Skip to main content
. 2023 Jun 26;24(8):e14082. doi: 10.1002/acm2.14082

TABLE 5.

Error budgets for this study.

Budgets of uncertainty

Uncertainty

(%, k = 1)

Comments
Charge measurement of electrometers 0.08 Provided by the ANTM*‐accredited facility
Repeatability (measurements) 0.01 Obtained from measured data
Time linearity of output charge 0.01 Obtained from catalog
Uncertainty of zero drift 0.1 Obtained from catalog
Temperature coefficient of zero drift (/°C) 0.015 Obtained from catalog
Temperature coefficient of output current (/°C) 0.003 Obtained from catalog
Combined uncertainty (k = 1) 0.13 Root Mean Square
* Association for Nuclear Technology in Medicine