Skip to main content
. 2023 Aug 4;2023:3116223. doi: 10.1155/2023/3116223

Table 2.

Summary of the validation report of the proposed methods.

Parameters I II III IV V
Linear equation y = 1.7418x − 0.0134 y = 2.6599x − 0.0328 y = 2.5719x − 0.0266 y = 1.7827x − 0.0168 y = 165113x − 26088
R 0.9999 0.9999 0.9998 0.9999 0.9999
Linearity range (μg/mL) 0.10–2.03 0.10–2.03 0.10–2.03 0.10–2.03 0.34–27.31 ng/mL
LOD (μg/mL) 0.005 0.005 0.005 0.005 0.05 ng/mL
LOQ (μg/mL) 0.01 0.01 0.01 0.01 0.17 ng/mL
Precision (%) (n = 6) 0.75 1.08 0.42 1.05 0.36
Repeatability (%) (n = 6) 3.65 3.36 2.25 5.22 8.72
Stability (%) 2.10 4.04 1.53 2.55 0.57
Accuracy at low level (n = 3)
 Recovery (%) 102.99 101.96 104.80 100.98 91.18
 RSD (%) 3.09 3.09 3.64 2.30 1.89
Accuracy at medium level (n = 3)
 Recovery (%) 102.68 98.35 104.20 101.69 111.27
 RSD (%) 3.69 0.70 1.83 1.78 0.58
Accuracy at high level (n = 3)
 Recovery (%) 102.53 100.84 105.78 103.28 102.96
 RSD (%) 2.28 1.89 1.93 1.83 0.73