Skip to main content
. 2023 Aug 2;17(1):92. doi: 10.1186/s13065-023-00990-7

Table 1.

Chromatographic conditions and Parameters of system suitability of the proposed HPLC method for the determination of PAR and DAN

Chromatographic conditions Column Hypersil C18 column
Mobile phase 40:60 (v/v) of ethanol: water, 0.2% TEA at pH 4.5 ± 0.02 and 0.8 mL/min flowrate
Detection UV detection 290 nm
Parameters of system suitability
Parameter PAR DAN
No of theoretical plates, N 1936 4500
Selectivity factor, α 12.66
Resolution, Rs 8.75
Retention time (tR), min 2.2 5.7
Tailing factor (T), 5% of the peak height 0.97 1.14
Asymmetry factor, 10% of the peak height 1.11 1.21
Run time, min 6