Table 1.
Scanner | GE revolution CT | Canon ACQUILLION One | Siemens somatom edge | |||
---|---|---|---|---|---|---|
Scan mode | NCCT | ULDCT | NCCT | ULDCT | NCCT | ULDCT |
Tube voltage | 100 kV | 100 kV | 100 kV | 100 kV | 100 kVp | 100 kVp SN100* |
Image quality metric | Noise Index 22 | Noise Index 85 | N/A | SD 35 | Quality ref mAs 100 | Quality ref mAs 160 |
Scan time | 0.5 s | 0.5 s | 0.35 s | 0.35 s | 0.5 s | 0.5 s |
Pitch | 0.5 | 0.5 | 1.388 | 1.388 | 1.2 | 1.2 |
Collimation | 128 × 0.625 mm | 128 × 0.625 mm | 80 × 0.5 mm | 80 × 0.5 mm | 128 × 0.6 mm | 128 × 0.6 mm |
Kernel | Lung, mediastinum | Lung, mediastinum | Lung, mediastinum | Lung, mediastinum | BL57, BR38 | BL57, BR38 |
Image reconstruction mode** | ASIR-V 40% | ASIR-V 50% | AIDR-3D enhanced | AIDR-3D enhanced | ADMIRE 3 | ADMIRE 4 |
Slice thickness | 1.25/2.5 mm | 1.25/2.5 mm | 3 mm | 3 mm | 3 mm | 3 mm |
Recon increment | 0.5/2.0 mm | 0.5/2.0 mm | 2.4 mm | 2.4 mm | 3 mm | 3 mm |
Multiplanar reformat | Coronal/sagittal 2.5 mm |
Coronal/sagittal 2.5 mm |
Coronal/sagittal 3 mm |
Coronal/sagittal 3 mm |
Coronal/sagittal 3 mm |
Coronal/sagittal 3 mm |