Authors Method Target Parameters Cutoff Values (HU) SE (%) SP (%) AC (%)
Ohno Y, et al. [5] 320-detector row CT Pulmonary nodules Perfusion (mL/100 mL/min) calculated with single-input maximum slope method 40.0 98 (42/43) 79 (26/33) 90 (68/76)
Extraction fraction (mL/100 mL/min) 2.0 88 (38/43) 82 (27/33) 86 (65/76)
Blood volume (mL/100 mL) 2.0 86 (37/43) 54 (18/33) 72 (55/76)
FDG-PET/CT SUVmax 2.0 91 (39/43) 52 (17/33) 74 (56/76)
Ohno Y, et al. [12] 320-detector row CT Pulmonary nodules Total perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 40 86.0 (49/57) 79.5 (31/39) 83.3 (80/96)
Perfusion (mL/100 mL/min) calculated with single-input maximum slope method 20 64.9 (37/57) 69.2 (27/39) 66.7 (64/96)
FDG-PET/CT SUVmax 2.5 63.2 (36/57) 56.4 (22/39) 60.4 (58/96)
Ohno Y, et al. [15] 320-detector row CT Pulmonary nodules Total perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 29 92 (123/133) 71 (60/85) 84 (183/218)
Nodule perfusion (mL/100 mL/min) calculated with single-input maximum slope method 10 91 (121/133) 28 (24/85) 67 (145/218)
Dynamic first-pass CE-perfusion MRI for 1.5T system Maximum relative enhancement 0.13 92 (123/133) 49 (42/85) 76 (165/218)
Slope of enhancement 0.016 93 (124/133) 49 (42/85) 76 (166/218)
FDG-PET/CT SUVmax 2 89 (119/133) 31 (26/85) 67 (145/218)
Ohno Y, et al. [20] 320-detector row CT Lymph node metastasis in NSCLC analyzed per node Total perfusion (mL/100 mL/min) calculated with slope of enhancement dual-input maximum slope method 58 54.2 (32/59) 89.8 (53/59) 72.0 (85/118)
Systemic arterial perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 4.1 98.3 (58/59) 56.4 (51/59) 92.4 (109/118)
Permeability surface (mL/100 mL/min) assessed with Patlak plot method 8.7 50.8 (30/59) 94.9 (56/59) 72.9 (86/118)
Distribution volume (mL/100 mL) assessed with Patlak plot method 0.37 84.7 (50/59) 44.1 (26/59) 64.34 (76/118)
FDG-PET/CT SUVmax 2.9 74.6 (44/59) 91.5 (54/559) 83.1 (98/118)
Seki S, et al. [22] 320-detector row CT Therapeutic outcome prediction for NSCLC Total perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 29.2 78.3 (18/23) 85 (17/20) 81.4 (35/43)
Pulmonary arterial perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 15.5 65.2 (15/23) 80 (16/20) 72.1 (31/43)
Systemic arterial perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 11 82.6 (19/23) 80 (16/20) 81.4 (35/43)
Dynamic first-pass CE-perfusion MRI at 3T system Total perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 37.5 69.6 (16/23) 95 (19/20) 81.4 (35/43)
Pulmonary arterial perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 16.3 65.2 (15/23) 80 (16/20) 72.1 (35/43)
Systemic arterial perfusion (mL/100 mL/min) calculated with dual-input maximum slope method 16.5 82.6 (19/23) 80 (16/20) 81.4 (35/43)
FDG-PET/CT SUVmax 5.7 87.0 (20/23) 76.9 (14/20) 79.1(34/43)