. 2023 Apr 22;24(8):e14001. doi: 10.1002/acm2.14001

TABLE 3.

Error detection capabilities of various gamma criteria in PD measurements evaluated per treatment field

Portal dosimetry
QA criterion γ2%/2 mm γ2%/1 mm γ1%/1 mm
AUC 0.988 0.982 0.993
Std. uncertainty AUC 0.004 0.005 0.003
Optimum threshold [%] 99.6 94.5 87.5

Besides the AUC and the corresponding standard uncertainty each criterion's optimum decision threshold is stated.