estímulo segue pelas vias normais de condução especializada (His-Purkinje), evitando a dissincronia induzida pela estimulação muscular do VD. 124

Sharma et al. demonstraram, em estudo não randomizado com 202 pacientes em seguimento de 2 anos, que a estimulação pelo feixe de His foi superior à estimulação convencional pela ponta do VD. Nos pacientes com mais de 40% de percentual de estimulação ventricular, houve redução significativa da necessidade de internação por IC (15% vs. 2% p = 0,02). 125

A estimulação do feixe de His apresenta algumas limitações, que incluem dificuldade técnica de localização do sítio mais apropriado para estimulação (maior tempo cirúrgico), limiares de estimulação mais elevados, menor amplitude de sinal intracavitário e possibilidade de inibição anormal por cross-sensing . 126

A estimulação direta do ramo esquerdo ou de região próxima por via septal profunda (eletrodo entregue via bainha, perfurando até o lado esquerdo do septo interventricular) é alternativa viável para manutenção de QRS estreito e prevenção de dissincronia. Apesar da diferença técnica entre a captura direta do ramo esquerdo e a captura da região do ramo esquerdo, do ponto de vista funcional, a estimulação da região do ramo esquerdo é capaz de promover sincronismo equivalente à estimulação direta do feixe de His.

Mafi-Rad et al. demonstraram, em uma pequena serie de 10 pacientes com DNS, a viabilidade de estimular a região esquerda do septo interventricular resultando em QRS estreito, com padrão de atraso final pelo ramo direito, com limiares estáveis e melhor performance hemodinâmica (mensurada pela dP/dT) do que com a estimulação apical do VD. 127 Mais tarde, Huang demonstrou a possibilidade de captura direta do ramo esquerdo por meio da estimulação septal profunda, corrigindo o BRE em pacientes com essa abordagem, e estabeleceu critérios para definição de captura do ramo esquerdo que incluem pelo menos três critérios: 1) presença de potencial de ramo esquerdo captado no sinal do EGM do eletrodo; 2) tempo de ativação da parede livre do VE (LVAT) menor que 90ms sem modificação com energia de estimulação de 2V ou 5V; 3) padrão de BRD incompleto ao ECG; 4) evidência de captura seletiva e não seletiva do ramo esquerdo; 5) evidência de estimulação direta do ramo esquerdo através de eletrodo concomitante no His ou no septo esquerdo. 128 , 129

A estimulação fisiológica através do feixe de His ou do ramo esquerdo vem sendo utilizada com eficiência em vários cenários de bradiarritmias com necessidade de estimulação ventricular ( Tabela 12 ).

Tabela 12. – Recomendações para estimulação fisiológica (feixe de His, ramo esquerdo) para tratamento de bradiarritmias.

  Classe de recomendação Nível de evidência
DNS com indicação de MP convencional em paciente com retardo de condução intraventricular IIa C
FA permanente com indicação de ablação da junção AV para controle de FC IIa C
BAV sem disfunção sistólica de VE IIb B

AV: atrioventricular; BAV: bloqueio atrioventricular; DNS: doença do nó sinusal; FA: fibrilação atrial; FC: frequência cardíaca; VE: ventrículo esquerdo.

Ambas as estratégias de estimulação (His, ramo esquerdo) têm demonstrado segurança, estabilidade e melhores resultados de medidas de sincronia, duração do QRS e tendência de melhora da função ventricular nas séries que compararam com a estimulação pelo VD. Quando comparadas as duas técnicas, a estimulação direta do feixe de His é capaz de promover um QRS normal e, aparentemente, ainda mais fisiológico, mas à custa de tempo maior de implante, limiares mais elevados e menor valor de onda R que a estimulação do ramo esquerdo. 130 , 131

A evolução da tecnologia das ferramentas de implante, a disponibilidade de geradores dedicados que possibilitem algoritmos específicos para detecção e gasto de energia proporcional e os resultados de estudos controlados a longo prazo determinarão o papel das técnicas de estimulação fisiológica possivelmente como preferencial em futuro próximo. 132

2.12. Estimulação sem Cabo-eletrodo (Leadless Pacemaker)

A estimulação cardíaca artificial não é isenta de problemas. A incidência de complicações com cabos-eletrodos e o gerador de pulsos, especialmente em loja subcutânea, aumenta ao longo dos anos de seguimento e pode afetar mais de 10% dos portadores desses dispositivos.

O implante de sistemas convencionais associa-se a risco de pneumotórax, hemotórax, deslocamento dos cabos-eletrodos, oclusão venosa, insuficiência da valva tricúspide e infecção (que, por vezes, pode cursar com endocardite). Ademais, o implante subcutâneo também se associa à ocorrência de hematoma da loja e infecção, que podem ocorrer não só na primeira cirurgia, mas também no momento da troca do gerador. 133 , 134 De todas as complicações descritas, a endocardite infecciosa merece especial destaque pela elevada morbidade e mortalidade associadas. 135 , 136

Sendo o cabo-eletrodo de MP a fonte principal de problemas e complicações, é natural que a tecnologia tenha avançado no sentido de privilegiar soluções em que fosse possível dispensá-lo. Neste sentido, o MP sem cabo-eletrodo ( leadless pacemaker ) é uma evolução tecnológica que traz algumas vantagens potenciais em relação aos sistemas convencionais. Inicialmente, foram lançados no mercado dois sistemas – Nanostim ( Abbott Medical Inc. Abbott Park, IL, USA ) e MICRA TPS ( Medtronic, Inc., MN, USA ), mas, atualmente, apenas o MICRA TPS é comercializado.

Trata-se de um uma pequena “cápsula” (26mm de comprimento por 6,7mm de diâmetro) em que está contido o eletrodo e o gerador de pulsos, que é implantado na cavidade cardíaca por via transvenosa. Todo o processo é relativamente simples e envolve a cateterização da veia femoral, passando uma bainha (de grande calibre – 24F) pela veia cava inferior até ao átrio direito. Um sistema de entrega segue depois, dentro dessa bainha, e permite libertar o dispositivo no VD. A posição inicial recomendada para o implante era o apex do VD, mas atualmente tem sido preconizado que seja implantado no septo interventricular. 137

O sistema MICRA foi avaliado em um ensaio clínico prospetivo, multicêntrico (MICRA IDE), de braço único, que incluiu 725 pacientes (idade média 75,9 ± 10,9 anos, 58,8% do sexo masculino) com indicação de implante de MP definitivo de câmara única. O objetivo principal foi avaliar a eficácia (limiar de captura no seguimento de 6 meses) e segurança (complicações maiores). O implante foi bem-sucedido em 99,2% dos casos. Ocorreram 28 complicações maiores em 3,4% dos pacientes, tendo sido registados perfuração ou derrame pericárdico (1,6%), complicações no local de acesso vascular (0,7%) e limiar de estimulação elevado (0,3%). Não ocorreram deslocamentos ou embolizações do dispositivo. Os valores médios de onda R, limiar de estimulação e impedância permaneceram estáveis. Nesse estudo, a taxa de complicações foi comparada com uma população de mais de 2.000 pacientes (controles históricos de outros ensaios clínicos de MP convencionais da mesma marca), tendo-se verificado menor taxa de complicações maiores (HR 0,49 IC95% 0,33-0,75, p = 0,001), incluindo menor número de hospitalizações (2,3% vs. 3,9%) e de necessidade de revisão de sistema (0,4% vs. 3,5%). Destaca-se, ainda, baixa taxa de infecções, não relacionados com o implante ou com a presença do dispositivo. 138 Esses resultados foram depois confirmados no seguimento anual dessa população. 139 A análise dessa mesma coorte mostrou melhoria de parâmetros de qualidade de vida aos 3 e 12 meses e elevados níveis de satisfação. No mesmo estudo, o sistema MICRA foi associado a menos restrições na atividade que os sistemas convencionais. 140

O sistema MICRA foi também já avaliado em registos do “mundo real”. O maior desses registos é multicêntrico (96 centros em 20 países) e pretende incluir 1.830 doentes. O desfecho primário do estudo é a ocorrência de complicações nos primeiros 30 dias pós procedimento. Os resultados dos primeiros 795 doentes foram já publicados e demonstraram elevada taxa de sucesso no implante (99,6%), com baixa taxa de complicações maiores (1,5%). Cerca de 20% dos pacientes desse registo tinham contraindicação para implante de sistema convencional (sobretudo por problemas relacionados com acesso vascular). Na população estudada, ocorreram 5 derrames pericárdicos (dois com necessidade de drenagem). 141 Pequenos estudos unicêntricos de seguimento de centro único que avaliam o leadless pacemaker no mundo real também foram publicados e confirmam a elevada taxa de sucesso no implante, assim como baixa taxa de complicações. 142 - 144

A segurança do implante do sistema também foi avaliada em populações especiais, como pacientes em hemodiálise (HD) e após extração de dispositivos convencionais por infecção.

O dispositivo Nanostim foi também avaliado em estudo multicêntrico e observacional, 145 , 146 cujos resultados obtidos até 6 meses de seguimento não diferem dos resultados obtidos com os estudos do MICRA. O motivo para ter sido retirado do mercado tem a ver com falência inesperada da bateria, que impede estimulação ventricular e comunicação com o dispositivo em cerca de 0,5% dos casos. 147

Além das complicações intraoperatórias que já foram referidas, existem duas situações que constituem, ainda, uma área de incerteza. Em primeiro lugar, ainda não está claro se o dispositivo tem interferência ou não na função da válvula tricúspide. Em um estudo inicial, com 2 meses de seguimento de 23 pacientes, concluiu-se que não existia interferência na função da válvula tricúspide. 148 Outro estudo mais recente, que incluiu 53 pacientes, concluiu que o leadless pacemaker interfere na função da válvula e pode causar ou agravar insuficiência tricúspide. 149 O mecanismo mais provável é a interferência mecânica do dispositivo com o aparelho subvalvular, sendo a lesão aguda (durante o implante) da válvula ou a dissincronia induzida pelo MP as causas menos prováveis. Os autores desse estudo descrevem, ainda, que os doentes em que o dispositivo fica em posição septal (a posição atualmente recomendada, pelo menor risco de perfuração) são aqueles em que mais vezes se observou insuficiência tricúspide, provavelmente pela maior proximidade à válvula e ao aparelho subvalvular.

Outra área de incerteza é a atitude a tomar no fim da vida do gerador. Está publicada a experiência mundial com o explante precoce do sistema MICRA – entre o primeiro e o 95ª dia – com bons resultados e baixa taxa de complicações. 150 Está também publicado um relato de um caso clínico em que foi possível extrair o dispositivo, por via percutânea, 4 anos após o implante. 151 É, no entanto, desconhecido qual vai ser o comportamento em seguimento de longo prazo e se será possível a extração (ou se adiciona-se outro dispositivo em outra localização). 152 , 153

É necessário, também, não esquecer que, com o aparecimento de novas tecnologias, podem surgir novas complicações não descritas – que poderão aparecer quando existirem seguimentos mais longos. Importante também recordar que não existe, até a data presente, qualquer comparação randomizada entre sistemas de MP convencionais e leadless pacemaker .

O dispositivo disponível atualmente é de câmara única e permite resposta em frequência (VVIR). As indicações, em termos gerais, são as mesmas de um MP de câmara única – de forma muito genérica: bradicardia sintomática em que se considera não ser necessário eletrodo atrial ( Tabela 13 ).

Tabela 13. – Recomendações para implante de marca-passo (MP) sem eletrodo.

  Classe de recomendação Nível de evidência
Leadless pacemaker é recomendado em pacientes com complicação de dispositivos convencionais (fratura de eletrodo, após extração por infecção), nos quais existem obstruções venosas críticas IIa C
Leadless pacemaker é uma opção aceitável em pacientes com fibrilação atrial com baixa resposta ventricular como primeira opção desde que discutidas com o paciente as vantagens e as limitações IIb B
Leadless pacemaker é uma opção aceitável em pacientes em ritmo sinusal, com indicação de MP com expectativa de pouca estimulação, desde que discutidas com o paciente as vantagens e as limitações (p. ex., pausas sinusais raras ou BAV paroxístico raro) Leadless pacemaker pode ser considerado em pacientes com BAVT em que se consideraria estimulação unicameral ventricular (muito idosos, pouca atividade, acamados, desde que discutidas as vantagens e as limitações) IIb C
Leadless pacemaker de câmara única ventricular em pacientes com BAVT, em ritmo sinusal, candidatos à estimulação convencional de dupla câmara quando é desejável a manutenção do sincronismo atrioventricular III C
Leadless pacemaker não está indicado em crianças e jovens (incerteza quanto à conduta quando o gerador chega ao fim de vida útil) III C

BAV: bloqueio atrioventricular; BAVT: bloqueio atrioventricular total; MP: marca-passo.

Um dos principais fatores limitantes à utilização desse tipo de dispositivo é o preço elevado, como ficou demonstrado em survey recente da EHRA. 154

Conceitualmente, os melhores candidatos seriam aqueles com contraindicações relativas aos dispositivos convencionais, por exemplo, por ausência de acessos vasculares, pacientes com risco de nova cirurgia, hemodiálise e após infecção de sistemas convencionais. Existem estudos (a maioria de centro único) que demostram eficácica do leadless pacemaker nesses subgrupos. 155 - 157 Esses pacientes estão também representados nos estudos de “vida real” já referidos.

A técnica de implante é distinta da técnica convencional; por isso, é importante ter experiência em estimulação cardíaca, em acessos vasculares femorais, em manipulação de bainhas de grande calibre e manipulação de eletrodos no VD. A perfuração cardíaca não é frequente (pode ocorrer no VD, mas também no átrio e na aurícula) e, quando ocorre, faz com que exija habitualmente resolução cirúrgica de emergência. 158 , 159

3. Recomendações para Implante de Marca-passo Multissítio/Terapia de Ressincronização Cardíaca (TRC)

3.1. Paciente em Ritmo Sinusal

A despeito da TFO, muitos pacientes com IC com FEVE reduzida (ICFEr) evoluem com persistência de sintomas e importante disfunção sistólica do VE. Os fatores mais comuns relacionados com a baixa resposta à TFO são a insuficiência mitral moderada ou grave, reserva funcional miocárdica reduzida e a dissincronia ventricular (DV). Com relação a esta última, a TRC, por meio da estimulação cardíaca atriobiventricular, associada ou não ao CDI, tem sido considerada excelente opção terapêutica para pacientes com BRE. A TRC tem o propósito de corrigir disfunções eletromecânicas em pacientes com ICFEr que apresentem a DV e, por consequência, melhorar a performance do VE.

O ECG de superfície é o método de eleição na pesquisa de DV e seleção de pacientes para a TRC. Apesar de os métodos de imagem, como o ecocardiograma, serem capazes de detectar a DV mecânica, o estudo PROSPECT (Predictors of Response to CRT) demonstrou que o doppler tecidual não conseguiu identificar os pacientes respondedores à TRC. 160 O estudo COMPANION (Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure) 161 e o CARE-HF (Cardiac Resynchronisation in Heart Failure Study) 162 foram os primeiros estudos randomizados de larga escala testando a TRC em desfechos clínicos de mortalidade total e taxa de hospitalização. Os achados desses estudos demonstraram aumento da sobrevida proporcionado pelo acréscimo da TRC à TFO.

Esses resultados foram confirmados por metanálise publicada em 2006, que incluiu oito ensaios clínicos com o total de 3.380 pacientes. 163 Em seguimento médio de 29,4 meses, foram observados 524 óbitos, com redução marcante da mortalidade (OR:0,72, IC95% 0,59 a 0,88) e da taxa de hospitalizações por IC (OR: 0,55, 95%IC 0,44 a 0,68) com a TRC. Em todos os estudos incluídos, houve melhora significativa da qualidade de vida (3 a 6 meses), apesar da heterogeneidade dos critérios de tempo de avaliação. Ademais, o número necessário para tratar (NNT) foi estimado em 11 (necessário implantar 11 dispositivos para salvar 1 vida em 2,5 anos). Considerando-se a longevidade média dos aparelhos de TRC (6 anos), seria necessário implantar 5 dispositivos para evitar 1 óbito. 164

Esses estudos embasaram as primeiras indicações da TRC como terapêutica coadjuvante à TFO na ICFEr avançada (FEVE ≤ 35% e classe funcional (CF) NYHA III ou IV a despeito da TFO por mais de 3 meses) que apresentassem DV, constatada pela presença de distúrbio da condução intraventricular (DCIV) no ECG. Ressalta-se que o estudo COMPANION também demonstrou maiores benefícios clínicos com o acréscimo do CDI à TRC (TRC-D).

Subsequentemente, foram publicados os estudos MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy), 165 REVERSE (Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction) 166 e RAFT (Resynchronization–Defibrillation for Ambulatory Heart Failure Trial). 167 Esses estudos compararam o TRC-D versus o CDI isolado, em pacientes com FEVE ≤ 40% (REVERSE) ou ≤ 30% (MADIT-CRT e RAFT) e CF NYHA I-II (REVERSE e MADIT-CRT) ou II-III (RAFT). Os resultados, ressaltados em metanálise, permitiram comprovar os benefícios da TRC, não só como terapêutica adicional à TFO, mas também ao CDI na redução da mortalidade total. 168

Em relação aos pacientes com ICFEr assintomáticos (CF NYHA I) ou em CF NYHA II incluídos em cinco estudos clínicos randomizados, metanálise demonstrou redução significativa da mortalidade total e taxa de hospitalização por IC nos pacientes com CF NYHA II. 165 - 167 , 169 , 170 Destaca-se que apenas 9% dos pacientes estudados tinham CF NYHA I e, nessa população, a TRC reduziu significativamente a taxa de hospitalização por IC, mas não a mortalidade total. 171 Esses resultados demonstram que a TRC, instituída precocemente nos pacientes com ICFEr assintomáticos, pode reduzir a progressão da IC, possivelmente por meio do remodelamento reverso ventricular. Entretanto, os potenciais benefícios da TRC em pacientes com ICFEr e CF NYHA I devem ser cuidadosamente avaliados em relação aos possíveis eventos adversos e custos associados ao implante da TRC.

A duração do QRS ≥ 120ms como ponto de corte na indicação da TRC foi baseada nos critérios de inclusão dos estudos COMPANION 161 e CARE-HF. 162 Porém, em 2013, houve as publicações dos resultados do estudo ECHO CRT que demonstraram aumento da mortalidade cardiovascular no subgrupo de pacientes com QRS < 130ms submetidos a TRC. 172 , 173 Esses achados foram corroborados em metanálise que demonstrou pouco benefício da TRC nos pacientes com QRS < 140ms. De fato, quanto maior a duração do QRS, melhor é a resposta à TRC. 174 , 175 Ademais, os pacientes com BRE e duração do QRS ≥ 150ms são os que mais se beneficiam da TRC. 165 , 176 , 177 Esse dado também foi confirmado na metanálise que reuniu 12.638 pacientes de 13 grandes estudos, corroborando o benefício da TRC em pacientes com BRE e o maior risco de morte por falência da bomba cardíaca nos pacientes com QRS mais alargados. 175

Em relação ao tipo de DCIV, subanálise de grandes estudos sugerem que pacientes com QRS largo não BRE apresentam pior resposta à TRC. Dados do estudo REVERSE, comparando o ressincronizador cardíaco ligado versus desligado, mostrou que pacientes não BRE apresentaram ausência de remodelamento reverso do VE, independentemente da duração do QRS. 178 Da mesma forma, estudo utilizando a população do MADIT-CRT também demonstrou ausência de benefícios clínicos da TRC-D em 537 pacientes com ICFEr leve e não BRE, independentemente da morfologia e da duração do QRS. 179

Importante ressaltar que poucos pacientes com bloqueio do ramo direito (BRD) foram incluídos nos grandes estudos, dificultando uma conclusão definitiva em relação aos efeitos da TRC nessa população. 180

Por outro lado, apesar de as subanálises desses estudos não observarem benefícios da TRC em pacientes não BRE, recente estudo observacional, de mundo real, utilizando informações do National Cardiovascular Data Registry (EUA), avaliou a resposta clínica da TRC-D versus CDI em 11.505 pacientes não BRE. Este estudo demonstrou que, em pacientes com DCIV inespecífico com QRS ≥ 150ms, a TRC reduziu mortalidade e taxa de hospitalização por IC, enquanto aqueles com QRS < 150ms tiveram má resposta clínica à TRC, com aumento de mortalidade e taxa de hospitalização. 181 Finalmente, esse mesmo estudo constatou aumento no risco de morte e na taxa de hospitalização por IC em pacientes com BRD tratados com a TRC-D. Essa conclusão foi consistente com resultados de Bilchick et al., 182 publicado em 2010, que utilizaram as informações do Medicare. Adicionalmente, Pastore et al. 183 demonstraram que pacientes com BRD típico, definido classicamente como QRS > 120ms de duração, padrão rsr’, rsR’ ou rSR’ na derivação V1 ou V2 e onda S maior que onda R ou > 40ms de duração nas derivações D1 e aVL, 184 apresentavam má resposta à TRC. Portanto, os dados disponíveis na literatura até o momento apontam para a indicação da TRC em pacientes com ICFEr e DCIV inespecífico com duração do QRS ≥ 150ms, e reforçam a necessidade de maior cautela na indicação desta terapêutica nos pacientes com BRD típico.

As indicações de TRC em pacientes em ritmo sinusal estão listadas na Tabela 14 .

Tabela 14. – Recomendações para indicação de terapia de ressincronização cardíaca (TRC) em pacientes com ritmo sinusal.

  Classe de recomendação Nível de evidência
TRC é recomendada para pacientes com ICFEr sintomática em ritmo sinusal, com BRE e QRS ≥150ms, com FEVE ≤ 35% a despeito da TFO I A
TRC é recomendada para pacientes com ICFEr sintomática em ritmo sinusal, com BRE e QRS entre 130 e 149ms, com FEVE ≤ 35% a despeito da TFO IIa B
TRC pode ser útil em pacientes com ICFEr com CF NYHA III ou IV ambulatorial, ritmo sinusal, com distúrbio de condução intraventricular não BRE e QRS ≥150ms, com FEVE ≤35% a despeito TFO IIa B
TRC pode ser útil em pacientes com ICFEr com CF NYHA III ou IV ambulatorial, ritmo sinusal, com distúrbio de condução intraventricular não BRE e QRS entre 130-149ms, com FEVE ≤35% a despeito de TFO IIb B
TRC não é recomendada em pacientes com QRS <130ms, sem outra indicação para estimulação de ventrículo direito III A

TRC: terapia de ressincronização cardíaca; ICFEr: insuficiência cardíaca com fração de ejeção reduzida; BRE: bloqueio de ramo esquerdo; FEVE: fração de ejeção do ventrículo esquerdo; TFO: terapia farmacológica otimizada; NYHA: New York Heart Association; CF: classe funcional.

3.2. Paciente com Fibrilação Atrial

A prevalência de FA em pacientes com IC varia de acordo com a gravidade, ocorrendo em 5% dos pacientes em classe funcional I (NYHA) e chegando a 40% em pacientes em classe funcional IV (NYHA). 185

Os dados sobre a TRC em pacientes com FA e IC são limitados, mas sugerem benefícios ainda que sejam menores que em pacientes em ritmo sinusal. Isso ocorre devido a algumas peculiaridades relacionadas com a FA, tais como perda do sincronismo atrioventricular, maior risco de falha da estimulação ventricular sincronizada devido à dificuldade de controle da frequência cardíaca e ocorrência de batimentos de fusão e pseudofusão, maior incidência de disparo de choques de CDI (apropriados ou inapropriados), além de maior incidência de hospitalização e mortalidade. 186 - 190

Ritmo de FA ocorre em cerca de um quarto dos pacientes submetidos à TRC; entretanto, a maioria dos ensaios clínicos controlados e randomizados que demonstraram o benefício da TRC excluiu pacientes com FA (p. ex., CARE-HF, COMPANION). 191

No estudo CARE-HF, que comparou TRC com TFO, embora a mortalidade tenha sido mais elevada em pacientes que desenvolveram FA durante o seguimento, esses pacientes se beneficiaram da TRC quando considerados os principais objetivos do estudo. 192

A associação de TRC com CDI (TRC-D) não foi superior ao CDI isoladamente no subgrupo de 229 pacientes com FA do estudo RAFT. Entretanto, menos de um terço dos pacientes recebeu mais que 95% de estimulação biventricular nos 6 meses de seguimento.

Uma metanálise 193 que incluiu cinco estudos (quatro coortes prospectivas e o ensaio clínico randomizado MUSTIC 194 ) comparou as respostas à TRC em 797 pacientes em ritmo sinusal e 367 pacientes em FA (ablação de nó AV foi realizada em 56 pacientes). Não houve diferença na melhora da classe funcional (NYHA) entre os pacientes, embora os resultados do teste de caminhada de 6 minutos e qualidade de vida (escore de Minnesota) tenham sido melhores no grupo em ritmo sinusal.

Em outra metanálise, 195 com 23 estudos observacionais, que incluiu 7.495 pacientes submetidos à TRC, os pacientes com FA (25%) tiveram maior taxa de não respondedores (34,5% vs. 26,7%) e maior taxa de mortalidade (10,8% vs. 7,1%) quando comparados com pacientes em ritmo sinusal. Ademais, a presença de FA foi associada com menor impacto da TRC na qualidade de vida, no teste de caminhada de 6 minutos e no volume diastólico final do ventrículo esquerdo, mas com resultado semelhante na melhora da fração de ejeção do ventrículo esquerdo.

O estudo MUSTIC-AF 196 incluiu 59 pacientes com IC e FA com bradicardia (estimulação ventricular com QRS estimulado ≥ 200ms), randomizados para estimulação de VD ou biventricular, com crossover no terceiro mês. Ablação do nó AV foi realizada em 63% dos pacientes. O estudo teve como limitação um número baixo de crossover , o que limitou qualquer conclusão (somente 39%). Não houve diferença significativa na tolerância ao esforço e pico de consumo de oxigênio entre os dois grupos quando se considerou a intenção de tratar. Entretanto, ao considerar os 37 pacientes que receberam terapia adequada (97% a 100% de estimulação biventricular), encontrou-se um aumento significativo na distância da caminhada em 6 minutos e no pico de consumo de oxigênio.

Os estudos PAVE, 197 OPSITE 198 e AVIL CLS/CRT 199 mostraram que a TRC acrescentou um modesto, mas significante, efeito na qualidade de vida, classe funcional (NYHA) e FEVE em comparação com estimulação apical de VD, em pacientes com FA submetidos à ablação do nó AV, com vários graus de disfunção de VE.

O registo ADHERE 200 comparou pacientes com IC, QRS ≥ 120ms e FEVE ≤ 35% submetidos a TRC-CDI (n = 4.471) com pacientes que não foram submetidos a implante de dispositivo (n = 4.888). O uso de TRC-D foi associado a menor risco de mortalidade e internação hospitalar. Esta associação foi observada também no subgrupo de 3.357 pacientes com FA.

O benefício da TRC requer estimulação biventricular na maior parte do tempo, evitando ao máximo a condução intrínseca. Em pacientes com FA com condução atrioventricular rápida, essa condição pode ser dificultada.

Boriani et al. 201 avaliaram 1.404 pacientes submetidos à TRC em seguimento médio de 18 meses. Todos estavam em ritmo sinusal no momento da inclusão no estudo, tendo sido documentada FA em 443 pacientes (32%). A duração dos episódios variou de >10 minutos a semanas e ocorreu tanto em pacientes sem história (22%) quanto em pacientes com história de FA (16%). A porcentagem de estimulação biventricular no grupo que apresentou FA foi de 95% versus 98% no total de pacientes. Quando os pacientes que apresentaram em FA estavam em ritmo sinusal, a porcentagem de estimulação biventricular foi de 98% versus 71% durante os episódios de FA (p < 0,001). Estimulação biventricular < 95% foi definida como subótima, a qual foi correlacionada com a ocorrência de FA persistente ou permanente (p < 0,001), e frequência ventricular não controlada (p = 0,002). A porcentagem de estimulação biventricular foi inversamente proporcional à frequência cardíaca em pacientes com FA, reduzindo em 7% para cada aumento de 10 batimentos na frequência ventricular.

A importância da alta taxa de estimulação biventricular foi confirmada em uma grande coorte de 36.395 pacientes que participaram do US LATITUDE Patient Management System, na qual os pacientes foram seguidos por monitoramento remoto. 202 A mortalidade foi inversamente proporcional à porcentagem de estimulação biventricular, tanto em ritmo sinusal quanto em FA ou estimulação atrial. A maior redução na mortalidade foi observada com estimulação biventricular > 98%. Pacientes com estimulação biventricular > 99,6% tiveram redução de mortalidade de 24% (p < 0,001), enquanto os que tiveram estimulação biventricular < 94,8% tiveram aumento de 19%. O tempo ideal de estimulação biventricular foi ≥ 98,7%.

No estudo APAF, prospectivo e multicêntrico, foram incluídos 186 pacientes com FA permanente sintomática com frequência ventricular não controlada ou IC refratária, disfunção sistólica de VE e QRS largo. 203 Todos foram submetidos a ablação do nó AV e implante de MP multissítio e randomizados para TRC guiada por ecocardiograma (97 pacientes) ou estimulação apical de VD (89 pacientes). No seguimento médio de 20 meses, a TRC reduziu o evento composto por hospitalização ou morte por IC, ou piora da IC (11% no grupo TRC vs. 26% no grupo estimulação apical de VD). A TRC reduziu hospitalização e piora da IC. A mortalidade total foi similar entre os grupos.

A ablação do nó atrioventricular (AV) elimina a condução intrínseca, resultando em estimulação biventricular em 100% do tempo em pacientes com TRC. Esta estratégia foi avaliada em uma série de 673 pacientes (FEVE ≤ 35%, NYHA ≥ II, QRS > 120ms). 187 Entre os 162 pacientes com FA permanente dessa coorte, 48 receberam medicação para controle da frequência cardíaca e 114 pacientes foram submetidos à ablação do nó AV. Em 4 anos de seguimento, o remodelamento reverso e a tolerância ao esforço foram similares entre os pacientes com FA e ritmo sinusal. Entre pacientes com FA, o benefício da TRC foi observado somente nos pacientes que tinham sido submetidos à ablação do nó AV. A despeito da estimulação biventricular >85% do tempo, os pacientes com controle medicamentoso da FC não tiveram melhora na função do VE e da capacidade funcional.

Gasparini e colaboradores demonstraram, em um estudo observacional, que ablação do nó AV associada à TRC melhorou significativamente a sobrevida quando comparada à TRC isolada. 204 Entre os 1.285 pacientes avaliados, 243 estavam em FA. O controle da FC (85% de estimulação biventricular) foi realizado por ablação do nó AV em 188 pacientes, e terapia medicamentosa em 55. No seguimento de 34 meses, a mortalidade foi significativamente menor em pacientes submetidos à ablação do nó AV (4,3% vs. 15,2% HR = 0,26 para mortalidade geral e 0,15 para mortalidade por IC). Esses resultados sugerem que a meta em TRC deve ser 100% de estimulação biventricular, visando ao benefício máximo da terapia.

O estudo CERTIFY ratificou a importância da ablação do nó AV em pacientes com FA submetidos à TRC. O estudo comparou a evolução clínica dos pacientes com FA permanente submetidos à TRC combinada com ablação do nó AV (n = 443) ou controle de frequência com medicamentos (n = 895) com pacientes em ritmo sinusal (n = 6.046). Os resultados mostraram, em seguimento médio de 37 meses, mortalidade por todas as causas (6,8 vs . 6,1 por 100 pessoas ano) e mortalidade cardiovascular (4,2 vs. 4,0) similares nos pacientes com FA + ablação do nó AV e ritmo sinusal. Em contraste, os pacientes com FA + controle medicamentoso tiveram maior mortalidade total e cardiovascular (11,3 e 8,1, respectivamente; p < 0,001) ( Tabela 15 ).

Tabela 15. – Recomendações para indicação de terapia de ressincronização cardíaca (TRC) em pacientes com fibrilação atrial.

  Classe de recomendação Nível de evidência
TRC deve ser considerada em pacientes com FA permanente, FEVE reduzida (< 50%) e indicação de ablação do nó AV para controle de frequência cardíaca I B
TRC deve ser considerada em pacientes com FA permanente, FEVE ≤ 35% e CF NYHA III e IV a despeito da terapia medicamentosa otimizada, com BRE e QRS ≥ 130ms, assegurando estratégia que permita estimulação biventricular > 95% IIa B
Ablação da junção atrioventricular deve ser considerada em casos de incompleta estimulação biventricular (< 95%) IIa B

CF: classe funcional; FEVE: fração de ejeção do ventrículo esquerdo; NYHA: New York Heart Association; TRC: terapia de ressincronização cardíaca.

3.3. Upgrade de Marca-passo Convencional

Pacientes submetidos à estimulação crônica do VD por MP convencional ou CDI podem evoluir com disfunção sistólica progressiva do VE em decorrência de dissincronia elétrica e mecânica. Dados de registros indicam que a disfunção do VE induzida por MP pode ocorrer em 12 a 30% dos pacientes. 205 , 206 Para esse diagnóstico, deve-se documentar alta taxa de estimulação ventricular, afastadas outras causas de disfunção de VE, como isquemia miocárdica, valvopatias e arritmias sem controle adequado.

Estudos não randomizados indicam que a TRC pode reverter a disfunção induzida por MP. 207 - 209 A melhora ou recuperação da FEVE observada nesses estudos ocorreu em até 86% dos pacientes. Assim, portadores de MP ou CDI com alta taxa de estimulação ventricular, que apresentam piora clínica e/ou ecocardiográfica, podem ser considerados para upgrade de sistema para TRC ( Tabela 16 ).

Tabela 16. – Recomendações para indicação de terapia de ressincronização cardíaca (TRC) upgrade em portadores de marca-passo convencional.

  Classe de recomendação Nível de evidência
TRC para pacientes com MP convencionais, com alta taxa de estimulação ventricular direita (> 40%), que evoluem com piora clínica e/ou ecocardiográfica, mesmo com tratamento otimizado, afastadas outras causas de disfunção ventricular esquerda (FEVE ≤ 35%). IIa B

MP: marca-passo; TRC: terapia de ressincronização cardíaca.

3.4. Na Indicação de Maca-passo Antibradicardia (1º Implante)

O possível efeito deletério da estimulação ventricular com MP convencional pode justificar a indicação de TRC como opção de escolha para o tratamento de bradicardias.

Estudos sugerem que a disfunção de VE induzida por MP (12 a 30%) é maior quando a taxa de estimulação ventricular excede a 40%, ou até mesmo 20%. A FEVE prévia ao implante e a duração do QRS também foram preditores da ocorrência de disfunção ventricular induzida por MP.

Ensaios clínicos randomizados avaliaram se a TRC seria superior à estimulação convencional do VD para reduzir a ocorrência de remodelamento do VE e desfechos clínicos em pacientes com FEVE > 35%. 210 , 211 Esses estudos foram avaliados conjuntamente em metanálise que comparou a estimulação convencional do VD com a estimulação biventricular ou do feixe de His. 212 A estimulação biventricular, em comparação com a estimulação do VD, foi associada a maior FEVE e redução de volumes finais sistólico e diastólico do VE (maior probabilidade de benefício com FEVE entre 36% e 52%). A melhora no teste de caminhada de 6 minutos ocorreu de modo significativo apenas em um estudo que incluiu pacientes com FA permanente que foram submetidos à ablação de nó AV (pacientes com FEVE < 45% e com IC NYHA II/III apresentaram melhor resultado no teste). Os dados relacionados à estimulação do feixe de His estão apresentados em outra sessão deste documento.

Importantes estudos não foram inseridos nesta metanálise por incluírem também pacientes com FEVE < 35%. O estudo BLOCK HF incluiu pacientes com indicação de MP por bloqueio AV, com IC NYHA I-III e FEVE ≤ 50%. 56 Os pacientes foram randomizados para estimulação convencional do VD ou estimulação biventricular. A FEVE média foi de 43%, sendo que 87% apresentavam FEVE > 35%. O desfecho primário composto de morte por qualquer causa, atendimento de urgência por IC com necessidade de diurético intravenoso e aumento de 15% ou mais no volume sistólico final do VE ocorreu mais frequentemente no grupo de estimulação do VD. O estudo APAF incluiu pacientes com FA permanente submetidos à ablação de nó AV e implante de TRC. 204 Os pacientes foram randomizados para estimulação biventricular ou estimulação de VD. O desfecho primário composto por morte por IC, hospitalização ou piora da IC ocorreu em 26% dos pacientes no grupo VD e 11% no grupo TRC. O benefício da estimulação biventricular ocorreu independentemente da FEVE e da duração do QRS. A FEVE média dos pacientes foi de 38%, sendo que 53% apresentavam FE > 35%.

Um estudo brasileiro (COMBAT 55 ), prospectivo, multicêntrico, randomizado, duplo-cego e cruzado, incluiu 60 pacientes com CF II-IV (NYHA), FEVE < 40% e bloqueio AV como indicação de MP. Todos os pacientes foram submetidos a implante de ressincronizador cardíaco; no entanto, a cada 3 meses, alternaram estimulação biventricular com estimulação convencional de VD (grupo A: VD-TRC-VD e grupo B: TRC-VD-TRC). Ao final de cada trimestre, os pacientes foram avaliados, tendo sido encontrado melhora significativa da qualidade de vida, CF, e volume sistólico final do VE com a TRC em comparação com a estimulação do VD. Ademais, a taxa de mortalidade foi maior com estimulação de VD.

As recomendações de TRC para pacientes com indicação de MP convencional para tratamento de bradiarritmias estão resumidas na Tabela 17 .

Tabela 17. – Recomendações para indicação de terapia de ressincronização cardíaca (TRC) a pacientes com indicação de marca-passo convencional.

  Classe de recomendação Nível de evidência
TRC para pacientes com bloqueio atrioventricular, com indicação de implante de marca-passo definitivo e disfunção sistólica de VE (FEVE < 40%). I A

FEVE: fração de ejeção do ventrículo esquerdo; TRC: terapia de ressincronização cardíaca.

3.5. Na Indicação de Cardio-desfibrilador Implantável (TRC-D)

O implante de CDI está indicado em diferentes condições clínicas como estratégia de prevenção primária e secundária de morte súbita (ver item 4). Muitos desses pacientes apresentam também disfunção de VE e bloqueio de ramo esquerdo (BRE), atendendo a critérios de indicação para TRC (ver item 3). Desta forma, há pacientes que podem se beneficiar de ambas as terapias (TRC-D).

Ensaios clínicos randomizados avaliaram os resultados do implante de CDI ou TRC-D em pacientes com disfunção de VE e DCIV. Nos estudos CONTAK, MIRACLE ICD, 213 MIRACLE ICD II e RethinQ, 214 todos os pacientes foram submetidos a implante de TRC-D, sendo que um grupo permanecia com a TRC ativada, e o outro, não.

No estudo CONTAK, foram incluídos 490 pacientes com indicação convencional para implante de CDI, com FEVE ≤ 35%, NYHA II-IV, QRS ≥ 120ms. Não houve diferença significativa no desfecho primário composto incluindo morte por qualquer causa, hospitalização por IC e arritmia ventricular com necessidade de intervenção do CDI. A TRC-D melhorou de modo significativo o consumo de O2, teste da caminhada e FEVE. O MIRACLE ICD incluiu 369 pacientes, também com indicação convencional para implante de CDI, com FEVE ≤ 35%, NYHA III-IV, QRS ≥ 130ms. A TRC-D foi associada a melhora da qualidade de vida e consumo de O2. No MIRACLE ICD II, foram incluídos 186 pacientes, indicação convencional de CDI, FE ≤ 35%, NYHA II, QRS ≥ 130ms. A TRC-D foi associada à redução dos volumes do VE, melhora da FEVE e da classe funcional (NYHA). No estudo RethinQ, foram incluídos 172 pacientes com indicação e CDI, FE ≤ 35%, NYHA III, QRS ≤ 130ms e evidência de dissincronia por ecocardiograma. Não houve benefícios da TRC-D para o desfecho de consumo O2.

Já nos estudos MADIT CRT 165 e RAFT, 167 os pacientes foram submetidos a implante de TRC-D ou de CDI isolado. No MADIT CRT, participaram 1.820 pacientes com FEVE ≤ 30%, NYHA I-II, QRS ≥ 130ms. Houve redução significativa do desfecho combinado de morte por qualquer causa ou evento não fatal de IC com a TRC-D. No estudo RAFT, foram incluídos 1.798 pacientes com FEVE ≤ 30%, NYHA II-III, QRS ≥ 120ms ou QRS estimulado por MP ≥ 200ms. A TRC-D reduziu significativamente o desfecho combinado de morte por qualquer causa ou internação por IC. Como desfecho secundário, foram analisados separadamente morte e internações por IC, havendo redução significativa dos dois desfechos.

Análise conjunta dos resultados desses ensaios clínicos evidenciou que a TRC-D reduziu mortalidade total em comparação com CDI isolado (risco relativo 0,84; IC95% 0,73-0,96). 215 Também foi demonstrada redução significativa de hospitalizações (RR 0,75; IC95% 0,64-0,88).

A decisão clínica sobre implante de TRC-D em pacientes com indicação de CDI deve levar em consideração o padrão de bloqueio intraventricular e a duração do complexo QRS. Resultados de metanálises apontam que os benefícios da TRC se restringem principalmente a pacientes que apresentam padrão de BRE. 216 , 217 Em análise do estudo RAFT, 167 o benefício sobre desfecho primário ocorreu apenas nos pacientes com QRS > 150ms. Nos pacientes com padrão de BRE, houve relação contínua entre duração do QRS e benefício clínico. Nos pacientes com padrão não BRE, o benefício ocorreu apenas com QRS > 160ms ( Tabela 18 ).

Tabela 18. – Recomendações para indicação de terapia de ressincronização cardíaca (TRC) em pacientes com indicação de cardioversor-desfibrilador implantável (CDI).

  Classe de recomendação Nível de evidência
Em pacientes com indicação de CDI que apresentem BRE com QRS ≥ 150ms, FEVE reduzida e IC sintomática sob TMO I A
TRC para pacientes com bloqueio atrioventricular, com indicação de implante de CDI e disfunção sistólica de VE IIa B

BRE: bloqueio de ramo esquerdo; CDI: cardioversor-desfibrilador implantável; FEVE: fração de ejeção do ventrículo esquerdo; IC: insuficiência cardíaca; TMO: terapia medicamentosa otimizada; TRC: terapia de ressincronização cardíaca; VE: ventrículo esquerdo.

3.6. Estimulação Direta do Sistema Excito-condutor Cardíaco

A TRC é um tratamento não farmacológico bem estabelecido para tratamento de pacientes com IC sintomáticos, FEVE reduzida e QRS largo. A despeito do avanço dessa modalidade terapêutica, ainda se observa taxa de não respondedores de 20 a 40%. 218 Neste cenário, a estimulação direta do sistema de excito-condutor cardíaco (feixe de His ou ramo esquerdo) pode ser a alternativa útil. 219 - 225

Em 2012, Pichardo et al ., descreveram uma série de 16 pacientes com IC grave em que houve falha na estimulação do VE através do seio coronário. Nessa série, foi possível obter-se a correção do distúrbio de condução (BRE) com a estimulação direta do feixe de His em 81% dos casos. 226 Lustgarten et al . apresentaram outro estudo que envolveu 29 pacientes, randomizados para estimulação do feixe de His ou TRC convencional. Os resultados quanto ao teste de caminhada de 6min, classe funcional, questionário de qualidade de vida e FEVE foram semelhantes entre os grupos. 227

Em uma metanálise que envolveu 11 estudos e 494 pacientes submetidos à estimulação do feixe de His, observou-se sucesso do implante em 82,4%. Foram incluídos pacientes com FA em programação de ablação do nó AV e indicação de TRC. A estimulação hissiana demonstrou, em pequenos estudos observacionais, resultados promissores apontando a necessidade de estudos randomizados. 228

Um recente estudo randomizado comparou a estimulação hissiana com a estimulação biventricular convencional em 41 pacientes. Dos 21 pacientes randomizados para estimulação do feixe de His, 10 (48%) necessitaram de crossover para o grupo convencional. O principal motivo para a troca da terapia foi a falta de correção do distúrbio de condução, mais distal, com a estimulação do feixe de His. Este achado está em linha com estudos subsequentes que demonstraram correção do bloqueio de ramo em torno de 60% dos casos com a estimulação hissiana. Não obstante, 26% dos pacientes do grupo de TRC convencional também cruzaram para o grupo da estimulação do His devido a dificuldades técnicas. Os pacientes do grupo de estimulação hissiana tiveram maior estreitamento do QRS (174 ± 18ms para 125 ± 22ms, p < 0,001 vs. 165 ± 17ms para 164 ± 30ms, p = 0,82) e melhora de parâmetros ecocardiográficos superior em comparação ao grupo TRC (80% dos pacientes de estimulação hissiana apresentaram aumento absoluto > 5% FEVE). As medidas de diâmetros cardíacos e volumes foram semelhantes em ambos os grupos. Estes achados sugerem que as terapias possam ser complementares, principalmente devido a dificuldades anatômicas do seio coronário e impossibilidade de correção do distúrbio da condução em todos os pacientes com BRE. 229

Zhang et al . descreveram uma série de 11 pacientes consecutivos com BRE e indicação clássica de TRC, que foram submetidos à estimulação direta do ramo esquerdo. Observou-se expressivo encurtamento do QRS (129,09 ± 15,94ms em comparação ao QRS nativo de 180,00 ± 15,86ms, p < 0,01). Os autores concluíram que a estimulação septal profunda em pacientes com disfunção sistólica e BRE é factível, resultando em melhora funcional e remodelamento reverso. 230 , 231 Huang et al . acompanharam uma coorte prospectiva de 63 pacientes com miocardiopatia dilatada, BRE e FEVE < 35% submetidos à estimulação do ramo esquerdo, em que foi possível corrigir o distúrbio de condução em 61 pacientes. Em 1 ano de seguimento, houve aumento significativo da FEVE, remodelamento reverso e melhora da classe funcional. Ademais, 75% dos pacientes foram classificados como hiper-respondedores, com normalização da FEVE. 232

Wu et al ., em estudo não randomizado, avaliaram 137 pacientes com IC sintomática e QRS largo acompanhados por 1 ano. Nesse estudo, 49 pacientes foram submetidos á estimulação do feixe de His, 32 do ramo esquerdo, e 54 pacientes foram submetidos à estimulação biventricular. Os pacientes submetidos à estimulação do sistema excito-condutor (feixe de His ou ramo esquerdo) obtiveram maior incremento da FEVE em comparação à estimulação biventricular (23,9%, 24% e 16,7% respectivamente, p < 0,05). A melhora funcional foi semelhante nos dois grupos de estimulação do sistema excito-condutor; contudo, os limiares de estimulação foram menores com a estimulação do ramo esquerdo (0,49 vs. 1,35, p < 0,001). 233

Em recente e elegante revisão sobre estimulação direta do sistema excito-condutor cardíaco, Sharma et al . propõem que essa modalidade terapêutica possa ser utilizada como estratégia inicial para TRC ou como terapia de resgate nos casos de insucesso da técnica convencional, por cateterizarão do seio coronário. Também consideram a associação dos métodos (estimulação simultânea His-TRC) como um novo conceito de estimulação para TRC em casos selecionados com distúrbios mais difusos da condução intraventricular 234 ( Tabela 19 ).

Tabela 19. – Recomendações para indicação de estimulação direta do sistema excito-condutor cardíaco como alternativa à terapia de ressicronização cardíaca (TRC).

  Classe de recomendação Nível de evidência
BAV e indicação de MP e FEVE < 50%, é razoável a estimulação fisiológica IIa B
IC sintomática, FEVE ≤ 35%, QRS ≥ 130ms é possível considerar a estimulação direta do sistema de condução como alternativa à TRC convencional IIa C
Em não respondedores à TRC, é razoável considerar a estimulação direta do sistema de condução como terapia de resgate IIb C

BAV: bloqueio atrioventricular; FEVE: fração de ejeção do ventrículo esquerdo; IC: insuficiência cardíaca; MP: marcapa-passo; TRC: terapia de ressicronização cardíaca.

3.6.1. Modulação de Contratilidade Cardíaca

A modulação da contratilidade cardíaca (MCC) é uma opção terapêutica utilizada para IC em pacientes que não têm indicação convencional para a TRC, com QRS estreito (< 130ms) e FEVE entre 25% e 45%. Nessa modalidade de terapia, promove-se a estimulação artificial do septo interventricular direito com alta voltagem 30 a 40ms após a ativação dos cardiomiócitos durante o período refratário absoluto. Em teoria, essa estimulação otimiza a dinâmica do cálcio, aumentando a contratilidade ventricular, resultando em melhora na tolerância ao exercício e capacidade funcional dos pacientes. 235 Não está claro ainda, entretanto, se a MCC deve ser indicada rotineiramente em pacientes que não são bons candidatos (QRS estreito) ou são não respondedores à TRC, ou, ainda, se deve ser usada em conjunto com os ressincronizadores cardíacos convencionais. 236 Os estudos até então publicados são pequenos e, talvez por isso, incapazes para detectar diferenças significativas. 237

4. Recomendações para Indicação de Cardioversor-desfibrilador Implantável

4.1. Prevenção Primária de Morte Súbita

4.1.1. Miocardiopatia Isquêmica

A morte súbita (MS) causada por arritmias ventriculares é uma das principais causas de morte em pacientes com IC com FEVE reduzida (ICFER), em especial na população de pacientes com cardiopatia isquêmica em que a incidência de fibrose miocárdica ventricular e, consequentemente, circuitos de reentrada é mais prevalente. 238 , 239

É importante frisar que os estudos que avaliaram o impacto do CDI na cardiopatia isquêmica definiram essa doença, de forma geral, como disfunção ventricular secundária a pelo menos uma lesão severa em uma das três principais artérias coronárias, ou história prévia documentada de IAM. 240 - 242

Vários estudos importantes testaram o impacto do CDI na profilaxia primária para MS em pacientes com cardiopatia isquêmica. Os estudos MADIT (Multicenter Automatic Defibrillator Implantation Trial) 243 e MUSTT (Multicenter Unsustained Tachycardia Trial) 244 testaram o CDI em pacientes com TVNS, disfunção ventricular esquerda e com indução de arritmias ventriculares sustentadas no EEF. O estudo MADIT incluiu pacientes com FEVE < 35%, classe funcional (NYHA) I, II e III, história prévia de IAM, TVNS assintomáticas registradas e arritmias ventriculares induzíveis ao EEF, refratárias ao uso de procainamida, ou antiarrítmico equivalente. Os 196 pacientes incluídos foram randomizados para implante de CDI versus tratamento médico otimizado (TMO), seguimento médio de 27 meses. A mortalidade no grupo CDI foi 15,7% e, no grupo TMO, 38,6%, com redução de risco relativo para mortalidade total o grupo CDI de 64% (RC 0,46; IC 95%, 0,26-0,82; p = 0,009). O estudo MUSTT incluiu pacientes com FE ≤ 40%, CF I, II e III e registro de TVNS assintomáticas. O objetivo inicial do estudo MUSTT foi comparar a eficácia de fármacos antiarrítmicos capazes de suprimir as arritmias ventriculares no EEF versus placebo. Devido aos resultados do MADIT, o protocolo foi modificado para implante de CDI nos casos em que havia indução de arritmias ventriculares e falha na reversão com pelo menos um antiarrítmico. O resultado do estudo MUSTT demonstrou que a diminuição na mortalidade não foi significativa com o uso de antiarrítmico, mas sim com o CDI com redução relativa de risco de morte de 76% (RC 0,24; IC95%: 0,13-0,45; p < 0,001).

Embora o estudo eletrofisiológico não seja utilizado de forma corrente para indicação de CDI devido ao baixo valor preditivo negativo, os estudos MADIT e MUSTT apresentaram resultados importantes nessa população (NNT para redução de mortalidade total = 4,2).

O estudo MADIT II randomizou pacientes com FE ≤ 30%, CF I, II e III e IAM há mais de 30 dias, para TMO vs CDI. Foram randomizados 1.232 pacientes e, em um seguimento médio de 20 meses, a mortalidade no grupo CDI foi 19% e, no grupo TMO, 24%, com redução relativa de risco de mortalidade total de 31% (RC 0,69; IC95% 0,51-0,93; p = 0,016). 245 O estudo SCD-HeFT ampliou os critérios de inclusão, randomizando pacientes com FE ≤ 35%, CF II e III, isquêmicos e não isquêmicos. Durante seguimento médio de 45,5 meses, o grupo CDI teve redução relativa da mortalidade de 23% (RC 0,77; IC95% 0,62-0,96; p = 0,007), sendo que 52% da população incluída foi de pacientes com cardiopatia isquêmica. 246

Por outro lado, os estudos que testaram o implante precoce do CDI após revascularização ou eventos isquêmicos miocárdicos foram neutros e até mesmo negativos em alguns desfechos secundários. O estudo CABG (Coronary Artery Bypass Graft) Patch randomizou 900 pacientes para implante de CDI profilático no intraoperatório de cirurgia de revascularização miocárdica, em pacientes com < 80 anos, FE < 36% e alterações no ECG de alta resolução. Após seguimento médio de 32 ± 16 meses, o estudo foi neutro em relação à mortalidade total (6,7% no grupo CDI e 4,6% no grupo controle, RC 1,07; IC95% 0,81-1,42; p = 0,64). 247 Análise posterior demonstrou aumento da taxa de infecção no grupo do CDI (2,2% vs. 0,4%; p < 0,05). 246 O estudo DINAMIT (Defibrillator after Acute Myocardial Infarction) randomizou 332 pacientes para implante de CDI e 342 para o grupo sem CDI, 6 a 40 dias após IAM. 243 Os pacientes incluídos deveriam apresentar FE ≤ 35% e diminuição da variabilidade cardíaca ao Holter . Durante seguimento de 30 ± 13 meses, não houve diferença na mortalidade total entre os grupos, com 62 mortes no grupo CDI e 58 mortes no grupo controle (RC 1,08; IC95% 0,76-1,55; p = 0,66).

Desde a publicação do estudo MERIT-HF (Metoprolol CR/XL Randomized Intervention Trial in Congestive Heart Failure), sabe-se que a taxa de MS decresce de forma proporcional com a piora da CF (NYHA). 248 Na análise de subgrupos dos estudos SCD-HeFT e MADIT II, pacientes em CF I e II foram os que mais se beneficiaram do CDI, ao passo que, em pacientes em CF III, o benefício não foi tão importante. Não há ensaios clínicos robustos que demonstrem benefício do CDI em pacientes em CF IV (apenas dados de coortes retrospectivas de pacientes em lista de transplante ou submetidos a implante de dispositivos de assistência ventricular [DAV]). Em um estudo retrospectivo que incluiu 1.089 pacientes em lista de transplante cardíaco, 550 possuíam CDI (216 para prevenção primária, 334 para prevenção secundária). Durante seguimento médio de somente 8 meses, morreram 39 pacientes (18%) do grupo CDI (profilaxia primária), 89 (27%) do grupo CDI (profilaxia secundária) e 162 (30%) do grupo sem CDI. Na análise multivariada, a presença do CDI foi preditor independente para diminuição de mortalidade (RC 0,4; IC95% 0,19-0,85; p = 0,016). 248 O mesmo achado foi encontrado analisando a população do UNOS (United Network for Organ Sharing) em seguimento de 1999 a 2014, em que foram incluídos 32l.599 pacientes, com seguimento médio de 154 dias. Nessa população, 3.638 pacientes (11%) morreram na lista de transplante cardíaco, tendo sido de 9% a mortalidade no grupo CDI e 15% no grupo sem CDI, com redução relativa de risco de 13% (RC 0,87; IC 95% 0,80-0,94 p < 0,0001). 249 Nesse mesmo estudo de coorte, no subgrupo de pacientes submetidos a implante de DAV (9.478 pacientes), a presença do CDI foi associada à redução de risco relativo de mortalidade de 19% (RC 0,81; IC95% 0,70 -0,94). Uma revisão sistemática analisou o CDI em pacientes com DAV, incluindo 937 pacientes (em 93% dos casos, o DAV foi utilizado como ponte para transplante). Durante seguimento médio de 7 meses, 16% dos pacientes morreram no grupo CDI e 26% no grupo sem CDI, com redução relativa de risco para morte de 39% (RC 0,61; IC95% 0,46-0,82; p < 0,01) 250 ( Tabela 20 ).

Tabela 20. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na prevenção primária na cardiopatia isquêmica.
  Classe de recomendação Nível de evidência
CDI é recomendado em pacientes com história de IAM > 40 dias ou cardiopatia isquêmica crônica, sob tratamento farmacológico ótimo, sem isquemia miocárdica passível de tratamento por revascularização cirúrgica ou percutânea e expectativa de vida de pelo menos 1 ano e que apresentem FEVE ≤ 35% e CF II-III, ou FEVE ≤ 30% e CF I, II ou III I A
CDI é recomendado em pacientes com história de IAM > 40 dias ou com cardiopatia isquêmica crônica, sob tratamento farmacológico ótimo, sem isquemia miocárdica passível de tratamento por revascularização cirúrgica ou percutânea e expectativa de vida de pelo menos 1 ano, que apresentem FEVE ≤ 40%, TVNS espontânea e indução de TVS no EEF I B
CDI pode ser considerado em pacientes com história de IAM > 40 dias ou com cardiopatia isquêmica crônica, sob tratamento farmacológico ótimo, sem isquemia miocárdica passível de tratamento por revascularização cirúrgica, ou percutânea e candidatos à lista de transplante ou implante de DAV IIa B
CDI não é indicado em pacientes com IAM < 40 dias de evolução, ou indicação de revascularização III B
CDI não é indicado em pacientes com ICFEr CF IV refratários ao tratamento e que não sejam candidatos a transplante ou implante de DAV III C

CDI: cardioversor-desfibrilador implantável; CF: classe funcional; IAM: infarto agudo do miocárdio; DAV: dispositivo de assistência ventricular; EEF: estudo eletrofisiológico; FEVE: fração de ejeção do ventrículo esquerdo; TVNS: taquicardia ventricular não sustentada; TVS taquicardia ventricular sustentada.

4.1.2. Miocardiopatia Não Isquêmica

A IC é uma condição clínica muito prevalente, com elevada morbidade e mortalidade. Em 20% a 30% dos casos, a etiologia é definida como não isquêmica, o que significa que há ausência de lesões significativas na angiografia coronariana ou resultado negativo em método de imagem para investigação de isquemia. A causa da disfunção ventricular esquerda pode ser desconhecida, sendo chamada miocardiopatia dilatada idiopática, ou pode ser atribuída a fatores como infecção viral, hipertensão arterial sistêmica, exposição a agentes potencialmente tóxicos (quimioterápicos, álcool), doença de Chagas, doenças infiltrativas, periparto, valvulopatias, doenças genéticas e autoimunes.

Mesmo que avanços na terapêutica da miocardiopatia não isquêmica (MNI) tenham trazido redução significativa da mortalidade nas últimas décadas, a morte súbita cardíaca (MSC) permanece como problema importante, sendo responsável por 30% dos óbitos. 251 As estratégias de prevenção primária de MSC nos pacientes com MNI incluem tratamento farmacológico, CDI e TRC. Ensaios clínicos randomizados demonstraram que o emprego de fármacos (betabloqueadores, sacubitril/valsartana e espironolactona) reduz significativamente as taxas de MSC neste grupo de pacientes. 252

A estratificação de risco inclui avaliação clínica e laboratorial. Quanto pior a classe funcional (CF, NYHA), maior o risco absoluto de mortalidade geral e de MSC. A MSC é a causa de 64% dos óbitos de pacientes em CF II, 50% em CF III e 33% em CF IV (progressão da IC é a causa de 50% dos óbitos em CF IV). A ocorrência de síncope é importante fator de risco para MSC em pacientes com MNI. 253 Outras variáveis clínicas associadas a maior risco de eventos arrítmicos nesta população são o não uso de betabloqueadores e a pressão arterial sistólica. 254 , 255 Exames laboratoriais, como hemoglobina, ácido úrico e peptídio natriurético atrial (BNP), aparecem como preditores de mortalidade e eventos arrítmicos em alguns estudos. 256

A redução da FEVE é considerada o principal fator de risco para MSC e mortalidade total em pacientes com IC. Poucos trabalhos, porém, avaliaram a FEVE como fator de risco para MSC em pacientes com MNI. O estudo MACAS (Marburg Cardiomyopathy Study), coorte prospectiva com 343 pacientes com MNI, demonstrou que, para cada 10% de redução da FEVE, houve risco relativo de 2,28 para eventos arrítmicos maiores (pacientes em ritmo sinusal).

A prevalência de QRS largo em pacientes com IC varia de 20% a 50%, estando associado ao aumento de MSC e mortalidade total; entretanto, nos estudos de coorte específicos de pacientes com MNI, não foi demonstrada relação significativa entre prolongamento do QRS e aumento do risco de MSC. 257 , 258

O Holter pode ser útil na avaliação de risco por meio da análise de presença de taquicardia ventricular não sustentada (TVNS) e medidas da atividade autonômica (variabilidade da frequência cardíaca [VFC] e turbulência da frequência cardíaca [TFC]). A incidência de TVNS em pacientes com MNI varia de 30% a 79% e sua utilização na estratificação de risco de eventos arrítmicos é controversa. 259

Dados de metanálise indicam que variáveis derivadas do teste ergoespirométrico como consumo de oxigênio (VO2), inclinação do equivalente ventilatório de CO2 (VE/VCO2 slope ) e presença de ventilação periódica indicam de modo independente aumento do risco de eventos combinados (incluindo mortalidade total, mortalidade cardíaca, transplante cardíaco, hospitalização e necessidade de dispositivo de assistência ventricular). 260 , 261

Na cardiopatia isquêmica, o estudo eletrofisiológico (EEF) com estimulação ventricular programada mostrou-se capaz de identificar pacientes com risco de eventos arrítmicos graves. 246 , 248 Por outro lado, na MNI, os resultados são controversos e as diretrizes geralmente não recomendam a realização rotineira do EEF para estratificação de risco nessa população.

Há vários estudos avaliando a associação entre mutações genéticas com a fisiopatologia e o prognóstico de pacientes com MNI, particularmente naqueles com doença familiar. 262 Entre as condições melhor investigadas, estão as mutações do gene da lâmina A/C (LMNA). Tais mutações são encontradas em 6% a 8% dos casos de MNI, podendo chegar a 30% nos casos de associação com doença do sistema de condução e envolvimento da musculatura esquelética.

A fibrose miocárdica, importante substrato arritmogênico, está presente em cerca de 44% dos pacientes com MNI, de acordo com recente metanálise que incluiu 34 estudos e 4.554 pacientes. 263 Esses pacientes apresentam maior mortalidade, arritmias ventriculares e hospitalizações por IC. 264 Para cada percentual de aumento no volume de realce tardio, há um aumento estimado de risco para mortalidade ou eventos arrítmicos de 3% a 20% (realce tardio de 8%: razão de chances em análise univariada de 8,23-2,84-23,8). Gulati et al. 265 publicaram os resultados da maior coorte de pacientes com MNI submetidos à RM, em que o desfecho combinado MSC e PCR recuperada ocorreu em 29,6% dos pacientes com fibrose miocárdica e em 7,0% dos pacientes sem fibrose. Para esse desfecho, a presença de fibrose apresentou hazard ratio de 4,61 (IC95% 2,75-7,74; p < 0,001) e a extensão da fibrose um hazard ratio de 1,10 (IC95% 1,05-1,16; p < 0,001). Esses resultados indicam que a RM pode ser útil na estratificação de risco de pacientes com MNI.

Diversos estudos avaliaram o impacto do CDI em pacientes com MNI. O maior deles foi o SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial), 245 que incluiu pacientes com cardiopatia isquêmica e não isquêmica, com IC NYHA II-III. No seguimento dos pacientes que implantaram CDI, 33,2% receberam algum choque, 22,4% receberam choque apropriado e 10,7% receberam apenas choques inapropriados. 266 No DANISH Trial, estudo randomizado com pacientes com MNI que tinha como desfecho primário mortalidade por qualquer causa, 556 receberam CDI e 560 receberam somente tratamento clínico otimizado. 267 Após mediana de seguimento de 67,6 meses, o desfecho primário ocorreu em 21,6% dos pacientes no grupo CDI, e 23,4% no grupo controle, sem diferença significativa (p = 0,28).

As recomendações atuais para implante de CDI na MNI estão listadas na Tabela 21 .

Tabela 21. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na prevenção primária na cardiopatia não isquêmica.
  Classe de recomendação Nível de evidência
Paciente com FEVE ≤ 35%, tratamento clínico otimizado, CF NYHA II-III e expectativa de vida >1 ano I B
Deve ser considerado na presença de alterações genéticas de alto risco (especialmente Lamina A/C) associado a dois ou mais fatores: FEVE ≤ 45%, TVNS, mutação de alto risco e sexo masculino IIa B
IC CF NYHA IV refratária sem perspectiva de transplante ou DAC III C

CDI: cardioversor-desfibrilador implantável; CF: classe funcional; DAC: doença arterial coronária; FEVE: fração de ejeção do ventrículo esquerdo; IC: insuficiência cardíaca; NYHA: New York Heart Association; TVNS: taquicardia ventricular não sustentada.

4.1.3. Cardiomiopatia Hipertrófica

A cardiomiopatia hipertrófica (CMH) é uma doença genética causada por uma mutação autossômica dominante em genes que codificam as proteínas dos sarcômeros, com prevalência em torno de 1:500 indivíduos. 268 , 269 É caracterizada pela presença de graus variados de hipertrofia ventricular esquerda assimétrica, na ausência de condições que possam resultar em sobrecarga que expliquem as alterações, podendo ocasionar IC diastólica, obstrução da via de saída do VE, arritmias atriais e ventriculares e, em alguns casos, a morte súbita cardíaca (MSC). 270 , 271 A maioria dos pacientes não apresenta qualquer sintoma, e a MSC, não raramente, é a primeira manifestação da doença. 272 - 276

Teare publicou, em 1958, os primeiros relatos de uma série de oito pacientes com hipertrofia miocárdica assimétrica (na época, ainda sem nomenclatura bem-definida) e considerou a possibilidade diagnóstica de hamartoma muscular, tendo correlacionado os achados anatômicos com maior ocorrência de MSC em adultos jovens. O aspecto anatomopatológico era de desarranjo grosseiro dos feixes musculares com hipertrofia da fibra muscular individualmente e dos seus núcleos. 277

Os pacientes com diagnóstico de CMH apresentam aproximadamente 1% de risco anual para MSC; porém, alguns pacientes podem ter esse risco bem maior de acordo com determinadas características de risco. 278 Na era “pré-CDI”, a taxa de mortalidade girava em torno de 1,5% ao ano; com a introdução do CDI, essa taxa tem sido reduzida para 0,5% ao ano. 279 , 280

Maron et al. publicaram, em 2019, um estudo longitudinal e unicêntrico, envolvendo uma grande coorte de 2.094 pacientes com CMH com seguimento de 17 anos. Dentre 527 pacientes que tinham pelo menos um fator de risco convencional e que receberam CDI para prevenção primária, 15,6% apresentaram terapias apropriadas (TV/FV), correspondendo a quase 50 vezes o número de eventos em comparação com o grupo sem CDI. 281

A CMH é a causa mais comum de MSC em indivíduos < 40 anos de idade, e a maioria dos episódios é devido à fibrilação ventricular; dessa forma, a abordagem mais efetiva para reduzir a mortalidade de pacientes de alto risco é com o implante de CDI, que não é isento de complicações e pode ocasionar desconforto, estresse psicológico e tem custos elevados. 282 , 283 A seleção de pacientes pode ser difícil em virtude de características individuais, manifestações clínicas, história famíliar e definições dos fatores de risco, além de a MSC ser um achado infrequente na prática clínica. 284 , 285 A maior probabilidade de benefício do CDI é baseada em testes não invasivos incluindo a história clínica, ECG, teste de esforço, Holter , ecocardiograma e ressonância magnética do coração (RMC). Os fatores de risco de MSC convencionais são: história familiar de MSC relacionada à CMH, síncope inexplicada ocorrendo em até 6 meses da avaliação, taquicardia ventricular não sustentada (TVNS), espessura septal ≥ 30mm e modificadores de risco incluindo resposta hipotensora ao teste de esforço, fibrose do VE e IC com FEVE < 50%. 286 Recomenda-se que a estratificação de risco para indicação de CDI deve ser realizada periodicamente a cada 1 ou 2 anos em pacientes com CMH. 287

A indicação de CDI na CMH não está embasada em estudos clínicos randomizados, mas, sim, em dados de estudos observacionais. Adicionalmente, estudos com pacientes com CMH portadores de CDI demonstraram que a taxa de eventos potencialmente fatais, com terapias apropriadas do dispositivo, ocorrem em 12%/ano na prevenção secundária e em 4%/ano na prevenção primária. 288 Neste contexto, a probabilidade de ocorrência de terapias apropriadas parece ser similar em pacientes com 1, 2, 3 ou mais fatores de risco convencionais (prevenção primária), levando à conclusão de que a presença de um único marcador pode justificar o implante do CDI. Dentre os fatores de risco convencionais, a história familiar de MSC, definitivamente, ou provavelmente, relacionada a CMH, em parentes de primeiro grau com idade ≤ 50 anos, principalmente na infância e adolescência, tem importância bastante significativa. 289 Outro marcador de risco de MSC consiste na extensão e na magnitude da hipertrofia, sendo mais importante quando ≥ 30mm; espessamento limítrofe (28 a 29mm) pode ser considerado a critério do cardiologista. Spirito et al. demonstraram, em 480 pacientes, que a incidência de MSC foi aproximadamente duas vezes maior a cada aumento de 5mm na espessura miocárdica ventricular, sendo de 1,8%/ano em casos com espessura ≥ 30mm. 290

A presença de síncope inexplicada, sendo improvável ou afastada a possibilidade de síncope vasovagal e não relacionada à obstrução da via de saída do VE, foi fortemente associada ao risco de MSC em um estudo com pacientes com CMH, principalmente se ocorreu até 6 meses da avaliação inicial, com o risco 5 vezes maior do que aqueles sem síncope. Episódios remotos (superior a 5 anos da avaliação inicial) não se correlacionaram a aumento no risco de MSC. 291

A TVNS é definida com a presença de 3 ou mais episódios com 3 ou mais batimentos ventriculares repetitivos e/ou 1 ou mais episódios prolongados com 10 ou mais batimentos a 130bpm ou mais, detectados em Holter de 24 a 48 horas. A incidência de TVNS tem sido relatada em 20% a 46% de pacientes com CMH. Episódios de taquicardia ventricular (TV) estão claramente associados a MSC em pacientes com CMH; entretanto, os dados são menos robustos em demonstrar que a presença isolada de TVNS seja um fator de risco independente. Por outro lado, o risco aumenta na presença de modificadores de risco, especialmente fibrose do VE. 292 , 293

O aconselhamento genético é importante em pacientes com CMH. A identificação do portador de mutação genética específica pode auxiliar na investigação da doença em parentes próximos. Indivíduos com genes positivos podem desenvolver CMH; por isso, devem ser acompanhados com atenção para a presença de características da doença ou fatores de risco ao longo do tempo. 294

Existe evidência cada vez mais forte da relação entre fibrose miocárdica (RMC) e risco de MSC, sendo considerado modificador de risco. 295 - 297 Em estudo brasileiro envolvendo pacientes com CMH de alto risco portadores de CDI, Shiozaki et al. demonstraram que a presença de fibrose miocárdica foi identificada em 96,4% desses pacientes, com taxa de fibrose média de 15,96%, sugerindo que esse fator possa ter maior sensibilidade em relação aos demais marcadores de risco convencionais. 298 Chan RH et al. demonstraram na CMH que a detecção de área de fibrose maior que 15% da massa ventricular esquerda se associou ao dobro do risco de MSC em pacientes considerados inicialmente de baixo risco. 299

Klopotowski et al. avaliaram prospectivamente 328 pacientes com CMH submetidos à RMC com o objetivo de avaliar a importância da localização da fibrose como ferramenta auxiliar na estratificação de risco de MSC. Áreas de realce tardio indicando a presença de fibrose além da região do septo interventricular em paciente com CMH foram associadas a maior risco de MSC ou equivalente, como TV instável ou terapia apropriada do CDI. Esse estudo sugere, considerando a calculadora de risco elaborada pela Sociedade Europeia de Cardiologia (ESC), que, em pacientes classificados de risco intermediário, a presença de fibrose além da região do septo interventricular pode identificar o paciente com maior benefício ao uso do CDI, favorecendo a indicação do dispositivo. 300 - 302

O uso da calculadora de escore de risco na CMH tem sido incentivado pela ESC, mas tem baixa sensibilidade para a decisão de implante de CDI em pacientes de alto risco. A estratégia da sociedade americana de cardiologia de analisar fatores de risco individualmente ou associado a modificadores de risco em cada paciente com CMH apresenta sensibilidade de 95% em predizer eventos de TV potencialmente fatais, sendo superior ao modelo matemático do escore de risco da ESC que apresenta sensibilidade em torno de 34%. Por outro lado, a calculadora da ESC apresenta maior sensibilidade em identificar os pacientes verdadeiramente de baixo risco, com menor probabilidade de eventos (em torno de 92% comparado a 78% da sociedade americana), evitando implantes de CDI desnecessários.

Resposta anormal ou hipotensão ao exercício ocorre em 1 a cada 3 pacientes com CMH. O mecanismo reflete queda exacerbada na resistência vascular sistêmica devido à disfunção autonômica e/ou obstrução dinâmica da via de saída do VE. Em pacientes jovens, a resposta anormal da PA está associada a aumento do risco de MSC. 303

Aneurisma apical de VE identificado por ecocardiografia ou RMC, independentemente do tamanho, pode estar associado a maior risco de TV sustentada monomórfica. 304 Rowin et al. avaliaram, retrospectivamente, 1.940 pacientes com CMH e identificaram aneurisma de VE em 93 pacientes (4,8%) com taxa de eventos adversos de 6,4%/ano, correspondendo a 3 vezes mais que nos pacientes sem aneurisma, incluindo MSC, terapia apropriada do CDI, eventos tromboembólicos e evolução terminal de IC com FEVE < 50%. O CDI foi indicado como prevenção primária em 54 destes pacientes, considerando o aneurisma isolado como fator de risco em 19 casos, ocorrendo terapia apropriada do CDI para TV/FV em 20%. A taxa de eventos arrítmicos foi de aproximadamente 5%/ano, sendo mais que 5 vezes a taxa de ocorrência em pacientes sem aneurisma, sugerindo equivalência a outros marcadores de risco convencionais nos pacientes com CMH de alto risco. 305

O CDI subcutâneo tem vantagens potenciais especialmente em jovens, uma vez que preserva o sistema venoso e evita complicações crônicas dos cabos-eletrodos (desde que não seja necessário estimulação ventricular). Por outro lado, a eficácia do CDI subcutâneo em abortar a FV em pacientes com CMH ainda permanece incerta. 306

Por fim, as pesquisas que avaliaram o papel do estudo eletrofisiológico (EEF) na estratificação de risco para MSC em pacientes com CMH não observaram benefício nessa estratégia; com isso, não deve ser indicado com essa finalidade 307 ( Figura 2 e Tabela 22 ).

Figura 2. – Algoritmo de prevenção primária da MSC no portador de CMH. fração de ejeção do ventrículo esquerdo; HF+: história familiar positiva; m: meses; MSC: morte súbita cardíaca; PA: pressão arterial; TVNS: taquicardia ventricular não sustentada; VE: ventrículo esquerdo.

Figura 2

Tabela 22. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na prevenção primária na cardiomiopatia hipertrófica.
  Classe de recomendação Nível de evidência
Pacientes com pelo menos 1 fator de risco maior IIa B
Pacientes com TVNS ou hipotensão arterial anormal no esforço, na presença de fatores de risco adicionais é razoável indicar CDI desde que expectativa de vida >1 ano IIa B
Estratificação de risco invasiva com EEF III B

Fatores de risco: Espessura de parede > 30mm; História familiar de MS; TVNS; presença de realce tardio na RM; síncope há menos de 5 anos; aneurisma de VE; FEVE <50%. CDI: cardioversor-desfibrilador implantável; EEF: estudo eletrofisiológico; TVNS: taquicardia ventricular não sustentada.

4.1.4. Cardiomiopatia Chagásica

A doença de Chagas (DCh) é causada pelo protozoário parasita Trypanosoma cruzi , transmitido aos seres humanos pelas fezes de um inseto hematófago, da família Triatominae , na maioria dos casos. 308 Geralmente, a infecção ocorre na infância e a fase aguda tem período de incubação de 1 a 2 semanas, podendo durar até 3 meses. Segue-se a fase crônica, na qual, por muito tempo (2 a 4 décadas, na maioria das vezes), os pacientes apresentam apenas sorologia positiva, sem sintomas ou outros sinais de doença clinicamente aparente. 309 , 310 Tais pacientes apresentam, portanto, a chamada forma indeterminada da DCh, cujo prognóstico é essencialmente benigno. 311 , 312 Enquanto, por mecanismos patogenéticos ainda incompletamente entendidos, 313 muitos pacientes permanecem por toda a vida com essa forma da doença, cerca de 30% a 50% dos indivíduos infectados evoluem para as formas determinadas: cardíaca, digestiva ou mista.

A cardiopatia chagásica crônica (CCC) tem peculiaridades fisiopatológicas muito acentuadas e constitui a forma clínica mais comum e mais grave da DCh, sendo responsável por expressiva morbimortalidade na América Latina e em países com expressiva imigração. 314

Estima-se que 8 a 10 milhões de pessoas estejam infectadas pelo Trypanosoma cruzi na na América Latina e em outros países. 315 - 317 Considerando o pior cenário, com base nas estimativas anteriores, pode-se deduzir que 3 a 5 milhões de indivíduos infectados manifestarão formas clínicas da doença em sua fase crônica.

A taxa de mortalidade anual média atribuída à CCC é estimada em 4%, podendo variar de 1% a 10% conforme estratificação de risco embasada em características clínicas e exames cardiológicos simples. 318

Além dos critérios empregados na estratificação de risco, diversos marcadores de pior prognóstico têm sido identificados por vários autores, especialmente no que se refere à MSC em diversos contextos clínicos. 319 - 338 Características como pré-síncope e síncope, disfunção ventricular esquerda e IC, taquicardia ventricular sustentada (TVS) ou não sustentada (TVNS), bradiarritmia grave (DNS e BAV) e parada cardíaca recuperada foram identificadas como marcadores de risco de MSC. Por outro lado, extrassístoles ventriculares isoladas ( Holter ) e bloqueio de ramo direito não interferem significativamente no prognóstico da CCC.

A MSC, responsável por aproximadamente 55% a 65% de todas as causas de óbitos, frequentemente se associa a manifestações de IC, mas pode também ocorrer em pacientes com disfunção de VE assintomática. 339 - 341 A IC refratária é causa de morte em cerca de 25% a 30% dos pacientes. A correlação entre os estágios da CCC e causas de mortalidade foi descrita recentemente: a MSC é mais prevalente no estágio III da doença, enquanto prevalência de morte por IC aumenta progressivamente do estágio I a IV.

O principal mecanismo de morte súbita na CCC é arritmogênico, sendo que a TVS com fibrilação ventricular (FV) subsequente é responsável pela imensa maioria dos eventos letais. 342 Nesse sentido, as anormalidades estruturais da CCC (com inflamação, morte celular e fibrose reativa e reparativa) constituem o substrato anatômico ideal porque promovem bloqueios unidirecionais e áreas de condução lenta propícias para desencadeamento de reentrada elétrica. Os disparadores que incidem sobre esse substrato anatômico, as extrassístoles ventriculares, também invariavelmente presentes, completam os elementos essenciais para a instalação da taquiarritmia ventricular por reentrada. 314 , 329 , 331 , 333 , 336 , 338 , 345 Assim, TVNS pode ocorrer em cerca de 40% dos pacientes com CCC e alterações regionais da mobilidade segmentar, e em praticamente todos os pacientes com disfunção sistólica global de VE e IC. 343 A TVS, de prognóstico mais ominoso, ocorre espontaneamente e pode ser reproduzida em cerca de 80% a 85% dos pacientes durante estudo eletrofisiológico. 319 , 328 , 329

O BAV total é outra causa, mas menos comum, de MSC na CCC, como consequência da degeneração necrótica e fibrose difusa, predominantemente na região atrioventricular. 330

Como já referido, a MSC também pode ser resultante de tromboembolismo pulmonar maciço ou de tromboembolismo sistêmico em órgãos vitais. Excepcionalmente, a MSC pode ocorrer em consequência de rompimento de aneurisma apical de VE.

Rassi et al. desenvolveram um escore para estratificação de risco para mortalidade de pacientes com CCC, com base em variáveis clínicas e exames cardiológicos básicos. 322 Este escore também foi aplicado por outros investigadores 344 em coorte retrospectiva (149 pacientes) que também propuseram que a presença de TV (teste ergométrico ou Holter ), FEVE < 0,50 e QRS > 150ms (ECGAR) possa identificar pacientes com CCC e risco de morte em 5 anos. A ausência ou presença de um fator caracterizaria grupo de baixo risco; risco intermediário quando ocorrem dois fatores; e alto risco quando há presença dos três fatores.

A prevenção primária de MSC em pacientes com CCC inclui, em tese, o uso de amiodarona ou CDI. 345 Nao há, entretanto, evidência científica que sustente a indicação de CDI na prevenção primária de MSC na CCC. Há muitas particularidades patogenéticas e fisiopatológicas que dificultam qualquer comparação direta com os resultados da literatura em outras cardiopatias. 317 A peculiaridade mais marcante é a de que muitos pacientes com CCC, mesmo com função de VE preservada, já possuírem substrato para arritmias potencialmente letais 331 , 345 - 348 ( Tabela 23 ).

Tabela 23. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na prevenção primária na cardiopatia chagásica crônica.
  Classe de recomendação Nível de evidência
TVS estável com FEVE <35% em tratamento clínico otimizado I C
TVS estável com FEVE >35% em tratamento clínico otimizado IIa C
TVNS; FEVE <35% em tratamento clínico otimizado IIb C
Pacientes em CF NYHA IV refratária não candidatos a transplante cardíaco III C

CDI: cardioversor-desfibrilador implantável; CF: classe funcional; FEVE: fração de ejeção do ventrículo esquerdo; NYHA: New York Heart Association; TVS: taquicardia ventricular sustentada; TVNS: taquicardia ventricular não sustentada.

O estudo CHAGASICS, em andamento, é um ensaio clínico randomizado, multicêntrico, nacional e aberto, que tem como objetivo comparar a eficácia do CDI versus amiodarona na prevenção primária de morte por todas as causas, em pacientes com CCC e TVNS, estratificados pelo escore de Rassi. 322 O CHAGASICS deverá ser a evidência científica necessária para definir os critérios para uso da amiodarona e/ou indicação de CDI em pacientes com CCC que não apresentaram desfechos clínicos potencialmente fatais.

4.1.5. Cardiomiopatia Arritmogênica do Ventrículo Direito

A cardiomiopatia arritmogênica do VD (CAVD) é uma cardiomiopatia de herança autossômica dominante e penetrância variável, que provoca mutação de genes que codificam proteínas de adesão celular, os desmossomos. A CAVD afeta predominantemente o VD, mas pode afetar o VE em cerca de 0,5% dos casos, determinando substituição do tecido miocárdico por fibrose e tecido adiposo. Tais alterações estruturais frequentemente causam arritmias ventriculares e morte súbita cardíaca (MSC). 346 , 347 A prevalência estimada da CAVD varia de 1:1.000 a 1:5.000 pessoas na população geral, e é importante causa de MSC em atletas e adultos jovens. 348 Arritmias ventriculares, síncope e MSC ocorrem particularmente na 2ª e 3ª décadas de vida, geralmente durante atividade física. Síncope é relatada em 16% a 39% dos pacientes com CAVD no momento do diagnóstico, está frequentemente relacionada à atividade física e tem sido associada a maior risco de arritmias. 349

Extrassístoles ventriculares frequentes, taquicardia ventricular não sustentada (TVNS) e taquicardia ventricular sustentada (TVS) são preditores importantes de eventos cardíacos. A ocorrência de TVS é preditora importante de MSC e de terapias apropriadas do CDI. A MSC pode ser a primeira manifestação da CAVD. 350 , 351

As arritmias ventriculares têm origem geralmente no VD (morfologia de BRE), mas o eixo do QRS durante a TVS geralmente difere da via de saída do VD (VSVD); muitos pacientes podem ter QRS de múltiplas morfologias. 352

As regiões de tecido fibrogorduroso criam áreas de ativação ventricular retardada, causando deflexões fracionadas no final do complexo QRS (ondas épsilon ) e potenciais tardios ao ECGAR. Em pacientes com suspeita de CAVD, o ECGAR pode ser útil para diagnóstico e estratificação de risco (Classe IIa, nível de evidência B). Achados anormais ao ECGAR estão relacionados com a GRAVIDADE da CAVD na RMC e ocorrência de eventos adversos. 353

A RMC fornece informações quanto à função ventricular, tamanho das cavidades cardíacas, anormalidades segmentares e extensão da fibrose com a técnica de realce tardio com gadolíneo. Com esta técnica, foi demonstrado acometimento biventricular em 34% a 56% dos pacientes e acometimento isolado do VE em 4% a 9%. Áreas de fibrose pela RMC têm relação com a localização do substrato da arritmia ventricular identificado pelo mapeamento eletrofisiológico endocárdico e epicárdico. 354

O estudo eletrofisiológico (EEF) tem valor incerto na CAVD assintomática como preditor do risco de MSC (Classe IIb, nível de evidência B). Em pacientes com CDI para prevenção primária, a indução de TVS não é preditor de choques apropriados. 355 Na Tabela 24 , estão listados os critérios para diagnóstico da CAVD. 355

Tabela 24. – Critérios revisados para diagnóstico da CAVD.
I. Disfunção regional ou global e alterações estruturais
Critério Maior Ecocardiograma
Acinesia segmentar de VD, discinesia ou aneurisma e um dos seguintes (final da diástole):
VSVD no ELPE ≥ 32mm (≥ 19mm/m2 corrigido para ASC)
VSVD no ECPE ≥ 36mm (≥ 21mm/m2 corrigido para ASC)
Mudança na área fraccional ≤ 33%
Ressonância magnética cardíaca
Acinesia segmentar de VD ou discinesia ou contração dissincrônica do VD e um dos seguintes:
Razão do volume diastólico final do VD pela ASC ≥ 110mL/m2 (homens) ou ≥ 100mL/m2 (mulheres)
Fração de ejeção do VD ≤ 40%
Angiografia do VD
Acinesia segmentar de VD, discinesia ou aneurisma
Critério Menor Ecocardiograma
Acinesia segmentar de VD ou discinesia e um dos seguintes (final da diástole):
VSVD no ELPE ≥ 29 a < 32mm (≥ 16 a < 19mm/m2 corrigido para ASC)
VSVD no ECPE ≥ 32 a < 36mm (≥ 18 a < 21mm/m2 corrigido para ASC)
Mudança na área fraccional > 33% a ≤ 40%
RMC
Acinesia segmentar de VD ou discinesia ou contração dissincrônica do VD e um dos seguintes:
Razão do volume diastólico final do VD pela ASC ≥ 100 a < 110mL/m2 (homens) ou ≥ 90 a < 100 mL/m2 (mulheres)
Fração de ejeção do VD >40% a ≤45%
II. Caracterização do tecido da parede
Critério maior Menos que 60% de miócitos residuais pela análise morfométrica (ou < 50% se estimado), com substituição fibrosa da parede livre do VD em mais de uma amostra, com ou sem substituição gordurosa na biópsia endomiocárdica
Critério menor 60% a 75% de miócitos residuais pela análise morfométrica (ou 50% a 65% se estimado), com substituição fibrosa da parede livre do VD em mais de uma amostra, com ou sem substituição gordurosa na biópsia endomiocárdica
III. Alterações na repolarização
Critério maior Ondas T invertidas nas derivações precordiais (V1, V2 e V3) ou além em indivíduos > 14 anos (na ausência de BRD completo QRS ≥ 20ms)
Critério menor Ondas T invertidas nas derivações V1 e V2 em indivíduos > 14 anos (na ausência de BRD completo) ou em V4, V5 e V6
Ondas T invertidas nas derivações V1, V2, V3 e V4 em indivíduos > 14 anos na presença de BRD completo
IV. Alterações na despolarização/condução
Critério maior Onda épsilon (sinais de baixa amplitude entre o final do QRS e o início da onda T) nas derivações V1 a V3
Critério menor Potenciais tardios no ECGAR em ≥ 1 dos 3 parâmetros na ausência de duração QRS ≥ 110ms no ECG
Duração do QRS filtrado ≥ 114ms
Duração do QRS terminal <40 µV (duração do sinal de baixa amplitude) ≥ 38ms
Voltagem da raiz quadrada média dos 40ms terminais ≤ 20 µV
Duração da ativação terminal do QRS ≥ 55ms medido do nadir da onda S ao final do QRS, incluindo R', nas derivações V1, V2 ou V3 na ausência de BRD completo
V. Arritmias
Critério maior TVNS ou TVS com morfologia de BRE e eixo superior (QRS negativo ou indeterminado nas derivações DII, DIII e aVF e positivo em aVL)
Critério Menor TVNS ou TVS da VSVD, morfologia de BRE e eixo inferior (QRS positivo nas derivações DII, DIII e aVF e negativo em aVL) ou eixo desconhecido
> 500 extrassístoles ventriculares em 24 horas (Holter)
VI. História familiar
Critério maior CAVD confirmada em parente de primeiro grau diagnosticado conforme esses critérios
CAVD confirmada em necrópsia ou cirurgia em parente de primeiro grau
Identificação de mutação patogênica categorizada como associada ou provavelmente associada à CAVD
Critério menor História de CAVD em parente de primeiro grau, o qual não é possível determinar se este familiar preenche estes critérios diagnósticos
Morte súbita em em parente de primeiro grau < 35 anos com suspeita de CAVD
CAVD confirmada patologicamente por estes critérios em parente de segundo grau

ASC: área de superfície corpórea; BRD: bloqueio de ramo direito; BRE: bloqueio de ramo esquerdo; CAVD: cardiomiopatia arritmogênica do ventrículo direito; ECG: eletrocardiograma; ECGAR: eletrocardiograma de alta resolução; ECPE: eixo curto para-esternal; ELPE: eixo longo para-esternal; RMC: ressonância magnética do coração; TVNS: taquicardia ventricular não sustentada; TVS: taquicardia ventricular sustentada; VD: ventrículo direito; VSVD: via de saída do ventrículo direito.

Os testes genéticos realizados em probandos com suspeita de CAVD são positivos em 30% a 54%. Importante ressaltar que um teste negativo não exclui a doença, e um teste positivo não define a terapêutica. A CAVD é detectada em aproximadamente 35% a 40% dos parentes de primeiro grau, e o rastreamento clínico com ECG, Holter, teste ergométrico e exames de imagem cardíaca pode identificar familiares sob risco de CAVD. 356

Pacientes assintomáticos e sem arritmias ventriculares devem receber apenas betabloqueadores e avaliados periodicamente quanto a função ventricular e ocorrência de arritmias. 357 Não há estudos randomizados avaliando a melhor opção de antiarrítmicos para tratamento da TVS. Um estudo observacional demonstrou supressão de TV induzida em 58% dos pacientes e apenas 10% apresentaram recorrência da arritmia com o uso de sotalol. 358 Em outro registro observacional, betabloqueadores ou sotalol não estiveram associados à redução de arritmias ventriculares; amiodarona foi superior para sua prevenção em pequena coorte. 359 Ablação de TVS reduz a sua recorrência, mas não elimina a necessidade de implante de CDI. 360

Pacientes com história de MSC abortada, TVS mal tolerada e síncope têm maior risco de MSC, com taxa anual >10%, sendo indicado implante de CDI. Fatores de risco para MSC ou choque apropriado reportados em diferentes coortes são: TVS, síncope inexplicada, TVNS frequente, história familiar de MSC precoce, comprometimento extenso do VD, QRS muito prolongado, realce tardio na RMC, disfunção de VE e indução de TVS no EEF. 361 Revisão sistemática recente incluindo 610 pacientes seguidos por 3,8 anos (média) demonstrou taxa anual de 9,5% e 3,7% de choques apropriados e inapropriados, respectivamente. 362 , 363

A indicação de CDI na CAVD para prevenção primária é medida de difícil avaliação e deve contar com uma avaliação clínica detalhada considerando história familiar, severidade da disfunção de VD e VE, complicações a longo prazo do CDI, impacto psicológico e econômico. A Tabela 25 e o Figura 3 listam as indicações de CDI na CAVD para prevenção primária.

Tabela 25. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na prevenção primária na cardiopatia arritmogênica do ventrículo direito.
  Classe de recomendação Nível de evidência
CAVD e MSC abortada ou TVS se sobrevida esperada maior que 1 ano I B
CAVD e disfunção importante de VD ou FEVE ≤ 35% se sobrevida esperada maior que 1 ano I B
CAVD e síncope presumidamente por arritmia ventricular, CDI pode ser útil se sobrevida esperada maior que 1 ano IIa B
CAVD e TVS bem tolerada, avaliando o benefício e risco de complicações a longo prazo do CDI IIa C

CAVD: cardiopatia arritmogênica do ventrículo direito; CDI: cardioversor-desfibrilador implantável; FEVE: fração de ejeção do ventrículo esquerdo; MSC: morte súbita cardíaca; TVS: taquicardia ventricular sustentada; VD: ventrículo direito.

Figura 3. – Fluxograma de estratificação e indicação de implante de CDI na CAVD. CDI: cardioversor-desfibrilador implantável; CAVD: cardiopatia arritmogênica do ventrículo direito; MSC: morte súbita cardíaca; TVS: taquicardia ventricular sustentada; TVNS: taquicardia ventricular não sustentada; VD: ventrículo direito; VE: ventrículo esquerdo.

Figura 3

Um novo modelo preditor de arritmias ventriculares na CAVD foi recentemente publicado. 364 As variáveis preditoras foram: sexo masculino, idade, síncope nos últimos 6 meses, TVNS prévia, número de extrassístoles ventriculares no Holter -24h, número de derivações com onda T invertida nas derivações inferiores e anteriores e fração de ejeção do VD. Esse novo modelo determinou maior refinamento na seleção de pacientes para implante de CDI quando comparado ao Fluxograma da International Task Force, publicado em 2015, reduzindo em 20,6% a indicação de implante. 365 Os autores do novo modelo disponibilizaram uma calculadora de risco online (www.arvcrisk.com) que calcula o risco de arritmia ventricular em 5 anos; embora não determine o limiar de risco aceitável para implante do CDI, acredita-se que o modelo auxilia no processo de decisão para prevenção primária.

4.1.6. Miocardiopatia Não Compactada

A miocardiopatia não compactada (MNC) é uma rara anormalidade congênita, caracterizada pela formação de trabeculações proeminentes e recessos intertrabeculares profundos nos ventrículos esquerdo (VE) e direito (VD), que ocorre durante a fase de embriogênese do endomiocárdio (entre a 5ª e 8ª semana de vida fetal), atingindo mais comumente o ápice do VE. 363 O comprometimento de ambos os ventrículos ocorre em 22% a 38% dos pacientes. A não compactação do VE ocorre em associação com outras cardiopatias congênitas ou de forma isolada. 366

A herança genética surge padrão autossômico dominante em pelo menos 30% a 50% dos pacientes; vários genes que causam não compactação do VE já foram identificados. Esses genes geralmente codificam proteínas do sarcômero (aparelho contrátil) ou citoesqueléticas. 367

A apresentação clínica da MNC é heterogênea, variando de casos totalmente assintomáticos até casos com manifestações graves e fatais como IC, tromboembolismo, BAV e intraventricular, arritmia ventricular e morte súbita cardíaca (MSC). Idade, diâmetro diastólico final do VE, IC sintomática, FA permanente ou persistente, bloqueio de ramo e doença neuromuscular associada são preditores de maior mortalidade. 368

O ecocardiograma (ECO) é rotineiramente realizado na investigação inicial, e o uso de contraste pode melhorar a sensibilidade do diagnóstico. 369 A RMC possibilita a visibilização da proporção de miocárdio não compactado e compactado e permite a identificação de trombos e fibrose miocárdica. 370

A MSC é a principal causa de morte na MNC, podendo ocorrer em qualquer idade; não existem ferramentas diagnósticas precisas para a estratificação de risco nesses pacientes. Arritmias ventriculares são relatadas em 38% a 47% e a MSC ocorre em 13% a 18% dos pacientes. 371 O exame histológico evidencia continuidade entre o endocárdio ventricular e os recessos intertrabeculares profundos que podem facilitar a arritmogênese pela formação de circuitos reentrantes subjacentes ao tecido cicatricial, predominando no ápice do VE e segmentos médio-apicais. 372

Steffel et al . demonstraram que a indução de TV sustentada no EEF para estratificação de risco tem pouco valor na MCN; por outro lado, a não indução pode identificar pacientes de baixo risco. 373 A ablação por cateter endo e/ou epicárdica parece ser útil em portadores de CDI com arritmias ventriculares frequentes. 374 , 375 A taxa de choques apropriados nesses pacientes, na prevenção secundária, é de 33% a 37% em seguimento médio de 34 a 40 meses. 376

Não existem dados convincentes demonstrando que a não compactação do VE, por si só, seja suficiente para indicação de implante de CDI. Tal indicação deve ser guiada pela gravidade da disfunção sistólica do VE e presença de arritmia ventricular sustentada (como na cardiomiopatia dilatada idiopática). 377 , 378 No entanto, Caliskan et al ., em estudo que incluiu 77 pacientes com MNC, demonstraram que, entre os pacientes que receberam CDI para prevenção secundária, a disfunção de VE ou sua dilatação não eram proeminentes, tornando esses critérios frágeis para a indicação de prevenção primária. Por outro lado, a presença de TVNS foi mais frequente na prevenção secundária em relação aos pacientes que receberam CDI para prevenção primária ou não foram submetidos ao implante de CDI. Outros fatores de risco que devem ser considerados são história familiar e ocorrência de síncope. 378

Apesar de não haver estudos prospectivos que abordem a prevenção de MSC em MNC, existem dados de estudos observacionais suficientes que suportam a indicação do CDI como estratégia razoável para prevenção de MSC nesses pacientes 379 ( Tabela 26 ).

Tabela 26. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) na Prevenção Primária na Miocardiopatia não compactada.
  Classe de recomendação Nível de evidência
Pacientes com MNC que apresentem FEVE ≤35%, classe funcional de NYHA II-III e expectativa de vida de pelo menos 1 ano devem receber o CDI como estratégia de prevenção da MSC, baseado em recomendações para CMD I B
Pacientes com diagnóstico de MNC com função ventricular normal com fatores de risco incluindo: TVNS, história familiar de MSC e síncope podem receber o CDI como estratégia de prevenção da MSC IIb C
Pacientes com MNC com função ventricular normal e sem fatores de risco não devem ser considerados candidatos a implante de CDI III C
Realização de EEF para estratificação de risco III C

CDI: cardioversor-desfibrilador implantável; CMD: cardiomiopatia dilatada; EEF: estudo eletrofisiológico; FEVE: fração de ejeção do ventrículo esquerdo; MNC: miocárdio não compactado; NYHA: New York Heart Association; TVNS: taquicardia ventricular não sustetantada.

4.1.7. Síndrome do QT Longo e Síndrome do QT Curto Congênito

A síndrome do QT longo congênito é caracterizada por prolongamento do intervalo QT e arritmias ventriculares polimórficas frequentemente deflagradas por estímulo adrenérgico. 380

Centenas de mutações já foram descritas em mais de 13 genes diferentes responsáveis por canais iônicos restauradores do potencial de repouso da célula miocárdica. O padrão de herança pode ser autossômico dominante ( Romano-Ward ), como nos tipos LQT 1 a 6; autossômico recessivo, associado a surdez congênita e intervalos QT muito prolongados ( Lange-Nielsen ); e autossômico dominante, associado a defeitos extracardíacos em associação a dismorfismos e paralisia periódica hipo ou hipercalemica ( Andersen-Tawil , LQT7). 379 , 381

O risco de morte súbita depende de uma série de fatores como o tipo de mutação responsável pelo fenótipo, a duração do intervalo QT e a presença de sintomas. Pacientes com intervalo QT muito prolongado (QTc > 500ms) ou com síncope recorrente podem ter risco anual de morte súbita de até 5%.

A síndrome do QT curto é definida por intervalo QTc < 340ms ou < 360ms em sobreviventes de parada cardíaca por FV/TV, história familiar de morte súbita com menos de 40 anos de idade, presença de mutação confirmada ou história familiar de QT curto. Trata-se de doença rara em que mutações em genes de canais de potássio podem ser encontrados em até 20% dos casos. 382

Alguns pacientes com QT curto podem se beneficiar do uso de quinidina. Sobreviventes de parada cardíaca devem ser submetidos a implante de CDI, enquanto os assintomáticos devem ser acompanhados. As recomendações para implante de CDI para prevenção primária na síndrome do QT longo e QT curto estão listadas na Tabela 27 .

Tabela 27. – Indicações de cardioversor-desfibrilador implantável (CDI) na prevenção primária no QT longo e QT curto.
  Classe de recomendação Nível de evidência
Pacientes com QT longo refratário a betabloqueadores e/ou a simpatectomia, com síncope ou TV polimórfica recorrente I B
Pacientes assintomáticos com QTc > 500ms associado a betabloqueador podem se beneficiar de CDI, associado ou não à simpatectomia IIb C
QT longo com estudo genético comprovando LQT2 ou LQT3, associado a betabloqueador IIb B
Pacientes com QTc < 330ms assintomáticos e história de morte súbita familiar IIb C

CDI: cardioversor-desfibrilador implantável; TV: taquicardia ventricular.

4.1.8. Síndrome de Brugada

A síndrome de Brugada é caracterizada pelo supradesnivelamento do segmento ST > 2mm nas derivações V1 e V2 ( coved type ) no 2º, 3º ou 4º espaço intercostal, espontâneo ou induzido por bloqueador de canal de sódio (ajmalina, procainamida), associado à ocorrência de arritmia ventricular polimórfica, síncope ou parada cardíaca.

Pacientes com o padrão eletrocardiográfico típico (“tipo 1”) espontâneo associado à síncope inexplicada ou parada cardíaca recuperada são os que apresentam maior risco de morte súbita. O implante de CDI está associado à redução do risco nesses pacientes sintomáticos 383 ( Tabela 28 ).

Tabela 28. – Indicações de cardioversor-desfibrilador implantável (CDI) na prevenção primária na síndrome de Brugada.
  Classe de recomendação Nível de evidência
TVS ou síncope de provável causa arrítmica e ECG com padrão Brugada tipo 1 espontâneo. I B
Síncope de provável causa arrítmica e ECG com padrão Brugada tipo 1 induzido farmacologicamente. IIa B
TVS induzida por EVP com um ou dois extraestímulos em dois sítios, em pacientes assintomáticos com padrão Brugada tipo 1 espontâneo ao ECG IIb C

CDI: cardioversor-desfirbrilador implantável; ECG: eletrocardiograma; EVP: estimulação ventricular programada; TVS: taquicardia ventricular sustentada.

O aspecto fenotípico da síndrome de Brugada está associado a defeitos genéticos detectáveis em até 30% dos casos. O gene SCN5A está envolvido na maioria das alterações encontradas, mas a pesquisa genética negativa não afasta o diagnóstico. 384

Diversos fatores podem deflagrar as manifestações eletrocardiográficas ou precipitar episódios arrítmicos, como febre, agentes anestésicos e diversos psicotrópicos (www.brugadadrugs.org).

Pacientes assintomáticos têm menor risco de morte súbita. O papel da estimulação ventricular programada (EVP) na estratificação de risco é controversa. Brugada et al . encontraram associação entre a indução de TV polimórfica com aplicação de até 2 extraestímulos no VD e o risco de morte em pacientes assintomáticos. Já a indução de arritmia com 3 extraestímulos reduz a especificidade e deve ser evitada. Outros autores encontraram redução do valor preditivo positivo da EVP ao longo do tempo. 385 - 387

Pacientes com tempestade elétrica e disparo e terapias de choque pelo CDI podem se beneficiar de controle clínico com quinidina e com ablação epicárdica de áreas de ativação anormal do VD identificados no mapeamento eletroanatômico. 388 , 389

4.1.9. Taquicardia Ventricular Polimórfica Catecolaminérgica (TVPC)

A TVPC é uma rara doença arritmogênica, de origem genética, caracterizada por taquicardia ventricular bidirecional e polimórfica, adrenérgico-dependente. A prevalência estimada é de 1 em 10.000 indivíduos. Já foram identificados dois tipos de genes: uma variante dominante secundária à mutação no gene do receptor cardíaco do rianodinio (RyR2) e uma rara variante recessiva causada por mutação do gene da calsequestrina cardíaca (CASQ2). 390

A manifestação clínica usualmente ocorre na 1ª ou 2ª década de vida, sendo desencadeada por exercício físico ou estresse emocional. Em geral, o ECG e o ecocardiograma são normais, mas o teste ergométrico desencadeia arritmias atriais e ventriculares (bidirecional e polimórfica).

O tratamento de escolha é o uso de betabloqueador em dose máxima tolerável. A flecainida e a simpatectomia cardíaca esquerda podem ser associadas como terapias coadjuvantes.

O CDI deve ser considerado nos casos que apresentam parada cardíaca, síncopes recorrentes ou taquicardia ventricular a despeito da TFO 391 ( Tabela 29 ).

Tabela 29. – Indicações de cardioversor-desfibrilador implantável (CDI) na prevenção primária na taquicardia ventricular polimórfica catecolaminérgica.
  Classe de recomendação Nível de evidência
TVPC com síncope ou TVS, apesar do uso de betabloqueador em dose máxima tolerada, ou que tenham contraindicação ao uso de betabloqueador e expectativa de vida > 1 ano IIa C
TVPC assintomática que apresente boa resposta ao tratamento com betabloqueador III C

CDI: cardioversor-desfirbrilador implantável; TVPC: taquicardia ventricular polimórfica catecolaminérgica; TVS: taquicardia ventricular sustentada.

Habitualmente, o CDI deve ser programado com longo intervalo de detecção, porque a dor e o estresse do choque podem desencadear mais arritmias e, consequentemente, tempestade elétrica. A decisão do implante deve levar em conta a alta probabilidade de disparo de choques (apropriados e inapropriados), além da chance de complicações devido à baixa faixa etária dos pacientes.

Em revisão sistemática recentemente publicada, a incidência de choque foi de 40%, tempestade elétrica 19,6%, mortalidades após implante atingiu 1,4% e complicações adicionais chegaram a 32,4% (fratura de eletrodo, endocardite e revisões cirúrgicas), em um faixa etária média de 15 anos (11 a 21 anos). 392

4.1.10. Taquicardia Ventricular Idiopática

Arritmias ventriculares em pacientes com coração estruturalmente normal são, na grande maioria das vezes, benignas. Contudo, uma pequena parcela de pacientes pode apresentar formas malignas de taquicardia ventricular monomórfica, polimórfica e até mesmo fibrilação ventricular.

Muitas dessas taquicardias são desencadeadas por ectopias ventriculares originadas em regiões muito similares àquelas de caráter benigno (via de saída, cúspide aórtica, sistema His-Purkinge, anel mitral e músculos papilares). O mecanismo exato dessas arritmias ventriculares malignas ainda não é completamente esclarecido. A anisotropia associada a condução lenta e bloqueio funcional causados por focos arritmogênicos rápidos provavelmente causa degeneração do ritmo para TV polimórfica e FV.

Características de alto risco estão relacionadas com a ocorrência de síncope ou parada cardíaca e achados eletrocardiográficos de intervalo de acoplamento curto da primeira ou segunda extrassístole, TVNS com ciclos curtos, QRS alargado (em TV ou sinusal) e TV polimórfica. 393 , 394 A Tabela 30 lista as recomendações para prevenção primária com implante de CDI em pacientes com arritmias ventriculares idiopáticas.

Tabela 30. – Indicações de cardioversor-desfibrilador implantável (CDI) na prevenção primária na taquicardia ventricular idiopatica.
  Classe de recomendação Nível de evidência
Pacientes com síncope e TVNS com características malignas e expectativa de vida >1 ano IIb C
Pacientes com TV idiopática assintomática, com características benignas, que apresentem boa resposta ao tratamento farmacológico ou ablativo III C

CDI: cardioversor-desfibrilador implantável; TVNS: taquicardia ventricular não sustentada; TV: taquicardia ventricular.

4.2. Prevenção Secundária de Morte Súbita

4.2.1. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada

4.2.1.1. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada na Presença de Cardiopatia Estrutural

A parada cardiorrespiratória (PCR) por taquicardia ventricular ou fibrilação ventricular (TV/FV) e, subsequentemente, a morte súbita cardíaca (MSC) constituem um grave problema de saúde pública, representando aproximadamente 50% de todas as mortes cardiovasculares. Adicionalmente, dos pacientes que apresentam PCR fora do ambiente hospitalar, a taxa de sobrevida é bastante baixa, entre 6 e 10%. 395 Além disso, esses poucos pacientes que conseguem sobreviver à PCR por TV/FV apresentam alto risco de recorrência de taquiarritmias potencialmente fatais, tornando fundamental a adoção de medidas preventivas, que inclui o tratamento da cardiopatia de base, das comorbidades, do uso de fármacos antiarrítmicos e da seleção de pacientes para implante de CDI. 396 , 397

Nos pacientes com cardiopatia isquêmica ou dilatada, é bastante conhecido o papel protetor de fármacos como os betabloqueadores, inibidores da enzima conversora de angiotensina (IECA), bloqueadores do receptor da angiotensina (BRA) e estatinas, que reduzem a mortalidade total, a mortalidade cardiovascular e a mortalidade súbita. 398

Uma metanálise 399 envolvendo mais que 35.000 pacientes com disfunção ventricular esquerda (FEVE < 40%), demostrou que fármacos como betabloqueadores, IECA, BRA e antagonistas dos receptores mineralocorticoides reduzem o risco de MSC quando comparados ao placebo (HR: 0,89, IC95%: 0,82-0,98, p = 0,02). Quando associado à TFO, o implante de CDI traz benefício adicional ao tratamento clínico, reduzindo ainda mais a taxa de MSC (HR; 0,39, IC95%: 0,30-0,51, p < 0,0001). Mais recentemente, a combinação de um inibidor (LCZ696) da neprilisina com BRA (sacubitril/valsartana) mostrou-se ainda mais eficiente que o enalapril na redução tanto da mortalidade por IC quanto da mortalidade arrítmica. 258 Questiona-se se o LCZ696 teria ação antiarrítmica primária ou se a redução das arritmias cardíacas seria resultado de melhora clínica da IC. 400

O emprego de fármacos antiarrítmicos foi, durante muitos anos, a principal estratégia de prevenção secundária de MSC, ainda que fundamentado em poucos estudos e com elevada taxa de recorrência de eventos. Até o início dos anos 1990, aceitava-se que antiarrítmicos da Classe I (p. ex., quinidina, flecainida, encainida) reduziam as extrassístoles ventriculares e a mortalidade. Com a demonstração subsequente dos efeitos deletérios desses fármacos pós-IAM e na IC, a amiodarona passou a ser a escolha para esses pacientes. O estudo CASCADE envolveu 228 pacientes recuperados de PCR que foram randomizados para tratamento empírico com amiodarona ou fármacos da classe I, orientadas por EEF ou Holter -24h. 401 Em seguimento de 6 anos, a sobrevida livre de eventos (morte cardíaca ou TV) foi de 41% no grupo amiodarona versus 20% no grupo de terapia convencional. No entanto, a falta de grupo placebo não permite concluir se os resultados decorreram de benefício do uso da amiodarona ou dos riscos associados às outras drogas antiarrítmicas.

O CDI é considerado o principal avanço para a prevenção secundária da MSC. Seus benefícios foram avaliados em uma série de ensaios clínicos randomizados. O estudo AVID 402 comparou a terapia antiarrítmica (amiodarona ou sotalol) versus CDI em 1.016 pacientes recuperados de PCR por TV/FV, com TV associada à síncope ou com instabilidade hemodinâmica e FEVE < 40%. A sobrevida foi significativamente maior no grupo CDI em 1 ano (89,3% vs. 82,3%), 2 anos (81,6% vs. 74,7%) e 3 anos (75,4% e 64,1%), p < 0,02. A principal crítica a esse estudo refere-se ao maior número de pacientes em uso de betabloqueador no grupo CDI em relação ao grupo de terapia antiarrítmica. Em análise posterior, observou-se que o benefício do CDI ocorreu principalmente nos pacientes com FEVE mais baixa. 403 Nos pacientes com FEVE > 35%, não houve diferença de sobrevida significativa. Nos pacientes com FEVE entre 20% e 34%, a sobrevida em 1 ano foi de 89,6% versus 79,8% e em 2 anos de 82,5% versus 71,8% (p < 0,05). Nos pacientes com FEVE < 20%, a sobrevida em 1 ano foi de 82,4% versus 73% e, em 2 anos 71,6% versus 63,8%, sem diferença significativa.

O estudo CIDS 404 avaliou o uso de amiodarona versus CDI em 659 pacientes com FV documentada, PCR recuperada, TV associada à síncope, TV > 150 bpm/min com pré-síncope ou angina e FEVE < 35% ou síncope associada à TV indutível ou com episódio de TV espontânea documentada. A mortalidade total após seguimento médio de 4 anos foi de 27% no grupo CDI e 33% no grupo amiodarona, sendo que essa diferença não foi significativa. Em análise subsequente, foi demonstrado que, em pacientes com dois dos seguintes critérios: FEVE < 35%, CF III ou IV e idade > 70 anos, o implante do CDI foi superior. 405 Após acompanhamento médio de 5,6 ± 2,6 anos, a mortalidade foi de 47% no grupo amiodarona comparada a 27% no grupo CDI (p = 0,002).

O estudo CASH 406 incluiu 349 pacientes recuperados de PCR que foram randomizados para tratamento com propafenona, amiodarona, metoprolol ou implante de CDI. O tratamento com propafenona foi suspenso após uma análise interina verificar aumento de mortalidade em comparação aos pacientes com CDI. Após seguimento médio de 2 anos, a mortalidade total foi de 12,1% no grupo CDI versus 19,6% nos grupos amiodarona e metoprolol combinados, sendo que a diferença igualmente não foi significativa.

A metanálise que avaliou os resultados desses três estudos demonstrou 50% de redução relativa na mortalidade arrítmica (p < 0,0001) e 28% na mortalidade total entre os pacientes com CDI em comparação aos que receberam tratamento antiarrítmico, com NNT = 29 (p < 0,00006). 407 O benefício foi maior em pacientes com FEVE < 35% e considerando seguimento de 6 anos (pacientes com CDI tiveram aumento de sobrevida de 4,4 meses).

O estudo MAVERIC comparou terapia guiada por EEF (antiarrítmicos, CDI) versus amiodarona empírica em pacientes com TV sustentada ou PCR recuperada. 408 Os resultados demostraram redução de mortalidade no grupo de pacientes com CDI. A realização de EEF não demonstrou benefício.

As recomendações para implante de CDI na prevenção secundária são baseadas nesses estudos ( Tabela 31 ).

Tabela 31. – Indicações de cardioversor-desfibrilador implantável (CDI) para pacientes recuperados de parada cardíaca ou taquicardia ventricular sustentada na presença de cardiopatia estrutural.
  Classe de recomendação Nível de evidência
Parada cardíaca por TV/FV de causa não reversível, com FEVE ≤35% e expectativa de vida de pelo menos 1 ano I A
TVS espontânea com comprometimento hemodinâmico ou síncope, de causa não reversível com FEVE ≤35% e expectativa de vida de pelo menos 1 ano I A
Sobreviventes de parada cardíaca, por TV/FV de causa não reversível, com FEVE ≥35% e expectativa de vida de pelo menos 1 ano IIa B
Pacientes com TVS espontânea, de causa não reversível, com FEVE ≥ 35%, refratária a outras terapêuticas e expectativa de vida de pelo menos 1 ano IIa B
Pacientes com síncope de origem indeterminada com indução de TVS hemodinamicamente instável e expectativa de vida de pelo menos 1 ano IIa B
TV incessante III C

CDI: cardioversor-desfibrilador implantável; FEVE: fração de ejeção do ventrículo esquerdo; FV: fibrilação ventricular; TV: taquicardia ventricular; TVS: taquicardia ventricular sustentada.

4.2.1.2. Recuperados de Parada Cardíaca ou Taquicardia Ventricular Sustentada na Ausência de Cardiopatia Estrutural

As canalopatias representam um grupo de patologias geneticamente determinadas que envolvem diversos tipos de disfunções dos canais iônicos dos cardiomiócitos, seja aumentando ou reduzindo suas funções e gerando desequilíbrio iônico que aumentam o risco de taquiarritmias potencialmente fatais e de MSC. 409

A gama de mutações genéticas é extremamente ampla, com grande sobreposição de expressões fenotípicas. São classificadas como canalopatias a síndrome do QT longo congênito (SQTL), síndrome de Brugada (SB), taquicardia ventricular catecolaminérgica (TVC), síndrome do QT curto (SQTC), síndrome do ponto J (SPJ), repolarização precoce (RP). A fibrilação ventricular idiopática, a síndrome de morte súbita arrítmica e a síndrome de morte súbita na infância, ainda que possam ter outros mecanismos envolvidos que não somente os genéticos, são discutidos também nessa categoria, pela manifestação arrítmica predominante de TV/FV na ausência de cardiopatia estrutural. Não está no escopo desta Diretriz a discussão detalhada de particularidades de cada expressão fenotípica das canalopatias, sendo possível encontrar extensa referência na literatura. 410 , 411

Pacientes com canalopatias que sobreviveram à PCR apresentam elevado risco de novo episódio de MSC. O implante de CDI nesses pacientes reduz o risco de MSC, sendo observada taxa de terapia apropriada nessa população entre 8% e 33%. Ainda, de maneira geral, pacientes com canalopatias e que apresentem síncope ou taquicardia ventricular apesar do uso de medicações apropriadas para cada condição, usualmente também têm indicação para implante de CDI, a menos que condições específicas, como idade muito prematura ou baixo peso, sejam considerados de risco para o implante do dispositivo. 412 A decisão clínica para o implante de CDI também deve considerar outras opções terapêuticas ou terapias coadjuvantes em situações especiais, como a denervação simpática esquerda na síndrome do QT longo ( Tabela 32 ).

Tabela 32. – Indicações de cardioversor-desfibrilador implantável (CDI) para pacientes recuperados de parada cardíaca ou taquicardia ventricular sustentada na ausência de cardiopatia estrutural.
  Classe de recomendação Nível de evidência
Pacientes com canalopatia e PCR por TV/FV, com expectativa de vida de pelo menos 1 ano I B
Pacientes com canalopatia, que evoluem com TVS ou síncope, apesar do uso de fármacos, com expectativa de vida de pelo menos 1 ano IIa B
Pacientes com SQTLc que evoluem com síncope ou TVS, apesar do uso de betabloqueador em dose máxima tolerada IIa B
Pacientes com SB e alterações eletrocardiográficas espontâneas, síncope presumidamente de causa arrítmica e expectativa de vida de pelo menos 1 ano IIa C
Pacientes com SB e documentação de TVS, com ou sem síncope e expectativa de vida de pelo menos 1 ano IIa C
Pacientes com SB e alterações eletrocardiográficas induzidas por fármacos, síncope de origem indeterminada e expectativa de vida de pelo menos 1 ano IIb C
Pacientes com TVPC que evoluem com síncope ou TVS, apesar do uso de betabloqueador em dose máxima tolerada e expectativa de vida de pelo menos 1 ano IIa C

CDI: cardioversor-desfibrilador implantável; FV: fibrilação ventricular; PCR: parada cardiorrespiratória; SB: síndrome de Brugada; SQTLc: síndrome do QT longo congênito; TV: taquicardia ventricular; TVPC: taquicardia ventricular polimórfica catecolaminérgica; TVS: taquicardia ventricular sustentada.

4.2.2. Síncope e Taquicardia/Fibrilação Ventricular no Estudo Eletrofisiológico

A síncope é um sintoma raro, exuberante e com grande diversidade de etiologia, mecanismos fisiopatológicos e prognósticos, podendo ser o único sintoma que precede a morte súbita. É definida como a perda súbita e transitória da consciência devido à hipoperfusão cerebral, com restabelecimento rápido, espontâneo e completo da consciência.

Na presença de cardiopatia, a ocorrência de síncope, mesmo de origem indeterminada, pode indicar a necessidade de implante de CDI para prevenção de MSC, devido à grande relação dessa manifestação com arritmias ventriculares potencialmente fatais. 413

Em algumas situações o EEF pode ser útil para confirmar a causa arrítmica da síncope. Na cardiopatia isquêmica com FEVE reduzida, a indução de taquicardia ventricular sustentada (TVS) é preditor de risco de MSC; 414 a indução de fibrilação ventricular é considerada achado inespecífico. 415 Por outro lado, a não indução de TVS no EEF identifica um grupo de menor risco 416 ( Tabela 33 ).

Tabela 33. – Indicações de CDI para pacientes com síncope e taquicardia fibrilação ventricular induzidas no EEF.
  Classe de recomendação Nível de evidência
Paciente com cardiomiopatia isquêmica e FEVE > 35% apresentando síncope de origem indeterminada, com indução de TVS ao EEF29-30 I B
Paciente com cardiomiopatia não isquêmica sem indicação para prevenção primária de MS, apresentado síncope, com indução de TVS ao EEF31-32 IIa B

CDI: cardioversor-desfibrilador implantável; EEF: estudo eletrofisiológco; FEVE: fração de ejeção do ventrículo esquerdo; TVS: taquicardia ventricular sustentada.

4.3. Crianças, Adolescentes e Cardiopatia Congênita

As indicações de CDI em crianças não têm sido adequadamente contempladas nas últimas diretrizes internacionais devido ao limitado número de trabalhos publicados sobre o assunto até o momento. Dessa forma, algumas publicações de pequenas séries têm direcionado a maioria das indicações atuais. 417 Frequentemente, as indicações de CDI na faixa etária pediátrica seguem critérios similares aos adultos, tanto para prevenção primária como para prevenção secundária, em que pese a importância do bom senso na avaliação desses pacientes. 418

Recentemente, SOBRAC e DCC-CP publicaram a Diretriz de Arritmias Cardíacas em Crianças e Cardiopatias Congênitas, a qual veio normatizar o diagnóstico e o tratamento de crianças com arritmias cardíacas. 419

É extremamente importante que o médico busque alternativas para evitar a indicação de CDI em crianças, sem que isso represente aumento de risco. Deve-se esgotar o tratamento clínico adequado e, quando indicada, a ablação dos focos arritmogênicos deve ser considerada.

Não existem evidências que suportam a utilização rotineira de CDI na população pediátrica com base apenas na disfunção ventricular esquerda, como ocorre em algumas situações da população adulta. Devido à dimensão do gerador de pulsos e calibre do cabo-eletrodo de choque, a indicação de CDI nessa população deve ser exceção em virtude de vários fatores: dificuldade técnica para implante devido à dimensão do corpo da criança, limitação para confecção da loja do gerador (muitas vezes no abdome), reduzidas opções de via de acesso, risco de tromboses/obstruções venosas e maior risco de extrusão. 420

Para o seguimento clínico e eletrônico, além do ECG e da avaliação rotineira por telemetria, é fundamental a avaliação radiológica periódica, para acompanhar o comportamento do cabo-eletrodo de acordo com o crescimento da criança. Esse cabo deve ser implantado deixando uma curva redundante para permitir o crescimento do paciente sem a necessidade de múltiplas intervenções. 421

A programação eletrônica também apresenta diferenças em relação ao adulto. A função de MP deve considerar frequência básica de estimulação apropriada à idade e à cardiopatia, habitualmente variando de 90 a 160ppm (adolescentes costumam acompanhar protocolos de adultos). Outros fatores importantes são a programação de intervalo AV adaptativo e período refratário atrial pós-ventricular curto, mas adequado para prevenir a ocorrência de taquicardia por reentrada eletrônica, uma vez que esses pacientes costumam ter boa condução ventriculoatrial.

A programação eletrônica do CDI e o seguimento clínico devem ser ainda mais cuidadosos. 422 A facilidade que as crianças apresentam em desenvolver taquicardia sinusal com frequência elevada pode resultar em choques inapropriados com grande impacto psicológico. Choques inapropriados, principalmente logo após o implante, podem afetar a confiança no equipamento e na equipe médica, com desenvolvimento de síndrome do pânico de difícil controle. 423 - 428

Os critérios de discriminação de arritmias e a frequência de detecção devem ser rigorosamente bem ajustados. É recomendável que as terapias sejam restritas a choques, evitando-se o uso de terapias antitaquicardia (ATP; do inglês, antitachycardia pacing ), visto que são raras as situações em que as arritmias potencialmente fatais se apresentam como TV monomórfica. 429 - 431 Quando ocorrem TV monomórficas, devem ser tratadas com ablação sempre que possível. 432 , 433 As TV polimórficas respondem melhor ao disparo precoce de choques, além disso, essas arritmias facilmente degeneram para FV durante as tentativas de ATP.

Em adolescentes, o CDI subcutâneo pode ser opção bastante interessante, visto que esse sistema não utiliza cabos-eletrodos intravasculares. 434 - 437 Essa opção, porém, pode gerar desconforto local e problemas estéticos em pacientes magros devido ao maior tamanho do gerador e presença do cabo-eletrodo paraesternal. Outra limitação do CDI subcutâneo é a incapacidade de estimular cronicamente o coração para tratar bradicardia. Choques inapropriados têm sido descritos com essas próteses em virtude de sua maior suscetibilidade à detecção de ruídos extracardíacos. 438 , 439

Na Tabela 34 , estão dispostas as principais indicações de implante de CDI em crianças e adolescentes.

Tabela 34. – Recomendações para implante de cardioversor-desfibrilador implantável (CDI) em crianças, adolescentes e na presença de cardiopatia congênita.

  Classe de recomendação Nível de evidência
Recuperado de PCR, excluídas causas reversíveis; I B
TV instável com disfunção ventricular, excluídas causas reversíveis A
TV sintomática após avaliação hemodinâmica e anatômica, afastada a possibilidade de correção cirúrgica ou por ablação que pode mudar ou eliminar o substrato arritmogênico C
Síncope recorrente com disfunção ventricular ou TV induzida IIa B
Síncope recorrente associada a SQTL ou TV polimórfica catecolaminérgica em uso de betabloqueador em doses otimizadas B
QT longo congênito, assintomático, não aderente ao tratamento ou história familiar de morte súbita C
Cardiomiopatia hipertrófica obstrutiva, assintomática, com critérios de alto risco C
Cardiopatia arritmogênica de VD com envolvimento ventricular extenso (VD e/ou VE), com TV, história familiar de morte súbita ou síncope indeterminada C
TV incessante ou por causas reversíveis III C

PCR: parada cardiorrespiratória; SQTL: síndrome do QT longo; TV: taquicardia ventricular; VD: ventrículo direito; VE: ventrículo esquerdo.

Atualmente, devido à evolução do tratamento cirúrgico das cardiopatias congênitas, adultos jovens têm apresentado maior sobrevida. Esses pacientes podem apresentar arritmias ventriculares complexas ou até mesmo a MSC, devido à presença de cicatriz miocárdica secundária à própria cardiopatia congênita de base ou devido a manipulação cirúrgica.

Existe extensa correlação entre anormalidades hemodinâmicas residuais e a ocorrência de TV em pacientes submetidos à correção cirúrgica de tetralogia de Fallot (T4F). Hipertrofia e dilatação do VD, além da obstrução ou regurgitação residual da via de saída do VD, são consideradas fatores de risco para a ocorrência de TV e MSC. 440 - 446 Abordagem híbrida, combinando a estratégia cirúrgica para o reparo das alterações estruturais com ablação da arritmia guiada pelo mapeamento pré-operatório ou intraoperatório, tem sido utilizada com sucesso na tentativa de reduzir a incidência de arritmias. 447 - 449 A troca da válvula pulmonar isoladamente nos pacientes com T4F resulta em melhora hemodinâmica e funcional, mas não elimina o risco de TV, sendo necessária avaliação pós-operatória de risco de MSC e eventual indicação de CDI. 450 , 451

Aproximadamente 50% dos implantes de CDI em adultos com cardiopatia congênita são indicados para prevenção secundária, entre 36 e 41 anos de idade. 452 , 453 Esses pacientes apresentam choques apropriados entre 3 e 6% ao ano e taxas de complicações (26% a 45%) e choques inapropriados (15% a 25%) maiores que outras populações. 450 , 451 , 454 - 457 Por isso, o custo-benefício e o impacto psicológico devem ser considerados quando da indicação de CDI nesta população.

Os desafios do implante de CDI em adultos com cardiopatia congênita incluem a complexidade anatômica, os desvios intracardíacos e acesso vascular limitado ao coração. O CDI subcutâneo pode ser uma boa opção para esses pacientes. 456

A indicação de CDI para prevenção primária em pacientes com cardiopatia congênita é controversa. Kairy et al . propuseram um escore de risco para pacientes submetidos à correção cirúrgica de T4F em que a pontuação > 5 seria suficiente para a indicação de CDI. Os critérios considerados nesse escore incluíram: cirurgia de desvio sistêmico-pulmonar paliativo prévia (2), indução de TVS ao EEF (2), QRS ≥ 180ms (1), ventriculotomia (2), TVNS (2) e pressão diastólica final do VE ≥ 12mmHg (3). 457

Os pacientes com reparo cirúrgico de T4F representam cerca de 50% dos implantes de CDI na população de cardiopatia congênita do adulto. Nessa população, choques apropriados ocorrem em até 7,7%/ano na prevenção primária e 9,8%/ano na prevenção secundária. 458 A TVS induzida no EEF em pacientes com cardiopatia congênita parece não se correlacionar com a ocorrência de choques apropriados. 458 A ablação por cateter de TVS monomórfica recorrente pode ser alternativa eficaz, dispensando o implante de CDI em alguns casos. 459 - 463

O maior risco de MSC em pacientes com cardiopatia congênita operada, de acordo com dados de grandes coortes, está na população com transposição das grandes artérias com switch atrial, anomalia da válvula tricúspide tipo Ebstein , estenose aórtica e fisiologia univentricular. 464 - 467

Os pacientes com antecedente de cirurgia de Senning ou Mustard apresentam maior risco de MSC principalmente durante o esforço físico. Nesses pacientes, o switch atrial pode resultar em aumento do volume e consequente estenose das veias pulmonares e aumento das pressões diastólicas finais. 468 Além disso, isquemia e infarto do VD já foram identificados em estudos de perfusão miocárdica em mais de 40% desses pacientes. 469 , 470 Os fatores de risco para PCR em pacientes com switch atrial incluem o fechamento prévio de CIV, sintomas de IC, arritmia atrial, FEVD < 30% a 35% e QRS > 140ms. 471 , 472 Em estudo multicêntrico que avaliou pacientes com switch atrial após implante de CDI, a falta de betabloqueadores foi associada a alto risco de terapia apropriada do CDI. 464 As arritmias atriais frequentemente precedem a TVS em pacientes com transposição, devendo o tratamento para taquicardia atrial ser intensificado. 473 , 474

O risco de MSC é maior entre os pacientes com doença cardíaca congênita do adulto ( Tabela 35 ) em comparação com a população geral; a idade média de óbito varia de 30 a 49 anos. 473 - 476 Pacientes com cardiopatia congênita de complexidade moderada ou grave apresentam risco ainda maior de MSC, correspondendo a aproximadamente 25% das causas de morte cardíaca. 477 , 478

Tabela 35. – Cardiopatia congênita e risco de morte súbita cardíaca.

Cardiopatia congênita Incidência de morte súbita Características de alto risco
Simples
Defeito do septo atrial < 1,5 % Estimulação ventricular; dilatação do ventrículo direito; hipertensão pulmonar; gene NKX2.5
Defeito do septo ventricular < 3% Estimulação ventricular; dilatação do ventrículo direito; hipertensão pulmonar; gene NKX2.5
Moderada
Tetralogia de Fallot 1,4% a 8,3% Síncope inexplicada; arritmia ventricular complexa ou taquicardia Ventricular sustentada; QRS >180ms; indução de taquicardia ventricular sustentada; taquicardia atrial; disfunção do ventrículo esquerdo; dilatação importante do ventrículo direito; insuficiência ou estenose pulmonar severa
Estenose aórtica 3% a 20 % Síncope inexplicada; hipertrofia do VE importante; estenose aórtica com gradiente médio >40mmHg; disfunção ventricular
Coarctação da aorta 2% Aneurisma no local do reparo; estenose aórtica; hipertensão arterial sistêmica; doença coronariana precoce
Anomalia de Ebstein 3% a 6 % Cardiomegalia; fibrilação atrial; taquicardia com complexo QRS largo; insuficiência mitral; dilatação da via de saída do VD
Severa
Transposição das grandes artérias Swicht atrial 3% a 9,5%; Swicth arterial 1%; Corrigida congenitamente 17% a 25% Swicht atrial mustard; fechamento prévio de comunicação interventricular; síncope inexplicada; taquicardia atrial; estenose do óstio coronariano; disfunção ventricular sistêmica; insuficiência tricúspide
Truncus Arteriosus 4% Múltiplas cirurgias de reparo; anomalia das coronárias; disfunção e/ou hipertrofia ventricular
Cirurgia de Fontan para fisiologia de ventrículo único 2,8 a 5,4 % Taquicardia atrial; sobrevida longa; hepatopatia perdedora de proteína; ascite

VD: ventrículo direito; VE: ventrículo esquerdo.

Defeitos septais com história familiar, cardiomiopatia ou bloqueio do sistema de condução podem estar relacionados com mutação do gene NKX2-5, que está associado a risco de MSC precoce, justificando implante de CDI quando o teste genético é positivo. 479 - 481

Pacientes com formas complexas de cardiopatia congênita e várias intervenções cirúrgicas nas primeiras décadas de vida e aqueles que apresentam hipertrofia com subsequente isquemia subendocárdica apresentam maior risco de arritmias ventriculares potencialmente fatais. Outros fatores de risco para MSC em pacientes com cardiopatia congênita incluem maior complexidade da cardiopatia, arritmias ventriculares e supraventriculares, aumento progressivo da duração do QRS, disfunção ventricular sistêmica e disfunção ventricular subpulmonar. História de síncope inexplicada em adultos com cardiopatia congênita de complexidade moderada ou grave pode ser indício de risco de MSC, devendo ser considerado o EEF para avaliar a necessidade de CDI. 457

Os adultos na faixa etária de 40 a 50 anos de idade representam 40% a 67% dos pacientes com cardiopatia congênita que recebem CDI para prevenção primária. Nesses pacientes, choques apropriados ocorrem em 14% a 22% nos primeiros 3 a 5 anos de acompanhamento. 482 Em pacientes sem acesso vascular ou cirurgia de Fontan prévia, o risco de implante de CDI epicárdico pode superar os benefícios potenciais, devendo-se considerar a possibilidade de CDI subcutâneo ou transplante cardíaco. 478

A segurança em relação ao uso de fármacos antiarrítmicos em pacientes com cardiopatia congênita pode ser influenciada pela presença de hipertrofia e disfunção ventricular. O uso de flecainida foi associado à pró-arritmia em 5,8% dos pacientes e MSC em 3,9% em um estudo. 482 Por outro lado, a amiodarona é geralmente reservada para pacientes com arritmias sintomáticas ou para prevenir piora da função ventricular. 483 , 484

As principais recomendações para indicação de CDI em pacientes adultos com cardiopatia congênita estão listadas na Tabela 36 .

Tabela 36. – Recomendações para indicação de cardioversor-desfibrilador implantável (CDI) em pacientes adultos com cardiopatia congênita.

  Classe de recomendação Nível de evidência
Pacientes adultos com cardiopatia congênita e arritmia ventricular complexa ou sustentada na presença de lesões hemodinâmicas residuais importantes, o tratamento da lesão residual, se possível, é indicado antes da consideração do implante de CDI I B
Em pacientes adultos com cardiopatia congênita e TV instável, recomenda-se o implante de CDI após avaliação e tratamento adequado das lesões residuais/disfunção ventricular, com expectativa de sobrevida superior a 1 ano I B
Em pacientes adultos com reparo de tetralogia de Fallot e TV ou FV induzida ou TVS espontânea, o implante de um CDI é razoável ao se esperar uma sobrevida superior a 1 ano IIa B
Em pacientes adultos com doença cardíaca congênita de alta complexidade e que tenha sido submetido a reparo cirúrgico apresentando arritmia ventricular frequente e/ou complexa, betabloqueador pode ser benéfico para reduzir o risco de morte súbita IIa B
Em pacientes adultos com doença cardíaca congênita, antecedente de reparo cirúrgico de complexidade moderada ou grave, com síncope inexplicada e disfunção ventricular moderada ou hipertrofia acentuada, o CDI ou EEF indicando CDI por TVS induzível é razoável se houver sobrevida esperada superior a 1 ano IIa B
Em pacientes adultos com cardiopatia congênita, FEVE < 35% e sintomas de IC, mesmo com fatores de risco adicionais, o implante de CDI pode ser considerado se houver expectativa de sobrevida significativa superior a 1 ano IIb B
TV incessante ou de causas reversíveis III C

CDI: cardioversor-desfibrilador implantável; EEF: estudo eletrofisiológico; FEVE: fração de ejeção do ventrículo esquerdo; FV: fibrilação ventricular; IC: insuficiência cardíaca; TV: taquicardia ventricular; TVS: taquicardia ventricular sustentada.

4.4. Escolha do Tipo de CDI e Modo de Estimulação

Uma vez indicado o implante de CDI para prevenção de MSC, as próximas etapas consistem em escolher a técnica de implante (transvenosa, epimiocárdica ou subcutânea) e escolher o modo de estimulação (ventricular, atrioventricular, biventricular ou atriobiventricular).

4.4.1. Técnica de Implante

Na ausência de comunicação intracardíaca, em pacientes com peso superior a 15kg, habitualmente, a preferência é pela técnica transvenosa. 485 Caso não seja necessário o suporte de bradicardia com indicação de MP (BAV, DNS), o CDI subcutâneo pode ser boa opção.

4.4.2. Modo de Estimulação

Nos pacientes que necessitam da função MP, a escolha do modo de estimulação é fundamental. Modos que priorizam a preservação da condução atrioventricular e intraventricular espontâneas estão associados a menor incidência de FA e remodelamento ventricular relacionado ao BRE induzido pelo estímulo do VD. 486 Ademais, em pacientes que já apresentam remodelamento do VE no momento do implante (FEVE ≤ 40% e diâmetro diastólico final do VE ≥ 60mm) e que necessitarão de estimulação ventricular, a estimulação biventricular é superior à estimulação isolada do VD. 487 , 488

Assim sendo, diante da indicação de CDI, a escolha do modo de estimulação deve considerar se o cronotropismo e a condução atrioventricular são normais, se a condução intraventricular ocorre com padrão de BRE espontâneo ou induzido por MP e se existe ou não remodelamento do VE.

Em pacientes com DNS e condução atrioventricular e intraventricular normais, pode-se optar tanto por CDI unicamerais quanto bicamerais, desde que sejam programados algoritmos de busca de condução intrínseca para se evitar a dissincronia induzida pelo MP.

4.5. Custo-efetividade do CDI na Prevenção Primária e Secundária de Morte Súbita

Os CDI aumentam a sobrevida de pacientes com disfunção ventricular esquerda e risco de morte súbita cardíaca (MSC). No entanto, os custos dessa terapia são elevados e constituem limitação para sua aplicação. Esses custos referem-se ao dispositivo propriamente, gastos hospitalares, honorários médicos, complicações, reinternações e trocas de gerador de pulsos e cabos-eletrodos.

A análise de custo-efetividade é definida pelo custo em moeda corrente por QALY ( quality-adjusted life years ou anos de vida ajustados por qualidade). 489 Outra forma de análise de custo-efetividade refere-se ao custo por anos de vida ganhos. 490 Essas análises variam de acordo com as condições socioeconômicas e culturais da população estudada.

A custo-efetividade deve ser analisada, a partir de trabalhos multicêntricos, em termos de mortalidade. O CDI, por exemplo, não pode ser custo-efetivo se não proporcionar melhora de sobrevida; dessa forma, a perspectiva de sobrevida deve ser fator primordial nessa análise.

O risco de morte por causas não arrítmicas também deve ser levado em conta, uma vez que os CDI não podem preveni-las. Implantes em pacientes com alta morbimortalidade podem ter relação de custo-efetividade desfavorável. A relação de custo-efetividade torna-se desfavorável quando a taxa de sobrevida dos candidatos a CDI for <1 ano. Em pacientes idosos, alguns estudos sugerem que a sobrevida deva ser >5 anos para alcançar boa custo-efetividade. 491 Não existem estudos de custo-efetividade do CDI em doenças menos frequentes como cardimiopatia hipertrófica e canalopatias.

4.5.1. Prevenção Primária

Sanders et al., 492 analisando os resultados dos estudos de prevenção primária MADIT, MADIT II, 247 COMPANION, 161 MUSTT, 493 SCD-HeFT 245 e DEFINITE, 494 estimaram que o uso do CDI adicionaria 1,01 a 2,99 QALYs a um custo de US$ 68.300,00 a US$ 101.500,00. Em comparação com populações controles e considerando-se uma troca do dispositivo a cada 5 anos, a custo-efetividade foi estimada em US$ 30.000,00 a 70.200,00 por cada QALY ganho. Os autores estimaram que essa custo-efetividade seria mantida abaixo de U$ 100.000,00 dólares por QALY, contanto que os CDI mantenham a redução da mortalidade por um período de 7 anos ou mais.

Nos estudos DINAMIT e CABG Patch trial, 250 também analisados por Sanders et al., as populações tratadas com CDI não tiveram redução de mortalidade em comparação com os controles e, portanto, o CDI não foi custo-efetivo. No estudo DINAMIT, o CDI foi testado em população com 6 a 40 dias pós infarto do miocárdio, com FEVE ≤ 35% e baixa variabilidade da frequência cardíaca. O desfecho primário (morte por qualquer causa) não foi diferente entre o grupo que recebeu CDI e o grupo controle. No estudo CABG Patch, a população consistiu em pacientes com doença aterosclerótica coronariana com FE ≤ 35%, ECG de alta resolução alterado e que se submeteram à cirurgia de revascularização miocárdica (CRVM). O implante profilático de CDI durante a CRVM também não mostrou redução do desfecho primário de morte por qualquer causa. Dessa forma, o implante profilático de CDI em pacientes com alto risco de morte súbita (FE ≤ 35%, ECG de alta resolução alterado, variabilidade de RR deprimida) não foi custo-efetivo nos primeiros 40 dias pós infarto ou imediatamente após CRVM.

Um estudo brasileiro de Ribeiro et al., publicado em 2010, avaliou a custo-efetividade em pacientes com IC sob a perspectiva da saúde pública e de saúde suplementar (a efetividade foi aferida em QALY). 495 A relação de custo-efetividade foi de R$ 68.318 por QALY no cenário da saúde pública e R$ 90.942 por QALY na saúde suplementar. Os autores concluíram que o custo do CDI, o tempo para troca do gerador e a efetividade do CDI foram as variáveis mais influentes na análise realizada. Em cenário com pacientes mais complexos, como os do estudo MADIT, a custo-efetividade foi bem mais favorável no cenário público (R$ 23.739,00 por QALY) do que no privado (R$ 33.592,00). 496

Outro trabalho brasileiro, de Matos et al. (2007), analisou a relação de custo-efetividade do CDI comparado com o tratamento medicamentoso. 497 O estudo utilizou a unidade de custo por ano de vida ganho (AVG). O custo por AVG alcançado foi de R$ 20.530,00 (US$ 9.550,00) na época. Esse indicador de efetividade foi calculado com base nos parâmetros de custo incremental de R$ 54.200,00 e expectativa de vida de 2,64 anos, decorrente do uso do CDI comparado com tratamento clínico. Concluiu-se que o índice de custo-efetividade foi favorável sob as condições da realidade brasileira.

No Reino Unido, Buxton et al. (2006) encontraram valores de £ 57.000 por AVG e £ 76.000 por QALY em longo período de acompanhamento. 498 Os autores concluíram que o uso do CDI poderia ser custo-efetivo se utilizado para pacientes com baixa FEVE, incorporando subgrupos de mais alto risco, mas não para uso generalizado.

Em 2009, Cowie et al. realizaram metanálise de estudos de prevenção primária, no contexto europeu, em pacientes com FEVE reduzida e indicações conforme a diretrizes europeias. Nesse cenário, os autores também encontraram boa relação de custo-efetividade. 499 A estimativa média de AVG e QALY foi de 1.88 e 1.57, respectivamente, e a estimativa média de custo por QALY foi de € 31.717. Tais achados foram reproduzidos em outro registro europeu em análise de prevenção primária. 500

Gialama et al. (2014), em revisão sistemática de avaliação econômica sobre o assunto, mostraram que o CDI pode apresentar boa custo-efetividade em grupos selecionados, sendo comparável a outras terapias cardiovasculares e não cardiovasculares já estabelecidas. 496 Fatores como eficácia e segurança, impacto na qualidade de vida, custo do dispositivo (implante e trocas), características dos pacientes e risco de MSC foram as variáveis influenciadoras nessa análise.

4.5.2. Prevenção Secundária

Larsen et al . analisaram a custo-efetividade do CDI no estudo AVID, em que o CDI foi comparado ao tratamento antiarrítmico (principalmente amiodarona) em pacientes que sobreviveram à TVS ou FV. 501 A custo-efetividade do CDI por “anos de vida ganhos” foi calculada em U$ 66.677,00 dólares (IC95% US$ 30.761,00 a US$ 154.768,00) comparada ao tratamento antiarrítmico para o período de 3 anos do estudo. A projeção para 6 e 20 anos manteve os custos estimados em cerca de US$ 68.000 e US$ 80.000 por anos de vida ganhos. Em análise de subgrupo, o CDI foi mais custo-efetivo nos pacientes com FV e menos com FE >35%.

Thjissen et al . 502 avaliaram a custo-efetividade do CDI e demonstraram resultado aceitável em comparação com outros tratamentos pelo sistema de saúde pública, como eritropoetina em pacientes dialíticos, certas quimioterapias para leucemia em idosos, transplante de pulmões e neurocirurgias para tumores malignos intracranianos. O custo por QALY foi semelhante ao de transplante de coração, hemodiálise e diálise peritoneal. Deve-se considerar, ainda, que alguns fatores podem reduzir consideravelmente a custo-efetividade, como complicações, infecções e comorbidades que reduzam a sobrevida do paciente e a longevidade do CDI.

Choques apropriados ou inapropriados podem reduzir a sobrevida e a qualidade de vida e, portanto, a custo-efetividade. Vários estudos avaliaram a importância da programação do CDI, com tempo de detecção de TVS mais prolongado e frequência cardíaca de detecção mais elevada. Essas programações foram capazes de prevenir choques inapropriados e choques “desnecessários”, com melhora de sobrevida e/ou redução de hospitalização. 503 , 504 Dessa forma, programações menos agressivas são capazes de melhorar a custo-efetividade dos CDI.

Mealings et al. analisaram 13 estudos de custo-efetividade de CDI e ressincronizadores e usaram um método analítico para adaptação do tratamento aos custos do Reino Unido. 505 Os autores avaliaram a custo-efetividade em vários subgrupos de pacientes, com base em critérios clínicos como a classe funcional, a duração do QRS, a idade, a presença de BRE e a etiologia isquêmica. Considerando-se um custo aceitável até o limite de £ 30.000 por QALY, os CDI foram custo-efetivos em pacientes com IC e disfunção sistólica ventricular esquerda, em classe funcional NYHA < IV e QRS < 120ms. Para pacientes com QRS entre 120 e 149ms, o CDI foi custo-efetivo apenas nas NYHA I e II. Para pacientes em NYHA IV, a custo-efetividade só foi comprovada para CDI associado ao ressincronizador em pacientes com BRE e QRS >120ms.

Em relação aos pacientes muito idosos, particularmente > 80 anos, a eficácia clínica e a custo-efetividade do CDI são duvidosas. A idade média de entrada dos pacientes nos estudos de prevenção primária e secundária foi de 58 a 66 anos e 58 a 65 anos, respectivamente. No entanto, estima-se que cerca de 28% dos pacientes elegíveis para implante de CDI tenham mais de 80 anos. 506 Dados de mundo real revelam que cerca de 8% a 12% dos implantes nos EUA e Canadá ocorrem em pacientes com mais de 80 anos. A relação morte súbita/morte por qualquer causa decresce com a idade, sendo de 0,51 em idade < 50 anos e de 0,26 para > 80 anos. 507 Como o número de acionamentos apropriados do CDI é semelhante em todas as faixas etárias, tanto na prevenção primária quanto na secundária, a relação morte súbita/morte por qualquer causa diminui no idoso devido ao aumento das mortes relacionadas às demais comorbidades.

Pellegrini et al. estudaram o impacto da idade no momento do implante do CDI na sobrevida. 491 Os pacientes foram estratificados para faixa etária < 65 anos, 65 a 75 anos e > 75 anos. 501 Os pacientes com mais de 75 anos tiveram sobrevida média de 5,3 anos após o implante do CDI (metade dos outros dois grupos). Os autores calcularam que, para uma sobrevida menor que 5 anos, o custo por QALY subiria de U$ 34.000 a U$ 70.200 (no estudo de Sanders 4 ) para U$ 90.000 a U$ 250.000. Nesse caso, o CDI não seria custo-efetivo se o paciente morrer em menos de 5 anos após o implante.

A relação de custo-efetividade para o CDI no Brasil e em países em desenvolvimento precisa ser analisada dentro do contexto socioeconômico, levando-se em conta aspectos locais, PIB, eficácia e complicações. Nesse sentido, deve-se priorizar situações que incluam pacientes de maior risco de morte por arritmia,função ventricular esquerda mais comprometida e menos comorbidades.

A redução do preço dos dispositivos e baterias mais duradouras pode aumentar significativamente a relação de custo-efetividade. Da mesma forma, todos os esforços devem ser realizados para se evitar choques inapropriados ou desnecessários, o que melhora a qualidade de vida (impacto positivo na avaliação do índice QALY) e aumenta a longevidade da bateria.

5. Recomendações para Monitor de Eventos (Loop Recorder) Implantável

O monitor de eventos implantável é um dispositivo que possibilita o monitoramento contínuo do ritmo cardíaco independentemente da participação ativa do paciente. Com capacidade de armazenamento de eventos diversos (bradicardia, taquiarritmia, pausas) e bateria com durabilidade de até cerca de 3 a 4 anos, o monitor de eventos ou looper implantável é uma ferramenta diagnóstica bastante atrativa para investigação de sintomas pouco frequentes (p. ex., menos que 1 vez por mês) com características suspeitas de serem atribuídos a bradi ou taquiarritmias. 508

Em pacientes com síncope inexplicada, em que a investigação não invasiva inicial com ECG, Holter de 24 horas ou monitorização estendida não tenham esclarecido a natureza dos sintomas, o looper implantável demonstrou ser superior à estratégia convencional de investigação, incluindo o Tilt Test e o estudo eletrofisiológico invasivo. Particularmente em pacientes idosos e com distúrbio da condução intraventricular, a principal causa encontrada nesses estudos foi por bradiarritmia. Nesses casos, bradiarritmia foi encontrada em até 41% dos casos, sendo 70% delas BAVT intermitente. 509

Em pacientes com AVC isquêmico criptogênico, em que não há documentação de FA, a busca ativa com ECG seriado e monitorização prolongada pode detectar episódios silenciosos de FA em até cerca de 23% dos casos. 510 A detecção de FA nesses pacientes pode determinar mudança no tratamento, o que pode significar respaldo para anticoagulação plena por tempo indeterminado. Contudo, faltam estudos randomizados que corroborem a eficácia da terapia anticoagulante em pacientes com FA silenciosa detectada por monitorização prolongada no AVC criptogênico.

O estudo Crystal AF randomizou 441 pacientes com AVC criptogênico após investigação inicial, para implante de monitor de eventos ou rastreamento convencional. 511 Em 6 meses de seguimento, 8,9% dos pacientes com monitor implantado tiveram registro de FA com mais de 30s de duração. Em 12 meses, esse número chegou a 12%, enquanto, no grupo de seguimento convencional, foi detectado FA em 2% (p < 0,001).

O STROKE-AF trial incluiu 496 pacientes acima de 50 anos de idade e foi apresentado no International Stroke Conference (ISC) 2021 ( late-breaking abstract 6). Entre os pacientes com monitor implantado, episódios de FA com duração superior a 2 minutos ocorreram em 12% versus 1,8% (p < 0,001). Pacientes com monitor implantado receberam maior número de terapia de anticoagulação e menor recorrência de AVC. Esses dados, embora limitados, pois não são estudos controlados para avaliar a comparação entre duas estratégias de intervenção terapêutica, sugerem algum benefício na monitorização prolongada.

As recomendações atuais para implante de Looper estão listadas na Tabela 37 .

Tabela 37. – Indicações de monitor de eventos implantáveis ( loop recorder )’.

  Classe de recomendação Nivel de evidência
Síncope recorrente de origem indeterminada, sem indicação formal de MP ou CDI, após investigação clínica e laboratorial inconclusiva I A
Em pacientes com palpitações recorrentes de provável causa arrítmica, em que outros métodos diagnósticos não demonstraram correlação com os sintomas IIa B
AVC criptogênico para detecção de FA em pacientes com investigação não invasiva negativa ou inconclusiva IIa B
Suspeita de síncope reflexa recorrente, com episódios frequentes e severos IIa B
Epilepsia cujo tratamento é ineficaz IIb B
Quedas inexplicadas IIb B

AVC: acidente vascular cerebral; CDI: cardiovesor-desfibrilador implantável; FA: fibrilação atrial; MP: marca-passo.

6. Recomendações para Avaliação e Programação Eletrônica dos DCEI

6.1. Marca-passo Convencional

A programação eletrônica de MP deve seguir os seguintes princípios básicos: 502 , 512

  • Restituir ou preservar a frequência cardíaca basal de repouso e adaptá-la ao esforço, restringindo a estimulação artificial à condição para a qual o MP foi indicado, evitando sua operação em condições em que não há benefício comprovado.

  • Preservar a condução atrioventricular intrínseca, quando possível.

  • Aumentar a longevidade da bateria do gerador de pulsos, sem perda de segurança e com benefício clínico ao paciente.

  • Detectar arritmias e disfunções do sistema.

A abordagem de portadores de MP consiste em avaliação clínica e eletrônica. Além da pesquisa de antecedentes pessoais (incluindo medicamentos em uso) e familiares, sintomatologia e exame físico, a avaliação clínica deve incluir o ECG de 12 derivações, fundamental para avaliar funções de sensibilidade, captura e arritmias. O ecocardiograma, habitualmente realizado antes do implante, pode ser essencial durante o seguimento para monitorar o remodelamento do VE em virtude de possíveis efeitos deletérios da estimulação crônica do VD e da síndrome de MP.

A avaliação eletrônica é realizada por meio de telemetria e deve contemplar o gerador de pulsos, os cabos-eletrodos e a recuperação de informações armazenadas na memória do dispositivo, principalmente eventos arrítmicos e disfunções.

A interrogação do sistema permite avaliar a duração da bateria do gerador, integridade dos cabos-eletrodos e medidas dos limiares de estimulação e sensibilidade. A inibição temporária do MP confirma o ritmo intrínseco, fundamental para a melhor programação do sistema. Dados estatísticos relacionados a cada câmara cardíaca e os eventos arrítmicos devem ser acessados, bem como o registro dos eletrogramas intracavitários.

A escolha do modo de estimulação deve considerar o ritmo intrínseco do paciente: sinusal normal, FA, DNS e/ou BAV.

6.1.1. Doença do Nó Sinusal

A estimulação unicameral ventricular (VVI) foi amplamente utilizada inicialmente, independentemente do tipo de bradicardia, devido a simplicidade e segurança. Mais de um quarto dos pacientes em estimulação VVI, contudo, desenvolve síndrome do MP (estimulação ventricular que ocasiona condução atrial retrógrada, resultando em sintomas como dispneia, palpitações, tonturas e sinais de baixo débito cardíaco), com comprometimento significativo da qualidade de vida. Dessa forma, na DNS, a estimulação atrial permite condução AV e IV espontâneas, evitando perda do sincronismo atrioventricular, síndrome de MP e dissincronia IV secundária à estimulação do VD. 513 , 514

A estimulação atrial pode ser realizada em modo AAI ou DDD; neste último, pode-se preservar a condução intrínseca por meio de algoritmos específicos. O modo DDD tem mais complicações relacionadas a desposicionamento de cabos-eletrodos quando comparado ao VVI; por outro lado, dispositivos AAI apresentam o dobro de reoperações em relação ao modo DDD, muitas vezes, devido ao desenvolvimento de BAV (progressão da lesão). BAV em pacientes com DNS ocorre em 0,6% a 1,9% ao ano, resultando em necessidade de mudança do sistema para DDD. 515 , 516

O modo AAI está relacionado a menor ocorrência de FA e eventos tromboembólicos em comparação ao modo VVI em pacientes com DNS. Resultados similares são observados em modo DDD, que também se associa a menores taxas de FA e melhor qualidade de vida que o modo VVI. Tais benefícios, no entanto, não impactam em desfechos de mortalidade, IC ou morte cardiovascular. 33 - 35

O efeito deletério da estimulação artificial do VD pode resultar em IC e pior sobrevida como consequência de dissincronia induzida. Por isso, em MP bicameral, é fundamental a programação de algoritmos de preservação da condução AV intrínseca, que prolongam automaticamente o intervalo AV ou promovem a mudança do modo de estimulação para AAI (com back-up ventricular), a fim de evitar estimulação desnecessária de VD em pacientes com condução AV preservada. Dados iniciais com esses algoritmos apontam redução significativa do percentual de estimulação ventricular (99% para 9%, p < 0,001) e de redução de FA (40%). Pacientes com BAV de 1º grau associado a DNS podem perder esses benefícios quando o intervalo PR é muito prolongado. 517

No estudo DANPACE, que comparou 1.415 pacientes com DNS, o modo DDDR com IAV até 220ms foi associado a menor ocorrência de FA paroxística; IAV muito longo se associou a regurgitação mitral, aumento de pré-carga e FA, sugerindo que há um limite para o prolongamento do IAV. Dessa forma, habitualmente, não é recomendado programar IAV superior a 220ms.

Outra ferramenta importante na doença DNS é o acionamento do sensor de variação de frequência (R). Esses sensores têm por objetivo aumentar FC em situações de aumento da demanda metabólica, como o exercício físico. Três pequenos estudos demonstraram melhora na qualidade de vida e tolerância ao esforço com a ativação do sensor, cujos resultados não foram reproduzidos no estudo ADEPT. 507 , 518

A função automatic mode switch (AMS) consiste na reversão do modo DDD(R) para VVI(R), em caso de surgimento de FA. Apesar de não haver fortes evidência comprovando seu benefício, recomenda-se a programação, principalmente em pacientes com FA paroxística, para alívio de sintomas. 519

6.1.2. Bloqueio Atrioventricular

No BAV, a estimulação do VD, necessária, é usualmente realizada em modo DDD ou VVI. O modo DDD mantém sincronismo AV mas está relacionado a mais complicações (6,2% vs. 3,2%), especialmente desposicionamento, aumento de limiares e infecção. 520

Em estudos que comparam os modos DDD e VVI em pacientes com BAVT e DNS (PASE, CTOPP), o modo DDD não se associou a redução de mortalidade e internações cardiovasculares. O estudo CTOPP evidenciou redução de FA com o modo DDD (benefício maior em pacientes com DNS); entretanto, 26% dos casos em modo VVI apresentaram síndrome de MP, com necessidade de crossover . Esses pacientes apresentaram melhora significativa após reprogramação para modo DDD. 31 , 33 Em pacientes acima de 70 anos de idade, o modo DDD parece não ser superior ao VVI em pacientes com BAVT durante seguimento de 3 anos (inclusive síndrome de MP); dessa forma, esse modo é alternativa aceitável para idosos com baixa expectativa de vida e restrição a atividades físicas.

6.1.3. Fibrilação Atrial

Em FA permanente, quando não há perspectiva de reversão para ritmo sinusal, apenas a câmara ventricular necessita de estimulação. Neste caso, o modo VVI(R) é recomendado. O sensor de variação de frequência se associa a melhor capacidade funcional e qualidade de vida em pequenos estudos. 521 , 522

6.1.4. Síncope Neuromediada e Síndrome do Seio Carotídeo

Síncope neuromediada com resposta cardioinibitória caracteriza-se por períodos de bradicardia intermitente, necessitando de curtos períodos de estimulação artificial, com frequência básica elevada para compensar a súbita instabilidade que ocorre durante o evento. Nesses casos, a estimulação deve ser de curta duração, apenas durante os episódios sintomáticos (função histerese). Os modos utilizados podem ser DDI, DVI ou DDD com algoritmo para preservação de condução intrínseca. O modo VVI esteve mais associado à ocorrência de síncope e pré-síncope que a estimulação bicameral (DDD e DVI) em alguns estudos. 72 , 73

Algoritmos como Rate Drop Response (RDR)® e Sudden Bradi Response (SBR)® identificam reduções abruptas da frequência cardíaca, instituindo frequência de intervenção acelerada a intervalos programáveis. Tais algoritmos são eficazes na redução de sintomas em pacientes com síncope neuromediada (cardioinibitória), em comparação a tratamento convencional sem MP. Embora não tenham sido testados contra outros modos de estimulação, tais algoritmos são eficazes e possibilitam que o MP seja programado para manter-se inibido a maior parte do tempo. No estudo ISSUE III, a estimulação DDD + RDR reduziu em 57% a chance de recorrência da síncope. O RDR foi programado para intervir quando a FC chegasse a 40bpm ou apresentasse queda de 20 batimentos em relação à FC basal (90bpm por 1 minuto). 523

O algoritmo Closed Loop Selection (CLS)® utiliza a bioimpedância intramiocárdica para avaliar a variação da contratilidade miocárdica para prever o início da síncope e instituir a intervenção (o aumento da contratilidade miocárdica ocorre na fase inicial dos episódios de síncope). 520 , 524

A função de busca automática de limiar de captura tem se mostrado segura, podendo prolongar a longevidade do gerador de pulsos em 60%, reduzindo custos em 42% em 10 anos. Por isso, salvo exceções, deve ser rotineiramente habilitado 525 ( Tabela 38 ).

Tabela 38. – Recomendações de programação eletrônica de marca-passo (MP) convencional.
  Classe de recomendação Nível de evidência
Modo DDD, com IAV máximo 220ms com algoritmos de busca da condução intrínseca em DNS I A
Modo DDD em BAVT, para evitar síndrome de MP I B
Modo VVIR em FA permanente I C
Automatic mode-switch deve ser rotineiramente programado I C
Sensor de variação de frequência em pacientes com incompetência cronotrópica IIa B
Busca automática de limiar de estimulação deve ser rotineiramente programada para prolongar sobrevida do gerador IIa B
Algoritmos específicos, histerese e IAV prolongado para evitar estimulação desnecessária em SN e HSC IIa B

BAVT: bloqueio atrioventricular total; DNS: doença do nó sinusal; FA: fibrilação atrial; HSC: hipersensibilidade do seio carotídeo; IAV: intervalo atrioventricular; MP: marca-passo; SN: síncope neuromediada.

6.2. Terapia de Ressincronização Cardíaca

A avaliação de portadores de TRC deve seguir os princípios de um MP convencional, acrescentando-se a abordagem de parâmetros específicos à correção da dissincronia. Dessa forma, o ECG periódico permite avaliar se a estimulação biventricular está ativa, e o ecocardiograma, realizado após 90 dias do implante e repetido ao longo do seguimento, deve documentar o remodelamento reverso nos respondedores.

O padrão de ativação ventricular dependerá do posicionamento dos cabos-eletrodos, assim como da precocidade de ativação de cada câmara. A ativação do VE resulta em eixo à direita, com padrão qR ou Qr em D1 e ondas r ou R em V1 (ver Figura 4 ). Na estimulação biventricular, ocorre fusão da estimulação de ambos os ventrículos. O padrão qR ou Qr em D1 é visualizado em 90% das estimulações biventriculares. A perda da onda q ou Q em D1 é altamente sugestiva de perda de captura do VE. Ondas r ou R em V1 estão presentes em 65% a 93% das estimulações biventriculares. 526 , 527

Figura 4. – Padrão QS em D1 e r inicial em V1 comprovando ativação precoce da parede lateral do VE em estimulação biventricular.

Figura 4

A perda da ressincronização pode ocorrer ao esforço, na presença de FA de alta resposta e devido ao encurtamento do intervalo AV; Holter -24h e teste ergométrico podem ser úteis para identificar tais cenários. Aumento do limiar de captura do VE é responsável por 10% dos casos de perda da ressincronização. 517 Essa alteração deve ser suspeitada pela perda do padrão Qr ou qR em D1 ao ECG12 e confirmada com o teste do limiar de estimulação. Os sistemas atuais dispõem da função de busca automática do limiar de captura que ajudam na identificação de variações dos limiares não detectados em avaliação convencional.

A programação dos parâmetros básicos deve seguir os princípios dos protocolos dos grandes estudos de TRC: DDD 50bpm, IAV sentido 100 a 120ms, intervalo VV 0ms (estimulação biventricular simultânea), com sensor de variação de frequência desligado. Pacientes com FA devem ser programados em DDI na presença de eletrodo atrial, VVI se não houver eletrodo atrial, e DDD se houver FA paroxística. Nesses casos, a frequência sugerida é de 60bpm na ausência de incompetência cronotrópica.

O modo de estimulação utilizado na maioria dos grandes estudos foi DDD ou VDD 35 a 60bpm, com objetivo de reduzir estimulação atrial, o que poderia comprometer o sincronismo atrioventricular em casos de condução intra-atrial retardada, prejudicando o enchimento ventricular. Diretrizes de IC recomendam o uso de fármacos para redução da FC quando está acima de 70bpm apesar de betabloqueadores, corroborando a programação de frequências básicas baixas e sensores de frequência desligados rotineiramente. 528

Intervalo AV curto, 100 a 120ms, visa atingir estimulação biventricular próximo a 100%, evitando perda de captura biventricular associada a encurtamento do intervalo PR. O ajuste dos intervalos AV e VV pelo ecocardiograma, ou por outros métodos, é habitualmente reservado aos não respondedores, uma vez que falta consenso quanto à real utilidade desses métodos quando aplicados rotineiramente. 529

A frequência máxima de sincronismo atrioventricular deve ser programada ao máximo, considerando-se a FC máxima predita para idade e eventuais limitações relacionadas à cardiopatia de base.

Pacientes com estimulação acima de 93% têm redução de 44% nas taxas de mortalidade e internação por IC (desfechos combinados), sendo os melhores resultados alcançados acima de 98%. Batimentos de fusão e pseudofusão podem superestimar o percentual de estimulação biventricular. Em pacientes com FA, quando o controle da FC não é alcançado com o tratamento clínico otimizado, a ablação do nó AV deve ser realizada, uma vez que se associa à redução de mortalidade. 206 A ablação da FA paroxística e persistente (isolamento das veias pulmonares) deve ser considerada em pacientes com IC. 520 A análise detalhada das situações clínicas que favorecem a indicação de ablação foge do escopo desta diretriz.

Extrassístoles ventriculares estão relacionadas à redução na taxa de estimulação biventricular e redução do remodelamento reverso, mesmo com incidência relativamente baixa. Fármacos antiarrítmicos e, eventualmente, ablação devem ser considerados em pacientes não respondedores. 530

Intervalos atrioventriculares curtos resultam em sístole ventricular precoce, não permitindo que a fase de contração atrial ocorra integralmente (onda A truncada). Nesses casos é necessário prolongar o intervalo AV até que a onda A fique evidente. Intervalos AV prolongados, inversamente, geram fusão das ondas E e A. Nesses casos, é necessário encurtamento do intervalo AV.

Dois métodos são habitualmente recomendados para o ajuste do intervalo AV: o método interativo e o método de Ritter. No método interativo, programa-se intervalo AV longo (200ms) e reduz-se gradativamente (20ms por vez) até 60ms analisando o fluxo mitral. O menor intervalo AV capaz de manter as ondas E e A sepradas (sem fusão), sem deformação da onda A e mantendo 40ms de distância do final da onda A ao início do QRS é o intervalo AV ótimo. O método de Ritter é realizado aferindo o intervalo QA (início do QRS ao final da onda A) em dois intervalos AV diferentes, um curto (60ms) e um longo (200ms). O intervalo AV ideal é calculado pela fórmula IAV = IAV longo – (QA[IAVcurto] – QA[IAV longo]). Assim, recomenda-se realizar ECO com avaliação do fluxo mitral após implante para analisar o sincronismo atrioventricular: se as ondas E e A se apresentarem separadas e o intervalo do final da onda A acima de 40ms, não há necessidade de ajuste de intervalo AV.

O intervalo VV pode ser programado empiricamente, ou ajustado também por ECO, ECG e algoritmos específicos. Com ECO, o ajuste é realizado testando-se diversos intervalos e avaliando a dissincronia. O intervalo que resultar em menor dissincronia deve ser o intervalo VV programado. Deve-se testar estimulação simultânea, estimulação precoce no VE e diferentes intervalos VV, 60, 40, 20ms. Posteriormente, deve-se testar os mesmos intervalos com estimulação precoce do VD. Os métodos mais utilizados são o modo M, com ou sem doppler tissular, e a velocidade de encurtamento longitudinal do VE, aferida pelo Doppler tissular.

Assim como o ajuste do intervalo AV, a programação do intervalo VV guiado por ECO deve ser realizada em não respondedores e em condições específicas.

Alguns dispositivos dispõem de algoritmos automáticos de ajuste de IAV e IVV, cuja eficácia ainda é controversa. De qualquer forma, não parecem ser inferiores ao ajuste empírico ou guiado pelo ECO.

Correlação entre diminuição da duração do QRS com a estimulação biventricular e a taxa de respondedores foi evidenciada em estudos retrospectivos, dando suporte à hipótese de que ajuste de intervalo AV e VV visando a QRS mais curto pode aumentar a taxa de resposta à TRC. 531

A estimulação multiponto fundamenta-se no princípio de estimular as regiões com ativação mais tardia do VE, por meio de eletrodo quadripolar, especialmente áreas basais e apicais, permitindo ativação de maior massa ventricular de forma mais rápida e homogênea. O estudo MPP trial comparou a estimulação multipontos com eletrodo quadripolar versus estimulação convencional, com resultados semelhantes entre as duas formas de programação, significância estatística para não inferioridade. No entanto, entre os pacientes com estimulação multiponto com 30mm de distância entre os dois pontos de estimulação de VE e com o menor intervalo (5ms), houve menor taxa de não respondedores. Esses resultados foram reproduzidos na primeira fase do estudo MORE-CRT e, mais uma vez, os pacientes com programação de 30mm de distância entre os pontos de estimulação do VE associado ao menor intervalo intra e interventricular apresentaram melhores resultados.

6.3. Cardioversor-desfibrilador Implantável

A programação do CDI deve ser voltada a quatro princípios básicos: 1) reduzir a mortalidade por meio da terapia efetiva na reversão de arritmias ventriculares potencialmente fatais; 2) priorizar a reversão de arritmias ventriculares por meio das terapias antitaquicardia sem choque (ATP), sempre que possível; 3) evitar choques inapropriados; e 4) reduzir ao máximo o percentual de estimulação artificial do VD (terapia antibradicardia).

As terapias apropriadas para reversão de FV e TV sustentadas são os pilares da redução de mortalidade na intervenção com CDI. Para isso, devem ser programadas terapias escalonadas em diferentes zonas de frequência, classificadas como TV (1 ou 2 zonas) e FV. As terapias programáveis incluem choques (até 35 ou 40J) e terapias com estimulação artificial de 3 a 20 pulsos com frequência superior à da taquicardia ( antitachycardia pacing [ATP]) que podem reverter TV monomórficas sem a aplicação de choques, de modo indolor e com redução do dano miocárdico eventualmente provocado pelos choques.

A efetividade dos choques na reversão de arritmias ventriculares costumava ser testada no intraoperatório, por meio da realização de limiar de desfibrilação (indução de FV, seguida de disparo de choque para reversão da circulação efetiva, com energia pelo menos 10J abaixo na energia máxima programável). Estudos subsequentes comprovaram ser desnecessária essa estratégia, uma vez que medidas intraoperatórias normais (limiar de estimulação, impedância e onda R) se correlacionam com efetividade adequada na reversão de arritmias de ocorrência espontânea. 532 Assim, é possível evitar eventuais intercorrências decorrentes da indução de FV no intraoperatório e dano miocárdico provocado pelos choques. A programação da zona de FV, portanto, deve incluir choques de energia máxima possível, com inversão da polaridade entre um choque e o outro. Nas zonas de TV, podem ser programados choques de energia mais baixa, que habitualmente são precedidos por tentativas de ATP. 533

É bem conhecida a eficácia do ATP como primeira linha nas arritmias ventriculares. Taquicardias ventriculares monomórficas, organizadas, com ciclo estável e especialmente sem repercussão hemodinâmica, podem ser facilmente revertidas com ATP, incluindo pulsos com intervalo fixo ( Burst ) ou com aceleração entre os pulsos ( Ramp ).

Algumas arritmias instáveis, mesmo em zonas de alta frequência (na faixa de FV), podem ser interrompidas com ATP antes do disparo de choques programados naquela zona. Nesse caso, uma tentativa de ATP durante ou antes do carregamento da energia de choque é programada; em caso de reversão da arritmia, o choque é abortado. O estudo PainFREE II utilizou o ATP como primeira linha de tratamento em zona de 188 a 250bpm, com redução significativa de 71% do risco relativo de choque, sem comprometimento da segurança dos pacientes. 534

A programação adequada de detecção e terapia escalonada é capaz de reduzir choques inapropriados, proporcionar mais terapia apropriada com ATP e reduzir a mortalidade. 535 Para isso, os princípios básicos de programação devem incluir:

  1. Zona de FV programada com frequência acima de 233bpm (nos dispositivos Medtronic, acima de 188), com pelo menos 30 batimentos em 40 (x em y) para detecção. Essa estratégia permite evitar choques em arritmias não sustentadas e choques inapropriados em situações de ruídos intermitentes, dupla contagem ou extrassistolia isolada.

  2. Em pacientes de prevenção primária, uma única zona de detecção de FV pode ser suficiente. Zonas de monitoramento de TV sem terapias (monitor) podem ser programadas a critério médico. Nos casos de prevenção secundária, terapias focadas para TV devem ser programadas com corte de detecção de 10-20bpm menores que a frequência da taquicardia documentada. De acordo com o critério clínico, podem ser programada zonas de terapia com frequência mais baixa, na dependência do risco de TV mais lenta, sempre priorizando o ATP. 536

  3. Algoritmos de monitoramento de ruídos e danos dos cabos-eletrodos devem ser programados, assim como recursos de autoajuste e supressão de oversensing, como a detecção de onda T.

  4. Programação adequada dos algoritmos de discriminação de arritmias supraventriculares, especificamente no critério de avaliação da morfologia nos dispositivos unicamerais e na avaliação de algoritmos baseados na relação atrioventricular nos dispositivos bicamerais. É interessante que os limitadores de tempo, como SRD ( Boston Scientific ) e Timeout , sejam desabilitados, pois esses recursos ignoram a discriminação de um evento classificado como TSV após o período preestabelecido e liberam a terapia que seria inapropriada.

Finalmente, é preciso avaliar cuidadosamente a necessidade concomitante de estimulação cardíaca artificial para bradiarritmia. A maioria dos portadores de CDI não requer terapia antibradicardia, especialmente nos casos de prevenção primária. Contudo, é sabido que a estimulação convencional do VD aumenta o risco de disfunção ventricular e a mortalidade. A programação deve, sempre que possível, priorizar a diminuição do percentual de estimulação artificial do VD. Para isso, em CDI unicameral, deve-se programar em modo VVI com 40ppm e, nos bicamerais, deve-se priorizar a estimulação atrial isolada através de algoritmos de minimização de estimulação ventricular (p. ex., RYTHMIQ, MVP, IRS plus) ou da programação de intervalo AV longo o suficiente para evitar a estimulação ventricular desnecessária. Em pacientes com necessidade de estimulação ventricular por bloqueio da condução AV, deve-se considerar, de acordo com a função ventricular, a possibilidade de estimulação de sítios alternativos como a estimulação biventricular (TRC) ou do sistema excito-condutor (His/ramo esquerdo). 537

6.4. Monitor de Eventos Implantável ( Loop Recorder )

O monitor de eventos implantável deve ser programado de maneira adequada a detectar a atividade elétrica ventricular, sem perdas de sensibilidade sinal ( undersensing ) e sem oversensing de ruídos que prejudiquem a identificação do ritmo.

A capacidade de detecção automática, independentemernte do acionamento do paciente, a memória com capacidade de armazenamento de eletrogramas e a disponibilidade de bateria com autonomia de até 3 a 4 anos tornam essa ferramenta muito útil para o esclarecimento de eventos arrítmicos não registrados em exames usuais.

A técnica de implante é semelhante entre os modelos atualmente disponíveis, sendo necessário confirmar, no intraoperatório, a adequação do sinal elétrico na posição escolhida para o posicionamento do dispositivo. Após a confirmação da captação adequada de sinal, é necessário ajustar a programação para detecção de arritmias de forma individualizada. 538 Alguns fabricantes sugerem programação inicial empírica, conforme:

  1. Pausas: 3s

  2. Bradicardia: FC ≤ 30bpm por mais de 4 batimentos consecutivos

  3. Taquicardia: FC acima da FC máxima predita para a idade (220- idade) por 16 ou mais batimentos;

  4. Fibrilação atrial: > 2 minutos classificado como ritmo de FA.

É necessária a adjudicação cuidadosa dos episódios registrados no monitor, uma vez que falsas detecções podem estar presentes na memória. Episódios classificados como FA, por exemplo, podem ser mal classificados em virtude de variações de intervalo RR por extrassistolia ventricular ou undersensing intermitente do QRS. Em estudo recente, a análise de 695 transmissões espontâneas ou agendadas, Afzal et al. encontraram até 81% de falsos eventos. 539 A programação adequada do critério de detecção e a análise posterior dos traçados é fundamental na otimização do monitoramento.

6.5. Monitoramento Remoto (Via Web )

O monitoramento remoto via web é uma realidade no seguimento dos portadores de DCEI. A transmissão de dados é possível via conexão do dispositivo a uma banda larga de internet ou, em dispositivos mais recentes, via bluetooth conectado a um smartphone. Por meio dessa tecnologia, é possível acessar diversos parâmetros de programação como frequência e modo de estimulação, energia de estimulação, parâmetros de detecção e sensibilidade, assim como registros de diagnósticos e status da bateria.

A transmissão de dados precisa ser ajustada, pois algumas informações serão transmitidas de forma ativa após acionamento de algum alarme, ou de forma passiva através do acionamento pelo paciente ou, ainda, de forma programada, mediante cronograma de transmissões. O acesso remoto às informações é disponibilizado ao serviço que acompanha o paciente mediante acesso privado ao servidor do sistema e resguardadas as garantias de privacidade de dados do paciente.

7. Recomendações para Prevenção e Tratamento de Infecções e Explante de DCEI

7.1. Prevenção e Tratamento de Infecções

Publicações recentes têm mostrado aumento da incidência de processos infecciosos relacionados a DCEI. Fatores demográficos e clínicos, como o envelhecimento populacional e comorbidades, podem influenciar tanto as contaminações hematogênicas quanto aquelas diretamente relacionadas ao implante e troca dos dispositivos. Um levantamento recente conduzido pela EHRA demonstrou que infecções relacionadas a DCEI são mais frequentes após reoperações, inclusive trocas isoladas de gerador de pulsos. 540

Os consensos mais recentes têm chamado a atenção para a necessidade de padronização de condutas e para a formação de times de especialistas para a abordagem desse tipo particular e pouco frequente de infecção, com a finalidade de mitigar as controvérsias entre especialistas, ainda frequentes. 541 - 546

Processos infecciosos que envolvem DCEI manifestam-se de duas formas principais: envolvimento da loja do gerador de pulsos ou exclusivamente intravascular. O acometimento exclusivo da loja é mais frequente, ocorrendo em aproximadamente 60% dos casos (geralmente por contaminação durante a cirurgia ou manipulação subsequente). A erosão da pele tardiamente pode ocorrer devido a/ou resultar em infecção da loja; em ambos os casos, a infecção pode progredir para infecção sistêmica. O acometimento da loja associado à infecção intravascular ocorre em aproximadamente 20% das infecções e habitualmente é secundário à demora ou a condutas inadequadas. O acometimento intravascular exclusivo também ocorre em cerca de 20% dos casos, por contaminação sanguínea na maior parte das vezes. 547 Essa contaminação pode ocorrer durante bacteremia causada por foco infeccioso distante, como tromboflebite séptica, osteomielite, pneumonia, infecção do sítio cirúrgico, cateteres vasculares contaminados ou infecção bacteriana originada da pele, boca, trato gastrointestinal ou urinário.

Um consenso de especialistas, encabeçado pela EHRA e endossado por outras sociedades internacionais, teve como objetivo principal definir a terminologia que deve ser utilizada em registros e estudos clínicos para a abordagem terapêutica das infecções e remoção de DCEI. 548 Na Tabela 39 , encontram-se as adaptações para a língua portuguesa da terminologia recomendada para a diversas apresentações clínicas.

Tabela 39. – Tipos de infecção relacionadas ao DCEI.

Tipos de infecção relacionados ao DCEI
  Cenários clínicos Definição
Local Infecção incisional superficial Envolve apenas pele e tecido subcutâneo sem acometer o DCEI
Infecção isolada da loja Presença de sinais clínicos de inflamação restritos à loja do gerador de pulsos (eritema, calor, flutuação, deiscência da ferida, fragilidade tecidual ou drenagem purulenta) com hemoculturas negativas
Erosão de loja Extrusão completa ou parcial de gerador de pulsos ou de cabo-eletrodo através da pele
Sistêmica Bacteremia Hemoculturas positivas associadas ou não a sintomas ou sinais de infecção sistêmica
Infecção ou erosão de loja com bacteremia associada Infecção ou erosão de loja de gerador com hemoculturas positivas, na ausência de vegetação valvar ou aderida a cabo-eletrodo
Endocardite relacionada ao DCEI sem infecção de loja Bacteremia e vegetação valvar ou aderida a cabo-eletrodo, na ausência de infecção de loja
Endocardite relacionada ao DCEI com infecção de loja Bacteremia e vegetação valvar ou aderida a cabo-eletrodo, na presença de infecção de loja
Bacteremia oculta com provável infecção relacionada ao DCEI Bacteremia em portador de DCEI, quando excluídas outras causas infecciosas

DCEI: dispositivo cardíaco eletrônico implantável.

O diagnóstico definitivo de infecção relacionada a DCEI baseia-se em três achados principais: 1) presença de coleção purulenta ou exteriorização do DCEI ao exame clínico; 2) crescimento de microrganismos em hemoculturas e 3) presença de vegetações na valva tricúspide ou em cabos-eletrodos evidenciada pelo ecocardiograma transesofágico (ETE). Quando não se consegue definir o diagnóstico da infecção relacionada ao DCEI com esses critérios, outros exames complementares (como PET-CT) podem ser necessários. Os critérios modificados da Duke University para diagnóstico de infecção de DCEI estão elencados nas Tabelas 40 e 41.

A comprovação de que o DCEI está definitivamente contaminado é fundamental para o tratamento adequado do paciente, uma vez que, comprovada a contaminação, sua remoção completa será fundamental para o sucesso do tratamento. Por outro lado, se não houver o DCEI, estiver livre de contaminação e o processo infeccioso estiver relacionado a outro foco, a remoção desnecessária do dispositivo implicará custo desnecessário e risco cirúrgico relacionado à extração dos cabos-eletrodos. O fluxograma para diagnóstico e tratamento de infecções de DCEI está representado na Figura 5 .

Figura 5. – Fluxograma para diagnóstico e do tratamento de infecção de DCEI. DCEI: dispositivo cardíaco eletrônico implantável; ECO: ecocardiograma; ETT: ecocardiograma transtorácico; ETE: ecocardiograma transesofágico; FDG: fluordeoxiglicose; PET/CT: tomografia computadorizada por emissão de pósitrons; SPECT/CT WBC: tomografia computadorizada por emissão de fóton único (single photon emission computed tomography) com leucócitos marcados.

Figura 5

Exames de imagem são importantes tanto para o diagnóstico quanto para a condução do tratamento. Nesse sentido, algumas informações obtidas por imagem podem ser relevantes: 1) identificação do tipo de DCEI; 2) identificação de cabos-eletrodos abandonados; 3) achado de vegetações intracardíacas e seu tamanho; 4) sinais sugestivos de embolização séptica para os pulmões.

No caso de febre em portador de DCEI em que não se consegue definir infecção pela avaliação da loja do gerador, hemoculturas ou ecocardiografia transesofágica e exames de imagem baseados na captação de radiofármacos podem ser importantes.

Embora a remoção completa do gerador de pulsos e de todos os cabos-eletrodos seja essencial, o tratamento da infecção deve ser feito, fundamentalmente, com o uso de antimicrobianos. A escolha do antibiótico deve ser estabelecida a partir das culturas de sangue, de fragmentos da loja e dos cabos-eletrodos removidos. Quando não é possível a definição do microrganismo, o uso empírico de antibióticos deve ser definido por critérios clínicos. Da mesma forma, o tempo de tratamento também deve ser definido em função do quadro clínico, sempre contado a partir da remoção completa do DCEI ( Figura 6 ).

Figura 6. – Duração do tratamento e necessidade de remoção do sistema nas diferentes apresentações de infecção relacionada a DCEI. DCEI: dispositivo cardíaco eletrônico implantável.

Figura 6

A remoção completa do DCEI é fundamental para evitar a recorrência da infecção. A extração dos cabos-eletrodos, entretanto, raramente deve ser considerada emergência, mesmo em choque séptico. À exceção dos implantes recentes que costuma ser tecnicamente mais fácil, a extração somente deverá ser realizada quando o paciente estiver com boas condições hemodinâmicas e com o quadro infeccioso estabilizado, face aos riscos associados ao procedimento (aderências às veias e ao coração).

A técnica de extração dos cabos-eletrodos deve ser preferencialmente transvenosa, exceto quando os cabos-eletrodos são epicárdicos ou quando houver vegetação intracavitária maior que 2,5cm em seu maior diâmetro. As recomendações para remoção do gerador de pulsos e dos cabos-eletrodos estão listadas na Tabela 42 .

Tabela 42. – Recomendações para a remoção de DCEI.

  Classe de recomendação Nível de evidência
Remoção completa do sistema em pacientes com infecção relacionada ao DCEI (sistêmica ou local) confirmada I B
Remoção completa do DCEI nos casos de bacteremia por S. aureus, CoNS, Cutibacterium spp. ou fungemia por Candida spp., afastados outros sítios de infecção I C
Remoção completa do DCEI nos casos de bacteremia por Pneumococcus spp. ou bactéria gram-negativa que não Pseudomonas ou Serratia , quando houver bacteremia recorrente/persistente, apesar da antibioticoterapia apropriada, na ausência de outra fonte identificável da recorrência ou da persistência da infecção I C
Remoção completa do DCEI em pacientes com endocardite infecciosa, com ou sem envolvimento definitivo do sistema DCEI IIa C
Remoção completa do DCEI nos casos de bacteremia por Streptococcus sp. alfa ou beta-hemolítico ou Enterococcus spp. como tratamento de primeira escolha e, no caso de bacteremia recorrente/persistente, apesar da antibioticoterapia apropriada, como tratamento de segunda escolha IIb C

DCEI: dispositivo cardíaco eletrônico implantável.

Tabela 40. – Critérios modificados para o diagnóstico de infecções relacionadas a DCEI.

Infecção de loja do gerador confirmada Presença de um dos seguintes achados: 1) edema, eritema, calor, dor e secreção purulenta; 2) formação de fístula; 3) deformação, aderência e erosão iminente da pele; 4) exposição do gerador ou de cabo-eletrodo
Endocardite relacionada ao DCEI confirmada Presença de: 1) dois critérios maiores; 2) um critério maior e três menores
Endocardite relacionada ao DCEI provável Presença de: 1) um critério maior e um menor; 2) TRÊS critérios menores
Endocardite relacionada ao DCEI descartada Ausência dos critérios anteriormente mencionados

DCEI: dispositivo cardíaco eletrônico implantável.

Tabela 41. – Critérios maiores e menores para diagnóstico de infecção relacionada a DCEI.

CRITÉRIOS MAIORES
Microbiologia A. Hemoculturas positivas para microrganismos típicos encontrados na infecção relacionada ao DCEI e/ou EI (estafilococos coagulase-negativos, S. aureus )
B. Microrganismos compatíveis com o EI em duas hemoculturas separadas:
a. Streptococus Viridans, Streptococcus gallolyticus ( S. bovis ), grupo HACEK, S. aureus
b. Enterococos adquiridos na comunidade, na ausência de um foco primário
C. Microrganismos compatíveis com o EI de hemoculturas persistentemente positivas:
a. ≥2 hemoculturas positivas de amostras de sangue colhidas >12 h
b. Três amostras ou a maioria de ≥ 4 culturas de sangue separadas (primeira e última amostras coletadas com ≥ 1 hora de intervalo)
c. Hemocultura positiva única para Coxiella burnetii ou título de anticorpo IgG de fase I >1: 800
Imagem positiva para infecção relacionada ao DCEI e/ou EI A. Ecocardiograma (incluindo intracardíaco) positivo para:
a. Infecção relacionada ao DCEI:
I. Infecção clínica do gerador/loja
II. Vegetação em cabo-eletrodo
b. EI em válvula
I. Vegetações
II. Abscesso, pseudoaneurisma, fístula intracardíaca
III. Perfuração valvar ou aneurisma
IV. Nova deiscência parcial de prótese valvar
B. FDG PET/CT (deve-se tomar cuidado em caso de implantes recentes) ou detecção radioativa de SPECT/CT WBC de atividade anormal no local da loja do gerador, ao longo dos condutores ou no local da válvula
C. Vazamento para-valvar definido por TC cardíaca
CRITÉRIOS MENORES
  a. Predisposição: como condições cardíacas predisponentes (p. ex., defeitos cardíacos estruturais preexistentes, insuficiência valvar tricúspide recente) ou uso de drogas injetáveis
b. Febre (temperatura > 38°C)
c. Fenômenos vasculares (incluindo os detectados apenas por imagem): embolia arterial importante, embolia pulmonar séptica, aneurisma infeccioso (micótico), hemorragia intracraniana, hemorragias conjuntivais e lesões de Janeway
d. Evidência microbiológica: hemocultura positiva que não atende a um critério principal, como observado anteriormente, ou evidência sorológica de infecção ativa por organismo compatível com EI ou cultura de loja de gerador ou cultura de cabo-eletrodo (extraída por loja não infectada).

EI: endocardite infecciosa; DCEI: dispositivo cardíaco eletrônico implantável; EIC: ecocardiograma com ultrassom intracardíaco; PET/CT: tomografia computadorizada por emissão de pósitrons; SPECT/CT WBC: tomografia computadorizada por emissão de fóton único (single photon emission computed tomography) com leucócitos marcados; lesões de Janeway: lesões hemorrágicas, indolores, embolização séptica.

O implante do novo DCEI deve ser realizado somente após remissão completa do processo infeccioso, e deve ser definido em função do quadro clínico. Até que o quadro infeccioso seja totalmente debelado, pacientes dependentes de estimulação artificial devem ser mantidos com MP temporário. Pacientes não dependentes devem permanecer sob monitoramento do ritmo cardíaco, até que o implante seja realizado. Em alguns casos, o implante do novo DCEI pode não ser necessário devido à alteração no padrão da doença ou por mudança de conduta. Por isso, a reavaliação da necessidade do DCEI é sempre fundamental. As recomendações relacionadas ao implante do novo DCEI estão listadas na Tabela 43 .

Tabela 43. – Recomendações para implante de novo DCEI, após extração.

  Classe de recomendação Nível de evidência
Reavaliação da indicação para o reimplante após a extração do dispositivo I C
O novo dispositivo deve ser implantado, preferencialmente, em outro sítio, podendo ser contralateral, femoral ou epicárdico I C
O reimplante deve ser evitado ou postergado, sempre que possível, até que os sintomas e sinais de infecção sistêmica e local estejam resolvidos IIa C
Cabo-eletrodo com mecanismo de fixação ativa para estimulação cardíaca temporária, ipsilateral ao sistema removido, em pacientes dependentes de marca-passo enquanto aguardam o reimplante IIa C

DCEI: dispositivo cardíaco eletrônico implantável.

Vários fatores de risco para o desenvolvimento de infecção relacionada ao DCEI têm sido detectados. Esses fatores podem estar relacionados ao próprio indivíduo, a procedimentos médicos realizados ou ao próprio DCEI. Os principais fatores para infeção de DCEI estão listados na Tabela 44 .

Tabela 44. – Fatores de risco que predispõem à infecção cardíaca relacionada a DCEI.

Fatores relacionados ao paciente
Doença renal terminal (DRT)a
História de infecção de dispositivo
Febre antes do implante
Uso de corticosteroides
Insuficiência renalb
Doença pulmonar obstrutiva crônica (DPOC)
Classe NYHA ≥ 2
Doenças de pele
Malignidade
Diabetes melito
Ponte de heparina
Insuficiência cardíaca (IC)
Anticoagulantes orais
Fatores relacionados ao procedimento
Duração do procedimento
Hematoma
Reposicionamento de cabos
Operador inexperientec
Estimulação temporária
Troca do gerador/revisão/ upgrade
Troca do gerador
Antibioticoprofilaxiad
Fatores relacionados ao dispositivo
Cabo(s) eletrodo(s) epicárdico(s)
Loja abdominal
Presença ≥ 2 cabos-eletrodos
Dispositivo dupla-câmara

DPOC: doença pulmonar obstrutiva crônica; DPOC: doença pulmonar obstrutiva crônica; DRT: doença renal terminal; IC: insuficiência cardíaca; NA: não disponível; NYHA: New York Heart Association. Parâmetros de risco que foram estatisticamente significativos para dados retrospectivos e prospectivos são mostrados. Análises restritas a dados prospectivos apenas para os mesmos parâmetros (se disponível) também são mostradas. Adaptada de Polyzos et al.549 aTaxa de filtração glomerular (TFG) < 60 mL/min ou depuração da creatinina (CrCL) < 60 mL/min. bGFR ≤ 15 mL/min ou hemodiálise ou diálise peritoneal. c < 100 procedimentos anteriores. dA estimativa do efeito combinado dos estudos randomizados foi de 0,26 [0,13, 0,52].

Cuidados preventivos são fundamentais para a redução da ocorrência de infecções relacionadas a procedimentos médicos nesses indivíduos. Na Tabela 45 e na Figura 7 , estão listadas as principais medidas preventivas recomendadas,

Tabela 45. – Lista de medidas preventivas recomendadas para infecção relacionada ao DCEI.

  Classe de recomendação Nível de evidência
Medidas pré-operatórias    
Postergar o implante de DCEI em pacientes com infecção I C
Evitar estimulação transvenosa temporária e acesso venoso central sempre que possível. Caso utilizados, idealmente, devem ser removidos antes da introdução de nova prótese I A
Medidas para evitar hematoma de loja: interromper sempre que possível o uso de antiplaquetários e, diante de anticoagulantes orais, evitar a " ponte" com heparina, interrompendo o uso durante o implante, caso possivel I B
A profilaxia antibiótica é recomendada no espaço de tempo de 30 minutos a 1 hora antes da incisão cirúrgica para cefazolina e de 90 a 120 minutos para vancomicina I A
Medidas perioperatórias
Uso de envelope antibiótico em situações de alto risco de infecção* IIa B
Instilar antisséptico e/ou antibióticos em loja de gerador IIb C
Medidas pós-operatórias
Uso de antibioticoterapia pós-operatória IIb B
Drenagem ou evacuação do hematoma (exceto na presença de tensão tecidual, deiscência da ferida ou dor intensa) III B

* Conforme definido pelo estudo WRAP-IT:550 pacientes submetidos a revisão de loja ou cabo-eletrodo, troca de gerador, mudança de modo de estimulação, implante inicial de CRT-D e os com fatores de risco listados na Tabela 44. DCEI: dispositivo cardíaco eletrônico implantável.

Figura 7. – Medidas preventivas para redução da infecção relacionada a DCEI considerando-se os fatores de risco, modificáveis ou não, já determinados na literatura. ACO: anticoagulação oral; CDI: cardioversor-desfibriador implantável; DCEI: dispositivo cardíaco eletrônico implantável; DPOC: doença pulmonar obstrutiva crônica; NYHA: New York Heart Association; TRC: terapia de ressincronização cardíaca.

Figura 7

Tabela 46. – Definições para abordagens e técnicas de extração.

Técnica Definição
Transvenosa Extração de cabo-eletrodo por via intravascular (percutânea) realizada através de uma veia central (subclávia, jugular ou femoral)
  • - Abordagem pelo local de entrada do cabo-eletrodo na veia

Extração de cabo-eletrodo, por tração ou contratração, utilizando-se a mesma veia e local por onde o cabo-eletrodo foi introduzido
  • - Abordagem transjugular ou transfemoral

Extração com ferramenta de captura intravascular introduzida pelas veias jugular ou femoral, direita ou esquerda
Transtorácica Extração de cabo-eletrodo com abertura da cavidade torácica por esternotomia mediana, toracotomia lateral, incisão subxifoide, inclusive por abordagem híbrida, com ou sem circulação extracorpórea

Idosos, crianças e adultos com cardiopatias congênitas representam subgrupos que merecem atenção especial quanto à possibilidade de infecção de DCEI. Loja submuscular em pacientes com pouco tecido subcutâneo para prevenção de erosão da pele é fundamental. Em pacientes pediátricos, especialmente com cardiopatia congênita, o operador deve ter experiência em abordagens cirúrgicas múltiplas e alternativas. CDI extravascular ou subcutâneo deve ser considerado em criança de menor idade, pacientes com cardiopatia congênita e com acesso venoso limitado ou inexistente. 551 , 552

Curioso destacar que a análise de registros retrospectivos 553 , 554 demonstra taxa de infecção de DCEI superior ao reportado em estudos prospectivos 537 , 555 - 558 (3,4% vs. 1,2%). Este fenômeno pode estar relacionado à maior adesão a procedimentos preventivos nos estudos clínicos em comparação ao que ocorre na prática clínica diária. Dentre os fatores operatórios mais relacionados ao risco de infecção, destacam-se o hematoma de loja, os procedimentos de longa duração e as reintervenções para reposicionamento de cabo-eletrodo. Especificamente sobre reoperações para troca de gerador de pulsos, correção de disfunção de cabos-eletrodos ou mudança de modo de estimulação, o adequado tratamento da loja do gerador de pulsos, seja pela remoção completa da carapaça fibrosa ou pelo uso de envelope antibacteriano, reduz a ocorrência de infecções.

A profilaxia antibiótica pré-operatória, com o uso de uma dose de cefalosporina de primeira geração (cefazolina), é fortemente recomendável, o que não ocorre quanto ao uso sistemático de antibiótico no período pós-operatório. 11

O intervalo de tempo entre o diagnóstico e o tratamento adequado da infecção relacionada a DCEI é fundamental. Dados da literatura mostram que se a remoção do dispositivo for feita até 3 dias após a hospitalização, tanto o tempo de internação quanto a mortalidade hospitalar são significativamente reduzidos. Nesse sentido, quando não existem dados suficientes para o estabelecimento do diagnóstico de infecção, baseada na tríade composta por sinais infecciosos na loja do gerador de pulsos, crescimento bacteriano nas hemoculturas e identificação de vegetações no ecocardiograma esofágico, outros recursos devem ser utilizados.

Recomenda-se a varredura por [18F] PET/CT FDG ou cintilografia WBC radiomarcada ou TC com contraste se houver suspeita de endocardite infecciosa relacionada ao DCEI, hemoculturas positivas e ecocardiografia negativa ou nos casos de bacteremia por S. aureus na presença de DCEI. Punção aspirativa e desbridamento cirúrgico em casos de infecção da loja do gerador, como tentativa para evitar a extração de cabo-eletrodo, devem ser fortemente desencorajados.

Para monitorar o número de casos de infecção relacionada a DCEI e o resultado das atitudes preventivas e terapêuticas, é fundamental a implementação de grandes registros, não voluntários, de fácil preenchimento e com grande qualidade. Cada serviço deve estabelecer rotinas para o diagnóstico preciso e tratamento em tempo adequado. A reavaliação constante do desempenho de cada centro é altamente recomendável.

7.2. Remoção de Cabos-eletrodos de Dispositivos Cardíacos Eletrônicos Implantáveis

A necessidade de remover cabos-eletrodos de DCEI cresceu nos últimos anos em função, principalmente, de dois fatores: 1) crescimento da taxa de infecções relacionadas a DCEI e 2) desenvolvimento dos MP e CDI multissítios, que utilizam maior número de cabos-eletrodos.

As indicações para a remoção de cabos-eletrodos podem ser: (a) obrigatórias, como no tratamento de infecções; (b) necessárias, para obtenção de acesso para novos cabos-eletrodos em pacientes com oclusões venosas; ou (c) opcionais, como em pacientes com acesso venoso adequado, quando são submetidos à substituição de cabo-eletrodo.

Como a maioria dos dispositivos necessita de acesso venoso para o implante dos cabos-eletrodos, as técnicas de extração transvenosa são as mais utilizadas. Na atualidade, a abertura do tórax para a remoção de cabos-eletrodos é muito pouco utilizada, sendo necessária, quase exclusivamente, para a retirada de cabos-eletrodos epicárdicos ou para a correção de complicações que ocorrem em extrações transvenosas.

O panorama atual da extração de cabos-eletrodos mostra indicações e técnicas operatórias bem estabelecidas. As ferramentas utilizadas estão bem desenvolvidas e os resultados dos diferentes procedimentos são bem conhecidos, com altas taxas de sucesso. Complicações catastróficas, entretanto, podem ocorrer durante procedimentos de extração. Tais complicações, embora raras, são potencialmente letais e costumam requerer cirurgia a céu aberto de emergência.

Neste item, são apresentadas as recomendações para extração em pacientes não infectados, uma vez que o manejo de infecções já foi abordado.

A remoção de cabos-eletrodos epicárdicos é feita, obrigatoriamente, pela reabertura da cavidade torácica, preferencialmente pelo mesmo acesso por onde o eletrodo foi implantado. A remoção de cabos-eletrodos transvenosos deve ser feita, preferencialmente, por acesso intravascular. Excepcionalmente, pode ser necessária uma abordagem transtorácica, com ou sem o auxílio de circulação extracorpórea, como na falha da extração transvenosa ou quando existem grandes vegetações aderidas aos cabos-eletrodos. A escolha do tipo de abordagem para a extração transvenosa de cabos-eletrodos depende, fundamentalmente, da possibilidade de se obter acesso ao cabo-eletrodo alvo da remoção. Infelizmente, muitos pacientes são portadores de cabos-eletrodos completamente intravasculares, por ter apresentado fratura espontânea do cabo ou por iatrogenia durante procedimentos de remoção.

Quando o cabo-eletrodo a ser removido está íntegro ou apresenta um segmento extravascular, por menor que esse segmento seja, devemos utilizar a abordagem pelo local de entrada do eletrodo na veia ( venous entry site approach ). Essa abordagem consiste na introdução na veia de uma bainha que é guiada pelo próprio cabo-eletrodo a ser removido. Essa bainha é utilizada para desfazer as aderências que se formam entre o cabo-eletrodo e o endotélio venoso ou o endocárdio. Quando o paciente tem mais que um cabo-eletrodo transvenoso implantado, frequentemente, existem aderências entre os cabos. Após se desfazerem todas as aderências e se atingir o local onde o eletrodo está fixado ao coração, essa mesma bainha é utilizada para se fazer pressão contra o músculo cardíaco enquanto se traciona o cabo-eletrodo (manobra de contratração). Existem várias ferramentas especificamente desenvolvidas para esse tipo de abordagem, a saber:

  • Guias de travamento ( locking stylets ), que são estiletes revestidos por uma fina malha de aço que se expande na luz do cabo-eletrodo conferindo a este a sustentação necessária para sua tração.

  • Bainhas para dissecção de aderência e contratração:

  • Bainhas mecânicas não energizadas ( non-powered sheaths ): conjuntos de tubos rígidos metálicos e tubos flexíveis de teflon ou polipropileno, utilizados para desfazer aderências por dissecção romba, com intensidade determinada pela força da mão do médico que faz o procedimento.

  • Bainhas mecânicas com lâminas rotacionais: seccionam as aderências e são ativadas por um gatilho na mão do operador ou por um motor elétrico.

  • Bainhas energizadas por raios LASER : seccionam as aderências por foto-ablação.

Quando o cabo-eletrodo que precisa ser removido não apresenta um segmento extravascular, torna-se obrigatória a sua captura intravascular. Existem ferramentas com o formato de laço ou de cesta, feitas com arames metálicos muito maleáveis, desenvolvidas para capturar esses fragmentos. Tais ferramentas são introduzidas habitualmente por punção das veias femoral ou jugular. Uma vez capturado, o cabo-eletrodo pode ser tracionado diretamente. Em casos específicos, pode ser necessário associar manobra de contratração após a captura do cabo-eletrodo. 559 As Tabelas 46 e 47 mostram as definições recomendadas para as ferramentas e para as abordagens usadas em extração de cabos-eletrodos.

O termo remoção tem sido utilizado, genericamente, para designar a retirada de cabos-eletrodos de DCEI, independentemente do tipo de abordagem. Pode ser realizada pela simples tração de um cabo-eletrodo transvenoso, sem a utilização de nenhuma ferramenta; por toracotomia, para a retirada de cabo-eletrodo epicárdico ou por toracotomia com o auxílio de circulação extracorpórea para a retirada de cabo-eletrodo transvenoso. O termo extração deve ser utilizado exclusivamente para os casos em que técnicas e ferramentas são utilizadas para: 1) dilatar o trajeto por onde o cabo-eletrodo passa no interior das veias; 2) desfazer aderências; 3) realizar manobra de contratração ou 4) capturar fragmentos de cabos-eletrodos no interior dos vasos ou das cavidades cardíacas.

A conclusão de um procedimento de remoção ou de extração de cabo-eletrodo pode resultar em: 1) retirada completa do cabo-eletrodo-alvo; 2) retirada parcial; ou 3) insucesso da retirada. Dependendo do tipo de indicação para a retirada do cabo-eletrodo, pode-se considerar que houve sucesso clínico do procedimento, mesmo que nem todo o cabo-eletrodo tenha sido removido. Considera-se que houve falha do procedimento quando: 1) não é obtido o sucesso clínico; 2) ocorre qualquer complicação permanentemente incapacitante ou 3) ocorre a morte do paciente. Na Tabela 48 , estão listadas as definições estabelecidas pelo Consenso EHRA 2018 com a devida concordância das demais sociedades representativas dessa área do conhecimento.

Tabela 48. – Definições para procedimentos realizados e resultados da remoção de cabos-eletrodos.

Conceito Definição
Procedimento  
Remoção de cabo-eletrodo Retirada de cabo-eletrodo realizando-se apenas a tração direta, sem utilização de ferramentas
wExtração de cabo-eletrodo Retirada de pelo menos um cabo-eletrodo com o uso de ferramentas (listadas na Tabela 47)
Resultado  
Relacionado ao cabo-eletrodo  
  • - Retirada completa do cabo-eletrodo

Remoção ou extração de cabo-eletrodo com sua retirada integral
  • - Retirada incompleta do cabo-eletrodo

Remoção ou extração de cabo-eletrodo em que parte deste fica retida no paciente
Relacionado ao paciente  
  • - Sucesso cirúrgico completo

Retirada de todos os cabos-eletrodos alvos do procedimento, sem qualquer complicação permanentemente incapacitante ou morte relacionada ao procedimento
  • - Sucesso clínico

Retenção de uma pequena parte de um cabo-eletrodo, que não afete negativamente o resultado do procedimento. Esse fragmento pode ser a ponta ou uma pequena parte (<4cm) do cabo-eletrodo (condutor, isolante ou ambos), desde que esse não represente risco de perfuração, evento embólico ou perpetuação de infecção, na ausência de qualquer complicação permanentemente incapacitante ou morte relacionada ao procedimento
  • - Falha do procedimento

Impossibilidade de alcançar sucesso clínico, ocorrência de qualquer complicação permanentemente incapacitante ou morte relacionada ao procedimento.

Tabela 47. – Definições usadas para as diversas ferramentas de extração.

Ferramentas e técnica Definição
Bainha para dissecção e contratração  
  • - Mecânica não energizada

Conjuntos compostos por tubos rígidos metálicos e tubos flexíveis de teflon ou polipropileno
  • - Mecânica rotacional548

Lâminas cortantes metálicas movidas por ativação manual ou elétrica
  • - LASER560

Construídas com microtubos de vidro, condutores da luz produzida por fonte (externa) de emissão do Excimer-LASER
Guia de travamento (Locking stylet) Estiletes revestidos por malha de aço que dão rigidez e sustentação aos cabos-eletrodos
Ferramentas de captura intravascular Cateteres em forma de laço561 ( snares ), cesta ( baskets ) e outras ferramentas utilizadas para capturar fragmentos intravasculares
Balões para oclusão 562 e dilatação vascular Conjuntos compostos por guias e balões para dilatação ou oclusão venosa.
Acessórios complementares Extensor de cabo, espiral de compressão563

Diversos motivos podem justificar a desativação de um cabo-eletrodo: 1) perda da capacidade de estimular adequadamente o coração; 2) necessidade de se mudar o tipo de dispositivo e 3) problemas relacionados à sua fabricação.

Cabos-eletrodos não infectados podem ser abandonados in loco a critério da equipe médica que opera o paciente. Existem, entretanto, desvantagens de se abandonar um cabo-eletrodo: 1) risco de fenômenos trombóticos; 2) restrições para realização de exames de RM; 3) aumento do risco em extração futura, uma vez que, quanto maior for o tempo de permanência do cabo-eletrodo, maior será o risco de insucesso da extração. O principal argumento para se abandonar in loco um cabo-eletrodo não infectado é o risco de complicações graves associadas ao procedimento de extração. Uma publicação recente mostra que a expectativa de sobrevivência no primeiro ano que se segue a um procedimento de troca de eletrodo é semelhante para os casos em que se realiza a extração ou o abandono in loco do cabo-eletrodo. 564 Diante disso, na atualidade, a decisão de se extrair ou não um cabo-eletrodo que será desativado depende, essencialmente, da expertise de cada serviço. Na Tabela 49 , encontram-se as definições estabelecidas pelo consenso EHRA 2018 para cabos-eletrodos não infectados.

Tabela 49. – Definições de termos para cabos eletrodos não infectados.

Cabos eletrodos não infectados Definições
Função do cabo-eletrodo Qualquer função do cabo-eletrodo, incluindo estimulação, sensibilidade e/ou desfibrilação
Falência do cabo-eletrodo Perda total da função do cabo-eletrodo
Disfunção do cabo-eletrodo Cabo-eletrodo não utilizável para estimulação e/ou desfibrilação devido à perda de sua integridade funcional
Cabo-eletrodo abandonado Cabo-eletrodo deixado no coração e não conectado a um DCEI. Pode ser normofuncionante ou não
Recall558,565-568 Retirada ou desativação de um cabo-eletrodo por problemas relacionados à sua fabricação, por exigência de agências regulatórias ou por orientação voluntária do fabricante
  • - Recall classe 1

Produtos perigosos ou defeituosos com probabilidade razoável de causar sérios problemas de saúde ou morte (p. ex., curto-circuito sem aviso)
  • - Recall classe 2

Produtos que podem causar um problema de saúde temporário ou com baixa probabilidade de causar ameaça de natureza grave (p. ex., esgotamento prematuro da bateria)
  • - Recall classe 3

Produtos com pouca probabilidade de causar qualquer reação adversa à saúde, mas que violam as normas regulatórias

DCEI: dispositivo cardíaco eletrônico implantável.

A durabilidade de um DCEI depende tanto de aspectos relacionados à sua fabricação (material e/ou desenho) quanto de seu modo de utilização. Cabos-eletrodos, especificamente, sofrem influência direta da técnica operatória utilizada, que pode influenciar negativamente sua durabilidade. Independentemente desses fatos, estratégias específicas de tecnovigilância devem ser adotadas pelos fabricantes e pelas agências regulatórias para avaliar a durabilidade dos componentes dos DCEI.

Ao se optar pelo abandono in loco de cabos-eletrodos por disfunção, necessidade de mudança do modo de estimulação ou, de maneira inadequada, para o tratamento de uma infecção relacionada a DCEI, cuidados devem ser seguidos, uma vez que a forma como esse cabo-eletrodo é abandonado pode dificultar sua extração no futuro. Recomendações para o abandono de cabos-eletrodos estão listadas na Tabela 50 .

Tabela 50. – Recomendações para o manuseio de cabos-eletrodos abandonados, desnecessários ou disfuncionais.

  Classe de recomendação Nível de evidência
Ao abandonar um cabo-eletrodo, deve-se deixá-lo em condição que evite sua retração dentro da veia e que permita sua extração no futuro I C
Diante de um cabo-eletrodo clinicamente desnecessário ou disfuncional, tanto a opção de abandono como a de extração podem ser consideradas dentro da estratégia cirúrgica569-571 IIa B

Em determinadas situações clínicas, a retirada de um cabo-eletrodo não infectado pode ser obrigatória, tais como: 1) para o tratamento da síndrome da veia cava superior provocada pela presença de cabos-eletrodos; 2) para o tratamento de arritmia cardíaca grave provocada mecanicamente por um fragmento de cabo-eletrodo; 3) para evitar uma lesão cardíaca por um cabo-eletrodo fraturado ou 4) para permitir o tratamento radioterápico na região em que o dispositivo está implantado.

Outras vezes, a extração de um cabo-eletrodo pode ser necessária, como nos casos em que oclusão ou obstrução venosa grave impedem a passagem de um novo cabo-eletrodo.

Em muitos casos, entretanto, a extração dos cabos-eletrodos é opcional, e pode ser definida por um conjunto de fatores menos objetivos, tais como: 1) idade do paciente ou sua expectativa de vida; 2) necessidade futura de realizar exames de RM; 3) risco de desenvolver obstruções venosas graves; ou 4) risco de infecção pela via hematogênica, como ocorre em pacientes com insuficiência renal dialítica. Nessas ocasiões, a expertise do profissional que realiza o procedimento é mandatória para a definição da conduta. Na Tabela 51 , estão listadas as recomendações para a extração de cabos-eletrodos não infectados.

Tabela 51. – Recomendações para a retirada de cabos-eletrodos por causas não infecciosas.

  Classe de recomendação Nível de evidência
Trombose/Comprometimento vascular    
Eventos embólicos clinicamente significativos atribuíveis à presença de cabo-eletrodo ou fragmento deste572-574 I C
Estenose ou oclusão de VCS575 que impeça a passagem de um novo cabo-eletrodo I C
Necessidade de tratamento com stent para evitar o aprisionamento do cabo-eletrodo576,577 I C
Outros    
Arritmias induzidas por cabo-eletrodos ou fragmento578 I C
Dispositivo implantado em local que interfira no tratamento de uma doença maligna579 IIa C
Superpopulação de eletrodos: > 4 eletrodos de um lado ou > 5 eletrodos através da VCS569,580,581 IIa C
Eletrodos abandonados que causem interferência no funcionamento de um dispositivo implantado582,583 IIa C
Eletrodos que, devido ao seu design ou falha, representam uma ameaça ou malefício ao paciente, se deixados no local584-587 IIb C
Condições que contraindiquem a ressonância magnética: cabos-eletrodos ou fragmentos abandonados; cabos-eletrodos não condicionais para ressonância magnética588-593 IIb C
Retirada definitiva do dispositivo por decisão compartilhada IIb C

VCS: veia cava superior.

Durante um procedimento de extração transvenosa de cabos-eletrodo, veias ou estruturas cardíacas podem ser lesionadas. Lesões das veias axilares ou subclávias, dos troncos venosos braquiocefálicos ou da veia cava superior podem provocar hemorragia grave com necessidade de transfusão sanguínea ou até mesmo correção cirúrgica. Avulsão muscular do átrio direito ou do VD, assim como perfuração de veia tributária do seio coronário, pode acarretar tamponamento cardíaco. A laceração da veia cava superior em seu trajeto extrapericárdico, entretanto, é a complicação catastrófica mais frequente e mais letal. Outras complicações, como arritmias cardíacas autolimitadas, pneumotórax ou retenção de fragmentos de cabos-eletrodos, também podem ocorrer e necessitar de cuidados específicos.

De maneira geral, as complicações são agrupadas em maiores e menores, em função de sua gravidade e do tipo de correção que demandam. Na Tabela 52 , estão listadas as complicações conforme sua classificação e incidência. 577

Tabela 52. – Classificação e incidência das complicações perioperatórias mais frequentes*.

Complicações Incidência %
Maiores 0,19–1,80
Morte 0,19–1,20
Avulsão cardíaca 0,19–0,96
Laceração vascular 0,16–0,41
Parada respiratória 0,20
Acidente vascular cerebral 0,07–0,08
Derrame pericárdico que requer intervenção 0,23–0,59
Hemotórax que requer intervenção 0,07–0,20
Parada cardíaca 0,07
Tromboembolismo que requer intervenção 0,07
Lesão valvar tricúspide com repercussão hemodinâmica grave 0,03
Embolia pulmonar maciça 0,08
Menores 0,06–6,20
Derrame pericárdico sem necessidade de intervenção 0,07–0,16
Hematoma que requer drenagem 0,90–1,60
Trombose venosa que requer intervenção médica 0,10–0,21
Necessidade de reparo vascular no local do acesso venoso 0,07–0,13
Migração de fragmento de cabo-eletrodo sem sequelas 0,20
Sangramento que requer transfusão de sangue 0,08–1,00
Fístula arteriovenosa que requer intervenção 0,16
Pneumotórax que requer drenagem torácica 1,10
Agravamento da insuficiência valvar tricúspide 0,02–0,59
Embolia pulmonar 0,24–0,59

* Fonte: adaptada de Kusumoto et al., 2017.594

Vários estudos têm sido desenhados para identificar fatores de risco determinantes da morbimortalidade relacionada à extração transvenosa de cabos-eletrodos. Esses estudos têm demonstrado baixa taxa de ocorrência de complicações catastróficas e de morte perioperatória, não permitindo a identificação de fatores de risco para esses eventos. Por outro lado, vários fatores demográficos, clínicos e cirúrgicos estão associados à mortalidade nos 30 primeiros dias que se seguem a um procedimento de extração. Fatores associados a complicações e morte tardias também foram identificados. 30 , 595 - 603 Na Tabela 53 , estão listados os fatores de risco já identificados e seu impacto na morbimortalidade.

Tabela 53. – Fatores de risco para morte e complicações associadas à remoção de cabos-eletrodos.

Fator Risco associado ao fator
Idade Aumenta a mortalidade em 1,05 vez
Sexo feminino Aumenta o risco de complicações maiores em 4,5 vezes
Baixo índice de massa corporal (< 25kg/m2)
  • Aumenta a mortalidade de 30 dias em 1,8 vez

  • Aumenta o número de complicações relacionadas à extração

História de AVC Aumenta o risco de complicações maiores em 2,0 vezes
Disfunção grave do VE Aumenta o risco de complicações maiores em 2,0 vezes
IC avançada
  • Aumenta o risco de mortalidade em 30 dias de 1,3 a 8,5 vezes

  • Aumenta a mortalidade em 1 ano de seguimento em 3,0 vezes

Disfunção renal
  • Disfunção renal terminal aumenta o risco de morte em 30 dias em 4,8 vezes

  • Creatinina > 2,0 aumenta a mortalidade hospitalar e mortalidade em 1 ano em 2,0 vezes

Diabetes melito Aumenta a mortalidade hospitalar e mortalidade geral em 1,7 vez
Coagulopatia
  • INR elevado aumenta risco de complicações maiores em 2,7 vezes e a mortalidade de 30 dias em 1,3 vez

  • Uso de anticoagulantes aumenta mortalidade de 1 ano em 1,8 vez

Contagem de plaquetas Plaquetopenia aumenta risco de complicações maiores em 1,7 vez
Anemia Aumenta 3,3 vezes o risco de morte em 30 dias
Número de cabos-eletrodos extraídos
  • Aumenta 3,5 vezes o risco de qualquer complicação

  • Aumenta em 1,6 vez a mortalidade a longo prazo

Presença de CDI de duplo coil Aumenta 2,7 vezes a mortalidade em 30 dias
Extração por infecção
  • Aumenta em 2,7 a 30 vezes a mortalidade em 30 dias

  • Aumenta de 5,0 a 9,7 vezes a mortalidade em 1 ano

  • PCR >7,2mg/L aumenta a mortalidade em 30 dias Aumenta

  • 3,5 vezes a mortalidade geral

Experiência do operador Aumenta 2,6 vezes a número de complicações relacionadas ao procedimento
Cirurgia cardíaca prévia Reduz a incidência de complicações maiores

AVC: acidente vascular cerebral; CDI: cardioversor-desfibrilador implantável; IC: insuficiência cardíaca; INR: Razão Normalizada Internacional; PCR: proteína C reativa; VE: ventrículo esquerdo.

Diante da dificuldade de se prever a ocorrência de complicações catastróficas perioperatórias, torna-se fundamental a prevenção da morte relacionada a esses eventos, o que implica treinamento dos profissionais envolvidos e capacitação técnica dos serviços que realizam extração de cabos-eletrodo. Uma revisão sistemática recente 604 mostra a estreita relação entre o número de procedimentos realizados pelo médico e a taxa de complicações associadas à extração de eletrodos. Operadores em início de experiência devem realizar os primeiros 40 procedimentos de extração transvenosa sob a supervisão de operadores mais experientes. Recomenda-se um volume mínimo de 20 procedimentos de extração transvenosa por ano, a fim de manutenção da capacidade técnica, para todos os operadores.

8. Recomendações para prevenção de interferências eletromagnéticas

8.1. Cirurgia com Uso de Eletrocautério

A eletrocirurgia compreende o uso de corrente alternada de alta frequência (200 KHz a 2,2 MHz), que é convertida em calor ao sofrer resistência ao passar pelos tecidos, o que permite efetuar os efeitos desejáveis: coagulação e corte. O bisturi elétrico é usado na maioria das especialidades cirúrgicas.

Por ser mais efetiva, a eletrocirurgia monopolar é a mais empregada na prática. Nessa modalidade, o eletrodo ativo está no sítio cirúrgico (haste do bisturi), enquanto o eletrodo indiferente é uma placa colocada em contato com a pele do paciente, em local afastado. A corrente flui entre os eletrodos, atravessando o corpo.

Um número crescente de pacientes com DCEI é submetido a intervenções cirúrgicas, o que expõe esses pacientes às interferências eletromagnéticas. A eletrocirurgia monopolar pode provocar várias anormalidades nos DCEI, tais como reprogramação do gerador de pulsos, inibição temporária da estimulação, deflagração de estimulação em frequência elevada, depleção da bateria e falha de estimulação, dano ao circuito, aumento de limiares, e disparo de terapia inapropriada (choques) em caso de CDI. 605

Para minimizar os riscos do uso do eletrocautério, alguns cuidados devem ser tomados no perioperatório: (1) a aplicação do bisturi monopolar deve ser intermitente, em pulsos de curta duração e com menor energia; (2) a placa indiferente deve ser posicionada em local de modo que a energia não flua através do gerador ou eletrodos.

Em geral, quando o sítio cirúrgico é localizado acima da cicatriz umbilical ou a uma distância menor que 15cm do gerador, o uso de bisturi monopolar deve ser evitado. Nesse cenário, deve-se preferir o bisturi bipolar, que é seguro, mas que também não deve ser aplicado diretamente sobre o gerador. 606 , 607 Em cirurgia de cabeça e pescoço, a placa indiferente do bisturi monopolar deve ser colocada na parte posterior do ombro contralateral à da loja do dispositivo. Assim, por exemplo, no caso de gerador implantado na região infraclavicular esquerda, a placa do bisturi deve ser colocada sob o ombro direto.

Com a finalidade de proteger o paciente e o DCEI dos efeitos indesejáveis do eletrocautério, duas abordagens têm sido empregadas: aposição de magneto sobre o gerador de pulsos e reprogramação do dispositivo antes do procedimento. Em caso de MP, o uso do magneto durante a cirurgia é opção quando o gerador desativar o circuito de sensibilidade sob efeito magnético (modo assíncrono) e apresentar bateria em bom estado. 608 , 609 A aplicação do ímã sobre a loja do gerador causa a reversão do MP para o modo assíncrono, ou seja, desabilita a sensibilidade e modifica a frequência de estimulação para a frequência magnética, que é geralmente maior que a frequência básica programada.

Em caso de CDI, um estudo randomizado comparou a aplicação do magneto versus a reprogramação em pacientes com CDI submetidos à cirurgia utilizando eletrocautério monopolar, a uma distância maior de 15 cm do gerador; os autores concluíram que as duas estratégias são seguras. 610 A aplicação do magneto sobre a loja do CDI desabilita somente a terapia para as taquiarritmias, não alterando a função de MP. No paciente com CDI e dependente de MP, o dispositivo deve ser reprogramado para modo assíncrono, antes do procedimento. A reprogramação do DCEI deve ser realizada logo antes da intervenção cirúrgica e revertida para a programação basal logo após o término do procedimento ( Tabela 54 ).

Tabela 54. – Recomendações para prevenção de interferências eletromagnéticas em cirurgia com uso de eletrocautério.

  Classe de recomendação Nível de evidência
  • A eletrocirurgia monopolar pode provocar várias anormalidades nos DCEI, e deve-se usar, preferencialmente, o eletrocautério bipolar

  • No paciente com CDI e não dependente de marca-passo, pode-se realizar a aplicação do magneto sobre a loja do CDI com segurança, para desabilitar as terapias para as taquiarritmias e evitar choques inapropriados

I A
Em caso de marca-passo, o uso do magneto durante a cirurgia é uma opção quando o gerador desativar o circuito de sensibilidade sob efeito magnético (modo assíncrono) e apresentar bateria em bom estado I B
  • Em cirurgia de cabeça e pescoço, a placa indiferente, em se tratando de bisturi monopolar, deve ser colocada na parte posterior do ombro contralateral à da loja do dispositivo

  • No paciente com CDI e dependente de marca-passo, as terapias são desativadas, e o dispositivo deve ser reprogramado para um modo assíncrono antes do procedimento

  • A reprogramação do DCEI deve ser realizada logo antes da intervenção cirúrgica e revertida para a programação basal logo após o término do procedimento

I C
Em casos de cirurgias em sítio acima do umbigo, deve-se utilizar um bisturi bipolar. Se não for possível evitar o uso do bisturi monopolar, o gerador deve ser reprogramado IIa C

CDI: cardioversor-desfibrilador implantável; DCEI: dispositivos cardíacos eletrônicos implantáveis.

8.2. Ressonância Magnética

A ressonância magnética (RM) tem se apresentado como ferramenta diagnóstica cada vez mais útil e acessível, com crescente relevância para avaliação diagnóstica e prognóstica.

O número de exames de RM apresentou um crescimento substancial nos últimos 20 anos, com mais de 60 milhões de exames anualmente ao redor do mundo. Estima-se que um paciente tem 50% a 75% de probabilidade de receber indicação de realização de RM após implante de um DCEI durante o tempo de vida do dispositivo. 594 , 611

O ambiente de realização do exame pode ser dividido em zonas, conforme descrito por Kanal et al . 612 e adotado pelo 2017 HRS expert consensus. 593 A zona 4 refere-se à sala do exame, sendo o espaço com maior risco para pacientes e equipe de saúde, incluindo o risco potencial de deslocamento de objetos metálicos. A zona 3 é o espaço fora da sala de exame, incluindo a sala de controle. Existe, nessa área, um risco potencial, devendo ser reservada a pessoal treinado. A zona 2 inclui área de recepção, e a zona 1 corresponde às regiões acessíveis ao público geral.

Com relação aos DCEI durante a realização de RM, podem ser definidos como: 613

  • Seguros : não representam perigo para realização da RM.

  • Condicionados : não representam perigo para realização da RM, desde que atendidas determinadas especificações. Tais condições podem incluir parâmetros como: região do corpo a ser examinada, força do campo magnético, gradiente espacial, tempo de exposição ao campo magnético, campo de radiofrequência e taxa de absorção específica. Condições adicionais podem ser requeridas, incluindo uso de combinações específicas de gerador e eletrodos, bem como modo de programação do dispositivo. As condições específicas podem variar entre fabricantes e entre aparelhos de um mesmo fabricante.

  • Não condicionados : representam perigo para realização da RM. Incluem todos os sistemas de estimulação cardíaca que não são condicionados para realização do exame. Isso inclui geradores condicionados associados a eletrodos não condicionados ou sistemas totalmente condicionados implantados em pacientes que não preenchem todas as condições específicas de uso, como aqueles com eletrodos abandonados.

Os DCEI não são classificados como seguro , e os novos DCEI que foram construídos com tecnologia apropriada são considerados condicionados para realização do exame.

Na prática clínica, os aparelhos precisam ser programados antes da realização de RM, com um nível de segurança bastante aceitável.

A interação do campo magnético estático, gradiente magnético e radiofrequência sobre o DCEI pode prejudicar o funcionamento de componentes eletrônicos, causar migração ou deslocamento de componentes do sistema, geração de corrente de energia que pode danificar o aparelho e/ou o miocárdio e provocar oversensing, undersensing e arritmias. A influência desses parâmetros nos DCEI pode ser dividida em dois grupos: os que prejudicam o funcionamento do DCEI de maneira transitória e os que o fazem de modo permanente. 614 , 615

As respostas a essas fontes de interferência podem ser variadas:

  • Campo magnético estático : deslocamento do dispositivo, ativação de sensores, perda súbita da função do dispositivo, mudanças no ECG.

  • Gradiente de campo magnético : indução de arritmias (raro); o versensing ou undersensing .

  • Campo de radiofrequência : aquecimento tissular adjacente aos eletrodos; indução de arritmias (raro); reprogramação do dispositivo ( reset ); interações de oversensing ou undersensing .

  • Efeitos combinados : perda súbita da função do dispositivo; alteração de função (parâmetros); forças mecânicas (vibração); reset do dispositivo; dano do gerador e/ou dos eletrodos.

  • Relacionados à imagem : artefatos que prejudiquem a imagem adequada do dispositivo.

As interações potenciais entre DCEI e interferência eletromagnética pela RM incluem:

  • Campo magnético induzido e torque devido a materiais ferromagnéticos : movimento de gerador é extremamente improvável pelo confinamento e pelo tecido subcutâneo adjacente. Eletrodos não contêm material ferromagnético suficiente para causar movimentos.

  • Corrente elétrica induzida por gradiente de campo magnético : gradientes de campo magnético podem induzir corrente, o que pode levar a captura miocárdica e, potencialmente, causar arritmias atriais ou ventriculares.

  • Calor e dano tecidual : campos de radiofrequência podem levar a aquecimento de componentes não condicionados, levando a aquecimento e dano térmico ao tecido adjacente (ablação funcional). Mudanças de sensibilidade e limiar de captura podem ocorrer como resultado do dano tecidual próximo aos eletrodos.

  • Efeitos na atividade do dispositivo : o dispositivo de estimulação cardíaca pode ser programado por meio da colocação de um ímã, liberando o dispositivo para a interação. Os campos magnéticos podem, portanto, afetar a atividade de um dispositivo condicional, com possibilidade de alterar a programação do dispositivo.

  • Reset elétrico : a interferência eletromagnética de alta energia pode levar a reset elétrico. Pode ocorrer acionamento de um modo de demanda de backup . Os parâmetros de reset de energia variam de acordo com o fornecedor e o tipo de dispositivo e podem incluir um conjunto de variações. Inibição da função de estimulação por sinais gerados por RM ou energia de estimulação abaixo do limiar (bipolar ou unipolar) em um paciente dependente de MP pode ocorrer. Além disso, o status da bateria pode ser afetado, principalmente para dispositivos próximos de um intervalo de substituição eletiva (ERI), que pode resultar em funções não confiáveis.

  • Funções inapropriadas e terapias : interferências eletromagnéticas por pulsos de radiofrequência ou mudanças rápidas de gradientes de campo magnético podem causar oversensing que podem levar à inibição inapropriada da estimulação e possibilidade de assistolia em pacientes dependentes de estimulação do MP, ou indução de terapias levando a choques inapropriados em pacientes com CDI.

Esses efeitos são influenciados por vários fatores, incluindo força do campo magnético, potência de RF, posição do paciente e do dispositivo em relação à máquina de RM, características do dispositivo e tamanho do paciente.

Tradicionalmente, a realização de RM em portadores de DCEI costumava ser contraindicada. O primeiro sistema condicionado para a realização de RM foi introduzido na Europa em 2010 e liberado pelo Food and Drug Administration (FDA) em 2011, nos EUA. 613 , 616

Para tornarem-se condicionados para RM, os DCEI sofreram mudanças estruturais (p. ex., uso de materiais não ferromagnéticos) e alterações de software que reduzem ou eliminam potenciais efeitos adversos. Uma vez acionada a programação especial (modo RM), o aparelho reverte o modo de estimulação para assíncrono e aumenta a energia de estimulação para evitar inibição da estimulação ou falha de captura durante o exame. Nos CDI, a função antitaquicardia é desabilitada temporariamente. No momento do exame, portanto, esses pacientes estão desprotegidos em caso de arritmias ventriculares.

A decisão de realizar RM em um paciente com DCEI envolve riscos e benefícios; assim, os fatores potenciais de risco devem ser identificados. Em pacientes com DCEI condicionais, o exame pode ser realizado sem riscos adicionais ao paciente, seguindo recomendações e protocolos estabelecidos. 617 , 618

Antes da realização do exame, é importante identificar o ritmo de base do paciente e se o paciente é dependente de MP, ativar a programação específica para a realização de RM; confirmar se todo o sistema é condicional para RM; verificar presença de eletrodos abandonados ou epicárdicos.

De modo geral, a maioria dos sistemas é aprovada para exames de RM de 1,5T, gradiente com taxa de variação de 200 T/m/s, SAR máxima de SAR 2 W/kg, número limitado de sequências e comprimento de imagens. Em novos dispositivos, o exame pode ser também seguro sob condições mais amplas. A maioria dos sistemas novos permite o exame de RM do corpo inteiro.

Um sistema condicional de RM consiste em uma combinação de eletrodos e gerador que foi especificamente testada para garantir condições de uso seguras durante o exame. A presença de qualquer componente do dispositivo que não atenda aos critérios de condicionalidade de RM o classifica como não condicional. Isso inclui um gerador condicional de RM combinado com componentes não condicionais e sistemas de dispositivos que combinam componentes individuais de eletrodos e gerador condicionais de RM de vários fabricantes, dado que essas não são combinações especificamente testadas em conjunto para a segurança do exame.

A rotulagem de condicional também especifica a localização do gerador: localização peitoral para sistemas transvenosos. Outros exemplos de componentes não condicionais incluem eletrodos epicárdicos, eletrodos abandonados, eletrodos fraturados ou um dispositivo não cardíaco ativo.

A programação do dispositivo fora do modelo de programação condicional para RM também torna o dispositivo não condicional. O estado da bateria deve estar adequado para considerar o dispositivo condicional ( Tabela 55 ).

Tabela 55. – Recomendações para prevenção de interferências eletromagnéticas com relação à ressonância magnética.

  Classe de recomendação Nível de evidência
  • DCEI condicionais para RM devem ser assim considerados apenas quando a especificação do produto for respeitada, o que inclui a programação do “modo RM” apropriado e a realização do exame com os pré-requisitos especificados para o dispositivo

  • A RM em um paciente com um sistema condicional deve sempre ser realizada no contexto de um fluxo de trabalho institucional padronizado, rigorosamente aplicado, seguindo as condições técnicas apropriadas

  • Recomenda-se aos pacientes com sistema condicional de RM que ECG e monitoração por oximetria de pulso continuem até a o término da observação do paciente, ou até que outras configurações clinicamente apropriadas do dispositivo sejam restauradas

I A
  • Recomenda-se para o paciente com DCEI não condicional que a avaliação do dispositivo seja realizada imediatamente antes e após a RM, com documentação de limiar de estimulação, amplitude das ondas P e R e impedância de cabo-eletrodo utilizando um protocolo padronizado Um desfibrilador/monitor (com função de estimulação externa) e um sistema de programação de DCEI específico do fabricante devem estar imediatamente disponíveis na área adjacente à sala de RM, enquanto o dispositivo não condicional estiver programado para aquisição das imagens.

  • Recomendam-se que monitoramento contínuo do ECG e oximetria de pulso sejam utilizados enquanto o dispositivo não condicional da RM estiver programado para aquisição imagens. Recomenda-se pessoal com a habilidade para suporte avançado de vida para acompanhamento do paciente com DCEI não condicional para RM, até que seja avaliado e declarado estável para retornar a status não monitorado

  • Para pacientes com um DCEI não condicional para RM dependente de estimulação (MP ou CDI), recomendam-se que: a) um médico com capacidade para implantar MP temporário esteja imediatamente disponível nas instalações; b) um médico com a capacidade de programar DCEI esteja imediatamente disponível nas instalações

  • Recomenda-se que o paciente com um DCEI não condicional para RM e dependente de estimulação seja programado para um modo de estimulação assíncrona com desativação de recursos adaptativos (sensor de resposta à frequência) durante o exame. A frequência de estimulação deve ser apropriada para evitar estimulação competitiva

  • Todas as detecções de taquicardia para pacientes com CDI devem ser desativadas antes da RM

I B
  • Recomenda-se aos pacientes com sistema condicional para RM a assistência de pessoal com habilidade para realizar suporte avançado de vida. No caso de atendimento ao paciente, este deve ocorrer pelo período até o DCEI ser reprogramado, ou até ser avaliado e declarado estável para retornar ao status não monitorado

  • O médico responsável pela RM deve ser informado da presença de paciente com um DCEI não condicional

  • Recomenda-se que o monitoramento do ECG e da oximetria de pulso seja continuado até o final do exame ou até que as configurações do dispositivo sejam reprogramadas

  • Todos os esforços de reanimação e tratamentos de emergência que envolvam o uso de um desfibrilador/monitor, sistema de programação de DCEI ou qualquer outro equipamento inseguro para RM devem ser realizados após a transferência do paciente para fora da zona 4

I C
  • É razoável que pacientes com um sistema não condicional sejam submetidos à RM se não houver eletrodo fraturado, epicárdico ou abandonado, e a RM é o melhor teste para a condição. Deve existir um protocolo institucional e um médico responsável pela RM e pelo DCEI

  • É razoável a realização de RM imediatamente após o implante de um eletrodo ou gerador de um dispositivo não condicional, se for clinicamente justificado

  • Para um paciente com DCEI não condicional e que não depende da estimulação, é possível programar o dispositivo para qualquer modo sem estimulação (OVO/ODO) ou para um modo inibido (DDI/VVI), com desativação de recursos avançados ou adaptativos durante o exame de RM

IIa B
  • É razoável a realização de RM em paciente com sistema condicional implantado mais recentemente do que o período isento de condicionalidade do sistema, com base na avaliação de risco e benefício para esse paciente

  • Para os pacientes com DCEI não condicionals, é possível realizar RM repetidas quando necessário e sem restrição quanto ao intervalo mínimo entre os estudos de imagem ou o número máximo de estudos realizados

  • É razoável programar pacientes com DCEI não condicional que não depende da estimulação para um modo de estimulação assíncrona (VOO/DOO) com desativação de recursos avançados ou adaptáveis durante o exame e com taxa de estimulação que evite a estimulação competitiva

  • Para pacientes com DCEI não condicional, é razoável agendar uma avaliação completa do dispositivo dentro de 1 semana nos casos de aumento do limiar de estimulação > 1,0V, diminuição da amplitude da onda P ou da onda R > 50%, alteração da impedância do eletrodo > 50ohms e alteração da impedância do eletrodo de choque > 5ohms.

IIa C

CDI: cardioversor-desfibrilador implantável; DCEI: dispositivos cardíacos eletrônicos implantáveis; ECG: eletrocardiograma; MP: marca-passo; RM: ressonância magnética.

8.3. Radioterapia

Número crescente de pacientes que se submetem à RT é portador de DCEI. Embora a ocorrência de disfunções induzidas por RT seja rara, recomendações de segurança são importantes.

A radiação ionizante pode causar interferência em componentes do CMOS (do inglês, complementary metal-oxide-semiconductor ) do gerador. A geração secundária de nêutrons é a maior preditora de disfunção dos DCEI no contexto da RT. Os geradores de pulsos mais modernos têm menor consumo de energia e circuitos menores, constituídos de metal semicondutor. Isso leva a maior suscetibilidade desses dispositivos a possíveis danos causados por radiação ionizante. 593 , 619

Altas doses de radiação, especialmente com energia > 6MV, podem ocasionar erros de software e hardware . Esses distúrbios são geralmente transitórios, como inibição da estimulação, alterações na sensibilidade ou estimulação inapropriada na frequência máxima do sensor. A reversão para o modo de backup ( reset ), que pode ser corrigida com a reprogramação, é uma das disfunções mais relatadas. Defeitos permanentes também podem ocorrer, como a perda da telemetria e depleção prematura da bateria. A falência do dispositivo, com perda completa da função do DCEI, já foi descrita in vitro . 593 , 619 - 621

É importante considerar, ainda, que os defeitos nos DCEI podem aparecer semanas ou meses após o término da RT (defeitos latentes). 622 Disfunções dos dispositivos são relatadas em até 3% dos cursos de RT. Eventos clinicamente relevantes são muito raros, dependem do tipo de dispositivo e da tolerância do paciente às alterações. Por exemplo, o paciente dependente de MP pode apresentar bradicardia e sintomas relacionados. 593 , 619

O planejamento da RT deve considerar as especificações do DCEI, bem como as características dos pacientes (dependente ou não de estimulação, histórico de taquicardia ou fibrilação ventricular (TV/FV) ( Tabela 56 ).

Tabela 56. – Recomendações para prevenção de interferências eletromagnéticas relacionadas à radioterapia.

  Classe de indicação Nível de evidência
  • Antes de iniciar a RT, completa avaliação do DCEI deve ser efetuada e a equipe responsável pelo tratamento deve ser informada:
    1. Se o dispositivo é um MP ou CDI
    2. Se o paciente é dependente da estimulação
    3. Frequência mínima de estimulação programada
    4. Frequências máximas programadas ( tracking e de sensor)
  • Tratamento com RT que não produz nêutrons é preferível àquela que produz nêutron, pelo menor risco de causar disfunções do DCEI, como reset do gerador

  • Avaliação completa do DCEI deve ser realizada semanalmente, em paciente submetido à RT que produz nêutrons

  • Avaliação completa do DCEI deve ser realizada ao término de um curso de tratamento radioterápico

I B
São recomendadas a interrogação e a avaliação do DCEI a 1, 3 e 6 meses após o término da RT pelo risco de danos latentes I C
A relocação do DCEI pode ser indicada se o gerador estiver situado no caminho do feixe da radiação IIa C
O reposicionamento do DCEI não é recomendado quando a dose cumulativa recebida pelo dispositivo é <5Gy III B

CDI: cardioversor-desfibrilador implantável; DCEI: dispositivos cardíacos eletrônicos implantáveis; MP: marca-passo; RT: radioterapia.

9. Conclusão

Muitas evidências científicas surgiram desde a publicação das últimas diretrizes brasileiras de DCEI da SBC/SOBRAC. A evolução da tecnologia e do próprio conhecimento deve estar alinhada com a prática clínica e com a atenção à saúde pública. Dessa maneira, o presente documento destaca a evolução do tratamento das arritmias cardíacas, mas não se furta em destacar a necessidade premente do uso racional dos recursos financeiros em favor do bem maior, qual seja, a saúde coletiva.

Footnotes

Nota : Estas diretrizes se prestam a informar e não a substituir o julgamento clínico do médico que, em última análise, deve determinar o tratamento apropriado para seus pacientes.

Referências

  • 1.Gualandro DM, Yu PC, Caramelli B, Marques AC, Calderaro D, Fornari LS, et al. 3ª Diretriz de Avaliação Cardiovascular Perioperatória da Sociedade Brasileira de Cardiologia. Arq. Bras. Cardiol. . 2017;109(3 suppl 1):1–104. doi: 10.5935/abc.20170140. [DOI] [Google Scholar]
  • 2.Oliveira JC, Martinelli M, Nishioka SA, Varejão T, Uipe D, Pedrosa AA, et al. Efficacy of Antibiotic Prophylaxis Before the Implantation of Pacemakers and Cardioverter-Defibrillators: Results of a Large, Prospective, Randomized, Double-Blinded, Placebo-Controlled Trial. Circ Arrhythm Electrophysiol . 2009;2(1):29–34. doi: 10.1161/CIRCEP.108.795906. [DOI] [PubMed] [Google Scholar]
  • 3.Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral J, et al. 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing. Heart Rhythm . 2016;13(2):50–86. doi: 10.1016/j.hrthm.2015.11.018. [DOI] [PubMed] [Google Scholar]
  • 4.Gleva MJ, Robinson M, Poole J. The Saga of Defibrillation Testing: When Less Is More. Curr Cardiol Rep . 2018;20(6) doi: 10.1007/s11886-018-0987-6. 44. [DOI] [PubMed] [Google Scholar]
  • 5.Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, et al. 2013 ESC Guidelines on Cardiac Pacing and Cardiac Resynchronization Therapy: The Task Force on Cardiac Pacing and Resynchronization Therapy of the European Society of Cardiology (ESC). Developed in Collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J. . 2013;34(29):2281–2329. doi: 10.1093/eurheartj/eht150. [DOI] [PubMed] [Google Scholar]
  • 6.Mond HG, Irwin M, Morillo C, Ector H. The World Survey of Cardiac Pacing and Cardioverter Defibrillators: Calendar Year 2001. Pacing Clin Electrophysiol . 2004;27(7):955–964. doi: 10.1111/j.1540-8159.2004.00565.x. [DOI] [PubMed] [Google Scholar]
  • 7.Ferrer MI. The Sick Sinus Syndrome in Atrial Disease. JAMA . 1968;206(3):645–646. [PubMed] [Google Scholar]
  • 8.Gillis AM. Ellenbogen KA, Wilkoff BL, Kay GN, Lau CP, Wilkoff BL, Auricchio A.editors . Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy . Philadelphia: Elsevier; 2017. Pacing for Sinus Node Disease; pp. 375–398. [Google Scholar]
  • 9.Melzer C, Witte J, Reibis R, Bondke HJ, Combs W, Stangl K, et al. Predictors of Chronotropic Incompetence in the Pacemaker Patient Population. Europace . 2006;8(1):70–75. doi: 10.1093/europace/euj017. [DOI] [PubMed] [Google Scholar]
  • 10.Nielsen JC, Thomsen PE, Højberg S, Møller M, Vesterlund T, Dalsgaard D, et al. A Comparison of Single-Lead Atrial Pacing with Dual-Chamber Pacing in Sick Sinus Syndrome. Eur Heart J. . 2011;32(6):686–696. doi: 10.1093/eurheartj/ehr022. [DOI] [PubMed] [Google Scholar]
  • 11.Lamas GA, Lee K, Sweeney M, Leon A, Yee R, Ellenbogen K, et al. The Mode Selection Trial (MOST) in Sinus Node Dysfunction: Design, Rationale, and Baseline Characteristics of the First 1000 Patients. Am Heart J. . 2000;140(4):541–551. doi: 10.1067/mhj.2000.109652. [DOI] [PubMed] [Google Scholar]
  • 12.Semelka M, Gera J, Usman S. Sick Sinus Syndrome: A Review. Am Fam Physician . 2013;87(10):691–696. [PubMed] [Google Scholar]
  • 13.Dobrzynski H, Boyett MR, Anderson RH. New Insights Into Pacemaker Activity: Promoting Understanding of Sick Sinus Syndrome. Circulation . 2007;115(14):1921–1932. doi: 10.1161/CIRCULATIONAHA.106.616011. [DOI] [PubMed] [Google Scholar]
  • 14.Thery C, Gosselin B, Lekieffre J, Warembourg H. Pathology of Sinoatrial Node. Correlations with Electrocardiographic Findings in 111 Patients. Am Heart J . 1977;93(6):735–740. doi: 10.1016/s0002-8703(77)80070-7. [DOI] [PubMed] [Google Scholar]
  • 15.John RM, Kumar S. Sinus Node and Atrial Arrhythmias. Circulation . 2016;133(19):1892–1900. doi: 10.1161/CIRCULATIONAHA.116.018011. [DOI] [PubMed] [Google Scholar]
  • 16.Benson DW, Wang DW, Dyment M, Knilans TK, Fish FA, Strieper MJ, et al. Congenital Sick Sinus Syndrome Caused by Recessive Mutations in the Cardiac Sodium Channel Gene (SCN5A) J Clin Invest . 2003;112(7):1019–1028. doi: 10.1172/JCI18062. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Jones SA, Boyett MR, Lancaster MK. Declining into Failure: The Age-Dependent Loss of the L-type Calcium Channel Within the Sinoatrial Node. Circulation . 2007;115(10):1183–1190. doi: 10.1161/CIRCULATIONAHA.106.663070. [DOI] [PubMed] [Google Scholar]
  • 18.Yeh YH, Burstein B, Qi XY, Sakabe M, Chartier D, Comtois P, et al. Funny Current Downregulation and Sinus Node Dysfunction Associated with Atrial Tachyarrhythmia: A Molecular Basis for Tachycardia-Bradycardia Syndrome. Circulation . 2009;119(12):1576–1585. doi: 10.1161/CIRCULATIONAHA.108.789677. [DOI] [PubMed] [Google Scholar]
  • 19.Elvan A, Wylie K, Zipes DP. Pacing-Induced Chronic Atrial Fibrillation Impairs Sinus Node Function in Dogs. Electrophysiological Remodeling. Circulation . 1996;94(11):2953–2960. doi: 10.1161/01.cir.94.11.2953. [DOI] [PubMed] [Google Scholar]
  • 20.Byrne EA, Fleg JL, Vaitkevicius PV, Wright J, Porges SW. Role of Aerobic Capacity and Body Mass Index in the Age-Associated Decline in Heart Rate Variability. J Appl Physiol . 1996;81(2):743–750. doi: 10.1152/jappl.1996.81.2.743. [DOI] [PubMed] [Google Scholar]
  • 21.Elias J., Neto Great Arteries Contribution in Orthostasis Cardiovascular Adaptation. Arq Bras Cardiol . 2006;87(2):209–222. doi: 10.1590/s0066-782x2006001500023. [DOI] [PubMed] [Google Scholar]
  • 22.den Hoed M, Eijgelsheim M, Esko T, Brundel BJ, Peal DS, Evans DM, et al. Identification of Heart Rate-Associated Loci and Their Effects on Cardiac Conduction and Rhythm Disorders. Nat Genet . 2013;45(6):621–631. doi: 10.1038/ng.2610. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Mezzano V, Liang Y, Wright AT, Lyon RC, Pfeiffer E, Song MY, et al. Desmosomal Junctions are Necessary for Adult Sinus Node Function. Cardiovasc Res . 2016;111(3):274–286. doi: 10.1093/cvr/cvw083. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Gauer RL. Evaluation of Syncope. Am Fam Physician . 2011;84(6):640–650. [PubMed] [Google Scholar]
  • 25.Maddala RNM, Ashwal AJ, Rao MS, Padmakumar R. Chronic Lithium Intoxication: Varying Electrocardiogram Manifestations. Indian J Pharmacol . 2017;49(1):127–129. doi: 10.4103/ijp.IJP20416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Brignole M, Moya A, Lange FJ, Deharo JC, Elliott PM, Fanciulli A, et al. 2018 ESC Guidelines for the Diagnosis and Management of Syncope. Eur Heart J . 2018;39(21):1883–1948. doi: 10.1093/eurheartj/ehy037. [DOI] [PubMed] [Google Scholar]
  • 27.Monté CP, Monté CJ, Boon P, Arends J. Epileptic Seizures Associated with Syncope: Ictal Bradycardia and Ictal Asystole. Epilepsy Behav . 2019;90:168–171. doi: 10.1016/j.yebeh.2018.10.027. [DOI] [PubMed] [Google Scholar]
  • 28.Elias J, Kuniyoshi R, Carloni WV, Borges MR, Peixoto CA, Pimentel D. Glossopharyngeal Neuralgia Associated with Cardiac Syncope. Arq Bras Cardiol . 2002;78(5):510–519. doi: 10.1590/s0066-782x2002000500008. [DOI] [PubMed] [Google Scholar]
  • 29.Hayashi H, Sumiyoshi M, Yasuda M, Komatsu K, Sekita G, Kawano Y, et al. Prevalence of the Brugada-Type Electrocardiogram and Incidence of Brugada Syndrome in Patients with Sick Sinus Syndrome. Circ J. . 2010;74(2):271–277. doi: 10.1253/circj.cj-09-0455. [DOI] [PubMed] [Google Scholar]
  • 30.Kusumoto FM, Schoenfeld MH, Barrett C, Edgerton JR, Ellenbogen KA, Gold MR, et al. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients with Bradycardia and Cardiac Conduction Delay: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm . 2019;16(9):128–226. doi: 10.1016/j.hrthm.2018.10.037. [DOI] [PubMed] [Google Scholar]
  • 31.Lamas GA, Orav EJ, Stambler BS, Ellenbogen KA, Sgarbossa EB, Huang SK, et al. Quality of Life and Clinical Outcomes in Elderly Patients Treated with Ventricular Pacing as Compared with Dual-Chamber Pacing. Pacemaker Selection in the Elderly Investigators. N Engl J Med . 1998;338(16):1097–1104. doi: 10.1056/NEJM199804163381602. [DOI] [PubMed] [Google Scholar]
  • 32.Andersen HR, Nielsen JC, Thomsen PE, Thuesen L, Mortensen PT, Vesterlund T, et al. Long-Term Follow-up of Patients from a Randomised Trial of Atrial Versus Ventricular Pacing for Sick-Sinus Syndrome. Lancet . 1997;350(9086):1210–1216. doi: 10.1016/S0140-6736(97)03425-9. [DOI] [PubMed] [Google Scholar]
  • 33.Connolly SJ, Kerr CR, Gent M, Roberts RS, Yusuf S, Gillis AM, et al. Effects of Physiologic Pacing Versus Ventricular Pacing on the Risk of Stroke and Death Due to Cardiovascular Causes. Canadian Trial of Physiologic Pacing Investigators. N Engl J Med . 2000;342(19):1385–1391. doi: 10.1056/NEJM200005113421902. [DOI] [PubMed] [Google Scholar]
  • 34.Kerr CR, Connolly SJ, Abdollah H, Roberts RS, Gent M, Yusuf S, et al. Canadian Trial of Physiological Pacing: Effects of Physiological Pacing During Long-Term Follow-up. Circulation . 2004;109(3):357–362. doi: 10.1161/01.CIR.0000109490.72104.EE. [DOI] [PubMed] [Google Scholar]
  • 35.Lamas GA, Lee KL, Sweeney MO, Silverman R, Leon A, Yee R, et al. Ventricular Pacing or Dual-Chamber Pacing for Sinus-Node Dysfunction. N Engl J Med. . 2002;346(24):1854–1862. doi: 10.1056/NEJMoa013040. [DOI] [PubMed] [Google Scholar]
  • 36.Healey JS, Toff WD, Lamas GA, Andersen HR, Thorpe KE, Ellenbogen KA, et al. Cardiovascular Outcomes with Atrial-Based Pacing Compared with Ventricular Pacing: Meta-Analysis of Randomized Trials, Using Individual Patient Data. Circulation . 2006;114(1):11–17. doi: 10.1161/CIRCULATIONAHA.105.610303. [DOI] [PubMed] [Google Scholar]
  • 37.Nahlawi M, Waligora M, Spies SM, Bonow RO, Kadish AH, Goldberger JJ. Left Ventricular Function During and After Right Ventricular Pacing. J Am Coll Cardiol . 2004;44(9):1883–1888. doi: 10.1016/j.jacc.2004.06.074. [DOI] [PubMed] [Google Scholar]
  • 38.Murakami Y, Tsuboi N, Inden Y, Yoshida Y, Murohara T, Ihara Z, et al. Difference in Percentage of Ventricular Pacing Between Two Algorithms for Minimizing Ventricular Pacing: Results of the IDEAL RVP (Identify the Best Algorithm for Reducing Unnecessary Right Ventricular Pacing) Study. Europace . 2010;12(1):96–102. doi: 10.1093/europace/eup252. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Ellenbogen KA. Pacing Therapy for Prevention of Atrial Fibrillation. Heart Rhythm . 2007;4(3):84–87. doi: 10.1016/j.hrthm.2006.12.005. [DOI] [PubMed] [Google Scholar]
  • 40.Davies MJ, Anderson RH, Becker AE. The Conduction System of The Heart . London: Butterworth; 1993. [Google Scholar]
  • 41.Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA, 3rd, Freedman RA, Gettes LS, et al. 2012 ACCF/AHA/HRS Focused Update Incorporated into the ACCF/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol . 2013;61(3):6–75. doi: 10.1016/j.jacc.2012.11.007. [DOI] [PubMed] [Google Scholar]
  • 42.Carroz P, Delay D, Girod G. Pseudo-Pacemaker Syndrome in a Young Woman with First-Degree Atrio-Ventricular Block. Europace . 2010;12(4):594–596. doi: 10.1093/europace/eup373. [DOI] [PubMed] [Google Scholar]
  • 43.Brignole M, Deharo JC, Roy L, Menozzi C, Blommaert D, Dabiri L, et al. Syncope Due to Idiopathic Paroxysmal Atrioventricular Block: Long-Term Follow-up of a Distinct Form of Atrioventricular Block. J Am Coll Cardiol . 2011;58(2):167–173. doi: 10.1016/j.jacc.2010.12.045. [DOI] [PubMed] [Google Scholar]
  • 44.Marti-Almor J, Cladellas M, Bazan V, Altaba C, Guijo M, Delclos J, et al. Long-Term Mortality Predictors in Patients with Chronic Bifascicular Block. Europace . 2009;11(9):1201–1207. doi: 10.1093/europace/eup181. [DOI] [PubMed] [Google Scholar]
  • 45.Pinsky WW, Gillette PC, Garson A, Jr, McNamara DG. Diagnosis, Management, and Long-Term Results of Patients with Congenital Complete Atrioventricular Block. Pediatrics . 1982;69(6):728–733. [PubMed] [Google Scholar]
  • 46.Dewey RC, Capeless MA, Levy AM. Use of Ambulatory Electrocardiographic Monitoring to Identify High-Risk Patients with Congenital Complete Heart Block. N Engl J Med . 1987;316(14):835–839. doi: 10.1056/NEJM198704023161403. [DOI] [PubMed] [Google Scholar]
  • 47.Stecker EC, Fendrick AM, Knight BP, Aaronson KD. Prophylactic Pacemaker Use to Allow Beta-Blocker Therapy in Patients with Chronic Heart Failure with Bradycardia. Am Heart J. . 2006;151(4):820–828. doi: 10.1016/j.ahj.2005.06.007. [DOI] [PubMed] [Google Scholar]
  • 48.Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse Effect of Ventricular Pacing on Heart Failure and Atrial Fibrillation Among Patients with Normal Baseline QRS Duration in a Clinical Trial of Pacemaker Therapy for Sinus Node Dysfunction. Circulation . 2003;107(23):2932–2937. doi: 10.1161/01.CIR.0000072769.17295.B1. [DOI] [PubMed] [Google Scholar]
  • 49.Dretzke J, Toff WD, Lip GY, Raftery J, Fry-Smith A, Taylor R. Dual Chamber Versus Single Chamber Ventricular Pacemakers for Sick Sinus Syndrome and Atrioventricular Block. Cochrane Database Syst Rev . 2004;2004(2) doi: 10.1002/14651858.CD003710.pub2. CD003710. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Toff WD, Camm AJ, Skehan JD. Single-Chamber Versus Dual-Chamber Pacing for High-Grade Atrioventricular Block. N Engl J Med . 2005;353(2):145–155. doi: 10.1056/NEJMoa042283. [DOI] [PubMed] [Google Scholar]
  • 51.Ellenbogen KA, Stambler BS, Orav EJ, Sgarbossa EB, Tullo NG, Love CA, et al. Clinical Characteristics of Patients Intolerant to VVIR Pacing. Am J Cardiol . 2000;86(1):59–63. doi: 10.1016/s0002-9149(00)00828-6. [DOI] [PubMed] [Google Scholar]
  • 52.Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-Chamber Pacing or Ventricular Backup Pacing in Patients with an Implantable Defibrillator: The Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA . 2002;288(24):3115–3123. doi: 10.1001/jama.288.24.3115. [DOI] [PubMed] [Google Scholar]
  • 53.Kiehl EL, Makki T, Kumar R, Gumber D, Kwon DH, Rickard JW, et al. Incidence and Predictors of Right Ventricular Pacing-Induced Cardiomyopathy in Patients with Complete Atrioventricular Block and Preserved Left Ventricular Systolic Function. Heart Rhythm . 2016;13(12):2272–2278. doi: 10.1016/j.hrthm.2016.09.027. [DOI] [PubMed] [Google Scholar]
  • 54.Albertsen AE, Mortensen PT, Jensen HK, Poulsen SH, Egeblad H, Nielsen JC. Adverse Effect of Right Ventricular Pacing Prevented by Biventricular Pacing During Long-Term Follow-up: A Randomized Comparison. Eur J Echocardiogr . 2011;12(10):767–772. doi: 10.1093/ejechocard/jer136. [DOI] [PubMed] [Google Scholar]
  • 55.Martinelli M, Filho, Siqueira SF, Costa R, Greco OT, Moreira LF, D’avila A, et al. Conventional Versus Biventricular Pacing in Heart Failure and Bradyarrhythmia: The COMBAT Study. J Card Fail . 2010;16(4):293–300. doi: 10.1016/j.cardfail.2009.12.008. [DOI] [PubMed] [Google Scholar]
  • 56.Curtis AB, Worley SJ, Adamson PB, Chung ES, Niazi I, Sherfesee L, et al. Biventricular Pacing for Atrioventricular Block and Systolic Dysfunction. N Engl J Med . 2013;368(17):1585–1593. doi: 10.1056/NEJMoa1210356. [DOI] [PubMed] [Google Scholar]
  • 57.Fahy GJ, Pinski SL, Miller DP, McCabe N, Pye C, Walsh MJ, et al. Natural History of Isolated Bundle Branch Block. Am J Cardiol . 1996;77(14):1185–1190. doi: 10.1016/s0002-9149(96)00160-9. [DOI] [PubMed] [Google Scholar]
  • 58.Aro AL, Anttonen O, Tikkanen JT, Junttila MJ, Kerola T, Rissanen HA, et al. Intraventricular Conduction Delay in a Standard 12-Lead Electrocardiogram as a Predictor of Mortality in the General Population. Circ Arrhythm Electrophysiol . 2011;4(5):704–710. doi: 10.1161/CIRCEP.111.963561. [DOI] [PubMed] [Google Scholar]
  • 59.McAnulty JH, Rahimtoola SH, Murphy E, DeMots H, Ritzmann L, Kanarek PE, et al. Natural History of “High-Risk” Bundle-Branch Block: Final Report of a Prospective Study. N Engl J Med . 1982;307(3):137–143. doi: 10.1056/NEJM198207153070301. [DOI] [PubMed] [Google Scholar]
  • 60.Haataja P, Anttila I, Nikus K, Eskola M, Huhtala H, Nieminen T, et al. Prognostic Implications of Intraventricular Conduction Delays in a General Population: The Health 2000 Survey. Ann Med . 2015;47(1):74–80. doi: 10.3109/07853890.2014.985704. [DOI] [PubMed] [Google Scholar]
  • 61.Linzer M, Prystowsky EN, Divine GW, Matchar DB, Samsa G, Harrell F, Jr, et al. Predicting the Outcomes of Electrophysiologic Studies of Patients with Unexplained Syncope: Preliminary Validation of a Derived Model. J Gen Intern Med . 1991;6(2):113–120. doi: 10.1007/BF02598305. [DOI] [PubMed] [Google Scholar]
  • 62.Sugrue DD, Gersh BJ, Holmes DR, Jr, Wood DL, Osborn MJ, Hammill SC. Symptomatic “Isolated” Carotid Sinus Hypersensitivity: Natural History and Results of Treatment with Anticholinergic Drugs or Pacemaker. J Am Coll Cardiol . 1986;7(1):158–162. doi: 10.1016/s0735-1097(86)80274-1. [DOI] [PubMed] [Google Scholar]
  • 63.Brito HL, Jr, Hachul DT. Souza OF, Pereira LSM, Maia IG.editors . O Sistema Holter e Outros Métodos nas Arritmias Cardíacas . Rio de Janeiro: Revinter; 2001. Teste de Inclinação na Síncope; pp. 299–313. [Google Scholar]
  • 64.Morillo CA, Camacho ME, Wood MA, Gilligan DM, Ellenbogen KA. Diagnostic Utility of Mechanical, Pharmacological and Orthostatic Stimulation of the Carotid Sinus in Patients with Unexplained Syncope. J Am Coll Cardiol . 1999;34(5):1587–1594. doi: 10.1016/s0735-1097(99)00365-4. [DOI] [PubMed] [Google Scholar]
  • 65.Brignole M, Menozzi C, Lolli G, Bottoni N, Gaggioli G. Long-Term Outcome of Paced and Nonpaced Patients with Severe Carotid Sinus Syndrome. Am J Cardiol . 1992;69(12):1039–1043. doi: 10.1016/0002-9149(92)90860-2. [DOI] [PubMed] [Google Scholar]
  • 66.Claesson JE, Kristensson BE, Edvardsson N, Währborg P. Less Syncope and Milder Symptoms in Patients Treated with Pacing for Induced Cardioinhibitory Carotid Sinus Syndrome: A Randomized Study. Europace . 2007;9(10):932–936. doi: 10.1093/europace/eum180. [DOI] [PubMed] [Google Scholar]
  • 67.Sugrue DD, Gersh BJ, Holmes DR, Jr, Wood DL, Osborn MJ, Hammill SC. Symptomatic “Isolated” Carotid Sinus Hypersensitivity: Natural History and Results of Treatment with Anticholinergic Drugs or Pacemaker. J Am Coll Cardiol . 1986;7(1):158–162. doi: 10.1016/s0735-1097(86)80274-1. [DOI] [PubMed] [Google Scholar]
  • 68.Brignole M, Menozzi C. The Natural History of Carotid Sinus Syncope and the Effect of Cardiac Pacing. Europace . 2011;13(4):462–464. doi: 10.1093/europace/euq516. [DOI] [PubMed] [Google Scholar]
  • 69.Kenny RA, Richardson DA, Steen N, Bexton RS, Shaw FE, Bond J. Carotid Sinus Syndrome: A Modifiable Risk Factor for Nonaccidental Falls in Older Adults (SAFE PACE) J Am Coll Cardiol . 2001;38(5):1491–1496. doi: 10.1016/s0735-1097(01)01537-6. [DOI] [PubMed] [Google Scholar]
  • 70.Maggi R, Menozzi C, Brignole M, Podoleanu C, Iori M, Sutton R, et al. Cardioinhibitory Carotid Sinus Hypersensitivity Predicts an Asystolic Mechanism of Spontaneous Neurally Mediated Syncope. Europace . 2007;9(8):563–567. doi: 10.1093/europace/eum092. [DOI] [PubMed] [Google Scholar]
  • 71.Shen WK, Sheldon RS, Benditt DG, Cohen MI, Forman DE, Goldberger ZD, et al. 2017 ACC/AHA/HRS Guideline for the Evaluation and Management of Patients With Syncope: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation . 2017;136(5):60–122. doi: 10.1161/CIR.0000000000000499. [DOI] [PubMed] [Google Scholar]
  • 72.Madigan NP, Flaker GC, Curtis JJ, Reid J, Mueller KJ, Murphy TJ. Carotid Sinus Hypersensitivity: Beneficial Effects of Dual-Chamber Pacing. Am J Cardiol . 1984;53(8):1034–1040. doi: 10.1016/0002-9149(84)90632-5. [DOI] [PubMed] [Google Scholar]
  • 73.Brignole M, Sartore B, Barra M, Menozzi C, Lolli G. Is DDD Superior to VVI Pacing in Mixed Carotid Sinus Syndrome? An Acute and Medium-Term Study. Pacing Clin Electrophysiol . 1988;11(11):1902–1910. doi: 10.1111/j.1540-8159.1988.tb06327.x. [DOI] [PubMed] [Google Scholar]
  • 74.Sutton R. Pacing in Patients with Carotid Sinus and Vasovagal Syndromes. Pacing Clin Electrophysiol . 1989;12(7):1260–1263. doi: 10.1111/j.1540-8159.1989.tb01982.x. [DOI] [Google Scholar]
  • 75.McLeod CJ, Trusty JM, Jenkins SM, Rea RF, Cha YM, Espinosa RA, et al. Method of Pacing Does Not Affect the Recurrence of Syncope in Carotid Sinus Syndrome. Pacing Clin Electrophysiol . 2012;35(7):827–833. doi: 10.1111/j.1540-8159.2012.03375.x. [DOI] [PubMed] [Google Scholar]
  • 76.Morley CA, Perrins EJ, Grant P, Chan SL, McBrien DJ, Sutton R. Carotid Sinus Syncope Treated by Pacing. Analysis of Persistent Symptoms and Role of Atrioventricular Sequential Pacing. Br Heart J . 1982;47(5):411–418. doi: 10.1136/hrt.47.5.411. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Sutton R, Peterson M, Brignole M, Raviele A, Menozzi C, Giani P. Proposed Classification for Tilt Induced Vasovagal Syncope. Eur J Cardiac Pacing Electrophysiol . 1992;2(3):180–183. [Google Scholar]
  • 78.Connolly SJ, Sheldon R, Roberts RS, Gent M. The North American Vasovagal Pacemaker Study (VPS). A Randomized Trial of Permanent Cardiac Pacing for the Prevention of Vasovagal Syncope. J Am Coll Cardiol . 1999;33(1):16–20. doi: 10.1016/s0735-1097(98)00549-x. [DOI] [PubMed] [Google Scholar]
  • 79.Ammirati F, Colivicchi F, Santini M. Permanent Cardiac Pacing Versus Medical Treatment for the Prevention of Recurrent Vasovagal Syncope: A Multicenter, Randomized, Controlled Trial. Circulation . 2001;104(1):52–57. doi: 10.1161/hc2601.091708. [DOI] [PubMed] [Google Scholar]
  • 80.Sutton R, Brignole M, Menozzi C, Raviele A, Alboni P, Giani P, Moya A. Dual-Chamber Pacing in the Treatment of Neurally Mediated Tilt-Positive Cardioinhibitory Syncope: Pacemaker Versus no Therapy: A Multicenter Randomized Study. The Vasovagal Syncope International Study (VASIS) Investigators. Circulation . 2000;102(3):294–299. doi: 10.1161/01.cir.102.3.294. [DOI] [PubMed] [Google Scholar]
  • 81.Connolly SJ, Sheldon R, Thorpe KE, Roberts RS, Ellenbogen KA, Wilkoff BL, et al. Pacemaker Therapy for Prevention of Syncope in Patients with Recurrent Severe Vasovagal Syncope: Second Vasovagal Pacemaker Study (VPS II): A Randomized Trial. JAMA . 2003;289(17):2224–2229. doi: 10.1001/jama.289.17.2224. [DOI] [PubMed] [Google Scholar]
  • 82.Brignole M, Menozzi C, Moya A, Andresen D, Blanc JJ, Krahn AD, et al. Pacemaker Therapy in Patients with Neurally Mediated Syncope and Documented Asystole: Third International Study on Syncope of Uncertain Etiology (ISSUE-3): A Randomized Trial. Circulation . 2012;125(21):2566–2571. doi: 10.1161/CIRCULATIONAHA.111.082313. [DOI] [PubMed] [Google Scholar]
  • 83.Palmisano P, Zaccaria M, Luzzi G, Nacci F, Anaclerio M, Favale S. Closed-Loop Cardiac Pacing vs. Conventional Dual-Chamber Pacing with Specialized Sensing and Pacing Algorithms for Syncope Prevention in Patients with Refractory Vasovagal Syncope: Results of a Long-Term Follow-Up. Europace . 2012;14(7):1038–1043. doi: 10.1093/europace/eur419. [DOI] [PubMed] [Google Scholar]
  • 84.Maron MS, Olivotto I, Betocchi S, Casey SA, Lesser JR, Losi MA, et al. Effect of Left Ventricular Outflow Tract Obstruction on Clinical Outcome in Hypertrophic Cardiomyopathy. N Engl J Med . 2003;348(4):295–303. doi: 10.1056/NEJMoa021332. [DOI] [PubMed] [Google Scholar]
  • 85.Maron BJ. Hypertrophic Cardiomyopathy: A Systematic Review. JAMA . 2002;287(10):1308–1320. doi: 10.1001/jama.287.10.1308. [DOI] [PubMed] [Google Scholar]
  • 86.Nishimura RA, Holmes DR., Jr Clinical Practice. Hypertrophic Obstructive Cardiomyopathy. N Engl J Med . 2004;350(13):1320–1327. doi: 10.1056/NEJMcp030779. [DOI] [PubMed] [Google Scholar]
  • 87.Nishimura RA, Hayes DL, Ilstrup DM, Holmes DR, Jr, Tajik AJ. Effect of Dual-Chamber Pacing on Systolic and Diastolic Function in Patients with Hypertrophic Cardiomyopathy. Acute Doppler Echocardiographic and Catheterization Hemodynamic Study. J Am Coll Cardiol . 1996;27(2):421–430. doi: 10.1016/0735-1097(95)00445-9. [DOI] [PubMed] [Google Scholar]
  • 88.Betocchi S, Losi MA, Piscione F, Boccalatte M, Pace L, Golino P, et al. Effects of Dual-Chamber Pacing in Hypertrophic Cardiomyopathy on Left Ventricular Outflow Tract Obstruction and on Diastolic Function. Am J Cardiol . 1996;77(7):498–502. doi: 10.1016/s0002-9149(97)89344-7. [DOI] [PubMed] [Google Scholar]
  • 89.Slade AK, Sadoul N, Shapiro L, Chojnowska L, Simon JP, Saumarez RC, et al. DDD Pacing in Hypertrophic Cardiomyopathy: A Multicentre Clinical Experience. Heart . 1996;75(1):44–49. doi: 10.1136/hrt.75.1.44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Kappenberger L, Linde C, Daubert C, McKenna W, Meisel E, Sadoul N, et al. Pacing in Hypertrophic Obstructive Cardiomyopathy. A Randomized Crossover Study. PIC Study Group. Eur Heart J . 1997;18(8):1249–1256. doi: 10.1093/oxfordjournals.eurheartj.a015435. [DOI] [PubMed] [Google Scholar]
  • 91.Nishimura RA, Trusty JM, Hayes DL, Ilstrup DM, Larson DR, Hayes SN, et al. Dual-Chamber Pacing for Hypertrophic Cardiomyopathy: A Randomized, Double-Blind, Crossover Trial. J Am Coll Cardiol . 1997;29(2):435–441. doi: 10.1016/s0735-1097(96)00473-1. [DOI] [PubMed] [Google Scholar]
  • 92.Mickelsen S, Bathina M, Hsu P, Holmes J, Kusumoto FM. Doppler Evaluation of the Descending Aorta in Patients with Hypertrophic Cardiomyopathy: Potential for Assessing the Functional Significance of Outflow Tract Gradients and for Optimizing Pacemaker Function. J Interv Card Electrophysiol . 2004;11(1):47–53. doi: 10.1023/B:JICE.0000035929.84238.2f. [DOI] [PubMed] [Google Scholar]
  • 93.Ommen SR, Nishimura RA, Squires RW, Schaff HV, Danielson GK, Tajik AJ. Comparison of Dual-Chamber Pacing Versus Septal Myectomy for the Treatment of Patients with Hypertrophic Obstructive Cardiomyopathy: A Comparison of Objective Hemodynamic and Exercise End Points. J Am Coll Cardiol . 1999;34(1):191–196. doi: 10.1016/s0735-1097(99)00173-4. [DOI] [PubMed] [Google Scholar]
  • 94.Maron BJ, Nishimura RA, McKenna WJ, Rakowski H, Josephson ME, Kieval RS. Assessment of Permanent Dual-Chamber Pacing as a Treatment for Drug-Refractory Symptomatic Patients with Obstructive Hypertrophic Cardiomyopathy. A Randomized, Double-Blind, Crossover Study (M-PATHY) Circulation . 1999;99(22):2927–2933. doi: 10.1161/01.cir.99.22.2927. [DOI] [PubMed] [Google Scholar]
  • 95.Qintar M, Morad A, Alhawasli H, Shorbaji K, Firwana B, Essali A, et al. Pacing for Drug-Refractory or Drug-Intolerant Hypertrophic Cardiomyopathy. CD008523 Cochrane Database Syst Rev . 2012;2012(5) doi: 10.1002/14651858.CD008523.pub2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96.Topilski I, Sherez J, Keren G, Copperman I. Long-Term Effects of Dual-Chamber Pacing with Periodic Echocardiographic Evaluation of Optimal Atrioventricular Delay in Patients with Hypertrophic Cardiomyopathy >50 Years of Age. Am J Cardiol . 2006;97(12):1769–1775. doi: 10.1016/j.amjcard.2006.01.040. [DOI] [PubMed] [Google Scholar]
  • 97.Perloff JK. O’Rourke RA.editor . Current Problems in Cardiology . Chicago: Yearbook; 1986. The Heart in Neuromuscular Disease; pp. 513–557. [PubMed] [Google Scholar]
  • 98.Lazarus A, Varin J, Babuty D, Anselme F, Coste J, Duboc D. Long-Term Follow-up of Arrhythmias in Patients with Myotonic Dystrophy Treated by Pacing: A Multicenter Diagnostic Pacemaker Study. J Am Coll Cardiol . 2002;40(9):1645–1652. doi: 10.1016/s0735-1097(02)02339-2. [DOI] [PubMed] [Google Scholar]
  • 99.Groh WJ, Groh MR, Saha C, Kincaid JC, Simmons Z, Ciafaloni E, et al. Electrocardiographic Abnormalities and Sudden Death in Myotonic Dystrophy Type 1. N Engl J Med . 2008;358(25):2688–2697. doi: 10.1056/NEJMoa062800. [DOI] [PubMed] [Google Scholar]
  • 100.Mehra R, Benjamin EJ, Shahar E, Gottlieb DJ, Nawabit R, Kirchner HL, et al. Association of Nocturnal Arrhythmias with Sleep-Disordered Breathing: The Sleep Heart Health Study. Am J Respir Crit Care Med . 2006;173(8):910–916. doi: 10.1164/rccm.200509-1442OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 101.Shepard JW, Jr, Garrison MW, Grither DA, Dolan GF. Relationship of Ventricular Ectopy to Oxyhemoglobin Desaturation in Patients with Obstructive Sleep Apnea. Chest . 1985;88(3):335–340. doi: 10.1378/chest.88.3.335. [DOI] [PubMed] [Google Scholar]
  • 102.Becker H, Brandenburg U, Peter JH, Von Wichert P. Reversal of Sinus Arrest and Atrioventricular Conduction Block in Patients with Sleep Apnea During Nasal Continuous Positive Airway Pressure. Am J Respir Crit Care Med . 1995;151(1):215–218. doi: 10.1164/ajrccm.151.1.7812557. [DOI] [PubMed] [Google Scholar]
  • 103.Pachon MJC, Pachon MEI, Lobo TJ, Pachon MJC, Pachon MZ, Vargas RN, Manrique RM, Jatene AD. Syncopal High-Degree AV Block Treated with Catheter RF Ablation Without Pacemaker Implantation. Pacing Clin Electrophysiol . 2006;29(3):318–322. doi: 10.1111/j.1540-8159.2006.00340.x. [DOI] [PubMed] [Google Scholar]
  • 104.Waddell-Smith KE, Skinner JR. Update on the Diagnosis and Management of Familial Long QT Syndrome. Heart Lung Circ . 2016;25(8):769–776. doi: 10.1016/j.hlc.2016.01.020. [DOI] [PubMed] [Google Scholar]
  • 105.Mönnig G, Eckardt L, Wedekind H, Haverkamp W, Gerss J, Milberg P, et al. Electrocardiographic Risk Stratification in Families with Congenital Long QT Syndrome. Eur Heart J . 2006;27(17):2074–2080. doi: 10.1093/eurheartj/ehl159. [DOI] [PubMed] [Google Scholar]
  • 106.Gil MAC, Rubira JCG. Who Was the Creator of Bazett’s Formula? Rev Esp Cardiol . 2008;61(8):896–897. [PubMed] [Google Scholar]
  • 107.Earle N, Crawford J, Smith W, Hayes I, Shelling A, Hood M, et al. Community Detection of Long QT Syndrome with a Clinical Registry: An Alternative to ECG Screening Programs? Heart Rhythm . 2013;10(2):233–238. doi: 10.1016/j.hrthm.2012.10.043. [DOI] [PubMed] [Google Scholar]
  • 108.Sakaguchi T, Shimizu W, Itoh H, Noda T, Miyamoto Y, Nagaoka I, et al. Age- and Genotype-Specific Triggers for Life-Threatening Arrhythmia in the Genotyped Long QT Syndrome. J Cardiovasc Electrophysiol . 2008;19(8):794–799. doi: 10.1111/j.1540-8167.2008.01138.x. [DOI] [PubMed] [Google Scholar]
  • 109.Chockalingam P, Crotti L, Girardengo G, Johnson JN, Harris KM, van der Heijden JF, et al. Not All Beta-Blockers are Equal in the Management of Long QT Syndrome Types 1 and 2: Higher Recurrence of Events Under Metoprolol. J Am Coll Cardiol . 2012;60(20):2092–2099. doi: 10.1016/j.jacc.2012.07.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 110.Priori SG, Blomström-Lundqvist C, Mazzanti A, Blom N, Borggrefe M, et al. 2015 ESC Guidelines for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death. The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology. G Ital Cardiol . 2016;17(2):108–170. doi: 10.1714/2174.23496. [DOI] [PubMed] [Google Scholar]
  • 111.DiBiase A, Tse TM, Schnittger I, Wexler L, Stinson EB, Valantine HA. Frequency and Mechanism of Bradycardia in Cardiac Transplant Recipients and Need for Pacemakers. Am J Cardiol . 1991;67(16):1385–1389. doi: 10.1016/0002-9149(91)90469-2. [DOI] [PubMed] [Google Scholar]
  • 112.Woo GW, Schofield RS, Pauly DF, Hill JA, Conti JB, Kron J, et al. Incidence, Predictors, and Outcomes of Cardiac Pacing After Cardiac Transplantation: An 11-year Retrospective Analysis. Transplantation . 2008;85(8):1216–1218. doi: 10.1097/TP.0b013e31816b677c. [DOI] [PubMed] [Google Scholar]
  • 113.Weiss ES, Nwakanma LU, Russell SB, Conte JV, Shah AS. Outcomes in Bicaval Versus Biatrial Techniques in Heart Transplantation: An Analysis of the UNOS Database. J Heart Lung Transplant . 2008;27(2):178–183. doi: 10.1016/j.healun.2007.11.003. [DOI] [PubMed] [Google Scholar]
  • 114.Melton IC, Gilligan DM, Wood MA, Ellenbogen KA. Optimal Cardiac Pacing After Heart Transplantation. Pacing Clin Electrophysiol . 1999;22(10):1510–1527. doi: 10.1111/j.1540-8159.1999.tb00356.x. [DOI] [PubMed] [Google Scholar]
  • 115.Holt ND, McComb JM. Cardiac Transplantation and Pacemakers: When and What to Implant. Card Electrophysiol Rev . 2002;6(1-2):140–151. doi: 10.1023/a:1017972129833. [DOI] [PubMed] [Google Scholar]
  • 116.Raghavan C, Maloney JD, Nitta J, Lowry RW, Saliba WI, Cocanougher B, et al. Long-Term Follow-up of Heart Transplant Recipients Requiring Permanent Pacemakers. J Heart Lung Transplant . 1995;14(6):1081–1089. [PubMed] [Google Scholar]
  • 117.Scott CD, McComb JM, Dark JH, Bexton RS. Permanent Pacing After Cardiac Transplantation. Br Heart J . 1993;69(5):399–403. doi: 10.1136/hrt.69.5.399. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 118.Nielsen JC, Thomsen PE, Højberg S, Møller M, Riahi S, Dalsgaard D, et al. Atrial Fibrillation in Patients with Sick Sinus Syndrome: The Association with PQ-Interval and Percentage of Ventricular Pacing. Europace . 2012;14(5):682–689. doi: 10.1093/europace/eur365. [DOI] [PubMed] [Google Scholar]
  • 119.Sweeney MO, Bank AJ, Nsah E, Koullick M, Zeng QC, Hettrick D, et al. Minimizing Ventricular Pacing to Reduce Atrial Fibrillation in Sinus-Node Disease. N Engl J Med . 2007;357(10):1000–1008. doi: 10.1056/NEJMoa071880. [DOI] [PubMed] [Google Scholar]
  • 120.Cheng S, Keyes MJ, Larson MG, McCabe EL, Newton-Cheh C, Levy D, et al. Long-Term Outcomes in Individuals with Prolonged PR Interval or First-Degree Atrioventricular Block. JAMA . 2009;301(24):2571–2577. doi: 10.1001/jama.2009.888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 121.Lamas GA, Knight JD, Sweeney MO, Mianulli M, Jorapur V, Khalighi K, et al. Impact of Rate-Modulated Pacing on Quality of Life and Exercise Capacity--Evidence from the Advanced Elements of Pacing Randomized Controlled Trial (ADEPT) Heart Rhythm . 2007;4(9):1125–1132. doi: 10.1016/j.hrthm.2007.05.021. [DOI] [PubMed] [Google Scholar]
  • 122.Heckman L, Vijayaraman P, Luermans J, Stipdonk AMW, Salden F, Maass AH, et al. Novel Bradycardia Pacing Strategies. Heart . 2020;106(24):1883–1889. doi: 10.1136/heartjnl-2020-316849. [DOI] [PubMed] [Google Scholar]
  • 123.Kaye G. The Desire for Physiological Pacing: Are We There Yet? J Cardiovasc Electrophysiol . 2019;30(12):3025–3038. doi: 10.1111/jce.14248. [DOI] [PubMed] [Google Scholar]
  • 124.Deshmukh P, Casavant DA, Romanyshyn M, Anderson K. Permanent, Direct His-Bundle Pacing: A Novel Approach to Cardiac Pacing in Patients with Normal His-Purkinje Activation. Circulation . 2000;101(8):869–877. doi: 10.1161/01.cir.101.8.869. [DOI] [PubMed] [Google Scholar]
  • 125.Sharma PS, Dandamudi G, Naperkowski A, Oren JW, Storm RH, Ellenbogen KA, et al. Permanent His-Bundle Pacing is Feasible, Safe, and Superior to Right Ventricular Pacing in Routine Clinical Practice. Heart Rhythm . 2015;12(2):305–312. doi: 10.1016/j.hrthm.2014.10.021. [DOI] [PubMed] [Google Scholar]
  • 126.Muthumala A, Vijayaraman P. Clinical Outcomes of His-Purkinje Conduction System Pacing. Pacing Clin Electrophysiol . 2021;44(1):5–14. doi: 10.1111/pace.14050. [DOI] [PubMed] [Google Scholar]
  • 127.Mafi-Rad M, Luermans JG, Blaauw Y, Janssen M, Crijns HJ, Prinzen FW, et al. Feasibility and Acute Hemodynamic Effect of Left Ventricular Septal Pacing by Transvenous Approach Through the Interventricular Septum. Circ Arrhythm Electrophysiol . 2016;9(3):e003344. doi: 10.1161/CIRCEP.115.003344. [DOI] [PubMed] [Google Scholar]
  • 128.Huang W, Su L, Wu S, Xu L, Xiao F, Zhou X, et al. A Novel Pacing Strategy with Low and Stable Output: Pacing the Left Bundle Branch Immediately Beyond the Conduction Block. Can J Cardiol . 2017;33(12):1736.e1–1736.e3. doi: 10.1016/j.cjca.2017.09.013. [DOI] [PubMed] [Google Scholar]
  • 129.Huang W, Chen X, Su L, Wu S, Xia X, Vijayaraman P. A Beginner’s Guide to Permanent Left Bundle Branch Pacing. Heart Rhythm . 2019;16(12):1791–1796. doi: 10.1016/j.hrthm.2019.06.016. [DOI] [PubMed] [Google Scholar]
  • 130.Hua W, Fan X, Li X, Niu H, Gu M, Ning X, et al. Comparison of Left Bundle Branch and His Bundle Pacing in Bradycardia Patients. JACC Clin Electrophysiol . 2020;6(10):1291–1299. doi: 10.1016/j.jacep.2020.05.008. [DOI] [PubMed] [Google Scholar]
  • 131.Hu Y, Li H, Gu M, Hua W, Niu H, Zhang N, et al. Comparison Between His-Bundle Pacing and Left Bundle Branch Pacing in Patients with Atrioventricular Block. J Interv Card Electrophysiol . 2021;62(1):63–73. doi: 10.1007/s10840-020-00869-w. [DOI] [PubMed] [Google Scholar]
  • 132.Sun JY, Sha YQ, Sun QY, Qiu Y, Shao B, Ni YH, et al. The Long-Term Therapeutic Effects of His-Purkinje System Pacing on Bradycardia and Cardiac Conduction Dysfunction Compared with Right Ventricular Pacing: A systematic Review and Meta-Analysis. J Cardiovasc Electrophysiol . 2020;31(5):1202–1210. doi: 10.1111/jce.14445. [DOI] [PubMed] [Google Scholar]
  • 133.Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC. Complications After Cardiac Implantable Electronic Device Implantations: An Analysis of a Complete, Nationwide Cohort in Denmark. Eur Heart J . 2014;35(18):1186–1194. doi: 10.1093/eurheartj/eht511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 134.Clémenty N, Fernandes J, Carion PL, Léotoing L, Lamarsalle L, Wilquin-Bequet F, et al. Pacemaker Complications and Costs: A Nationwide Economic Study. J Med Econ . 2019;22(11):1171–1178. doi: 10.1080/13696998.2019.1652186. [DOI] [PubMed] [Google Scholar]
  • 135.Özcan C, Raunsø J, Lamberts M, Køber L, Lindhardt TB, Bruun NE, et al. Infective Endocarditis and Risk of Death After Cardiac Implantable Electronic Device Implantation: A Nationwide Cohort Study. Europace . 2017;19(6):1007–1014. doi: 10.1093/europace/euw404. [DOI] [PubMed] [Google Scholar]
  • 136.Greenspon AJ, Patel JD, Lau E, Ochoa JA, Frisch DR, Ho RT, et al. 16-year Trends in the Infection Burden for Pacemakers and Implantable Cardioverter-Defibrillators in the United States 1993 to 2008. J Am Coll Cardiol . 2011;58(10):1001–1006. doi: 10.1016/j.jacc.2011.04.033. [DOI] [PubMed] [Google Scholar]
  • 137.El-Chami MF, Roberts PR, Kypta A, Omdahl P, Bonner MD, Kowal RC, et al. How to Implant a Leadless Pacemaker With a Tine-Based Fixation. J Cardiovasc Electrophysiol . 2016;27(12):1495–1501. doi: 10.1111/jce.13092. [DOI] [PubMed] [Google Scholar]
  • 138.Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A Leadless Intracardiac Transcatheter Pacing System. N Engl J Med . 2016;374(6):533–541. doi: 10.1056/NEJMoa1511643. [DOI] [PubMed] [Google Scholar]
  • 139.Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al. Long-Term Performance of a Transcatheter Pacing System: 12-Month Results from the Micra Transcatheter Pacing Study. Heart Rhythm . 2017;14(5):702–709. doi: 10.1016/j.hrthm.2017.01.035. [DOI] [PubMed] [Google Scholar]
  • 140.Tjong FVY, Beurskens NEG, Groot JR, Waweru C, Liu S, Ritter P, et al. Health-Related Quality of Life Impact of a Transcatheter Pacing System. J Cardiovasc Electrophysiol . 2018;29(12):1697–1704. doi: 10.1111/jce.13726. [DOI] [PubMed] [Google Scholar]
  • 141.Roberts PR, Clementy N, Al Samadi F, Garweg C, Martinez-Sande JL, Iacopino S, et al. A Leadless Pacemaker in the Real-World Setting: The Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm . 2017;14(9):1375–1379. doi: 10.1016/j.hrthm.2017.05.017. [DOI] [PubMed] [Google Scholar]
  • 142.Vaidya VR, Dai M, Asirvatham SJ, Rea RF, Thome TM, Srivathsan K, et al. Real-World Experience with Leadless Cardiac Pacing. Pacing Clin Electrophysiol . 2019;42(3):366–373. doi: 10.1111/pace.13601. [DOI] [PubMed] [Google Scholar]
  • 143.Martínez-Sande JL, García-Seara J, Rodríguez-Mañero M, Fernández-López XA, González-Melchor L, Redondo-Diéguez A, et al. The Micra Leadless Transcatheter Pacemaker. Implantation and Mid-term Follow-up Results in a Single Center. Rev Esp Cardiol . 2017;70(4):275–281. doi: 10.1016/j.rec.2016.11.027. [DOI] [PubMed] [Google Scholar]
  • 144.Valiton V, Graf D, Pruvot E, Carroz P, Fromer M, Bisch L, et al. Leadless Pacing Using the Transcatheter Pacing System (Micra TPS) in the Real World: Initial Swiss Experience from the Romandie Region. Europace . 2019;21(2):275–280. doi: 10.1093/europace/euy195. [DOI] [PubMed] [Google Scholar]
  • 145.Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, et al. Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker. N Engl J Med . 2015;373(12):1125–1135. doi: 10.1056/NEJMoa1507192. [DOI] [PubMed] [Google Scholar]
  • 146.Sperzel J, Defaye P, Delnoy PP, Garcia Guerrero JJ, Knops RE, Tondo C, et al. Primary safety results from the LEADLESS Observational Study. Europace . 2018;20(9):1491–1497. doi: 10.1093/europace/eux359. [DOI] [PubMed] [Google Scholar]
  • 147.Lakkireddy D, Knops R, Atwater B, Neuzil P, Ip J, Gonzalez E, et al. A Worldwide Experience of the Management of Battery Failures and Chronic Device Retrieval of the Nanostim Leadless Pacemaker. Heart Rhythm . 2017;14(12):1756–1763. doi: 10.1016/j.hrthm.2017.07.004. [DOI] [PubMed] [Google Scholar]
  • 148.Salaun E, Tovmassian L, Simonnet B, Giorgi R, Franceschi F, Koutbi-Franceschi L, et al. Right Ventricular and Tricuspid Valve Function in Patients Chronically Implanted with Leadless Pacemakers. Europace . 2018;20(5):823–828. doi: 10.1093/europace/eux101. [DOI] [PubMed] [Google Scholar]
  • 149.Beurskens NEG, Tjong FVY, Bruin-Bon RHA, Dasselaar KJ, Kuijt WJ, Wilde AAM, et al. Impact of Leadless Pacemaker Therapy on Cardiac and Atrioventricular Valve Function Through 12 Months of Follow-Up. Circ Arrhythm Electrophysiol . 2019;12(5):e007124. doi: 10.1161/CIRCEP.118.007124. [DOI] [PubMed] [Google Scholar]
  • 150.Afzal MR, Daoud EG, Cunnane R, Mulpuru SK, Koay A, Hussain A, et al. Techniques for Successful Early Retrieval of the Micra Transcatheter Pacing System: A Worldwide Experience. Heart Rhythm . 2018;15(6):841–846. doi: 10.1016/j.hrthm.2018.02.008. [DOI] [PubMed] [Google Scholar]
  • 151.Kiani S, Merchant FM, El-Chami MF. Extraction of a 4-Year-Old Leadless Pacemaker with a Tine-Based Fixation. HeartRhythm Case Rep . 2019;5(8):424–425. doi: 10.1016/j.hrcr.2019.05.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 152.Chen K, Zheng X, Dai Y, Wang H, Tang Y, Lan T, et al. Multiple Leadless Pacemakers Implanted in the Right Ventricle of Swine. Europace . 2016;18(11):1748–1752. doi: 10.1093/europace/euv418. [DOI] [PubMed] [Google Scholar]
  • 153.Omdahl P, Eggen MD, Bonner MD, Iaizzo PA, Wika K. Right Ventricular Anatomy Can Accommodate Multiple Micra Transcatheter Pacemakers. Pacing Clin Electrophysiol . 2016;39(4):393–397. doi: 10.1111/pace.12804. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 154.Boveda S, Lenarczyk R, Haugaa KH, Iliodromitis K, Finlay M, Lane D, et al. Use of Leadless Pacemakers in Europe: Results of the European Heart Rhythm Association survey. Europace . 2018;20(3):555–559. doi: 10.1093/europace/eux381. [DOI] [PubMed] [Google Scholar]
  • 155.Beurskens NEG, Tjong FVY, Dasselaar KJ, Kuijt WJ, Wilde AAM, Knops RE. Leadless Pacemaker Implantation After Explantation of Infected Conventional Pacemaker Systems: A Viable Solution? Heart Rhythm . 2019;16(1):66–71. doi: 10.1016/j.hrthm.2018.07.006. [DOI] [PubMed] [Google Scholar]
  • 156.El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless Pacemaker Implant in Patients with Pre-Existing Infections: Results from the Micra postapproval Registry. J Cardiovasc Electrophysiol . 2019;30(4):569–574. doi: 10.1111/jce.13851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 157.El-Chami MF, Clementy N, Garweg C, Omar R, Duray GZ, Gornick CC, et al. Leadless Pacemaker Implantation in Hemodialysis Patients: Experience With the Micra Transcatheter Pacemaker. JACC Clin Electrophysiol . 2019;5(2):162–170. doi: 10.1016/j.jacep.2018.12.008. [DOI] [PubMed] [Google Scholar]
  • 158.Tjong FV, Reddy VY. Permanent Leadless Cardiac Pacemaker Therapy: A Comprehensive Review. Circulation . 2017;135(15):1458–1470. doi: 10.1161/CIRCULATIONAHA.116.025037. [DOI] [PubMed] [Google Scholar]
  • 159.Lloyd MS, El-Chami MF, Nilsson KR, Jr, Cantillon DJ. Transcatheter/leadless pacing. Heart Rhythm . 2018;15(4):624–628. doi: 10.1016/j.hrthm.2017.12.004. [DOI] [PubMed] [Google Scholar]
  • 160.Chung ES, Leon AR, Tavazzi L, Sun JP, Nihoyannopoulos P, Merlino J, et al. Results of the Predictors of Response to CRT (PROSPECT) trial. Circulation . 2008;117(20):2608–2616. doi: 10.1161/CIRCULATIONAHA.107.743120. [DOI] [PubMed] [Google Scholar]
  • 161.Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T, et al. Cardiac-Resynchronization Therapy with or Without an Implantable Defibrillator in Advanced Chronic Heart Failure. N Engl J Med . 2004;350(21):2140–2150. doi: 10.1056/NEJMoa032423. [DOI] [PubMed] [Google Scholar]
  • 162.Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L, et al. The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart Failure. N Engl J Med . 2005;352(15):1539–1549. doi: 10.1056/NEJMoa050496. [DOI] [PubMed] [Google Scholar]
  • 163.Freemantle N, Tharmanathan P, Calvert MJ, Abraham WT, Ghosh J, Cleland JG. Cardiac Resynchronisation for Patients with Heart Failure due to Left Ventricular Systolic Dysfunction -- A Systematic Review and Meta-Analysis. Eur J Heart Fail . 2006;8(4):433–440. doi: 10.1016/j.ejheart.2005.11.014. [DOI] [PubMed] [Google Scholar]
  • 164.Martinelli M, Filho, Zimerman LI, Lorga AM, Vasconcelos JTM, Rassi A., Jr Guidelines for Implantable Electronic Cardiac Devices of the Brazilian Society of Cardiology. Arq Bras Cardiol . 2007;89(6):210–238. [Google Scholar]
  • 165.Moss AJ, Hall WJ, Cannom DS, Klein H, Brown MW, Daubert JP, et al. Cardiac-Resynchronization Therapy for the Prevention of Heart-Failure Events. N Engl J Med . 2009;361(14):1329–1338. doi: 10.1056/NEJMoa0906431. [DOI] [PubMed] [Google Scholar]
  • 166.Linde C, Abraham WT, Gold MR, St Sutton MJ, Ghio S, Daubert C, et al. Randomized Trial of Cardiac Resynchronization in Mildly Symptomatic Heart Failure Patients and in Asymptomatic Patients with Left Ventricular Dysfunction and Previous Heart Failure Symptoms. J Am Coll Cardiol . 2008;52(23):1834–1843. doi: 10.1016/j.jacc.2008.08.027. [DOI] [PubMed] [Google Scholar]
  • 167.Tang AS, Wells GA, Talajic M, Arnold MO, Sheldon R, Conolly S, et al. Cardiac-Resynchronization Therapy for Mild-to-Moderate Heart Failure. N Engl J Med . 2010;363(25):2385–2395. doi: 10.1056/NEJMoa1009540. [DOI] [PubMed] [Google Scholar]
  • 168.Wells G, Parkash R, Healey JS, Talajic M, Arnold JM, Sullivan S, et al. Cardiac Resynchronization Therapy: A Meta-Analysis of Randomized Controlled Trials. CMAJ . 2011;183(4):421–429. doi: 10.1503/cmaj.101685. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 169.Abraham WT, Young JB, León AR, Adler S, Bank AJ, Hall SA, et al. Effects of Cardiac Resynchronization on Disease Progression in Patients with Left Ventricular Systolic Dysfunction, an Indication for an Implantable Cardioverter-Defibrillator, and Mildly Symptomatic Chronic Heart Failure. Circulation . 2004;110(18):2864–2868. doi: 10.1161/01.CIR.0000146336.92331. [DOI] [PubMed] [Google Scholar]
  • 170.Higgins SL, Hummel JD, Niazi IK, Giudici MC, Worley SJ, Saxon LA, et al. Cardiac Resynchronization Therapy for the Treatment of Heart Failure in Patients with Intraventricular Conduction Delay and Malignant Ventricular Tachyarrhythmias. J Am Coll Cardiol . 2003;42(8):1454–1459. doi: 10.1016/s0735-1097(03)01042-8. [DOI] [PubMed] [Google Scholar]
  • 171.Adabag S, Roukoz H, Anand IS, Moss AJ. Cardiac Resynchronization Therapy in Patients with Minimal Heart Failure: A Systematic Review and Meta-Analysis. J Am Coll Cardiol . 2011;58(9):935–941. doi: 10.1016/j.jacc.2011.05.022. [DOI] [PubMed] [Google Scholar]
  • 172.Ruschitzka F, Abraham WT, Singh JP, Bax JJ, Borer JS, Brugada J, et al. Cardiac-Resynchronization Therapy in Heart Failure with a Narrow QRS Complex. N Engl J Med . 2013;369(15):1395–1405. doi: 10.1056/NEJMoa1306687. [DOI] [PubMed] [Google Scholar]
  • 173.Steffel J, Robertson M, Singh JP, Abraham WT, Bax JJ, Borer JS, et al. The Effect of QRS Duration on Cardiac Resynchronization Therapy in Patients with a Narrow QRS Complex: A Subgroup Analysis of the EchoCRT Trial. Eur Heart J . 2015;36(30):1983–1989. doi: 10.1093/eurheartj/ehv242. [DOI] [PubMed] [Google Scholar]
  • 174.Cleland JG, Abraham WT, Linde C, Gold MR, Young JB, Claude Daubert J, et al. An Individual Patient Meta-Analysis of Five Randomized Trials Assessing the Effects of Cardiac Resynchronization Therapy on Morbidity and Mortality in Patients with Symptomatic Heart Failure. Eur Heart J . 2013;34(46):3547–3556. doi: 10.1093/eurheartj/eht290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 175.Woods B, Hawkins N, Mealing S, Sutton A, Abraham WT, Beshai JF, et al. Individual Patient Data Network Meta-Analysis of Mortality Effects of Implantable Cardiac Devices. Heart . 2015;101(22):1800–1806. doi: 10.1136/heartjnl-2015-307634. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 176.Birnie DH, Ha A, Higginson L, Sidhu K, Green M, Philippon F, et al. Impact of QRS Morphology and Duration on Outcomes After Cardiac Resynchronization Therapy: Results from the Resynchronization-Defibrillation for Ambulatory Heart Failure Trial (RAFT) Circ Heart Fail . 2013;6(6):1190–1198. doi: 10.1161/CIRCHEARTFAILURE.113.000380. [DOI] [PubMed] [Google Scholar]
  • 177.Masoudi FA, Ponirakis A, Yeh RW, Maddox TM, Beachy J, Casale PN, et al. Cardiovascular Care Facts: A Report from the National Cardiovascular Data Registry: 2011. J Am Coll Cardiol . 2013;62(21):1931–1947. doi: 10.1016/j.jacc.2013.05.099. [DOI] [PubMed] [Google Scholar]
  • 178.Gold MR, Thébault C, Linde C, Abraham WT, Gerritse B, Ghio S, et al. Effect of QRS Duration and Morphology on Cardiac Resynchronization Therapy Outcomes in Mild Heart Failure: Results from the Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction (REVERSE) Study. Circulation . 2012;126(7):822–829. doi: 10.1161/CIRCULATIONAHA.112.097709. [DOI] [PubMed] [Google Scholar]
  • 179.Biton Y, Kutyifa V, Cygankiewicz I, Goldenberg I, Klein H, McNitt S, et al. Relation of QRS Duration to Clinical Benefit of Cardiac Resynchronization Therapy in Mild Heart Failure Patients Without Left Bundle Branch Block: The Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy Substudy. Circ Heart Fail . 2016;9(2):e002667. doi: 10.1161/CIRCHEARTFAILURE.115.002667. [DOI] [PubMed] [Google Scholar]
  • 180.Cleland JG, Mareev Y, Linde C. Reflections on EchoCRT: Sound Guidance on QRS Duration and Morphology for CRT? Eur Heart J . 2015;36(30):1948–1951. doi: 10.1093/eurheartj/ehv264. [DOI] [PubMed] [Google Scholar]
  • 181.Kawata H, Bao H, Curtis JP, Minges KE, Mitiku T, Birgersdotter-Green U, et al. Cardiac Resynchronization Defibrillator Therapy for Nonspecific Intraventricular Conduction Delay Versus Right Bundle Branch Block. J Am Coll Cardiol . 2019;73(24):3082–3099. doi: 10.1016/j.jacc.2019.04.025. [DOI] [PubMed] [Google Scholar]
  • 182.Bilchick KC, Kamath S, DiMarco JP, Stukenborg GJ. Bundle-Branch Block Morphology and Other Predictors of Outcome after Cardiac Resynchronization Therapy in Medicare Patients. Circulation . 2010;122(20):2022–2030. doi: 10.1161/CIRCULATIONAHA.110.956011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 183.Pastore G, Morani G, Maines M, Marcantoni L, Bolzan B, Zanon F, et al. Patients with Right Bundle Branch Block and Concomitant Delayed Left Ventricular Activation Respond to Cardiac Resynchronization Therapy. Europace . 2018;20(11):171–178. doi: 10.1093/europace/eux362. [DOI] [PubMed] [Google Scholar]
  • 184.Surawicz B, Childers R, Deal BJ, Gettes LS, Bailey JJ, Gorgels A, et al. AHA/ACCF/HRS Recommendations for the Standardization and Interpretation of the Electrocardiogram: Part III: Intraventricular Conduction Disturbances: A Scientific Statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society. Endorsed by the International Society for Computerized Electrocardiology. J Am Coll Cardiol . 2009;53(11):976–981. doi: 10.1016/j.jacc.2008.12.013. [DOI] [PubMed] [Google Scholar]
  • 185.Maisel WH, Stevenson LW. Atrial Fibrillation in Heart Failure: Epidemiology, Pathophysiology, and Rationale for Therapy. Am J Cardiol . 2003;91(6A):2D–8D. doi: 10.1016/s0002-9149(02)03373-8. [DOI] [PubMed] [Google Scholar]
  • 186.Khadjooi K, Foley PW, Chalil S, Anthony J, Smith RE, Frenneaux MP, et al. Long-Term Effects of Cardiac Resynchronisation Therapy in Patients with Atrial Fibrillation. Heart . 2008;94(7):879–883. doi: 10.1136/hrt.2007.129429. [DOI] [PubMed] [Google Scholar]
  • 187.Gasparini M, Auricchio A, Regoli F, Fantoni C, Kawabata M, Galimberti P, et al. Four-Year Efficacy of Cardiac Resynchronization Therapy on Exercise Tolerance and Disease Progression: The Importance of Performing Atrioventricular Junction Ablation in Patients with Atrial Fibrillation. J Am Coll Cardiol . 2006;48(4):734–743. doi: 10.1016/j.jacc.2006.03.056. [DOI] [PubMed] [Google Scholar]
  • 188.Gasparini M, Galimberti P, Leyva F. Complete Atrioventricular Block DOES Reduce Mortality in Patients with Atrial Fibrillation Treated with Cardiac Resynchronization Therapy. Eur J Heart Fail . 2014;16(1):114. doi: 10.1002/ejhf.25. [DOI] [PubMed] [Google Scholar]
  • 189.Gasparini M, Leclercq C, Lunati M, Landolina M, Auricchio A, Santini M, et al. Cardiac Resynchronization Therapy in Patients with Atrial Fibrillation: The CERTIFY Study (Cardiac Resynchronization Therapy in Atrial Fibrillation Patients Multinational Registry) JACC Heart Fail . 2013;1(6):500–507. doi: 10.1016/j.jchf.2013.06.003. [DOI] [PubMed] [Google Scholar]
  • 190.Marijon E, Jacob S, Mouton E, Defaye P, Piot O, Delarche N, et al. Frequency of Atrial Tachyarrhythmias in Patients Treated by Cardiac Resynchronization (from the Prospective, Multicenter Mona Lisa Study) Am J Cardiol . 2010;106(5):688–693. doi: 10.1016/j.amjcard.2010.04.025. [DOI] [PubMed] [Google Scholar]
  • 191.Tolosana JM, Madrid AH, Brugada J, Sitges M, Bolao IG, Lozano IF, et al. Comparison of Benefits and Mortality in Cardiac Resynchronization Therapy in Patients with Atrial Fibrillation versus Patients in Sinus Rhythm (Results of the Spanish Atrial Fibrillation and Resynchronization [SPARE] Study) Am J Cardiol . 2008;102(4):444–449. doi: 10.1016/j.amjcard.2008.04.008. [DOI] [PubMed] [Google Scholar]
  • 192.Hoppe UC, Casares JM, Eiskjaer H, Hagemann A, Cleland JG, Freemantle N, et al. Effect of Cardiac Resynchronization on the Incidence of Atrial Fibrillation in Patients with Severe Heart Failure. Circulation . 2006;114(1):18–25. doi: 10.1161/CIRCULATIONAHA.106.614560. [DOI] [PubMed] [Google Scholar]
  • 193.Upadhyay GA, Choudhry NK, Auricchio A, Ruskin J, Singh JP. Cardiac Resynchronization in Patients with Atrial Fibrillation: A Meta-Analysis of Prospective Cohort Studies. J Am Coll Cardiol . 2008;52(15):1239–1246. doi: 10.1016/j.jacc.2008.06.043. [DOI] [PubMed] [Google Scholar]
  • 194.Linde C, Leclercq C, Rex S, Garrigue S, Lavergne T, Cazeau S, et al. Long-Term Benefits of Biventricular Pacing in Congestive Heart Failure: Results from the MUltisite STimulation in Cardiomyopathy (MUSTIC) Study. J Am Coll Cardiol . 2002;40(1):111–118. doi: 10.1016/s0735-1097(02)01932-0. [DOI] [PubMed] [Google Scholar]
  • 195.Wilton SB, Leung AA, Ghali WA, Faris P, Exner DV. Outcomes of Cardiac Resynchronization Therapy in Patients with Versus Those Without Atrial Fibrillation: A Systematic Review and Meta-Analysis. Heart Rhythm . 2011;8(7):1088–1094. doi: 10.1016/j.hrthm.2011.02.014. [DOI] [PubMed] [Google Scholar]
  • 196.Leclercq C, Walker S, Linde C, Clementy J, Marshall AJ, Ritter P, Djiane P, et al. Comparative Effects of Permanent Biventricular and Right-Univentricular Pacing in Heart Failure Patients with Chronic Atrial Fibrillation. Eur Heart J . 2002;23(22):1780–1787. doi: 10.1053/euhj.2002.3232. [DOI] [PubMed] [Google Scholar]
  • 197.Doshi RN, Daoud EG, Fellows C, Turk K, Duran A, Hamdan MH, et al. Left Ventricular-Based Cardiac Stimulation Post AV Nodal Ablation Evaluation (the PAVE Study) J Cardiovasc Electrophysiol . 2005;16(11):1160–1165. doi: 10.1111/j.1540-8167.2005.50062.x. [DOI] [PubMed] [Google Scholar]
  • 198.Brignole M, Gammage M, Puggioni E, Alboni P, Raviele A, Sutton R, et al. Comparative Assessment of Right, Left, and Biventricular Pacing in Patients with Permanent Atrial Fibrillation. Eur Heart J . 2005;26(7):712–722. doi: 10.1093/eurheartj/ehi069. [DOI] [PubMed] [Google Scholar]
  • 199.Orlov MV, Gardin JM, Slawsky M, Bess RL, Cohen G, Bailey W, et al. Biventricular Pacing Improves Cardiac Function and Prevents Further Left Atrial Remodeling in Patients with Symptomatic Atrial Fibrillation After Atrioventricular Node Ablation. Am Heart J . 2010;159(2):264–270. doi: 10.1016/j.ahj.2009.11.012. [DOI] [PubMed] [Google Scholar]
  • 200.Khazanie P, Hammill BG, Qualls LG, Fonarow GC, Hammill SC, Heidenreich PA, et al. Clinical Effectiveness of Cardiac Resynchronization Therapy versus Medical Therapy Alone Among Patients with Heart Failure: Analysis of the ICD Registry and ADHERE. Circ Heart Fail . 2014;7(6):926–934. doi: 10.1161/CIRCHEARTFAILURE.113.000838. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 201.Boriani G, Gasparini M, Landolina M, Lunati M, Proclemer A, Lonardi G, et al. Incidence and Clinical Relevance of Uncontrolled Ventricular Rate During Atrial Fibrillation in Heart Failure Patients Treated with Cardiac Resynchronization Therapy. Eur J Heart Fail . 2011;13(8):868–876. doi: 10.1093/eurjhf/hfr046. [DOI] [PubMed] [Google Scholar]
  • 202.Hayes DL, Boehmer JP, Day JD, Gilliam FR, 3rd, Heidenreich PA, Seth M, et al. Cardiac Resynchronization Therapy and the Relationship of Percent Biventricular Pacing to symptoms and Survival. Heart Rhythm . 2011;8(9):1469–1475. doi: 10.1016/j.hrthm.2011.04.015. [DOI] [PubMed] [Google Scholar]
  • 203.Brignole M, Botto G, Mont L, Iacopino S, De Marchi G, Oddone D, et al. Cardiac Resynchronization Therapy in Patients Undergoing Atrioventricular Junction Ablation for Permanent Atrial Fibrillation: A Randomized Trial. Eur Heart J . 2011;32(19):2420–2429. doi: 10.1093/eurheartj/ehr162. [DOI] [PubMed] [Google Scholar]
  • 204.Gasparini M, Auricchio A, Metra M, Regoli F, Fantoni C, Lamp B, et al. Long-Term Survival in Patients Undergoing Cardiac Resynchronization Therapy: The Importance of Performing Atrio-Ventricular Junction Ablation in Patients with Permanent Atrial Fibrillation. Eur Heart J . 2008;29(13):1644–1652. doi: 10.1093/eurheartj/ehn133. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 205.Ahmed M, Gorcsan J, 3rd, Marek J, Ryo K, Haugaa K, R Ludwig D, et al. Right Ventricular Apical Pacing-Induced Left Ventricular Dyssynchrony is Associated with a Subsequent Decline in Ejection Fraction. Heart Rhythm . 2014;11(4):602–608. doi: 10.1016/j.hrthm.2013.12.020. [DOI] [PubMed] [Google Scholar]
  • 206.Cherian TS, Upadhyay GA. Right Ventricular Pacing and Cardiac Resynchronization Devices. Card Electrophysiol Clin . 2018;10(1):31–42. doi: 10.1016/j.ccep.2017.11.004. [DOI] [PubMed] [Google Scholar]
  • 207.Khurshid S, Obeng-Gyimah E, Supple GE, Schaller R, Lin D, Owens AT, et al. Reversal of Pacing-Induced Cardiomyopathy Following Cardiac Resynchronization Therapy. JACC Clin Electrophysiol . 2018;4(2):168–177. doi: 10.1016/j.jacep.2017.10.002. [DOI] [PubMed] [Google Scholar]
  • 208.Gage RM, Burns KV, Bank AJ. Echocardiographic and Clinical Response to Cardiac Resynchronization Therapy in Heart Failure Patients with and Without Previous Right Ventricular Pacing. Eur J Heart Fail . 2014;16(11):1199–1205. doi: 10.1002/ejhf.14. [DOI] [PubMed] [Google Scholar]
  • 209.Stankovic I, Prinz C, Ciarka A, Daraban AM, Mo Y, Aarones M, et al. Long-Term Outcome After CRT in the Presence of Mechanical Dyssynchrony Seen with Chronic RV Pacing or Intrinsic LBBB. JACC Cardiovasc Imaging . 2017;10(10 Pt A):1091–1099. doi: 10.1016/j.jcmg.2016.08.015. [DOI] [PubMed] [Google Scholar]
  • 210.Stockburger M, Gómez-Doblas JJ, Lamas G, Alzueta J, Fernández-Lozano I, Cobo E, et al. Preventing Ventricular Dysfunction in Pacemaker Patients Without Advanced Heart Failure: Results from a Multicentre International Randomized Trial (PREVENT-HF) Eur J Heart Fail . 2011;13(6):633–641. doi: 10.1093/eurjhf/hfr041. [DOI] [PubMed] [Google Scholar]
  • 211.Yu CM, Fang F, Luo XX, Zhang Q, Azlan H, Razali O. Long-Term Follow-Up Results of the Pacing to Avoid Cardiac Enlargement (PACE) trial. Eur J Heart Fail . 2014;16(9):1016–1025. doi: 10.1002/ejhf.157. [DOI] [PubMed] [Google Scholar]
  • 212.Slotwiner DJ, Raitt MH, Del-Carpio Munoz F, Mulpuru SK, Nasser N, Peterson PN. Impact of Physiologic Pacing Versus Right Ventricular Pacing Among Patients with Left Ventricular Ejection Fraction Greater Than 35%: A Systematic Review for the 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients with Bradycardia and Cardiac Conduction Delay: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol . 2019;74(7):988–1008. doi: 10.1016/j.jacc.2018.10.045. [DOI] [PubMed] [Google Scholar]
  • 213.Young JB, Abraham WT, Smith AL, Leon AR, Lieberman R, Wilkoff B, et al. Combined Cardiac Resynchronization and Implantable Cardioversion Defibrillation in Advanced Chronic Heart Failure: The MIRACLE ICD Trial. JAMA . 2003;289(20):2685–2694. doi: 10.1001/jama.289.20.2685. [DOI] [PubMed] [Google Scholar]
  • 214.Beshai JF, Grimm RA, Nagueh SF, Baker JH, 2nd, Beau SL, Greenberg SM, et al. Cardiac-Resynchronization Therapy in Heart Failure with Narrow QRS Complexes. N Engl J Med . 2007;357(24):2461–2471. doi: 10.1056/NEJMoa0706695. [DOI] [PubMed] [Google Scholar]
  • 215.National Institute for Health and Care Excellence . Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy for Arrhythmias and Heart Failure . London: NICE; 2014. [Google Scholar]
  • 216.Sipahi I, Chou JC, Hyden M, Rowland DY, Simon DI, Fang JC. Effect of QRS Morphology on Clinical Event Reduction with Cardiac Resynchronization Therapy: Meta-Analysis of Randomized Controlled Trials. Am Heart J . 2012;163(2):260–7.e3. doi: 10.1016/j.ahj.2011.11.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 217.Cunnington C, Kwok CS, Satchithananda DK, Patwala A, Khan MA, Zaidi A, et al. Cardiac Resynchronisation Therapy is not Associated with a Reduction in Mortality or Heart Failure Hospitalisation in Patients with Non-Left Bundle Branch Block QRS Morphology: Meta-Analysis of Randomised Controlled Trials. Heart . 2015;101(18):1456–1462. doi: 10.1136/heartjnl-2014-306811. [DOI] [PubMed] [Google Scholar]
  • 218.Zhu H, Zou T, Zhong Y, Yang C, Ren Y, Wang F. Prevention of Non-Response to Cardiac Resynchronization Therapy: Points to Remember. Heart Fail Rev . 2020;25(2):269–275. doi: 10.1007/s10741-019-09834-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 219.Auricchio A, Heggermont WA. Technology Advances to Improve Response to Cardiac Resynchronization Therapy: What Clinicians Should Know. Rev Esp Cardiol . 2018;71(6):477–484. doi: 10.1016/j.rec.2018.01.006. [DOI] [PubMed] [Google Scholar]
  • 220.Arnold AD, Shun-Shin MJ, Keene D, Howard JP, Sohaib SMA, Wright IJ, et al. His Resynchronization Versus Biventricular Pacing in Patients With Heart Failure and Left Bundle Branch Block. J Am Coll Cardiol . 2018;72(24):3112–3122. doi: 10.1016/j.jacc.2018.09.073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 221.Vijayaraman P. Cardiac Resynchronization Therapy Using Permanent His-Bundle Pacing: Are We There Yet? Heart Rhythm . 2017;14(9):1362–1363. doi: 10.1016/j.hrthm.2017.05.024. [DOI] [PubMed] [Google Scholar]
  • 222.Shan P, Su L, Chen X, Xu L, Ni X, Huang W. Direct His-Bundle Pacing Improved Left Ventricular Function and Remodelling in a Biventricular Pacing Nonresponder. Can J Cardiol . 2016;32(12):1577.e1–1577.e4. doi: 10.1016/j.cjca.2015.10.024. [DOI] [PubMed] [Google Scholar]
  • 223.Ajijola OA, Upadhyay GA, Macias C, Shivkumar K, Tung R. Permanent His-Bundle Pacing for Cardiac Resynchronization Therapy: Initial Feasibility Study in lieu of Left Ventricular Lead. Heart Rhythm . 2017;14(9):1353–1361. doi: 10.1016/j.hrthm.2017.04.003. [DOI] [PubMed] [Google Scholar]
  • 224.Leyton-Mange JS, Mela T. Novel Pacing Strategies for Heart Failure Management. 64 Curr Treat Options Cardiovasc Med . 2017;19(8) doi: 10.1007/s11936-017-0561-3. [DOI] [PubMed] [Google Scholar]
  • 225.Scherlag BJ, Papaila A. Permanent His Bundle Pacing to Replace Biventricular Pacing for Cardiac Resynchronization Therapy. Med Hypotheses . 2017;109:77–79. doi: 10.1016/j.mehy.2017.09.026. [DOI] [PubMed] [Google Scholar]
  • 226.Barba-Pichardo R, Sánchez AM, Fernández-Gómez JM, Moriña-Vázquez P, Venegas-Gamero J, Herrera-Carranza M. Ventricular Resynchronization Therapy by Direct His-Bundle Pacing Using an Internal Cardioverter Defibrillator. Europace . 2013;15(1):83–88. doi: 10.1093/europace/eus228. [DOI] [PubMed] [Google Scholar]
  • 227.Lustgarten DL, Crespo EM, Arkhipova-Jenkins I, Lobel R, Winget J, Koehler J, et al. His-Bundle Pacing versus Biventricular Pacing in Cardiac Resynchronization Therapy Patients: A Crossover Design Comparison. Heart Rhythm . 2015;12(7):1548–1557. doi: 10.1016/j.hrthm.2015.03.048. [DOI] [PubMed] [Google Scholar]
  • 228.Qian Z, Zou F, Wang Y, Qiu Y, Chen X, Jiang H, et al. Permanent His Bundle Pacing in Heart Failure Patients: A Systematic Review and Meta-Analysis. Pacing Clin Electrophysiol . 2019;42(2):139–145. doi: 10.1111/pace.13565. [DOI] [PubMed] [Google Scholar]
  • 229.Upadhyay GA, Vijayaraman P, Nayak HM, Verma N, Dandamudi G, Sharma PS, et al. On-Treatment Comparison Between Corrective His Bundle Pacing and Biventricular Pacing for Cardiac Resynchronization: A Secondary Analysis of the His-SYNC Pilot Trial. Heart Rhythm . 2019;16(12):1797–1807. doi: 10.1016/j.hrthm.2019.05.009. [DOI] [PubMed] [Google Scholar]
  • 230.Zhang W, Huang J, Qi Y, Wang F, Guo L, Shi X, et al. Cardiac Resynchronization Therapy by Left Bundle Branch Area Pacing in Patients with Heart Failure and Left Bundle Branch Block. Heart Rhythm . 2019;16(12):1783–1790. doi: 10.1016/j.hrthm.2019.09.006. [DOI] [PubMed] [Google Scholar]
  • 231.Tan NY, Witt CM, Oh JK, Cha YM. Left Bundle Branch Block: Current and Future Perspectives. Circ Arrhythm Electrophysiol . 2020;13(4):e008239. doi: 10.1161/CIRCEP.119.008239. [DOI] [PubMed] [Google Scholar]
  • 232.Huang W, Wu S, Vijayaraman P, Su L, Chen X, Cai B, et al. Cardiac Resynchronization Therapy in Patients With Nonischemic Cardiomyopathy Using Left Bundle Branch Pacing. JACC Clin Electrophysiol . 2020;6(7):849–858. doi: 10.1016/j.jacep.2020.04.011. [DOI] [PubMed] [Google Scholar]
  • 233.Wu S, Su L, Vijayaraman P, Zheng R, Cai M, Xu L, et al. Left Bundle Branch Pacing for Cardiac Resynchronization Therapy: Nonrandomized On-Treatment Comparison With His Bundle Pacing and Biventricular Pacing. Can J Cardiol . 2021;37(2):319–328. doi: 10.1016/j.cjca.2020.04.037. [DOI] [PubMed] [Google Scholar]
  • 234.Sharma PS, Vijayaraman P, Ellenbogen KA. Permanent His Bundle Pacing: Shaping the Future of Physiological Ventricular Pacing. Nat Rev Cardiol . 2020;17(1):22–36. doi: 10.1038/s41569-019-0224-z. [DOI] [PubMed] [Google Scholar]
  • 235.Kahwash R, Burkhoff D, Abraham WT. Cardiac Contractility Modulation in Patients with Advanced Heart Failure. Expert Rev Cardiovasc Ther . 2013;11(5):635–645. doi: 10.1586/erc.13.48. [DOI] [PubMed] [Google Scholar]
  • 236.Nägele H, Behrens S, Eisermann C. Cardiac Contractility Modulation in Non-Responders to Cardiac Resynchronization Therapy. Europace . 2008;10(12):1375–1380. doi: 10.1093/europace/eun257. [DOI] [PubMed] [Google Scholar]
  • 237.Borggrefe MM, Lawo T, Butter C, Schmidinger H, Lunati M, Pieske B, et al. Randomized, Double Blind Study of Non-Excitatory, Cardiac Contractility Modulation Electrical Impulses for Symptomatic Heart Failure. Eur Heart J . 2008;29(8):1019–1028. doi: 10.1093/eurheartj/ehn020. [DOI] [PubMed] [Google Scholar]
  • 238.Myerburg RJ, Kessler KM, Castellanos A. Sudden Cardiac Death: Epidemiology, Transient Risk, and Intervention Assessment. Ann Intern Med . 1993;119(12):1187–1197. doi: 10.7326/0003-4819-119-12-199312150-00006. [DOI] [PubMed] [Google Scholar]
  • 239.Greenberg H, Case RB, Moss AJ, Brown MW, Carroll ER, Andrews ML, et al. Analysis of Mortality Events in the Multicenter Automatic Defibrillator Implantation Trial (MADIT-II) J Am Coll Cardiol . 2004;43(8):1459–1465. doi: 10.1016/j.jacc.2003.11.038. [DOI] [PubMed] [Google Scholar]
  • 240.Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. Amiodarone or an Implantable Cardioverter-Defibrillator for Congestive Heart Failure. N Engl J Med . 2005;352(3):225–237. doi: 10.1056/NEJMoa043399. [DOI] [PubMed] [Google Scholar]
  • 241.Moss AJ, Hall WJ, Cannom DS, Daubert JP, Higgins SL, Klein H, et al. Improved Survival with an Implanted Defibrillator in Patients with Coronary Disease at High Risk for Ventricular Arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med . 1996;335(26):1933–1940. doi: 10.1056/NEJM199612263352601. [DOI] [PubMed] [Google Scholar]
  • 242.Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection Fraction. N Engl J Med . 2002;346(12):877–883. doi: 10.1056/NEJMoa013474. [DOI] [PubMed] [Google Scholar]
  • 243.Hohnloser SH, Kuck KH, Dorian P, Roberts RS, Hampton JR, Hatala R, et al. Prophylactic Use of an Implantable Cardioverter-Defibrillator After Acute Myocardial Infarction. N Engl J Med . 2004;351(24):2481–2488. doi: 10.1056/NEJMoa041489. [DOI] [PubMed] [Google Scholar]
  • 244.Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A Randomized Study of the Prevention of Sudden Death in Patients with Coronary Artery Disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med . 1999;341(25):1882–1890. doi: 10.1056/NEJM199912163412503. [DOI] [PubMed] [Google Scholar]
  • 245.The MERIT HF study group Effect of Metoprolol CR/XL in Chronic Heart Failure: Metoprolol CR/XL Randomised INTERVENTION Trial in Congestive Heart Failure (MERIT-HF) Lancet . 1999;353(9169):2001–2007. doi: 10.1016/S0140-6736(99)04440-2. [DOI] [PubMed] [Google Scholar]
  • 246.Spotnitz HM, Herre JM, Raza ST, Hammon JW, Jr, Baker LD, Jr, Fitzgerald DM, et al. Effect of Implantable Cardioverter-Defibrillator Implantation on Surgical Morbidity in the CABG Patch Trial. Surgical Investigators of the Coronary Artery Bypass Graft Patch Trial. Circulation . 1998;98(19) Suppl:II77–II80. [PubMed] [Google Scholar]
  • 247.Bigger JT., Jr Prophylactic Use of Implanted Cardiac Defibrillators in Patients at High Risk for Ventricular Arrhythmias after Coronary-Artery Bypass Graft Surgery. Coronary Artery Bypass Graft (CABG) Patch Trial Investigators. N Engl J Med . 1997;337(22):1569–1575. doi: 10.1056/NEJM199711273372201. [DOI] [PubMed] [Google Scholar]
  • 248.Fröhlich GM, Holzmeister J, Hübler M, Hübler S, Wolfrum M, Enseleit F, et al. Prophylactic Implantable Cardioverter Defibrillator Treatment in patiEnts with End-Stage Heart Failure Awaiting Heart Transplantation. Heart . 2013;99(16):1158–1165. doi: 10.1136/heartjnl-2013-304185. [DOI] [PubMed] [Google Scholar]
  • 249.Vakil K, Duval S, Cogswell R, Eckman P, Levy WC, Anand I, t al S. Impact of Implantable Cardioverter-Defibrillators on Waitlist Mortality Among Patients Awaiting Heart Transplantation: An UNOS/OPTN Analysis. JACC Clin Electrophysiol . 2017;3(1):33–40. doi: 10.1016/j.jacep.2016.07.010. [DOI] [PubMed] [Google Scholar]
  • 250.Vakil K, Kazmirczak F, Sathnur N, Adabag S, Cantillon DJ, Kiehl EL, et al. Implantable Cardioverter-Defibrillator Use in Patients With Left Ventricular Assist Devices: A Systematic Review and Meta-Analysis. JACC Heart Fail . 2016;4(10):772–779. doi: 10.1016/j.jchf.2016.05.003. [DOI] [PubMed] [Google Scholar]
  • 251.Koutalas E, Kanoupakis E, Vardas P. Sudden Cardiac Death in Non-Ischemic Dilated Cardiomyopathy: A Critical Appraisal of Existing and Potential Risk Stratification Tools. Int J Cardiol . 2013;167(2):335–341. doi: 10.1016/j.ijcard.2012.07.014. [DOI] [PubMed] [Google Scholar]
  • 252.McMurray JJ, Packer M, Desai AS, Gong J, Lefkowitz MP, Rizkala AR, et al. Angiotensin-Neprilysin Inhibition Versus Enalapril in Heart Failure. N Engl J Med . 2014;371(11):993–1004. doi: 10.1056/NEJMoa1409077. [DOI] [PubMed] [Google Scholar]
  • 253.Middlekauff HR, Stevenson WG, Stevenson LW, Saxon LA. Syncope in Advanced Heart Failure: High Risk of Sudden Death Regardless of Origin Of Syncope. J Am Coll Cardiol . 1993;21(1):110–116. doi: 10.1016/0735-1097(93)90724-f. [DOI] [PubMed] [Google Scholar]
  • 254.Grimm W, Christ M, Bach J, Müller HH, Maisch B. Noninvasive Arrhythmia Risk Stratification in Idiopathic Dilated Cardiomyopathy: Results of the Marburg Cardiomyopathy Study. Circulation . 2003;108(23):2883–2891. doi: 10.1161/01.CIR.0000100721.52503.85. [DOI] [PubMed] [Google Scholar]
  • 255.Anselmino M, De Ferrari GM, Massa R, Manca L, Tritto M, Molon G, et al. Predictors of Mortality and Hospitalization for Cardiac Causes in Patients with Heart Failure and Nonischemic Heart Disease: A Subanalysis of the ALPHA Study. Pacing Clin Electrophysiol . 2009;32(Suppl 1):S214–S218. doi: 10.1111/j.1540-8159.2008.02286.x. [DOI] [PubMed] [Google Scholar]
  • 256.Scott PA, Barry J, Roberts PR, Morgan JM. Brain Natriuretic Peptide for the Prediction of Sudden Cardiac Death and Ventricular Arrhythmias: A Meta-Analysis. Eur J Heart Fail . 2009;11(10):958–966. doi: 10.1093/eurjhf/hfp123. [DOI] [PubMed] [Google Scholar]
  • 257.Goldberger JJ, Cain ME, Hohnloser SH, Kadish AH, Knight BP, Lauer MS, et al. American Heart Association/American College of Cardiology Foundation/Heart Rhythm Society Scientific Statement on Noninvasive Risk Stratification Techniques for Identifying Patients at risk for Sudden Cardiac Death: A Scientific Statement from the American Heart Association Council on Clinical Cardiology Committee on Electrocardiography and Arrhythmias and Council on Epidemiology and Prevention. Heart Rhythm . 2008;5(10):e1–21. doi: 10.1016/j.hrthm.2008.05.031. [DOI] [PubMed] [Google Scholar]
  • 258.Hofmann T, Meinertz T, Kasper W, Geibel A, Zehender M, Hohnloser S, et al. Mode of Death in Idiopathic Dilated Cardiomyopathy: A Multivariate Analysis of Prognostic Determinants. Am Heart J . 1988;116(6 Pt 1):1455–1463. doi: 10.1016/0002-8703(88)90728-4. [DOI] [PubMed] [Google Scholar]
  • 259.Okutucu S, Oto A. Risk Stratification in Nonischemic Dilated Cardiomyopathy: Current perspectives. Cardiol J . 2010;17(3):219–229. [PubMed] [Google Scholar]
  • 260.Cahalin LP, Chase P, Arena R, Myers J, Bensimhon D, Peberdy MA, et al. A Meta-Analysis of the Prognostic Significance of Cardiopulmonary Exercise Testing in Patients with Heart Failure. Heart Fail Rev . 2013;18(1):79–94. doi: 10.1007/s10741-012-9332-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 261.Guazzi M, Raimondo R, Vicenzi M, Arena R, Proserpio C, Braga SS, et al. Exercise Oscillatory Ventilation May Predict Sudden Cardiac Death in Heart Failure Patients. J Am Coll Cardiol . 2007;50(4):299–308. doi: 10.1016/j.jacc.2007.03.042. [DOI] [PubMed] [Google Scholar]
  • 262.Piran S, Liu P, Morales A, Hershberger RE. Where Genome Meets Phenome: Rationale for Integrating Genetic and Protein Biomarkers in the Diagnosis and Management of Dilated Cardiomyopathy and Heart Failure. J Am Coll Cardiol . 2012;60(4):283–289. doi: 10.1016/j.jacc.2012.05.005. [DOI] [PubMed] [Google Scholar]
  • 263.McCrohon JA, Moon JC, Prasad SK, McKenna WJ, Lorenz CH, Coats AJ, et al. Differentiation of Heart Failure Related to Dilated Cardiomyopathy and Coronary Artery Disease Using Gadolinium-Enhanced Cardiovascular Magnetic Resonance. Circulation . 2003;108(1):54–59. doi: 10.1161/01.CIR.0000078641.19365.4C. [DOI] [PubMed] [Google Scholar]
  • 264.Becker MAJ, Cornel JH, van de Ven PM, van Rossum AC, Allaart CP, Germans T. The Prognostic Value of Late Gadolinium-Enhanced Cardiac Magnetic Resonance Imaging in Nonischemic Dilated Cardiomyopathy: A Review and Meta-Analysis. JACC Cardiovasc Imaging . 2018;11(9):1274–1284. doi: 10.1016/j.jcmg.2018.03.006. [DOI] [PubMed] [Google Scholar]
  • 265.Gulati A, Jabbour A, Ismail TF, Guha K, Khwaja J, Raza S, et al. Association of Fibrosis with Mortality and Sudden Cardiac Death in Patients with Nonischemic Dilated Cardiomyopathy. JAMA . 2013;309(9):896–908. doi: 10.1001/jama.2013.1363. [DOI] [PubMed] [Google Scholar]
  • 266.Poole JE, Johnson GW, Hellkamp AS, Anderson J, Callans DJ, Raitt MH, et al. Prognostic Importance of Defibrillator Shocks in Patients with Heart Failure. N Engl J Med . 2008;359(10):1009–1017. doi: 10.1056/NEJMoa071098. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 267.Køber L, Thune JJ, Nielsen JC, Haarbo J, Videbæk L, Korup E, et al. Defibrillator Implantation in Patients with Nonischemic Systolic Heart Failure. N Engl J Med . 2016;375(13):1221–1230. doi: 10.1056/NEJMoa1608029. [DOI] [PubMed] [Google Scholar]
  • 268.Semsarian C, Ingles J, Maron MS, Maron BJ. New Perspectives on the Prevalence of Hypertrophic Cardiomyopathy. J Am Coll Cardiol . 2015;65(12):1249–1254. doi: 10.1016/j.jacc.2015.01.019. [DOI] [PubMed] [Google Scholar]
  • 269.Maron BJ, Gardin JM, Flack JM, Gidding SS, Kurosaki TT, Bild DE. Prevalence of Hypertrophic Cardiomyopathy in a General Population of Young Adults. Echocardiographic analysis of 4111 subjects in the CARDIA Study. Coronary Artery Risk Development in (Young) Adults. Circulation . 1995;92(4):785–789. doi: 10.1161/01.cir.92.4.785. [DOI] [PubMed] [Google Scholar]
  • 270.McKenna WJ, Deanfield JE. Hypertrophic Cardiomyopathy: An Important Cause of Sudden Death. Arch Dis Child . 1984;59(10):971–975. doi: 10.1136/adc.59.10.971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 271.Minami Y, Haruki S, Hagiwara N. Phenotypic Overlap in Hypertrophic Cardiomyopathy: Apical Hypertrophy, Midventricular Obstruction, and Apical Aneurysm. J Cardiol . 2014;64(6):463–469. doi: 10.1016/j.jjcc.2014.03.003. [DOI] [PubMed] [Google Scholar]
  • 272.Maron BJ, Roberts WC, Epstein SE. Sudden Death in Hypertrophic Cardiomyopathy: A Profile of 78 Patients. Circulation . 1982;65(7):1388–1394. doi: 10.1161/01.cir.65.7.1388. [DOI] [PubMed] [Google Scholar]
  • 273.Maron BJ, Bonow RO, Cannon RO, 3rd, Leon MB, Epstein SE. Hypertrophic Cardiomyopathy. Interrelations of Clinical Manifestations, Pathophysiology, and Therapy (1) N Engl J Med . 1987;316(13):780–789. doi: 10.1056/NEJM198703263161305. [DOI] [PubMed] [Google Scholar]
  • 274.Maron BJ, Bonow RO, Cannon RO, 3rd, Leon MB, Epstein SE. Hypertrophic Cardiomyopathy. Interrelations of Clinical Manifestations, Pathophysiology, and Therapy (2) N Engl J Med . 1987;316(14):844–852. doi: 10.1056/NEJM198704023161405. [DOI] [PubMed] [Google Scholar]
  • 275.Spirito P, Bellone P. Natural History of Hypertrophic Cardiomyopathy. Br Heart J . 1994;72(6) Suppl:S10–S12. doi: 10.1136/hrt.72.6_suppl.s10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 276.Wigle ED, Rakowski H, Kimball BP, Williams WG. Hypertrophic Cardiomyopathy. Clinical Spectrum and Treatment. Circulation . 1995;92(7):1680–1692. doi: 10.1161/01.cir.92.7.1680. [DOI] [PubMed] [Google Scholar]
  • 277.Teare D. Asymmetrical Hypertrophy of the Heart in Young Adults. Br Heart J . 1958;20(1):1–8. doi: 10.1136/hrt.20.1.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 278.Maron BJ, Olivotto I, Spirito P, Casey SA, Bellone P, Gohman TE, et al. Epidemiology of Hypertrophic Cardiomyopathy-Related Death: Revisited in a Large Non-Referral-Based Patient Population. Circulation . 2000;102(8):858–864. doi: 10.1161/01.cir.102.8.858. [DOI] [PubMed] [Google Scholar]
  • 279.Maron BJ, Casey SA, Poliac LC, Gohman TE, Almquist AK, Aeppli DM. Clinical Course of Hypertrophic Cardiomyopathy in a Regional United States cohort. JAMA . 1999;281(7):650–655. doi: 10.1001/jama.281.7.650. [DOI] [PubMed] [Google Scholar]
  • 280.Maron BJ, Rowin EJ, Casey SA, Link MS, Lesser JR, Chan RH, et al. Hypertrophic Cardiomyopathy in Adulthood Associated With Low Cardiovascular Mortality With Contemporary Management Strategies. J Am Coll Cardiol . 2015;65(18):1915–1928. doi: 10.1016/j.jacc.2015.02.061. [DOI] [PubMed] [Google Scholar]
  • 281.Maron MS, Rowin EJ, Wessler BS, Mooney PJ, Fatima A, Patel P, et al. Enhanced American College of Cardiology/American Heart Association Strategy for Prevention of Sudden Cardiac Death in High-Risk Patients With Hypertrophic Cardiomyopathy. JAMA Cardiol . 2019;4(7):644–657. doi: 10.1001/jamacardio.2019.1391. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 282.Syska P, Przybylski A, Chojnowska L, Lewandowski M, Sterliński M, Maciag A, et al. Implantable Cardioverter-Defibrillator in Patients with Hypertrophic Cardiomyopathy: Efficacy and Complications of the Therapy in Long-Term Follow-Up. J Cardiovasc Electrophysiol . 2010;21(8):883–889. doi: 10.1111/j.1540-8167.2009.01716.x. [DOI] [PubMed] [Google Scholar]
  • 283.Schinkel AF, Vriesendorp PA, Sijbrands EJ, Jordaens LJ, ten Cate FJ, Michels M. Outcome and Complications After Implantable Cardioverter Defibrillator Therapy in Hypertrophic Cardiomyopathy: Systematic Review and Meta-Analysis. Circ Heart Fail . 2012;5(5):552–559. doi: 10.1161/CIRCHEARTFAILURE.112.969626. [DOI] [PubMed] [Google Scholar]
  • 284.Elliott PM, Sharma S, Varnava A, Poloniecki J, Rowland E, McKenna WJ. Survival After Cardiac Arrest or Sustained Ventricular Tachycardia in Patients with Hypertrophic Cardiomyopathy. J Am Coll Cardiol . 1999;33(6):1596–1601. doi: 10.1016/s0735-1097(99)00056-x. [DOI] [PubMed] [Google Scholar]
  • 285.Maron BJ, Spirito P, Shen WK, Haas TS, Formisano F, Link MS, et al. Implantable Cardioverter-Defibrillators and Prevention of Sudden Cardiac death in Hypertrophic Cardiomyopathy. JAMA . 2007;298(4):405–412. doi: 10.1001/jama.298.4.405. [DOI] [PubMed] [Google Scholar]
  • 286.Maron BJ, Maron MS. Contemporary Strategies for Risk Stratification and Prevention of Sudden Death with the Implantable Defibrillator in Hypertrophic Cardiomyopathy. Heart Rhythm . 2016;13(5):1155–1165. doi: 10.1016/j.hrthm.2015.12.048. [DOI] [PubMed] [Google Scholar]
  • 287.Bittencourt MI, Cader SA, Araújo DV, Salles ALF, Albuquerque FN, Spineti PPM, et al. Morte Súbita na Cardiomiopatia Hipertrófica. Int J Cardiovasc Sci . 2016;29(6):504–511. [Google Scholar]
  • 288.Maron BJ. Clinical Course and Management of Hypertrophic Cardiomyopathy. N Engl J Med . 2018;379(7):655–658. doi: 10.1056/NEJMra1710575. [DOI] [PubMed] [Google Scholar]
  • 289.Maron BJ, Lipson LC, Roberts WC, Savage DD, Epstein SE. “Malignant” Hypertrophic Cardiomyopathy: Identification of a Subgroup of Families with Unusually Frequent Premature Death. Am J Cardiol . 1978;41(7):1133–1140. doi: 10.1016/0002-9149(78)90870-6. [DOI] [PubMed] [Google Scholar]
  • 290.Spirito P, Bellone P, Harris KM, Bernabo P, Bruzzi P, Maron BJ. Magnitude of Left Ventricular Hypertrophy and Risk of Sudden Death in Hypertrophic Cardiomyopathy. N Engl J Med . 2000;342(24):1778–1785. doi: 10.1056/NEJM200006153422403. [DOI] [PubMed] [Google Scholar]
  • 291.Spirito P, Autore C, Rapezzi C, Bernabò P, Badagliacca R, Maron MS, et al. Syncope and Risk of Sudden Death in Hypertrophic Cardiomyopathy. Circulation . 2009;119(13):1703–1710. doi: 10.1161/CIRCULATIONAHA.108.798314. [DOI] [PubMed] [Google Scholar]
  • 292.Monserrat L, Elliott PM, Gimeno JR, Sharma S, Penas-Lado M, McKenna WJ. Non-Sustained Ventricular Tachycardia in Hypertrophic Cardiomyopathy: An Independent Marker of Sudden Death Risk in Young Patients. J Am Coll Cardiol . 2003;42(5):873–879. doi: 10.1016/s0735-1097(03)00827-1. [DOI] [PubMed] [Google Scholar]
  • 293.Wang W, Lian Z, Rowin EJ, Maron BJ, Maron MS, Link MS. Prognostic Implications of Nonsustained Ventricular Tachycardia in High-Risk Patients With Hypertrophic Cardiomyopathy. Circ Arrhythm Electrophysiol . 2017;10(3):e004604. doi: 10.1161/CIRCEP.116.004604. [DOI] [PubMed] [Google Scholar]
  • 294.Maron BJ, Rowin EJ, Casey SA, Maron MS. How Hypertrophic Cardiomyopathy Became a Contemporary Treatable Genetic Disease With Low Mortality: Shaped by 50 Years of Clinical Research and Practice. JAMA Cardiol . 2016;1(1):98–105. doi: 10.1001/jamacardio.2015.0354. [DOI] [PubMed] [Google Scholar]
  • 295.Green JJ, Berger JS, Kramer CM, Salerno M. Prognostic Value of Late Gadolinium Enhancement in Clinical Outcomes for Hypertrophic Cardiomyopathy. JACC Cardiovasc Imaging . 2012;5(4):370–377. doi: 10.1016/j.jcmg.2011.11.021. [DOI] [PubMed] [Google Scholar]
  • 296.O’Hanlon R, Grasso A, Roughton M, Moon JC, Clark S, Wage R, et al. Prognostic Significance of Myocardial Fibrosis in Hypertrophic Cardiomyopathy. J Am Coll Cardiol . 2010;56(11):867–874. doi: 10.1016/j.jacc.2010.05.010. [DOI] [PubMed] [Google Scholar]
  • 297.Rubinshtein R, Glockner JF, Ommen SR, Araoz PA, Ackerman MJ, Sorajja P, et al. Characteristics and Clinical Significance of Late Gadolinium Enhancement by Contrast-Enhanced Magnetic Resonance Imaging in Patients with Hypertrophic Cardiomyopathy. Circ Heart Fail . 2010;3(1):51–58. doi: 10.1161/CIRCHEARTFAILURE.109.854026. [DOI] [PubMed] [Google Scholar]
  • 298.Shiozaki AA, Senra T, Arteaga E, Pita CG, Martinelli M, Filho, Avila LF, et al. Myocardial Fibrosis in Patients with Hypertrophic Cardiomyopathy and High Risk for Sudden Death. Arq Bras Cardiol . 2010;94(4):535–540. doi: 10.1590/s0066-782x2010005000017. [DOI] [PubMed] [Google Scholar]
  • 299.Chan RH, Maron BJ, Olivotto I, Pencina MJ, Assenza GE, Haas T, et al. Prognostic Value of Quantitative Contrast-Enhanced Cardiovascular Magnetic Resonance for the Evaluation of Sudden Death Risk in Patients with Hypertrophic Cardiomyopathy. Circulation . 2014;130(6):484–495. doi: 10.1161/CIRCULATIONAHA.113.007094. [DOI] [PubMed] [Google Scholar]
  • 300.O’Mahony C, Jichi F, Pavlou M, Monserrat L, Anastasakis A, Rapezzi C, et al. A Novel Clinical Risk Prediction Model for Sudden Cardiac Death in Hypertrophic Cardiomyopathy (HCM risk-SCD) Eur Heart J . 2014;35(30):2010–2020. doi: 10.1093/eurheartj/eht439. [DOI] [PubMed] [Google Scholar]
  • 301.Elliott PM, Anastasakis A, Borger MA, Borggrefe M, Cecchi F, Charron P, et al. 2014 ESC Guidelines on Diagnosis and Management of Hypertrophic Cardiomyopathy: The Task Force for the Diagnosis and Management of Hypertrophic Cardiomyopathy of the European Society of Cardiology (ESC) Eur Heart J . 2014;35(39):2733–2779. doi: 10.1093/eurheartj/ehu284. [DOI] [PubMed] [Google Scholar]
  • 302.Klopotowski M, Kukula K, Malek LA, Spiewak M, Polanska-Skrzypczyk M, Jamiolkowski J, et al. The Value of Cardiac Magnetic Resonance and Distribution of Late Gadolinium Enhancement for Risk Stratification of Sudden Cardiac Death in Patients with Hypertrophic Cardiomyopathy. J Cardiol . 2016;68(1):49–56. doi: 10.1016/j.jjcc.2015.07.020. [DOI] [PubMed] [Google Scholar]
  • 303.Geske JB, Ommen SR, Gersh BJ. Hypertrophic Cardiomyopathy: Clinical Update. JACC Heart Fail . 2018;6(5):364–375. doi: 10.1016/j.jchf.2018.02.010. [DOI] [PubMed] [Google Scholar]
  • 304.Maron MS, Finley JJ, Bos JM, Hauser TH, Manning WJ, Haas TS, et al. Prevalence, Clinical Significance, and Natural History of Left Ventricular Apical Aneurysms in Hypertrophic Cardiomyopathy. Circulation . 2008;118(15):1541–1549. doi: 10.1161/CIRCULATIONAHA.108.781401. [DOI] [PubMed] [Google Scholar]
  • 305.Rowin EJ, Maron BJ, Haas TS, Garberich RF, Wang W, Link MS, et al. Hypertrophic Cardiomyopathy With Left Ventricular Apical Aneurysm: Implications for Risk Stratification and Management. J Am Coll Cardiol . 2017;69(7):761–773. doi: 10.1016/j.jacc.2016.11.063. [DOI] [PubMed] [Google Scholar]
  • 306.Weinstock J, Bader YH, Maron MS, Rowin EJ, Link MS. Subcutaneous Implantable Cardioverter Defibrillator in Patients With Hypertrophic Cardiomyopathy: An Initial Experience. J Am Heart Assoc . 2016;5(2):e002488. doi: 10.1161/JAHA.115.002488. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 307.Behr ER, Elliott P, McKenna WJ. Role of Invasive EP Testing in the Evaluation and Management of Hypertrophic Cardiomyopathy. Card Electrophysiol Rev . 2002;6(4):482–486. doi: 10.1023/a:1021161114347. [DOI] [PubMed] [Google Scholar]
  • 308.Chagas C. Nouvelle Espèce de Trypanosomiase Humaine. Bull Soc Path Exotique . 1909;2(6):304–307. [PubMed] [Google Scholar]
  • 309.Dias JCP. História Natural da Doença de Chagas. Arq Bras Cardiol . 1995;65(4):359–366. [PubMed] [Google Scholar]
  • 310.Rassi A, Jr, Rassi A, Marin JA., Neto Chagas disease. Lancet . 2010;375(9723):1388–1402. doi: 10.1016/S0140-6736(10)60061-X. [DOI] [PubMed] [Google Scholar]
  • 311.Dias JC. The Indeterminate Form of Human Chronic Chagas’ Disease A Clinical Epidemiological Review. Rev Soc Bras Med Trop . 1989;22(3):147–156. doi: 10.1590/s0037-86821989000300007. [DOI] [PubMed] [Google Scholar]
  • 312.Barretto AC, Ianni BM. The Undetermined Form of Chagas’ Heart Disease: Concept and Forensic Implications. Sao Paulo Med J . 1995;113(2):797–801. doi: 10.1590/s1516-31801995000200010. [DOI] [PubMed] [Google Scholar]
  • 313.Marin JA, Neto, Cunha E, Neto, Maciel BC, Simões MV. Pathogenesis of Chronic Chagas Heart Disease. Circulation . 2007;115(9):1109–1123. doi: 10.1161/CIRCULATIONAHA.106.624296. [DOI] [PubMed] [Google Scholar]
  • 314.Schmunis GA. Epidemiology of Chagas Disease in Non-Endemic Countries: The Role of International Migration. Mem Inst Oswaldo Cruz . 2007;102(Suppl 1):75–85. doi: 10.1590/s0074-02762007005000093. [DOI] [PubMed] [Google Scholar]
  • 315.Dias JC, Prata A, Correia D. Problems and Perspectives for Chagas Disease Control: In Search of a Realistic Analysis. Rev Soc Bras Med Trop . 2008;41(2):193–196. doi: 10.1590/s0037-86822008000200012. [DOI] [PubMed] [Google Scholar]
  • 316.Moncayo A, Silveira AC. Current Epidemiological Trends for Chagas Disease in Latin America and Future Challenges in Epidemiology, Surveillance and Health Policy. Mem Inst Oswaldo Cruz . 2009;104(Suppl 1):17–30. doi: 10.1590/s0074-02762009000900005. [DOI] [PubMed] [Google Scholar]
  • 317.Rassi A, Jr, Dias JC, Marin JA, Neto, Rassi A. Challenges and opportunities for primary, secondary, and tertiary prevention of Chagas’ disease. Heart . 2009;95(7):524–534. doi: 10.1136/hrt.2008.159624. [DOI] [PubMed] [Google Scholar]
  • 318.Rassi A, Jr, Rassi A, Little WC, Xavier SS, Rassi SG, Rassi AG, et al. Development and Validation of a Risk Score for Predicting Death in Chagas’ Heart Disease. N Engl J Med . 2006;355(8):799–808. doi: 10.1056/NEJMoa053241. [DOI] [PubMed] [Google Scholar]
  • 319.Dias JC, Kloetzel K. The Prognostic Value of the Electrocardiographic Features of Chronic Chagas’ Disease. Rev Inst Med Trop Sao Paulo . 1968;10(3):158–162. [PubMed] [Google Scholar]
  • 320.Espinosa R, Carrasco HA, Belandria F, Fuenmayor AM, Molina C, González R, et al. Life Expectancy Analysis in Patients with Chagas’ Disease: Prognosis After One Decade (1973-1983) Int J Cardiol . 1985;8(1):45–56. doi: 10.1016/0167-5273(85)90262-1. [DOI] [PubMed] [Google Scholar]
  • 321.Rassi A. Curva Atuarial da Taquicardia Ventricular Sustentada na Cardiopatia Chagásica Crônica ; Anais do IV Simpósio Brasileiro de Arritmias Cardíacas; Recife. 1987. [Google Scholar]
  • 322.Santana OO. Arritmia Ventricular e Evolução Clínica de Pacientes na Fase Crônica da Doença de Chagas . Salvador (BA): Universidade Federal da Bahia; 1987. dissertation. [Google Scholar]
  • 323.Guerra HAC. Fatores Prognósticos en la Evolución de la Cardiopatia Chagásica Crônica. Rev. Fed. Argent. Cardiol. . 1988;17:247–250. [Google Scholar]
  • 324.Rassi SG, Rassi A, Jr, Rassi AG, Lima AMC, Jatene JA, Rassi A. Avaliação da Síncope e da Pré-Sincope na Cardiopatia Chagásica Crônica através da Estimulação Elétrica Programada. In: Anais do II Congresso da Sociedade Latino-Americana de Estimulação Cardíaca . Porto Alegre (RS): Sociedade Latino-Americana de Estimulação Cardíaca; 1989. [Google Scholar]
  • 325.Rassi SG, Rassi A, Jr, Jatene JA, Lima AMC, Ghannam VM, Rassi A. Significado Clínico da Indução de Fibrilação Ventricular, Flutter Ventricular e Taquicardia Ventricular Polimórfica Sustentados ao Estudo Eletrofisiológico. Arq Bras Cardiol . 1991;57(Suppl C):2 [Google Scholar]
  • 326.Rassi A, Rassi A, Jr, Faria GHDC, Rassi AG, Rassi SG, Callender K, et al. História Natural do Bloqueio Atrioventricular Total de Etiologia Chagásica. Arq Bras Cardiol . 1992;59(Suppl II) 191. [Google Scholar]
  • 327.Rassi A, Jr, Rassi AG, Rassi SG, Rassi L, Jr, Rassi A. Relação entre Sintomas, Disfunção Ventricular e Arritmia Ventricular na Cardiopatia Chagásica Crônica. Arq Bras Cardiol . 1992;59(Suppl II) 182. [Google Scholar]
  • 328.Carrasco HA, Parada H, Guerrero L, Duque M, Durán D, Molina C. Prognostic Implications of Clinical, Electrocardiographic and Hemodynamic Findings in Chronic Chagas’ Disease. Int J Cardiol . 1994;43(1):27–38. doi: 10.1016/0167-5273(94)90087-6. [DOI] [PubMed] [Google Scholar]
  • 329.Sosa E, Scanavacca M. Rev Soc Cardiol . 2. Vol. 4. Estado de Säo Paulo; 1994. Estudo Eletrofisiológico na Cardiopatia Chagásica Crônica; pp. 168–176. [Google Scholar]
  • 330.Moraes AP, Moffa PJ, Sosa EA, Belloti G, Pastore CA, Lima EV, et al. Eletrocardiograma de Alta Resolução na Cardiopatia Chagásica Crônica. Rev Soc Cardiol Estado São Paulo . 1994;4:177–182. [Google Scholar]
  • 331.Martinelli M, Filho, Sosa E, Nishioka S, Scanavacca M, Bellotti G, Pileggi F. Clinical and Electrophysiologic Features of Syncope in Chronic Chagasic Heart Disease. J Cardiovasc Electrophysiol . 1994;5(7):563–570. doi: 10.1111/j.1540-8167.1994.tb01297.x. [DOI] [PubMed] [Google Scholar]
  • 332.Mady C, Cardoso RH, Barretto AC, Luz PL, Bellotti G, Pileggi F. Survival and Predictors of Survival in Patients with Congestive Heart Failure due to Chagas’ cardiomYopathy. Circulation . 1994;90(6):3098–3102. doi: 10.1161/01.cir.90.6.3098. [DOI] [PubMed] [Google Scholar]
  • 333.Bestetti RB, Dalbo CM, Freitas OC, Teno LA, Castilho OT, Oliveira JS. Noninvasive Predictors of Mortality for Patients with Chagas’ Heart Disease: A Multivariate Stepwise Logistic Regression Study. Cardiology . 1994;84(4-5):261–267. doi: 10.1159/000176409. [DOI] [PubMed] [Google Scholar]
  • 334.Paola AA, Gomes JA, Terzian AB, Miyamoto MH, Martinez Fo EE. Ventricular Tachycardia During Exercise Testing as a Predictor of Sudden Death in Patients with Chronic Chagasic Cardiomyopathy and Ventricular Arrhythmias. Br Heart J . 1995;74(3):293–295. doi: 10.1136/hrt.74.3.293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 335.Bestetti RB, Dalbo CM, Arruda CA, Correia D, Filho, Freitas OC. Predictors of Sudden Cardiac Death for Patients with Chagas’ Disease: A Hospital-Derived Cohort Study. Cardiology . 1996;87(6):481–487. doi: 10.1159/000177142. [DOI] [PubMed] [Google Scholar]
  • 336.Silva RMFL. Valor Preditivo das Variáveis Clínicas e Eletrofisiológicas nos Pacientes com Cardiopatia Chagásica Crônica e Taquicardia Ventricular Não-Sustentada Análise Terapêutica . São Paulo (SP): Universidade Federal de São Paulo; 1997. dissertation. [Google Scholar]
  • 337.Garzon SAC, Lorga AM, Jacob JLB, Greco OT, Vitola J, Machado NCS, et al. Predictors of Mortality in Chronic Chagas’ Heart Disease: Long Term Follow up of 987 Subjects for up to 22 Years. J Am Coll Cardiol . 1998;31(suppl C) 107C. [Google Scholar]
  • 338.Rassi A, Jr, Waktare JEP, Rassi SG, Rassi A. Chagas heart disease: long term prognostic significance of nonsustained tachycardia and left ventricular dysfunction. Pace . 1999;22(Part II):862–870. [Google Scholar]
  • 339.Carrasco HA, Guerrero L, Parada H, Molina C, Vegas E, Chuecos R. Ventricular Arrhythmias and Left Ventricular Myocardial Function in Chronic Chagasic Patients. Int J Cardiol . 1990;28(1):35–41. doi: 10.1016/0167-5273(90)90006-q. [DOI] [PubMed] [Google Scholar]
  • 340.Lopes ER. Sudden Death in Patients with Chagas Disease. Mem Inst Oswaldo Cruz . 1999;94(Suppl 1):321–324. doi: 10.1590/s0074-02761999000700061. [DOI] [PubMed] [Google Scholar]
  • 341.Rassi A, Jr, Rassi SG, Rassi A. Sudden Death in Chagas’ Disease. Arq Bras Cardiol . 2001;76(1):75–96. doi: 10.1590/s0066-782x2001000100008. [DOI] [PubMed] [Google Scholar]
  • 342.Mendoza I, Moleiro F, Marques Morte Súbita na Doença de Chagas. Arq Bras Cardiol . 1992;59:3–4. [PubMed] [Google Scholar]
  • 343.Rassi A, Jr, Rassi AG, Rassi SG, Rassi L, Jr, Rassi A. Arritmias Ventriculares na Doença de Chagas. Particularidades Diagnósticas, Prognósticas e Terapêuticas. Arq Bras Cardiol . 1995;65(4):377–387. [PubMed] [Google Scholar]
  • 344.Rocha MO, Ribeiro ALP. A Risk Score for Predicting Death in Chagas’ Heart Disease. N Engl J Med . 2006;355(23):2488–2489. doi: 10.1056/NEJMc062580. [DOI] [PubMed] [Google Scholar]
  • 345.Andrade JP, Marin JA, Neto, Paola AAV, Vilas-Boas F, Oliveira GMM, Bacal F, et al. I Diretriz Latino Americana para o Diagnóstico e Tratamento da Cardiopatia Chagásica. Arq Bras Cardiol . 2011;97(2):1–48. doi: 10.1590/S0066-782X2011001600001. [DOI] [PubMed] [Google Scholar]
  • 346.Marcus FI, McKenna WJ, Sherrill D, Basso C, Bauce B, Bluemke DA, et al. Diagnosis of Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia: Proposed Modification of the Task Force Criteria. Eur Heart J . 2010;31(7):806–814. doi: 10.1093/eurheartj/ehq025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 347.Basso C, Corrado D, Marcus FI, Nava A, Thiene G. Arrhythmogenic Right Ventricular Cardiomyopathy. Lancet . 2009;373(9671):1289–1300. doi: 10.1016/S0140-6736(09)60256-7. [DOI] [PubMed] [Google Scholar]
  • 348.Basso C, Corrado D, Thiene G. Cardiovascular Causes of Sudden Death in Young Individuals Including Athletes. Cardiol Rev . 1999;7(3):127–135. doi: 10.1097/00045415-199905000-00009. [DOI] [PubMed] [Google Scholar]
  • 349.Marcus FI, Zareba W, Calkins H, Towbin JA, Basso C, Bluemke DA, et al. Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia Clinical Presentation and Diagnostic Evaluation: Results from the North American Multidisciplinary Study. Heart Rhythm . 2009;6(7):984–992. doi: 10.1016/j.hrthm.2009.03.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 350.Link MS, Laidlaw D, Polonsky B, Zareba W, McNitt S, Gear K, et al. Ventricular Arrhythmias in the North American Multidisciplinary Study of ARVC: Predictors, Characteristics, and Treatment. J Am Coll Cardiol . 2014;64(2):119–125. doi: 10.1016/j.jacc.2014.04.035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 351.Piccini JP, Dalal D, Roguin A, Bomma C, Cheng A, Prakasa K, et al. Predictors of Appropriate Implantable Defibrillator Therapies in Patients with Arrhythmogenic Right Ventricular Dysplasia. Heart Rhythm . 2005;2(11):1188–1194. doi: 10.1016/j.hrthm.2005.08.022. [DOI] [PubMed] [Google Scholar]
  • 352.Hoffmayer KS, Machado ON, Marcus GM, Yang Y, Johnson CJ, Ermakov S, et al. Electrocardiographic Comparison of Ventricular Arrhythmias in Patients with Arrhythmogenic Right Ventricular Cardiomyopathy and Right Ventricular Outflow Tract Tachycardia. J Am Coll Cardiol . 2011;58(8):831–838. doi: 10.1016/j.jacc.2011.05.017. [DOI] [PubMed] [Google Scholar]
  • 353.Kamath GS, Zareba W, Delaney J, Koneru JN, McKenna W, Gear K, et al. Value of the Signal-Averaged Electrocardiogram in Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia. Heart Rhythm . 2011;8(2):256–262. doi: 10.1016/j.hrthm.2010.10.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 354.Riele AS, Bhonsale A, James CA, Rastegar N, Murray B, Burt JR, et al. Incremental Value of Cardiac Magnetic Resonance Imaging in Arrhythmic Risk Stratification of Arrhythmogenic Right Ventricular Dysplasia/Cardiomyopathy-Associated Desmosomal Mutation Carriers. J Am Coll Cardiol . 2013;62(19):1761–1769. doi: 10.1016/j.jacc.2012.11.087. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 355.Marcus FI, McKenna WJ, Sherrill D, Basso C, Bauce B, Bluemke DA, et al. Diagnosis of Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia: Proposed Modification of the Task Force Criteria. Circulation . 2010;121(13):1533–1541. doi: 10.1161/CIRCULATIONAHA.108.840827. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 356.Quarta G, Muir A, Pantazis A, Syrris P, Gehmlich K, Garcia-Pavia P, et al. Familial Evaluation in Arrhythmogenic Right Ventricular Cardiomyopathy: Impact of Genetics and Revised Task Force Criteria. Circulation . 2011;123(23):2701–2709. doi: 10.1161/CIRCULATIONAHA.110.976936. [DOI] [PubMed] [Google Scholar]
  • 357.Saeed M, Homoud MK, Wang PJ, Estes NA, 3rd, Link MS. Role of Invasive Electrophysiologic Testing in Risk Stratification for Sudden Cardiac Death. J Invasive Cardiol . 2001;13(11):758–762. [PubMed] [Google Scholar]
  • 358.Wichter T, Borggrefe M, Haverkamp W, Chen X, Breithardt G. Efficacy of Antiarrhythmic Drugs in Patients with Arrhythmogenic Right Ventricular Disease. Results in Patients with Inducible and Noninducible Ventricular Tachycardia. Circulation . 1992;86(1):29–37. doi: 10.1161/01.cir.86.1.29. [DOI] [PubMed] [Google Scholar]
  • 359.Marcus GM, Glidden DV, Polonsky B, Zareba W, Smith LM, Cannom DS, et al. Efficacy of Antiarrhythmic Drugs in Arrhythmogenic Right Ventricular Cardiomyopathy: A Report from the North American ARVC Registry. J Am Coll Cardiol . 2009;54(7):609–615. doi: 10.1016/j.jacc.2009.04.052. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 360.Philips B, Riele AS, Sawant A, Kareddy V, James CA, Murray B, et al. Outcomes and Ventricular Tachycardia Recurrence Characteristics after Epicardial Ablation of Ventricular Tachycardia in arrhythmogenic Right Ventricular Dysplasia/Cardiomyopathy. Heart Rhythm . 2015;12(4):716–725. doi: 10.1016/j.hrthm.2014.12.018. [DOI] [PubMed] [Google Scholar]
  • 361.Schinkel AF. Implantable Cardioverter Defibrillators in Arrhythmogenic Right Ventricular Dysplasia/Cardiomyopathy: Patient Outcomes, Incidence of Appropriate and Inappropriate Interventions, and Complications. Circ Arrhythm Electrophysiol . 2013;6(3):562–568. doi: 10.1161/CIRCEP.113.000392. [DOI] [PubMed] [Google Scholar]
  • 362.Lemola K, Brunckhorst C, Helfenstein U, Oechslin E, Jenni R, Duru F. Predictors of Adverse Outcome in Patients with Arrhythmogenic Right Ventricular Dysplasia/Cardiomyopathy: Long Term Experience of a Tertiary Care Centre. Heart . 2005;91(9):1167–1172. doi: 10.1136/hrt.2004.038620. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 363.Kalavakunta JK, Tokala H, Gosavi A, Gupta V. Left Ventricular Noncompaction and Myocardial Fibrosis: A Case Report. Int Arch Med . 2010;3(20) doi: 10.1186/1755-7682-3-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 364.Cadrin-Tourigny J, Bosman LP, Nozza A, Wang W, Tadros R, Bhonsale A, et al. A New Prediction Model for Ventricular Arrhythmias in Arrhythmogenic Right Ventricular Cardiomyopathy. Eur Heart J . 2019;40(23):1850–1858. doi: 10.1093/eurheartj/ehz103. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 365.Corrado D, Wichter T, Link MS, Hauer RN, Marchlinski FE, Anastasakis A, et al. Treatment of Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia: An International Task Force Consensus Statement. Circulation . 2015;132(5):441–453. doi: 10.1161/CIRCULATIONAHA.115.017944. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 366.Bhatia NL, Tajik AJ, Wilansky S, Steidley DE, Mookadam F. Isolated Noncompaction of the Left Ventricular Myocardium in Adults: A Systematic Overview. J Card Fail . 2011;17(9):771–778. doi: 10.1016/j.cardfail.2011.05.002. [DOI] [PubMed] [Google Scholar]
  • 367.Oechslin E, Jenni R. Left Ventricular Non-Compaction Revisited: A Distinct Phenotype with Genetic Heterogeneity? Eur Heart J . 2011;32(12):1446–1456. doi: 10.1093/eurheartj/ehq508. [DOI] [PubMed] [Google Scholar]
  • 368.Lofiego C, Biagini E, Pasquale F, Ferlito M, Rocchi G, Perugini E, et al. Wide Spectrum of Presentation and Variable Outcomes of Isolated Left Ventricular Non-Compaction. Heart . 2007;93(1):65–71. doi: 10.1136/hrt.2006.088229. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 369.Jenni R, Oechslin E, Schneider J, Jost CA, Kaufmann PA. Echocardiographic and pathoanatomical characteristics of isolated left ventricular non-compaction: a step towards classification as a distinct cardiomyopathy. Heart . 2001;86(6):666–671. doi: 10.1136/heart.86.6.666. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 370.Petersen SE, Selvanayagam JB, Wiesmann F, Robson MD, Francis JM, Anderson RH, et al. Left Ventricular Non-Compaction: Insights from Cardiovascular Magnetic Resonance Imaging. J Am Coll Cardiol . 2005;46(1):101–105. doi: 10.1016/j.jacc.2005.03.045. [DOI] [PubMed] [Google Scholar]
  • 371.Weiford BC, Subbarao VD, Mulhern KM. Noncompaction of the Ventricular Myocardium. Circulation . 2004;109(24):2965–2971. doi: 10.1161/01.CIR.0000132478.60674.D0. [DOI] [PubMed] [Google Scholar]
  • 372.Oechslin EN, Jost CA, Rojas JR, Kaufmann PA, Jenni R. Long-Term Follow-Up of 34 Adults with Isolated Left Ventricular Noncompaction: A Distinct Cardiomyopathy with Poor Prognosis. J Am Coll Cardiol . 2000;36(2):493–500. doi: 10.1016/s0735-1097(00)00755-5. [DOI] [PubMed] [Google Scholar]
  • 373.Steffel J, Kobza R, Namdar M, Wolber T, Brunckhorst C, Lüscher TF, et al. Electrophysiological Findings in Patients with Isolated Left Ventricular Non-Compaction. Europace . 2009;11(9):1193–1200. doi: 10.1093/europace/eup187. [DOI] [PubMed] [Google Scholar]
  • 374.Caliskan K, Kardos A, Szili-Torok T. Empty Handed: A Call For an International Registry of Risk Stratification to Reduce the ‘Sudden-Ness’ of Death in Patients with Non-Compaction Cardiomyopathy. Europace . 2009;11(9):1138–1139. doi: 10.1093/europace/eup228. [DOI] [PubMed] [Google Scholar]
  • 375.Sohns C, Ouyang F, Volkmer M, Metzner A, Nürnberg JH, Ventura R, et al. Therapy of Ventricular Arrhythmias in Patients Suffering from Isolated Left Ventricular Non-Compaction Cardiomyopathy. Europace . 2019;21(6):961–969. doi: 10.1093/europace/euz016. [DOI] [PubMed] [Google Scholar]
  • 376.Caliskan K, Szili-Torok T, Theuns DA, Kardos A, Geleijnse ML, Balk AH, et al. Indications and Outcome of Implantable Cardioverter-Defibrillators for Primary and Secondary Prophylaxis in Patients with Noncompaction Cardiomyopathy. J Cardiovasc Electrophysiol . 2011;22(8):898–904. doi: 10.1111/j.1540-8167.2011.02015.x. [DOI] [PubMed] [Google Scholar]
  • 377.Jefferies JL, Wilkinson JD, Sleeper LA, Colan SD, Lu M, Pahl E, et al. Cardiomyopathy Phenotypes and Outcomes for Children With Left Ventricular Myocardial Noncompaction: Results From the Pediatric Cardiomyopathy Registry. J Card Fail . 2015;21(11):877–884. doi: 10.1016/j.cardfail.2015.06.381. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 378.Priori SG, Blomström-Lundqvist C, Mazzanti A, Blom N, Borggrefe M, Camm J, et al. 2015 ESC Guidelines for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC). Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC) Eur Heart J . 2015;36(41):2793–2867. doi: 10.1093/eurheartj/ehv316. [DOI] [PubMed] [Google Scholar]
  • 379.Jons C, Moss AJ, Goldenberg I, Liu J, McNitt S, Zareba W, et al. Risk of Fatal Arrhythmic Events in Long QT Syndrome Patients After Syncope. J Am Coll Cardiol . 2010;55(8):783–788. doi: 10.1016/j.jacc.2009.11.042. [DOI] [PubMed] [Google Scholar]
  • 380.Hobbs JB, Peterson DR, Moss AJ, McNitt S, Zareba W, Goldenberg I, et al. Risk of Aborted Cardiac Arrest or Sudden Cardiac Death During Adolescence in the Long-QT Syndrome. JAMA . 2006;296(10):1249–1254. doi: 10.1001/jama.296.10.1249. [DOI] [PubMed] [Google Scholar]
  • 381.Mazzanti A, Maragna R, Vacanti G, Monteforte N, Bloise R, Marino M, et al. Interplay Between Genetic Substrate, QTc Duration, and Arrhythmia Risk in Patients With Long QT Syndrome. J Am Coll Cardiol . 2018;71(15):1663–1671. doi: 10.1016/j.jacc.2018.01.078. [DOI] [PubMed] [Google Scholar]
  • 382.Mazzanti A, Kanthan A, Monteforte N, Memmi M, Bloise R, Novelli V, et al. Novel Insight Into the Natural History of Short QT Syndrome. J Am Coll Cardiol . 2014;63(13):1300–1308. doi: 10.1016/j.jacc.2013.09.078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 383.Nannenberg EA, Sijbrands EJ, Dijksman LM, Alders M, van Tintelen JP, Birnie M, et al. Mortality of Inherited Arrhythmia Syndromes: Insight Into Their Natural History. Circ Cardiovasc Genet . 2012;5(2):183–189. doi: 10.1161/CIRCGENETICS.111.961102. [DOI] [PubMed] [Google Scholar]
  • 384.Priori SG, Wilde AA, Horie M, Cho Y, Behr ER, Berul C, et al. HRS/EHRA/APHRS Expert Consensus Statement on the Diagnosis and Management of Patients with Inherited Primary Arrhythmia Syndromes: Document Endorsed by HRS, EHRA, and APHRS in May 2013 and by ACCF, AHA, PACES, and AEPC in June 2013. Heart Rhythm . 2013;10(12):1932–1963. doi: 10.1016/j.hrthm.2013.05.014. [DOI] [PubMed] [Google Scholar]
  • 385.Priori SG, Gasparini M, Napolitano C, Della Bella P, Ottonelli AG, Sassone B, et al. Risk Stratification in Brugada Syndrome: Results of the PRELUDE (PRogrammed ELectrical stimUlation preDictive valuE) registry. J Am Coll Cardiol . 2012;59(1):37–45. doi: 10.1016/j.jacc.2011.08.064. [DOI] [PubMed] [Google Scholar]
  • 386.Probst V, Veltmann C, Eckardt L, Meregalli PG, Gaita F, Tan HL, et al. Long-Term Prognosis of patients Diagnosed with Brugada Syndrome: Results from the FINGER Brugada Syndrome Registry. Circulation . 2010;121(5):635–643. doi: 10.1161/CIRCULATIONAHA.109.887026. [DOI] [PubMed] [Google Scholar]
  • 387.Sroubek J, Probst V, Mazzanti A, Delise P, Hevia JC, Ohkubo K, et al. Programmed Ventricular Stimulation for Risk Stratification in the Brugada Syndrome: A Pooled Analysis. Circulation . 2016;133(7):622–630. doi: 10.1161/CIRCULATIONAHA.115.017885. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 388.Zhang P, Tung R, Zhang Z, Sheng X, Liu Q, Jiang R, et al. Characterization of the Epicardial Substrate for Catheter Ablation of Brugada Syndrome. Heart Rhythm . 2016;13(11):2151–2158. doi: 10.1016/j.hrthm.2016.07.025. [DOI] [PubMed] [Google Scholar]
  • 389.Talib AK, Takagi M, Shimane A, Nakano M, Hayashi T, Okajima K, et al. Efficacy of Endocardial Ablation of Drug-Resistant Ventricular Fibrillation in Brugada Syndrome: Long-Term Outcome. Circ Arrhythm Electrophysiol . 2018;11(8):e005631. doi: 10.1161/CIRCEP.117.005631. [DOI] [PubMed] [Google Scholar]
  • 390.Roston TM, Yuchi Z, Kannankeril PJ, Hathaway J, Vinocur JM, Etheridge SP, et al. The Clinical and Genetic Spectrum of Catecholaminergic Polymorphic Ventricular Tachycardia: Findings from an International Multicentre Registry. Europace . 2018;20(3):541–547. doi: 10.1093/europace/euw389. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 391.Priori SG, Blomström-Lundqvist C, Mazzanti A, Blom N, Borggrefe M, Camm J, et al. 2015 ESC Guidelines for the management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC)Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC) Europace . 2015;17(11):1601–1687. doi: 10.1093/europace/euv319. [DOI] [PubMed] [Google Scholar]
  • 392.Roston TM, Jones K, Hawkins NM, Bos JM, Schwartz PJ, Perry F, et al. Implantable Cardioverter-Defibrillator Use in Catecholaminergic Polymorphic Ventricular Tachycardia: A Systematic Review. Heart Rhythm . 2018;15(12):1791–1799. doi: 10.1016/j.hrthm.2018.06.046. [DOI] [PubMed] [Google Scholar]
  • 393.Tan AY, Ellenbogen K. Ventricular Arrhythmias in Apparently Normal Hearts: Who Needs an Implantable Cardiac Defibrillator? Card Electrophysiol Clin . 2016;8(3):613–621. doi: 10.1016/j.ccep.2016.04.010. [DOI] [PubMed] [Google Scholar]
  • 394.Chokr MO, Darrieux FC, Hardy CA, Hachul DT, Britto AV, Melo SL, et al. Short-Coupled Variant of “Torsades de Pointes” and Polymorphic Ventricular Tachycardia. Arq Bras Cardiol . 2014;102(6):60–64. doi: 10.5935/abc.20140075. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 395.Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS Guideline for Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm . 2018;15(10):73–189. doi: 10.1016/j.hrthm.2017.10.036. [DOI] [PubMed] [Google Scholar]
  • 396.Myerburg RJ, Kessler KM, Estes D, Conde CA, Luceri RM, Zaman L, et al. Long-Term Survival After Prehospital Cardiac Arrest: Analysis of Outcome During an 8 Year Study. Circulation . 1984;70(4):538–546. doi: 10.1161/01.cir.70.4.538. [DOI] [PubMed] [Google Scholar]
  • 397.Furukawa T, Rozanski JJ, Nogami A, Moroe K, Gosselin AJ, Lister JW. Time-Dependent Risk of and Predictors for Cardiac Arrest Recurrence in Survivors of Out-of-Hospital Cardiac Arrest with Chronic Coronary Artery Disease. Circulation . 1989;80(3):599–608. doi: 10.1161/01.cir.80.3.599. [DOI] [PubMed] [Google Scholar]
  • 398.Priori SG, Aliot E, Blomstrom-Lundqvist C, Bossaert L, Breithardt G, Brugada P, et al. Task Force on Sudden Cardiac Death of the European Society of Cardiology. Eur Heart J . 2001;22(16):1374–1450. doi: 10.1053/euhj.2001.2824. [DOI] [PubMed] [Google Scholar]
  • 399.Peck KY, Lim YZ, Hopper I, Krum H. Medical Therapy Versus Implantable Cardioverter -Defibrillator in Preventing Sudden Cardiac death in Patients with left Ventricular Systolic Dysfunction and Heart Failure: A Meta-Analysis of > 35,000 Patients. Int J Cardiol . 2014;173(2):197–203. doi: 10.1016/j.ijcard.2014.02.014. [DOI] [PubMed] [Google Scholar]
  • 400.Sarrias A, Bayes-Genis A. Is Sacubitril/Valsartan (Also) an Antiarrhythmic Drug? Circulation . 2018;138(6):551–553. doi: 10.1161/CIRCULATIONAHA.118.034755. [DOI] [PubMed] [Google Scholar]
  • 401.Greene HL. The CASCADE Study: Randomized Antiarrhythmic Drug Therapy in Survivors of Cardiac Arrest in Seattle. CASCADE Investigators. Am J Cardiol . 1993;72(16):70–74. doi: 10.1016/0002-9149(93)90966-g. [DOI] [PubMed] [Google Scholar]
  • 402.Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular Arrhythmias. N Engl J Med . 1997;337(22):1576–1583. doi: 10.1056/NEJM199711273372202. [DOI] [PubMed] [Google Scholar]
  • 403.Domanski MJ, Sakseena S, Epstein AE, Hallstrom AP, Brodsky MA, Kim S, et al. Relative Effectiveness of the Implantable Cardioverter-Defibrillator and Antiarrhythmic Drugs in Patients with Varying Degrees of Left Ventricular Dysfunction Who Have Survived Malignant Ventricular Arrhythmias. AVID Investigators. Antiarrhythmics Versus Implantable Defibrillators. J Am Coll Cardiol . 1999;34(4):1090–1095. doi: 10.1016/s0735-1097(99)00327-7. [DOI] [PubMed] [Google Scholar]
  • 404.Connolly SJ, Gent M, Roberts RS, Dorian P, Roy D, Sheldon RS, et al. Canadian Implantable Defibrillator Study (CIDS): A Randomized Trial of the Implantable Cardioverter Defibrillator Against Amiodarone. Circulation . 2000;101(11):1297–1302. doi: 10.1161/01.cir.101.11.1297. [DOI] [PubMed] [Google Scholar]
  • 405.Sheldon R, Connolly S, Krahn A, Roberts R, Gent M, Gardner M. Identification of Patients Most Likely to Benefit from Implantable Cardioverter-Defibrillator Therapy: The Canadian Implantable Defibrillator Study. Circulation . 2000;101(14):1660–1664. doi: 10.1161/01.cir.101.14.1660. [DOI] [PubMed] [Google Scholar]
  • 406.Kuck KH, Cappato R, Siebels J, Rüppel R. Randomized Comparison of Antiarrhythmic Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Cardiac Arrest: The Cardiac Arrest Study Hamburg (CASH) Circulation . 2000;102(7):748–754. doi: 10.1161/01.cir.102.7.748. [DOI] [PubMed] [Google Scholar]
  • 407.Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, et al. Meta-Analysis of the Implantable Cardioverter Defibrillator Secondary Prevention Trials. AVID, CASH and CIDS Studies. Antiarrhythmics vs Implantable Defibrillator Study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J . 2000;21(24):2071–2078. doi: 10.1053/euhj.2000.2476. [DOI] [PubMed] [Google Scholar]
  • 408.Lau EW, Griffith MJ, Pathmanathan RK, Ng GA, Clune MM, Cooper J, et al. The Midlands Trial of Empirical Amiodarone Versus Electrophysiology-Guided Interventions and Implantable Cardioverter-Defibrillators (MAVERIC): A Multi-Centre Prospective Randomised Clinical Trial on the Secondary Prevention of Sudden Cardiac Death. Europace . 2004;6(4):257–266. doi: 10.1016/j.eupc.2004.03.009. [DOI] [PubMed] [Google Scholar]
  • 409.Lieve KV, Wilde AA. Inherited Ion Channel Diseases: A Brief Review. Europace . 2015;17(Suppl 20):1–6. doi: 10.1093/europace/euv105. [DOI] [PubMed] [Google Scholar]
  • 410.Ackerman MJ, Priori SG, Willems S, Berul C, Brugada R, Calkins H, et al. HRS/EHRA expert consensus statement on the state of genetic testing for the channelopathies and cardiomyopathies this document was developed as a partnership between the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA) Heart Rhythm . 2011;8(8):1308–1339. doi: 10.1016/j.hrthm.2011.05.020. [DOI] [PubMed] [Google Scholar]
  • 411.Schwartz PJ, Ackerman MJ, Wilde AAM. Channelopathies as Causes of Sudden Cardiac Death. Card Electrophysiol Clin . 2017;9(4):537–549. doi: 10.1016/j.ccep.2017.07.005. [DOI] [PubMed] [Google Scholar]
  • 412.Fuganti CJ, Melo CS, Moraes AV, Jr, Pachon-Mateos JC, Pereira WL, Galvão SS, Filho, et al. Diretrizes Brasileiras de Dispositivos Cardíacos Eletrônicos Implantáveis do Departamento de Estimulação Cardíaca Artificial (DECA) da Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV) Relampa . 2015;28(2) Supl:1–62. [Google Scholar]
  • 413.Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS Guideline for Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation . 2018;138(13):210–271. doi: 10.1161/CIR.0000000000000548. [DOI] [PubMed] [Google Scholar]
  • 414.Olshansky B, Hahn EA, Hartz VL, Prater SP, Mason JW. Clinical Significance of Syncope in the Electrophysiologic Study Versus Electrocardiographic Monitoring (ESVEM) Trial. The ESVEM Investigators. Am Heart J . 1999;137(5):878–886. doi: 10.1016/s0002-8703(99)70412-6. [DOI] [PubMed] [Google Scholar]
  • 415.Link MS, Kim KM, Homoud MK, Estes NA, 3rd, Wang PJ. Long-Term Outcome of Patients with Syncope Associated with Coronary Artery Disease and a Nondiagnostic Electrophysiologic Evaluation. Am J Cardiol . 1999;83(9):1334–1337. doi: 10.1016/s0002-9149(99)00096-x. [DOI] [PubMed] [Google Scholar]
  • 416.Mittal S, Hao SC, Iwai S, Stein KM, Markowitz SM, Slotwiner DJ, et al. Significance of Inducible Ventricular Fibrillation in Patients with Coronary Artery Disease and Unexplained Syncope. J Am Coll Cardiol . 2001;38(2):371–376. doi: 10.1016/s0735-1097(01)01379-1. [DOI] [PubMed] [Google Scholar]
  • 417.Berul CI, Moak JP. Implantable Cardioverter-Defibrillators in Children: Innovation to Design a Pediatric ICD. J Innov Card Rhythm Manag . 2011;2:179–185. doi: 10.19102/icrm.2011.020202. [DOI] [Google Scholar]
  • 418.DeWitt ES, Triedman JK, Cecchin F, Mah DY, Abrams DJ, Walsh EP, et al. Time Dependence of Risks and Benefits in Pediatric Primary Prevention Implantable Cardioverter-Defibrillator Therapy. Circ Arrhythm Electrophysiol . 2014;7(6):1057–1063. doi: 10.1161/CIRCEP.114.001569. [DOI] [PubMed] [Google Scholar]
  • 419.Magalhães LP, Guimarães I, Melo SL, Mateo E, Andalaft RB, Xavier L, et al. Diretriz de Arritmias Cardíacas em Crianças e Cardiopatias Congênitas SOBRAC e DCC - CP. Arq Bras Cardiol . 2016;107(1) Suppl 3:1–58. doi: 10.5935/abc.20160103. [DOI] [PubMed] [Google Scholar]
  • 420.Atallah J, Erickson CC, Cecchin F, Dubin AM, Law IH, Cohen MI, et al. Multi-Institutional Study of Implantable Defibrillator Lead Performance in Children and Young Adults: Results of the Pediatric Lead Extractability and Survival Evaluation (PLEASE) Study. Circulation . 2013;127(24):2393–2402. doi: 10.1161/CIRCULATIONAHA.112.001120. [DOI] [PubMed] [Google Scholar]
  • 421.Moss AJ, Schuger C, Beck CA, Brown MW, Cannom DS, Daubert JP, et al. Reduction in Inappropriate Therapy and Mortality Through ICD Programming. N Engl J Med . 2012;367(24):2275–2283. doi: 10.1056/NEJMoa1211107. [DOI] [PubMed] [Google Scholar]
  • 422.Garnreiter JM, Pilcher TA, Etheridge SP, Saarel EV. Inappropriate ICD Shocks in Pediatrics and Congenital Heart Disease Patients: Risk Factors and Programming Strategies. Heart Rhythm . 2015;12(5):937–942. doi: 10.1016/j.hrthm.2015.01.028. [DOI] [PubMed] [Google Scholar]
  • 423.Sears SF, Hazelton AG, St Amant J, Matchett M, Kovacs A, Vazquez LD, et al. Quality of Life in Pediatric Patients with Implantable Cardioverter Defibrillators. Am J Cardiol . 2011;107(7):1023–1027. doi: 10.1016/j.amjcard.2010.11.027. [DOI] [PubMed] [Google Scholar]
  • 424.DeMaso DR, Lauretti A, Spieth L, van der Feen JR, Jay KS, Gauvreau K, et al. Psychosocial Factors and Quality of Life in Children and Adolescents with Implantable Cardioverter-Defibrillators. Am J Cardiol . 2004;93(5):582–587. doi: 10.1016/j.amjcard.2003.11.022. [DOI] [PubMed] [Google Scholar]
  • 425.Uzark K, Jones K, Slusher J, Limbers CA, Burwinkle TM, Varni JW. Quality of Life in Children with Heart Disease as Perceived by Children and Parents. Pediatrics . 2008;121(5):1060–1067. doi: 10.1542/peds.2006-3778. [DOI] [PubMed] [Google Scholar]
  • 426.Lemon J, Edelman S. Psychological Adaptation to ICDs and the Influence of Anxiety Sensitivity. Psychol Health Med . 2007;12(2):163–171. doi: 10.1080/13548500500448478. [DOI] [PubMed] [Google Scholar]
  • 427.Vazquez LD, Kuhl EA, Shea JB, Kirkness A, Lemon J, Whalley D, et al. Age-Specific Differences in Women with Implantable Cardioverter Defibrillators: An International Multi Center Study. Pacing Clin Electrophysiol . 2008;31(12):1528–1534. doi: 10.1111/j.1540-8159.2008.01223.x. [DOI] [PubMed] [Google Scholar]
  • 428.Walker RL, Campbell KA, Sears SF, Glenn BA, Sotile R, Curtis AB, et al. Women and the Implantable Cardioverter Defibrillator: A Lifespan Perspective on Key Psychosocial Issues. Clin Cardiol . 2004;27(10):543–546. doi: 10.1002/clc.4960271019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 429.Kaski JP, Esteban MTT, Lowe M, Sporton S, Rees P, Deanfield JE, McKenna WJ, Elliott PM, et al. Outcomes After Implantable Cardioverter-Defibrillator Treatment in Children with Hypertrophic Cardiomyopathy. Heart . 2007;93(3):372–374. doi: 10.1136/hrt.2006.094730. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 430.Stephenson EA, Batra AS, Knilans TK, Gow RM, Gradaus R, Balaji S, et al. A Multicenter Experience with Novel Implantable Cardioverter Defibrillator Configurations in the Pediatric and Congenital Heart Disease Population. J Cardiovasc Electrophysiol . 2006;17(1):41–46. doi: 10.1111/j.1540-8167.2005.00271.x. [DOI] [PubMed] [Google Scholar]
  • 431.Kalra Y, Radbill AE, Johns JA, Fish FA, Kannankeril PJ. Antitachycardia Pacing Reduces Appropriate and Inappropriate Shocks in Children and Congenital Heart Disease Patients. Heart Rhythm . 2012;9(11):1829–1834. doi: 10.1016/j.hrthm.2012.06.042. [DOI] [PubMed] [Google Scholar]
  • 432.Van Hare GF, Javitz H, Carmelli D, Saul JP, Tanel RE, Fischbach PS, et al. Prospective Assessment After Pediatric Cardiac Ablation: Demographics, Medical Profiles, and Initial Outcomes. J Cardiovasc Electrophysiol . 2004;15(7):759–770. doi: 10.1046/j.1540-8167.2004.03645.x. [DOI] [PubMed] [Google Scholar]
  • 433.Seixo F, Rossi R, Adração P, Cavaco D, Santos KR, Morgado FB, et al. Percutaneous Catheter Ablation of Arrhythmias in Children. Rev Port Cardiol . 2008;27(11):1419–1426. [PubMed] [Google Scholar]
  • 434.Lieberman R, Havel WJ, Rashba E, DeGroot PJ, Stromberg K, Shorofsky SR. Acute Defibrillation Performance of a Novel, Non-Transvenous Shock Pathway in Adult ICD Indicated Patients. Heart Rhythm . 2008;5(1):28–34. doi: 10.1016/j.hrthm.2007.08.030. [DOI] [PubMed] [Google Scholar]
  • 435.Bardy GH, Smith WM, Hood MA, Crozier IG, Melton IC, Jordaens L, et al. An Entirely Subcutaneous Implantable Cardioverter-Defibrillator. N Engl J Med . 2010;363(1):36–44. doi: 10.1056/NEJMoa0909545. [DOI] [PubMed] [Google Scholar]
  • 436.Radbill AE, Triedman JK, Berul CI, Fynn-Thompson F, Atallah J, Alexander ME, et al. System Survival of Nontransvenous Implantable Cardioverter-Defibrillators Compared to Transvenous Implantable Cardioverter-Defibrillators in Pediatric and Congenital Heart Disease Patients. Heart Rhythm . 2010;7(2):193–198. doi: 10.1016/j.hrthm.2009.10.014. [DOI] [PubMed] [Google Scholar]
  • 437.Ertuğrul İ, Karagöz T, Aykan H, Yıldırım I, Özer S, Karagöz H, et al. Subcutaneous Defibrillator Implantation in Pediatric Patients. Anatol J Cardiol . 2016;16(8):630–634. doi: 10.5152/AnatolJCardiol.2015.6589. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 438.Griksaitis MJ, Rosengarten JA, Gnanapragasam JP, Haw MP, Morgan JM. Implantable Cardioverter Defibrillator Therapy in Paediatric Practice: A Single-Centre UK Experience with Focus on Subcutaneous Defibrillation. Europace . 2013;15(4):523–530. doi: 10.1093/europace/eus388. [DOI] [PubMed] [Google Scholar]
  • 439.Pettit SJ, McLean A, Colquhoun I, Connelly D, McLeod K. Clinical Experience of Subcutaneous and Transvenous Implantable Cardioverter Defibrillators in Children and Teenagers. Pacing Clin Electrophysiol . 2013;36(12):1532–1538. doi: 10.1111/pace.12233. [DOI] [PubMed] [Google Scholar]
  • 440.Diller GP, Kempny A, Liodakis E, Alonso-Gonzalez R, Inuzuka R, Uebing A, et al. Left Ventricular Longitudinal Function Predicts Life-Threatening Ventricular Arrhythmia and Death in Adults with Repaired Tetralogy of Fallot. Circulation . 2012;125(20):2440–2446. doi: 10.1161/CIRCULATIONAHA.111.086983. [DOI] [PubMed] [Google Scholar]
  • 441.Gatzoulis MA, Till JA, Somerville J, Redington AN. Mechanoelectrical Interaction in Tetralogy of Fallot. QRS Prolongation Relates to Right Ventricular Size and Predicts Malignant Ventricular Arrhythmias and Sudden Death. Circulation . 1995;92(2):231–237. doi: 10.1161/01.cir.92.2.231. [DOI] [PubMed] [Google Scholar]
  • 442.Knauth AL, Gauvreau K, Powell AJ, Landzberg MJ, Walsh EP, Lock JE, et al. Ventricular Size and Function Assessed by Cardiac MRI Predict Major Adverse Clinical Outcomes Late After Tetralogy of Fallot Repair. Heart . 2008;94(2):211–216. doi: 10.1136/hrt.2006.104745. [DOI] [PubMed] [Google Scholar]
  • 443.Deal BJ, Scagliotti D, Miller SM, Gallastegui JL, Hariman RJ, Levitsky S. Electrophysiologic Drug Testing in Symptomatic Ventricular Arrhythmias After Repair of Tetralogy of Fallot. Am J Cardiol . 1987;59(15):1380–1385. doi: 10.1016/0002-9149(87)90924-6. [DOI] [PubMed] [Google Scholar]
  • 444.Garson A, Jr, Porter CB, Gillette PC, McNamara DG. Induction of Ventricular Tachycardia During Electrophysiologic Study After Repair of Tetralogy of Fallot. J Am Coll Cardiol . 1983;1(6):1493–1502. doi: 10.1016/s0735-1097(83)80054-0. [DOI] [PubMed] [Google Scholar]
  • 445.Khairy P, Aboulhosn J, Gurvitz MZ, Opotowsky AR, Mongeon FP, Kay J, et al. Arrhythmia Burden in Adults with Surgically Repaired Tetralogy of Fallot: A Multi-Institutional Study. Circulation . 2010;122(9):868–875. doi: 10.1161/CIRCULATIONAHA.109.928481. [DOI] [PubMed] [Google Scholar]
  • 446.Valente AM, Gauvreau K, Assenza GE, Babu-Narayan SV, Schreier J, Gatzoulis MA, et al. Contemporary Predictors of Death and Sustained Ventricular Tachycardia in Patients with Repaired Tetralogy of Fallot Enrolled in the INDICATOR Cohort. Heart . 2014;100(3):247–253. doi: 10.1136/heartjnl-2013-304958. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 447.Harrison DA, Harris L, Siu SC, MacLoghlin CJ, Connelly MS, Webb GD, et al. Sustained Ventricular Tachycardia in Adult Patients Late After Repair of Tetralogy of Fallot. J Am Coll Cardiol . 1997;30(5):1368–1373. doi: 10.1016/s0735-1097(97)00316-1. [DOI] [PubMed] [Google Scholar]
  • 448.Rotes AS, Connolly HM, Warnes CA, Ammash NM, Phillips SD, Dearani JA, et al. Ventricular Arrhythmia Risk Stratification in Patients with Tetralogy of Fallot at the Time of Pulmonary Valve Replacement. Circ Arrhythm Electrophysiol . 2015;8(1):110–116. doi: 10.1161/CIRCEP.114.001975. [DOI] [PubMed] [Google Scholar]
  • 449.Therrien J, Siu SC, Harris L, Dore A, Niwa K, Janousek J, et al. Impact of Pulmonary Valve Replacement on Arrhythmia Propensity Late After Repair of Tetralogy of Fallot. Circulation . 2001;103(20):2489–2494. doi: 10.1161/01.cir.103.20.2489. [DOI] [PubMed] [Google Scholar]
  • 450.Santharam S, Hudsmith L, Thorne S, Clift P, Marshall H, Bono J. Long-Term Follow-up of Implantable Cardioverter-Defibrillators in Adult Congenital Heart Disease Patients: Indications and Outcomes. Europace . 2017;19(3):407–413. doi: 10.1093/europace/euw076. [DOI] [PubMed] [Google Scholar]
  • 451.Khanna AD, Warnes CA, Phillips SD, Lin G, Brady PA. Single-Center Experience with Implantable Cardioverter-Defibrillators in Adults with Complex Congenital Heart Disease. Am J Cardiol . 2011;108(5):729–734. doi: 10.1016/j.amjcard.2011.04.020. [DOI] [PubMed] [Google Scholar]
  • 452.Kella DK, Merchant FM, Veledar E, Book W, Lloyd MS. Lesion-Specific Differences for Implantable Cardioverter Defibrillator Therapies in Adults with Congenital Heart Disease. Pacing Clin Electrophysiol . 2014;37(11):1492–1498. doi: 10.1111/pace.12434. [DOI] [PubMed] [Google Scholar]
  • 453.Yap SC, Roos-Hesselink JW, Hoendermis ES, Budts W, Vliegen HW, Mulder BJ, et al. Outcome of Implantable Cardioverter Defibrillators in Adults with Congenital Heart Disease: A Multi-Centre Study. Eur Heart J . 2007;28(15):1854–1861. doi: 10.1093/eurheartj/ehl306. [DOI] [PubMed] [Google Scholar]
  • 454.Koyak Z, Groot JR, Van Gelder IC, Bouma BJ, van Dessel PF, Budts W, et al. Implantable Cardioverter Defibrillator Therapy in Adults with Congenital Heart Disease: Who is at Risk of Shocks? Circ Arrhythm Electrophysiol . 2012;5(1):101–110. doi: 10.1161/CIRCEP.111.966754. [DOI] [PubMed] [Google Scholar]
  • 455.Berul CI, Van Hare GF, Kertesz NJ, Dubin AM, Cecchin F, Collins KK, et al. Results of a Multicenter Retrospective Implantable Cardioverter-Defibrillator Registry of Pediatric and Congenital Heart Disease Patients. J Am Coll Cardiol . 2008;51(17):1685–1691. doi: 10.1016/j.jacc.2008.01.033. [DOI] [PubMed] [Google Scholar]
  • 456.Moore JP, Mondésert B, Lloyd MS, Cook SC, Zaidi AN, Pass RH, et al. Clinical Experience with the Subcutaneous Implantable Cardioverter-Defibrillator in Adults with Congenital Heart Disease. Circ Arrhythm Electrophysiol . 2016;9(9):e004338. doi: 10.1161/CIRCEP.116.004338. [DOI] [PubMed] [Google Scholar]
  • 457.Khairy P, Dore A, Poirier N, Marcotte F, Ibrahim R, Mongeon FP, et al. Risk Stratification in Surgically Repaired Tetralogy of Fallot. Expert Rev Cardiovasc Ther . 2009;7(7):755–762. doi: 10.1586/erc.09.38. [DOI] [PubMed] [Google Scholar]
  • 458.Khairy P, Harris L, Landzberg MJ, Viswanathan S, Barlow A, Gatzoulis MA, et al. Implantable Cardioverter-Defibrillators in Tetralogy of Fallot. Circulation . 2008;117(3):363–370. doi: 10.1161/CIRCULATIONAHA.107.726372. [DOI] [PubMed] [Google Scholar]
  • 459.Kapel GF, Reichlin T, Wijnmaalen AP, Piers SR, Holman ER, Tedrow UB, et al. Re-Entry Using Anatomically Determined Isthmuses: A Curable Ventricular Tachycardia in Repaired Congenital Heart Disease. Circ Arrhythm Electrophysiol . 2015;8(1):102–109. doi: 10.1161/CIRCEP.114.001929. [DOI] [PubMed] [Google Scholar]
  • 460.Kapel GF, Reichlin T, Wijnmaalen AP, Tedrow UB, Piers SR, Schalij MJ, et al. Left-Sided Ablation of Ventricular Tachycardia in Adults with Repaired Tetralogy of Fallot: A Case Series. Circ Arrhythm Electrophysiol . 2014;7(5):889–897. doi: 10.1161/CIRCEP.114.001661. [DOI] [PubMed] [Google Scholar]
  • 461.Kapel GF, Sacher F, Dekkers OM, Watanabe M, Blom NA, Thambo JB, et al. Arrhythmogenic Anatomical Isthmuses Identified by Electroanatomical Mapping are the Substrate for Ventricular Tachycardia in Repaired Tetralogy of Fallot. Eur Heart J . 2017;38(4):268–276. doi: 10.1093/eurheartj/ehw202. [DOI] [PubMed] [Google Scholar]
  • 462.van Zyl M, Kapa S, Padmanabhan D, Chen FC, Mulpuru SK, Packer DL, et al. Mechanism and Outcomes of Catheter Ablation for Ventricular Tachycardia in Adults with Repaired Congenital Heart Disease. Heart Rhythm . 2016;13(7):1449–1454. doi: 10.1016/j.hrthm.2016.03.002. [DOI] [PubMed] [Google Scholar]
  • 463.Zeppenfeld K, Schalij MJ, Bartelings MM, Tedrow UB, Koplan BA, Soejima K, et al. Catheter Ablation of Ventricular Tachycardia After Repair of Congenital Heart Disease: Electroanatomic Identification of the Critical Right Ventricular Isthmus. Circulation . 2007;116(20):2241–2252. doi: 10.1161/CIRCULATIONAHA.107.723551. [DOI] [PubMed] [Google Scholar]
  • 464.Raissadati A, Nieminen H, Haukka J, Sairanen H, Jokinen E. Late Causes of Death After Pediatric Cardiac Surgery: A 60-Year Population-Based Study. J Am Coll Cardiol . 2016;68(5):487–498. doi: 10.1016/j.jacc.2016.05.038. [DOI] [PubMed] [Google Scholar]
  • 465.Diller GP, Kempny A, Alonso-Gonzalez R, Swan L, Uebing A, Li W, et al. Survival Prospects and Circumstances of Death in Contemporary Adult Congenital Heart Disease Patients Under Follow-Up at a Large Tertiary Centre. Circulation . 2015;132(22):2118–2125. doi: 10.1161/CIRCULATIONAHA.115.017202. [DOI] [PubMed] [Google Scholar]
  • 466.Nieminen HP, Jokinen EV, Sairanen HI. Causes of Late Deaths After Pediatric Cardiac Surgery: A Population-Based Study. J Am Coll Cardiol . 2007;50(13):1263–1271. doi: 10.1016/j.jacc.2007.05.040. [DOI] [PubMed] [Google Scholar]
  • 467.Zomer AC, Vaartjes I, Uiterwaal CS, van der Velde ET, van den Merkhof LF, Baur LH, et al. Circumstances of Death in Adult Congenital Heart Disease. Int J Cardiol . 2012;154(2):168–172. doi: 10.1016/j.ijcard.2010.09.015. [DOI] [PubMed] [Google Scholar]
  • 468.Lange R, Hörer J, Kostolny M, Cleuziou J, Vogt M, Busch R, et al. Presence of a Ventricular Septal Defect and the Mustard Operation are Risk Factors for Late Mortality After the Atrial Switch Operation: Thirty Years of Follow-up in 417 Patients at a Single Center. Circulation . 2006;114(18):1905–1913. doi: 10.1161/CIRCULATIONAHA.105.606046. [DOI] [PubMed] [Google Scholar]
  • 469.Lubiszewska B, Gosiewska E, Hoffman P, Teresińska A, Rózański J, Piotrowski W, et al. Myocardial Perfusion and Function of the Systemic Right Ventricle in Patients After Atrial Switch Procedure for Complete Transposition: Long-Term Follow-up. J Am Coll Cardiol . 2000;36(4):1365–1370. doi: 10.1016/s0735-1097(00)00864-0. [DOI] [PubMed] [Google Scholar]
  • 470.Millane T, Bernard EJ, Jaeggi E, Howman-Giles RB, Uren RF, Cartmill TB, et al. Role of Ischemia and Infarction in Late Right Ventricular Dysfunction After Atrial Repair of Transposition of the Great Arteries. J Am Coll Cardiol . 2000;35(6):1661–1668. doi: 10.1016/s0735-1097(00)00585-4. [DOI] [PubMed] [Google Scholar]
  • 471.Schwerzmann M, Salehian O, Harris L, Siu SC, Williams WG, Webb GD, et al. Ventricular Arrhythmias and Sudden Death in Adults After a Mustard Operation for Transposition of the Great Arteries. Eur Heart J . 2009;30(15):1873–1879. doi: 10.1093/eurheartj/ehp179. [DOI] [PubMed] [Google Scholar]
  • 472.Engelings CC, Helm PC, Abdul-Khaliq H, Asfour B, Bauer UM, Baumgartner H, et al. Cause of Death in Adults with Congenital Heart Disease - An Analysis of the German National Register for Congenital Heart Defects. Int J Cardiol . 2016;211:31–36. doi: 10.1016/j.ijcard.2016.02.133. [DOI] [PubMed] [Google Scholar]
  • 473.Buber J, Ackley TJ, Daniels CJ, Roble SL, Mah ML, Kamp AN, et al. Outcomes Following the Implantation of Cardioverter-Defibrillator for Primary Prevention in Transposition of the Great Arteries After Intra-Atrial Baffle Repair: A Single-Centre Experience. Europace . 2016;18(7):1016–1022. doi: 10.1093/europace/euv297. [DOI] [PubMed] [Google Scholar]
  • 474.Backhoff D, Kerst G, Peters A, Lüdemann M, Frische C, Horndasch M, et al. Internal Cardioverter Defibrillator Indications and Therapies after Atrial Baffle Procedure for d-Transposition of the Great Arteries: A Multicenter Analysis. Pacing Clin Electrophysiol . 2016;39(10):1070–1076. doi: 10.1111/pace.12933. Oct. [DOI] [PubMed] [Google Scholar]
  • 475.Oechslin EN, Harrison DA, Connelly MS, Webb GD, Siu SC. Mode of Death in Adults with Congenital Heart Disease. Am J Cardiol . 2000;86(10):1111–1116. doi: 10.1016/s0002-9149(00)01169-3. [DOI] [PubMed] [Google Scholar]
  • 476.Verheugt CL, Uiterwaal CS, van der Velde ET, Meijboom FJ, Pieper PG, van Dijk AP, et al. Mortality in Adult Congenital Heart Disease. Eur Heart J . 2010;31(10):1220–1229. doi: 10.1093/eurheartj/ehq032. [DOI] [PubMed] [Google Scholar]
  • 477.Silka MJ, Hardy BG, Menashe VD, Morris CD. A Population-Based Prospective Evaluation of Risk of Sudden Cardiac Death After Operation for Common Congenital Heart Defects. J Am Coll Cardiol . 1998;32(1):245–251. doi: 10.1016/s0735-1097(98)00187-9. [DOI] [PubMed] [Google Scholar]
  • 478.Fish FA, Gillette PC, Benson DW., Jr Proarrhythmia, Cardiac Arrest and Death in Young Patients Receiving Encainide and Flecainide. The Pediatric Electrophysiology Group. J Am Coll Cardiol . 1991;18(2):356–365. doi: 10.1016/0735-1097(91)90586-x. [DOI] [PubMed] [Google Scholar]
  • 479.Hassan OKA, Fahed AC, Batrawi M, Arabi M, Refaat MM, DePalma SR, et al. NKX2-5 Mutations in an Inbred Consanguineous Population: Genetic and Phenotypic Diversity. Sci Rep . 2015;5:8848. doi: 10.1038/srep08848. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 480.El Malti R, Liu H, Doray B, Thauvin C, Maltret A, Dauphin C, et al. A Systematic Variant Screening in Familial Cases of Congenital Heart Defects Demonstrates the Usefulness of Molecular Genetics in this Field. Eur J Hum Genet . 2016;24(2):228–236. doi: 10.1038/ejhg.2015.105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 481.Ellesøe SG, Johansen MM, Bjerre JV, Hjortdal VE, Brunak S, Larsen LA. Familial Atrial Septal Defect and Sudden Cardiac Death: Identification of a Novel NKX2-5 Mutation and a Review of the Literature. Congenit Heart Dis . 2016;11(3):283–290. doi: 10.1111/chd.12317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 482.Vehmeijer JT, Brouwer TF, Limpens J, Knops RE, Bouma BJ, Mulder BJ, et al. Implantable Cardioverter-Defibrillators in Adults with Congenital Heart Disease: A Systematic Review and Meta-Analysis. Eur Heart J . 2016;37(18):1439–1448. doi: 10.1093/eurheartj/ehv735. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 483.Stan MN, Sathananthan M, Warnes CA, Brennan MD, Thapa P, Bahn RS. Amiodarone-Induced Thyrotoxicosis in Adults with Congenital Heart Disease--Clinical Presentation and Response to Therapy. Endocr Pract . 2014;20(1):33–40. doi: 10.4158/EP13059.OR. [DOI] [PubMed] [Google Scholar]
  • 484.Thorne SA, Barnes I, Cullinan P, Somerville J. Amiodarone-Associated Thyroid Dysfunction: Risk Factors in Adults with Congenital Heart Disease. Circulation . 1999;100(2):149–154. doi: 10.1161/01.cir.100.2.149. [DOI] [PubMed] [Google Scholar]
  • 485.Ellenbogen KA, Wilkoff BL, Kay GN, Lau CP, Auricchio A. Clinical Cardiac Pacing Defibrillation, and Resynchronization Therapy . 5th ed. Philadelphia: Elsevier; 2017. [Google Scholar]
  • 486.Gillis AM, Russo AM, Ellenbogen KA, Swerdlow CD, Olshansky B, Al-Khatib SM, et al. HRS/ACCF Expert Consensus Statement on Pacemaker Device and Mode Selection. J Am Coll Cardiol . 2012;60(7):682–703. doi: 10.1016/j.jacc.2012.06.011. [DOI] [PubMed] [Google Scholar]
  • 487.Kindermann M, Hennen B, Jung J, Geisel J, Böhm M, Fröhlig G. Biventricular Versus Conventional Right Ventricular Stimulation for Patients with Standard Pacing Indication and Left Ventricular Dysfunction: The Homburg Biventricular Pacing Evaluation (HOBIPACE) J Am Coll Cardiol . 2006;47(10):1927–1937. doi: 10.1016/j.jacc.2005.12.056. [DOI] [PubMed] [Google Scholar]
  • 488.Yu CM, Chan JY, Zhang Q, Omar R, Yip GW, Hussin A, et al. Biventricular Pacing in Patients with Bradycardia and Normal Ejection Fraction. N Engl J Med . 2009;361(22):2123–2134. doi: 10.1056/NEJMoa0907555. [DOI] [PubMed] [Google Scholar]
  • 489.Boriani G, Cimaglia P, Biffi M, Martignani C, Ziacchi M, Valzania C, et al. Cost-Effectiveness of Implantable Cardioverter-Defibrillator in Today’s World. Indian Heart J . 2014;66(Suppl 1):101–104. doi: 10.1016/j.ihj.2013.12.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 490.Mark DB, Simons TA. Economics and Cost-Effectiveness in Evaluating the Value of Cardiovascular Therapies. Fundamentals of Economic Analysis. Am Heart J . 1999;137(5):38–40. doi: 10.1016/s0002-8703(99)70426-6. [DOI] [PubMed] [Google Scholar]
  • 491.Pellegrini CN, Lee K, Olgin JE, Turakhia MP, Tseng ZH, Lee R, et al. Impact of Advanced Age on Survival in Patients with Implantable Cardioverter Defibrillators. Europace . 2008;10(11):1296–1301. doi: 10.1093/europace/eun253. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 492.Sanders GD, Hlatky MA, Owens DK. Cost-Effectiveness of Implantable Cardioverter-Defibrillators. N Engl J Med . 2005;353(14):1471–1480. doi: 10.1056/NEJMsa051989. [DOI] [PubMed] [Google Scholar]
  • 493.Buxton AE, Lee KL, DiCarlo L, Gold MR, Greer GS, Prystowsky EN, et al. Electrophysiologic Testing to Identify Patients with Coronary Artery Disease Who are at Risk for Sudden Death. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med . 2000;342(26):1937–1945. doi: 10.1056/NEJM200006293422602. [DOI] [PubMed] [Google Scholar]
  • 494.Kadish A, Dyer A, Daubert JP, Quigg R, Estes NA, Anderson KP, et al. Prophylactic Defibrillator Implantation in Patients with Nonischemic Dilated Cardiomyopathy. N Engl J Med . 2004;350(21):2151–2158. doi: 10.1056/NEJMoa033088. [DOI] [PubMed] [Google Scholar]
  • 495.Ribeiro RA, Stella SF, Zimerman LI, Pimentel M, Rohde LEP, Polanczyk CA. Custo-Efetividade de Cardiodesfibriladores Implantáveis no Brasil nos Setores Público e Privado. Arq. Bras. Cardiol . 2010;95(5):577–586. doi: 10.1590/S0066-782X2010005000134. [DOI] [PubMed] [Google Scholar]
  • 496.Gialama F, Prezerakos P, Maniadakis N. The Cost Effectiveness of Implantable Cardioverter Defibrillators: A Systematic Review of Economic Evaluations. Appl Health Econ Health Policy . 2014;12(1):41–49. doi: 10.1007/s40258-013-0069-2. [DOI] [PubMed] [Google Scholar]
  • 497.Matos AJ. Análise da Relação Custo-Efetividade do Tratamento com DCI - Desfibrilador Cardioversor Implantável . São Paulo (SP): Universidade de São Paulo; 2007. dissertation. [Google Scholar]
  • 498.Buxton M, Caine N, Chase D, Connelly D, Grace A, Jackson C, et al. A Review of the Evidence on the Effects and Costs of Implantable Cardioverter Defibrillator Therapy in Different Patient Groups, and Modelling of Cost-Effectiveness and Cost-utility for These Groups in a UK Context. Health Technol Assess . 2006;10(27):1–164. doi: 10.3310/hta10270. Aug. [DOI] [PubMed] [Google Scholar]
  • 499.Cowie MR, Marshall D, Drummond M, Ferko N, Maschio M, Ekman M, et al. Lifetime Cost-Effectiveness of Prophylactic Implantation of a Cardioverter Defibrillator in Patients with Reduced Left Ventricular Systolic Function: Results of Markov Modelling in a European Population. Europace . 2009;11(6):716–726. doi: 10.1093/europace/eup068. [DOI] [PubMed] [Google Scholar]
  • 500.Thijssen J, van den Akker van Marle ME, Borleffs CJ, van Rees JB, Bie MK, et al. Cost-Effectiveness of Primary Prevention Implantable Cardioverter Defibrillator Treatment: Data from a Large Clinical Registry. Pacing Clin Electrophysiol . 2014;37(1):25–34. doi: 10.1111/pace.12238. [DOI] [PubMed] [Google Scholar]
  • 501.Larsen G, Hallstrom A, McAnulty J, Pinski S, Olarte A, Sullivan S, et al. Cost-Effectiveness of the Implantable Cardioverter-Defibrillator Versus Antiarrhythmic Drugs in Survivors of Serious Ventricular Tachyarrhythmias: Results of the Antiarrhythmics Versus Implantable Defibrillators (AVID) Economic Analysis Substudy. Circulation . 2002;105(17):2049–2057. doi: 10.1161/01.cir.0000015504.57641.d0. [DOI] [PubMed] [Google Scholar]
  • 502.Palmisano P, Ziacchi M, Biffi M, Ricci RP, Landolina M, Zoni-Berisso M, et al. Clinically Oriented Device Programming in Bradycardia Patients: Part 2 (Atrioventricular Blocks and Neurally Mediated Syncope). Proposals from AIAC (Italian Association of Arrhythmology and Cardiac Pacing) J Cardiovasc Med (Hagerstown) . 2018;19(4):170–180. doi: 10.2459/JCM.0000000000000629. [DOI] [PubMed] [Google Scholar]
  • 503.Scott PA, Silberbauer J, McDonagh TA, Murgatroyd FD. Impact of Prolonged Implantable Cardioverter-Defibrillator Arrhythmia Detection Times on Outcomes: A Meta-Analysis. Heart Rhythm . 2014;11(5):828–835. doi: 10.1016/j.hrthm.2014.02.009. [DOI] [PubMed] [Google Scholar]
  • 504.Tan VH, Wilton SB, Kuriachan V, Sumner GL, Exner DV. Impact of Programming Strategies Aimed at Reducing Nonessential Implantable Cardioverter Defibrillator Therapies on Mortality: A Systematic Review and Meta-Analysis. Circ Arrhythm Electrophysiol . 2014;7(1):164–170. doi: 10.1161/CIRCEP.113.001217. [DOI] [PubMed] [Google Scholar]
  • 505.Mealing S, Woods B, Hawkins N, Cowie MR, Plummer CJ, Abraham WT, et al. Cost-Effectiveness of Implantable Cardiac Devices in Patients with Systolic Heart Failure. Heart . 2016;102(21):1742–1749. doi: 10.1136/heartjnl-2015-308883. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 506.Barra S, Providência R, Paiva L, Heck P, Agarwal S. Implantable Cardioverter-Defibrillators in the Elderly: Rationale and Specific Age-Related Considerations. Europace . 2015;17(2):174–186. doi: 10.1093/europace/euu296. [DOI] [PubMed] [Google Scholar]
  • 507.Padeletti L, Pieragnoli P, Di Biase L, Colella A, Landolina M, Moro E, et al. Is a Dual-Sensor Pacemaker Appropriate in Patients with Sino-Atrial Disease? Results from the DUSISLOG Study. Pacing Clin Electrophysiol . 2006;29(1):34–40. doi: 10.1111/j.1540-8159.2006.00301.x. [DOI] [PubMed] [Google Scholar]
  • 508.Sulke N, Sugihara C, Hong P, Patel N, Freemantle N. The Benefit of a Remotely Monitored Implantable Loop Recorder as a First Line Investigation in Unexplained Syncope: The EaSyAS II Trial. Europace . 2016;18(6):912–918. doi: 10.1093/europace/euv228. [DOI] [PubMed] [Google Scholar]
  • 509.Moya A, García-Civera R, Croci F, Menozzi C, Brugada J, Ammirati F, et al. Diagnosis, Management, and Outcomes of Patients with Syncope and Bundle Branch Block. Eur Heart J . 2011;32(12):1535–1541. doi: 10.1093/eurheartj/ehr071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 510.Sposato LA, Cipriano LE, Saposnik G, Vargas ER, Riccio PM, Hachinski V. Diagnosis of Atrial Fibrillation After Stroke and Transient Ischaemic Attack: A Systematic Review and Meta-Analysis. Lancet Neurol . 2015;14(4):377–387. doi: 10.1016/S1474-4422(15)70027-X. [DOI] [PubMed] [Google Scholar]
  • 511.Sanna T, Diener HC, Passman RS, Di Lazzaro V, Bernstein RA, Morillo CA, et al. Cryptogenic Stroke and Underlying Atrial Fibrillation. N Engl J Med . 2014;370(26):2478–2486. doi: 10.1056/NEJMoa1313600. [DOI] [PubMed] [Google Scholar]
  • 512.Ziacchi M, Palmisano P, Biffi M, Ricci RP, Landolina M, Zoni-Berisso M, et al. Clinically Oriented Device Programming in Bradycardia Patients: Part 1 (Sinus Node Disease). Proposals from AIAC (Italian Association of Arrhythmology and Cardiac Pacing) J Cardiovasc Med (Hagerstown) . 2018;19(4):161–169. doi: 10.2459/JCM.0000000000000630. [DOI] [PubMed] [Google Scholar]
  • 513.Leong DP, Mitchell AM, Salna I, Brooks AG, Sharma G, Lim HS, et al. Long-Term Mechanical Consequences of Permanent Right Ventricular Pacing: Effect of Pacing Site. J Cardiovasc Electrophysiol . 2010;21(10):1120–1126. doi: 10.1111/j.1540-8167.2010.01804.x. [DOI] [PubMed] [Google Scholar]
  • 514.Flevari P, Leftheriotis D, Fountoulaki K, Panou F, Rigopoulos AG, Paraskevaidis I, et al. Long-Term Nonoutflow Septal Versus Apical Right Ventricular Pacing: Relation to Left Ventricular Dyssynchrony. Pacing Clin Electrophysiol . 2009;32(3):354–362. doi: 10.1111/j.1540-8159.2008.02244.x. [DOI] [PubMed] [Google Scholar]
  • 515.Andersen HR, Thuesen L, Bagger JP, Vesterlund T, Thomsen PE. Prospective Randomised Trial of Atrial Versus Ventricular Pacing in Sick-Sinus Syndrome. Lancet . 1994;344(8936):1523–1528. doi: 10.1016/s0140-6736(94)90347-6. [DOI] [PubMed] [Google Scholar]
  • 516.Sulke N, Chambers J, Dritsas A, Sowton E. A Randomized Double-Blind Crossover Comparison of Four Rate-Responsive Pacing Modes. J Am Coll Cardiol . 1991;17(3):696–706. doi: 10.1016/s0735-1097(10)80186-x. [DOI] [PubMed] [Google Scholar]
  • 517.Cheng A, Landman SR, Stadler RW. Reasons for Loss of Cardiac Resynchronization Therapy Pacing: Insights from 32 844 Patients. Circ Arrhythm Electrophysiol . 2012;5(5):884–888. doi: 10.1161/CIRCEP.112.973776. [DOI] [PubMed] [Google Scholar]
  • 518.Capucci A, Boriani G, Specchia S, Marinelli M, Santarelli A, Magnani B. Evaluation by Cardiopulmonary Exercise Test of DDDR Versus DDD Pacing. Pacing Clin Electrophysiol . 1992;15(11):1908–1913. doi: 10.1111/j.1540-8159.1992.tb02992.x. [DOI] [PubMed] [Google Scholar]
  • 519.Palmisano P, Dell’Era G, Russo V, Zaccaria M, Mangia R, Bortnik M, et al. Effects of Closed-Loop Stimulation vs. DDD Pacing on Haemodynamic Variations and Occurrence of Syncope Induced by Head-Up Tilt Test in Older Patients with Refractory Cardioinhibitory Vasovagal Syncope: The Tilt Test-Induced REsponse in Closed-loop Stimulation Multicentre, Prospective, Single Blind, Randomized Study. Europace . 2018;20(5):859–866. doi: 10.1093/europace/eux015. [DOI] [PubMed] [Google Scholar]
  • 520.Marrouche NF, Brachmann J, Andresen D, Siebels J, Boersma L, Jordaens L, et al. Catheter Ablation for Atrial Fibrillation with Heart Failure. N Engl J Med . 2018;378(5):417–427. doi: 10.1056/NEJMoa1707855. [DOI] [PubMed] [Google Scholar]
  • 521.Lau CP, Rushby J, Leigh-Jones M, Tam CY, Poloniecki J, Ingram A, et al. Symptomatology and Quality of Life in Patients with Rate-Responsive Pacemakers: A Double-Blind, Randomized, Crossover Study. Clin Cardiol . 1989;12(9):505–512. doi: 10.1002/clc.4960120907. [DOI] [PubMed] [Google Scholar]
  • 522.Oto MA, Müderrisoglu H, Ozin MB, Korkmaz ME, Karamehmetoglu A, Oram A, et al. Quality of Life in Patients with Rate Responsive Pacemakers: A Randomized, Cross-Over Study. Pacing Clin Electrophysiol . 1991;14(5):800–806. doi: 10.1111/j.1540-8159.1991.tb04110.x. [DOI] [PubMed] [Google Scholar]
  • 523.Benditt DG, Sutton R, Gammage MD, Markowitz T, Gorski J, Nygaard GA, et al. Clinical Experience with Thera DR Rate-Drop Response Pacing Algorithm in Carotid Sinus Syndrome and Vasovagal Syncope. The International Rate-Drop Investigators Group. Pacing Clin Electrophysiol . 1997;20(3):832–839. doi: 10.1111/j.1540-8159.1997.tb03916.x. [DOI] [PubMed] [Google Scholar]
  • 524.Occhetta E, Bortnik M, Vassanelli C. The DDDR Closed Loop Stimulation for the Prevention of Vasovagal Syncope: Results from the INVASY Prospective Feasibility Registry. Europace . 2003;5(2):153–162. doi: 10.1053/eupc.2002.0292. [DOI] [PubMed] [Google Scholar]
  • 525.Biffi M, Bertini M, Saporito D, Belotti G, Quartieri F, Piancastelli M, et al. Automatic Management of Atrial and Ventricular Stimulation in a Contemporary Unselected Population of Pacemaker Recipients: The ESSENTIAL Registry. Europace . 2016;18(10):1551–1560. doi: 10.1093/europace/euw021. [DOI] [PubMed] [Google Scholar]
  • 526.Refaat M, Mansour M, Singh JP, Ruskin J, Heist EK. Electrocardiographic Characteristics in Right Ventricular vs Biventricular Pacing in Patients with Paced Right Bundle-Branch Block QRS Pattern. J Electrocardiol . 2011;44(2):289–295. doi: 10.1016/j.jelectrocard.2010.08.003. [DOI] [PubMed] [Google Scholar]
  • 527.Fazelifar A, Jorfi F, Haghjoo M. Electrocardiographic Patterns in Biventricular Pacing Delivered by Second-Generation Cardiac Resynchronization Devices. Indian Pacing Electrophysiol J . 2018;18(1):13–19. doi: 10.1016/j.ipej.2017.10.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 528.Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JGF, Coats AJS, et al. 2016 ESC Guidelines for the Diagnosis and Treatment of Acute and Chronic Heart Failure: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure of the European Society of Cardiology (ESC)Developed with the Special Contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J . 2016;37(27):2129–2200. doi: 10.1093/eurheartj/ehw128. [DOI] [PubMed] [Google Scholar]
  • 529.Abraham WT, Gras D, Yu CM, Guzzo L, Gupta MS. Rationale and Design of a Randomized Clinical Trial to Assess the Safety and Efficacy of Frequent Optimization of Cardiac Resynchronization Therapy: The Frequent Optimization Study Using the QuickOpt Method (FREEDOM) Trial. Am Heart J . 2010;159(6):944–948. doi: 10.1016/j.ahj.2010.02.034. [DOI] [PubMed] [Google Scholar]
  • 530.Ruwald AC, Aktas MK, Ruwald MH, Kutyifa V, McNitt S, Jons C, et al. Postimplantation Ventricular Ectopic Burden and Clinical Outcomes in Cardiac Resynchronization Therapy-Defibrillator Patients: A MADIT-CRT Substudy. Ann Noninvasive Electrocardiol . 2018;23(2):e12491. doi: 10.1111/anec.12491. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 531.Trucco E, Tolosana JM, Arbelo E, Doltra A, Castel MÁ, Benito E, et al. Improvement of Reverse Remodeling Using Electrocardiogram Fusion-Optimized Intervals in Cardiac Resynchronization Therapy: A Randomized Study. JACC Clin Electrophysiol . 2018;4(2):181–189. doi: 10.1016/j.jacep.2017.11.020. [DOI] [PubMed] [Google Scholar]
  • 532.Bänsch D, Bonnemeier H, Brandt J, Bode F, Svendsen JH, Táborský M, et al. Intra-Operative Defibrillation Testing and Clinical Shock Efficacy in Patients with Implantable Cardioverter-Defibrillators: The NORDIC ICD Randomized Clinical Trial. Eur Heart J . 2015;36(37):2500–2507. doi: 10.1093/eurheartj/ehv292. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 533.Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM. A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators: Results from the Prospective Randomized Multicenter EMPIRIC Trial. J Am Coll Cardiol . 2006;48(2):330–339. doi: 10.1016/j.jacc.2006.03.037. [DOI] [PubMed] [Google Scholar]
  • 534.Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, et al. Prospective Randomized Multicenter Trial of Empirical Antitachycardia Pacing Versus Shocks for Spontaneous Rapid Ventricular Tachycardia in Patients with Implantable Cardioverter-Defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) Trial Results. Circulation . 2004;110(17):2591–2596. doi: 10.1161/01.CIR.0000145610.64014.E4. [DOI] [PubMed] [Google Scholar]
  • 535.Stiles MK, Fauchier L, Morillo CA, Wilkoff BL. 2019 HRS/EHRA/APHRS/LAHRS Focused Update to 2015 Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing. Europace . 2019;21(9):1442–1443. doi: 10.1093/europace/euz065. [DOI] [PubMed] [Google Scholar]
  • 536.Jiménez-Candil J, Durán O, Núñez J, Bravo L, Hernández J, Martín-García A, et al. Effectiveness of First Versus Successive Antitachycardia Pacing Attempts: Predictors and Clinical Consequences. J Interv Card Electrophysiol . 2019;56(3):349–357. doi: 10.1007/s10840-019-00624-w. [DOI] [PubMed] [Google Scholar]
  • 537.Ahsan SY, Saberwal B, Lambiase PD, Koo CY, Lee S, Gopalamurugan AB, et al. A Simple Infection-Control Protocol to Reduce Serious Cardiac Device Infections. Europace . 2014;16(10):1482–1489. doi: 10.1093/europace/euu126. [DOI] [PubMed] [Google Scholar]
  • 538.Lee R, Mittal S. Utility and Limitations of Long-Term Monitoring of Atrial Fibrillation Using an Implantable Loop Recorder. Heart Rhythm . 2018;15(2):287–295. doi: 10.1016/j.hrthm.2017.09.009. [DOI] [PubMed] [Google Scholar]
  • 539.Afzal MR, Mease J, Koppert T, Okabe T, Tyler J, Houmsse M, et al. Incidence of False-Positive Transmissions During Remote Rhythm Monitoring with Implantable Loop Recorders. Heart Rhythm . 2020;17(1):75–80. doi: 10.1016/j.hrthm.2019.07.015. [DOI] [PubMed] [Google Scholar]
  • 540.Traykov V, Bongiorni MG, Boriani G, Burri H, Costa R, Dagres N, et al. Clinical Practice and Implementation of Guidelines for the Prevention, Diagnosis and Management of Cardiac Implantable Electronic Device Infections: Results of a Worldwide Survey Under the Auspices of the European Heart Rhythm Association. Europace . 2019;21(8):1270–1279. doi: 10.1093/europace/euz137. [DOI] [PubMed] [Google Scholar]
  • 541.Bongiorni MG, Burri H, Deharo JC, Starck C, Kennergren C, Saghy L, et al. 2018 EHRA Expert Consensus Statement on Lead Extraction: Recommendations on Definitions, Endpoints, Research Trial Design, and Data Collection Requirements for Clinical Scientific Studies and Registries: Endorsed by APHRS/HRS/LAHRS. Europace . 2018;20(7):1217. doi: 10.1093/europace/euy050. [DOI] [PubMed] [Google Scholar]
  • 542.Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG, et al. European Heart Rhythm Association (EHRA) International Consensus Document on How to Prevent, Diagnose, and Treat Cardiac Implantable Electronic Device Infections-Endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in Collaboration with the European Association for Cardio-Thoracic Surgery (EACTS) Eur J Cardiothorac Surg . 2020;57(1):1–31. doi: 10.1093/ejcts/ezz296. [DOI] [PubMed] [Google Scholar]
  • 543.Erba PA, Habib G, Glaudemans AWJM, Miro JM, Slart RHJA. The Round Table Approach in Infective Endocarditis & Cardiovascular Implantable Electronic Devices Infections: Make Your E-Team Come True. Eur J Nucl Med Mol Imaging . 2017;44(7):1107–1108. doi: 10.1007/s00259-017-3679-3. [DOI] [PubMed] [Google Scholar]
  • 544.Sollini M, Berchiolli R, Bolton RCD, Rossi A, Kirienko M, Boni R, et al. The “3M” Approach to Cardiovascular Infections: Multimodality, Multitracers, and Multidisciplinary. Semin Nucl Med . 2018;48(3):199–224. doi: 10.1053/j.semnuclmed.2017.12.003. [DOI] [PubMed] [Google Scholar]
  • 545.Habib G, Lancellotti P, Antunes MJ, Bongiorni MG, Casalta JP, Del Zotti F, et al. 2015 ESC Guidelines for the Management of Infective Endocarditis: The Task Force for the Management of Infective Endocarditis of the European Society of Cardiology (ESC) Eur Heart J . 2015;36(44):3075–3128. doi: 10.1093/eurheartj/ehv319. [DOI] [PubMed] [Google Scholar]
  • 546.Mahmood M, Kendi AT, Farid S, Ajmal S, Johnson GB, Baddour LM, et al. Role of 18F-FDG PET/CT in the Diagnosis of Cardiovascular Implantable Electronic Device Infections: A meta-analysis. J Nucl Cardiol . 2019;26(3):958–970. doi: 10.1007/s12350-017-1063-0. [DOI] [PubMed] [Google Scholar]
  • 547.Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N, et al. Risk Factors Related to Infections of Implanted Pacemakers and Cardioverter-Defibrillators: Results of a Large Prospective Study. Circulation . 2007;116(12):1349–1355. doi: 10.1161/CIRCULATIONAHA.106.678664. [DOI] [PubMed] [Google Scholar]
  • 548.Mazzone P, Migliore F, Bertaglia E, Facchin D, Daleffe E, Calzolari V, et al. Safety and Efficacy of the New Bidirectional Rotational Evolution® Mechanical Lead Extraction Sheath: Results from a Multicentre Italian Registry. Europace . 2018;20(5):829–834. doi: 10.1093/europace/eux020. [DOI] [PubMed] [Google Scholar]
  • 549.Polyzos KA, Konstantelias AA, Falagas ME. Risk Factors for Cardiac Implantable Electronic Device Infection: A Systematic Review and Meta-Analysis. Europace . 2015;17(5):767–777. doi: 10.1093/europace/euv053. [DOI] [PubMed] [Google Scholar]
  • 550.Tarakji KG, Mittal S, Kennergren C, Corey R, Poole JE, Schloss E, et al. Antibacterial Envelope to Prevent Cardiac Implantable Device Infection. N Engl J Med . 2019;380(20):1895–1905. doi: 10.1056/NEJMoa1901111. [DOI] [PubMed] [Google Scholar]
  • 551.D’Souza BA, Epstein AE, Garcia FC, Kim YY, Agarwal SC, Belott PH, et al. Outcomes in Patients with Congenital Heart Disease Receiving the Subcutaneous Implantable-Cardioverter Defibrillator: Results from a Pooled Analysis from the IDE Study and the EFFORTLESS S-ICD Registry. JACC Clin Electrophysiol . 2016;2(5):615–622. doi: 10.1016/j.jacep.2016.02.008. [DOI] [PubMed] [Google Scholar]
  • 552.Schneider AE, Burkhart HM, Ackerman MJ, Dearani JA, Wackel P, Cannon BC. Minimally Invasive Epicardial Implantable Cardioverter-Defibrillator Placement for Infants and Children: An Effective Alternative to the Transvenous Approach. Heart Rhythm . 2016;13(9):1905–1912. doi: 10.1016/j.hrthm.2016.06.024. [DOI] [PubMed] [Google Scholar]
  • 553.Clémenty N, Carion PL, Léotoing L, Lamarsalle L, Wilquin-Bequet F, Brown B, et al. Infections and Associated Costs Following Cardiovascular Implantable Electronic Device Implantations: A Nationwide Cohort Study. Europace . 2018;20(12):1974–1980. doi: 10.1093/europace/eux387. [DOI] [PubMed] [Google Scholar]
  • 554.Ludwig S, Theis C, Brown B, Witthohn A, Lux W, Goette A. Incidence and Costs of Cardiac Device Infections: Retrospective Analysis Using German Health Claims Data. J Comp Eff Res . 2018;7(5):483–492. doi: 10.2217/cer-2017-0080. [DOI] [PubMed] [Google Scholar]
  • 555.Uslan DZ, Gleva MJ, Warren DK, Mela T, Chung MK, Gottipaty V, et al. Cardiovascular Implantable Electronic Device Replacement Infections and Prevention: Results from the REPLACE Registry. Pacing Clin Electrophysiol . 2012;35(1):81–87. doi: 10.1111/j.1540-8159.2011.03257.x. [DOI] [PubMed] [Google Scholar]
  • 556.Biffi M, Ammendola E, Menardi E, Parisi Q, Narducci ML, De Filippo P, et al. Real-Life Outcome of Implantable Cardioverter-Defibrillator and Cardiac Resynchronization Defibrillator Replacement/Upgrade in a Contemporary Population: Observations from the Multicentre DECODE Registry. Europace . 2019;21(10):1527–1536. doi: 10.1093/europace/euz166. [DOI] [PubMed] [Google Scholar]
  • 557.Krahn AD, Longtin Y, Philippon F, Birnie DH, Manlucu J, Angaran P, et al. Prevention of Arrhythmia Device Infection Trial: The PADIT Trial. J Am Coll Cardiol . 2018;72(24):3098–3109. doi: 10.1016/j.jacc.2018.09.068. [DOI] [PubMed] [Google Scholar]
  • 558.U.S. Food and Drug Administration . Medical Devices: Recalls, Corrections and Removals (Devices) Rockville: U.S. Food and Drug Administration; c2021. [cited 2010 November 16]. Internet. http://www.fda.gov/medicaldevices/deviceregulationandguidance/postmarketrequirements/recallscorrectionsandremovals/default.htm.cali . [Google Scholar]
  • 559.Bongiorni MG, Soldati E, Zucchelli G, Di Cori A, Segreti L, De Lucia R, et al. Transvenous Removal of Pacing and Implantable Cardiac Defibrillating Leads Using Single Sheath Mechanical Dilatation and Multiple Venous Approaches: High Success Rate and Safety in More than 2000 Leads. Eur Heart J . 2008;29(23):2886–2893. doi: 10.1093/eurheartj/ehn461. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 560.Kennergren C, Bucknall CA, Butter C, Charles R, Fuhrer J, Grosfeld M, et al. Laser-Assisted Lead Extraction: The European Experience. Europace . 2007;9(8):651–656. doi: 10.1093/europace/eum098. [DOI] [PubMed] [Google Scholar]
  • 561.Bracke FA, Dekker L, van Gelder BM. The Needle’s Eye Snare as a Primary Tool for Pacing Lead Extraction. Europace . 2013;15(7):1007–1012. doi: 10.1093/europace/eus426. [DOI] [PubMed] [Google Scholar]
  • 562.Tsang DC, Azarrafiy R, Pecha S, Reichenspurner H, Carrillo RG, Hakmi S. Long-Term Outcomes of Prophylactic Placement of an Endovascular Balloon in the Vena Cava for High-Risk Transvenous Lead Extractions. Heart Rhythm . 2017;14(12):1833–1838. doi: 10.1016/j.hrthm.2017.08.003. [DOI] [PubMed] [Google Scholar]
  • 563.Starck CT, Stepuk A, Holubec T, Steffel J, Stark JW, Falk V. Compression Coil Provides Increased Lead Control in Extraction Procedures. Europace . 2015;17(3):499–503. doi: 10.1093/europace/euu272. [DOI] [PubMed] [Google Scholar]
  • 564.Zeitler EP, Wang Y, Dharmarajan K, Anstrom KJ, Peterson ED, Daubert JP, et al. Outcomes 1 Year After Implantable Cardioverter-Defibrillator Lead Abandonment Versus Explantation for Unused or Malfunctioning Leads: A Report from the National Cardiovascular Data Registry. Circ Arrhythm Electrophysiol . 2016;9(7):e003953. doi: 10.1161/CIRCEP.116.003953. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 565.Burri H, Combescure C. Management of Recalled Implantable Cardioverter-Defibrillator Leads at Generator Replacement: A Decision Analysis Model for Fidelis Leads. Europace . 2014;16(8):1210–1217. doi: 10.1093/europace/eut425. [DOI] [PubMed] [Google Scholar]
  • 566.U.S. Food and Drug Administration . What is a Medical Device Recall? Rockville: U.S. Food and Drug Administration; c2021. [cited 2014 June 02]. Internet. http://www.fda.gov/MedicalDevices/Safety/ListofRecalls/ucm329946.htm . [Google Scholar]
  • 567.Kramer DB, Tan YT, Sato C, Kesselheim AS. Postmarket Surveillance of Medical Devices: A Comparison of Strategies in the US, EU, Japan, and China. PLoS Med . 2013;10(9):e1001519. doi: 10.1371/journal.pmed.1001519. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 568.European Commission DG Health and Consumers (SANCO) Guidelines on a Medical Device Vigilance System . Bruxelas: European Union; c2021. [cited 2018 March 14]. http://ec.europa.eu/DocsRoom/documents/15506 . [Google Scholar]
  • 569.Suga C, Hayes DL, Hyberger LK, Lloyd MA. Is There an Adverse Outcome from Abandoned Pacing Leads? J Interv Card Electrophysiol . 2000;4(3):493–499. doi: 10.1023/a:1009860514724. [DOI] [PubMed] [Google Scholar]
  • 570.Zhang S, Kriza C, Schaller S, Kolominsky-Rabas PL. Recalls of Cardiac Implants in the Last Decade: What Lessons can we Learn? PLoS One . 2015;10(5):e0125987. doi: 10.1371/journal.pone.0125987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 571.Amelot M, Foucault A, Scanu P, Gomes S, Champ-Rigot L, Pellissier A, et al. Comparison of Outcomes in Patients with Abandoned Versus Extracted Implantable Cardioverter Defibrillator Leads. Arch Cardiovasc Dis . 2011;104(11):572–577. doi: 10.1016/j.acvd.2011.08.004. [DOI] [PubMed] [Google Scholar]
  • 572.Rijal S, Shah RU, Saba S. Extracting Versus Abandoning Sterile Pacemaker and Defibrillator Leads. Am J Cardiol . 2015;115(8):1107–1110. doi: 10.1016/j.amjcard.2015.01.537. [DOI] [PubMed] [Google Scholar]
  • 573.Khairy P, Landzberg MJ, Gatzoulis MA, Mercier LA, Fernandes SM, Côté JM, et al. Transvenous Pacing Leads and Systemic Thromboemboli in Patients with Intracardiac Shunts: A Multicenter Study. Circulation . 2006;113(20):2391–2397. doi: 10.1161/CIRCULATIONAHA.106.622076. [DOI] [PubMed] [Google Scholar]
  • 574.Larsen JM, Theuns DA, Thøgersen AM. Paradoxical Thromboembolic Stroke During Extraction of a Recalled St Jude Medical Riata Defibrillator Lead with Conductor Externalization. Europace . 2014;16(2):240. doi: 10.1093/europace/eut164. [DOI] [PubMed] [Google Scholar]
  • 575.Noheria A, Ponamgi SP, Desimone CV, Vaidya VR, Aakre CA, Ebrille E, et al. Pulmonary Embolism in Patients with Transvenous Cardiac Implantable Electronic Device Leads. Europace . 2016;18(2):246–252. doi: 10.1093/europace/euv038. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 576.Fu HX, Huang XM, Zhong L, Osborn MJ, Bjarnason H, Mulpuru S, et al. Outcome and Management of Pacemaker-Induced Superior Vena Cava Syndrome. Pacing Clin Electrophysiol . 2014;37(11):1470–1476. doi: 10.1111/pace.12455. [DOI] [PubMed] [Google Scholar]
  • 577.Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction. Heart Rhythm . 2017;14(12):503–551. doi: 10.1016/j.hrthm.2017.09.001. [DOI] [PubMed] [Google Scholar]
  • 578.Riley RF, Petersen SE, Ferguson JD, Bashir Y. Managing Superior Vena Cava Syndrome as a Complication of Pacemaker Implantation: A Pooled Analysis of Clinical Practice. Pacing Clin Electrophysiol . 2010;33(4):420–425. doi: 10.1111/j.1540-8159.2009.02613.x. [DOI] [PubMed] [Google Scholar]
  • 579.Lee JC, Epstein LM, Huffer LL, Stevenson WG, Koplan BA, Tedrow UB. ICD Lead Proarrhythmia Cured by Lead Extraction. Heart Rhythm . 2009;6(5):613–618. doi: 10.1016/j.hrthm.2009.01.039. [DOI] [PubMed] [Google Scholar]
  • 580.Indik JH, Gimbel JR, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD, et al. 2017 HRS Expert Consensus Statement on Magnetic Resonance Imaging and Radiation Exposure in Patients with Cardiovascular Implantable Electronic Devices. Heart Rhythm . 2017;14(7):97–153. doi: 10.1016/j.hrthm.2017.04.025. [DOI] [PubMed] [Google Scholar]
  • 581.Celikyurt U, Agacdiken A, Bozyel S, Argan O, Sade I, Vural A, et al. Assessment of Shoulder Pain and Shoulder Disability in Patients with Implantable Cardioverter-Defibrillator. J Interv Card Electrophysiol . 2013;36(1):91–94. doi: 10.1007/s10840-012-9753-7. [DOI] [PubMed] [Google Scholar]
  • 582.Gula LJ, Ames A, Woodburn A, Matkins J, McCormick M, Bell J, et al. Central Venous Occlusion is not an Obstacle to Device Upgrade with the Assistance of Laser Extraction. Pacing Clin Electrophysiol . 2005;28(7):661–666. doi: 10.1111/j.1540-8159.2005.00163.x. [DOI] [PubMed] [Google Scholar]
  • 583.Pfitzner P, Trappe HJ. Oversensing in a Cardioverter Defibrillator System Caused by Interaction Between Two Endocardial Defibrillation Leads in the Right Ventricle. Pacing Clin Electrophysiol . 1998;21(4):764–768. doi: 10.1111/j.1540-8159.1998.tb00136.x. [DOI] [PubMed] [Google Scholar]
  • 584.Valentino V, Greenberg YJ, Saunders P, Yang F. An Unusual Interaction Between an Abandoned Pacing Lead and an ICD Lead. Heart Rhythm . 2015;12(6):1400–1401. doi: 10.1016/j.hrthm.2015.02.021. [DOI] [PubMed] [Google Scholar]
  • 585.Kay GN, Brinker JA, Kawanishi DT, Love CJ, Lloyd MA, Reeves RC, et al. Risks of Spontaneous Injury and Extraction of an Active Fixation Pacemaker Lead: Report of the Accufix Multicenter Clinical Study and Worldwide Registry. Circulation . 1999;100(23):2344–2352. doi: 10.1161/01.cir.100.23.2344. [DOI] [PubMed] [Google Scholar]
  • 586.Brunner MP, Cronin EM, Jacob J, Duarte VE, Tarakji KG, Martin DO, et al. Transvenous Extraction of Implantable Cardioverter-Defibrillator Leads Under Advisory--A Comparison of Riata, Sprint Fidelis, and Non-Recalled Implantable Cardioverter-Defibrillator Leads. Heart Rhythm . 2013;10(10):1444–1450. doi: 10.1016/j.hrthm.2013.06.021. [DOI] [PubMed] [Google Scholar]
  • 587.Hauser RG, Almquist AK. Learning from our Mistakes? Testing New ICD Technology. N Engl J Med . 2008;359(24):2517–2519. doi: 10.1056/NEJMp0805359. [DOI] [PubMed] [Google Scholar]
  • 588.Maytin M, Wilkoff BL, Brunner M, Cronin E, Love CJ, Grazia Bongiorni M, et al. Multicenter Experience with Extraction of the Riata/Riata ST ICD Lead. Heart Rhythm . 2014;11(9):1613–1618. doi: 10.1016/j.hrthm.2014.05.014. [DOI] [PubMed] [Google Scholar]
  • 589.Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H, et al. Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients with Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 Tesla. Circulation . 2006;114(12):1277–1284. doi: 10.1161/CIRCULATIONAHA.105.607655. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 590.Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, et al. A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients with Implanted Cardiac Devices. Ann Intern Med . 2011;155(7):415–424. doi: 10.7326/0003-4819-155-7-201110040-00004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 591.Mollerus M, Albin G, Lipinski M, Lucca J. Magnetic Resonance Imaging of Pacemakers and Implantable Cardioverter-Defibrillators without Specific Absorption Rate Restrictions. Europace . 2010;12(7):947–951. doi: 10.1093/europace/euq092. [DOI] [PubMed] [Google Scholar]
  • 592.Cohen JD, Costa HS, Russo RJ. Determining the Risks of Magnetic Resonance Imaging at 1.5 tesla for Patients with Pacemakers and Implantable Cardioverter Defibrillators. Am J Cardiol . 2012;110(11):1631–1636. doi: 10.1016/j.amjcard.2012.07.030. [DOI] [PubMed] [Google Scholar]
  • 593.Wazni O, Epstein LM, Carrillo RG, Love C, Adler SW, Riggio DW, et al. Lead Extraction in the Contemporary Setting: The LExICon Study: An Observational Retrospective Study of Consecutive Laser Lead Extractions. J Am Coll Cardiol . 2010;55(6):579–586. doi: 10.1016/j.jacc.2009.08.070. [DOI] [PubMed] [Google Scholar]
  • 594.Kalin R, Stanton MS. Current Clinical Issues for MRI Scanning of Pacemaker and Defibrillator Patients. Pacing Clin Electrophysiol . 2005;28(4):326–328. doi: 10.1111/j.1540-8159.2005.50024.x. [DOI] [PubMed] [Google Scholar]
  • 595.Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, et al. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N Engl J Med . 2017;376(8):755–764. doi: 10.1056/NEJMoa1603265. [DOI] [PubMed] [Google Scholar]
  • 596.Maytin M, Carrillo RG, Baltodano P, Schaerf RH, Bongiorni MG, Di Cori A, et al. Multicenter Experience with Transvenous Lead Extraction of Active Fixation Coronary Sinus Leads. Pacing Clin Electrophysiol . 2012;35(6):641–647. doi: 10.1111/j.1540-8159.2012.03353.x. [DOI] [PubMed] [Google Scholar]
  • 597.Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Turk KT, Reeves R, et al. Intravascular Extraction of Problematic or Infected Permanent Pacemaker Leads: 1994-1996. U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol . 1999;22(9):1348–1357. doi: 10.1111/j.1540-8159.1999.tb00628.x. [DOI] [PubMed] [Google Scholar]
  • 598.Tarakji KG, Wazni OM, Harb S, Hsu A, Saliba W, Wilkoff BL. Risk Factors for 1-year Mortality Among Patients with Cardiac Implantable Electronic Device Infection Undergoing Transvenous Lead Extraction: The Impact of the Infection Type and the Presence of Vegetation on Survival. Europace . 2014;16(10):1490–1495. doi: 10.1093/europace/euu147. [DOI] [PubMed] [Google Scholar]
  • 599.Ji SY, Gundewar S, Palma EC. Subclavian Venoplasty May Reduce Implant Times and Implant Failures in the Era of Increasing Device Upgrades. Pacing Clin Electrophysiol . 2012;35(4):444–448. doi: 10.1111/j.1540-8159.2011.03303.x. [DOI] [PubMed] [Google Scholar]
  • 600.Agarwal SK, Kamireddy S, Nemec J, Voigt A, Saba S. Predictors of Complications of Endovascular Chronic Lead Extractions from Pacemakers and Defibrillators: A Single-Operator Experience. J Cardiovasc Electrophysiol . 2009;20(2):171–175. doi: 10.1111/j.1540-8167.2008.01283.x. [DOI] [PubMed] [Google Scholar]
  • 601.Merchant FM, Levy MR, Kelli HM, Hoskins MH, Lloyd MS, Delurgio DB, et al. Predictors of Long-Term Survival Following Transvenous Extraction of Defibrillator Leads. Pacing Clin Electrophysiol . 2015;38(11):1297–1303. doi: 10.1111/pace.12733. [DOI] [PubMed] [Google Scholar]
  • 602.Hamid S, Arujuna A, Ginks M, McPhail M, Patel N, Bucknall C, et al. Pacemaker and Defibrillator Lead Extraction: Predictors of Mortality During Follow-up. Pacing Clin Electrophysiol . 2010;33(2):209–216. doi: 10.1111/j.1540-8159.2009.02601.x. [DOI] [PubMed] [Google Scholar]
  • 603.Cronin EM, Brunner MP, Tan CD, Rodriguez ER, Rickard J, Martin DO, et al. Incidence, Management, and Outcomes of the Arteriovenous Fistula Complicating Transvenous Lead Extraction. Heart Rhythm . 2014;11(3):404–411. doi: 10.1016/j.hrthm.2013.11.024. [DOI] [PubMed] [Google Scholar]
  • 604.Di Monaco A, Pelargonio G, Narducci ML, Manzoli L, Boccia S, Flacco ME, et al. Safety of Transvenous Lead Extraction According to Centre Volume: A Systematic Review and Meta-analysis. Europace . 2014;16(10):1496–1507. doi: 10.1093/europace/euu137. [DOI] [PubMed] [Google Scholar]
  • 605.Pinski SL, Trohman RG. Interference in Implanted Cardiac Devices, Part II. Pacing Clin Electrophysiol . 2002;25(10):1496–1509. doi: 10.1046/j.1460-9592.2002.01496.x. [DOI] [PubMed] [Google Scholar]
  • 606.Lin Y, Melby DP, Krishnan B, Adabag S, Tholakanahalli V, Li JM. Frequency of Pacemaker Malfunction Associated with Monopolar Electrosurgery During Pulse Generator Replacement or Upgrade Surgery. J Interv Card Electrophysiol . 2017;49(2):205–209. doi: 10.1007/s10840-017-0241-y. [DOI] [PubMed] [Google Scholar]
  • 607.Crossley GH, Poole JE, Rozner MA, Asirvatham SJ, Cheng A, Chung MK, et al. The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the Perioperative Management of Patients with Implantable Defibrillators, Pacemakers and Arrhythmia Monitors: Facilities and Patient Management this Document was Developed as a Joint Project with the American Society of Anesthesiologists (ASA), and in Collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS) Heart Rhythm . 2011;8(7):1114–1154. doi: 10.1016/j.hrthm.2010.12.023. [DOI] [PubMed] [Google Scholar]
  • 608.Jacob S, Panaich SS, Maheshwari R, Haddad JW, Padanilam BJ, John SK. Clinical Applications of Magnets on Cardiac Rhythm Management Devices. Europace . 2011;13(9):1222–1230. doi: 10.1093/europace/eur137. [DOI] [PubMed] [Google Scholar]
  • 609.Rodriguez-Blanco YF, Souki F, Tamayo E, Candiotti K. Magnets and Implantable Cardioverter Defibrillators: What’s the Problem? Ann Card Anaesth . 2013;16(1):54–57. doi: 10.4103/0971-9784.105372. [DOI] [PubMed] [Google Scholar]
  • 610.Gifford J, Larimer K, Thomas C, May P, Stanhope S, Gami A. Randomized Controlled Trial of Perioperative ICD Management: Magnet Application Versus Reprogramming. Pacing Clin Electrophysiol . 2014;37(9):1219–1224. doi: 10.1111/pace.12417. [DOI] [PubMed] [Google Scholar]
  • 611.Sutton R, Kanal E, Wilkoff BL, Bello D, Luechinger R, Jenniskens I, et al. Safety of Magnetic Resonance Imaging of Patients with a New Medtronic EnRhythm MRI SureScan Pacing System: Clinical Study Design. Trials . 2008;9:68. doi: 10.1186/1745-6215-9-68. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 612.Kanal E, Barkovich AJ, Bell C, Borgstede JP, Bradley WG, Jr, Froelich JW, et al. ACR Guidance Document on MR Safe Practices: 2013. J Magn Reson Imaging . 2013;37(3):501–530. doi: 10.1002/jmri.24011. [DOI] [PubMed] [Google Scholar]
  • 613.Verma A, Ha AC, Dennie C, Essebag V, Exner DV, Khan N, et al. Canadian Heart Rhythm Society and Canadian Association of Radiologists Consensus Statement on Magnetic Resonance Imaging with Cardiac Implantable Electronic Devices. Can Assoc Radiol J . 2014;65(4):290–300. doi: 10.1016/j.carj.2014.08.001. [DOI] [PubMed] [Google Scholar]
  • 614.Rosatti SFC, Araújo-Moreira FM, Trevelin LC, Rosa Ressonância Magnética de Tórax em Portadores de Dispositivos Cardíacos Eletrônicos Implantáveis Condicionais para Ressonância Magnética: Contraindicação Clássica ou Exame Seguro. Relampa . 2015;28(1):3–11. [Google Scholar]
  • 615.Fetter J, Aram G, Holmes DR, Jr, Gray JE, Hayes DL. The Effects of Nuclear Magnetic Resonance Imagers on External and Implantable Pulse Generators. Pacing Clin Electrophysiol . 1984;7(4):720–727. doi: 10.1111/j.1540-8159.1984.tb05602.x. [DOI] [PubMed] [Google Scholar]
  • 616.Nordbeck P, Ertl G, Ritter O. Magnetic Resonance Imaging Safety in Pacemaker and Implantable Cardioverter Defibrillator Patients: How Far Have We Come? Eur Heart J . 2015;36(24):1505–1511. doi: 10.1093/eurheartj/ehv086. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 617.Gimbel JR, Bello D, Schmitt M, Merkely B, Schwitter J, Hayes DL, et al. Randomized Trial of Pacemaker and Lead System for Safe Scanning at 1.5 Tesla. Heart Rhythm . 2013;10(5):685–691. doi: 10.1016/j.hrthm.2013.01.022. [DOI] [PubMed] [Google Scholar]
  • 618.Wilkoff BL, Bello D, Taborsky M, Vymazal J, Kanal E, Heuer H, et al. Magnetic Resonance imagi|ng in Patients with a Pacemaker System Designed for the Magnetic Resonance Environment. Heart Rhythm . 2011;8(1):65–73. doi: 10.1016/j.hrthm.2010.10.002. [DOI] [PubMed] [Google Scholar]
  • 619.Yeung C, Chacko S, Glover B, Campbell D, Crystal E, Ben-Dov N, et al. Radiotherapy for Patients with Cardiovascular Implantable Electronic Devices: A Review. Can J Cardiol . 2018;34(3):244–251. doi: 10.1016/j.cjca.2017.11.023. [DOI] [PubMed] [Google Scholar]
  • 620.Grant JD, Jensen GL, Tang C, Pollard JM, Kry SF, Krishnan S, et al. Radiotherapy-Induced Malfunction in Contemporary Cardiovascular Implantable Electronic Devices: Clinical Incidence and Predictors. JAMA Oncol . 2015;1(5):624–632. doi: 10.1001/jamaoncol.2015.1787. [DOI] [PubMed] [Google Scholar]
  • 621.Zaremba T, Jakobsen AR, Thøgersen AM, Oddershede L, Riahi S. The Effect of Radiotherapy Beam Energy on Modern Cardiac Devices: An in Vitro Study. Europace . 2014;16(4):612–616. doi: 10.1093/europace/eut249. [DOI] [PubMed] [Google Scholar]
  • 622.Salerno F, Gomellini S, Caruso C, Barbara R, Musio D, Coppi T, et al. Management of Radiation Therapy Patients with Cardiac Defibrillator or Pacemaker. Radiol Med . 2016;121(6):515–520. doi: 10.1007/s11547-015-0616-z. [DOI] [PubMed] [Google Scholar]
Arq Bras Cardiol. 2023 Jan 16;120(1):e20220892. [Article in English]

Brazilian Guidelines for Cardiac Implantable Electronic Devices – 2023

Ricardo Alkmim Teixeira 1, Alexsandro Alves Fagundes 2, José Mário Baggio Junior 3, Júlio César de Oliveira 4, Paulo de Tarso Jorge Medeiros 5, Bruno Pereira Valdigem 5, Luiz Antônio Castilho Teno 6, Rodrigo Tavares Silva 7,8, Celso Salgado de Melo 9, Jorge Elias Neto 10, Antonio Vitor Moraes Júnior 11,12, Anisio Alexandre Andrade Pedrosa 13, Fernando Mello Porto 14, Hélio Lima de Brito Júnior 15, Thiago Gonçalves Schroder e Souza 16, José Carlos Pachón Mateos 5, Luis Gustavo Belo de Moraes 17, Alexander Romeno Janner Dal Forno 18, Andre Luiz Buchele D’Avila 18, Diogo Alberto de Magalhães Cavaco 19, Ricardo Ryoshim Kuniyoshi 20,21, Mauricio Pimentel 22, Luiz Eduardo Montenegro Camanho 23, Eduardo Benchimol Saad 23,24, Leandro Ioschpe Zimerman 25, Eduardo Bartholomay Oliveira 26, Mauricio Ibrahim Scanavacca 13, Martino Martinelli Filho 13, Carlos Eduardo Batista de Lima 27,28, Giselle de Lima Peixoto 29, Francisco Carlos da Costa Darrieux 13, Jussara de Oliveira Pinheiro Duarte 30, Silas dos Santos Galvão Filho 31, Eduardo Rodrigues Bento Costa 32, Enrique Indalécio Pachón Mateo 33, Sissy Lara De Melo 13, Thiago da Rocha Rodrigues 34, Eduardo Arrais Rocha 35, Denise Tessariol Hachul 13, Adalberto Menezes Lorga Filho 36, Silvana Angelina D’Orio Nishioka 13, Eduardo Barreto Gadelha 37, Roberto Costa 38, Veridiana Silva de Andrade 39, Gustavo Gomes Torres 40, Nestor Rodrigues de Oliveira Neto 40, Fernando Antonio Lucchese 41, Henrique Murad 42, José Wanderley Neto 43, Paulo Roberto Slud Brofman 44, Rui M S Almeida 45, João Carlos Ferreira Leal 46

Brazilian Guidelines for Cardiac Implantable Electronic Devices – 2023
The report below lists declarations of interest as reported to the SBC by the experts during the period of the development of these statement, 2019/2022.
Expert Type of relationship with industry
Adalberto Menezes Lorga Filho
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc.

  • From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Vatis / Libbs; Lixiana / Daiichi Sankyo; Eliquis / Pfizer: Atrial fibrillation.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Daiichi Sankyo: Lixiana

Alexander Romeno Janner Dal Forno
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc.

  • From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biotronik: Pacemaker, Abbott: Atrial fibrillation.

  • B - Research funding under your direct/personal responsibility (directed to the department or institution) from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Proadi.

Alexsandro Alves Fagundes
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer: Xarelto, Medtronic, Sankyo, Libbs.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Boston.

Andre Luiz Buchele D’Avila Nothing to be declared
Anisio Alexandre Andrade Pedrosa
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Medtronic / Biomedical / Biocath / Imagem: Proctor.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Medtronic, Biocath, Biomedical.

Antonio Vitor Moraes Júnior
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biotronik.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Xarelto / Bayer: Atrial fibrillation.

Bruno Pereira Valdigem Nothing to be declared
Carlos Eduardo Batista de Lima
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Daiichi Sankyo / Lixiana: Atrial fibrillation.

  • C - Personal research funding paid by the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - RIVER AF trial, Thrombosis Research Institute / Brazilian Clinical Research.

Celso Salgado de Melo Nothing to be declared
Denise Tessariol Hachul Nothing to be declared
Diogo Alberto de Magalhães Cavaco
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Boston Scientific/S-ICD.

Eduardo Arrais Rocha
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - ABBOTT: Remote monitoring class.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biotronik / Bayer: Congress.

  • Any economically relevant equity interest in companies in the healthcare or education industry or in any companies competing with or supplying to SBC:

  • - Health Area.

Eduardo Barreto Gadelha Nothing to be declared
Eduardo Bartholomay Oliveira
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biotronik.

Eduardo Benchimol Saad
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biosense Webster: Catheter ablation.

Eduardo Rodrigues Bento Costa Nothing to be declared
Enrique Indalécio Pachón Mateo Nothing to be declared
Fernando Antônio Lucchese Nothing to be declared
Fernando Mello Porto Nothing to be declared
Francisco Carlos da Costa Darrieux
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer / Boehringer Ingelheim / Pfizer / Libbs / Biolab: Anticoagulation and atrial fibrillation.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer / Pfizer: Congresses and Virtual Courses.

Giselle de Lima Peixoto Nothing to be declared
Gustavo Gomes Torres
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biotronik.

Hélio Lima de Brito Júnior
  • Other relationships

  • Any economically relevant equity interest in companies in the healthcare or education industry or in any companies competing with or supplying to SBC:

  • - Health Area.

Henrique Murad Nothing to be declared
Jorge Elias Neto Nothing to be declared
João Carlos Ferreira Leal
  • Financial declaration

  • B - Research funding under your direct/personal responsibility (directed to the department or institution) from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Braile Biomédica: Inovare® Alpha - Biological Valve Prosthesis / Artivion-Neomex: Evita open plus, hybrid stent.

José Carlos Pachón Mateos Nothing to be declared
José Mário Baggio Junior
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Abbott: Webinar Lecture; Biotronik: Satellite Symposium at Congress; Medtronic: Satellite Symposium at Congress.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Medtronic: Training on leadless pacing and physiological stimulation; Biotronik: Congress; Abbott: Product training.

José Wanderley Neto Nothing to be declared
Júlio César de Oliveira
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Infinity.

Jussara de Oliveira Pinheiro Duarte Nothing to be declared
Leandro Ioschpe Zimerman
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer/Xarelto; Daiichi Sankyo/Lixiana; Libbs/Propafenona e Amiodarona; Pfizer/Eliquis.

Luis Gustavo Belo de Moraes
  • Other relationships Any economically relevant equity interest in companies in the healthcare or education industry or in any companies competing with or supplying to SBC:

  • - Sócio cotista da Cardioritmo Serviços Médicos Ltda e Luis Belo Serviços Médicos.

Luiz Antônio Castilho Teno Nothing to be declared
Luiz Eduardo Montenegro Camanho Nothing to be declared
Martino Martinelli Filho Nothing to be declared
Mauricio Ibrahim Scanavacca
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Johnson & Johnson.

  • B - Research funding under your direct/personal responsibility (directed to the department or institution) from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Milestone Pharmaceuticals: etripamil NS for the treatment of Paroxysmal.

  • C - Personal research funding paid by the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Medtronic, J&J and Abbott.

Mauricio Pimentel
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer: Anticoagulation.

Nestor Rodrigues de Oliveira Neto Nothing to be declared
Paulo Roberto Slud Brofman Nothing to be declared
Paulo de Tarso Jorge Medeiros Nothing to be declared
Ricardo Alkmim Teixeira
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Boehringer-Ingelheim: Pradaxa, Jardiance; Daiichi-Sankyo: Lixiana; Abbott: Implantable Electronic Heart Devices; Biotronik: Implantable Electronic Heart Devices; Medtronic: Implantable Electronic Heart Devices; Biomedical: Lead Extraction from Implantable Electronic Cardiac Devices.

  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Biomedical: Lead Extraction from Implantable Electronic Cardiac Devices.

Ricardo Ryoshim Kuniyoshi Nothing to be declared
Roberto Costa
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Boston Scientific: Speaker and consultant.

  • Other relationships

  • Participation in procurement committees for supplies or drugs in health institutions or any similar roles taken:

  • - Reviewer of the auctions of the Hospital das Clínicas of FMUSP.

Rodrigo Tavares Silva Nothing to be declared
Rui M. S. Almeida
  • Other relationships

  • Any economically relevant equity interest in companies in the healthcare or education industry or in any companies competing with or supplying to SBC:

  • - Health and Education Area.

Silas dos Santos Galvão Filho Nothing to be declared
Silvana Angelina D’Orio Nishioka
  • Other relationships

  • Funding of continuing medical education activities, including travel, accommodation and registration in conferences and courses, from the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Abbott.

  • Performance, in the previous year, as a medical auditor for health insurance companies or the like:

  • - For medical classes and training only.

Sissy Lara de Melo Nothing to be declared
Thiago da Rocha Rodrigues
  • Financial declaration

  • A - Economically relevant payments of any kind made to (i) you, (ii) your spouse/partner or any other person living with you, (iii) any legal person in which any of these is either a direct or indirect controlling owner, business partner, shareholder or participant; any payments received for lectures, lessons, training instruction, compensation, fees paid for participation in advisory boards, investigative boards or other committees, etc. From the brazilian or international pharmaceutical, orthosis, prosthesis, equipment and implants industry:

  • - Bayer / Pfizer: eliquis; Daichi-Sanchio: lixiana; Abbott: rivacrist; Bhoeringer-Ingelheim: jardiance; Libbs: Vatis.

Thiago Gonçalves Schroder e Souza Nothing to be declared
Veridiana Silva de Andrade Nothing to be declared

Contents

1. General Recommendations 8

1.1 Operating Room 8

1.1.1. Human Resources 8

1.1.2. Material Resources 9

1.1.2.1. Fluoroscopy 9

1.1.2.2. Monitoring 9

1.1.2.3. Surgical Instruments 9

1.2. CIED Evaluation and Programming Clinic 9

1.3. Clinical Evaluation Prior to CIED Implantation 9

1.4. Surgical Procedure and Types of CIED 9

2. Recommendations for Conventional Pacemaker Implantation 10

2.1. Sinus Node Disease 10

2.2. Atrioventricular and Intraventricular Blocks 12

2.2.1. Atrioventricular Blocks 12

2.2.2. Intraventricular Blocks with 1:1 Atrioventricular Conduction 14

2.3. Carotid Sinus Syndrome 14

2.4. Vasovagal Syndrome 16

2.5. Hypertrophic Cardiomyopathy 16

2.6. Neuromuscular Diseases 17

2.7 Obstructive Sleep Apnea Syndrome 18

2.8. Congenital Long QT Syndrome 18

2.9. Cardiac Transplantation 19

2.10. Choosing the Type of Pacemaker and Pacing Mode 20

2.11. Direct Stimulation of the Cardiac Conduction System (His Bundle, Left Bundle Branch) 20

2.12. Leadless Pacing 21

3. Recommendations for Multisite Pacemaker Implantation/Cardiac Resynchronization Therapy 22

3.1. Patients in Sinus Rhythm 22

3.2. Patients with Atrial Fibrillation 24

3.3. Conventional PM Upgrade 25

3.4. Indication for Antibradycardia Pacing (First Implant) 26

3.5. Indication for Implantable Cardioverter-defibrillator Combined with Cardiac Resynchronization Therapy 26

3.6. Direct Stimulation of the Cardiac Conduction System 27

3.6.1. Cardiac Contractility Modulation 28

4. Recommendations for Placement of Implantable Cardioverter-defibrillators 28

4.1. Primary Prevention of Sudden Death 28

4.1.1. Ischemic Cardiomyopathy 28

4.1.2. Nonischemic Cardiomyopathy 29

4.1.3. Hypertrophic Cardiomyopathy 30

4.1.4. Chagas Cardiomyopathy 32

4.1.5. Arrhythmogenic Right Ventricular Cardiomyopathy 33

4.1.6. Noncompaction Cardiomyopathy 34

4.1.7. Congenital Long and Short QT Syndromes 37

4.1.8. Brugada Syndrome 37

4.1.9. Catecholaminergic polymorphic ventricular tachycardia 38

4.1.10. Idiopathic Ventricular Tachycardia 38

4.2. Secondary prevention of sudden death 38

4.2.1. Recovered cardiac arrest or sustained ventricular tachycardia 38

4.2.1.1. Recovered cardiac arrest or sustained ventricular tachycardia in the presence of structural heart disease 38

4.2.1.2. Survivors of Cardiac Arrest or Sustained Ventricular Tachycardia in the Absence of Structural Heart Disease 40

4.2.2. Syncope and Ventricular Tachycardia/Fibrillation on Electrophysiology Study 40

4.3. Children, Adolescents, and Congenital Heart Disease 41

4.4. Choosing Implantable Cardioverter-defibrillator Type and Pacing Mode 44

4.4.1. Implantation Technique 44

4.4.2. Pacing Mode 44

4.5. Cost-effectiveness of Implantable Cardioverter-defibrillators in Primary and Secondary Prevention of Sudden Death 44

4.5.1. Primary Prevention 44

4.5.2. Secondary Preventio n 45

5. Recommendations for Implantable Loop Recorders 45

6. Recommendations for CIED Evaluation and Programming 46

6.1. Conventional Pacemakers 46

6.1.1. Sinus Node Disease 46

6.1.2. Atrioventricular Block 47

6.1.3. Atrial Fibrillation 47

6.1.4. Neurally Mediated Syncope and Carotid Sinus Syndrome 47

6.2. Cardiac Resynchronization Therapy 47

6.3. Implantable Cardioverter-defibrillator 49

6.4. Implantable Loop Recorder 50

6.5. Remote Monitoring (Online) 50

7. Recommendations for Prevention and Treatment of CIED infecTions and for System Removal 51

7.1. Prevention and Treatment of Infections 51

7.2. Lead Removal from Cardiac Implantable Electronic Devices 55

8. Recommendations for the prevention of electromagnetic interference 58

8.1. Surgery Using Electrocautery 59

8.2. Magnetic Resonance Imaging 60

8.3. Radiotherapy 62

9. Conclusion 62

References 64

1. General Recommendations

Although implantation techniques for cardiac implantable electronic devices (CIEDs) have been standardized and simplified, adequate settings and materials, as well as medical knowledge and surgical experience, remain necessary. Electrocardiographic (ECG) knowledge, especially of cardiac arrhythmias and the principles of cardiac electrophysiology, is essential.

1.1. Operating Room

Surgical procedures involving artificial cardiac pacing are performed by cardiovascular surgeons or cardiologists with cardiac pacing training (SOBRAC/SBC, ABEC/SBCCV, AMB). The procedures should be performed in an operating room or a catheterization/electrophysiology laboratory. The operating room should have adequate size, lighting, and ventilation, a sink for surgical hand antisepsis, and a dual-voltage electrical system (with a grounding system that prevents electromagnetic interference and protects the materials).

1.1.1. Human Resources

Professionals involved in surgical procedures for CIED implantation include:

  1. Attending and assisting physicians with cardiac pacing training

  2. Anesthesiologist

  3. Scrub nurse preferably with cardiac pacing training

  4. Nursing professional preferably with cardiac pacing training

  5. Pacemaker (PM) technician

  6. Radiology technician

1.1.2. Material Resources

1.1.2.1. Fluoroscopy

A fluoroscopy system (with an image intensifier) remains essential during CIED procedures. The equipment may be fixed, as in catheterization laboratories, or portable (C-arm). Image quality and image recording and mirroring resources facilitate the procedure, especially for cardiac resynchronization therapy (CRT). The image intensifier should allow visualization of small-caliber guidewires and movements from different views (including oblique views).

1.1.2.2. Monitoring

Continuous ECG monitoring should be performed, and ECG tracings may be stored. Available leads should allow proper intraoperative assessment of CRT and physiological pacing of the conduction system (His bundle, left bundle branch [LBB], deep septal). In these cases, an intracavitary ECG (polygraph) should be analyzed.

Noninvasive monitoring of blood pressure and pulse oximetry should be available.

1.1.2.3. Surgical Instruments
  1. A tray with the adequate surgical instruments

  2. Electrocautery device

  3. External cardioverter-defibrillator

  4. Temporary external PM

  5. Advanced life support system

  6. Materials and drugs for anesthesia and cardiovascular stability (analgesics, anesthetics, antiarrhythmics, vasoactive drugs, antibiotics etc.)

  7. Pulse generator, leads, introducers, and sheaths for catheterization of the coronary sinus and conduction system

  8. CIED-specific programmer and analyzer being used or to be implanted

  9. Ultrasound for venous access may reduce complications related to deep vein puncture (eg, accidental arterial puncture, pneumothorax)

  10. Transesophageal echocardiography: helpful during percutaneous lead extraction for early diagnosis of cardiac tamponade

1.2. CIED Evaluation and Programming Clinic

The physician in charge of a CIED follow-up clinic should have cardiac pacing training (SOBRAC/SBC, ABEC/SBCCV, AMB). The clinic’s structure should include:

  1. ECG machine

  2. CIED programmer devices from different manufacturers

  3. Uninterruptible power supply

  4. External cardioverter-defibrillator with transcutaneous PM

  5. Magnet

  6. Transthoracic echocardiogram

  7. Access to additional tests such as cardiac stress test, 24-hour Holter monitoring, and imaging tests (radiography, magnetic resonance imaging [MRI], myocardial scintigraphy, computed tomography). The tilt table test must be available at the clinic or at a referral facility.

  8. Access to an engineer specializing in cardiac pacing

1.3. Clinical Evaluation Prior to CIED Implantation

a) Initial clinical evaluation prior to CIED implantation should include:

Patient history and physical examination

Patient history should investigate signs and symptoms of cardiac arrhythmias, such as syncope, presyncope, dizziness, and palpitations, and signs and symptoms of heart failure (HF). A family history of sudden death is highly relevant, especially if occurring prematurely or affecting first-degree relatives.

Physical examination should include inspection, peripheral pulse palpation, blood pressure measurement, cardiac and carotid auscultation, heart rate, and peripheral perfusion.

If possible, oral anticoagulants and antiplatelets should be temporarily interrupted before the surgical procedure. 1 Preventive interruption of other drugs is generally unnecessary. Patients with signs of active infection must not undergo device implantation until the infection is resolved.

b) Preoperative additional tests

  1. Resting ECG;

  2. Chest radiograph (posteroanterior view plus left profile);

  3. Laboratory tests: complete blood count and coagulation profile tests are required for all patients. For procedures that require intravenous contrast (eg, CRT and venous obstruction), a renal function evaluation with electrolyte measurement should be performed. In patients with diabetes, fasting glucose should be assessed. Urinalysis and urine culture tests are indicated for patients with urinary complaints;

  4. Tests such as echocardiogram, 24-hour Holter monitoring, electrophysiology study (EPS), and upper-limb ultrasound or venography should be performed only if the clinical condition requires.

Patients should fast for at least 6 to 8 hours before surgery depending on the complexity of the procedure and type of anesthesia. Trichotomy, local antisepsis, and prophylactic antibiotic therapy 2 should follow institutional protocols.

1.4. Surgical Procedure and Types of CIED

a) Surgical procedures

Before surgery, medical and nursing staff must follow safe surgery protocols by checking patient’s name, date of birth, hospital record number, and laterality. The staff should also confirm the indication for the procedure and check preoperative test results.

CIED procedures should be performed in an operating room or a catheterization/electrophysiology laboratory under fluoroscopic guidance with continuous ECG monitoring, pulse oximetry, and intermittent or continuous blood pressure measurement. Devices such as subcutaneous implantable cardioverter-defibrillators (ICDs) and implantable loop recorders do not require the use of an image intensifier during implantation.

Anesthesia is either local, preferably combined with sedation, or general. The type of anesthesia depends on the complexity of the procedure, access route, and patient’s clinical status.

The choice of surgical access should consider the following: pulse generator implantation site, heart access for lead implantation (intravenous or epicardial route), and the possibility of implanting devices without transvenous leads (leadless PM and subcutaneous ICD). The use of a temporary PM and venous catheters, previous thoracic surgeries, need for radiotherapy, anatomical characteristics, skin infections, and dominant arm may also influence the choice of surgical strategy.

The pulse generator pocket is generally created in the pectoral region, although it may be abdominal in specific situations, and the device is placed in a subcutaneous or submuscular position. Venous access is achieved by cephalic vein dissection or puncture of the axillary, subclavian, jugular, or femoral vein. The number of leads varies according to the type of implanted device, usually ranging from one to three. Active-fixation (screw-in) leads are currently preferred to passive-fixation leads according to professional experience.

During the surgical procedure, pacing and sensing thresholds and lead impedance parameters should be obtained, and an endocavitary or epicardial electrogram should be performed. For patients undergoing ICD implantation, shock impedance should be measured; defibrillation threshold testing is optional for patients undergoing transvenous ICD implantation (usually unnecessary) but recommended for patients undergoing right pectoral subcutaneous ICD implantation. 3 - 5 Multisite pacing systems with lead implantation for left ventricular (LV) pacing through the carotid sinus require specific tools, such as sheaths, electrophysiology catheter, and venography catheter for choosing the best tributary vein for lead implantation.

The CIED implantation report should include patient identification, surgery description, technical data regarding the system, and if there were any complications (eg, pneumothorax, hemothorax, failure to capture and/or sense, bad lead-generator connection, lead dislodgement, right ventricle (RV) perforation, diaphragmatic stimulation, pocket hematoma, contamination, and arrhythmias). The Brazilian Pacemaker Registry must be completed.

Postoperative and inpatient evaluation

After CIED implantation, the patient should be clinically evaluated, and the implanted device should undergo electronic assessment. ECG and chest radiography should be performed to confirm proper device function and lead position and to identify any dysfunctions or complications.

Patients are generally hospitalized for 12 to 24 hours. Those undergoing procedures that do not require intravascular access (pulse generator replacement or subcutaneous device implantation) usually remain under postoperative observation for 6 to 12 hours (day hospital).

b) Types of CIED

The main types of CIED and their characteristics are summarized in Chart 1.

Chart 1. – Main types of CIED and their characteristics.

Device Characteristics
Pacemaker (PM)
  • ✓ Multiprogrammable electronic cardiac device capable of monitoring and promoting electrical cardiac stimulation, restoring atrioventricular synchrony and heart rate variability, detecting and recording arrhythmias, and treating atrial arrhythmias using mechanisms of atrial overpacing

  • ✓ PMs usually consist of a pulse generator, an electronic circuit, and leads that connect the generator to the endocardial surface. The devices may be single- or dual-chamber

  • ✓ Recently introduced in the market, leadless PMs are electrical stimulation systems in which all device units are contained in a single intracardiac component implanted via a transvenous sheath system

  • ✓ Bradyarrhythmias are the focus of PM therapy

Cardiac resynchronization therapy (CRT)
  • ✓ Cardiac device with antibradycardia pacing function similar to a PM but allowing multisite cardiac pacing, ie, left and right ventricular (biventricular) pacing, for intraventricular dyssynchrony correction

  • ✓ Adjuvant treatment of advanced heart failure is the primary focus of CRT

Implantable cardioverter-defibrillator (ICD)
  • ✓ Electronic cardiac device with the same functions of a conventional PM plus the ability to detect, record, and automatically treat potentially fatal ventricular tachyarrhythmias via shock therapy (cardioversion or defibrillation) or programmed ventricular pacing (overdrive or antitachycardia pacing)

  • ✓ Potentially fatal ventricular tachyarrhythmias associated or not with bradyarrhythmias may be treated with ICD

CRT + ICD ✓ Devices that combine CRT and ICD functions in a single device
Implantable loop recorder ✓ Subcutaneous cardiac device for prolonged cardiac monitoring (3 to 4 years) with the aim of detecting intermittent and sporadic cardiac arrhythmias in cases of unexplained syncope or cryptogenic stroke via continuous recording of the heart’s electrical activity

2. Recommendations for Conventional Pacemaker Implantation

2.1. Sinus Node Disease

Symptomatic sinus node dysfunctions are named sinus node disease (SND) and tend to be the most common indication for artificial cardiac pacing worldwide, accounting for approximately half of permanent PM implants. 6

From an ECG perspective, SND is characterized by the presence of one or more of the following manifestations: sinus bradycardia, sinus pause or arrest, sinoatrial block, and atrial tachyarrhythmias (especially flutter or atrial fibrillation [AF]) associated with bradyarrhythmias (sinus pauses), such as tachycardia-bradycardia syndrome and chronotropic incompetence (inadequate heart rate response to exercise or stress). 7 - 9

SND symptoms are related to low heart rate or to the duration of sinus pause. The most common symptoms include palpitation, tiredness, dyspnea, dizziness, and presyncope or syncope. Syncope is a common clinical symptom and may affect approximately 50% of patients referred to PM implantation due to SND. 10

Although SND may occur at any time in life, the incidence increases with age, affecting 1 in 600 patients older than 65 years. Men and women are affected equally. 11 - 13

SND pathophysiology is diverse and usually involves complex electrophysiological and structural remodeling. 14 , 15 Etiology can be divided into intrinsic and extrinsic factors.

Intrinsic causes of SND include inflammatory, infectious, and immunological processes, degenerative fibrosis, ion channel dysfunction, and sinoatrial node remodeling. Age-related idiopathic degenerative fibrosis is the most common intrinsic cause of SND. However, recent studies have demonstrated that an inherited ion channel dysfunction may also play a role in the genesis of age-related sinus dysfunction. 16 - 19 In addition, baroreflex response and heart rate variability are reduced in older patients. 20 , 21

Other intrinsic mechanisms of SND include infiltrative (eg, hemochromatosis, amyloidosis) and inflammatory (eg, sarcoidosis) diseases, HF and AF (anatomical and structural changes along the crista terminalis), and chronic coronary heart disease (sinoatrial nodal artery involvement). Genomic analyses have identified loci in proteins that interact with ion channels and channels related to normal and abnormal resting heart rates, providing insights into the mechanisms that control heart rate. 22 - 24

The main extrinsic causes of SND include pharmacological agents, metabolic disorders, and autonomic dysfunction. Betablockers, calcium channel blockers, digitalis, antiarrhythmics, and sympatholytic drugs are the pharmacological agents most commonly associated with the development of SND. 24 Lithium may also be associated with SND, often permanent. 25

Autonomic dysfunction with cardioinhibitory manifestations can mimic or intensify SND in vasovagal syncope and carotid sinus hypersensitivity. 26

Another rare cause of recurrent bradycardia and syncope is cardiac asystole induced by temporal lobe epilepsy. Although temporary PM implantation may be fundamental in the acute phase of this condition, specific management (surgical or pharmacological) of epilepsy may often regulate sinus dysfunction and/or atrioventricular (AV) block with no need for permanent PM implantation. The same considerations are applicable to another rare condition, namely glossopharyngeal neuralgia associated with cardiac asystole and syncope. 27 , 28

Metabolic abnormalities, such as severe systemic acidosis, hyperkalemia, hypokalemia, and hypocalcemia, may cause sinus bradycardia, which is uncommon in acute cases. Other possible extrinsic factors include hypothyroidism, hypoxia, hypothermia, and toxins. The association between Brugada syndrome and the occurrence of SND has also been described. 29

Patients with SND are usually asymptomatic in the initial phases of the disease and develop symptoms after several years. It is also important to identify cases of asymptomatic functional bradycardia, such as nocturnal bradycardia, in healthy young people and athletes, although this condition poses no health risks and requires no intervention.

Documenting a correlation between ECG changes and clinical manifestations is essential to SND diagnosis. Twelve-lead ECG or other methods, such as Holter monitoring (24-48 hours, 7 days) and external or implantable loop recorders, may be used. 30 Invasive EPS should not be regularly used in clinical practice because there are no conclusive data on the true indication for permanent PM implantation in patients with abnormal sinus node recovery time or sinoatrial conduction time.

Before choosing the best approach and deciding if PM implantation should be performed in patients with SND, establishing a correlation between bradycardia and clinical symptoms is crucial as well as identifying reversible causes.

In symptomatic SND without a reversible cause, permanent PM implantation is the treatment of choice, although there is no evidence that artificial cardiac pacing has any impact on survival or risk of sudden death in these patients compared with the general population. 30 However, PM implantation significantly increases quality of life, possibly reduces the risk of AF and systemic thromboembolism, allows the use of antiarrhythmics that may cause bradycardia, and provides continuous monitoring of heart rhythm. 31 The recommendations for permanent PM implantation in SND are described in Table 1 .

Table 1. – Recommendations for permanent pacemaker implantation in sinus node disease.

  Class of recommendation Level of evidence
Spontaneous SND (sinus bradycardia, sinus pause/arrest, sinoatrial block, or tachycardia-bradycardia syndrome) without a treatable cause or induced by necessary and irreplaceable drugs with documented symptoms of bradycardia (syncope, presyncope, dizziness, or tiredness/fatigue) I C
Spontaneous SND (sinus bradycardia, sinus pause/arrest, sinoatrial block, or tachycardia-bradycardia syndrome) without a treatable cause or induced by necessary and irreplaceable drugs without documented symptoms (syncope, presyncope, dizziness, or tiredness/fatigue) after a recommended investigation (presumptive diagnosis) Unexplained syncope with evidence of SND on EPS IIa C
Unexplained syncope with documented asymptomatic sinus pause > 6s IIb C
Asymptomatic patient with documented symptoms that are clearly unrelated to bradycardia; symptomatic patient with bradycardia due to reversible causes, including nonessential drugs III C

EPS: electrophysiology study; SND: sinus node disease.

Regarding pacing mode, major randomized studies have reported no improved survival with atrial (AAI) or AV (DDD) pacing compared with ventricular (VVI) pacing; however, benefits such as reductions in AF, syncope, and PM syndrome rates have been observed. 32 - 35 A systematic review of these major randomized studies identified a significant reduction in stroke (hazard ratio [HR]: 0.81) and AF (HR: 0.80) rates with AAI and/or DDD pacing compared with VVI pacing. 36 The effect of pacing mode on HF and stroke prevention and quality-of-life improvement is less evident. The Danish Multicenter Randomized Trial on Single Lead Atrial Pacing versus Dual Chamber Pacing in Sick Sinus Syndrome (DANPACE) 10 analyzed AAI vs DDD pacing in patients with SND and showed that AAIR pacing was associated with a higher incidence of paroxysmal AF (HR: 1.24) and a two-fold increase in reoperations compared with DDDR pacing. Reoperations were mostly due to the need for an upgrade from AAIR to DDDR pacing in patients who developed AV block during follow-up. Another relevant aspect in that study is that the benefit of AAI pacing may be attenuated in patients with long PR intervals that may trigger diastolic mitral regurgitation.

Atrial pacing (AAIR or DDDR) is the preferred pacing mode for patients with SND ( Table 2 ).

Table 2. – Recommendations for choosing a pacing mode in sinus node disease.

  Class of recommendation Level of evidence
1. AAI(R) in patients with normal AV conduction 2. DDD(R) in patients with advanced AV block I A
1. AAI(R) with automatic switch to DDD(R) in patients with intermittent advanced AV block I B
1. DDD with an algorithm that reduces ventricular pacing in patients with preserved AV conduction IIa B
1. VVI(R) in older people without retrograde VA conduction IIb B
1. VVI(R) in patients who may need only short periods of ventricular pacing or patients with significant comorbidities that affect survival or clinical events IIb C
1. VVI(R) in patients with retrograde VA conduction; VDD(R) 2. AAI(R) in patients with advanced AV block III C

AV: atrioventricular; VA: ventriculoatrial.

Most patients with SND present with preserved AV conduction. Conversely, RV pacing has been known to be associated with negative physiological consequences resulting from ventricular dyssynchrony (VD), such as LV remodeling, reduced LV ejection fraction (LVEF), and functional mitral regurgitation. 37 Therefore, algorithms that reduce unnecessary ventricular pacing, such as AV hysteresis and automatic switch from DDD to AAI mode, should be programmed in patients without associated AV block. 38

Algorithms that suppress AF occurrence, such as continuous atrial pacing (atrial overpacing) or atrial pacing induced by intrinsic atrial activity sensing, either alone or combined, have no proven benefits. Similarly, alternative pacing sites – such as Bachmann bundle pacing and dual-site or multisite atrial pacing – have failed to show consistent effects. 39

Current devices have one or more mechanisms of frequency response sensors usually based on body movement (piezoelectric crystals or accelerometers) or ventilation/minute volume. The main purpose of the sensors is to physiologically increase heart rate and not necessarily change clinical outcomes. Although devices cannot accurately assess atrial chronotropic response, they can provide indicators of atrial disease progression through rate and arrhythmia histograms, atrial pacing percentage, and patient daily activity. Such data may be useful for sensor programming. There is no evidence that the use of combined sensors (eg, accelerometer and minute ventilation) improves quality of life.

2.2. Atrioventricular and Intraventricular Blocks

2.2.1. Atrioventricular Blocks

An electrical stimulus originating in the sinus node is propagated through the myocardium by the conduction system. A delay or failure in propagating the stimulus between atria and ventricles characterizes an AV block. This change in stimulus propagation may be a pathological change or a functional phenomenon resulting from physiological refractoriness (an intrinsic characteristic of conduction system cells). 40

From an ECG perspective, AV blocks are classified into first-degree, second-degree (Mobitz I, Mobitz II, 2:1, advanced), and third-degree.

First-degree AV block is defined by a delay in stimulus conduction from the atrium to the ventricle with a PR interval > 200ms.

In type I second-degree AV block (Mobitz I), the block occurs after progressive prolongation of the PR interval (Wenckebach phenomenon), with a blocked P wave at the end. In type II second-degree AV block (Mobitz II), the P wave is suddenly blocked, ie, there is no progressive prolongation of the PR interval. When AV conduction occurs with a 2:1 ratio, second-degree block usually cannot be classified unequivocally as type I or type II without the support of autonomic maneuvers, drugs, or even invasive EPS. Advanced AV block refers to the blocking of two or more consecutive P waves with some conducted beats, which indicates some preservation of AV conduction. In AF with prolonged pauses (> 5s), advanced second-degree AV block should be considered.

Finally, third-degree AV block (complete AV block) is defined as the absence of AV conduction (complete dissociation between P waves and QRS complexes with atrial rate > ventricular rate). 41

There are numerous congenital and, most often, acquired conditions that may affect AV conduction. Degenerative causes are commonly observed in clinical practice and are associated with aging, hypertension, and diabetes mellitus. The most common infectious causes in Brazil are chronic Chagas myocarditis and, to a lesser extent, viral acute myocarditis, which may cause definitive intermittent acute blocks.

AV blocks resulting from inferior-wall ischemia or acute myocardial infarction as well as autonomic nervous system-mediated blocks may be reversible.

Iatrogenic causes, especially from pharmacological action, should also be considered depending on clinical status.

Anatomically, AV blocks are defined as AV nodal, intra-Hisian, or infra-Hisian according to block site. AV nodal block is associated with slower progression, a faster and more reliable junctional escape, and improved response to autonomic manipulation with the administration of atropine, isoproterenol, and epinephrine. In contrast, intra-Hisian or infra-Hisian AV blocks progress more rapidly and are associated with a slower and more unpredictable ventricular escape, a wider QRS with poor response to adrenergic activity, and vagal block. High-degree blocks (advanced or third-degree) have a higher risk of low output and severe asystole and require urgent therapy. In the setting of AF, complete AV block is characterized by a low ventricular response (< 50bpm) and a regular RR interval.

First-degree AV block is generally asymptomatic but may result in fatigue or exertion intolerance if the PR interval is long enough to allow loss of AV synchrony. This change is called pseudo-PM syndrome and may occur when the PR interval is > 300ms. 42 Likewise, type I second-degree AV block is often asymptomatic and affects healthy, active patients with or without a history of heart disease, particularly during parasympathetic activity. However, if it occurs frequently or during exercise, it may cause exertion intolerance or dizziness. 43

In 61% of patients with syncope and underlying bundle branch block or bifascicular block, significant and clinically relevant conduction abnormalities in the His-Purkinje system may be detected on EPS. 44

In patients with AV block, clinical evaluation may help identify transient or reversible causes, and treatment or resolution may make permanent artificial pacing unnecessary.

In congenital complete AV block, permanent PM implantation must be indicated when symptoms are present or when the child has a resting heart rate < 55bpm or < 70bpm, if associated with structural heart disease. 45 In asymptomatic patients, regular follow-up with additional tests that assess mean heart rate, QT interval, pauses, ventricular arrhythmia, intraventricular (IV) conduction disorders (IVCD), presence or emergence of structural heart disease, low cognitive development and low weight and height gain, and exercise intolerance should be conducted to evaluate the need for permanent PM implantation. 46 Prophylactic PM or ICD implantation is also recommended in some asymptomatic patients with neuromuscular dysfunctions or other genetic disorders.

In patients with bradycardia indicative of PM implantation and LV dysfunction, ICD implantation should be considered (see item 4).

PM implantation is not indicated for asymptomatic patients with permanent AF and low resting heart rate who have an appropriate chronotropic response while awake, regardless of pause occurrence and duration. Conversely, for symptomatic patients with prolonged pauses (> 3s) that may have resulted from an infranodal block, PM implantation is indicated. 30

Prophylactic PM implantation may be necessary especially in patients with HF or coronary artery disease (CAD) requiring chronic use of betablockers. 47

To determine the best type of device and pacing mode for patients with AV block, the following clinical variables should be considered: expected ventricular pacing percentage and LV systolic function (ie, LVEF).

Studies comparing pacing with preservation of sequential AV activation vs single-chamber ventricular pacing in patients with AV block (eg, Pacemaker Selection in the Elderly [PASE], Mode Selection Trial [MOST], The Canadian Trial of Physiological Pacing [CTOPP]) showed no significant reduction in mortality or stroke rates. 34 , 35 , 48

Conversely, a systematic review revealed that dual-chamber pacing is the preferred choice for reducing AF incidence and PM syndrome prevalence compared with single-chamber ventricular pacing (VVI). 49 However, the United Kingdom Pacing and Cardiovascular Events (UKPACE) trial (patients aged ≥75 years) found no benefits of AV pacing on mortality or AF and HF incidence compared with ventricular pacing. Similar stroke rates and higher rates of complications related to the surgical procedure were also observed in patients undergoing dual-chamber PM implantation (7.8% vs 3.5%; p < 0.001). 50 Therefore, the indication for single-chamber PM (VVI) is reasonable in frail patients or those with significant comorbidities, older age, a very sedentary lifestyle, or reduced daily need for pacing.

In patients with retrograde VA conduction, ventricular pacing may trigger symptoms of PM syndrome. In these cases, dual-chamber pacing should be preferred to avoid AV dyssynchrony. 51

The deleterious effects of chronic RV pacing have been demonstrated in several studies, although only a minority (5% to 9%) of patients with chronic RV pacing develop severe ventricular dysfunction with symptoms of HF. 52 - 54 Patients with LV dysfunction and indication for PM due to AV block were evaluated in the Conventional versus Multisite Pacing for Bradyarrhythmia Therapy (COMBAT) 55 (LVEF < 35%) and Biventricular versus Right Ventricular Pacing in Heart Failure Patients with Atrioventricular Block (BLOCK-HF) 56 (LVEF ≤ 50%) studies. The studies compared CRT vs conventional RV pacing and demonstrated clinical improvement (New York Heart Association [NYHA] functional class) and reverse LV remodeling (increased LVEF) with CRT, with a significant reduction in primary outcomes.

In patients with AF and LV dysfunction undergoing AV node ablation for heart rate control, CRT or conduction system (His bundle or LBB) pacing seems to be associated with better outcomes compared with conventional RV pacing. 30

The recommendations for PM implantation in AV block are summarized in Table 3 .

Table 3. – Recommendations for permanent pacemaker implantation in atrioventricular block.
  Class of recommendation Level of evidence
Acquired second-degree (Mobitz II), advanced, or third-degree AV block not associated with a reversible or physiological cause, regardless of presence of symptoms Second-degree (Mobitz II), advanced, or third-degree AV block or alternating bundle branch block (even if asymptomatic) persisting for at least 72 hours after AMI Advanced AV block or alternating bundle branch block after TAVI persisting for at least 24-48 hours Advanced AV block after AMI persisting for at least 5 days Symptomatic advanced AV block persisting for at least 5 days after heart valve surgery, revascularization, or AF surgery Symptomatic, congenital third-degree AV block Asymptomatic, congenital third-degree AV block associated with risk factors (pause > 3 times the baseline RR cycle, wide QRS, prolonged QTc, complex ventricular arrhythmia, mean heart rate < 50bpm, ventricular dysfunction) I B
Symptomatic, acquired second-degree AV block (Mobitz I) not associated with a reversible cause or nonessential drug Asymptomatic second-degree (Mobitz II), advanced, or third-degree AV block persisting for at least 5 days after heart valve surgery, revascularization, or AF surgery Permanent AF with a poor ventricular response with symptoms attributed to bradycardia I C
Congenital third-degree AV block in asymptomatic adults (age > 18 years) IIa B
In symptomatic patients clearly affected by significant first-degree AV block (pseudo-PM syndrome) IIa C

AF: atrial fibrillation; AMI: acute myocardial infarction; AV: atrioventricular; PM: pacemaker; QTc: corrected QT interval; TAVI: transcatheter aortic valve implantation.

2.2.2. Intraventricular Blocks with 1:1 Atrioventricular Conduction

QRS complex abnormalities (eg, fascicular blocks and bundle branch blocks) are caused by a conduction delay or a block within one or more branches of the His-Purkinje system.

Conduction delay or right bundle branch block (RBBB) associated with a block in one fascicle of the LBB is named bifascicular block (the same terminology is used for LBB block [LBBB]). Clinical conditions that may result in IV block include genetic/hereditary, inflammatory, infectious, infiltrative, metabolic, ischemic, and degenerative causes.

The presence of IV block alone is rarely associated with symptoms but may be a marker of structural heart disease; the presence or development of LBBB may result in cardiac dyssynchrony and progressive LV dysfunction. Some studies have demonstrated a correlation between LBBB and CAD and HF. 57 , 58 The prevalence of progression of LBBB or bifascicular block to advanced AV block is low, approximately 1% per year. 59 LBBB is usually associated with higher mortality compared with other IVCDs. 60 In some neuromuscular diseases, PM implantation is recommended in patients with IV block because of the high incidence of complete AV block and sudden cardiac death.

Although an EPS may identify high-risk conduction disorders, this study is characterized by a variable degree of sensitivity and some degree of risks. In patients with syncope, the presence of IV block is a predictor of electrophysiological abnormalities. 61

Regardless of symptoms, alternating bundle branch blocks (when QRS complex morphology spontaneously alternates between LBBB and RBBB) are also indicative of PM implantation because their presence suggests infranodal conduction system disease with high risk for severe complete AV block.

The recommendations for PM implantation in IV blocks are summarized in Table 4 .

Table 4. – Recommendations for permanent pacemaker implantation in intraventricular block.
  Class of recommendation Level of evidence
Permanent PM implantation is recommended in patients with syncope and bundle branch block with H-V interval ≥ 70ms or infranodal block on EPS without documented hemodynamically unstable VT Permanent PM implantation is recommended in patients with alternating bundle branch block regardless of symptoms New, persistent LBBB with QRS > 150ms for more than 48 hours after TAVI if PRi > 240ms I C
Patients with bifascicular block and unexplained syncope who did not undergo EPS (older people, frail patients, and recurrent syncope) IIb B
New, persistent LBBB with QRS > 150ms for more than 48 hours after TAVI if normal PRi IIb C
Asymptomatic patients with IV conduction disorder alone and 1:1 AV conduction in the absence of other indications for PM implantation III B

EPS: electrophysiology study; LBBB: left bundle branch block; PM: pacemaker; PRi: PR interval; TAVI: transcatheter aortic valve implantation; VT: ventricular tachycardia.

2.3. Carotid Sinus Syndrome

Carotid sinus syndrome (CSS) is characterized by a history of syncope associated with an exaggerated reflex response to mechanical stimulation of the carotid sinus occurring either spontaneously or with carotid sinus massage (CSM). 62 Syncope results from significant bradycardia and/or hypertension triggered by head movements or situations that cause involuntary compression of the neck and carotid sinus, although this correlation is not clinically evident in many patients. 63

CSS is diagnosed by a pause > 3s (sinus node arrest or AV block) or a fall in systolic blood pressure (SBP) ≥ 50mmHg in the absence of conductor system depressants combined with reproduction of syncope during right- and left-sided CSM (5 to 10 s) in patients aged > 40 years. The patient should be placed in the supine position and tilted (tilt table test). 64 Reproducing syncope during CSM increases test specificity, whereas performing CSM in the supine position increases test sensibility. 63

Reflex responses in CSS can be classified according to their hemodynamic profile into three types: cardioinhibitory (ventricular pause > 3s), mixed (ventricular pause > 3s associated with a fall in SBP ≥ 50mmHg), or vasodepressor (a fall in SBP ≥ 50mmHg alone). The incidence of CSS increases with age (< 50 years: ~0%, 50 to 59 years: 2,4%, 60 to 69 years: 9,1%, 70 to 79 years: 20%, and > 80 years: 40%). 63

The evidence supporting permanent PM implantation in patients with CSS is based on small, controlled studies and retrospective observational studies. 65 - 68

A literature review of 12 studies including 601 patients treated with PM and 305 controls found lower rates of syncopal recurrence in treated patients (0% to 20%) compared with controls (20% to 60%), although patient selection, patient position during CSM (horizontal or tilted), follow-up duration, and pacing mode were heterogeneous among studies. 68 A meta-analysis of three controlled studies with a mean follow-up duration of 3.3 years showed a significant reduction (76%) in syncopal recurrence in patients treated with PM compared with controls (9% and 38%, respectively; relative risk [RR]: 0.24; 95% CI 0.12-0.48). 68

The Syncope And Falls in the Elderly – Pacing And Carotid Sinus Evaluation (SAFE PACE) study 69 evaluated 175 older patients with recurrent unexplained falls, no reported loss of consciousness, and cardioinhibitory response during CSM and suggested that a diagnosis of CSS should be suspected in such cases. The group that was randomly assigned to permanent PM implantation showed a significantly reduced rate of events (syncope: 53%; falls: 70%; physical trauma: 70%) during follow-up.

A study in which the diagnosis of CSS also included spontaneous pauses recorded by an implantable loop recorder found a 98% reduction in syncope burden after PM implantation (1.68 episodes per patient/year [95% CI 1.66-1.70] vs 0.04 episodes per patient/year [95% CI 0.038-0.042]). 70 When initial investigation with CSM and head-up tilt table test is negative, unexplained syncope and vasovagal syncope in patients aged > 40 years may be diagnosed with the help of an implantable loop recorder. 71 The recommendations for PM implantation in CSS are summarized in Table 5 .

Table 5. – Recommendations for permanent pacemaker implantation in carotid sinus syndrome and vasovagal syncope.

  Class of recommendation Level of evidence
Patients aged > 40 years with recurrent syncope and documented symptomatic spontaneous pause > 3s (sinus pause and/or AV block) or asymptomatic pause > 6 s I A
Patients aged > 40 years with recurrent syncope and cardioinhibitory (pause > 3s, sinus pause, and/or AV block) or mixed (pause > 3s + hypotension) response to CSM in the absence of conduction system depressants IIa B
Patients aged > 40 years with recurrent syncope and a symptomatic pause > 3s (sinus pause and/or AV block) induced by the tilt table test Unexplained recurrent falls in patients aged > 40 years with cardioinhibitory response (pause > 3s, sinus pause, and/or AV block) to CSM and no prodromes IIb B
Patients with syncope and no documented cardioinhibitory response III B
Asymptomatic patients with cardioinhibitory response to CSM III C

AV: atrioventricular; CSM: carotid sinus massage.

Importantly, asymptomatic older patients may present with carotid sinus hypersensitivity but no characterization of CSS and no indication for PM implantation. Conversely, patients with recurrent unexplained falls and cardioinhibitory response to carotid sinus compression (carotid sinus hypersensitivity) may benefit from a permanent PM. 69

Regarding pacing mode in CSS, a study evaluating acute hemodynamic changes in patients with CSS undergoing CSM demonstrated that VVI (single-chamber) pacing was associated with an increased fall in SBP and an increased rate of persisting symptoms compared with DVI (dual-chamber) pacing. 72 Subsequent studies comparing long-term single-chamber vs dual-chamber pacing showed a trend towards lower rates of recurrent syncope and presyncope in patients with dual-chamber pacing 73 - 76 ( Table 6 ).

Table 6. – Recommendations for pacing mode in carotid sinus syndrome and vasovagal syncope.

  Class of recommendation Level of evidence
Dual-chamber PM (DDD, O/R) in CSS and vasovagal syncope IIa B

CSS: carotid sinus syndrome; PM: pacemaker.

2.4. Vasovagal Syndrome

Vasovagal syncope is characterized by a history of loss of consciousness associated with an exaggerated neurally-mediated reflex that progresses with a sudden reduction in cerebral blood flow secondary to vasodilation and/or reduced heart rate. In most cases, syncope is secondary to a sudden and significant fall in blood pressure followed by varying degrees of bradycardia and generally preceded by prodromal manifestations such as malaise, sweating, feeling hot, pallor, dizziness, and then fatigue. Vasovagal syncope is often triggered by significant emotional stress, fear, or pain and is the main cause of syncope, especially in young people. Predisposing factors include prolonged orthostatism, closed or hot environments, venipuncture, and physical trauma, among others.

Based on changes in blood pressure and heart rate, vasovagal response can be classified into three types: type 1, or mixed response (a significant fall in blood pressure followed by a slight reduction in heat rate); type 2, or cardioinhibitory response (a significant reduction in heart rate < 40bpm or asystole > 3s); and type 3, or vasodepressor response (a significant fall in blood pressure with no significant reduction in heart rate). 77

Despite sometimes being related to physical trauma and inability to perform activities that pose personal or collective risks, vasovagal syncope has a favorable long-term prognosis. Treatment is most often nonpharmacological, including guidance and changes in lifestyle. However, approximately 14% of patients present with severe forms of vasovagal syncope and require additional treatment (very frequent syncopal recurrence associated with impaired quality of life, short or absent prodromes, and increased risk of trauma, or with high-risk occupations such as professional driver, machine operator, airplane pilot, competitive athlete etc.). Patient age is the most important factor for choosing the best therapy.

There are few evidence-based therapeutic options for vasovagal syncope. Artificial cardiac pacing may be effective in patients with vasovagal syncope and dominant cardioinhibitory reflex; thus, clinical investigation should be focused on documenting the correlation between syncope and bradycardia.

The effectiveness of permanent PM insertion has been assessed in some randomized studies. The Vasovagal Pacemaker Study I (VPS-I) evaluated patients with recurrent syncope (≥6 syncopal spells) and a positive tilt table test (bradycardia < 60bpm or pause > 1s) who were randomly assigned to receive a PM (DDD mode with rate-drop response function) or clinical treatment. 78 Syncopal recurrence at 12 months was 22% (6/27) in the PM group vs 70% (19/27) in the clinical treatment group (RR 85.4%; 95% CI 59.7%-94.7%; 2p = 0.00002). The Syncope Diagnosis and Treatment (SYDIT) study randomly assigned patients with recurrent syncope (≥3 syncopal spells) and a positive tilt table test (bradycardia) to receive a PM (DDD mode with rate-drop response function) or clinical treatment (atenolol 100 mg/day). 79 Syncopal recurrence after a median follow-up of 135 days was 4.3% (2/46) in the PM group vs 25.5% (12/47) in the atenolol group (OR 0.133; 95% CI 0.028-0.632; p = 0.004). In the Vasovagal Syncope International Study (VASIS), patients were randomly assigned to receive a PM (DDI mode with rate hysteresis) or no treatment. 80 Syncopal recurrence after a mean follow-up of 3.7 years was 5% (1/19) in the MP group vs 61% (14/23) in the control group (p = 0.0006). The Vasovagal Pacemaker Study II (VPS-II) included patients with recurrent syncope (6 syncopal spells or over), although significant bradycardia during the tilt table test was not an inclusion criterion. 81 After PM implantation, patients were randomized to dual-chamber pacing (DDD) or sensing without pacing (ODO). Syncopal recurrence rates were 33% (16/48) in the DDD group vs 42% (22/52) in the ODO group, with no significant reduction in the risk of syncope (RR 30%; 95% CI -33% to 63%; p = 0.14). In the Third International Study on Syncope of Uncertain Etiology (ISSUE-3), the most important double-blind randomized trial published to date, patients aged > 40 years with spontaneous syncope documented by an implantable loop recorder associated with asystole > 3s or asystole > 6s without syncope were randomly assigned to dual-chamber pacing with rate-drop response function (PM ON) or sensing without pacing (PM OFF). 82 During a mean follow-up of 2 years, there was a significant reduction (57%) in the risk of syncopal recurrence (25% in PM ON arm vs 57% in the PM OFF arm; p = 0.039). All studies used dual-chamber pacing with rate-drop response function.

A small retrospective study compared conventional dual-chamber pacing with closed-loop stimulation (CLS) pacing and reported lower rates of syncopal recurrence in the latter. 83

Recommendations for permanent PM implantation in vasovagal syncope are summarized in Table 5 . Recommendations for choosing a pacing mode are shown in Table 6 .

2.5. Hypertrophic Cardiomyopathy

Hypertrophic cardiomyopathy (HCM) is a common genetic cardiovascular disease characterized by LV hypertrophy in the absence of other cardiac abnormalities or systemic disorders capable of producing a degree of hypertrophy equivalent to that found in patients with this condition. 84 In hypertrophic obstructive cardiomyopathy (HOCM), there is a pressure gradient in the LV outflow tract (LVOT) – larger gradients are associated with more severe symptoms and increased mortality. 85

For patients with symptoms caused by LVOT obstruction, therapeutic options include negative inotropic drugs, septal myotomy-myectomy operation, alcohol septal ablation, and heart transplantation. 86 In these patients, RV apical pacing changes the pattern of ventricular contraction and generates regional dyssynchrony. This results in late basal septal activation and reduced LV contractility, which reduces both the systolic anterior motion of the mitral valve and the LVOT pressure gradient. 87 , 88 However, VD caused by activation with a wide QRS complex alone reduces LV contractility and may lead to a reduced outflow tract gradient. Thus, in this case, the benefit is related to an adverse effect of the PM.

Reduced LVOT gradients with ventricular pacing have been demonstrated in three small randomized, controlled studies and several observational studies. However, improvement of symptoms and quality of life varied among studies. 89 - 93 A subgroup analysis suggests that patients aged > 65 years are more likely to benefit from dual-chamber pacing. 94

A Cochrane systematic review has concluded that the benefits of ventricular pacing in HCM are based on gradient measurements and that evidence regarding relevant clinical endpoints are lacking. 95 In addition, gradient reduction is generally smaller in ventricular pacing compared with septal myectomy or ablation. Thus, the indication of dual-chamber pacing solely to reduce the LVOT gradient is restricted to very specific conditions: patients aged > 65 years with moderate hypertrophy, defined symptoms due to LVOT obstruction, and no indication for ICD. 93

Overall, septal myectomy or ablation should be considered a first-line therapy in patients with HOCM and symptoms refractory to pharmacological treatment. Very severe cases may require heart transplantation.

In patients undergoing PM implantation for LVOT gradient reduction, a short AV interval must be programmed (AV interval in VAT = 100 ± 30 ms) to obtain maximal RV preexcitation without compromising diastolic ventricular filling. 96 In addition, maximal atrial pacing rate should be higher than the maximal rate exhibited by the patient during the stress test. Patients with HCM and very poor tolerance to an elevated heart rate are usually on beta-blockers, which results in a lower maximal rate during exertion; conversely, these patients are more susceptible to developing AF. Thus, automatic mode switching to DDI(R) mode should be used to avoid high-rate ventricular pacing in AF. If the atrial lead is inefficient in detecting AF, the maximal rate should be lowered.

Finally, a significant number of patients with HCM receive an ICD for preventing sudden death. For these patients, a dual-chamber DDD pacing device with a short AV interval may reduce the LVOT gradient and prevent or delay the need for additional interventions.

The recommendations for permanent PM implantation in patients with HCM are described in Table 7 .

Table 7. – Recommendations for permanent pacemaker implantation in hypertrophic cardiomyopathy.

  Class of recommendation Level of evidence
Patients in sinus rhythm with advanced, second-degree (Mobitz II), or third-degree AV block, spontaneous or after alcohol septal ablation or surgical myectomy (dual-chamber pacing is recommended) I B
Patients with HOCM and ICD indication should be considered for a dual-chamber pacing device IIa C
In patients with LVOT obstruction (gradient ≥50 mm Hg), dual-chamber pacing with a short AV interval may be considered in those who are refractory to clinical treatment, have no ICD indication, and cannot undergo alcohol septal ablation, surgical myectomy, or heart transplantation IIb C

AV: atrioventricular; HOCM: hypertrophic obstructive cardiomyopathy; ICD: implantable cardioverter-defibrillator; LVOT: left ventricle outflow tract.

2.6. Neuromuscular Diseases

Some neuromuscular conditions may cause progressive and insidious disease of the cardiac conduction system, such as Duchenne muscular dystrophy, facioscapulohumeral muscular dystrophy, X-linked dystrophy, myasthenia gravis, myotonic dystrophy, and Friedreich ataxia. 97

Most manifestations are related to infranodal conduction disorders, resulting in fascicular blocks and third-degree AV blocks. Such findings are typical of Kearns-Sayre syndrome (progressive external ophthalmoplegia with pigmentary retinopathy), Guillain-Barré syndrome, myotonic dystrophy, Becker muscular dystrophy, and facioscapulohumeral muscular dystrophy.

Myotonic dystrophy and Kearns-Sayre syndrome are associated with a high incidence of conduction system disease, which often progresses rapidly and cannot be predicted by ECG or intracavitary recordings. The disease almost always affects the His-Purkinje system and can lead to Stokes-Adams attacks or sudden death, except when prevented by PM implantation.

In a study of 49 patients with myotonic dystrophy (mean age: 46 years, HV interval ≥ 70ms), high-degree AV blocks were recorded in 47% of patients after PM implantation, despite no documented evidence of bradycardia at baseline. 98 The authors concluded that permanent PM implantation should be considered in patients with myotonic dystrophy and prolonged HV interval (≥ 70ms) even if they are asymptomatic or if bradycardia is found on ECG.

In patients with neuromuscular diseases, waiting for a documented AV block may result in significant risk of sudden death or syncope. Thus, permanent PM implantation should be considered early in the course of neuromuscular disease upon the presence of conduction abnormalities even if the patient is asymptomatic ( Table 8 ).

Table 8. – Recommendations for permanent pacemaker implantation in neuromuscular diseases.

  Class of recommendation Level of evidence
Symptomatic or asymptomatic patients with second-degree (type II or advanced) or third-degree AV block I B
Asymptomatic patients with first-degree AV block (PR interval > 240ms) or wide QRS complex (QRS > 120ms) (in myotonic dystrophy) IIb C

AV: atrioventricular.

ECG abnormalities such as nonsinus rhythm, QRS > 120ms, PRi > 240ms, second- or third-degree AV block, and atrial tachyarrhythmias were independent predictors of sudden death in patients with myotonic dystrophy type I. 99

2.7 Obstructive Sleep Apnea Syndrome

Bradyarrhythmias, such as sinus bradycardia, sinus pauses, advanced or type I second-degree AV block, and junctional escape rhythm, are common during sleep, especially in young healthy people who are physically fit. Direct and indirect data have shown an association with vagal hypertonia. In nearly all cases, these findings are physiological and require no specific treatment. However, cardiac arrhythmias have been observed in obstructive sleep apnea (OSA) syndrome. 100

The most significant cases are related to Pickwick syndrome, obesity, hypertension, metabolic syndrome, anatomic and/or functional airway obstruction (eg, constitutional, macroglossia, soft palate hypertrophy, and tonsillar and adenoid hypertrophy), chronic lung diseases, and neurological diseases, among others.

During apnea, airway obstruction leads to oxygen desaturation that may result in severe hypoxemia with consequent development of atrial and ventricular brady- and tachyarrhythmias. 101

Primary treatment is intended for apnea correction and weight loss. Continuous positive airway pressure devices for respiratory support during sleep may be of great importance, as their use has resolved bradyarrhythmias in a large number of cases. 102 Thus, there is no primary indication for PM implantation in OSA-related bradyarrhythmias.

In clinical practice, OSA is commonly observed in a setting characterized by vagal hypertonia, nocturnal bradyarrhythmia, and AF commonly triggered by bradycardia (bradycardia-tachycardia syndrome). 102 When airway obstruction correction is not sufficient to improve the status, radiofrequency ablation may be indicated (cardioneuroablation [CNA] and/or AF ablation). In exceptional cases, MP implantation may facilitate the treatment of AF (beta-blockers or other antiarrhythmics).

Increased vagal tone can be well characterized by temporal and spectral analyses of RR variability on 24-hour Holter monitoring with a comparison of wake and sleep periods. RR variability and the number of pauses > 2.5 s pre- and 1-year post-CNA were compared in a study including 18 patients with a history of OSA and tachycardia-bradycardia syndromes and/or vagal hypertonia. There was a significant reduction in RR variability (pre-CNA standard deviation of the NN intervals 131.2 ± 38 ms vs 91.9 ± 37 ms 11 months after CNA, p = 0.0001). Furthermore, the number of pauses significantly decreased from 6.5 ± 9.4 pre-CNA to 1.1 ± 3 within 11 months of CNA (p = 0.03). No patient received a PM. 103 Thus, PM implantation is currently reserved for patients with a compromised conduction system ( Table 9 ).

Table 9. – Recommendations for permanent pacemaker implantation in obstructive sleep apnea syndrome.

  Class of recommendation Level of evidence
Patients with nocturnal bradyarrhythmias (asymptomatic while awake), OSA syndrome, and no significant heart disease who received no specific treatment III C

OSA: obstructive sleep apnea.

2.8. Congenital Long QT Syndrome

Congenital long QT syndrome (LQTS) is a channelopathy caused by abnormal cardiac repolarization and features a prolonged QT interval, ventricular arrhythmias (ventricular extrasystole, polymorphic ventricular tachycardia, torsade de pointes), and a history of syncope and/or sudden death. There is generally a family history of LQTS in close relatives, and the most common pattern of inheritance is autosomal dominant (Romano-Ward syndrome), although the condition may also be autosomal recessive (very rare) with associated deafness. These two types account for 90% of cases; however, there are currently 14 known different types of congenital LQTS. 104

Some patients with LQST do not present with spontaneous clinical manifestations, although certain conditions, such as medication use, physical stress, and electrolyte changes, may be triggers.

The QT interval (QTi) can be measured in any lead, although leads II and V5are more commonly used. 105 The QTi is measured from the beginning of the QRS complex to the end of the T wave, excluding the U wave. Given that QTi physiology changes inversely to heart rate, the original value should be corrected for heart rate, most commonly with the Bazett formula. 106 In this formula, the measured QTi is divided by the square root of the preceding RR interval (the units are measured in seconds), resulting in the corrected QT (QTc) interval. The normal limits are 450ms and 460ms for men and women, respectively. A QTc interval > 480ms 4 minutes after the stress test is highly suggestive of LQST. Approximately 20% of cases with a positive genotype have a normal QTi.

In different types of congenital LQST, a delay in cellular repolarization occurs either due to a reduction in potassium channel function or an increase in sodium channel function (delayed channel inactivation) and manifests as an increased QTi. Electrophysiological abnormalities are apparently heterogeneous and become significantly prominent in the presence of conditions such as autonomic stimulation, physical and mental stress, electrolyte changes, drug action, and ischemia, among others, resulting in electrical instability, extrasystole, polymorphic tachycardia, torsade de pointes, ventricular fibrillation, and sudden death.

Whenever possible, the type of LQTS (specifically types I, II, and III) should be defined according to clinical and ECG manifestations, as there is a recommended treatment for each type.

Types I and II are the most common and originate from a decrease in potassium channel function, whereas type III is caused by an increase in sodium channel function. 107 Congenital LQTS manifestations typically appear during childhood or adolescence and usually affect male patients (adolescence) earlier than female patients (adulthood). Syncope is the most common manifestation and usually occurs between 5 and 15 years of age. A family history of sudden death is a strong predictor of mortality. Overall, the longer the QTc interval, the higher the risk of sudden death.

In congenital LQTS type I, arrhythmias are usually triggered by physical exertion, especially swimming. In type II, they are commonly induced by mental stress caused, for example, by loud noises, especially during rest or sleep. As for type III, arrhythmias typically occur during rest or sleep without a clear relationship with stress.

It is absolutely essential that patients avoid electrolyte disturbances such as hypokalemia and the use of drugs that may trigger fatal arrhythmias. Several websites provide a list of drugs that may potentially increase the QTc interval (http://www.crediblemeds.org) and should always be consulted before taking any medication.

All patients (symptomatic, asymptomatic, and “silent” carriers) should significantly reduce the level of physical activity. Competitive sports are contraindicated. Certain type-related triggers, such as strenuous swimming in LQTS type I and very loud noises in LQTS type II, should be avoided. 108 Noncompetitive recreational sports may be cautiously considered in patients with LQTS type III if there is easy access to an automated external defibrillator. Drugs and substances that prolong repolarization, such as potassium channel blockers, which can induce torsade de pointes even in asymptomatic cases, are absolutely contraindicated. Sympathomimetics should also be avoided. Patients are advised to carry a list of prohibited drugs.

Pharmacological treatment of LQTS types I and II is based on beta-blockers, and propranolol and nadolol are the most effective drugs. Metoprolol appears to be less effective and should be avoided. 109

Retrospective studies have demonstrated the unquestionable benefits of beta-blockers and denervation surgery (removal of the left stellate ganglion), with a mortality of 9% in the treated group vs 60% in the untreated group.

The QTc interval can be experimentally reduced with potassium-channel activators such as nicorandil in LQTS type I or spironolactone combined with oral potassium in LQTS type II. LQTS type III may benefit from sodium-channel blockers such as mexiletine and flecainide, which can shorten the QTc interval; however, flecainide may induce a Brugada phenotype. There are reports of successful treatment of electrical storms in LQTS type III with mexiletine, which is recommended by some professionals when the QTc interval is very long.

There is currently no primary indication for PM implantation for preventing sudden death in LQTS, given the greater safety provided by ICD. 110 However, a PM may be recommended for patients with AV block or ventricular arrhythmias triggered or aggravated by bradycardia or pauses, provided there is no history of recovered sudden death or high-risk signs such as congenital deafness, syncope, documented complex ventricular arrhythmias, family history of sudden death, female sex, and QTc > 0.60s. PM insertion combined with beta-blocker therapy is occasionally indicated to avoid drug-induced bradycardia. A cardiac pacing rate above the spontaneous sinus rate can reflexively inhibit sympathetic action and be quite helpful in controlling arrhythmic storms. A PM should only stimulate the atrium (avoid VD) and can be programmed to a higher rate (eg, 80 ppm), which shortens QTc duration. Furthermore, activation of the frequency response sensor can ensure chronotropic adaptation, which is impaired by beta-blockers ( Table 10 ).

Table 10. – Recommendations for permanent pacemaker implantation in congenital long QT syndrome.

  Class of recommendation Level of evidence
Low-risk patients (no high-risk conditions*) with bradyarrhythmia (sinus or AV block) aggravated or not by beta-blockers, especially in LQTS type III IIb C

*High-risk conditions in congenital LQTS: congenital deafness, syncope, documented complex ventricular arrhythmias, family history of sudden death, female sex, corrected QT interval >0.60 s. AV: atrioventricular; LQTS: long QT syndrome.

Denervation surgery is an alternative in patients with recurrent syncope despite the use of regular full-dose beta-blockers or when optimal drug therapy (ODT) cannot be prescribed (eg, asthma).

2.9. Cardiac Transplantation

The rates of permanent PM implantation after cardiac transplantation vary between 2% and 24% and have significantly decreased in patients undergoing bicaval anastomosis vs those undergoing biatrial anastomosis. 111 - 113

SND is the most commonly reported disorder in these patients, and possible causes include surgical trauma, damage to the sinoatrial artery due to trauma or ischemia, prolonged ischemic time of the donor heart, cardiac denervation, and baseline characteristics of the donor heart. 114

Approximately 10% of patients requiring a PM have an AV conduction disorder (especially second- or third-degree AV blocks), which are probably due to inadequate graft lpreservation. 115

Bradycardia often occurs in the early postoperative period following cardiac transplantation and affects approximately two thirds of patients, although it tends to remit spontaneously. However, if bradycardia persists for a few weeks and the patient develops symptoms, PM implantation may be necessary.

Studies have shown that several patients with SND and bradycardia implanted with a PM become device-independent after 3 months. However, patients with early AV block often require long-term cardiac pacing 116 , 117 ( Table 11 ).

Table 11. – Recommendations for permanent pacemaker implantation after cardiac transplantation.

  Class of recommendation Level of evidence
A pacemaker should be considered in symptomatic patients with bradyarrhythmias or chronotropic incompetence that are not likely to resolve spontaneously or, even if transient, may persist for months IIa C

2.10. Choosing the Type of Pacemaker and Pacing Mode

Randomized clinical trials have not reported any impact of atrial or AV pacing (AAI/DDD) on survival compared with ventricular pacing (VVI) alone; however, AF occurrence, syncopal incidence, and PM syndrome rates were reduced. 32 - 35 A systematic review reported a significant reduction in stroke (HR: 0.81) and AF (HR: 0.80) incidence rates with sequential AV pacing compared with ventricular pacing. 36

In patients with persistent bradycardia, dual-chamber pacing is the preferred choice. The Danish Multicenter Randomized Trial on Single Lead Atrial Pacing vs Dual Chamber Pacing in Sick Sinus Syndrome (DANPACE) found that AAIR pacing was associated with a higher incidence of paroxysmal AF (HR: 1.24) and a two-fold increase in reoperations compared with DDDR pacing. 118 Ventricular pacing should be avoided in patients in sinus rhythm as it may cause AF and worsening HF 119 ( Figure 1 ). Importantly, programming an excessively long AV interval to avoid ventricular pacing in patients with spontaneous AV conduction may be hemodynamically harmful in patients with a marked first-degree AV block and may result in diastolic mitral regurgitation, pseudo-PM syndrome, and lAF. 120

Figure 1. – Algorithm for preservation of intrinsic atrioventricular conduction. AF: atrial fibrillation; AVI: algorithm for intrinsic atrioventricular conduction preservation.

Figure 1

Activation of the rate variation sensor may be beneficial in patients with chronotropic incompetence. The Advanced Elements of Pacing Trial (ADEPT) compared the impact of DDDR vs DDD mode on quality-of-life improvement in 872 patients with chronotropic incompetence. At 6-month follow-up, patients randomly assigned to DDDR mode had a higher peak heart rate compared with those assigned to DDD mode (113.3 ppm x 101.1 ppm; p < 0.0001). 121 However, at 1 year, the Specific Activity Scale scores and the secondary quality-of-life endpoints did not differ significantly between groups.

2.11. Direct Stimulation of the Cardiac Conduction System (His Bundle, Left Bundle Branch)

LV remodeling and consequent LV dysfunction promoted by RV pacing-induced dyssynchrony warrants the search for alternative pacing sites in patients with bradyarrhythmias requiring artificial ventricular pacing.

The MOST study reported that RV apical pacing in patients with sinus dysfunction led to a significant increase in AF episodes and HF hospitalizations. 35 More importantly, these adverse events were directly associated with the cumulative percentage of ventricular pacing.

Consequently, the search for a pacing strategy that preserves interventricular and IV synchrony and corrects bradyarrhythmias is a relevant clinical necessity that has led to physiological pacing. 122 , 123

Direct stimulation of the conduction system is the most physiological form of artificial pacing because it reproduces the natural electrical activation of the heart. The stimulus travels through the specialized conduction pathways (His-Purkinje system) and avoids RV pacing-induced dyssynchrony. 124

A nonrandomized study of 202 patients who were followed-up for 2 years reported that His bundle pacing was superior to RV apical pacing. In patients with 40% of ventricular pacing, hospitalizations due to HF were significantly reduced (15% vs 2%, p = 0.02). 125

His bundle pacing has some limitations, including technical difficulties to identify the best pacing site (higher implant times), higher pacing thresholds, lower intracavitary sensing amplitudes, and abnormal inhibition by cross-sensing. 126

Deep septal pacing with direct capture of the LBB or a nearby area (the lead is inserted with a sheath into the left side of the interventricular septum) is a viable option that preserves narrow QRS complexes and prevents dyssynchrony. From a functional perspective, despite technical differences between direct capture of the LBB and capture of the LBB area, pacing of the LBB area promotes synchronism equivalent to that of direct His bundle pacing.

In a small series of 10 patients with SND, Mafi-Rad et al. demonstrated the feasibility of LV pacing of the interventricular septum resulting in a narrow RBBB-like QRS morphology, stable thresholds, and improved hemodynamic performance (measured by dP/dt) compared with RV apical pacing. 127 Subsequently, Huang et al. demonstrated the feasibility of direct LBB capture via deep septal pacing for LBBB correction. Some criteria for defining LBB capture were established, and patients should meet at least 3 of the following: 1) presence of LBB potential recorded on the lead electrogram; 2) LV activation time < 90ms with a stable pacing output of 2V or 5V; 3) incomplete RBBB on ECG; 4) evidence of selective and nonselective LBB capture; and 5) evidence of direct LBB pacing via a concomitant lead in the His bundle or the left septum. 128 , 129

Physiological pacing via His bundle or LBB has been effectively used in different settings of bradyarrhythmia requiring ventricular pacing ( Table 12 ).

Table 12. – Recommendations for physiological pacing (His bundle, left bundle branch) in the treatment of bradyarrhythmias.

  Class of recommendation Level of evidence
SND with an indication for a conventional pacemaker in patients with delayed intraventricular conduction IIa C
Permanent atrial fibrillation with an indication for AV junction ablation for heart rate control IIa C
AV block without LV systolic dysfunction IIb B

AV: atrioventricular; LV: left ventricular; SND: sinus node disease.

Compared with RV pacing, both His bundle and LBB pacing modalities have been shown to be safe and stable, to improve synchrony and QRS duration measurements, and to tend to improve ventricular function. Compared with LBB pacing, direct His bundle pacing generates a normal and apparently more physiological QRS complex, but at the expense of higher implant times, higher thresholds, and lower R-wave amplitudes. 130 , 131

The evolution of implantable technologies, the availability of generators with specific algorithms for detection of proportional energy expenditure, and the results of long-term controlled studies will potentially establish physiological pacing as the preferred strategy in the near future. 132

2.12. Leadless Pacing

Artificial cardiac pacing is not free of problems. The incidence of complications involving leads and pulse generators, especially in the subcutaneous pocket, increases over the follow-up period and may affect more than 10% of patients with CIEDs.

Implantation of conventional systems is associated with risk of pneumothorax, hemothorax, lead dislodgement, venous occlusion, tricuspid valve insufficiency, and infection (which may occasionally lead to endocarditis). Furthermore, subcutaneous implantation is associated with occurrence of pocket hematoma and infection not only during the first surgical procedure but also during the generator replacement procedure. 133 , 134 Of all complications listed, infective endocarditis warrants special attention because of its association with increased morbidity and mortality. 135 , 136

Since PM leads are the main source of problems and complications, implantable technologies have naturally advanced towards prioritizing solutions that do not require them. Within this context, leadless pacing has emerged as a technological evolution with potential advantages over conventional pacing. Two leadless systems were initially introduced in the market, Nanostim leadless cardiac pacemaker (Abbott Medical Inc. Abbott Park, IL, EUA) and Micra transcatheter pacing system (Medtronic, Inc., MN, EUA), but only the latter is currently available.

The Micra system is a small “capsule” (26mm in length x 6.7 mm in diameter) containing a lead and a pulse generator which is implanted in the cardiac chamber via transvenous route. The procedure is relatively simple and involves femoral vein catheterization. A large-caliber sheath (24 French) in introduced in the inferior vena cava and advanced into the right atrium. A delivery system is then inserted inside the sheath, allowing deployment of the device in the RV. Initially, the RV apex was the recommended site for device implantation, but interventricular septal implantation is currently recommended. 137

A multicenter, single-arm, prospective clinical trial evaluated the Micra system and included 725 patients (mean age was 75.9 ± 10.9 years, 58.8% were male) with an indication for single-chamber permanent PM implantation. The primary objective was to evaluate device efficacy (threshold capture at 6-month follow-up) and safety (major complications). Implantation was successful in 99.2% of cases. There were 28 major complications in 3.4% of patients, including perforation or pericardial effusion (1.6%), complications in the venous access site (0.7%), and elevated pacing threshold (0.3%). There were no dislodgements or device emboli. The mean values of R-wave amplitude, pacing threshold, and impedance remained stable. The rate of complications was compared with that of a population of more than 2,000 patients (historical controls from other clinical trials of same-brand conventional PMs). The Micra study found a lower rate of major complications (HR: 0.49, 95% CI 0.33-0.75, p = 0.001), including fewer hospitalizations (2.3% vs 3.9%) and fewer system revisions (0.4% vs 3.5%). There was also a low rate of infections, which were unrelated to the implantation or the device. 138 These results were then confirmed at 1-year follow-up. 139 Another analysis of the same cohort reported improved quality-of-life parameters at 3 and 12 months and high levels of patient satisfaction. The Micra system was also associated with less activity restrictions compared with conventional systems. 140

The Micra system has also been evaluated in “real-world” registries. The largest is a multicenter registry (96 centers in 20 countries) that aims to include 1,830 patients. The primary endpoint is occurrence of complications within 30 days of implantation. The published results of the first 795 patients showed a high implant success rate (99.6%) and a low rate of major complications (1.5%). Conventional system implantation was contraindicated in approximately 20% of patients, mainly due to vascular access problems. In addition, there were five pericardial effusion events in the study population, two of which had to be drained. 141 Small single-center studies evaluating leadless PMs in real-world settings have confirmed the findings of high implant success rates and low complication rates. 142 - 144

The safety of leadless system implantation has also been evaluated in specific populations, such as patients undergoing hemodialysis or conventional device extraction due to infection.

The Nanostim device was assessed in a multicenter observational study, 145 , 146 whose results at 6-month follow-up did not differ from those obtained in the Micra studies. However, Nanostim was withdrawn from the market because of unexpected rates of battery failure causing loss of pacing output and PM communication in approximately 0.5% of cases. 147

In addition to the previously mentioned intraoperative complications, there are two situations that remain uncertain. First, the impact of the device on tricuspid valve function is still unknown. An initial study of 23 patients reported no effect on tricuspid valve function at 2-month follow-up. 148 A more recent study including 53 patients revealed that leadless PMs affect valve function and may cause or worsen tricuspid regurgitation, 149 probably due to mechanical interference of the device in the subvalvular apparatus. Acute complications of the procedure and PM-induced dyssynchrony are the least likely causes. Patients with septal positioning of the leadless PM (currently recommended because of lower risk of perforation) were associated with tricuspid regurgitation, which was probably explained by greater proximity to the valve and the subvalvular apparatus.

Another situation that remains unresolved is how to manage the device at end of service. A worldwide experience with early retrieval of the Micra system – range 1 to 95 days – reported good outcomes and low complication rates. 150 A case report of percutaneous device extraction performed 4 years after implantation has also been published. 151 However, how the device will behave in the long term and whether extraction is possible (or whether another device should be implanted in another site) remain unknown. 152 , 153

Importantly, with the emergence of new technologies, unreported complications may arise as the follow-up periods become longer. In addition, there are no current randomized studies comparing conventional PMs with leadless PMs.

The currently available device is single-chamber and rate-responsive (VVIR). Overall, the indications for leadless PMs are the same as those for single-chamber PMs – mostly for symptomatic bradycardia not requiring an atrial lead ( Table 13 ).

Table 13. – Recommendations for leadless pacemaker implantation.

  Class of recommendation Level of evidence
A leadless PM should be recommended in patients with compromised conventional devices (lead fracture after extraction due to infection) and severe venous obstructions IIa C
A leadless PM is an acceptable option in patients with atrial fibrillation and low ventricular response, provided that the advantages and limitations are discussed with the patient IIb B
A leadless PM is an acceptable option in patients in sinus rhythm with an indication for PM implantation with limited stimulation, provided that the advantages and limitations are discussed with the patient (eg, rare sinus pauses or rare paroxysmal AV block) A leadless PM may be considered in patients with complete AV block who could be recommended single-chamber ventricular pacing (oldest-old patients, sedentary patients, bedridden patients), provided that the advantages and limitations are discussed with the patient IIb C
A leadless single-chamber ventricular PM should be recommended in patients in sinus rhythm with complete AV block who are candidates for conventional dual-chamber pacing when maintenance of AV synchronism is desirable III C
A leadless PM is not indicated for children and young people (there is uncertainty about how to manage the device at end of service) III C

AV: atrioventricular; PM: pacemaker.

One of the main factors limiting the use of leadless PMs is the high cost of the devices, as demonstrated in a recent survey by the European Heart Rhythm Association (EHRA). 154

Conceptually, the best candidates for leadless PM implantation would be those with contraindications to conventional devices, such as no vascular access, risk of new surgery, hemodialysis, and previous infection of conventional systems. Some studies (mostly single-center) have demonstrated the effectiveness of leadless PMs in this subgroup of patients, 155 - 157 who were also evaluated in the aforementioned “real-life” studies.

The implant technique for leadless PMs is different from that of conventional PMs; thus, physicians should be experienced in cardiac pacing, femoral vascular access, and handling of large-caliber sheaths and RV leads. Cardiac perforation is uncommon but, when it occurs, it may affect the RV or the atrium and auricle and usually requires emergency surgical intervention. 158 , 159

3. Recommendations for Multisite Pacemaker Implantation/Cardiac Resynchronization Therapy

3.1. Patients in Sinus Rhythm

Despite ODT, many patients with HF with reduced LVEF (HFrEF) develop persistent symptoms and significant LV systolic dysfunction. Moderate or severe mitral regurgitation, reduced myocardial functional reserve, and VD are the most common factors related to poor response to ODT. In VD settings, CRT alone (through atrio-biventricular pacing) or combined with an ICD has been considered an excellent therapeutic option for patients with LBBB. CRT is intended to correct electromechanical dysfunctions in patients with HFrEF and VD and, consequently, improve LV performance.

A surface ECG is the method of choice when investigating VD symptoms and selecting patients for CRT. Although imaging methods such as echocardiography can detect mechanical VD, tissue Doppler was unable to identify patients who responded to CRT in the Predictors of Response to CRT (PROSPECT) study. 160 The Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) 161 and the Cardiac Resynchronization in Heart Failure (CARE-HF) studies 162 were the first large-scale randomized trials to test the effect of CRT on the clinical outcomes of total mortality and hospitalization rate. The studies found that CRT increased survival when combined with ODT.

Those results were corroborated by a 2006 meta-analysis of 8 clinical trials including a total of 3,380 patients. 163 In a mean follow-up duration of 29.4 months, there were 524 deaths, with significant reductions in mortality (OR: 0.72, 95% CI 0.59-0.88) and HF hospitalization rates (OR: 0.55, 95% CI 0.44 to 0.68) with CRT. All trials reported a significant improvement in quality of life (3 to 6 months), although follow-up criteria were heterogenous among studies. Furthermore, the number needed to treat (NNT) was estimated at 11 (11 devices need to be implanted to save 1 life in 2.5 years). Considering the average longevity of CRT devices (6 years), five devices would need to be implanted to avoid 1 death. 164

Those trials supported the first recommendations for CRT as an adjunctive therapy to ODT in patients with advanced HFrEF (LVEF ≤ 35% and NYHA functional class III or IV despite ODT for more than 3 months) and VD. VD was confirmed by the presence of IVCD on ECG. It should be noted that the COMPANION study also observed greater clinical benefits when combining CRT with ICD (CRT-D).

The Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy (MADIT-CRT), 165 the Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction (REVERSE) study, 166 and the Resynchronization-Defibrillation for Ambulatory Heart Failure Trial (RAFT) were subsequently published. 167 These studies compared CRT-D vs ICD alone in patients with LVEF ≤ 40% (REVERSE) or ≤ 30% (MADIT-CRT and RAFT) and NYHA functional class I-II (REVERSE and MADIT-CRT) or II-III (RAFT). Study results, which were later corroborated by a meta-analysis, confirmed that both CRT plus ODT and CRT-D were beneficial in reducing total mortality. 168

A meta-analysis of 5 randomized clinical trials including patients with asymptomatic HFrEF (NYHA functional class I) or NYHA functional class II HF demonstrated a significant reduction in total mortality and HF hospitalization rates in patients with NYHA functional class II HF. 165 - 167 , 169 , 170 165 , 166 , 167 Only 9% of patients had NYHA functional class I HF, and CRT significantly reduced HF hospitalization but not total mortality among them. 171 The results show that early CRT in asymptomatic patients with HFrEF may reduce HF progression possibly by means of reverse ventricular remodeling. However, the potential benefits of CRT in patients with NYHA functional class I HFrEF should be carefully weighed against possible adverse events and costs associated with CRT implantation.

A QRS duration ≥ 120ms as a cutoff point for CRT indication is based on the inclusion criteria of the COMPANION 161 and CARE-HF studies. 162 However, the results of the Echocardiography Guided Cardiac Resynchronization Therapy (EchoCRT) study revealed that cardiovascular mortality increased in the subgroup of patients with QRS <130 ms undergoing CRT. 172 , 173 These findings were corroborated by a meta-analysis showing limited benefits of CRT in patients with QRS < 140ms. Indeed, the longer the QRS duration, the better the response to CRT. 174 , 175 Furthermore, patients with LBBB and a QRS duration ≥ 150ms benefit the most from CRT. 165 , 176 , 177 A meta-analysis of 13 large studies including 12,638 patients confirmed the benefit of CRT in patients with LBBB and the higher risk of death from heart pump failure in patients with wider QRS complexes. 175

As for IVCD type, a subanalysis of large studies suggests that patients with a wide QRS complex and no LBBB respond poorly to CRT. Data from the REVERSE study, which compared an active CRT group (CRT ON) vs a control group (CRT OFF), revealed that patients without LBBB had no reverse LV remodeling regardless of QRS duration. 178 Similarly, a study evaluating the MADIT-CRT population demonstrated no clinical benefits from CRT-D in 537 patients with mild HFrEF and no LBBB, regardless of QRS morphology and duration. 179

Importantly, only a small number of patients with RBBB were included in the large studies, which precludes any definitive conclusions about the effects of CRT on this patient population. 180

Conversely, although CRT was not beneficial in patients without LBBB, a recent real-world observational study using information from the National Cardiovascular Data Registry (USA) evaluated the clinical response to CRT-D vs ICD in 11,505 patients without LBBB. In patients with nonspecific IVCD and QRS ≥ 150ms, CRT reduced mortality and HF hospitalization rates, whereas those with QRS < 150ms responded poorly to CRT, with increased mortality and hospitalization rates. 181 There were also increases in risk of death and HF hospitalization rate in patients with RBBB treated with CRT-D. These findings are consistent with those reported by Bilchick et al., 182 whose study included Medicare data. Additionally, Pastore et al. 183 reported that patients with typical RBBB (classically defined as QRS duration > 120ms; rsr’, rsR’, or rSR’ in leads V1or V2; and S-wave duration > R wave or > 40ms in leads I and V6) 184 responded poorly to CRT. Therefore, the current literature suggests that CRT is indicated in patients with HFrEF, nonspecific IVCD, and a QRS duration ≥ 150ms and that greater caution is required when recommending this therapy for patients with typical RBBB.

Indications for CRT in patients in sinus rhythm are listed in Table 14 .

Table 14. – Recommendations for cardiac resynchronization therapy in patients in sinus rhythm.

  Class of recommendation Level of evidence
CRT is indicated in patients in sinus rhythm with symptomatic HFrEF, LBBB, QRS ≥ 150ms, and LVEF ≤ 35% despite ODT I A
CRT is indicated in patients in sinus rhythm with symptomatic HFrEF, LBBB, QRS between 130 and 149 ms, and LVEF ≤ 35% despite ODT IIa B
CRT may be helpful in outpatients in sinus rhythm with NYHA functional class III or IV HFrEF, non-LBBB intraventricular conduction disorder, QRS ≥ 150ms, and LVEF ≤ 35% despite ODT IIa B
CRT may be helpful in outpatients in sinus rhythm with NYHA functional class III or IV HFrEF, non-LBBB intraventricular conduction disorder, QRS between 130 and 149ms, and LVEF ≤ 35% despite ODT IIb B
CRT is not indicated in patients with QRS <130ms and no other indication for right ventricular pacing III A

CRT: cardiac resynchronization therapy; HFrEF: heart failure with reduced ejection fraction; LBBB: left bundle branch block; LVEF: left ventricular ejection fraction; NYHA: New York Heart Association; ODT: optimal drug therapy.

3.2. Patients with Atrial Fibrillation

The prevalence of AF in patients with HF varies according to HF severity. The condition affects 5% of patients with NYHA functional class I HF and 40% of those with NYHA functional class IV HF. 185

Data on the effect of CRT on patients with AF and HF are limited but suggest benefits, although less so than in patients in sinus rhythm. This is explained by some unique AF features, such as loss of AV synchronism, higher risk of synchronized ventricular pacing failure due to difficulties in controlling heart rate and occurrence of fusion and pseudofusion beats, and increased incidence of ICD shock delivery (appropriate or inappropriate), hospitalization, and mortality. 186 - 190

Approximately one-quarter of patients undergoing CRT develop AF; however, most randomized controlled trials demonstrating the benefit of CRT excluded patients with AF (eg, CARE-HF and COMPANION). 191

The CARE-HF study compared CRT vs ODT and found that, although mortality was higher in patients who developed AF during follow-up, they still benefited from CRT according to the primary objectives of the study. 192

CRT-D was not superior to ICD alone in the subgroup of 229 patients with AF in the Resynchronization-Defibrillation for Ambulatory Heart Failure Trial (RAFT). However, less than one-third of patients received more than 95% of biventricular pacing during 6-month follow-up.

A meta-analysis 193 of 5 studies (4 prospective cohorts and the MUSTIC randomized clinical trial 194 ) compared the impact of CRT on 797 patients in sinus rhythm vs 367 patients with AF (AV junction [AVJ] ablation was performed in 56 patients). Improvement in NYHA functional class did not differ between groups, although the Minnesota quality-of-life scores and the 6-minute walk test results were significantly improved in the sinus rhythm group.

In a meta-analysis 195 of 23 observational studies including 7,495 patients undergoing CRT, patients with AF (25%) showed higher rates of nonresponse to CRT (34.5% vs 26.7%) and mortality (10.8% vs 7.1%) compared with patients in sinus rhythm. Furthermore, the presence of AF was associated with lower impact of CRT on quality of life, 6-minute walk distance test and LV end-diastolic volume, but with similar result in the improvement of the LVEF.

The Multisite Stimulation in Cardiomyopathies (MUSTIC-AF) 196 study enrolled 59 patients with HF, AF, and bradycardia (ventricular pacing with paced QRS ≥ 200ms) who were randomly assigned to RV pacing or biventricular pacing with 3-month crossover phases. AV node ablation was performed in 63% of patients. Study limitations included the small number of patients that concluded the crossover phases (only 39%), which precludes any conclusions. In the intention-to-treat analysis, exercise tolerance and peak oxygen uptake did not differ between groups. However, in the 37 patients who received effective therapy (97-100% of biventricular pacing), 6-minute walked distance and peak oxygen uptake increased significantly.

In the Post AV Nodal Ablation Evaluation (PAVE), 197 Optimal Pacing Site (OPSITE), 198 and AV

Node Ablation with CLS and CRT Pacing Therapies for Treatment of AF (AVIL CLS/CRT) 199 studies, CRT moderately but significantly improved quality of life, NYHA functional class, and LVEF compared with RV apical pacing in patients with AF and varying degrees of LV dysfunction undergoing AV node ablation.

The Acute Decompensated Heart Failure National Registry (ADHERE) 200 compared patients with HF, QRS ≥ 120ms, and LVEF ≤ 35% undergoing CRT-D implantation (n = 4,471) vs those who did not undergo device implantation (n = 4,888). CRT-D was associated with lower risks of mortality and readmissions. The same association was observed in the subgroup of 3,357 patients with AF.

CRT requires biventricular pacing for most of the time to be beneficial, avoiding intrinsic conduction as much as possible. In patients with AF and rapid AV conduction, this requirement may be difficult to meet.

Boriani et al. 201 evaluated 1,404 patients undergoing CRT for a mean follow-up duration of 18 months. All patients were in sinus rhythm at the time of study enrollment, and 443 patients had documented AF (32%). AF episodes occurred in patients with (22%) and without (16%) a history of AF, lasting from 10 minutes to several weeks. Percent biventricular pacing was 95% in patients with AF vs 98% in the entire patient population. In patients with AF, percent biventricular pacing was 98% during sinus rhythm vs 71% during AF episodes (p < 0.001). Biventricular pacing <95% was defined as suboptimal and was associated with the occurrence of persistent or permanent AF (p < 0.001) and uncontrolled ventricular rate (p = 0.002) The percentage of biventricular pacing was inversely proportional to heart rate in patients with AF, decreasing by 7% for each 10-bpm increase in ventricular rate.

The importance of achieving high percentages of biventricular pacing was demonstrated in a large cohort of 36,395 patients included in the US LATITUDE Patient Management System who were remotely monitored. 202 Mortality was inversely proportional to the percentage of biventricular pacing both in sinus rhythm and AF or atrial pacing. The highest mortality reduction was observed in those with biventricular pacing < 98%. Patients with biventricular pacing > 99.6% had a mortality reduction of 24% (p < 0.001), whereas those with biventricular pacing < 94.8% had a mortality increase of 19%. Optimal percentage of biventricular pacing was set at ≥ 98.7%.

The multicenter, prospective Ablate and Pace in Atrial Fibrillation (APAF) study included 186 patients with symptomatic permanent AF, uncontrolled ventricular rate or refractory HF, LV systolic dysfunction, and a wide QRS complex. 203 All patients underwent AVJ ablation and multisite CRT implantation and were randomly assigned to receive echo-guided CRT (97 patients) or RV apical pacing (89 patients). During a median follow-up period of 20 months, CRT reduced the composite endpoint of death or hospitalization due to HF and worsening HF (11% in the CRT group vs 26% in the RV apical pacing group). CRT reduced hospitalization and worsening HF, while total mortality was similar in both groups.

AVJ ablation eliminates intrinsic conduction, resulting in 100% biventricular pacing in patients with CRT. This strategy was evaluated in a series of 673 patients (LVEF ≤ 35%, NYHA class ≥ II, QRS ≥ 120ms). 187 Among 162 patients with permanent AF in this cohort, 48 received rate control drugs and 114 underwent AVJ ablation. At 4-year follow-up, reverse remodeling and exercise tolerance were similar between patients with AF and those in sinus rhythm. In patients with AF, CRT benefited only those undergoing AVJ ablation. Despite > 85% biventricular pacing time, LV function and functional capacity were not improved in patients on rate control drugs.

In an observational study, Gasparini et al. reported that AVJ ablation combined with CRT significantly improved survival when compared with CRT alone. 204 Among 1,285 patients evaluated, 243 had AF. Rate control (85% biventricular pacing) was achieved by AVJ ablation in 188 patients and drug therapy in 55. At 34-month follow-up, mortality was significantly reduced in patients undergoing AVJ ablation (4.3% vs 15.2%, HR: 0.26 for overall mortality and 0.15 for HF mortality). These results suggest that CRT should achieve 100% biventricular pacing for maximum therapeutic benefit.

The Cardiac Resynchronization Therapy in Atrial Fibrillation Patients Multinational Registry (CERTIFY) corroborated the importance of AVJ ablation in patients with AF undergoing CRT. The study compared patients with permanent AF undergoing CRT combined with AVJ ablation or rate control drugs (n = 895) vs patients in sinus rhythm (n = 6,046). The results showed that, within 37 months, all-cause mortality (6.8 vs 6.1 per 100 person-years) and cardiovascular mortality (4.2 vs. 4.0) were similar in patients with AF plus AVJ ablation and patients in sinus rhythm. In contrast, patients with AF plus rate control drugs had higher total and cardiovascular mortality rates (11.3 and 8.1, respectively; p < 0.001) ( Table 15 ).

Table 15. – Recommendations for cardiac resynchronization therapy in patients with atrial fibrillation.

  Class of recommendation Level of evidence
CRT should be considered in patients with permanent AF, reduced LVEF (< 50%), and an indication for AVJ ablation for heart rate control I B
CRT should be considered in patients with permanent AF, reduced LVEF ≤ 35%, and NYHA functional class III and IV heart failure despite optimal drug therapy, LBBB, and QRS ≥ 130ms, with a strategy that allows > 95% biventricular pacing IIa B
AVJ ablation should be considered if biventricular pacing < 95% IIa B

AVJ: atrioventricular junction; CRT: cardiac resynchronization therapy; LVEF: left ventricular ejection fraction; NYHA: New York Heart Association.

3.3. Conventional PM Upgrade

Patients undergoing permanent RV pacing with a conventional PM or an ICD may develop progressive LV systolic dysfunction due to mechanical and electrical dyssynchrony. Registry data indicate that PM-induced LV dysfunction may occur in 12 to 30% of patients. 205 , 206 This condition is diagnosed by the presence of a high ventricular pacing rate and no other cause of LV dysfunction, such as untreated myocardial ischemia, valvular heart disease, or arrhythmias.

Nonrandomized studies have suggested that CRT may reverse PM-induced dysfunction. 207 - 209 LVEF recovery or improvement occurred in up to 86% of patients in those studies. Thus, patients with a PM or an ICD with high percent ventricular pacing showing worsening clinical and/or echocardiographic status may be candidates for CRT upgrade ( Table 16 ).

Table 16. – Recommendations for cardiac resynchronization therapy upgrade in patients with a conventional pacemaker.

  Class of recommendation Level of evidence
CRT is recommended in patients with a high percentage of right ventricular pacing (> 40%) showing worsening clinical and/or echocardiographic status despite optimal treatment, provided that other causes of left ventricular dysfunction are ruled out (LVEF ≤ 35%). IIa B

CRT: cardiac resynchronization therapy.

3.4. Indication for Antibradycardia Pacing (First Implant)

Possible deleterious effects of ventricular pacing with a conventional MP may justify the indication of CRT as an alternative treatment of bradycardias.

Studies have suggested that percent ventricular pacing > 40%, or even > 20%, is more likely to result in PM-induced LV dysfunction (12 to 30%). Pre-implant LVEF and QRS duration were additional predictors of PM-induced LV dysfunction.

Randomized clinical trials have evaluated whether CRT is superior to conventional RV pacing in reducing the occurrence of LV remodeling and improving clinical outcomes in patients with LVEF > 35%. 210 , 211 A meta-analysis of those studies compared conventional RV pacing vs biventricular or His bundle pacing. 212 Biventricular pacing compared with RV pacing was associated with increased LVEF and reduced end-diastolic and end-systolic LV volumes (patients with LVEF between 36% and 52% are more likely to benefit). Only 1 study reported significant improvement in the 6-minute walk distance test, which included patients with permanent AF who underwent AV node ablation (patients with LVEF < 45% and NYHA functional class II/III HF had better results). Data on His bundle pacing are summarized elsewhere in this guideline.

The meta-analysis excluded some important studies because they included patients with LVEF < 35%. The Biventricular versus Right Ventricular Pacing in Heart Failure Patients with Atrioventricular Block (BLOCK HF) study enrolled patients with indication for PM implantation due to AV block, NYHA functional class I-III HF, and LVEF ≤ 50%, 56 and they were randomly assigned to receive conventional RV pacing or biventricular pacing. Mean LVEF was 43%, and 87% of patients had LVEF > 35%. The primary composite outcome of death from any cause, urgent health care visit for HF requiring intravenous diuretic, and an increase ≥ 15% in LV end-systolic was more common in the RV pacing group. The APAF study evaluated patients with permanent AF undergoing AV junction (AVJ) ablation and CTR implantation who were randomly assigned to biventricular pacing or RV pacing. 204 The composite primary outcome of death from HF, hospitalization, or worsening HF occurred in 26% of patients in the RV group and 11% in the CRT group. Biventricular pacing was beneficial regardless of LVEF and QRS duration. Mean LVEF was 38%, and 53% of patients had LVEF > 35%.

COMBAT 55 was a prospective, multicenter, randomized, double-blind, crossover study that included 60 Brazilian patients with NYHA functional class II-IV HF, LVEF < 40%, and AV block as an indication for pacing. All patients underwent biventricular system implantation and were crossed over to conventional RV pacing after every 3 months (group A: RVP-BiVP-RVP and group B: BiVP-RVP-BiVP). Patients were evaluated at the end of each 3-month period. There were significant improvements in quality of life, functional class, and LV end-systolic volume compared with RV pacing. Also, the mortality rate was higher in the RV pacing group.

Recommendations for CRT in patients with a conventional PM indication for the treatment of bradyarrhythmias are summarized in Table 17 .

Table 17. – Recommendations for cardiac resynchronization therapy in patients with a conventional pacemaker indication.

  Class of recommendation Level of evidence
CRT is recommended in patients with atrioventricular block, a permanent pacemaker indication, and left ventricular systolic dysfunction (LVEF < 40%). I A

CRT: cardiac resynchronization therapy; LVEF: left ventricular ejection fraction.

3.5. Indication for Implantable Cardioverter-defibrillator Combined with Cardiac Resynchronization Therapy

ICD implantation is indicated in patients with different clinical conditions for primary or secondary prevention of sudden death (see item 4). Many of these patients also have LV dysfunction and LBBB and meet the criteria for CRT indication (see item 3). Thus, there are patients that may benefit from both therapies (CRT-D).

Randomized clinical trials have evaluated the effects of ICD and CRT-D implantation on patients with LV dysfunction and IVCD. All patients participating in the CONTAK CD, Multicenter InSync ICD Randomized Clinical Evaluation (MIRACLE ICD), 213 MIRACLE ICD II, and Cardiac Resynchronization Therapy in Patients with Heart Failure and Narrow QRS (RethinQ) studies 214 underwent CRT-D implantation. The groups were active CRT vs no pacing.

The CONTAK CD study evaluated 490 patients with a standard indication for ICD implantation, LVEF ≤ 35%, NYHA functional class II-IV HF, and QRS ≥ 120ms. There was no significant difference in the composite primary outcome of death from any cause, HF hospitalization, and ventricular arrhythmia requiring ICD intervention. CRT-D significantly improved oxygen uptake, 6-minute hall walk distance, and LVEF. The MIRACLE ICD study included 369 patients who also had a standard indication for ICD implantation, LVEF ≤ 35%, NYHA functional class III-IV HF, and QRS ≥ 130ms. CRT-D was associated with improved quality of life and oxygen uptake. The MIRACLE ICD II study included 186 patients with a standard indication for ICD, ejection fraction (EF) ≤ 35%, NYHA functional class II HF, and QRS ≥130ms. CRT-D was associated with reduced LV volumes and improved LVEF and NYHA functional class. The RethinQ enrolled 172 patients with an indication for ICD, EF ≤ 35%, NYHA functional class III HF, QRS ≤ 130 ms, and echocardiographic evidence of dyssynchrony. CRT-D did not improve the endpoint of peak oxygen consumption.

In the MADIT CRT 165 and RAFT studies, 167 patients were assigned to CRT-D or ICD alone. The MADIT CRT evaluated 1,820 patients with LVEF ≤ 30%, NYHA functional class I-II HF, and QRS ≥ 130ms. The composite endpoint of death from any cause or a nonfatal HF event was significantly reduced with CRT-D. The RAFT study included 1,798 patients with LVEF ≤ 30%, NYHA functional class II-III HF, and QRS ≥120 ms or paced QRS ≥200 ms. CRT-D significantly reduced the combined outcome of death from any cause or HF hospitalization. Mortality and HF hospitalization were analyzed separately as secondary outcomes and were significantly reduced.

A pooled analysis of the trial results showed that CRT-D reduced total mortality compared with ICD alone (RR 0.84; 95% CI 0.73-0.96). 215 There was also a significant reduction in hospitalizations (RR 0.75; 95% CI 0.64-0.88).

The clinical decision on whether to implant a CRT-D device in patients with an indication for ICD should consider the pattern of IV block and the QRS complex duration. Results from meta-analyses suggest that the benefits of CRT are mostly restricted to patients with an LBBB pattern. 216 , 217 The RAFT study 167 reported that CRT-D only reduced the primary outcome of death

from any cause or HF hospitalization in patients with QRS > 150ms. In patients with an LBBB pattern, there was a continuous relationship between QRS duration and clinical benefit, whereas patients with a non-LBBB pattern only benefited when QRS > 160ms ( Table 18 ).

Table 18. – Recommendations for cardiac resynchronization therapy in patients with an implantable cardioverter-defibrillator indication.

  Class of recommendation Level of evidence
CRT in patients with an ICD indication, LBBB, QRS ≥ 150ms, reduced LVEF, and symptomatic HF while on optimal drug therapy I A
CRT in patients with atrioventricular block, ICD indication, and left ventricular systolic dysfunction IIa B

CRT: cardiac resynchronization therapy; HF: heart failure; ICD: implantable cardioverter-defibrillator; LBBB: left bundle branch block; LVEF: left ventricular ejection fraction.

3.6. Direct Stimulation of the Cardiac Conduction System

CRT is a well-established nonpharmacological treatment for patients with symptomatic HF, reduced LVEF, and wide QRS complex. Although this therapeutic modality has advanced, 20 to 40% of patients still do not respond to CRT. 218 In this setting, direct stimulation of the cardiac conduction system (His bundle or LBB pacing) may be a useful alternative. 219 - 225

Barba-Pichardo et al. described in 2012 a series of 16 patients with severe HF in whom LV pacing via the coronary sinus was not achievable. Direct His-bundle pacing was able to correct the conduction disorder (LBBB) in 81% of cases. 226 In another study, Lustgarten et al. reported on 29 patients who were randomly assigned to His bundle pacing or conventional CRT. Six-minute hall walk distance, NYHA functional class, quality-of-life score, and LVEF were similar between groups. 227

A meta-analysis of 11 studies including 494 patients undergoing His bundle pacing reported a successful implantation rate of 82.4%. The studies evaluated patients with AF undergoing AV node ablation and patients with CRT indication. His bundle pacing had promising results in small, observational studies, which suggests the need for randomized trials. 228

A recent randomized trial compared His bundle pacing vs standard biventricular pacing in 41 patients. Of 21 patients randomly assigned to His bundle pacing, 10 (48%) had to cross over to biventricular pacing. The main reason for therapy crossover was that His bundle pacing failed to correct the conduction disorder. This finding is in line with those of subsequent studies that reported bundle branch block correction with His bundle pacing in approximately 60% of cases. Nevertheless, 26% of patients in the conventional CRT group crossed over to His bundle pacing due to technical difficulties. The His bundle pacing group had greater QRS narrowing (174±18 ms to 125 ± 22ms, p < 0.001 vs 165 ± 17 ms to 164 ± 30 ms, p = 0.82) and greater improvement of echocardiographic parameters compared with the CRT group (80% of patients with His bundle pacing had an absolute LVEF improvement > 5%). Measurements of cardiac diameters and volumes were similar in both groups. These findings suggest that CRT and His bundle pacing may be complementary therapies considering the anatomical challenges of the coronary sinus and the failure to correct the disorder in all patients with LBBB. 229

Zhang et al. described a series of 11 consecutive patients with LBBB and a conventional indication for CRT undergoing direct LBB pacing. QRS complex was significantly shortened (129.09 ± 15.94 ms compared with native QRS 180.00 ± 15.86 ms, p < 0.01). Deep septal pacing was found to be feasible in patients with systolic dysfunction and LBBB, resulting in functional improvement and reverse remodeling. 230 , 231 In a prospective cohort of 63 patients with nonischemic cardiomyopathy, LBBB, and LVEF < 35% undergoing LBB pacing, Huang et al. reported that LBB pacing was successful in 61 patients. Significant LVEF increase, reverse remodeling, and improved functional class were found at 1-year follow-up. Furthermore, a hyperresponse with normalized LVEF was observed in 75% of patients. 232

Wu et al. conducted a nonrandomized study of 137 patients with symptomatic HF and wide QRS complex who were followed up for 1 year. Forty-nine patients underwent His bundle pacing, 32 underwent LBB pacing, and 54 patients underwent biventricular pacing. Conduction system pacing (His bundle or LBB) showed greater LVEF improvement compared with biventricular pacing (23.9%, 24%, and 16.7%, respectively, p < 0.05). Functional improvement was similar in both groups of conduction system pacing; however, LBB pacing was associated with lower pacing thresholds (0.49 vs. 1.35, p < 0.001). 233

A recent review addressing direct stimulation of the cardiac conduction system suggests that His bundle pacing may be used as a primary strategy to achieve CRT or as a rescue strategy in patients in whom the standard technique (via coronary sinus catheterization) is unsuccessful. The combination of both methods (simultaneous His-CRT pacing) is considered a new concept of pacing for CRT in selected patients with more disseminated IVCDs 234 ( Table 19 ).

Table 19. – Recommendations for direct stimulation of the cardiac conduction system as an alternative to cardiac resynchronization therapy.

  Class of recommendation Level of evidence
Physiological pacing is reasonable in patients with AV block, a pacemaker indication, and LVEF < 50% IIa B
Direct stimulation of the cardiac conduction system may be considered an alternative to conventional CRT in patients with symptomatic HF, LVEF ≤ 35%, and QRS ≥ 130ms IIa C
Direct stimulation of the cardiac conduction system may be considered as a rescue therapy in patients who do not respond to CRT IIb C

AV: atrioventricular; CRT: cardiac resynchronization therapy; HF: heart failure; LVEF: left ventricular ejection fraction.

3.6.1. Cardiac Contractility Modulation

Cardiac contractility modulation (CCM) is a therapeutic option for HF in patients with no conventional indication for CRT, narrow QRS complex (< 130ms), and LVEF 25-45%. CCM promotes high-voltage (30 to 50ms) stimulation of the right interventricular septum after myocyte activation during the absolute refractory period. This type of stimulation theoretically optimizes calcium dynamics, resulting in increased ventricular contractility and improved exercise tolerance and functional capacity. 235 However, it remains unclear whether CCM should be routinely indicated in patients who are unfit candidates (narrow QRS) or who do not respond to CRT. It also remains unclear whether CCM should be used in combination with conventional CRTs. 236 The available studies are small and, perhaps because of that, were unable to detect significant differences. 237

4. Recommendations for Placement of Implantable Cardioverter-defibrillators

4.1. Primary Prevention of Sudden Death

4.1.1. Ischemic Cardiomyopathy

Sudden death from ventricular arrhythmias is one of the main causes of death in patients with HFrEF, especially in those with ischemic heart disease in whom the incidence of ventricular myocardial fibrosis and, consequently, reentry circuits is more prevalent. 238 , 239

Importantly, studies evaluating the impact of ICDs on ischemic heart disease have broadly defined this condition as ventricular dysfunction secondary to at least 1 severe lesion in 1 of the 3 main coronary arteries or a previous history of documented AMI. 240 - 242

Several important studies have evaluated the impact of ICDs on primary prophylaxis of sudden death in patients with ischemic heart disease. The MADIT 243 and MUSTT 244 studies tested ICD placement in patients with nonsustained ventricular tachycardia (NSVT), LV dysfunction, and inducible sustained ventricular arrhythmias (SVA) on EPS. The MADIT study enrolled 196 patients with LVEF < 35%, NYHA functional class I-III HF, a previous history of AMI, recorded asymptomatic NSVTs, and inducible SVAs on EPS refractory to procainamide or an equivalent antiarrhythmic drug. The patients were randomly assigned to ICD implantation vs optimized medical treatment (OMT) for a mean follow-up period of 27 months. The mortality rate was 15.7% in the ICD group and 38.6% in the OMT group, with a 64% relative risk reduction in total mortality in the ICD group (HR 0.46; 95% CI 0.26-0.82; p = 0.009). The MUSTT study evaluated patients with EF ≤ 40%, NYHA functional class I-III HF, and recorded asymptomatic NSVTs. The initial objective of the study was to compare the efficacy of antiarrhythmic medications capable of suppressing ventricular arrhythmias on EPS vs placebo. As a consequence of the MADIT results, the protocol was modified to ICD implantation in patients with inducible ventricular arrhythmias and reverse failure with at least one antiarrhythmic drug. Patients treated with antiarrhythmic medications did not show a significant decrease in mortality, whereas patients treated with an ICD had a 76% relative risk reduction in mortality (HR 0.24; 95% CI 0.13-0.45; p < 0.001).

Although the EPS is not normally used for ICD indication because of the low negative predictive value, the MADIT and MUSTT trials showed important results in this population (NNT for total mortality reduction = 4.2).

The MADIT II study randomly assigned 1,232 patients with EF ≤ 30%, NYHA functional class I-III HF, and AMI for more than 30 days to ODT vs ICD. During a mean follow-up period of 20 months, the mortality rate was 19% in the ICD group and 24% in the ODT group, with a 31% relative risk reduction in overall mortality (HR 0.69; 95% CI 0.51-0.93; p = 0.016). 245 The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) expanded the inclusion criteria and enrolled patients with ischemic and nonischemic heart disease, LVEF ≤ 35%, and NYHA functional class II and III HF. During a mean follow-up period of 45.5 months, there was a 23% relative risk reduction in mortality in the ICD group (HR 0.77; 95% CI 0.62-0.96; p = 0.007), in which 52% of patients had ischemic heart disease. 246

Conversely, studies evaluating early ICD implantation after revascularization or myocardial ischemic events found a neutral or even negative effect on some secondary outcomes. The Coronary Artery Bypass Graft (CABG) Patch trial randomly assigned 900 patients aged < 80 years with EF < 36% and signal-averaged ECG (SAECG) abnormalities to prophylactic ICD implantation at the time of bypass surgery. During a mean follow-up period of 32 ± 16 months, the results regarding overall mortality were neutral (6.7% in the ICD group and 4.6% in the control group, HR 1.07; 95% CI 0.81-1.42; p = 0.64). 247 A subsequent analysis showed an increased rate of infection in the ICD group (2.2% vs 0.4%; p < 0.05). 246 The Defibrillator after Acute Myocardial Infarction (DINAMIT) trial randomly assigned 332 patients to ICD therapy and 342 to no ICD therapy 6 to 40 days after AMI. 243 Inclusion criteria were patients with LVEF ≤ 35% and depressed heart-rate variability on Holter monitoring. In a mean follow-up duration of 30 ± 13 months, total mortality did not differ between groups, with 62 deaths in the ICD group and 58 deaths in the control group (HR 1.08; 95% CI 0.76-1.55; p = 0.66).

Since the publication of the Metoprolol CR/XL Randomized Intervention Trial in Congestive Heart Failure (MERIT-HF), the rate of sudden death has been known to decrease proportionally to worsening HF (NYHA class). 248 In a subgroup analysis of the SCD-HeFT and MADIT II trials, patients with NYHA functional class I and II HF benefited the most from ICD therapy, whereas patients with NYHA functional class III HF did not benefit significantly. There are no robust clinical trials demonstrating the benefit of ICD placement in patients with NYHA functional class IV HF, only data from retrospective cohorts of patients waiting for a transplant or undergoing ventricular assist device (VAD) implantation. A retrospective study including 1,089 patients on the heart transplant waiting list revealed that 550 patients had an ICD implant (216 for primary prevention and 334 for secondary prevention). In a mean follow-up duration of only 8 months, 39 patients (18%) in the primary prophylaxis ICD group, 89 (27%) in the secondary prophylaxis ICD group, and 162 (30%) in the no ICD group died. In multivariate analysis, the presence of an ICD was an independent predictor of reduced mortality (HR 0.4; 95% CI 0.19-0.85; p = 0.016). 248 The same finding was reported by a study that identified patients listed for heart transplantation between 1999 and 2014 in the United Network for Organ Sharing (UNOS) registry. Data from 32,599 patients were analyzed, with a mean follow-up duration of 154 days. A total of 3,638 patients (11%) died while waiting for a heart transplant, with a mortality rate of 9% in the ICD group and 15% in the no ICD group and a relative risk reduction of 13% (HR 0.87; 95% CI 0.80-0.94 p < 0.0001). 249 In the subgroup of patients undergoing VAD implantation (9,478 patients), the presence of an ICD was associated with a 19% relative risk reduction in mortality (HR 0.81; 95% CI 0.70-0 .94). A systematic review analyzed the use of ICDs in 937 patients with VADs (in 93% of the cases, the VAD was used as a bridge to transplantation). During a mean follow-up period of 7 months, 16% of patients in the ICD group and 26% in the no ICD group died, with a 39% relative risk reduction in mortality (HR 0.61; 95% CI 0.46-0.82; p < 0.01) 250 ( Table 20 ).

Table 20. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of ischemic heart disease.
  Class of recommendation Level of evidence
An ICD is recommended in patients with a history of AMI > 40 days or chronic ischemic heart disease receiving optimal drug therapy, no myocardial ischemia that could be treated by surgical or percutaneous revascularization, a life expectancy of at least 1 year, LVEF ≤ 35%, and NYHA functional class II-III HF (or LVEF ≤ 30% and NYHA functional class I-III HF) I A
An ICD is recommended in patients with a history of AMI > 40 days or chronic ischemic heart disease undergoing optimal drug therapy, no myocardial ischemia that can be treated by surgical or percutaneous revascularization, a life expectancy of at least 1 year, LVEF ≤ 40%, spontaneous NSVT, and inducible sustained ventricular arrhythmias on EPS I B
An ICD may be considered in patients with a history of AMI > 40 days or chronic ischemic heart disease receiving optimal drug therapy, no myocardial ischemia that could be treated by surgical or percutaneous revascularization, and who are candidates for heart transplantation or VAD implantation IIa B
ICD is not indicated in patients with AMI < 40 days or an indication for revascularization III B
ICD is not indicated in patients with NYHA functional class IV HFrEF refractory to treatment and who are not candidates for heart transplantation or VAD implantation III C

AMI: acute myocardial infarction; EPS: electrophysiology study; ICD: implantable cardioverter-defibrillator; LVEF: left ventricular ejection fraction; NSVT: nonsustained ventricular tachycardia; NYHA: New York Heart Association; SVT: sustained ventricular tachycardia; VAD: ventricular assist device.

4.1.2. Nonischemic Cardiomyopathy

HF is a highly prevalent clinical condition with significant morbidity and mortality. HF etiology is defined as nonischemic in 20% to 30% of cases, which means that no significant lesions are seen on coronary angiography or that the imaging test result for ischemia is negative. LV dysfunction may result from unknown causes (idiopathic dilated cardiomyopathy) or viral infections, hypertension, exposure to potentially toxic agents (chemotherapeutic drugs, alcohol), Chagas disease, infiltrative diseases, peripartum cardiomyopathy, valvular heart disease, or genetic and autoimmune diseases.

Although advances in the treatment of nonischemic cardiomyopathy (NICM) have significantly reduced mortality in recent decades, sudden cardiac death (SCD) remains a major problem that accounts for 30% of deaths. 251 Primary prevention strategies for SCD in patients with NICM include pharmacological treatment, ICD implantation, and CRT. Randomized clinical trials have shown that drug therapy (beta-blockers, sacubitril/valsartan, and spironolactone) significantly reduces the rates of SCD in this group of patients. 252

Risk stratification should include clinical and laboratory assessment. The worse the NYHA functional class, the greater the absolute risk of overall mortality and SCD. SCD accounts for 64%, 50%, and 33% of deaths in patients with NYHA functional class II, III, and IV HF, respectively (HF progression accounts for 50% of deaths in NYHA class IV patients). Syncope is an important risk factor for SCD in patients with NICM. 253 Other clinical variables associated with a higher risk of arrhythmic events in this population include not using beta-blockers and SBP. 254 , 255 Laboratory tests, such as hemoglobin, uric acid, and atrial natriuretic peptide (brain natriuretic peptide [BNP]), were predictors of mortality and arrhythmic events in some studies. 256

LVEF reduction is considered the main risk factor for SCD and total mortality in patients with HF. However, few studies have evaluated LVEF as a risk factor for SCD in patients with NICM. The Marburg Cardiomyopathy Study (MACAS), a prospective cohort of 343 patients with NICM, revealed that for every 10% reduction in LVEF, there was a relative risk of 2.28 for major arrhythmic events (patients in sinus rhythm).

The prevalence of wide QRS complexes among patients with HF ranges from 20% to 50% and is associated with increased SCD and total mortality rates; however, in cohort studies of patients with NICM, there was no significant relationship between QRS prolongation and increased risk of SCD. 257 , 258

Holter monitoring may be useful for risk assessment as it allows investigating the presence of NSVT and measuring autonomic activities (heart rate variability and heart rate turbulence). The incidence of NSVT among patients with NICM ranges from 30% to 79%, but its use in the risk stratification of arrhythmic events is controversial. 259

A meta-analysis has suggested that cardiopulmonary exercise testing variables such as oxygen uptake (VO2), minute ventilation/carbon dioxide production (VE/VCO2) slope, and presence of periodic breathing are independently associated with an increased risk of combined events. Combinations include total mortality, cardiac mortality, heart transplantation, hospitalization, and need for VAD. 260 , 261

In cases of ischemic heart disease, performing an EPS with programmed ventricular pacing has been shown to identify patients at risk of serious arrhythmic events. 246 , 248 Conversely, EPS results are controversial in cases of NICM, and guidelines do not usually recommend routine EPS for risk stratification in this patient population.

Several studies have evaluated the association between genetic mutations and the pathophysiology and prognosis of NICM, particularly in patients with familial disease. 262 Mutations in the lamin A/C ( LMNA ) gene are among the most studied conditions. Such mutations are found in 6% to 8% of NICM cases and may reach up to 30% when combined with conduction system diseases and skeletal muscle involvement.

According to a recent meta-analysis of 34 studies including 4,554 patients, approximately 44% of patients with NICM have myocardial fibrosis (an important arrhythmogenic substrate). 263 These patients show higher rates of mortality, ventricular arrhythmias, and HF hospitalizations. 264 For each percent increase in late gadolinium enhancement (LGE) volume, the risk of mortality or arrhythmic events is estimated to increase by 3% to 20% (8% LGE; HR in univariate analysis 8.23, 95% CI 2.84-23.8). In the largest cohort of patients with NICM undergoing CABG surgery, 265 Gulati et al. reported that the combined endpoint of SCD and aborted SCD occurred in 29.6% of patients with myocardial fibrosis and in 7.0% of patients without fibrosis. The HR for presence of fibrosis was 4.61 (95% CI 2.75-7.74; p < 0.001), and the HR for extent of fibrosis was 1.10 (95% CI 1.05-1.16; p < 0.001). These results suggest that MRI may be useful for the risk stratification of patients with NICM.

Several studies have evaluated the impact of ICDs on patients with NICM. The SCD-HeFT 245 is the largest study and included patients with ischemic and nonischemic heart disease and NYHA functional class II-III HF. Among patients implanted with an ICD, 33.2% received at least one shock, with 22.4% receiving only appropriate shocks and 10.7% receiving only inappropriate shocks. 266 In the DANISH study, a randomized trial of patients with NICM whose primary endpoint was all-cause mortality, 556 patients received ICD therapy and 560 received only ODT. 267 In a median follow-up duration of 67.6 months, the primary outcome occurred in 21.6% of the ICD group and in 23.4% of the control group, with no significant differences (p = 0.28).

The current recommendations for ICD implantation in patients with NICM are listed in Table 21 .

Table 21. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of nonischemic heart disease.
  Class of recommendation Level of evidence
Patients on optimal drug therapy with LVEF ≤ 35%, NYHA functional class II-III HF, and a life expectancy > 1 year I B
An ICD should be considered in the presence of high-risk genetic abnormalities (especially lamin A/C) combined with two or more of the following factors: LVEF ≤ 45%, NSVT, high-risk mutations, and male sex IIa B
Refractory NYHA functional class IV heart failure with no chance for heart transplantation or CAD III C

CAD: coronary artery disease; ICD: implantable cardioverter-defibrillator; LVEF: left ventricular ejection fraction; NSVT: nonsustained ventricular tachycardia; NYHA: New York Heart Association.

4.1.3. Hypertrophic Cardiomyopathy

Hypertrophic cardiomyopathy (HCM) is a genetic disease caused by an autosomal dominant mutation in genes encoding sarcomere proteins, with a prevalence of approximately 1:500. 268 , 269 HCM is characterized by the presence of varying degrees of asymmetric LV hypertrophy, provided that there are no other conditions that may explain this abnormality. It may result in diastolic HF, LVOT obstruction, atrial and ventricular arrhythmias, and, in some cases, SCD. 270 , 271 Most patients have no symptoms, with SCD often being the first manifestation of the disease. 272 - 276

In 1958, Teare reported on a series of eight patients with asymmetric myocardial hypertrophy (the nomenclature was not consolidated at the time) or hamartoma and correlated the anatomical findings with a higher occurrence of SCD in young adults. Pathology findings consisted of a coarse disarray of the muscle bundles with hypertrophy of the individual muscle fibers and their nuclei. 277

The annual risk of SCD in patients with an HCM diagnosis is approximately 1%; however, some patients may be at higher risk because of certain characteristics. 278 Before the ICD was introduced, the mortality rate was approximately 1.5% per year; after its introduction, the mortality rate has decreased to 0.5% per year. 279 , 280

Maron et al. conducted a longitudinal, single-center study including a large cohort of 2,094 patients with HCM followed up for 17 years. Of 527 patients with a primary prevention ICD implanted based on at least one conventional risk factor, 15.6% experienced appropriate ICD interventions (ventricular tachycardia [VT]/ventricular fibrillation [VF]), corresponding to almost 50 times the number of events in the non-ICD group. 281

HCM is the leading cause of SCD in patients aged < 40 years, mostly resulting from VF. Thus, ICD implantation is the most effective strategy to reduce mortality in high-risk patients despite higher costs and the possibility of complications, discomfort, and psychological stress. 282 , 283 Individual characteristics, clinical manifestations, family history, and definitions of risk factors may hinder patient selection, in addition to the fact that SCD is uncommon in clinical practice. 284 , 285 Patients who are more likely to benefit from ICDs are identified through noninvasive tests such as clinical history, ECG, stress test, Holter monitoring, echocardiography, and cardiac magnetic resonance (CMR). Conventional risk factors for SCD include a family history of HCM-related SCD, unexplained syncope occurring within 6 months of evaluation, NSVT, septal thickness ≥ 30mm, and risk modifiers such as a hypotensive response during stress test, LV fibrosis, and HF with LVEF < 50%. 286 Risk stratification for an ICD indication in patients with HCM should be performed periodically, every 1 or 2 years. 287

An ICD indication for HCM is not based on randomized clinical trials but rather on observational studies. Studies of patients with HCM and implanted ICDs reported that life-threatening events with appropriate ICD therapy occur at rates of 12% per year in secondary prevention and 4% per year in primary prevention. 288 In this setting, the likelihood of experiencing appropriate ICD therapies seems to be similar among patients with 1, 2, 3, or more conventional risk factors (primary prevention), which suggests that the presence of a single marker may justify ICD implantation. Among conventional risk factors, a family history of SCD likely or definitely due to HCM in first-degree relatives aged ≤ 50 years, especially during childhood or adolescence, is highly significant. 289 Another risk marker for SCD is the extent and magnitude of hypertrophy, especially when wall thickness ≥ 30mm; borderline thickening (28 to 29 mm) may be considered at the cardiologist’s discretion. Spirito et al. evaluated 480 patients and reported that SCD incidence was almost twice as high for each 5-mm increase in ventricular myocardial thickness, reaching 1.8% per year in those with wall thickness ≥ 30mm. 290

The presence of unexplained syncope (when syncopal episodes are unrelated to LVOT obstruction or when vasovagal syncope is unlikely or has been ruled out) has been strongly associated with SCD risk in patients with HCM, especially if occurring within 6 months of initial evaluation. These patients had a 5-fold higher risk compared with those without syncope. Remote episodes of syncope (> 5 years before initial evaluation) did not correlate with an increased risk of SCD. 291

NSVT is defined by the presence of 3 or more episodes with 3 or more repetitive ventricular beats and/or 1 or more prolonged episode with 10 or more beats at a rate of ≥ 130bpm on 24-hour or 48-hour Holter monitoring. Reported NSVT incidence in patients with HCM has ranged from 20% to 46%. In patients with HCM, VT episodes are undoubtedly associated with SCD; however, data are less robust for demonstrating that the presence of NSVT alone is an independent risk factor. Conversely, the risk increases in the presence of risk modifiers, especially LV fibrosis. 292 , 293

Genetic counseling is important in patients with HCM. Identifying carriers of specific genetic mutations may help investigating the disease in close relatives. Because patients with positive gene tests are likely to develop HCM, they should be closely monitored for symptoms and risk factors over time. 294

There is growing evidence of the relationship between myocardial fibrosis on CMR and SCD risk, which is considered a risk modifier. 295 - 297 A Brazilian study of high-risk patients with HCM and implanted ICDs found myocardial fibrosis in 96.4%, with a mean fibrosis rate of 15.96%. This suggests that fibrosis may be more sensitive than other conventional risk markers. 298 Chan et al. reported that LGE ≥ 15% of LV mass was associated with a 2-fold higher risk of SCD in patients otherwise considered low risk. 299

Klopotowski et al. prospectively analyzed 328 patients with HCM undergoing CMR to evaluate whether LGE location could be used as an auxiliary tool in the risk stratification of SCD. LGE suggesting the presence of fibrosis outside the interventricular septal region in patients with HCM was associated with an increased risk of SCD or its equivalent, such as unstable VT or appropriate ICD therapy. Considering the risk calculator developed by the European Society of Cardiology (ESC), the presence of fibrosis outside the interventricular septal region in intermediate-risk patients may help identify those who are more likely to benefit from ICD and thus favor an ICD indication. 300 - 302

Although the ESC has encouraged the use of the risk calculator in patients with HCM, the tool has a low sensitivity to determine whether an ICD should be implanted in high-risk patients. The American society strategy is to analyze risk factors alone or combined with risk modifiers in each patient with HCM. This strategy has a sensitivity of 95% for predicting potentially fatal VT events, being superior to the mathematical model of the ESC risk score, whose sensitivity is approximately 34%. Conversely, the ESC risk calculator is more sensitive for identifying patients who are truly low-risk (those with a lower likelihood of events), approximately 92% vs 78% of the American society strategy, and this avoids unnecessary ICD implantations.

Abnormal responses or exercise-induced hypotension affect 1 in 3 patients with HCM. The mechanism reflects an exacerbated fall in systemic vascular resistance due to an autonomic dysfunction and/or dynamic obstruction of the LVOT. In young patients, an abnormal blood pressure response is associated with an increased risk of SCD. 303

The identification of a LV apical aneurysm on echocardiography or CMR regardless of size may be associated with an increased risk of sustained monomorphic VT. 304 Rowin et al. retrospectively evaluated 1,940 patients with HCM and found LV aneurysms in 93 of them (4.8%). The adverse event rate was 6.4% per year, which is 3 times higher than that observed in patients without aneurysms, and included SCD, appropriate ICD therapy, thromboembolic events, and end-stage HF with LVEF < 50%. Fifty-four patients were implanted with an ICD for primary prevention, including 19 in whom the aneurysm alone was considered a risk factor. Appropriate ICD interventions for VT/VF were experienced by 20% of patients. Patients with apical aneurysms had an arrhythmic event rate of almost 5% per year, over 5-fold higher than that in patients without aneurysms, suggesting an equivalence to other conventional risk markers in high-risk HCM populations. 305

Subcutaneous ICDs are potentially advantageous, especially in young people, considering the device capability to preserve the venous system and avoid chronic lead complications (as long as ventricular pacing is not required). Conversely, the effectiveness of subcutaneous ICDs in aborting VF in patients with HCM remains uncertain. 306

Finally, studies evaluating the role of EPS in the risk stratification of SCD in patients with HCM found no benefits. Therefore, EPS should not be indicated for this purpose 307 ( Figure 2 and Table 22 ).

Figure 2. – Algorithm for primary prevention of SCD in patients with HCM. +FH: positive family history; HCM: hypertrophic cardiomyopathy; LV: left ventricular; LVEF: left ventricular ejection fraction; LVOT: left ventricular outflow tract; NSVT: nonsustained ventricular tachycardia; SCD: sudden cardiac death.

Figure 2

Table 22. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of hypertrophic cardiomyopathy.
  Class of Recommendation Level of evidence
Patients with at least one major risk factor IIa B
Patients with NSVT or abnormal arterial hypotension during exercise in the presence of additional risk factors, if life expectancy > 1 year IIa B
Invasive risk stratification with EPS III B

Risk factors: wall thickness > 30 mm, family history of sudden death, NSVT, late gadolinium enhancement on MRI, syncope < 5 years, LV aneurysm, and LVEF <50%. EPS: electrophysiology study; NSVT: nonsustained ventricular tachycardia.

4.1.4. Chagas Cardiomyopathy

Chagas disease is caused by the protozoan Trypanosoma cruzi , being mostly transmitted to humans through the feces of bloodsucking insects of the Triatominae family. 308 The infection generally occurs during childhood, and the acute phase has an incubation period of 1 to 2 weeks that can last up to 3 months. In the chronic phase, patients have no clinically apparent signs or symptoms of the disease, while their serological tests remain positive for 2 to 4 decades. 309 , 310 These patients have the indeterminate form of Chagas disease, whose prognosis is mostly favorable. 311 , 312 Although many patients, through mechanisms that are still not fully understood, 313 will carry the indeterminate form throughout life, approximately 30% to 50% of infected patients will develop one of the determinate forms (cardiac, digestive, or mixed).

Chronic Chagas cardiomyopathy (CCC) has very marked pathophysiological characteristics and is the most common and severe clinical form of the disease, with increased morbidity and mortality rates in Latin America and in countries with significant immigration. 314

Eight to 10 million people are estimated to be infected with Trypanosoma cruzi in Latin American and other countries. 315 - 317 Based on previous estimates and considering the worst-case scenario, 3 to 5 million infected individuals are expected to manifest clinical forms of the disease in the chronic phase.

The estimated average annual mortality rate of CCC is 4% but may range from 1% to 10% according to risk stratifications based on clinical characteristics and simple cardiac tests. 318

In addition to risk stratification criteria, several markers of poor prognosis have been identified by different studies, especially regarding SCD in different clinical settings. 319 - 338 Variables such as presyncope and syncope, LV dysfunction and HF, sustained ventricular tachycardia (SVT) or NSVT, severe bradyarrhythmia (SND and AV block), and recovered cardiac arrest were identified as risk markers for SCD. Conversely, ventricular extrasystoles alone on Holter monitoring and RBBB do not significantly affect the prognosis of CCC.

SCD accounts for approximately 55% to 65% of all death causes and is often associated with HF manifestations, although it may also occur in patients with asymptomatic LV dysfunction. 339 - 341 Refractory HF accounts for approximately 25% to 30% of deaths. The correlation between CCC stages and causes of death was recently described: SCD is more prevalent in stage III, whereas the prevalence of death from HF progressively increases from stage I to IV.

The main mechanism of sudden death in CCC is arrhythmogenic, and SVT with subsequent VF accounts for the vast majority of fatal events. 342 Thus, the structural abnormalities of CCC (with inflammation, cell death, and reactive and reparative fibrosis) represent the ideal anatomical substrate because they promote unidirectional blocks and slow conduction areas favorable to electrical reentry triggering. The triggers that affect this anatomical substrate, named ventricular extrasystoles, are invariably present and complete the key elements for the onset of reentry ventricular tachyarrhythmia. 314 , 329 , 331 , 333 , 336 , 338 , 345 Thus, NSVT may affect approximately 40% of patients with CCC and abnormal segmental mobility and practically all patients with global LV systolic dysfunction and HF. 343 SVT, which has a more ominous prognosis, occurs spontaneously and can be reproduced on EPS in approximately 80% to 85% of patients. 319 , 328 , 329

Complete AV block, although less common, is another cause of SCD in CCC, resulting from necrotic degeneration and diffuse fibrosis predominantly in the AV region. 330

As previously mentioned, SCD may also result from massive pulmonary thromboembolism or systemic thromboembolism in vital organs. Exceptionally, SCD may be due to a ruptured LV apical aneurysm.

Rassi et al. developed a risk score to predict death in patients with CCC based on clinical variables and routine cardiac tests. 322 The score was also used elsewhere 344 in a retrospective cohort of 149 patients. The hypothesis that the presence of VT on cardiac stress test or Holter monitoring, LVEF < 0.50, and QRS > 50ms on SAECG could identify patients with CCC at risk of death in 5 years was raised. Low-risk groups are characterized by the absence or presence of one risk factor, intermediate-risk groups by the presence of two risk factors, and high-risk groups by the presence of three risk factors.

Primary prevention of SCD in patients with CCC theoretically includes the use of amiodarone or ICD therapy. 345 However, there is no scientific evidence to support an ICD indication for primary prevention of SCD in CCC. There are many pathogenetic and pathophysiological characteristics that hinder any direct comparison with the results of studies evaluating other heart diseases. 317 The most significant characteristic is that many patients with CCC, even those with preserved LV function, already have substrates for potentially fatal arrhythmias 331 , 345 - 348 ( Table 23 ).

Table 23. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of chronic Chagas cardiomyopathy.
  Class of recommendation Level of evidence
Patients with stable SVT, LVEF < 35%, on optimal medical treatment I C
Patients with stable SVT, LVEF > 35%, on optimal medical treatment IIa C
Patients with NSVT, LVEF < 35%, on optimal medical treatment IIb C
Patients with refractory NYHA IV HF, not candidates for heart transplantation III C

LVEF: left ventricular ejection fraction; NYHA: New York Heart Association; NSVT: nonsustained ventricular tachycardia; SVT: sustained ventricular tachycardia; HF: heart failure.

The Chronic Use of Amiodarone against ICD Therapy in Chagas Cardiomyopathy for Primary Prevention of Death (CHAGASICS) study is an ongoing open-label, randomized, multicenter clinical trial whose aim is to compare the efficacy of ICD therapy vs amiodarone for the primary prevention of all-cause mortality in patients with CCC and NSVT stratified by the Rassi risk score. 322 The CHAGASICS trial is expected to provide the scientific evidence needed to define the criteria for amiodarone and/or ICD indication in patients with CCC without life-threatening clinical outcomes.

4.1.5. Arrhythmogenic Right Ventricular Cardiomyopathy

Arrhythmogenic RV cardiomyopathy (ARVC) has an autosomal dominant mode of inheritance and variable penetrance, which causes mutations in genes encoding cell adhesion proteins, named desmosomes. ARVC predominantly affects the RV but may also affect the LV in approximately 0.5% of cases, resulting in myocardial tissue replacement by fibrosis and adipose tissue. Such structural changes often cause ventricular arrhythmias and SCD. 346 , 347 The estimated prevalence of ARVC ranges from 1:1,000 to 1:5,000 in the general population, representing a leading cause of SCD in athletes and young adults. 348

Ventricular arrhythmias, syncope, and SCD occur particularly in the second and third decades of life, usually during a physical activity. Syncope is reported in 16% to 39% of patients with ARVC at the time of diagnosis, is often related to physical activity, and has been associated with an increased risk of arrhythmias. 349

Frequent ventricular extrasystoles, NSVT, and SVT are important predictors of cardiac events, and SVT is an important predictor of SCD and appropriate ICD therapies. SCD may be the first manifestation of ARVC. 350 , 351

Ventricular arrhythmias usually originate in the RV (LBBB morphology), but the QRS axis during SVT often differs from the RV outflow tract (RVOT). Many patients may have QRS complexes with different morphologies. 352

Regions of fibrofatty tissue create areas of delayed ventricular activation, resulting in fractional deflections at the end of the QRS complex (epsilon waves) and late potentials on SAECG. In patients with suspected ARVC, performing an SAECG may be useful for diagnosis and risk stratification (class of recommendation: IIa; level of evidence: B). On CMR, SAECG abnormalities are related to ARVC severity and the occurrence of adverse events. 353

CMR uses LGE to provide information regarding ventricular function, cardiac chamber size, segmental abnormalities, and extent of fibrosis. LGE has demonstrated biventricular and LV involvement alone in 34% to 56% and 4% to 9% of patients, respectively. LGE areas on CMR are related to the location of the ventricular arrythmia substrate, identified by endocardial and epicardial electrophysiological mapping. 354

The value of EPS as a risk predictor of SCD in asymptomatic ARVC is uncertain (class of recommendation: IIb, level of evidence: B). In patients implanted with an ICD for primary prevention, inducible SVT is not a predictor of appropriate shocks. 355 The diagnostic criteria for ARVC are listed in Table 24 . 355

Table 24. – Revised criteria for the diagnosis of arrhythmogenic right ventricular cardiomyopathy.
I. Regional or global dysfunction and structural abnormalities
Major criteria Echocardiography
RV segmental akinesia, dyskinesia, or aneurysm and one of the following (end-diastole):
RVOT on PLAX view ≥ 32mm (≥ 19mm/m2corrected for BSA)
RVOT on PSAX view ≥ 36mm (≥ 21mm/m2corrected for BSA)
Change in fractional area ≤ 33%
Cardiac magnetic resonance
RV segmental akinesia, dyskinesia, or dyssynchronous RV contraction and one of the following:
Ratio of RV end-diastolic volume to BSA ≥ 110 mL/m2(men) or ≥ 100 mL/m2(women)
RV ejection fraction ≤ 40%
RV angiography
RV segmental akinesia, dyskinesia, or aneurysm
Minor criteria Echocardiography
RV segmental akinesia or dyskinesia and one of the following (end-diastole):
RVOT on PLAX view ≥ 29 to < 32mm (≥ 16mm to < 19mm/m2corrected for BSA)
RVOT on PSAX view ≥ 32 to < 36mm (≥ 18 to < 21mm/m2corrected for BSA)
Change in fractional area > 33% to ≤ 40%
CMR
RV segmental akinesia, dyskinesia, or dyssynchronous RV contraction and one of the following:
Ratio of RV end-diastolic volume to BSA ≥100 to <110 mL/m2(men) or ≥ 90 to < 100mL/m2(women)
RV ejection fraction > 40% to ≤ 45%
II. Wall tissue characterization
Major criteria Less than 60% of residual myocytes by morphometric analysis (or < 50% if estimated), with RV free wall fibrous replacement in more than one sample, with or without fat replacement on endomyocardial biopsy
Minor criteria 60% to 75% of residual myocytes by morphometric analysis (or 50% to 65% if estimated), with RV free wall fibrous replacement in more than one sample, with or without fat replacement on endomyocardial biopsy
III. Changes in repolarization
Major criteria Inverted T waves in the precordial leads (V1, V2, and V3) or beyond in patients aged > 14 years (in the absence of complete RBBB QRS ≥ 20ms)
Minor criteria Inverted T waves in leads V1and V2in patients aged > 14 years (in the absence of complete RBBB) or leads V4, V5e V6
Inverted T waves in leads V1, V2, V3, and V4in patients aged > 14 years in the presence of complete RBBB
IV. Repolarization/conduction abnormalities
Major criteria Epsilon wave (low-amplitude signals between the end of the QRS complex and the onset of the T wave) in leads V1to V3
Minor criteria Late potentials by ≥ 1 of the following 3 SAECG parameters in the absence of QRS duration ≥ 110ms on ECG
Filtered QRS duration ≥ 114ms
Terminal QRS duration < 40µV (low-amplitude signal duration) ≥38 ms
Root-mean-square voltage of terminal 40 ms ≤ 20µV
Terminal activation duration of QRS ≥ 55ms measured from the nadir of the S wave to the end of the QRS, including R’, in V1, V2, or V3in the absence of complete RBBB
V. Arrhythmias  
Major criteria NSVT or SVT with LBBB morphology and superior axis (negative or indeterminate QRS in leads II, III, and aVF and positive QRS in lead aVL)
Minor criteria NSVT or SVT of the RVOT, LBBB morphology, and inferior (positive QRS in leads II, III, and aVF and negative QRS in lead aVL) or unknown axis
  > 500 ventricular extrasystoles in 24 hours (Holter monitoring)
VI. Family history
Major criteria ARVC confirmed in a first-degree relative diagnosed according to these criteria
ARVC confirmed during autopsy or surgery in a first-degree relative
Identification of a pathogenic mutation categorized as associated or probably associated with ARVC
Minor criteria History of ARVC in a first-degree relative in whom it is not possible to determine whether the diagnostic criteria were met
Sudden death in a first-degree relative aged < 35 years with suspected ARVC
ARVC pathologically confirmed by these criteria in a second-degree relative

ARVC: arrhythmogenic right ventricular cardiomyopathy; BSA: body surface area; CMR: cardiac magnetic resonance; ECG: electrocardiography; LBBB: left bundle branch block; NSVT: nonsustained ventricular tachycardia; PLAX: parasternal long axis; PSAX: parasternal short axis; RBBB: right bundle branch block; RV: right ventricle; RVOT: right ventricular outflow tract; SAECG: signal-averaged electrocardiogram; SVT: sustained ventricular tachycardia.

Genetic tests performed on probands with suspected ARVC are positive in 30% to 54% of cases. Importantly, a negative test does not rule out the disease, and a positive test does not define the course of treatment. ARVC is detected in approximately 35% to 40% of first-degree relatives, and clinical screening with ECG, Holter monitoring, stress test, and cardiac imaging may identify family members at risk for ARVC. 356

Asymptomatic patients without ventricular arrhythmias should receive only beta-blockers and undergo periodical evaluations of ventricular function and arrhythmia. 357 Randomized trials evaluating the best option among antiarrhythmic drugs for the treatment of SVT are lacking. An observational study reported suppression of inducible VT with sotalol in 58% of patients; only 10% of patients had recurrent arrhythmias. 358 In another observational registry, beta-blockers and sotalol were not associated with a reduction in ventricular arrhythmias, whereas amiodarone was superior in preventing them in a small cohort. 359 Finally, SVT ablation reduces recurrent arrhythmias but does not eliminate the need for ICD implantation. 360

Patients with a history of aborted SCD, poorly tolerated SVT, and syncope are at increased risk of SCD, with an annual rate > 10%. ICD implantation is indicated in these cases. Different cohorts have shown SVT, unexplained syncope, frequent NSVT, family history of early SCD, extensive RV involvement, very prolonged QRS complexes, LGE on CMR, LV dysfunction, and inducible SVT on EPS as risk factors for SCD or appropriate shock. 361 A recent systematic review including 610 patients followed up for a mean period of 3.8 years revealed annual rates of appropriate and inappropriate shocks of 9.5% and 3.7%, respectively. 362 , 363

An ICD indication for primary prevention of ARVC is difficult to assess and should rely on detailed clinical evaluation, including family history, RV and LV dysfunction severity, long-term ICD complications, and psychological and economic impacts. ICD indications for primary prevention of ARVC are listed in Table 25 and Figure 3 .

Table 25. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of arrhythmogenic right ventricular cardiomyopathy.
  Class of recommendation Level of evidence
ARVC and aborted SCD, or SVT if expected survival > 1 year I B
ARVC and major RV dysfunction, or LVEF ≤ 35% if expected survival > 1 year I B
In ARVC and syncope likely due to ventricular arrhythmia, an ICD may be useful if expected survival > 1 year IIa B
ARVC and well-tolerated SVT, provided that ICD benefits and long-term complications are evaluated IIa C

ARVC: arrhythmogenic RV cardiomyopathy; ICD: implantable cardioverter-defibrillator; LVEF: left ventricular ejection fraction; RV: right ventricular; SCD: sudden cardiac death; SVT: sustained ventricular tachycardia.

Figure 3. – Flow diagram for patient stratification and implantable cardioverter-defibrillator indication in arrhythmogenic right ventricular cardiomyopathy. ICD: implantable cardioverter-defibrillator; LV: left ventricle; NSVT: nonsustained ventricular tachycardia; RV: right ventricle; SCD: sudden cardiac death; SVT: sustained ventricular tachycardia.

Figure 3

A new model for predicting ventricular arrhythmias in ARVC was recently published. 364 Predictive variables included male sex, age, syncope in the last 6 months, previous NSVT, number of ventricular extrasystoles on 24-hour Holter monitoring, number of leads with inverted T wave in the inferior and anterior leads, and RV ejection fraction. This new model allowed greater refinement in patient selection for ICD implantation when compared with the 2015 International Task Force flowchart, reducing the rate of implant indication by 20.6%. 365 The authors of the new model have launched an online risk calculator (www.arvcrisk.com) that calculates the risk of ventricular arrhythmia in 5 years. Although it does not determine an acceptable risk threshold for ICD implantation, the model is believed to help in the decision-making process for primary prevention.

4.1.6. Noncompaction Cardiomyopathy

Noncompaction cardiomyopathy (NCCM) is a rare congenital abnormality characterized by the formation of prominent trabeculations and deep intertrabecular recesses in the LV and RV. It occurs during the endomyocardial morphogenesis phase (between weeks 5 and 8 of fetal life), most commonly reaching the LV apex. 363 Both ventricles are involved in 22% to 38% of patients. Left ventricular noncompaction (LVNC) occurs alone or in combination with other congenital heart diseases. 366

An autosomal dominant mode of inheritance is present in at least 30% to 50% of patients. Several genes that cause LVNC have already been identified, and they generally encode sarcomere (contractile apparatus) or cytoskeletal proteins. 367

The clinical manifestations of NCCM are heterogeneous, ranging from completely asymptomatic cases to cases of severe and fatal manifestations such as HF, thromboembolism, AV and IV blocks, ventricular arrhythmia, and SCD. Predictors of higher mortality include age, LV end-diastolic diameter, symptomatic HF, permanent or persistent AF, bundle branch block, and associated neuromuscular diseases. 368

Echocardiography is routinely performed during initial investigations, and contrast use may improve diagnostic sensitivity. 369 CMR allows visualization of noncompacted and compacted myocardial segments and is able to identify thrombi and myocardial fibrosis. 370

SCD is the leading cause of death in NCCM and may occur at any age. There are no diagnostic tools for accurate risk stratification in these patients. Ventricular arrhythmias are reported in 38% to 47% of cases, and SCD occurs in 13% to 18% of patients. 371 Histological examination shows continuity between the ventricular endocardium and the deep intertrabecular recesses; the latter may facilitate arrhythmogenesis by forming reentry circuits underlying the scar tissue, predominantly at the LV apex and mid-apical segments. 372

Steffel et al. showed that SVT inducibility on EPS has limited value in the risk stratification of NCCM; in contrast, noninducibility may identify low-risk patients. 373 Endocardial and/or epicardial catheter ablation seems to be useful in patients implanted with an ICD who have frequent ventricular arrhythmias. 374 , 375 The rate of appropriate shocks in these patients, in secondary prevention, ranged from 33% to 37% during a mean follow-up period of 34 to 40 months. 376

There are no convincing data demonstrating that LVNC alone is sufficient for an ICD indication. The indication should be guided by the severity of LV systolic dysfunction and the presence of sustained ventricular arrhythmia (similar to idiopathic dilated cardiomyopathy). 377 , 378 However, a study of 77 patients with NCCM showed that LV dysfunction and dilation were not prominent in patients receiving an ICD for secondary prevention, which means that these criteria are fragile for primary prevention indication. Conversely, the presence of NSVT was more common in patients receiving secondary prevention compared with those receiving an ICD for primary prevention or those who did not undergo ICD implantation. Other risk factors that should be considered are family history and syncope. 378

Although there are no prospective studies addressing SCD prevention in NCCM, there are sufficient data from observational studies to support an ICD indication as a reasonable strategy for SCD prevention in these patients 379 ( Table 26 ).

Table 26. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of noncompaction cardiomyopathy.
  Class of recommendation Level of evidence
Patients with NCCM, LVEF ≤ 35%, NYHA functional class II and III HF, and a life expectancy of at least 1 year should receive a primary prevention ICD for SCD based on DCM recommendations I B
Patients with an NCCM diagnosis, normal ventricular function, and risk factors such as NSVT, family history of SCD, and syncope may receive a primary prevention ICD for SCD based on DCM recommendations IIb C
Patients with NCCM, normal ventricular function, and no risk factors should not be considered candidates for an ICD III C
Performing an EPS for risk stratification III C

DCM: dilated cardiomyopathy; EPS: electrophysiology study; LVEF: left ventricular ejection fraction; MNC: noncompacted cardiomyopathy; NSVT: nonsustained ventricular tachycardia; NYHA: New York Heart Association.

4.1.7. Congenital Long and Short QT Syndromes

Congenital long QT syndrome (LQTS) is characterized by QT interval prolongation and polymorphic ventricular arrhythmias often triggered by adrenergic stimulation. 380

Hundreds of mutations have been described in more than 13 different genes encoding ion channels that restore cardiomyocyte resting potential. LQTS may be inherited via an autosomal dominant pattern of transmission (Romano-Ward), such as in LQTS types 1 to 6; an autosomal recessive pattern associated with congenital deafness and very prolonged QT intervals (Lange-Nielsen); or an autosomal dominant pattern associated with extracardiac disorders, dysmorphism, and hypokalemic or hyperkalemic periodic paralysis (Andersen-Tawil, LQTS type 7). 379 , 381

The risk of sudden death depends on several factors, such as the type of mutation causing the phenotype, QT interval duration, and the presence of symptoms. Patients with a very prolonged QT interval (QTc > 500ms) or recurrent syncope may be at an annual risk of sudden death of up to 5%.

Short QT syndrome (SQTS) is defined by a QTc interval < 340ms or < 360ms in survivors of cardiac arrest due to VF/VT, a family history of sudden death in those aged < 40 years, and the presence of a confirmed mutation or a family history of short QT. SQTS is a rare condition in which mutations in genes encoding potassium channels can be found in up to 20% of cases. 382

Some patients with short QT may benefit from quinidine. Survivors of cardiac arrest should undergo ICD implantation, whereas asymptomatic patients should be closely monitored. The recommendations for ICD implantation for primary prevention of long and short QT syndromes are listed in Table 27 .

Table 27. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of long and short QT syndromes.
  Class of recommendation Level of evidence
Patients with long QT refractory to beta-blockers and/or sympathectomy, with syncope or recurrent polymorphic VT I B
Asymptomatic patients with beta-blocker-related QTc > 500ms may benefit from an ICD alone or combined with sympathectomy IIb C
Beta-blocker-related long QT with genetic testing confirming LQTS types 2 or 3 IIb B
Asymptomatic patients with QTc < 330ms and a family history of sudden death IIb C

ICD: implantable cardioverter-defibrillator; LQTS: long QT syndrome; VT: ventricular tachycardia.

4.1.8. Brugada Syndrome

Brugada syndrome is characterized by ST-segment elevation > 2mm in leads V1and V2(coved type) in the second, third, or fourth intercostal space in combination with the occurrence of polymorphic ventricular arrhythmia, syncope, or cardiac arrest. ST-segment elevation may occur spontaneously or be induced by sodium channel blockers such as ajmaline and procainamide.

Patients with a spontaneous ECG pattern (type 1) associated with unexplained syncope or recovered cardiac arrest are at greatest risk of sudden death. ICD implantation is associated with risk reduction in symptomatic patients 383 ( Table 28 ).

Table 28. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of Brugada syndrome.
  Class of recommendation Level of evidence
SVT or syncope likely due to an arrhythmia and a spontaneous type 1 Brugada ECG pattern I B
Syncope likely due to an arrhythmia and a drug-induced type 1 Brugada ECG pattern IIa B
SVT induced by PES with one or two extrastimuli in two sites in asymptomatic patients with a spontaneous type 1 Brugada ECG pattern IIb C
 

ECG: electrocardiogram; PES: programmed electrical stimulation; SVT: sustained ventricular tachycardia.

The phenotype of Brugada syndrome is associated with detectable genetic defects in up to 30% of cases. The SCN5A gene is involved in most mutations, but a negative genetic test does not rule out the diagnosis. 384

Several factors may trigger ECG manifestations or precipitate arrhythmic episodes, such as fever, anesthetic agents, and various psychotropic drugs (www.brugadadrugs.org).

Asymptomatic patients are at lower risk of sudden death. The role of programmed electrical stimulation (PES) in risk stratification is controversial. Brugada et al. found an association between polymorphic VT induction with up to 2 extrastimuli in the RV and the risk of death in asymptomatic patients. Arrhythmia induction with 3 extrastimuli reduces specificity and should be avoided. Other studies reported reductions in the positive predictive value of PES over time. 385 - 387

Patients in electrical storm triggering ICD shock therapy may benefit from clinical management with quinidine and from epicardial ablation of abnormal RV activation regions identified by electroanatomical mapping. 388 , 389

4.1.9. Catecholaminergic polymorphic ventricular tachycardia

Catecholaminergic polymorphic ventricular tachycardia (CPVT) is a rare arrhythmogenic genetic disorder characterized by adrenergic-induced bidirectional and polymorphic VT. The estimated prevalence of the disease is 1:10,000. Two types of genes have been identified: a dominant variant secondary to a mutation in the cardiac ryanodine receptor gene ( RyR2 ) and a rare recessive variant caused by a mutation in the cardiac calsequestrin gene ( CASQ2 ). 390

Clinical manifestations usually develop in the first or second decades of life and are triggered by exercise or emotional stress. ECG and echocardiogram are generally normal, but stress tests trigger atrial and ventricular arrhythmias (bidirectional and polymorphic).

A maximally tolerated beta-blocker dose is the treatment of choice. Flecainide and left thoracoscopic cardiac sympathectomy may be used as adjunctive therapies.

ICD implantation should be considered in patients with cardiac arrest, recurrent syncope, or VT despite ODT 391 ( Table 29 ).

Table 29. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of catecholaminergic polymorphic ventricular tachycardia.
  Class of recommendation Level of evidence
Patients with CPVT and syncope or SVT despite maximally tolerated beta-blocker doses or patients with a beta-blocker contraindication and a life expectancy > 1 year IIa C
Patients with asymptomatic CPVT and a good response to beta-blocker treatment III C

CPVT: catecholaminergic polymorphic ventricular tachycardia; SVT: sustained ventricular tachycardia.

In most cases, the ICD should be programmed with a long detection interval given that shock-related pain and stress may trigger more arrhythmias and, consequently, an electrical storm. The decision to implant an ICD should consider the high probability of shock delivery (appropriate and inappropriate) and the chance of complications associated with young patient age.

A recent systematic review reported incidence rates of 40% for shocks, 19.6% for electrical storms, 1.4% for post-implant mortality, and 32.4% for additional complications (lead fracture, endocarditis, and surgical revisions) in patients whose mean age was 15 years (11 to 21 years). 392

4.1.10. Idiopathic Ventricular Tachycardia

Ventricular arrhythmias in patients with normally structured hearts are mostly benign. However, a small number of patients may have malignant forms of monomorphic or polymorphic VTs and even ventricular fibrillation.

Many of these tachycardias are triggered by ventricular ectopics originating in very similar locations compared to those of benign aspect (outflow tract, aortic cusp, His-Purkinje system, mitral annulus, and papillary muscles). The exact mechanism of malignant ventricular arrhythmias is not completely understood yet. Anisotropy associated with slow conduction and functional block caused by rapid arrhythmogenic foci likely results in rhythm degeneration to VF and polymorphic VT.

High-risk characteristics are related to syncope or cardiac arrest and ECG findings of a short coupling interval in the first or second extrasystole, NSVT with short cycles, wide QRS complex (in VT or sinus rhythm), and polymorphic VT. 393 , 394 The recommendations for ICD implantation for primary prevention of idiopathic ventricular arrhythmias are listed in Table 30 .

Table 30. – Recommendations for placement of implantable cardioverter-defibrillators for primary prevention of idiopathic ventricular tachycardia.
  Class of recommendation Level of evidence
Patients with syncope, malignant NSVT, and a life expectancy > 1 year IIb C
Patients with benign asymptomatic idiopathic VT and a good response to pharmacological or ablative treatment III C

NSVT: nonsustained ventricular tachycardia; VT: ventricular tachycardia.

4.2. Secondary prevention of sudden death

4.2.1. Recovered cardiac arrest or sustained ventricular tachycardia

4.2.1.1. Recovered cardiac arrest or sustained ventricular tachycardia in the presence of structural heart disease

Cardiac arrest due to VT/VF and subsequent SCD constitute a serious public health issue, accounting for approximately 50% of all cardiovascular deaths. Additionally, the survival rate of out-of-hospital cardiac arrest is significantly low, ranging from 6% to 10%. 395 The few patients who survive cardiac arrest due to VT/VF are at high risk of recurrence of potentially fatal tachyarrhythmias. Thus, preventive measures are essential and include treatment of underlying heart disease and comorbidities, use of antiarrhythmic drugs, and adequate patient selection for ICD implantation. 396 , 397

In patients with ischemic or dilated heart disease, the protective role of drugs such as beta-blockers, angiotensin-converting enzyme inhibitors (ACEIs), angiotensin receptor blockers (ARBs), and statins, which reduce total, cardiovascular, and sudden deaths, is well known. 398

A meta-analysis 399 including more than 35,000 patients with LV dysfunction (LVEF <40%) revealed that beta-blockers, ACEIs, ARBs, and mineralocorticoid receptor antagonists reduce the risk of SCD compared with placebo (HR: 0.89, 95% CI 0.82-0.98; p = 0.02). When combined with ODT, ICD implantation provides additional benefits by reducing SCD rates (HR: 0.39, 95% CI 0.30-0.51; p < 0.0001). More recently, the combination of a neprilysin inhibitor (LCZ696) with an ARB (sacubitril/valsartan) was shown to be even more efficient than enalapril in reducing both HF and arrhythmic deaths. 258 Thus, whether LCZ696 has a primary antiarrhythmic function or whether cardiac arrhythmia reduction results from clinical HF improvement remains unknown. 400

For several years, antiarrhythmic drugs were the main strategy for secondary SCD prevention, although their use was based on only a few studies that reported a high rate of recurrent events. Until the early 1990s, class I agents (eg, quinidine, flecainide, encainide) were believed to reduce ventricular extrasystoles and mortality. A subsequent demonstration of the deleterious effects of these drugs after AMI and in HF led amiodarone to become the treatment of choice in these patients. The Cardiac Arrest in Seattle: Conventional versus Amiodarone Drug Evaluation (CASCADE) study enrolled 228 survivors of cardiac arrest who were randomly assigned to empiric treatment with amiodarone or class I drugs guided by EPS or 24-hour Holter monitoring. 401 At 6-year follow-up, event-free survival (cardiac death or VT) was 41% in the amiodarone group vs 20% in the standard therapy group. However, the absence of a placebo group prevents a conclusion on whether the outcomes could be explained by amiodarone use or by the risks associated with other antiarrhythmic drugs.

The ICD is considered a major advance in secondary SCD prevention, and its benefits have been evaluated in several randomized clinical trials. The Antiarrhythmics Versus Implantable Defibrillators (AVID) 402 study compared antiarrhythmic therapy (amiodarone or sotalol) vs ICD in 1,016 patients with either recovered cardiac arrest due to VT/VF, VT causing syncope, or hemodynamic compromise and LVEF < 40%. Survival was significantly improved in the ICD group at 1 year (89.3% vs 82.3%), 2 years (81.6% vs 74.7%), and 3 years (75.4% vs 64.1%) (p < 0.02). The study was mostly criticized for the greater number of patients on beta-blockers in the ICD group compared with the antiarrhythmic therapy group. A subsequent analysis reported that ICD implantation mostly benefited patients with lower LVEF. 403 Survival did not differ significantly among patients with LVEF > 35%. In patients with LVEF between 20% and 34%, 1-year survival was 89.6% vs 79.8% and 2-year survival was 82.5% vs 71.8% (p < 0.05) in the ICD and antiarrhythmic groups, respectively. In patients with LVEF < 20%, 1-year survival was 82.4% vs 73% and 2-year survival was 71.6% vs 63.8%, with no significant differences.

The Canadian Implantable Defibrillator Study (CIDS) 404 evaluated amiodarone vs ICD in 659 patients with documented VF, recovered cardiac arrest, VT causing syncope, VT > 150bpm/min causing presyncope or angina in patients with LVEF <35%, or syncope associated with inducible VT or documented spontaneous VT. Total mortality after a mean follow-up period of 4 years was 27% in the ICD group vs 33% in the amiodarone group, with no significant differences. A subsequent analysis reported that ICD implantation was superior in patients with two of the following criteria: LVEF < 35%, NYHA functional class III or IV HF, and age > 70 years. 405 After a mean follow-up period of 5.6 ± 2.6 years, mortality was 47% in the amiodarone group vs 27% in the ICD group (p = 0.002).

The Cardiac Arrest Study Hamburg (CASH) 406 included 349 survivors of cardiac arrest who were randomly assigned to treatment with propafenone, amiodarone, metoprolol, or an ICD. Propafenone treatment was discontinued after an interim analysis found increased mortality compared with ICD implantation. After a mean follow-up period of 2 years, total mortality was 12.1% in the ICD group vs 19.6% in the amiodarone and metoprolol groups combined, with no significant differences.

A meta-analysis evaluating the results of those 3 studies demonstrated relative reductions of 50% in arrhythmic mortality (p < 0.0001) and 28% in total mortality among patients with an ICD compared with those who received antiarrhythmic treatment (NNT = 29, p < 0.00006). 407 Patients with LVEF <35% benefited the most at 6-year follow-up (patients with an ICD had an increased survival of 4.4 months).

The Midlands Trial of Empirical Amiodarone versus Electrophysiology-guided Interventions and Implantable Cardioverter-defibrillators (MAVERIC) compared EPS-guided therapies (antiarrhythmic drugs, ICD) vs empirical amiodarone in patients with SVT or recovered cardiac arrest. 408 The results showed reduced mortality in the ICD group and no benefits from performing an EPS.

The recommendations for ICD implantation in secondary prevention are based on those studies ( Table 31 ).

Table 31. – Recommendations for placement of implantable cardioverter-defibrillators in patients with recovered cardiac arrest or sustained ventricular tachycardia in the presence of structural heart disease.

  Class of recommendation Level of evidence
Cardiac arrest due to VT/VF without a reversible cause, LVEF ≤ 35%, and a life expectancy ≥ 1 year I A
Spontaneous SVT with hemodynamic compromise or syncope without a reversible cause, LVEF ≤ 35%, and a life expectancy ≥ 1 year I A
Survivors of cardiac arrest due to VT/VF without a reversible cause, LVEF ≤ 35%, and a life expectancy ≥ 1 year IIa B
Patients with spontaneous SVT without a reversible cause, LVEF ≤ 35% refractory to other treatments, and a life expectancy ≥ 1 year IIa B
Patients with syncope of unknown origin, inducible hemodynamically unstable SVT, and a life expectancy ≥ 1 year IIa B
Incessant VT III C

LVEF: left ventricular ejection fraction; SVT: sustained ventricular tachycardia; VF: ventricular fibrillation; VT: ventricular tachycardia.

4.2.1.2. Survivors of Cardiac Arrest or Sustained Ventricular Tachycardia in the Absence of Structural Heart Disease

Channelopathies are genetically determined diseases involving different types of cardiac ion channel dysfunctions, such as increased or reduced ion channel function and ionic imbalance These conditions increase the risk of potentially fatal tachyarrhythmias and SCD. 409

The range of genetic mutations is extremely wide, with a large overlap of phenotypic expressions. Congenital LQTS, Brugada syndrome, catecholaminergic polymorphic ventricular tachycardia (CPVT), SQTS, J-wave syndrome, and early repolarization are classified as channelopathies. Although other mechanisms may be involved in the development of idiopathic VF, sudden arrhythmic death syndrome, and sudden childhood death syndrome, these are included in this section because of the predominant arrhythmic manifestation of VT/VF in the absence of structural heart disease. A detailed discussion of the characteristics of each channelopathy phenotypic expression is outside the scope of this guideline, but an extensive literature is available. 410 , 411

Survivors of cardiac arrest with channelopathies are at a higher risk for a new SCD episode. ICD implantation in this setting reduces the risk of SCD, with a reported rate of appropriate therapy between 8% and 33%. Overall, patients with channelopathies and syncope or VT despite appropriate drug therapy have an ICD indication unless specific conditions, such as very young age or low weight, are considered high-risk for device implantation. 412 In special situations, the clinical decision to implant an ICD should also consider other therapeutic options or adjuvant therapies, such as left cardiac sympathetic denervation in LQTS ( Table 32 ).

Table 32. – Recommendations for placement of implantable cardioverter-defibrillators in patients with recovered cardiac arrest or sustained ventricular tachycardia in the absence of structural heart disease.

  Class of recommendation Level of evidence
Patients with a channelopathy, cardiac arrest due to VT/VF, and a life expectancy ≥ 1 year I B
Patients with a channelopathy who develop SVT or syncope despite drug therapy, with a life expectancy ≥ 1 year IIa B
Patients with congenital LQTS who develop syncope or SVT despite maximally tolerated beta-blocker dose IIa B
Patients with Brugada syndrome and spontaneous ECG abnormalities, syncope likely due to an arrhythmia, and a life expectancy ≥ 1 year IIa C
Patients with Brugada syndrome and documented SVT, with or without syncope, and a life expectancy ≥ 1 year IIa C
Patients with CPVT who progress with syncope or SVT despite maximally tolerated beta-blocker dose, with a life expectancy ≥ 1 year IIa C
Patients with Brugada syndrome and drug-induced ECG abnormalities, syncope of unknown origin, and a life expectancy ≥ 1 year IIb C

CPVT: catecholaminergic polymorphic ventricular tachycardia; ECG: electrocardiography; LQTS: long QT syndrome; SVT: sustained ventricular tachycardia; VF: ventricular fibrillation; VT: ventricular tachycardia.

4.2.2. Syncope and Ventricular Tachycardia/Fibrillation on Electrophysiology Study

Syncope is a rare, exuberant symptom with a range of etiologies, pathophysiological mechanisms, and prognoses, and may be the only symptom to precede sudden death. Syncope is defined as the sudden and transient loss of consciousness due to cerebral hypoperfusion, with rapid, spontaneous, and complete restoration of consciousness.

In the presence of heart disease, syncope (even of unknown origin) may indicate the need for an ICD for SCD prevention given the strong association between syncope and potentially fatal ventricular arrhythmias. 413

In some cases, performing an EPS may be useful to confirm an arrhythmic cause. Inducible SVT is a predictor of SCD risk in ischemic heart disease with reduced LVEF, 414 whereas inducible VF is considered a nonspecific finding. 415 Conversely, patients without inducible SVT are considered low-risk 416 ( Table 33 ).

Table 33. – Recommendations for placement of implantable cardioverter-defibrillators in patients with syncope and ventricular tachycardia/fibrillation induced on electrophysiology study.

  Class of recommendation Level of evidence
Patients with ischemic cardiomyopathy, LVEF > 35%, syncope of unknown origin, and inducible SVT on EPS29-30 I B
Patients with nonischemic cardiomyopathy, syncope, inducible SVT on EPS, and no indication for primary prevention of sudden death31-32 IIa B

EPS: electrophysiology study; LVEF: left ventricular ejection fraction; SVT: sustained ventricular tachycardia.

4.3. Children, Adolescents, and Congenital Heart Disease

ICD indications in children have not been properly addressed in recent international guidelines because of the limited number of published studies. Consequently, current indications are mostly based on some small case series. 417 ICD indications in pediatric patients often follow similar criteria to ICD indications in adults for both primary and secondary prevention strategies, although common sense is required when evaluating these patients. 418

The Brazilian Society of Cardiac Arrhythmias (SOBRAC) and the Department of Congenital Heart Diseases and Pediatric Cardiology (DCC-CP) have recently published the Cardiac Arrhythmias in Children and Congenital Heart Diseases Guideline with the aim of standardizing the diagnosis and treatment of children with cardiac arrhythmias. 419

It is extremely important that physicians seek alternatives to avoid an ICD indication in children, as long as they do not result in increased risk. Adequate options of clinical treatment should be exhausted and, when indicated, ablation of arrhythmogenic foci should be considered.

Unlike in the adult population, there is no evidence to support routine use of ICDs in pediatric patients based on LV dysfunction alone. ICD implantation in children should be an exception because pulse generator size and lead caliber are associated with technical difficulties due to small body size, limited sites for generator pocket creation (usually in the abdomen), limited access routes, risk of venous thrombosis/obstructions, and increased risk of extrusion. 420

Clinical and electronic follow-up should include ECG and routine assessment by telemetry as well as periodic radiological assessment, which is essential to monitor lead behavior as the child grows. Lead implantation should be careful to leave a redundant curve to allow for patient growth without the need for multiple interventions. 421

Electronic programming of the device in children also differs from that in adults. The PM function should consider a baseline pacing rate adjusted for age and type of heart disease, usually ranging from 90 to 160 ppm (programming used in adolescents usually follows adult protocols). Importantly, a rate-adaptive AV interval and a short but adequate post-ventricular atrial refractory period should be programmed to avoid tachycardia due to electronic reentry, given that these patients usually exhibit good ventriculoatrial conduction.

Electronic ICD programming and clinical follow-up should be even more careful. 422 Children may develop sinus tachycardia with increased heart rates very easily, which could result in inappropriate shock delivery with a major psychological impact. Inappropriate shocks, especially if immediately after implantation, may lead the patient to distrust the device and medical staff and develop difficult-to-treat panic syndrome. 423 - 428

The criteria for arrhythmic detection and classification must be rigorously defined. Therapies should be restricted to shocks, and antitachycardia pacing (ATP) should be avoided because potentially fatal arrhythmias rarely manifest as monomorphic VTs. 429 - 431 In case of monomorphic VT development, ablation should be conducted whenever possible. 432 , 433 Polymorphic VTs respond better to early shock delivery and commonly deteriorate to VF during ATP attempts.

In adolescents, subcutaneous ICD placement may be an interesting option given that subcutaneous systems do not require intravascular leads. 434 - 437 However, use of larger generators and parasternal leads may cause local discomfort and image issues in thinner patients. Another limitation of subcutaneous ICDs is the inability to provide long-term antibradycardia pacing. These devices have been associated with inappropriate shocks due to a greater susceptibility to extracardiac noise detection. 438 , 439

The main indications for ICD implantation in children and adolescents are summarized in Table 34 .

Table 34. – Recommendations for placement of implantable cardioverter-defibrillators in children and adolescents in the presence of congenital heart disease.

  Class of recommendation Level of evidence
Survivors of cardiac arrest, provided that reversible causes have been ruled out I B
Patients with stable VT and ventricular dysfunction, provided that reversible causes have been ruled out A
Patients with symptomatic VT who have undergone hemodynamic and anatomical evaluation, provided that surgical correction or ablation of arrhythmogenic substrate is not an option C
Patients with recurrent syncope and ventricular dysfunction or induced VT IIa B
Patients with recurrent syncope associated with LQTS or catecholaminergic polymorphic VT despite maximally tolerated beta-blocker dose B
Patients with asymptomatic congenital long QT unresponsive to treatment or a family history of sudden death C
Patients with asymptomatic hypertrophic obstructive cardiomyopathy, with high-risk features C
Patients with arrhythmogenic RV cardiomyopathy and extensive ventricular involvement (RV and/or LF), VT, a family history of sudden death, or syncope of unknown origin C
Incessant VT or VT due to a reversible cause III C

LQTS: long QT syndrome; LV: left ventricle; RV: right ventricle; VT: ventricular tachycardia.

The evolution of surgical treatment for congenital heart diseases has led to increased survival rates among young adults. The presence of myocardial scarring secondary to the underlying congenital heart disease or to surgical treatment may result in complex ventricular arrhythmias or even SCD in these patients.

There is a significant correlation between residual hemodynamic abnormalities and VT in patients undergoing surgical correction of tetralogy of Fallot (TOF). RV hypertrophy and dilation together with residual RVOT obstruction and regurgitation are considered risk factors for VT and SCD. 440 - 446 A hybrid approach combining surgical repair of structural abnormalities and arrhythmia ablation guided by pre- or intraoperative mapping has been successfully used for reducing the incidence of arrhythmias. 447 - 449 Pulmonary valve replacement alone in patients with TOF results in hemodynamic and functional improvement but does not eliminate VT risk. SCD risk assessment should be conducted postoperatively to evaluate whether an ICD should be indicated. 450 , 451

Approximately 50% of ICD implants in adults with congenital heart disease are indicated for secondary prevention in patients between 36 and 41 years of age. 452 , 453 The rate of appropriate shocks in these patients ranged from 3% to 6% per year, and the rates of inappropriate shocks (15% to 25%) and complications (26% to 45%) were higher than those of other populations. 450 , 451 , 454 - 457 Thus, an ICD indication in this population should consider cost-effectiveness and psychological impact.

ICD implantation in adults with congenital heart disease may be challenging because of anatomical complexity, intracardiac shunts, and limited vascular access. A subcutaneous ICD may be a good option for these patients. 456

An ICD indication for primary prevention in patients with congenital heart disease is controversial. Kairy et al. proposed a risk score for patients undergoing surgical correction of TOF in which a score > 5 is considered sufficient to indicate an ICD. The score included the following criteria: previous palliative shunt (2), inducible SVT on EPS (2), QRS ≥ 180ms (1), ventriculotomy (2), NSVT (2), and LV end-diastolic pressure ≥ 12mmHg (3). 457

Patients with surgically repaired TOF account for approximately 50% of ICD implantations in adults with congenital heart disease. Annual rates of appropriate shocks in this population have been reported at up to 7.7% in primary prevention and 9.8% in secondary prevention. 458 Inducible SVT on EPS in patients with congenital heart disease does not seem to correlate with the occurrence of appropriate shocks. 458 In some cases, catheter ablation of recurrent monomorphic SVT may be an effective alternative to prevent ICD implantation. 459 - 463

Among patients with operated congenital heart disease, those with transposition of the great arteries (TGA) via atrial switch procedure, Ebstein anomaly, aortic valve stenosis, and single ventricle physiology are at higher risk for SCD. 464 - 467

Patients with a previous Senning or Mustard procedure are at higher risk for SCD, especially during exercise. In these patients, atrial switch procedures may result in increased volume and consequent stenosis of the pulmonary veins and increased end-diastolic pressures. 468 Additionally, myocardial perfusion studies have identified RV ischemia and infarction in more than 40% of patients. 469 , 470 Risk factors for cardiac arrest in patients who had atrial switch procedures include previous ventricular septal defect closure, HF symptoms, atrial arrhythmia, right ventricular ejection fraction between < 30% and 35%, and QRS > 140ms. 471 , 472 A multicenter study evaluating patients who had atrial switch procedures and ICD implants reported that the lack of beta-blockers was associated with a high risk of appropriate ICD therapy. 464 Since atrial arrhythmias often precede SVT in patients with TGA, atrial tachycardia treatment should be intensified. 473 , 474

SCD risk is higher among patients with adult congenital heart disease ( Table 35 ) compared with the general population; mean age at death ranges from 30 to 49 years. 473 - 476 Patients with moderate to severe congenital heart disease are at even greater risk of SCD, accounting for approximately 25% of all causes of cardiac death. 477 , 478

Table 35. – Congenital heart disease and risk of sudden cardiac death.

Congenital heart disease Sudden death incidence High-risk features
Simple
Atrial septal defect < 1.5% Ventricular pacing; RV dilation; pulmonary hypertension; NKX2-5 gene
Ventricular septal defect < 3% Ventricular pacing; RV dilation; pulmonary hypertension; NKX2-5 gene
Moderate
Tetralogy of Fallot 1.4% to 8.3% Unexplained syncope; complex ventricular arrhythmia or sustained ventricular tachycardia; QRS >180 ms; inducible sustained ventricular tachycardia; atrial tachycardia; LV dysfunction; significant RV dilation; severe pulmonary insufficiency or stenosis
Aortic stenosis 3% to 20% Unexplained syncope; significant LV hypertrophy; aortic stenosis with mean gradient >40 mm Hg; ventricular dysfunction
Coarctation of the aorta 2% Aneurysm at the repair site; aortic stenosis; hypertension; early coronary artery disease
Ebstein anomaly 3% to 6% Cardiomegaly; atrial fibrillation; tachycardia with wide QRS complex; mitral regurgitation; RV outflow tract dilation
Severe
Transposition of the great arteries Atrial switch 3% to 9.5%; Arterial switch 1%; Congenitally corrected 17% to 25% Mustard atrial switch; previous ventricular septal defect closure; unexplained syncope; atrial tachycardia; coronary ostial stenosis; systemic ventricular dysfunction; tricuspid valve regurgitation
Truncus arteriosus 4% Multiple repair operations; coronary artery anomalies; ventricular dysfunction and/or hypertrophy
Fontan procedure for single ventricle physiology 2.8% to 5.4 % Atrial tachycardia; long-term survival; protein-losing enteropathy; ascites

LV: left ventricular; RV: right ventricular.

Family history and presence of septal defects, cardiomyopathy, or conduction system blocks may be related to a mutation in the NKX2-5 gene, which is associated with early SCD risk. A positive genetic test warrants ICD implantation. 479 - 481

Patients with complex forms of congenital heart disease and multiple surgical interventions in the first decades of life and patients with hypertrophy with subsequent subendocardial ischemia are at increased risk of potentially fatal ventricular arrhythmias. Other risk factors for SCD in patients with congenital heart disease include greater disease complexity, ventricular and supraventricular arrhythmias, progressive increase in QRS duration, systemic ventricular dysfunction, and subpulmonary ventricular dysfunction. A history of unexplained syncope in adults with moderate to severe congenital heart disease may be suggestive of SCD risk; in such cases, an EPS may be performed to assess the need for an ICD. 457

Adults aged 40 to 50 years account for 40% to 67% of patients with congenital heart disease who receive an ICD for primary prevention. In these patients, the rates of appropriate shocks ranged from 14% to 22% in the first 3 to 5 years of follow-up. 482 In patients without vascular access or with a previous Fontan procedure, the risks of epicardial ICD implantation may outweigh the potential benefits; therefore, subcutaneous ICD implantation or heart transplantation should be considered. 478

The safety of antiarrhythmic therapy in patients with congenital heart disease may be affected by the presence of ventricular hypertrophy and dysfunction. Flecainide use was associated with proarrhythmia in 5.8% of patients and SCD in 3.9%. 482 Conversely, amiodarone is usually reserved for patients with symptomatic arrhythmias or for prevention of worsening ventricular function. 483 , 484

The main recommendations for ICD placement in adult patients with congenital heart disease are listed in Table 36 .

Table 36. – Recommendations for placement of implantable cardioverter-defibrillators in adult patients with congenital heart disease.

  Class of recommendation Level of evidence
In adult patients with congenital heart disease and complex or sustained ventricular arrhythmia in the presence of significant residual hemodynamic abnormalities, ICD implantation is indicated only after treatment of abnormalities (if possible) I B
In adult patients with congenital heart disease and unstable VT, ICD implantation is recommended only after evaluation and adequate treatment of residual abnormalities/ventricular dysfunction, with expected survival > 1 year I B
In adult patients with repaired tetralogy of Fallot and inducible VT/VF or spontaneous SVT, ICD implantation is reasonable if expected survival > 1 year IIa B
In adult patients with highly complex congenital heart disease who underwent surgical repair and frequent and/or complex ventricular arrhythmias, beta-blockers may help reduce the risk of sudden death IIa B
In adult patients with previously repaired moderate to severe congenital heart disease, unexplained syncope and moderate ventricular dysfunction or marked hypertrophy, an ICD indication is reasonable if expected survival > 1 year in cases of inducible SVT on EPS IIa B
In adult patients with congenital heart disease, LVEF < 35%, and HF symptoms, ICD implantation may be considered if there is expected survival > 1 year (even in the presence of additional risk factors) IIb B
In patients with incessant VT or VT due to a reversible cause III C

EPS: electrophysiology study; HF: heart failure; ICD: implantable cardioverter-defibrillator; LVEF: left ventricular ejection fraction; SVT: sustained ventricular tachycardia; VF: ventricular fibrillation; VT: ventricular tachycardia.

Choosing Implantable Cardioverter-defibrillator Type and Pacing Mode

4.4. Choosing Implantable Cardioverter-defibrillator Type and Pacing Mode

Once an ICD is indicated for SCD prevention, the physician must choose the surgical technique for device implantation (transvenous, epicardial, or subcutaneous) and pacing mode (ventricular, atrioventricular, biventricular, or atrio-biventricular).

4.4.1. Implantation Technique

In the absence of an atrial septal defect, patients weighing > 15kg normally undergo transvenous implantation. 485 If antibradycardia pacing (AV block, SND) is not required, a subcutaneous ICD may be a good option.

4.4.2. Pacing Mode

In patients requiring a PM function, the choice of pacing mode is crucial. Pacing modes that prioritize the preservation of spontaneous AV and IV conductions are associated with a lower incidence of AF and ventricular remodeling related to RV pacing-induced LBBB. 486 Furthermore, in patients with LV remodeling at the time of implantation (LVEF ≤ 40% and LV end-diastolic diameter ≥ 60mm) requiring ventricular pacing, biventricular pacing is superior to RV pacing alone. 487 , 488

Therefore, after an ICD indication, the choice of pacing mode should consider whether chronotropism and AV conduction are normal, whether IV conduction follows a spontaneous or PM-induced LBBB pattern, and whether there is LV remodeling.

Patients with SND and normal AV and IV conductions may receive either single- or dual-chamber ICDs provided that intrinsic conduction search algorithms are programmed to avoid PM-induced dyssynchrony.

4.5. Cost-effectiveness of Implantable Cardioverter-defibrillators in Primary and Secondary Prevention of Sudden Death

Although ICDs increase survival in patients with LV dysfunction at risk of SCD, their use is limited by high therapy costs. These costs refer to the device itself, hospital expenses, medical fees, complications, readmissions, and pulse generator and lead replacement.

Cost-effectiveness analyses consider the cost in local currency per quality-adjusted life years (QALYs) 489 or life years gained (LYG), 490 varying according to socioeconomic and cultural factors specific to each study population.

Cost-effectiveness should be analyzed in terms of mortality, based on multicenter studies. For example, if the ICD does not improve survival, it is not cost-effective. Therefore, expected survival is a key factor in cost-effectiveness analyses.

The risk of death from nonarrhythmic causes should also be considered given that ICDs are not recommended for this purpose. The cost-effectiveness ratio becomes unfavorable when the survival rate of ICD candidates is < 1 year; therefore, ICD implantation in patients with high morbidity and mortality may not be cost-effective. In older patients, some studies suggest that the expected survival should be > 5 years to achieve cost-effectiveness. 491 Cost-effectiveness studies evaluating ICD implantation for less common diseases, such as hypertrophic cardiomyopathy and channelopathies, are lacking.

4.5.1. Primary Prevention

Sanders et al. 492 analyzed the results of the MADIT, MADIT II, 247 COMPANION, 161 MUSTT, 493 SCD-HeFT, 245 and DEFINITE 494 studies and projected that ICD use would provide a gain of 1.01 to 2.99 QALYs at a cost of US$68,300,00 to US$101,500.00. Assuming that the generator would be replaced every 5 years, cost-effectiveness was estimated at US$30,000.00 to US$70,200.00 for each QALY gained compared with the results for controls. The authors estimated that cost-effectiveness would be less than US$100,000 per QALY if the ICD reduced mortality for ≥7 years.

In another analysis conducted by Sanders et al., patients treated with an ICD in the DINAMIT and CABG Patch trials 250 did not experience mortality reductions compared with controls; therefore, ICD implantation was not cost-effective. In the DINAMIT trial, an ICD was implanted 6 to 40 days after myocardial infarction in patients with LVEF ≤ 35% and depressed heart rate variability. The primary outcome (death from any cause) did not differ between the ICD group and the control group. The CABG Patch trial enrolled patients with coronary heart disease, EF ≤ 35%, and abnormalities on SAECG undergoing CABG surgery. Prophylactic ICD implantation at the time of CABG surgery did not reduce the primary endpoint of death from any cause. Thus, prophylactic ICD implantation in patients at high risk for sudden death (EF ≤ 35%, abnormal SAECG, and depressed heart rate variability) was not cost-effective in the first 40 days post-infarction or immediately after CABG surgery.

A 2010 Brazilian study evaluated the cost-effectiveness of ICDs in patients with HF from both public and private health perspectives (effectiveness was measured in QALYs). 495 Ribeiro et al. 495 reported that the cost-effectiveness ratio was R$68,318 per QALY in the public setting and R$ 90,942 per QALY in the private setting. The variables with the highest impact in the analysis were the costs of ICD implantation, the frequency of generator replacement, and ICD effectiveness. In more complex study populations, such as the MADIT population, cost-effectiveness was much more favorable in the public setting (R$23,739.00 per QALY) than in the private setting (R$33,592.00 per QALY). 496

In a 2007 Brazilian study, Matos et al. evaluated the cost-effectiveness ratio of ICDs vs drug treatment per LYG. 497 The calculated cost per LYG was R$20,530.00 (US$9,550.00) at the time. This calculation was based on the parameters of an incremental R$54,200.00 cost and a life expectancy of 2.64 years gained from ICD implantation compared with clinical treatment. The cost-effectiveness ratio was considered favorable for Brazilian standards.

In the UK, Buxton et al. (2006) reported costs of £57,000 per LYG and £76,000 per QALY over a long-term follow-up period. 498 The authors concluded that cost-effectiveness would be more favorable in patients with low LVEF, including higher-risk subgroups, but not for routine use.

Cowie et al. conducted in 2009 a meta-analysis of primary prevention studies in the European setting including patients with reduced LVEF and ICD indications according to European guidelines. Prophylactic ICD implantation was found to have a good cost-effectiveness ratio. 499 Estimated mean LYG and QALY were 1.88 and 1.57, respectively, and mean estimated cost per QALY was €31,717. These findings were reproduced in another European registry evaluating a primary prevention setting. 500

In a systematic review of economic evaluations, Gialama et al. (2014) reported that ICDs may have a good cost-effectiveness ratio in selected patient groups, 496 comparable to other established therapies for cardiovascular and noncardiovascular diseases. Variables such as ICD efficacy and safety, device costs (implantation and replacement), patient characteristics, and SDC risk had the highest impact in the analysis.

4.5.2. Secondary Prevention

In the Antiarrhythmics Versus Implantable Defibrillators (AVID) study, Larsen et al. evaluated the cost-effectiveness of ICDs compared with antiarrhythmic drugs (mostly amiodarone) in survivors of SVT and VF. 501 At 3 years, ICD cost-effectiveness per LYG was estimated at $66,677.00 (95% CI $30,761.00 to $154,768.00) compared with antiarrhythmic drug therapy. The 6- and 20-year projections estimated costs of approximately $68,000 and $80,000 per LYG. In the subgroup analysis, ICD implantation was more cost-effective in patients with VF and less cost-effective in patients with EF > 35%.

Thijssen et al. 502 evaluated ICD cost-effectiveness and found acceptable results compared with those of other treatments available in the public health system, such as erythropoietin in patients on dialysis, some chemotherapies for leukemia in older patients, lung transplantation, and neurosurgery for malignant intracranial tumors. The cost per QALY was similar to those of heart transplantation, hemodialysis, and peritoneal dialysis. Importantly, some factors may considerably reduce cost-effectiveness, such as complications, infections, and comorbidities that negatively impact patient survival and ICD longevity.

Appropriate and inappropriate shocks may reduce survival and quality of life and, consequently, cost-effectiveness. Several studies have analyzed the importance of programming ICDs with longer SVT detection times and increased heart rate detection. These features were able to avoid inappropriate and “unnecessary” shocks, with improved survival rates and/or reduced hospitalization rates. 503 , 504 Thus, less aggressive programming may improve the cost-effectiveness of ICDs.

Mealings et al. analyzed 13 cost-effectiveness studies of ICDs and CRTs using an analytical method to adapt treatment costs to the UK setting. 505 Cost-effectiveness was evaluated in several subgroups of patients based on clinical criteria such as functional class, QRS duration, age, presence of LBBB, and ischemic etiology. At a maximum acceptable cost of £30,000 per QALY, ICDs were cost-effective in patients with HF, LV systolic dysfunction, NYHA functional class < IV, and QRS <120ms. In patients with QRS between 120 and 149 ms, the ICD was cost-effective only in those with NYHA functional class I and II HF. In patients with NYHA functional class IV HF, ICD was cost-effective only when combined with CRT in patients with LBBB and QRS >120ms.

In older patients, especially those aged >80 years, the clinical efficacy and cost-effectiveness of ICDs are uncertain. Mean patient age at the time of enrollment in primary and secondary prevention studies was 58 to 66 years and 58 to 65 years, respectively. However, approximately 28% of eligible patients for ICD implantation are estimated to be over 80 years of age. 506 Real-world data revealed that approximately 8% to 12% of implants in the USA and Canada are placed in patients aged > 80 years. The sudden death/death from any cause ratio decreases with age, being 0.51 in patients aged < 50 years and 0.26 in patients aged > 80 years. 507 Given that the number of appropriate ICD therapy is similar in all age groups in both primary and secondary prevention, the sudden death/death from any cause ratio decreases in older patients because of an increase in deaths from other comorbidities.

Pellegrini et al. evaluated the impact of age at the time of ICD implantation on survival. 491 Patients were categorized as < 65, 65 to 75, and > 75 years of age. 501 Mean survival after ICD implantation in patients aged > 75 years was 5.3 years (half than in the other groups). For survival rates < 5 years, the cost per QALY would increase from $34,000-$70,200 (Sanders 4 ) to $90,000-$250,000. In this scenario, the ICD would not be cost-effective if the patient died less than 5 years after implantation.

The cost-effectiveness ratio of ICDs in Brazil and other developing countries needs to be assessed within the socioeconomic context of each country considering local aspects, gross domestic product, effectiveness, and complications. Thus, patients with compromised LV function, fewer comorbidities, and at higher risk of death from arrhythmia should be prioritized.

Reducing the cost of long-lasting devices and batteries can significantly increase cost-effectiveness. Additionally, every effort should be made to avoid inappropriate or unnecessary shocks, which would improve quality of life (a positive impact on the QALY index assessment) and increase battery longevity.

5. Recommendations for Implantable Loop Recorders

Implantable loop recorders allow continuous heart rhythm monitoring regardless of active patient participation. With the ability to record different events (bradycardia, tachyarrhythmia, pauses) and a battery life of approximately 3 to 4 years, the implantable loop recorder is an extremely attractive diagnostic tool in the investigation of unusual symptoms (eg, less than once a month) that may be attributed to bradyarrhythmias or tachyarrhythmias. 508

In patients with unexplained syncope in whom initial noninvasive investigation with ECG, 24-hour Holter, or prolonged monitoring could not explain the nature of the symptoms, implantable loop recorders have been superior to conventional investigation strategies, such as tilt table test and invasive EPS. Several studies have identified bradyarrhythmia as the main cause of syncope, particularly in older patients with an IVCD. In these cases, bradyarrhythmia was identified in up to 41% of cases, 70% of which were intermittent complete AV blocks. 509

In patients with cryptogenic ischemic stroke without documented AF, active investigation with serial ECG and prolonged monitoring can detect silent AF episodes in up to 23% of cases. 510 AF detection in these patients may change the course of treatment, and full anticoagulation therapy may be required indefinitely. However, randomized trials are still needed to support the efficacy of anticoagulant therapy in patients with silent AF detected by prolonged monitoring in cryptogenic stroke.

The Cryptogenic Stroke and Underlying AF (Crystal AF) study randomly assigned 441 patients with cryptogenic stroke after initial investigation to monitoring by an implantable loop recorder or conventional follow-up. 511 At 6-month follow-up, AF > 30s was detected in 8.9% of patients with an implantable loop recorder. At 12 months, AF was detected in 12% of these patients vs 2% of patients with conventional follow-up (p < 0.001).

The Stroke of Known Cause and Underlying Atrial Fibrillation (STROKE-AF) trial enrolled 496 patients aged > 50 years and was presented at the 2021 International Stroke Conference (late-breaking abstract 6). AF episodes > 2 minutes occurred in 12% of patients with an implantable loop recorder vs 1.8% in the control group (p < 0.001). Patients with an implantable loop recorder received more anticoagulation therapy and had reduced stroke recurrence. Although not derived from controlled studies comparing different therapeutic interventions, the data suggest that prolonged monitoring may be beneficial.

The current recommendations for implantable loop recorders are listed in Table 37 .

Table 37. – Recommendations for implantable loop recorders.

  Class of recommendation Level of evidence
Patients with recurrent syncope of unknown origin and no formal indication for a pacemaker or an ICD, after inconclusive clinical and laboratory investigation I A
Patients with recurrent palpitations of probable arrhythmic cause in whom other diagnostic methods found no correlation with symptoms IIa B
Cryptogenic stroke for AF detection in patients with a negative or inconclusive noninvasive investigation IIa B
Patients with suspected recurrent reflex syncope presenting with frequent and severe episodes IIa B
Patients with epilepsy in whom treatment was ineffective IIb B
Patients with unexplained falls IIb B

AF: atrial fibrillation; ICD: implantable cardioverter-defibrillator.

6. Recommendations for CIED Evaluation and Programming

6.1. Conventional Pacemakers

PM programming should adhere to the following basic principles: 502 , 512

  • To preserve or restore baseline resting heart rate and adapt it to stress demands, as well as to restrict pacing to the condition for which it was indicated, avoiding pacing when there are no proven benefits.

  • To preserve intrinsic atrioventricular conduction whenever possible.

  • To increase the longevity of the pulse generator battery while providing clinical benefits to the patient without compromising safety.

  • To detect arrhythmias and system malfunctions.

The care of patients with a PM should include electronic and clinical evaluation. In addition to personal (including current medications) and family history, investigation of symptoms, and physical examination, a 12-lead ECG should be performed, which is essential to assess sensing, capture, and arrhythmia functions. Echocardiography, although usually performed before implantation, may be required in the follow-up period to monitor LV remodeling due to the possible deleterious effects of chronic RV pacing and PM syndrome.

The electronic evaluation is performed by telemetry and should include the pulse generator, the leads, and the retrieval of data stored in the device’s memory, especially arrhythmic events and malfunctions.

System interrogation allows the assessment of generator battery longevity, lead integrity, and pacing and sensing threshold measurements. Temporarily inhibiting the PM function allows the identification of intrinsic rhythm, which is essential for optimal system programming. Statistical data related to each cardiac chamber and to arrhythmic events should be accessed, as well as intracavitary electrogram recordings.

The choice of pacing mode should consider the patient’s intrinsic rhythm: normal sinus, AF, SND, and/or AV block.

6.1.1. Sinus Node Disease

Single-chamber ventricular pacing (VVI) was initially widely used regardless of bradycardia type because of its simplicity and safety. However, more than a quarter of patients with VVI pacing develop PM syndrome (ventricular pacing that causes retrograde atrial conduction and results in symptoms such as dyspnea, palpitations, dizziness, and signs of low cardiac output), with significantly impaired quality of life. Thus, atrial pacing allows spontaneous AV and IV conduction in SND, preventing loss of AV synchronism, PM syndrome, and IV dyssynchrony secondary to RV pacing. 513 , 514

Atrial pacing may be programmed to AAI or DDD mode. The DDD mode allows the preservation of intrinsic conduction through specific algorithms but has more complications, such as lead dislodgement, than the VVI mode. Conversely, AAI devices require twice as many reoperations compared with DDD devices, often due to AV block development (injury progression). AV blocks in patients with SND occur between 0.6% and 1.9% per year, requiring system upgrade to DDD mode. 515 , 516

AAI mode is associated with a lower occurrence of AF and thromboembolic events compared with VVI mode in patients with SND. Similar results were observed with DDD mode, which is also associated with lower rates of AF and better quality of life compared with VVI mode. However, these benefits have no impact on the outcomes of mortality, HF, and cardiovascular death. 33 - 35

The deleterious effects of RV pacing may result in HF and poor survival as a consequence of induced dyssynchrony. Thus, algorithms that preserve intrinsic AV conduction, automatically prolong the AV interval, or promote device upgrade to AAI mode (with ventricular backup) should be programmed when using a dual-chamber PM to avoid unnecessary RV pacing in patients with preserved AV conduction. Initial data suggest that these algorithms promote significant reductions in ventricular pacing percentage (99% to 9%, p < 0.001) and AF (40%). Patients with SND-related first-degree AV block may lose these benefits if the PR interval is too long. 517

In the DANPACE study, which enrolled 1,415 patients with SND, DDDR mode with a maximum AV interval of 220ms was associated with a lower occurrence of paroxysmal AF; a very prolonged AV interval was associated with mitral regurgitation, increased preload, and AF, suggesting that there is a limit to AV interval prolongation. Thus, programming AV intervals > 220ms is not usually recommended.

Sensors for rate-responsive pacing are another important PM feature in SND. These sensors aim to increase heart rate in cases of increased metabolic demand, such as during physical exercise. Three small studies reported improved quality of life and increased exercise tolerance during sensor activation; however, these findings were not reproduced in the ADEPT study. 507 , 518

The automatic mode switching (AMS) function reverses DDD(R) to VVI(R) mode in case of AF development. Although there is no strong evidence supporting the benefits of AMS, it should be programmed particularly in patients with paroxysmal AF for symptom relief. 519

6.1.2. Atrioventricular Block

RV pacing in AV block may be programmed to DDD or VVI mode. DDD mode maintains AV synchrony but is associated with more complications (6.2% vs 3.2%), especially lead dislodgement, threshold increases, and infection. 520

In studies comparing DDD vs VVI mode in patients with complete AV block and SND (PASE, CTOPP), DDD mode was not associated with reduced cardiovascular mortality or hospitalizations. The CTOPP study reported AF reduction with DDD mode (greater benefit in patients with SND); however, 26% of patients in VVI mode had PM syndrome and had to be crossed over to DDD mode. Patients’ conditions improved significantly after upgrade to DDD mode. 31 , 33 In patients aged > 70 years, DDD mode did not seem to be superior to VVI mode in those with complete AV block after a 3-year follow-up period (including PM syndrome). Therefore, DDD mode is a suitable alternative in older patients with low life expectancy and physical activity restrictions.

6.1.3. Atrial Fibrillation

In patients with permanent AF and no possibility of reversion to sinus rhythm, only ventricular pacing is required. In these cases, VVI(R) mode is recommended. Sensors for rate-responsive pacing have been associated with improved functional capacity and quality of life in small studies. 521 , 522

6.1.4. Neurally Mediated Syncope and Carotid Sinus Syndrome

Neurally mediated syncope with a cardioinhibitory response is characterized by periods of intermittent bradycardia, requiring limited pacing periods with increased baseline rate to compensate for the sudden instability occurring during the event. In these cases, pacing should last for only a short period of time, only during symptomatic episodes (hysteresis function). DDI, DVI, or DDD mode may be programmed in combination with an algorithm that preserves intrinsic conduction. VVI mode has been more closely associated with syncope and presyncope occurrence than dual-chamber pacing (DDD and DVI) in some studies. 72 , 73

Algorithms such as the Rate Drop Response (RDR)®and Sudden Brady Response (SBR)®can identify abrupt drops in heart rate and respond by applying an accelerated intervention rate at programmable intervals. These algorithms are effective for symptom improvement in patients with neurally mediated syncope (cardioinhibitory) compared with conventional treatment without a PM. Although these algorithms have not been compared with other pacing modes, they are effective and allow PM inhibition for most of the time. In the International Study on Syncope of Uncertain Etiology III (ISSUE III), DDD mode plus RDR reduced the chance of syncope recurrence by 57%. RDR was programmed to intervene if heart rate reached 40 bpm or dropped by 20 beats from baseline heart rate (90bpm for 1 minute). 523

The Closed Loop Selection (CLS)®algorithm uses intracardiac impedance measurements to assess myocardial contractility changes in order to predict syncope onset and initiate intervention (myocardial contractility increase occurs in the early phase of syncope episodes). 520 , 524

The automatic capture threshold testing feature has proven to be safe and may prolong the pulse generator longevity by 60%, reducing costs by 42% in 10 years. With few exceptions, these algorithms should be programmed routinely 525 ( Table 38 ).

Table 38. – Electronic programming recommendations for conventional pacemakers.
  Class of recommendation Level of evidence
DDD mode with a maximum AV interval of 220ms and algorithms for intrinsic conduction search in SND I A
DDD mode in complete AV block to avoid pacemaker syndrome I B
VVIR mode in permanent atrial fibrillation I C
The automatic mode switching function should be programmed routinely I C
Sensors for rate-responsive pacing in patients with chronotropic incompetence IIa B
The automatic capture threshold testing feature should be programmed routinely to prolong generator longevity IIa B
Specific algorithms, hysteresis function, and a prolonged AV interval should be programmed to avoid unnecessary pacing in neurally mediated syncope and carotid sinus hypersensitivity IIa B

AV: atrioventricular; SND: sinus node disease.

6.2. Cardiac Resynchronization Therapy

The evaluation of patients with CRT should follow the same principles for conventional PMs and include parameters specific to dyssynchrony correction. Thus, periodic ECG allows evaluating whether biventricular pacing is active, whereas conducting an echocardiogram 90 days after implantation and repeatedly throughout follow-up should document reverse remodeling in patients who respond to CRT.

The pattern of ventricular activation depends on lead placement and how early each chamber is activated. LV activation results in right axis deviation, with a qR or Qr pattern in lead I and r or R waves in V1(see Figure 4 ). Ventricular fusion occurs during biventricular pacing. A qR or Qr pattern in lead I was observed in 90% of patients with biventricular pacing. Loss of the q or Q wave in lead I is highly suggestive of LV capture loss. R or r waves in V1were observed in 65% to 93% of patients with biventricular pacing. 526 , 527

Figure 4. – QS pattern in D1 and initial r wave in V1 proving early activation of the LV lateral wall in biventricular pacing.

Figure 4

CRT loss may occur during exercise in the presence of AF with high ventricular rate or due to AV interval shortening; 24-hour Holter monitoring and exercise stress test may be useful to identify such cases. Increased LV capture thresholds account for 10% of cases of CRT pacing loss. 517 This change should be suspected when there is loss of Qr or qR pattern in lead I on 12-lead ECG and confirmed by a pacing threshold test. Current systems have an automatic capture threshold testing feature that helps to identify variations in pacing thresholds that have not been detected during conventional assessment.

Protocols from large CRT studies should be followed when programming basic parameters: DDD 50bpm, sensed AV interval of 100 to 120ms, and VV interval of 0ms (simultaneous biventricular pacing), with deactivation of sensors for rate-responsive pacing. Patients with AF should be programmed to DDI mode if there is an atrial lead, VVI if there is no atrial lead, and DDD if there is paroxysmal AF. In these cases, a rate of 60bpm is recommended in the absence of chronotropic incompetence.

Most large studies use DDD or VDD mode with a rate between 35 and 60bpm to reduce atrial pacing, which could compromise AV synchrony in patients with delayed interatrial conduction and impair ventricular filling. HF guidelines recommend the use of heart rate-lowering drugs when heart rate is > 70bpm despite the use of beta-blockers, which supports the programming of low baseline rates and routine deactivation of rate-responsive pacing sensors. 528

A short AV interval (100 to 120ms) aims to achieve a biventricular pacing percentage of approximately 100%, avoiding biventricular capture loss associated with PR interval shortening. AV and VV interval optimization by echocardiography or other methods is usually reserved for patients who do not respond to CRT, since there is no consensus on the real usefulness of these methods in routine practice. 529

The maximum rate of atrioventricular synchrony should be programmed to the maximum, considering the maximum predicted heart rate for age and any limitations related to the underlying heart disease.

Patients with pacing > 93% have a 44% reduction in mortality and HF hospitalization rates (combined outcomes), with optimal results being achieved with pacing > 98%. Fusion and pseudofusion beats may overestimate the biventricular pacing percentage. When heart rate control in patients with AF cannot be achieved by ODT, AV node ablation should be performed as it is associated with reduced mortality. 206 Paroxysmal and persistent AF ablation (pulmonary vein isolation) should be considered in patients with HF. 520 A detailed analysis of clinical situations with an indication for ablation is outside the scope of this Guideline.

Ventricular extrasystoles are related to reduced biventricular pacing percentages and reduced reverse remodeling, even with a relatively low incidence. Antiarrhythmics and, eventually, ablation should be considered in patients who do not respond to CRT. 530

Short AV intervals result in early ventricular systole, preventing the completion of the atrial contraction phase (A-wave truncation). In these cases, the AV interval should be prolonged until the A wave is evident. Conversely, prolonged AV intervals result in E and A wave fusion; in these cases, the AV interval should be shortened.

Two methods are commonly recommended for AV delay optimization: the interactive method and the Ritter method. In the interactive method, a long AV delay (200ms) is programmed and gradually reduced (20ms at a time) to 60ms while the mitral flow is observed. The shortest AV delay capable of maintaining the E and A waves separated (without fusion), without deformation of the A wave, and maintaining a 40ms distance from the end of the A wave to the beginning of the QRS complex is considered the optimal AV delay. Ritter’s method consists of measuring the QA interval (beginning of the QRS complex to the end of the A wave) with two different AV delays, one short (60ms) and one long (200ms). The optimal AV delay is calculated using the following formula: AV delay = long AV delay − (QA[shortAVdelay] − AQ[longAVdelay]). Thus, an echocardiogram with mitral flow assessment should be conducted after implantation to evaluate AV synchrony: if the E and A waves are separated and the interval at the end of the A wave is above 40 ms, there is no need for AV delay optimization.

The interventricular delay may be programmed empirically or optimized by echocardiography, ECG, or specific algorithms. Optimization by echocardiography is performed by testing different delays and assessing dyssynchrony. The delay resulting in the least dyssynchrony should be programmed. Simultaneous pacing, early LV pacing, and different interventricular delays (ie, 60, 40, and 20ms) should be tested. Subsequently, the same delays should be tested with early RV pacing. M mode, with or without tissue Doppler, and LV longitudinal shortening velocity, measured by tissue Doppler, are the most commonly used methods.

As in AV delay optimization, echocardiography-guided interventricular delay programming should be performed under specific conditions in patients who do not respond to CRT.

Some devices have automatic AV- and interventricular-delay optimization algorithms, whose effectiveness remains controversial but does not seem to be inferior to that of empirical or echocardiography-guided optimization.

A correlation between decreased QRS duration with biventricular pacing and the rate of CRT responders has been observed in retrospective studies, supporting the hypothesis that AV- and interventricular-delay optimization aiming at a shorter QRS may increase the response rate to CRT. 531

Multipoint pacing consists of stimulating regions (especially basal or apical) with delayed LV activation with a quadripolar lead, allowing capture of greater ventricular mass in a faster and more homogeneous way. The MultiPoint Pacing (MPP) trial compared multipoint pacing with a quadripolar lead vs conventional pacing. The results were similar in both groups, with a statistically significant value for noninferiority. However, there was a lower rate of nonresponders among patients with multipoint pacing programmed with an LV lead distance ≥ 30mm and the shortest delay (5ms). These results were reproduced in the first phase of the More Response on Cardiac Resynchronization Therapy With MultiPoint Pacing (MORE-CRT MPP) study. Patients programmed with an LV lead distance ≥ 30mm combined with the shortest intraventricular and interventricular delays also had better results.

6.3. Implantable Cardioverter-defibrillator

ICD programming should follow four basic principles: 1) reduce mortality by effectively reversing potentially fatal ventricular arrhythmias; 2) prioritize reversing ventricular arrhythmias with ATP whenever possible; 3) avoid inappropriate shocks; and 4) reduce RV pacing percentage as much as possible (antibradycardia pacing).

Appropriate therapies for sustained VF and VT termination are the cornerstone for mortality reduction in ICD intervention. To this end, tiered therapies should be programmed in different zones, which are classified into VT (one or two zones) and VF zones. Programmable therapies include shocks (up to 35 or 40 J), as well as pacing therapies with 3 to 20 pulses and a pacing rate faster than that of tachycardia (ATP) that can terminate monomorphic VTs without painful shocks, reducing shock-related myocardial damage.

The effectiveness of shock delivery in ventricular arrhythmia termination used to be evaluated intraoperatively by defibrillation threshold testing (VF induction followed by shock delivery for effective circulation reversal, with an output of at least 10J below the maximum programmable output). Subsequent studies have demonstrated that this strategy is unnecessary, since standard intraoperative measurements (pacing threshold, impedance, and R wave) are sufficient for effective termination of spontaneous arrhythmias. 532 Thus, certain complications due to intraoperative VF induction and shock-related myocardial damage can be avoided. VF zones should therefore be programmed with reversed polarity shocks with the highest possible energy. Conversely, VT zones may be programmed with lower-energy shocks, which are usually preceded by an ATP attempt. 533

The effectiveness of ATP as first-line treatment for VTs is well known. Monomorphic, organized VTs with stable cycles and no hemodynamic repercussions can be easily terminated by both burst ATP (with a fixed interpulse interval) and ramp ATP (with an automatically decreasing interpulse interval).

Some unstable arrhythmias, even in high-rate zones (in the VF range), may be terminated by ATP before the delivery of programmed shocks. In these cases, an ATP attempt should be programmed during or before shock energy load; if the arrhythmia is terminated, the shock should be aborted. The Pacing Fast VT Reduces Shock Therapies II (PainFREE II) study used ATP as first-line treatment in a fast VT zone (188 to 250bpm) and reported a significant relative reduction in the risk of shock of 71%, without compromising patient safety. 534

Adequate programming of detection and tiered therapy reduces inappropriate shocks, promotes higher rates of appropriate ATP, and reduces mortality. 535 To this end, basic programming principles should include:

  1. A VF zone programmed with a rate > 233bpm (> 188 for Medtronic devices), with at least 30 beats out of 40 (x in y) for detection. This strategy avoids shocks in nonsustained arrhythmias and inappropriate shocks due to intermittent noise, double counting, or extrasystoles.

  2. In primary prevention patients, a single VF detection zone may be sufficient. VT monitoring zones without therapies (monitoring) may be programmed at the physician’s discretion. In secondary prevention patients, VT-targeted therapies should be programmed with a detection cutoff of 10-20bpm lower than the documented tachycardia rate. Based on clinical criteria, low-rate therapy zones may be programmed according to the risk of slower VT, but ATP should always be prioritized. 536

  3. Noise and lead integrity monitoring algorithms should be programmed, as well as automatic adjustment and oversensing prevention features, such as T-wave detection.

  4. Adequate programming of supraventricular arrhythmia discrimination algorithms, particularly in the morphology discrimination criterion (single-chamber devices) and in the evaluation of algorithms based on the atrioventricular relationship (dual-chamber devices). Timers, such as the Sustained Rate Duration (Boston Scientific) and Timeout, should be deactivated, as they ignore the discrimination of events classified as SVT after the preestablished period and deliver inappropriate therapies.

Finally, the need for concomitant antibradycardia pacing must be carefully evaluated. Most patients with an ICD do not require antibradycardia therapy, especially for primary prevention. However, conventional RV pacing is known to increase the risk of ventricular dysfunction and mortality. ICD programming should, whenever possible, prioritize RV pacing percentage reduction. To this end, single-chamber ICDs should be programmed to VVI mode with 40 ppm, whereas dual-chamber ICDs should prioritize atrial pacing alone by using ventricular pacing minimization algorithms (eg, RYTHMIQ, MVP, IRS plus) or by programming an AV interval long enough to avoid unnecessary ventricular pacing. In patients requiring ventricular pacing due to AV conduction block, the possibility of pacing alternative sites, such as biventricular pacing (CRT) or conduction system pacing (His bundle/LBB), should be considered according to ventricular function. 537

6.4. Implantable Loop Recorder

Implantable loop recorders should be adequately programmed to detect ventricular electrical activity without undersensing and without noise oversensing, which could impair rhythm identification.

The automatic detection function, regardless of manual activation by the patient, a memory with electrogram storage capacity, and a battery life of 3 to 4 years help to identify arrhythmic events that have not been documented by conventional tests.

The implantation technique is similar for all available models. The adequacy of electrical signal at the position chosen for device placement should be confirmed intraoperatively. After confirmation of adequate signal capture, the implantable loop recorder should be programmed to detect arrhythmias on an individual basis. 538 Some manufacturers suggest an initial empirical programming as follows:

  1. Pauses: 3 s

  2. Bradycardia: heart rate ≤ 30bpm for more than 4 consecutive beats

  3. Tachycardia: heart rate > maximum predicted heart rate for age (220 - age) for 16 beats or more

  4. Atrial fibrillation: episodes > 2 minutes are characterized as AF rhythm

Careful adjudication of the episodes recorded by the implantable loop recorder is required, as false detections may have been stored. Episodes classified as AF, for example, may be misclassified due to RR interval variations resulting from ventricular extrasystoles or intermittent QRS undersensing. An analysis of 695 spontaneous or scheduled transmissions revealed a false-positive transmission rate of up to 81%. 539 Adequate programming of the detection criterion and subsequent analysis of tracings is essential for optimal monitoring.

6.5. Remote Monitoring (Online)

Online remote monitoring is a reality for patients with CIEDs. Data from the device can be transmitted via a broadband Internet connection. Newer devices allow data transmission via a Bluetooth connection in a smartphone. Remote monitoring allows access to several programming features, such as pacing rate and mode, pacing output, sensing and detection algorithms, as well as to diagnostic records and battery status.

Data transmission needs to be adjusted because information may be automatically transmitted by the device after a trigger or manually transmitted by the patient. A scheduled transmission may also be programmed. Remote access to device information is given to the service responsible for monitoring the patient through private access to the system’s server, provided that patients’ data privacy is protected.

7. Recommendations for Prevention and Treatment of CIED infecTions and for System Removal

7.1. Prevention and Treatment of Infections

Recent publications have shown an increased incidence of CIED-related infectious processes. Demographic and clinical factors, such as population aging and comorbidities, may influence both hematogenous seeding and direct contamination from device implantation and replacement. A recent EHRA survey reported that CIED infections are more frequent after reoperations, including those for isolated pulse generator replacement. 540

The most recent consensus statements have emphasized the need for standardizing management strategies and forming expert teams to address this particular and uncommon type of infection with the aim of mitigating the still frequent controversies over the topic. 541 - 546

CIED-related infectious processes manifest as either involvement of the pulse generator pocket or exclusive intravascular involvement. Exclusive pocket involvement is more frequent and accounts for approximately 60% of cases (generally resulting from contamination during surgery or subsequent manipulation). Late skin erosion may be due to or result in pocket infection; in both cases, this may progress to a systemic infection. Pocket involvement associated with intravascular infection accounts for approximately 20% of infections and is usually secondary to delayed or inadequate management. Exclusive intravascular involvement also represents approximately 20% of cases and results from bloodstream contamination in most cases. 547 Such contamination may occur during bacteremia caused by a distant infectious focus, such as septic thrombophlebitis, osteomyelitis, pneumonia, surgical site infection, contaminated vascular catheters, or bacterial infection originating from the skin, mouth, or gastrointestinal or urinary tract.

An expert consensus statement issued by the EHRA and endorsed by other international societies aimed to define the terminology that should be used in clinical studies and registries for the therapeutic approach to CIED infections and for system removal. 548 Table 39 shows the terminology recommended for different clinical presentations.

Table 39. – CIED-related infection types.

Infection types
  Clinical scenarios Definition
Local Superficial incisional infection Involves only skin and subcutaneous tissue, not the CIED
Isolated pocket infection Presence of clinical signs of inflammation limited to the pulse generator pocket (erythema, warmth, fluctuance, wound dehiscence, tenderness, or purulent drainage) with negative blood cultures
Isolated pocket erosion Complete or partial extrusion of pulse generator or lead through the skin
Systemic Bacteremia Positive blood cultures with or without systemic infection symptoms and signs
Pocket infection or erosion with associated bacteremia Generator pocket infection or erosion with positive blood cultures, without lead or valvular vegetation(s)
CIED-related endocarditis without pocket infection Bacteremia and lead or valvular vegetation(s) without pocket infection
Pocket infection with lead/valvular endocarditis Bacteremia and lead or valvular vegetation(s) with pocket infection
Occult bacteremia with probable CIED infection Bacteremia without an obvious source other than the CIED

CIED: cardiac implantable electronic device.

Table 40. – Modified criteria for diagnosis of CIED infections.

Definite generator pocket infection Presence of one of the following findings: 1) edema, erythema, warmth, pain, and purulent discharge; 2) fistula formation; 3) deformation, adherence, or threatened erosion of the skin; 4) generator or lead exposure
Definite CIED-related endocarditis Presence of 1) 2 major criteria; 2) 1 major and 3 minor criteria
Possible CIED-related endocarditis Presence of 1) 1 major and 1 minor criteria; 2) 3 minor criteria
Rejected ICED-related endocarditis Absence of the aforementioned criteria

CIED: cardiac implantable electronic device.

Table 41. – Major and minor criteria for diagnosis of CIED infections.

MAJOR CRITERIA
Microbiology A. Blood cultures positive for typical microorganisms found in CIED infection and/or IE (coagulase-negative staphylococci, S. aureus )
B. Microorganisms consistent with IE from 2 separate blood cultures:
a. Streptococcus viridans, Streptococcus gallolyticus ( S. bovis ), HACEK group, S. aureus
b. Community-acquired enterococci, in the absence of a primary focus
C. Microorganisms consistent with IE from persistently positive blood cultures:
a. ≥ 2 positive blood cultures of blood samples drawn > 12h apart
b. All of 3 or a majority of ≥ 4 separate blood cultures (first and last samples drawn ≥ 1 h apart)
c. Single positive blood culture for Coxiella burnetii or phase I IgG antibody titer > 1: 800
Imaging positive for CIED infections and/or IE A. Echocardiogram (including intracardiac) positive for:
a. CIED infection:
I. Clinical pocket/generator infection
II. Lead vegetation
b. Valve IE
I. Vegetations
II. Abscess, pseudoaneurysm, intracardiac fistula
III. Valvular perforation or aneurysm
IV. New partial dehiscence of prosthetic valve
B. [18F]FDG PET/CT (caution should be taken in case of recent implants) or radiolabeled WBC SPECT/CT detection of abnormal activity at pocket/generator site, along leads, or at valve site
C. Definite paravalvular leakage by cardiac CT
MINOR CRITERIA
  a. Predisposition such as predisposing heart condition (eg, preexisting structural heart defects, new-onset tricuspid valve regurgitation) or injection drug use
b. Fever (temperature > 38°C)
c. Vascular phenomena (including those detected only by imaging): major arterial emboli, septic pulmonary embolisms, infectious (mycotic) aneurysm, intracranial hemorrhage, conjunctival hemorrhages, and Janeway lesions
d. Microbiological evidence: positive blood culture which does not meet a major criterion as noted above or serological evidence of active infection with organism consistent with IE or pocket culture or lead culture (extracted by noninfected pocket).

CIED: cardiac implantable electronic device; IE: infective endocarditis; PET/CT: positron emission tomography/computed tomography; WBC SPECT/CT: white blood cell single-photon emission tomography/computed tomography; Janeway lesions: painless hemorrhagic lesions due to septic emboli.

A definite diagnosis of CIED infection is based on three major findings: 1) presence of purulent drainage or CIED exposure on clinical examination; 2) growth of microorganisms in blood cultures, and 3) presence of tricuspid valve or lead vegetations seen on transesophageal echocardiogram (TEE). When the diagnosis of CIED infection cannot be defined using these criteria, additional tests (eg, PET-CT) may be needed. The modified Duke University criteria for diagnosing CIED infections are listed in Tables 40 and 41.

Proof that the CIED is definitely contaminated is essential for proper treatment because, once contamination is confirmed, complete device removal will be necessary for a successful outcome. Conversely, if the CIED is free from contamination and the infectious process is related to a different focus, unnecessary device removal will imply avoidable cost and surgical risk related to lead extraction. A flowchart for diagnosing and treating CIED infections is shown in Figure 5 .

Figure 5. – Flowchart for diagnosis and treatment of CIED infections. CIED: cardiac implantable electronic device; FDG: fluorodeoxyglucose; PET/CT: positron emission tomography/computed tomography; TEE: transesophageal echocardiogram; TTE: transthoracic echocardiogram; WBC SPECT/CT: white blood cell single-photon emission tomography/computed tomography.

Figure 5

Imaging tests are useful for both diagnosis and treatment. Thus, some imaging information may be relevant, including 1) identification of CIED type; 2) identification of abandoned leads; 3) findings of intracardiac vegetations and their size; and 4) signs suggestive of septic pulmonary embolism.

In cases of fever in which a CIED pocket evaluation is unable to determine the infection, blood cultures or TEE and radiopharmaceutical-based imaging tests may be relevant.

Although complete removal of the pulse generator and all leads is essential, the infection must be treated with antimicrobial therapy. Antibiotic choice should be based on blood cultures and removed pocket/lead fragment cultures. When the microorganism cannot be identified, empirical antibiotic use should be defined by clinical criteria. Likewise, treatment duration should also be defined according to the clinical status, and the starting point should always be complete CIED removal ( Figure 6 ).

Figure 6. – Treatment duration and need for system removal in different presentations of CIED infection. CIED: cardiac implantable electronic device.

Figure 6

Complete CIED removal is critical to preventing recurrent infections. Lead extraction, however, should rarely be considered an emergency, even in cases of septic shock. With the exception of recent implants, which tend to be technically easier, extraction should only be performed when the hemodynamic status and the infection have stabilized, given the risks associated with the procedure (vein and heart adhesions).

Transvenous lead extraction should be the preferred technique, except when leads are epicardial or when intracavitary vegetations are greater than 2.5cm (largest diameter). Recommendations for removing the pulse generator and leads are listed in Table 42 .

Table 42. – Recommendations for CIED removal.

  Class of recommendation Level of evidence
Complete system removal in patients with definite CIED infection (systemic or local) I B
Complete CIED removal in patients with bacteremia with S. aureus, CoNS, Cutibacterium spp., or fungemia with Candida spp., after other infection sites are ruled out I C
Complete CIED removal in patients with bacteremia with Pneumococcus spp. or nonpseudomonal/Serratia gram-negative bacteria, in the presence of recurrent/persistent bacteremia despite appropriate antibiotic therapy when there is no other identifiable source for recurrence or persistent infection I C
Complete CIED removal in patients with infective endocarditis with or without definite involvement of the CIED system IIa C
Complete CIED removal in patients with bacteremia with alpha- or beta-hemolytic Streptococcus spp. or Enterococcus spp. performed as first-line treatment or, in case of recurrent/persistent bacteremia despite appropriate antibiotic therapy, as second-line treatment IIb C

CIED: cardiac implantable electronic device.

Implantation of a new CIED should only be performed after complete remission of the infectious process, based on the clinical status. Until the infection has completely resolved, patients dependent on artificial pacing should maintain treatment with a temporary PM. Nondependent patients should remain under cardiac rhythm monitoring until the implant is performed. In some cases, implantation of a new CIED may not be necessary because of a change in the disease pattern or a revision in the management strategy. Therefore, reassessing the need for a CIED is always essential. Recommendations for implantation of a new CIED are listed in Table 43 .

Table 43. – Recommendations for implantation of a new CIED.

  Class of recommendation Level of evidence
Reassessment of the indication for reimplantation after device extraction I C
Preferred access sites for implantation of a new device are the contralateral side, the femoral vein, or an epicardial approach I C
Whenever possible, reimplantation should be avoided or delayed until symptoms and signs of systemic and local infection have resolved IIa C
A temporary pacemaker with ipsilateral active fixation strategy in pacemaker-dependent patients as they wait for reimplantation IIa C

CIED: cardiac implantable electronic device.

Several risk factors for the development of CIED infection have been reported. These factors may be related to the patient, the medical procedures performed, or the CIED itself. Major factors for CIED infection are listed in Table 44 .

Table 44. – Risk factors predisposing patients to CIED infection.

Patient-related factors
End-stage renal diseasea
History of device infection
Fever prior to implantation
Corticosteroid use
Renal insufficiencyb
Chronic obstructive pulmonary disease
NYHA class ≥II
Skin disorders
Malignancy
Diabetes mellitus
Heparin bridging
Heart failure
Oral anticoagulants
Procedure-related factors
Procedure duration
Hematoma
Lead repositioning
Inexperienced operatorc
Temporary pacing
Device replacement/revision/upgrade
Generator change
Antibiotic prophylaxisd
Device-related factors
Epicardial leads
Abdominal pocket
≥ 2 leads
Dual-chamber device

NYHA: New York Heart Association. Risk parameters which were statistically significant for retrospective and prospective data are shown. Analyses restricted to prospective data only for the same parameters (if available) are also shown. Adapted from Polyzos et al.549 aGlomerular filtration rate (GFR) < 60mL/min or creatinine clearance (CrCL) < 60mL/min. bGFR ≤15 mL/min or hemodialysis or peritoneal dialysis. c <100 previous procedures. dThe pooled effect estimate from randomized studies was 0.26 [0.13, 0.52].

Preventive care is crucial to reducing the occurrence of procedure-related infections. Table 45 and Figure 7 show the most commonly recommended preventive measures.

Table 45. – List of recommended preventive measures for CIED infections.

  Class of recommendation Level of evidence
Preprocedural measures
Delay CIED implantation in patients with infection I C
Avoid temporary transvenous pacing and central venous access whenever possible. If used, they should ideally be removed prior to introducing a new device I A
Measures to avoid pocket hematoma: discontinue antiplatelet agents whenever possible and, in the case of oral anticoagulants, avoid heparin bridging, discontinuing the use during implantation if possible I B
Antibiotic prophylaxis is recommended within 30min to 1h of incision for cefazolin and within 90-120min for vancomycin I A
Periprocedural measures
Use of antibiotic envelope in high-risk situations for infection* IIa B
Instillation of antiseptic and/or antibiotics in the generator pocket IIb C
Postprocedural measures
Use of postoperative antibiotic therapy IIb B
Hematoma drainage or evacuation (except in the presence of tense tissue, wound dehiscence, or severe pain) III B

* As defined in the WRAP-IT study:550 patients undergoing pocket or lead revision, generator replacement, system upgrade, or initial CRT-D implantation, and those with risk factors as listed in Table 44. CIED: cardiac implantable electronic device.

Figure 7. – Preventive measures for reducing CIED infections considering modifiable and nonmodifiable risk factors as determined in the literature. CIED: cardiac implantable electronic device; COPD: chronic obstructive pulmonary disease; CRT: cardiac resynchronization therapy; ICD: implantable cardioverter-defibrillator; NYHA: New York Heart Association; OAC: oral anticoagulation.

Figure 7

Table 46. – Definitions for extraction approaches and techniques.

Technique Definition
Transvenous Intravascular (percutaneous) lead extraction performed through a central vein (subclavian, jugular, or femoral)
  • - Venous entry site approach

Lead extraction by traction or countertraction using the same vein and site through which the lead was introduced
  • - Transjugular or transfemoral approach

Extraction with an intravascular tool introduced through right or left jugular or femoral veins
Transthoracic Lead extraction by opening the thoracic cavity using median sternotomy, lateral thoracotomy, or subxiphoid incision, including a hybrid approach, with or without cardiopulmonary bypass

Older people, children, and adults with congenital heart disease constitute subgroups that deserve special attention regarding CIED infections. Submuscular positioning of pocket in patients with limited subcutaneous tissue is essential to preventing skin erosion. In pediatric patients, especially those with congenital heart disease, the operator should be experienced in multiple and alternative surgical approaches. Extravascular or subcutaneous ICD implantation should be considered in younger children, patients with congenital heart disease, and those with limited or no venous access. 551 , 552

Interestingly, retrospective registries 553 , 554 have reported higher CIED infection rates than those of prospective studies 537 , 555 - 558 (3.4% vs 1.2%). This phenomenon may reflect greater adherence to preventive procedures in clinical studies compared to daily clinical practice. The most important procedure-related factors for risk of infection include pocket hematoma, long procedure duration, and reinterventions for lead repositioning. Regarding reoperations for pulse generator replacement, lead dysfunction correction, or pacing mode change, appropriate management of the pulse generator pocket, either by complete removal of the fibrotic capsule or by use of antibacterial envelopes, reduces the number of infections.

Preoperative antibiotic prophylaxis with a single dose of first-generation cephalosporins (cefazolin) is strongly recommended, which is not the case with systematic postoperative antibiotic use. 11

The time interval between diagnosis and appropriate treatment of CIED infections is critical. Literature data show that if the device is removed within 3 days of hospitalization, both length of stay and in-hospital mortality are significantly reduced. Thus, when there are no sufficient data to establish the diagnosis of infection based on the triad of infectious signs in the pulse generator pocket, bacterial growth in blood cultures, and vegetations on TEE, additional resources should be employed.

[ 18 F]FDG PET/CT scanning or radiolabeled WBC scintigraphy or contrast-enhanced CT are recommended in cases of suspected CIED-related infective endocarditis, positive blood cultures, and negative echocardiogram, or in patients with S. aureus bacteremia in the presence of CIED. Needle aspiration and surgical debridement in cases of generator pocket infection, in an attempt to avoid lead extraction, should be strongly discouraged.

Large, nonvoluntary, easy-to-fill, high-quality registries are essential to monitoring the number of cases of CIED infection and the outcomes of preventive and therapeutic measures. Each center should establish routines for accurate diagnosis and timely treatment. Constant reassessment of the performance of each center is highly recommended.

7.2. Lead Removal from Cardiac Implantable Electronic Devices

There has been a growing demand for CIED lead removal in recent years because of two major factors: 1) increase in CIED infection rates and 2) development of multisite PM and ICD systems with a higher number of leads.

Indications for lead removal may be (a) mandatory for treating infections; (b) necessary for obtaining access for new leads in patients with venous occlusions; or (c) optional for performing lead replacement in patients with appropriate venous access.

As most devices require venous access for lead implantation, transvenous extraction techniques are most commonly used. Currently, open-chest lead removal is rarely performed as few scenarios require this method, such as removal of epicardial leads or correction of complications occurring in transvenous extraction procedures.

The current panorama of lead extraction shows well-established indications and operative techniques. The tools used are well developed, and the outcomes of different procedures are well known, with high success rates. Catastrophic complications, however, may occur during extraction procedures. Such complications, although rare, are potentially lethal and often require emergency open surgery.

This item presents the recommendations for extraction in noninfected patients, as the management of infections has been previously addressed.

Epicardial lead removal is necessarily performed by reopening the thoracic cavity preferably using the same access through which the lead was implanted. Transvenous lead removal should preferably be done by intravascular access. Exceptionally, a transthoracic approach may be used with or without cardiopulmonary bypass (eg, failure of transvenous extraction or presence of large lead vegetations). The choice of transvenous lead extraction approach depends essentially on the possibility of obtaining access to the lead targeted for removal. Unfortunately, many patients have entirely intravascular leads because of spontaneous lead fractures or iatrogenic events during removal procedures.

When the lead to be removed is intact or has an extravascular segment, however small this segment may be, the venous entry site approach must be used. This approach consists of introducing a sheath into the vein which is guided by the lead to be removed. This sheath is used to cut through adhesions that form between the lead and the venous endothelium or the endocardium. When the patient has more than one transvenous lead implanted, adhesions are often seen between the leads. After all adhesions have been removed and the site where the lead is attached to the heart has been reached, the same sheath is used to apply pressure against the heart muscle while pulling the lead (countertraction maneuver). There are several tools specifically developed for this type of approach, namely:

  • Locking stylets, which are stylets coated with a fine steel mesh that expands in the light of the lead, providing the lead with the necessary support for its traction.

  • Sheaths for adhesion dissection and countertraction:

  • Nonpowered sheaths are sets of rigid metal and flexible Teflon or polypropylene tubes that cut through adhesions by blunt dissection, with an intensity level determined by the strength of the operator’s hand.

  • Rotational mechanical sheaths are activated by a trigger in the operator’s hand or by an electric motor and cut through adhesions.

  • Laser sheaths use photoablation to cut through adhesions.

When the lead that needs to be removed does not have an extravascular segment, intravascular extraction is mandatory. There are tools in the shape of a loop or basket made with very malleable metal wires and designed to grasp these fragments. Such tools are usually introduced by puncturing the femoral or jugular veins. Once grasped, the lead can be pulled directly. Specific cases may require a combined countertraction maneuver after grasping the lead. 559 Tables 46 and 47 show recommended definitions for tools and approaches used in lead extraction.

The term removal has been generically used to refer to CIED lead removal regardless of the type of approach. It can be performed by simple traction of the transvenous lead without using any tools; by thoracotomy for removing epicardial leads; or by thoracotomy with cardiopulmonary bypass for removing transvenous leads. The term extraction should be reserved for cases requiring the use of techniques and tools to 1) dilate the venous path through which the lead passes; 2) cut through adhesions; 3) perform a countertraction maneuver; or 4) grasp lead fragments inside the heart vessels or chambers.

Completion of a lead removal or lead extraction procedure may result in 1) complete removal of the targeted lead; 2) partial removal; or 3) unsuccessful removal. Depending on the type of indication for lead removal, the procedure may be considered clinically successful even if the lead was not entirely removed. The procedure is considered unsuccessful when 1) clinical success is not achieved; 2) any permanently disabling complication occurs; or 3) the patient dies. Table 48 lists the definitions established by the 2018 EHRA consensus statement and endorsed by other representative societies.

Table 48. – Definitions for lead removal procedures and outcomes.

Concept Definition
Procedure  
Lead removal Lead removal using simple traction techniques and no specialized tools
Lead extraction Removal of at least one lead using specialized tools (listed in Table 47)
Outcome  
Lead-related  
  • - Complete lead removal

Lead explant or extraction with removal of all lead material
  • - Incomplete lead removal

Lead explant or extraction in which part of the lead remains in the patient’s body
Patient-related  
  • - Complete procedural success

Removal of all targeted leads, with no permanently disabling complication or procedure-related death
  • - Clinical procedural success

Retention of a small portion of a lead that does not negatively impact the outcome of the procedure. This may be the tip or a small part (< 4cm) of the lead (conductor coil, insulation, or both) provided that the fragment does not increase the risk of perforation, embolic events, or perpetuation of infection, in the absence of any permanently disabling complication or procedure-related death
  • - Procedural failure

Inability to achieve clinical success, or the development of any permanently disabling complication or procedure-related death

Table 47. – Definitions for extraction tools.

Tools and techniques Definition
Sheath for dissection and countertraction  
  • - Mechanical nonpowered

Sets of rigid metal and flexible Teflon or polypropylene tubes
  • - Rotational mechanical 548

Metal-cutting blades driven by manual or electrical activation
  • - Laser 560

Glass microtubes conducive of the light produced by an (external) emission source of the Excimer laser
Locking stylet Stylets coated with a steel mesh that provide the leads with rigidity and support
Intravascular extraction tools Snares, 561 baskets, and other tools used to grasp intravascular fragments
Vascular occlusion balloons 562 Sets of guidewires and balloons for venous dilation or occlusion
Accessory tools Lead extenders, compression coils 563

Leads may be deactivated for several reasons, such as 1) loss of ability to adequately stimulate the heart, 2) need to change the device type, and 3) manufacturing-related problems.

Noninfected leads may be abandoned in situ at the discretion of the surgical team. There are, however, disadvantages to abandoning a lead, including 1) risk of thrombotic phenomena, 2) limitations to MRI scans, and 3) increased risk in a future extraction procedure, since the longer the lead remains, the higher the risk of unsuccessful extraction. The main argument for abandoning a noninfected lead in situ is the risk of severe complications associated with the extraction procedure. A recent publication shows that 1-year expected survival following a lead replacement procedure is similar for cases of lead extraction vs in situ abandonment. 564 Therefore, the decision to extract or not a lead that will be deactivated essentially depends on the expertise of each center. Table 49 shows the EHRA 2018 consensus definitions for noninfected leads.

Table 49. – Definitions of terms for noninfected leads.

Noninfected leads Definitions
Lead function Any lead function, including pacing, sensing, and/or defibrillation
Lead failure Loss of any lead function
Nonfunctional lead Lead not usable for pacing and/or defibrillation due to loss of functional integrity
Abandoned lead Lead left in place in the heart and not connected to a CIED. It may be functional or nonfunctional
Recall558,565-568 Removal or correction of a lead due to manufacturing-related problems, regulatory agency requirements, or voluntary manufacturer guidance
  • - Class I

Dangerous or defective products with reasonable probability of causing serious health problems or death (eg, short circuit without warning)
  • - Class II

Products that might cause a temporary health problem or pose a slight threat of a serious nature (eg, premature battery depletion)
  • - Class III

Products that are unlikely to cause any adverse health reaction but that violate regulatory standards

CIED: cardiac implantable electronic device.

CIED durability depends on both manufacturing (material and/or design) and usage aspects. Leads are directly influenced by the operative technique used, which may negatively impact their durability. Nonetheless, specific surveillance strategies must be adopted by manufacturers and regulatory agencies to assess the durability of CIED components.

In cases of in situ lead abandonment due to a dysfunction or need to change the pacing mode, or, inappropriately, in cases of treatment of a CIED infection, caution is required because how leads are abandoned may hinder future extraction procedures. Recommendations for lead abandonment are listed in Table 50 .

Table 50. – Recommendations for managing abandoned, unnecessary, or dysfunctional leads.

  Class of recommendation Level of evidence
When a lead is abandoned, it should be left in a condition that prevents retraction inside the vein and that allows extraction in the future I C
In cases of clinically unnecessary or dysfunctional leads, both options (abandonment and extraction) can be considered within the surgical strategy569-571 IIa B

Noninfected lead removal may be mandatory in certain clinical situations, such as for 1) treatment of superior vena cava syndrome caused by the presence of leads; 2) treatment of severe cardiac arrhythmia mechanically caused by a lead fragment; 3) prevention of cardiac injury from a fractured lead; or 4) radiation therapy in the region where the device is implanted.

Also, lead extraction may be necessary in cases of severe venous occlusion or obstruction preventing the passage of a new lead.

In many cases, however, lead extraction is optional and can be defined by evaluation of less objective factors, such as 1) patient’s age or estimated life expectancy; 2) future need for MRI scans; 3) risk of developing severe venous obstructions; or 4) risk of infection via a hematogenous route (eg, patients with renal failure on dialysis). Such situations require the expertise of the professional who performs the procedure for defining an appropriate management strategy. Table 51 lists the recommendations for removal of noninfected leads.

Table 51. – Recommendations for lead removal in noninfectious situations.

  Class of recommendation Level of evidence
Thrombosis/Vascular involvement    
Clinically significant embolic events attributable to the presence of a lead or lead fragment572-574 I C
SVC stenosis or occlusion575preventing the passage of a new lead I C
Need for stenting to avoid lead entrapment576,577 I C
Others    
Lead- or fragment-induced arrhythmias578 I C
Device implanted in a site that interferes with the treatment of a malignant disease579 IIa C
Lead overpopulation: > 4 leads on one side or > 5 leads across the SVC569,580,581 IIa C
Abandoned leads interfering with the operation of an implanted device582,583 IIa C
Leads that due to their design or failure pose a threat or harm to the patient if left in place584-587 IIb C
Contraindications to magnetic resonance imaging: abandoned leads or fragments; nonconditional leads for magnetic resonance imaging588-593 IIb C
Permanent device removal by shared decision IIb C

SVC: superior vena cava.

During transvenous lead extraction procedures, veins or cardiac structures may be injured. Injuries to the axillary or subclavian veins, brachiocephalic veins, or superior vena cava may cause severe hemorrhage requiring blood transfusion or even surgical correction. Muscle avulsion of the right atrium or RV and perforation of coronary sinus tributaries may lead to cardiac tamponade. Extrapericardial laceration of the superior vena cava, however, is the most frequent and most lethal catastrophic complication. Other complications, such as self-limiting cardiac arrhythmias, pneumothorax, or lead fragment retention, may also occur and require specific care.

Complications are generally grouped into major and minor according to severity and type of correction required. Table 52 lists complications according to their classification and incidence. 577

Table 52. – Classification and incidence of the most frequent perioperative complications*.

Complications Incidence, %
Major 0.19–1.80
Death 0.19–1.20
Cardiac avulsion 0.19–0.96
Vascular laceration 0.16–0.41
Respiratory arrest 0.20
Cerebrovascular accident 0.07–0.08
Pericardial effusion requiring intervention 0.23–0.59
Hemothorax requiring intervention 0.07–0.20
Cardiac arrest 0.07
Thromboembolism requiring intervention 0.07
Flail tricuspid valve leaflet requiring intervention 0.03
Massive pulmonary embolism 0.08
Minor 0.06–6.20
Pericardial effusion without intervention 0.07–0.16
Hematoma requiring evacuation 0.90–1.60
Venous thrombosis requiring medical intervention 0.10–0.21
Vascular repair at venous entry site 0.07–0.13
Migrated lead fragment without sequelae 0.20
Bleeding requiring blood transfusion 0.08–1.00
Arteriovenous fistula requiring intervention 0.16
Pneumothorax requiring chest tube 1.10
Worsening tricuspid valve function 0.02–0.59
Pulmonary embolism 0.24–0.59

* Source: adapted from Kusumoto et al., 2017.594

Several studies have been designed to identify risk factors that determine morbidity and mortality in transvenous lead extraction procedures. These studies have reported low rates of catastrophic complications and perioperative death, not allowing proper identification of those risk factors. Conversely, several demographic, clinical, and surgical factors are associated with 30-day mortality following an extraction procedure. Factors associated with late complications and death have also been described. 30 , 595 - 603 Table 53 lists the identified risk factors and their impact on morbidity and mortality.

Table 53. – Risk factors for death and complications associated with lead removal.

Factor Risk associated with factor
Age Increases mortality by 1.05 times
Female sex Increases risk of major complications by 4.5 times
Low body mass index (<25 kg/m2)
  • Increases 30-day mortality by 1.8 times

  • Increases number of extraction-related complications

Stroke history Increases risk of major complications by 2.0 times
Severe LV dysfunction Increases risk of major complications by 2.0 times
Advanced HF
  • Increases 30-day mortality risk by 1.3 to 8.5 times

  • Increases 1-year mortality by 3.0 times

Renal dysfunction
  • End-stage renal disease increases 30-day risk of death by 4.8 times

  • Creatinine > 2.0 increases in-hospital mortality and 1-year mortality by 2.0 times

Diabetes mellitus Increases in-hospital mortality and overall mortality by 1.7 times
Coagulopathy
  • High INR increases risk of major complications by 2.7 times and 30-day mortality by 1.3 times

  • Anticoagulants increase 1-year mortality by 1.8 times

Platelet count Thrombocytopenia increases risk of major complications by 1.7 times
Anemia Increases 30-day risk of death by 3.3 times
Number of extracted leads
  • Increases risk of any complications by 3.5 times

  • Increases long-term mortality by 1.6 times

Presence of dual-coil ICD Increases 30-day mortality by 2.7 times
Extraction for infection
  • Increases 30-day mortality by 2.7 to 30 times Increases 1-year mortality by 5.0 to 9.7 times CRP > 7.2mg/L

  • increases 30-day mortality Increases overall mortality by 3.5 times

Operator’s experience Increases number of procedure-related complications by 2.6 times
Previous heart surgery Reduces incidence of major complications

CRP: C-reactive protein; ICD: implantable cardioverter-defibrillator; INR: international normalized ratio; LV: left ventricle; HF: heart failure.

Given the difficulty of predicting perioperative catastrophic complications, prevention of associated deaths becomes crucial, which implies training the staff and providing centers with the technical skills required for lead extraction. A recent systematic review 604 shows the close relationship between the volume of procedures performed at the center and the rate of complications associated with lead extraction. Naive operators should be supervised by more experienced operators during the first 40 transvenous extraction procedures. A minimum volume of 20 transvenous extraction procedures per year is recommended for all operators to maintain their technical skills.

8. Recommendations for the prevention of electromagnetic interference

8.1. Surgery Using Electrocautery

Electrosurgery uses high-frequency alternating current (200kHz to 2.2MHz), which is converted into heat when passing through tissue with sufficient resistance, allowing the desired effects to be achieved: coagulation and cutting. An electric scalpel is used in most surgical specialties.

Monopolar electrosurgery is the most effective and therefore most widely used technique in surgical practice. In this modality, the active electrode is located at the surgical site (on the surgical instrument), whereas the indifferent (return) electrode is a plate placed on the patient’s skin at a distance. The current flows between the electrodes, passing through the body.

An increasing number of patients with CIEDs are treated surgically, which exposes these patients to electromagnetic interference. Monopolar electrosurgery can cause a number of CIED abnormalities, such as pulse generator reprogramming, temporary pacing inhibition, high-frequency pacing triggering, battery depletion and pacing failure, circuit damage, threshold elevation, and triggering of inappropriate therapy (shocks) in the case of ICDs. 605

To minimize the risks of using electrocautery, some precautions should be taken perioperatively: (1) the monopolar probe should be used intermittently, with short bursts of current and low energy levels; (2) the indifferent plate should be positioned so that the current does not flow through the generator or electrodes.

In general, when the surgical site is located above the umbilicus or at a distance of less than 15 cm from the generator, the use of a monopolar probe should be avoided. In this scenario, a bipolar probe should be preferred as it is safer, but it should not be applied directly to the generator. 606 , 607 In head and neck surgery, the indifferent plate of the monopolar probe should be placed on the posterior shoulder contralateral to the device pocket. For example, if the generator is placed on the left infraclavicular region, the probe plate should be placed on the right shoulder.

In order to protect both patients and CIEDs from the undesirable effects of electrocautery, two approaches have been used: placing a magnet over the pulse generator and reprogramming the device before the procedure. In the case of PMs, magnet use during surgery is an option when circuit sensing has been deactivated by the generator under magnetic effect (asynchronous mode) and the battery is in good condition. 608 , 609 Placing a magnet over the generator pocket causes the PM to revert to asynchronous mode, that is, it disables the sensors and changes to magnet pacing rate, which is often higher than the programmed pacing rate.

In the case of ICDs, a randomized trial compared magnet application versus reprogramming in patients with ICDs undergoing surgery using monopolar electrocautery, at a distance greater than 15 cm from the generator; the authors concluded that both strategies are safe. 610 Magnet application to the ICD pocket only disables tachyarrhythmia therapy, it does not change PM function. In PM-dependent patients with ICDs, the device should be reprogrammed to asynchronous mode before the procedure. CIED reprogramming should be performed immediately before the surgical intervention and reverted to original programming immediately after the end of the procedure ( Table 54 ).

Table 54. – Recommendations for the prevention of electromagnetic interference in surgery using electrocautery.

  Class of recommendation Level of evidence
Monopolar electrosurgery can cause several CIED abnormalities, and bipolar electrocautery should be used preferably In patients with ICDs who are not pacemaker-dependent, a magnet can be safely placed over the ICD pocket to disable tachyarrhythmia therapy and avoid inappropriate shocks I A
In the case of pacemakers, magnet use during surgery is an option when circuit sensing has been deactivated by the generator under magnetic effect (asynchronous mode) and the battery is in good condition I B
In head and neck surgery using a monopolar probe, the indifferent plate should be placed on the posterior shoulder contralateral to the device pocket In pacemaker-dependent patients with ICDs, therapies are deactivated and the device should be reprogrammed to asynchronous mode before the procedure CIED reprogramming should be performed immediately before the surgical intervention and reverted to original programming immediately after the end of the procedure I C
When the surgical site is located above the umbilicus, a bipolar probe should be used. If the use of a monopolar probe cannot be avoided, the generator should be reprogrammed IIa C

CIED: cardiac implantable electronic device; ICD: implantable cardioverter-defibrillator.

8.2. Magnetic Resonance Imaging

MRI has emerged as an increasingly useful and accessible diagnostic tool, with growing relevance for diagnostic and prognostic evaluation.

The number of MRI scans has grown substantially over the past 20 years, with over 60 million scans performed worldwide every year. It is estimated that, after CIED implantation, a patient has a 50% to 75% probability of being indicated for an MRI over the lifetime of their device. 594 , 611

The MRI environment can be divided into zones, as described by Kanal et al . 612 and adopted by the 2017 HRS expert consensus. 593 Zone 4 refers to the scanner room, which is the space with the greatest risk for patients and health care staff, including the potential risk of dislodgement of metal objects. Zone 3 is the space outside the scanner room, including the scan control room. Because of the potential risks in this area, access to it must be limited to only trained personnel. Zone 2 includes the reception area, and zone 1 corresponds to areas freely accessible to the general public.

During MRI scanning, CIEDs can be defined as: 613

  • MRI safe : devices that pose no hazards for MRI scanning.

  • MRI conditional : devices that pose no hazards for MRI scanning, as long as specified conditions of use are met. These conditions include parameters such as: region of the body to be scanned, magnetic field strength, spatial gradient, magnetic field exposure time, radiofrequency field, and specific absorption rate. Additional conditions might be required, including the use of specific generator and electrode combinations as well as device programming mode. Specific conditions may vary among manufacturers and among devices from the same manufacturer.

  • MRI nonconditional : devices that pose hazards for MRI scanning. They include all CIED systems that are nonconditional for MRI scanning, such as MRI-conditional generators combined with nonconditional leads or MRI-conditional systems implanted in patients who do not meet all specified conditions of use, including patients with abandoned leads.

No CIED system is classified as MRI safe , and new CIEDs that have been developed with appropriate technology are considered MRI conditional for MRI scanning.

In clinical practice, devices need to be programmed before MRI scanning, with a highly acceptable level of safety.

Interactions of the static and gradient magnetic fields with radiofrequency on CIEDs can impair the functioning of electronic components, cause migration or dislodgement of system components, generate energy currents that might damage the device and/or the myocardium, and cause oversensing, undersensing, or arrhythmias. The influence of these parameters on CIEDs can be divided into two groups: transient and permanent impairment of the CIED operation. 614 , 615

Responses to these sources of interference may vary:

  • Static magnetic field : device dislodgement, sensor activation, sudden loss of device function, ECG changes.

  • Gradient magnetic field : induction of arrhythmias (rare), oversensing or undersensing.

  • Radiofrequency field : heating of tissue adjacent to lead electrodes, induction of arrhythmias (rare), device reprogramming (reset), oversensing or undersensing interactions.

  • Combined field effects : sudden loss of device function, alteration of device function (parameters), mechanical forces (vibration), device reset, damage to generator and/or leads.

  • Imaging-related : artifacts that prevent adequate device image visualization.

Potential interactions between CIEDs and MRI electromagnetic interference include:

  • Magnetic field-induced force and torque due to ferromagnetic materials : generator movement is extremely unlikely due to confinement and adjacent subcutaneous tissues. Leads do not contain sufficient ferromagnetic material to cause movement.

  • Gradient magnetic field-induced electrical current : gradient magnetic fields can induce current, which can lead to myocardial capture and potentially cause atrial or ventricular arrhythmias.

  • Heating and tissue damage : radiofrequency fields can lead to nonconditional CIED component heating, causing heating of and thermal damage to the adjacent tissue (functional ablation). Changes in sensing or capture thresholds can occur as a result of tissue damage near lead electrodes.

  • Effects on device activity : the CIED can be programmed by placement of a magnet, thus allowing device interactions. Magnetic fields might therefore affect the activity of a nonconditional device, possibly changing the programming of the device.

  • Electrical reset : High-energy electromagnetic interference can lead to electrical power-on reset, and a backup demand mode may be activated. Power-on reset parameters vary among vendors and device types and can include a set of variations. Inhibition of pacing function by MRI-generated signals or pacing at an output below threshold (bipolar or unipolar) in a PM-dependent patient may occur. Additionally, battery status can be affected, particularly for devices that are near an elective replacement interval (ERI), which may result in unreliable function.

  • Inappropriate function and therapies : electromagnetic interference from radiofrequency energy pulses or rapidly changing magnetic field gradients might cause oversensing that can lead to inappropriate pacing inhibition and possibility asystole in PM pacing-dependent patients, or induction of therapies leading to inappropriate shocks in patients with ICDs.

These effects are influenced by many factors, including magnetic field strength, radiofrequency power, position of the patient and device inside the MRI bore, device characteristics, and the size of the patient.

Traditionally, MRI scanning has been contraindicated in patients with CIEDs. The first MRI-conditional system was introduced in Europe in 2010 and approved by the Food and Drug Administration (FDA) in 2011 for use in the United States. 613 , 616

To render CIEDs MRI conditional, structural changes (eg, use of non-ferromagnetic materials) and software changes have been made to reduce or eliminate potential adverse effects. Once the special programming mode (MRI mode) has been activated, the device reverts to an asynchronous pacing mode and increases the pacing outputs to avoid inhibition of pacing during MRI scanning. In ICDs, the antitachycardia function is temporarily disabled. Therefore, patients with ICDs will be unprotected from ventricular arrhythmias during MRI scanning.

Because the decision to perform MRI scanning in a patient with a CIED system involves risks and benefits, potential risk factors should be identified. Patients with MRI-conditional CIEDs may undergo MRI scanning without additional risks if established recommendations and protocols are followed. 617 , 618

Before MRI scanning, it is important to identify the patient’s baseline rhythm and whether the patient is PM dependent, activate the specific MRI programming mode, confirm that the entire system is MRI conditional, and check for the presence of abandoned or epicardial leads.

In general, most CIED systems have been approved for MRI scanning with 1.5T, a gradient slew rate of 200 T/m/s, a maximum specific absorption rate of 2 W/kg, and a limited number and length of imaging sequences. New devices allow safe MRI scanning under broader conditions. Most new systems allow full-body MRI scanning.

An MRI-conditional system consists of a combination of leads and generator that has been specifically tested to ensure safe conditions of use during MRI scanning. The presence of any device component that does not meet the criteria for MRI conditionality renders the CIED MRI nonconditional. This includes an MRI-conditional generator combined with nonconditional components and device systems that combine individual MRI lead components and MRI-conditional generators from various manufacturers, as these are not combinations specifically tested together for MRI scanning safety.

Conditional labeling also specifies the location of the generator (eg, pectoral location for a transvenous system). Other examples of nonconditional components include epicardial leads, abandoned leads, fractured leads, and active noncardiac devices.

Programming of the device outside the MRI-conditional programming mode also renders the device MRI nonconditional. Battery status must be adequate to consider the device MRI conditional ( Table 55 ).

Table 55. – Recommendations for the prevention of electromagnetic interference in MRI scanning.

  Class of recommendation Level of evidence
  • MRI-conditional CIEDs should be considered MRI conditional only when the product labeling is adhered to, which includes programming the appropriate “MRI mode” and scanning with the prerequisites specified for the device MRI in a patient with an MRI-conditional system should always be performed in the context of a rigorously applied standardized institutional workflow, following appropriate technical conditions

  • It is recommended for patients with an MRI-conditional system that ECG and pulse oximetry monitoring be continued until patient observation is completed, or until other clinically appropriate device settings are restored

I A
  • It is recommended for patients with an MRI-nonconditional CIED that device evaluation be performed immediately before and after the scan, with documentation of pacing threshold, P- and R-wave amplitude, and lead impedance using a standardized protocol

  • A defibrillator/monitor (with external pacing function) and a manufacturer-specific CIED programming system should be immediately available in the area adjacent to the scanner room while the MRI-nonconditional device is programmed for imaging

  • It is recommended that continuous ECG and pulse oximetry monitoring be used while the MRI-nonconditional device is programmed for imaging

  • It is recommended that personnel skilled to perform advanced cardiac life support accompany the patient with an MRI-nonconditional CIED until assessed and declared stable to return to unmonitored status

  • For patients with an MRI-nonconditional CIED who are pacing-dependent (PM or ICD), it is recommended that: a) a physician skilled to implant a temporary PM be immediately available on the premises of the facility; b) a physician skilled to program the CIED be immediately available on the premises of the facility

  • It is recommended for patients with an MRI-unconditional CIED who are pacing-dependent that the device be programmed to an asynchronous pacing mode with deactivation of adaptive features (rate-response sensor) during MRI scanning. The appropriate pacing rate should be selected to avoid competitive pacing

  • All tachycardia detections for patients with an ICD should be disabled prior to MRI

I B
  • It is recommended for patients with an MRI-conditional system that personnel with advanced cardiac life support skills be in attendance. If advanced cardiac life support is required, the patient should be monitored for the duration of time the CIED is reprogrammed, or until the patient is assessed and declared stable to return to unmonitored status

  • The MRI-responsible physician should be informed of the presence of a patient with an MRI-nonconditional CIED

  • It is recommended that ECG and pulse oximetry monitoring be continued until the end of MRI scanning or until the device settings are reprogrammed

  • All resuscitative efforts and emergency treatments involving the use of a defibrillator/monitor, CIED programming system, or any other MRI-unsafe equipment should be performed after moving the patient outside zone 4

I C
  • It is reasonable for patients with an MRI-nonconditional system to undergo MRI if there are no fractured, epicardial, or abandoned leads, and MRI is the best test for the condition. There should be an institutional protocol and a responsible MRI physician and CIED physician

  • It is reasonable to perform MRI immediately after implantation of a lead or generator of an MRI-nonconditional device if clinically warranted

  • For a patient with an MRI-nonconditional CIED who is not pacing-dependent, it is possible to program the device to either a nonpacing mode (OVO/ODO) or to an inhibited mode (DDI/VVI), with deactivation of advanced or adaptive features during MRI scanning

IIa B
  • It is reasonable to perform MRI on a patient with an MRI-conditional system implanted more recently than the exempt period for conditionality of the system, based on the risk-benefit assessment for that patient

  • For patients with an MRI-nonconditional CIED, it is possible to perform repeat MRI when required, without restriction as to the minimum interval between imaging studies or the maximum number of studies performed

  • It is reasonable to program patients with an MRI-nonconditional CIED who are not pacing-dependent to an asynchronous pacing mode (VOO/DOO) with deactivation of advanced or adaptive features during scanning, and with a pacing rate that avoids competitive pacing

  • For patients with an MRI-nonconditional CIED, it is reasonable to schedule a complete device evaluation within 1 week for a pacing lead threshold increase > 1.0 V, P-wave or R-wave amplitude decrease > 50%, pacing lead impedance change > 50 Ω, and shock lead impedance change > 5 Ω

IIa C

CIED: cardiac implantable electronic device; ECG: electrocardiogram; ICD: implantable cardioverter-defibrillator; MRI: magnetic resonance imaging; PM: pacemaker.

8.3. Radiotherapy

An increasing number of patients undergoing radiotherapy have a CIED. Although radiotherapy-induced malfunction is rare, safety recommendations are important.

Ionizing radiation can interfere with the complementary metal oxide semiconductor (CMOS) components of the generator. The production of secondary neutrons is the strongest predictor of CIED malfunction in the setting of radiotherapy. Modern pulse generators have lower power consumption and smaller circuits, made of semiconductor metal. This renders modern devices more susceptible to possible damage caused by ionizing radiation. 596 , 619

High radiation doses, especially with energy > 6MV, can cause software and hardware errors. These disturbances are usually transient, such as pacing inhibition, sensing abnormalities, and inappropriate pacing at the maximum sensor rate. Reset to backup mode, which can be corrected with reprogramming, is one of the most reported malfunctions. Permanent damage to the device may also occur, such as loss of telemetry and premature battery depletion. CIED failure, with complete interruption of device functioning, has been described in vitro. 593 , 619 - 621

It is also important to consider that damage to the CIED may appear weeks or months after the end of radiotherapy (latent damage). 622 Device malfunction has been reported in up to 3% of radiotherapy courses. Clinically relevant events are rare and dependent on the type of device and patient’s tolerance to changes. For example, a PM-dependent patient may have bradycardia and associated symptoms. 593 , 619

Radiotherapy planning should consider the conditions specified for the CIED and patient characteristics, such as whether the patient is pacing-dependent or not and has a history of ventricular tachycardia or ventricular fibrillation (VT/VF) ( Table 56 ).

Table 56. – Recommendations for the prevention of electromagnetic interference in radiotherapy.

  Class of recommendation Level of evidence
  • Prior to the initiation of RT, a complete CIED evaluation should be performed and the treatment team should be informed of:
    1. Whether the device is a PM or ICD
    2. Whether the patient is pacing-dependent
    3. The minimum programmed pacing rate
    4. The maximum programmed tracking and sensor rates
  • Non-neutron-producing RT is preferred over neutron-producing treatment to minimize the risk of causing CIED malfunction, such as generator reset

  • Complete CIED evaluation should be performed weekly for patients undergoing neutron-producing RT

  • Complete CIED evaluation should be performed at the end of the course of RTs

I B
It is recommended that CIED interrogation and evaluation be performed at 1, 3, and 6 months after the end of RT due to the risk of latent damage I C
CIED relocation is recommended if the generator is situated in the path of the radiation beam IIa C
CIED relocation is not recommended for devices receiving a cumulative dose <5 Gy III B

CIED: cardiac implantable electronic device; ICD: implantable cardioverter-defibrillator; PM: pacemaker; RT: radiotherapy.

9. Conclusion

Much scientific evidence has emerged since the latest Brazilian guidelines for CIEDs were published by SOBRAC/SBC. Advances in technology and knowledge must be in line with clinical practice and public health care. In this respect, the present document highlights the evolution of the treatment of cardiac arrhythmias, but it does not shy away from highlighting the pressing need for the rational use of financial resources in favor of the greater good, that is, collective health.

Footnotes

Note: These guidelines are for information purposes and should not replace the clinical judgment of a physician, who must ultimately determine the appropriate treatment for each patient.


Articles from Arquivos Brasileiros de Cardiologia are provided here courtesy of Sociedade Brasileira de Cardiologia

RESOURCES