Skip to main content
. 2023 Apr 22;24(8):e14001. doi: 10.1002/acm2.14001

TABLE 2.

Error detection capabilities of various gamma criteria in PD and ArcCHECK measurements evaluated per plan

Portal dosimetry ArcCHECK
QA criterion γ2%/2 mm γ2%/1 mm γ1%/1 mm γ3%/3 mm γ3%/2 mm γ2%/2 mm
AUC 0.995 0.990 0.997 0.848 0.823 0.818
Std. uncertainty AUC 0.004 0.006 0.002 0.034 0.037 0.038
Optimum threshold [%] 99.0 92.0 86.0 99.3 98.4 95.3

Besides the AUC and the corresponding standard uncertainty each criterion's optimum decision threshold is stated.