Skip to main content
. 2023 Jun 26;24(8):e14077. doi: 10.1002/acm2.14077

TABLE 3.

The sensitivities and coefficient of determinations (R 2) at the given source‐detector distance (mean ± SD).

Source‐detector distance (cm) Sensitivity (cps/MBq) R 2
3 785.05 ± 39.97 1.00 ± 0.002
6 486.70 ± 30.84 1.00 ± 0.004
9 290.18 ± 21.69 0.99 ± 0.003
12 174.19 ± 24.50 0.99 ± 0.004
15 99.64 ± 10.83 0.99 ± 0.017
18 55.99 ± 5.13 0.98 ± 0.018