TABLE 3.

Summary of the calibration points, current, and time used for each electrometer.

Electrometer Range Calibration points
RT521R Input charge [nC] +200 +20 +5 −5 −20 −200
Current [nA] +20 +2 +0.2 −0.2 −2 −20
Time [s] 10 10 25 25 10 10
Model 35040 Input charge [nC] +200 +20 +5 −5 −20 −200
Current [nA] +20 +2 +0.2 −0.2 −2 −20
Time [s] 10 10 25 25 10 10
RAMTEC Duo L Input charge [nC] +10 +5 +1 −1 −5 −10
Current [nA] +0.2 +0.2 +0.2 −0.2 −0.2 −0.2
Time [s] 50 25 5 5 25 50
M Input charge [nC] +500 +250 +100 +50 +5 −5 −50 −100 −250 −500
Current [nA] +10 +10 +10 +10 +0.2 −0.2 −10 −10 −10 −10
Time [s] 50 25 10 5 25 25 5 10 25 50
UNIDOS‐E L Input charge [nC] +10 +5 +1 −1 −5 −10
Current [nA] +0.2 +0.2 +0.2 −0.2 −0.2 −0.2
Time [s] 50 25 5 5 25 50
M Input charge [nC] +500 +250 +100 +50 +5 −5 −50 −100 −250 −500
Current [nA] +10 +10 +10 +10 +0.2 −0.2 −10 −10 −10 −10
Time [s] 50 25 10 5 25 25 5 10 25 50