Skip to main content
. 2023 Jul 6;31(9):101697. doi: 10.1016/j.jsps.2023.101697

Table 7.

Validation Parameters of the UV–Visible Spectrophotometric Method of Paclitaxel Analysis.

Parameters Results
Linearity
Calibration Range 1 – 30 µg/ml
Wavelength (λ max) 237 nm
Regression Equation Y = 0.0288x – 0.1511
Correlation Co Efficient (R2) 0.993
Accuracy (% Recovery) Mean ± S.D; % RSD
Sample Without Excipients 98.46 ± 0.13; 0.13
Sample With Excipients 99.76 ± 0.27; 0.27
Stability (Amount Recovered) Mean ± S.D; % RSD
Day 01 (n = 3) 9.94 ± 0.04. 0.40
Day 02 (n = 3) 9.88 ± 0.03. 0.30
Day 03 (n = 3) 9.94 ± 0.02. 0.20