Design, characterization,
and performance assessment of the on-chip
integrated GFET device. (A) Schematics of the on-chip integrated GFET
biosensor platform with the antibody functionalized channel as the
sensing channel and a channel without antibody functionalization as
the control channel. (B) Illustration of the layout of on-chip integrated
GFET sensor, designed with 6 devices each on the left and right sides.
These devices are separated by the central common-on-chip integrated
Au electrode. (C) GFET device shift of the control and sensing channels
on three GFET chips tested for the same concentration of exosomes
in buffer, illustrating the importance of control channel. The GFET
device detection signal is calculated based on the difference between
the GFET response from control and sensing channels. Error bars are
determined by the standard deviation of multiple device measurements,
5 ≤ n ≤ 6 for each reading. (D) GFET
device detection signal comparison between the five chips tested for
the same concentration of exosomes in buffer, illustrating low device-to-device
variation and high chip-to-chip reproducibility of the GFET biosensor.
Error bars are determined by the standard deviation of multiple device
measurements, 4 ≤ n ≤ 6 for each reading.
(E) Device exclusion criteria for GFET reproducibility and validity.
Histogram of Dirac voltage for lab-tested GFET devices with Gaussian
distribution curve. Device exclusion criteria were set based on 444
lab-tested GFET devices, where the Dirac voltage for lab-tested devices
was VD = 0.71 ± 0.24 V. (F) Calibration
curve of the GFET biosensor for detection of various concentrations
of model cancerous exosomes in buffer. Error bars are determined by
the standard deviation of multiple device measurements, 4 ≤ n ≤ 24 for each reading. (G) Measurement results
for the GFET biosensor with healthy plasma and model cancer plasma,
illustrating excellent selectivity of the GFET biosensor. (****P < 0.0001). Data are mean ± standard deviation
(s.d.).