Abstract
Implanted neural stimulation and recording devices hold vast potential to treat a variety of neurological conditions, but the invasiveness, complexity, and cost of the implantation procedure greatly reduce access to an otherwise promising therapeutic approach. To address this need, a novel electrode that begins as an uncured, flowable prepolymer that can be injected around a neuroanatomical target to minimize surgical manipulation is developed. Referred to as the Injectrode, the electrode conforms to target structures forming an electrically conductive interface which is orders of magnitude less stiff than conventional neuromodulation electrodes. To validate the Injectrode, detailed electrochemical and microscopy characterization of its material properties is performed and the feasibility of using it to stimulate the nervous system electrically in rats and swine is validated. The silicone-metal-particle composite performs very similarly to pure wire of the same metal (silver) in all measures, including exhibiting a favorable cathodic charge storage capacity (CSCC) and charge injection limits compared to the clinical LivaNova stimulation electrode and silver wire electrodes. By virtue of its simplicity, the Injectrode has the potential to be less invasive, more robust, and more cost-effective than traditional electrode designs, which could increase the adoption of neuromodulation therapies for existing and new indications.
Keywords: biomaterials, injectrodes, neural interfaces, neuromodulation, peripheral nerves
1. Introduction
Electrical stimulation of the peripheral nervous system, often known as neuromodulation, bioelectronic medicines,[1] or electroceuticals,[2,3] is an increasingly prevalent clinical therapy. In recent years, the Federal Food and Drug Administration (FDA) has approved market applications for a myriad of new indications for bioelectronic medicines including diverse conditions such as obesity, sleep apnea, migraine, opioid withdrawal symptoms, dry eye, essential tremor, and hypertension. These supplement traditional market indications for bioelectronic medicines such as pain and overactive bladder.[4,5] Additional areas of research currently beginning early stage clinical testing include the use of therapeutic electrical stimulation to treat inflammation, metabolism, and endocrine disorders.[6–9] The increasing prevalence of bioelectronic medicines has led to a renewed interest in both developing novel devices as well as improving the mechanistic understanding of how neural interface technologies interact with the nervous system. This interest has resulted in numerous large-scale government funding programs such as the White House BRAIN Initiative, the NIH SPARC Program, and the DARPA HAPTIX and ElectRx Programs.[10]
Existing electrical stimulation devices designed for integrating with the human nervous system can be separated into noninvasive devices, minimally invasive hybrid strategies, and invasive devices. Noninvasive devices apply an electrical stimulation waveform through electrodes placed on the surface of the skin, which are typically intended to interface electrically with deeper neural structures (Figure 1B). Although some impressive therapeutic responses have been demonstrated using noninvasive stimulation paradigms targeted to specific nerves, the unintended stimulation of nerve and muscle fibers superficial to the deep target often leads to therapy limiting side effects.[11–13] Additionally, the electric field falls off rapidly with distance from the stimulating electrode, which limits the depth at which structures can be reliably stimulated, at least without intolerable activation of cutaneous pain fibers.[14]
Figure 1.

Peripheral nerve stimulation electrode types from least to most invasive, left-to-right. A) Simplified tissue schematic showing distinct layers of the skin muscle and nerve. B) Transcutaneous electric nerve stimulation (TENS), C) Percutaneous electric nerve stimulation (PENS), D) injected wireless stimulator (e.g., BION device), E) implanted nerve stimulator (e.g., spinal cord paddle electrode), and F) implanted cuff electrodes (e.g., vagus nerve stimulator).
Given the limitations of noninvasive stimulation strategies, implantable neural stimulation electrodes are commonly used to provide more precise activation of a nerve trunk, ganglia, nuclei, or other targets of interest. One of the most common implantable interfaces to stimulate the peripheral nervous system is the nerve cuff, which consists of one or multiple electrodes positioned on an insulating backer that is placed surgically with the electrode contact(s) surrounding the epineural surface of a nerve trunk (Figure 1F).[15] Although this approach can provide very specific target engagement, the placement of nerve cuffs requires surgical dissection down to the nerve trunk of interest through skin, muscle, smaller nerves, and microvasculature. Moreover, cuff implants typically require 360° dissection around the nerve trunk of interest to facilitate placement around the nerve. Additionally, stimulation leads must be tunneled to an implantable pulse generator (IPG) that has limited battery life. Multiple points of failure and the complexity of the implantation procedure increases both surgical risk and costs and may contribute to a large amount of variability in the therapeutic effectiveness of peripheral neuromodulation devices. For example, a 152-patient clinical study recently compared invasive dorsal root ganglion (DRG) stimulation to traditional epidural spinal cord stimulation (SCS) for intractable pain. 81.2% of the DRG patients received a greater than 50% decrease in back pain versus 55.7% with SCS, a significant difference. However, the adverse event rate related to the neural stimulator/device or the implant procedure was significantly higher in the DRG arm (36.8 to 26.4% and 46.1 to 26.3%, respectively). As both DRG and SCS require a complex electrode/lead/IPG system, implant costs ranging from $32k to $58k and annual maintenance costs ranging from $5k to $20k also pose significant barriers to treatment.[16]
Practically speaking, stimulation device selection involves carefully balancing trade-offs between efficacy and safety, including implantation invasiveness, efficacy, robustness, and the cost of the procedure. Although stimulation and implantation strategies spanning the spectrum from noninvasive to fully implanted (Figure 1) are available, there remains an unmet need for neural interface technologies that can be implanted through a minimally invasive procedure but can maintain a robust connection with complex and difficult to surgically dissect peripheral neuroanatomy.
To meet this need, we have developed a novel electrode, which we call the “Injectrode” (Figure 2). The Injectrode is flowable as a prepolymer and is injected via a syringe where it cures to form a highly conforming, compliant neural electrode in vivo. By flowing around the target neuroanatomy, the Injectrode can conform to a variety of targets to form different neural interfaces. For instance, the neural engagement of an invasive cuff electrode could be mimicked by injection of the electrode into the sheath around a nerve trunk. Alternatively, the Injectrode can be used to stimulate complex neural structures such as nerve plexus or those inside a foramen that may be very difficult to target with traditional cuff electrodes. Through the injection of this electrode, neural interfaces can be created using a minimally invasive surgical approach that avoids the risks of surgical complications associated with open cut downs and extensive dissection. Although, the Injectrode is eventually intended to be delivered in a minimally invasive fashion in a clinical population, the goal of this paper was to characterize the electrochemical properties of the composite material to demonstrate potential as a neural stimulation electrode as well as to show proof-of-concept stimulation of complex neuroanatomy in small and large animal models. Development of methods for precise and reproducible administration of the Injectrode, including pairing the Injectrode with a wireless stimulator, to form functional multicontact neural interfaces in a clinical setting is reserved for future work.
Figure 2.

The Injectrode Concept. A) A composite material consisting of (1) a prepolymer carrier and (2) conductive filler elements injected into the body near a target nerve. The mixture cures in place to form an electrically conductive interface. B) Wireless configuration (e.g., if connected with a RF antenna circuit), C) wired configuration (e.g., if directly connected to an IPG).
Herein, we present detailed electrochemical and microscopy characterization of the material properties of the Injectrode. Additionally, we validate the feasibility of electrically stimulating peripheral nerves with an in vivo curing composite material in preclinical animal models. Visual assessments were made of the surface morphology of electrodes via light microscopy and scanning electron microscopy. Cyclic voltammetry (CV), voltage excursions, and electrochemical impedance spectroscopy (EIS) were used to directly compare the electrochemical properties of the Injectrode to those of pure silver wire, as well as clinical LivaNova electrodes. Finally, we also present proof-of-concept experiments in small and large animal models to demonstrate the feasibility of this material as an in vivo neural interface.
2. Results and Discussion
Initial benchtop impedance measurements were undertaken to ascertain the percolation threshold of the material mixtures, which was found to occur at ≈65% w/v mix with respect to silver content (Figure S1, Supporting Information). Based on these results, an 80% w/v with respect to silver content was selected to provide a wide margin of error and used for all of the electrochemical and in vivo testing presented in this paper.
2.1. Electrode Imaging
Light microscopy and electron microscopy were performed in order to assess the surface properties of the material. For both types of imaging, a thin slice of material was cut from the surface of an Injectrode working electrode sample that had not undergone any electrochemical testing or stimulation. Imaging revealed that the silver particles were well distributed throughout the silicone matrix (Figure 3). Electron microscopy showed surface porosity (Figure 3C,D) as well as a complex arrangement of silver particles within in the silicone matrix.
Figure 3.

Light microscopy and scanning electron microscopy (SEM) of the surface of an Injectrode. A,B) Light microscopy of the Injectrode surface showing distribution of silver particles throughout the sample. C,D) SEM at comparable magnification showing sample topography and revealing porosity of the polymer matrix.
The porosity of the silicone matrix may act to increase electrolyte permeability, which could enable electrode/electrolyte interactions with particles deeper within the silicone matrix. This in turn would increase the fractal dimension of the material and thus the effective electrochemical surface area. For these reasons, the porosity of the material may be important to the functional capabilities of the Injectrode as a stimulation electrode particularly with respect to increasing the effective surface area available for charge injection. Thus, modifying the porosity of the cured Injectrode is one factor that could potentially be controlled in order to improving the stimulation characteristics of the electrode.
2.2. Electrochemical Characterization: Cyclic Voltammetry
Cyclic voltammetry is a classical electrochemical method for the characterization of stimulation electrodes. It provides information regarding the electrode interface under electrical load, the electrochemical conversion of species within the solution, and on transient changes due to oxidation or other processes at the electrode surface.[17]
For assessment of the LivaNova electrode, performing slow scan (50 mV s−1) cyclic voltammetry sweeps in phosphate buffered saline (PBS) over a voltage range known as the water window; the window between the voltages at which evolution of hydrogen and oxygen occur is common practice.[17,18] Commonly used values for assessment of Pt and Pt/Ir electrodes include −0.62 to 0.9 V versus Ag/AgCl[18] and −0.6 to 0.8 V versus Ag/AgCl.[17] These values are a function of the electrode material and the solution within which testing is conducted. For assessment in the context of new materials for neural stimulation, understanding the limits of the water window in physiologically relevant solution is important. Therefore, we performed a cyclic voltammetry sweep for silver wire and the Injectrode samples with an expanded range from −1.9 to 3 V versus Ag/AgCl. It is important to note that, silver particles were used as the conductive filler for the Injectrode due to silver’s high conductivity and low cost for iterating through different Injectrode fabrication and testing methods to establish this early proof of concept. Despite these benefits, silver is not an acceptable choice for the creation of a chronic neural interface as silver and silver chloride have well-understood toxic effects.[19] Although testing performed with silver particle based electrodes was sufficient for validating the Injectrode concept, future development will focus on the use of stainless steel, gold, or platinum filler particles that have been well validated in chronic neural interfaces.[18,20] We compared the Injectrode samples to a silver wire electrode with the same geometric surface area to assess equivalency of electrochemical behavior over the expanded voltammogram but did not conduct a detailed study of the electrochemical behavior over this range.
As has been described elsewhere, in PBS, pure silver wire exhibited a large peak near 0 V that has been attributed to the formation of silver chloride (AgCl).[21] The broadness of the silver chloride peak and the observation of multiple humps as well as differences in the peak locations between the normal and extended window CV sweeps may indicate the formation of other silver oxides or silver phosphates, particularly during the extended sweep.[21–23] Additionally, the reversal of the silver chloride reaction was observed on the anodic side of the CV sweep (see Figure 4A). Putative hydrogen evolution was observed near the anodic limit of the voltammogram (at ≈−1.7 V) and visually confirmed through the formation of microbubbles at the surface of the exposed silver wire electrodes and Injectrode samples. Interestingly, no clear oxygen evolution was observed even with the upper limit of the voltammogram sweep extended to 3.0 V. This phenomenon was not explored further. One of the primary purposes of the electrochemical characterization experiments was to show that the Injectrode behaves similarly to the base metal, in this case silver. This was clearly true via visual inspection, although the Injectrode exhibited slightly decreased peak heights compared to a silver wire electrode with the same geometric surface area, peak shapes, and locations for the Injectrode and base metal were similar in shape and location. This was the case for both the truncated sweep within the water window for platinum (Figure 4B) and the expanded sweeps performed on the silver wire electrode and Injectrode (Figure 4A).
Figure 4.

Cyclic voltammetry comparison. A) Mean cyclic voltammograms of five samples each of Injectrode and silver wire electrodes with matched geometric surface area. Scans were performed at 50 mV s−1 and over a range from −1.9 to 3.0 V. B) Mean cyclic voltammograms of five samples each of Injectrode and silver wire electrodes as well as one contact of a LivaNova PerenniaFLEX electrode. Voltammograms were scanned from −0.62 to 0.9 V at a rate of 50 mV s−1. Inset shows an expanded view of the voltammogram for the LivaNova electrode. C) Cathodic charge storage capacity (CSCC) calculated from the −0.62 to 0.9 V voltammograms for each set of electrodes. Injectrode and silver wire electrodes exhibit similar CSCC. D) The 1st and 15th voltammogram from one representative Injectrode sample showing spiking behavior in early voltammetry sweeps that goes away with the continued application of cyclic voltammograms. Insets show a zoomed view of this phenomenon. Shaded regions in panels (A) and (B) represent ±1 standard deviation. Error bars in panel (C) also represent standard deviation.
Despite the expanded water window of Ag, an −0.62 to 0.9 V versus Ag/AgCl voltammogram normally used for assessment of Pt/Ir was used instead of the expanded sweep for comparing the Injectrode voltammogram and charge storage capacity to that of platinum stimulation electrodes to provide a more fair comparison. The Injectrode and silver wire electrode both exhibited much larger cathodic charge storage capacity (CSCC) compared to a LivaNova electrode. This is largely due to increased charge injection through faradaic mechanisms. However, in light of the decrease in exposed silver available for faradaic reactions when looking at a 2D cross section of the Injectrode, as can be seen in the scanning electron microscopy (SEM) imaging shown in Figure 3C,D, the similarity between the CSCC of the silver wire and Injectrode samples was unexpected. This suggests that the Injectrode material has greater available 3D surface area due to the previously discussed porosity and permeability of the matrix to the electrolyte to achieve greater CSCC. This effect is similar to the increases in CSCC that have been observed with electrode coatings such as sputtering iridium oxide films (SIROF) and activated iridium oxide films AIROF on platinum.[17,24–27]
An interesting feature within the CV response that was inconsistent between silver wire electrodes and the Injectrode samples was observed. In extreme voltages of both the anodic and cathodic sweeps of the voltammogram, sharp peaks were initially observed during early CV sweeps (Figure 4D). These peaks are shown in detail in the lower inset of Figure 4D and are inconsistent with traditional faradaic reaction mechanisms that lead to “humps” in the CV, such as those described for silver/silver chloride. Additionally, these spikes appeared at inconsistent voltages on subsequent CVs, and as additional CVs were run on the Injectrode samples the frequency of these spikes decreased. An example of this is shown in Figure 4D, where extensive spikes are shown on the first complete CV scan but are not observed on the 15th scan. These observations suggest the spikes are not caused by reactions occurring at specific voltages but instead may indicate changes in the exposed surface of the Injectrode or the presence of contaminants from the manufacturing process. We speculate that these spikes may be indicative of dissolution of base silver particles less well-bound within the silicone matrix, or of the exposure of new metal particles available for faradaic reactions through the opening of new pores in the polymer matrix. However, as the overall area under the CV tended to increase over initial sweeps and then settle into a consistent value, this would suggest the loss of particles or changes in exposed surface area would not negatively impact electrical characteristics of the Injectrode, from a neuromodulation perspective. Loss of metal particles from the surface of the Injectrode could be minimized, at least to some extent, through future refinement of the particle sizes, polymer matrix, mixing procedure, or other factors. However, it is worth noting that large chunks of Pt/Ir have been reported in post mortem human tissue from chronically stimulated cochlear implants and that tissue damage is driven by the formation of specific Pt salts during stimulation that lead to local toxicity,[28] as opposed to larger chunks of unreacted Pt or Pt/Ir flaking off.[29–33] Whether loss of a small portion of conductive particles during the application of stimulation is an issue for strategies that leverage conductive particles to dope traditionally insulative matrices, such as the Injectrode, should be explored in future studies.
2.3. Electrochemical Characterization: Voltage Excursions
Although slow cyclic voltammetry provides unique information about electrochemical reactions that occur at the electrode/electrolyte interface, neuromodulation therapies typically employ charge balanced square wave current controlled pulses ranging from 30 to 1000 μs in duration. For that reason, voltage transients (VT) (or voltage excursions) are traditionally used to measure the polarization on the electrode itself during the application of cathodic leading, constant current, charged balanced biphasic pulses at a pulse width within normal clinical neuromodulation ranges.[34] For the LivaNova electrode, the current applied is increased until the cathodic polarization on the electrode itself approaches the cathodic limit for the evolution of hydrogen on a Pt/Ir electrode, and this current is adjusted for the geometric area of the electrode to determine the safe charge-injection limit. It is worth noting that for materials that differ from Pt or Pt/Ir the use of voltage excursions to determine unsafe levels of polarization needs to be reevaluated on an electrode material by electrode material basis. This is of particularly importance for materials through which there may be reversible safe electrochemical reactions induced at one level of polarization, and another unsafe electrochemical reaction occurring at a higher level of polarization.
Given the extended CVs performed to determine the water window for silver indicated the anodic and cathodic limits before the evolution of oxygen or hydrogen were outside the ranges of traditional Pt or Pt/Ir electrodes, and the use of silver particles within the composite material of the Injectrode is known to be an issue for direct translation, we decided to perform a conservative comparison of acceptable charge injection limits. This was done by intentionally limiting the acceptable polarization on the silver wire electrodes or Injectrode to the established cathodic limit for Pt or Pt/Ir electrodes. By limiting the polarization of the electrode to the safe limits for Pt or Pt/Ir before hydrogen evolution, we did not find the maximum charge injection limits of the Injectrode sample; however, this approach is sufficient to show the Injectrode concept can achieve neuromodulation relevant charge injection limits and is not inherently limited by the properties of the composite material. Given that the Injectrode silicone matrix is permeable to water (or PBS, or electrolyte solution in vivo) over time, the increased access to interior fractal dimensions during slow scan voltammetry may lead to a large CSCC calculation that is not relevant for standard neuromodulation pulses. As such, voltage transient measurements provide an additional method to assess the electrochemical behavior during more stereotypical short neuromodulation pulses. For this reason, it will be important to determine safe charge injection limits for a more clinically accepted conductor/carrier composition in time.
As shown in Figure 5A, the voltage transient on the LivaNova electrode first consists of a very sharp negative voltage deflection at the initiation of the cathodic current pulse, resulting from the voltage generated from application of the current to the resistive load from the ionic solution and internal properties of the Injectrode. There is then a slow, consistent change in voltage during the remainder of the cathodic current pulse, generated by the polarization of primarily capacitive Pt/Ir electrode (Figure 5D, green line). Charge density limits before reaching −0.6 V of electrode polarization were between 50 and 150 μC cm−2, consistent with the known charge density limits for Pt electrodes (Figure 5C). In comparison to the LivaNova electrode, there is almost no polarization of the silver wire electrodes or the Injectrodes during the application of a current pulse adjusted to match the same applied current density (see Figure 5A,B,D). Overall, Injectrodes and silver wire electrodes behaved similarly during the applications of voltage transients, which is another indicator that the Injectrode will behave similarly to metal electrodes made from the same material as the conductive filler in the electrode. Small differences, between electrodes, in the initial voltage caused by the resistance of the solution are expected, as the two electrode types have slightly different geometric surface areas and geometries. The uncompensated resistance is driven primarily by the ionic resistance at the cross-sectional area of the electrode/electrolyte interface, since ionic current flow further from the electrode is not limited to the cross-sectional area of the electrode/electrolyte interface and can take advantage of the much greater volume for ionic current flow in the bulk solution. The internal resistance of the Injectrode, compared to the silver wire electrode and LivaNova electrode may also contribute to these differences. Overall, the Injectrode and Silver Wire electrode show similar behavior during voltage transients, which is promising for the use of the Injectrode for neuromodulation. This suggests that future Injectrode compositions using conductors more appropriate for chronic stimulation, such as platinum, will exhibit charge injection limits that are similar or superior to existing neuromodulation electrodes.
Figure 5.

Voltage transients. Voltage transients were performed in phosphate buffered saline using a 500 μs pulse width and allowing at least 1000 pulses for the response to stabilize. A) Voltage transients performed at 500 through 3000 μA on one Injectrode sample that had previously undergone 20 cyclic voltammogram sweeps from −0.62 to 0.9 V. B) The sample measurements shown in panel (A) but performed with a silver wire working electrode with geometric surface area matched to the Injectrode sample. C) Voltage transients performed with a LivaNova PerenniaFLEX stimulation electrode. D) Voltage transients at 3000 μA from each electrode sample. To optimize visual comparisons of the progression of the voltage transients with respect to stimulation amplitude, the baseline for all voltage transients are centered on the vertical axis; however 0 V with respect to Ag/AgCl is indicated on each vertical axis for assessing offsets.
2.4. Electrochemical Characterization: Impedance Spectroscopy
In addition to providing a relative measure of impedance at different frequencies for different electrode materials, EIS provides additional information about the balance of resistive versus capacitive effects dominating current flow. Figure 6A is an EIS sweep between 0.1 Hz and 100 kHz for the LivaNova electrode. The LivaNova electrode EIS results can be explained using a simplified model assuming the electrode/electrolyte interfaces act as resistor and capacitor in parallel. This interface in turn is in series with the ionic resistance of the solution, and the impedance of the large Pt foil counter electrode is neglected due to the use of a three-electrode setup. At very high frequencies in the 10–50 kHz range, the capacitive element of the electrode/electrolyte interface is very low impedance with respect to both the resistive element of the electrode/electrolyte interface and the ionic resistance of the PBS solution. Consequently, the ionic resistance of the solution dominates the overall measured impedance as it is in series with the electrode, and the phase angle measured is 0° indicating current flow across a resistor. As the applied frequency decreases until another transition point 1–10 Hz, the impedance of the capacitive component of the electrode/electrolyte interface increases, and the phase angle increases toward purely capacitive current flow (90°). As the applied frequency goes below 1 Hz the resistance component of the electrode/electrolyte interface becomes dominant with respect to both the impedance of the parallel capacitive component and the solution resistance, and the overall measured impedance begins to dramatically increase with the phase angle trending back toward purely resistive values (0°).
Figure 6.

Electrochemical impedance spectroscopy comparison. Electrochemical impedance spectroscopy of five samples each of Injectrode and silver wire electrodes as well as one contacts of a LivaNova PerenniaFLEX electrode. A) Mean impedance of each set of electrodes as a function of frequency. B) Mean phase of each set of electrodes as a function of frequency. Impedance and phase of Injectrode samples and silver wire show similar behavior across frequencies. Shaded regions represent ±1 standard deviation.
For a silver wire electrode, the overall impedance at low frequencies is less than the LivaNova electrode, and the phase angle stays much closer to 0° above 100 Hz indicating current flow through a resistor (Figure 6B). This result is not surprising, as the silver wire has a much more facile mechanism for faradaic (resistive) charge transfer through the formation of AgCl than non-faradaic charge transfer (capacitive). The EIS curves for the Injectrode are very similar to those of a pure Ag electrode. The overall Injectrode impedance is slightly higher than the pure Ag wire at high frequencies, and the impedance becomes almost identical to Ag wire at low frequencies. This may suggest that more of the fractal dimensions of the silver particles are accessible to ions in the electrolyte solution through the silicone matrix at low frequencies. This low-frequency behavior is a known phenomenon for conductive polymer coating strategies intended to increase the effective fractal dimensions of a stimulating electrode, decreasing pore resistance by providing more time for the ions in solution to permeate the inner recesses of the conductive polymer matrix during each sinusoidal pulse.[24–26] Additionally, differences in fractal dimensions between Injectrode samples may have led to different amounts of charge transfer via capacitive and faradaic mechanisms between samples, thereby causing the increased variability in the low-frequency phase that was observed in comparison to the silver wire electrodes (Figure 6B). However, given the complex nature of the Injectrode/electrolyte interface caution should be observed in overinterpreting these results.
2.5. Mechanical Testing
The Injectrode was capable of undergoing large reversible deformations (Figure 7). Of four samples tested, one initial sample was tested to only 140% strain, the other three were tested to 200%, of which only one failed, at 167% strain. In comparison to the relevant neuroanatomy, the ulnar nerve is thought to undergo one of the largest deformations of nerves in the body and is only estimated to stretch by ≈29% strain under normal elbow flexion.[35] The Young’s modulus of the Injectrode was estimated from the slope of the elastic region in the stress–strain curves to be 72.1 ± 10.9 kPa. This is orders of magnitude less stiff than other materials currently utilized in neural interfaces, for example, steel (≈200 GPa), gold (74 GPa), polyimide (2.5 GPa), and PTFE (400 MPa).[36] The Young’s modulus of the Injectrode remains slightly higher than that of nerve tissue. Estimates of Young’s modulus of peripheral nerves vary from (20 kPa to 1 MPa), depending on the varied methods of measurement.[37]
Figure 7.

Tensile mechanical testing. A) Dog bone shaped samples of Injectrode material were prepared in acrylic molds (blue: Injectrode base material, gray: conductive silver Injectrode), modeled after ASTM D412-16, standardized testing of vulcanized rubber and other elastomeric materials. B) The elastic region of an Injectrode under tensile testing shown by a representative stress–strain curve; error bars show standard deviation (n = 4 samples). C) Photographs of Injectrode sample during testing, showing that the material is capable of elastically stretching to more than double its original length before plastic deformation occurred.
2.6. Surgical Delivery Proof of Concept
For proof of concept and visualization purposes, we demonstrated the ability to place Injectrodes with an open-cut down approach in a swine cadaver. The materials were easily extruded from 18 gauge needles affixed to syringes and formed cohesive and flexible structures around the nerves (Figure 8A). Importantly, the materials were capable of flowing around complex branched structures and conformed to the anatomy. The materials could also be injected into enclosed areas such as the tissue “sheath” overlying the neurovascular bundle (Figure 8B). While these procedures were performed with an open incision, the modality could be adapted to a minimally invasive approach mimicking similar endoscopic procedures performed today.
Figure 8.

Delivery to multiple complex nerve anatomies in swine. Injectrodes were used to stimulate nerves in a rat (N = 5) and a swine (N = 2) model following extensive cadaver testing to optimize delivery and easy handling under surgical-OR conditions. A) Delivered by syringe, nerve cuff diameters of 1 to 3 mm were achieved around peripheral nerves in a swine cadaver. In addition to encasing a single nerve with an injectable cuff, several nerves running in parallel—either in a bundle or at a neural plexus—were encapsulated and stimulated with a wire that was embedded within the Injectrode cuff. B) Furthermore, nerve bundles were encased by injecting the Injectrode into a nerve sheath commonly surrounding the nerve(s) and then gently manipulating (massaging) the nerve sheath, thereby distributing the Injectrode inside the sheath that functioned as a mold. Injectrodes were placed around nerves on the inside and on the outside of the nerve sheath in swine model.
2.7. Electrical Stimulation of Compound Motor Nerves (Brachial Plexus)
Injectrode cuffs were placed 360° around the brachial plexus in a rat model, just distal to the location where the nerve visibly branched in to the median, ulnar, and radial segments. A silver wire was embedded into the Injectrode cuff prior to curing and used to connect to the signal generator. A nerve recruitment curve was collected and supramaximal recruitment was determined to be beyond 2 mA of current amplitude. To ensure a stable system during tetanic stimulation experiments, at least twice the supramaximal current amplitude was used, in this case 5 mA pulses of 200 μs pulse width. Complete tetanic muscle contractions resulted from 30 Hz stimulation using a charge balanced biphasic waveform (200 μs pulse width, 5 mA supramaximal amplitude) as shown in Figure 9. The simulation effect was quantified by direct measurement of joint angles in captured video (Figure 9A).
Figure 9.

Acute tetanic stimulation rat brachial plexus. Injectrodes were placed in a bipolar arrangement on the left brachial plexus of a rat. A) Based on visual assessment, stimulation parameters were established to produce tetanic contraction of the forearm (30 Hz, 200 μs, 5 mA). B) A period of stimulation was applied to produce a contraction. C) Contraction was later quantified by measuring wrist joint angles via post hoc by video analysis.
This brief experiment demonstrated that the Injectrode material can form an electrically conductive interface with the nerve and produce activation of the fibers that can lead to a tetanic contraction of muscles. In addition, we demonstrated that the material could form an interface with the complex branched nerve structures of the brachial plexus.
2.8. Electrical Stimulation of Swine Vagus Nerve
In order to directly compare stimulation thresholds with as closely controlled electrode geometries as possible, we formed the Injectrode materials into bipolar cuffs of roughly the same size and electrode spacing as the clinically available LivaNova cuff electrode (Figure 10A). The goal of the experiment was to test the capability of the Injectrode to activate neural tissue in a relevant large animal model and not to verify a specific surgical approach. For this reason, the Injectrode contacts were preformed with the correct geometric dimensions on a stainless steel mesh while curing ex vivo around a connecting wire immediately prior to direct placement on the swine cervical vagus. Care was taken to ensure the stainless steel mesh was completely encased by Injectrode material; the backing was further encapsulated with polydimethylsiloxane (PDMS), which is the green material shown in Figure 10A. With this setup, we performed a current controlled dose–titration protocol, where the biphasic waveform was held constant, 200 μs pulse width at 30 Hz stimulation, and current was ramped up between 0 and 5 mA. The physiological parameter we measured was heart rate (BPM), per previously published protocols.[38] For a given amplitude, the Injectrode induced a proportionally lower change in heart rate compared to the LivaNova electrode (Figure 10B). The approximate voltage was estimated based on recorded cuff impedance values prior to stimulation. Interestingly, though not surprisingly, when corrected for voltage titration, the two electrodes performed nearly identically, as shown in Figure 10.
Figure 10.

Demonstration of acute swine vagal nerve stimulation. To evaluate effective stimulation parameters of the Injectrode compared to the commercial LivaNova electrode to induce heart rate reduction, Injectrodes were placed in an open-cut down procedure on the vagus nerve. To limit variables of the comparison, Injectrodes were fabricated outside the body. A) Stimulation dose–titration was performed with a current-controlled simulator. The experiment was performed first with a standard LivaNova electrode, followed by bipolar Injectrode contacts which were placed in the same location and configuration on the vagus nerve. B) Higher stimulation currents were required with the Injectrode compared to the LivaNova electrode to drive the same heart rate response. C) However, impedance of each electrode was measured to be 1350 Ω for the LivaNova electrode and 360 Ω for the Injectrode. Correcting for lower impedance of Injectrode connection, dose–titration on Voltage-BPM scale was nearly identical with one another, demonstrating a characteristic sigmoidal response relationship. In panels (B) and (C), spline curves were fit to each dataset to improve visualization of the relationship between stimulation and response.
2.9. Benefits, Challenges, and Future Directions of the Injectable Electrode
The Injectrode is a unique electrode technology that can be injected onto, into, or around a target structure, where it cures in vivo to form a conductive tissue interface. The fluidity of the prepolymer material when it is injected allows for the electrode to conform to the patient’s individual neuroanatomy and thus provides a scalable approach to the size and geometry of different target structures. This gives the Injectrode the potential to simplify the surgical approach and eliminate complicated electrode designs needed for difficult neuroanatomical targets such as the stimulation of a nerve plexus presented in Figure 9. Additionally, the elastomeric properties of the cured material have a Young’s modulus of under 100 kPa. Additionally, the elastomeric properties of the cured material have a Young’s modulus of under 100 kPa. While this is slightly stiffer than the peripheral nerves, it is still orders of magnitude less stiff than traditional neural interface materials made from solid wires with polymer insulation. Reducing the mechanical mismatch between the device and neural tissue, has, in other cases, reduced cyclic strains and stresses on the implanted device and surrounding anatomy.[39,40] Chronic implant studies are required to determine if this holds true with the Injectrode and if there are indeed any long-term advantages with the interface. Despite the benefits derived from the material properties of the Injectrode there are a number of design considerations that will need to be addressed by future studies, including the use of conductive metal particles with known chronic biocompatibility such as gold, platinum, or stainless steel.
A number or previous studies have taken advantage of silicone (PDMS) composites doped with conductive filler particles in order to create a conductive material,[41–43] including for use the development of neural electrodes.[44] Additionally, other in-body curing PDMS composite materials have been proposed in the literature[45,46] and are commonly used in small animal studies[47–50] due to PDMS being physiological inert and favorable curing properties; PDMS cures in an aqueous environment without the formation of toxic products or exothermic heat.[45] These advancements in conductive and in-body curing polymers have not been applied to create an injectable electrode, but they provide a basis for the use of silver particles as a conductive filler within a stretchable silicone substrate. Based on these previously published studies, and the low cost of silver for iterating through the proof-of-concept experiments presented in this paper, in the present proof-of-concept study silver particles were selected as the conductive filler in the composite material. Despite its use as an antimicrobial in certain medical applications, in high doses and during long-term contact with tissue, dissolution of silver metals and the formation of Ag salts are well known to be associated with cellular and organ toxicity.[19] As such, the exact Injectrode formulation used in this paper is not viable as a chronic neural interface. To overcome this limitation, we are currently testing Injectrode formulations based on the use of stainless steel, gold, platinum, and other nonmetallic filler particles to provide conductivity. Due to the need for future changes in the conductive filler used in the composite Injectrode material, we focused the electrochemical testing presented in this paper on demonstrating similarity to the base conductor (silver) and showing that the composite material construct is capable of neuromodulation relevant charge storage capacities and charge injection limits. This was sufficient for this proof-of-concept demonstration of feasibility, in future studies we intend to emphasize developing the neural interface in terms of both bio-electrochemical compatibility and material flexibility.[51]
Another important point of future development is the Injectrode implantation procedure. While the experiments presented herein were performed through an open cut down to the neuroanatomical target, the ultimate goal of the Injectrode concept is to enable the completion of these procedures through a minimally invasively surgical approach. Achieving this goal will require refinement and further development of surgical delivery tools and image guidance tools for accurately injecting multicontact Injectrodes in way that well-controlled and repeatable across surgeons. A particularly important aspect of non-invasive delivery will be the indirect visualization of the nerve target and associated instrumentation for accurately injecting the electrode followed imaging of the final Injectrode alignment for confirmation of target engagement. Typical imaging techniques for lidocaine nerve blocks utilize ultrasound imaging surrounding anatomical landmarks, for example, bony structures and muscle fascia planes. Similarly, the implantation of spinal cord stimulation electrodes utilizes fluoroscopic guidance of the electrode. For this reason, intraoperative ultrasound and fluoroscopic methods for visualizing the procedure are currently being explored; other imaging modalities may also be useful for intraoperative guidance or postoperative verification of targeting. In addition to implantation of the conductive Injectrode, nonconductive polymer may also need to be delivered as insulation in order to isolate Injectrode contacts or direct current flow to the target of interest. However, the individual procedures for administration of conductive and nonconductive materials to isolate the neuroanatomy of interest will need to be optimized for specific indications.
The principal ingredients of the Injectrode were selected to provide rapid proof of concept, and because they have been used in different forms within other previously FDA-approved implantable products. Although the FDA does not approve specific materials, as performance depends on specific use in the context of an entire system, familiarity for implantable use can streamline the regulatory process. Similarly, surgical glues that cure within the body are not uncommon in clinical use and provide a model to follow for conductive Injectrode based solutions.
2.10. Advantages and Disadvantages of the Injectrode Compared to Other Recent Innovations in Soft Electrodes
Over the past decade, there has been a renewed interest in the use of thin films and other “soft” electrode materials to create high-channel, yet low geometric area electrodes with a Young’s modulus closer to native tissue to minimize initial surgical trauma and the chronic immune response.[50–56] These electrode development strategies include the use of PDMS substrates with conductive traces and electrode contacts to develop stretchable multicontact electrode arrays for stimulation and recording.[44,57–60] Additionally, shape memory polymer (SMP) or other “shuttles” have been used to create a stiff carrier to facilitate implantation of electrodes into the brain, yet leave a soft electrode chronically to minimize the immune response of tissue.[50,54,56] Similarly, the Charles Lieber group has demonstrated that a ultraflexible open mesh electrode array can be injected into the rodent brain via syringe and minimize the chronic immune response, enabling high-quality chronic recordings for periods of at least 12 weeks.[61] More recently, the Romero-Ortega group has demonstrated that SMP cuffs can be surgically implanted directly on the rat sciatic and pelvic nerves and both record and intermittently stimulate for periods of at least 30 d, while minimizing the immune response compared to stiffer control devices.[62] In comparison to the Injectrode, these promising technologies have the clear advantage of higher channel counts, as well as predictable electrode geometries and spacing.
However, extensibility of these technologies to an embodiment compatible with minimally invasive delivery outside the brain yet capable of high-duty cycle continuous stimulation common for clinical neuromodulation therapies remains unclear. At present, the only high-density thin-film electrode technology that is FDA market approved for over 30 d use is the Argus II Retinal Prostheses from Second Sight. Thin film polymers are permeable to water, and tend to fail chronically as fluid ingress finds pinhole defects between depositions layers which eventually lead to adhesion failures between layers and ultimately insulation between adjacent electrodes.[63–66] Electrical stimulation through these devices exacerbates the failure modes by two mechanisms. First, electrical stimulation through an ultrathin platinum or platinum/iridium electrode causes transient mechanical deformation of the electrode, leading to lack of conformance to the insulation and failure of the lead traces connecting to the electrode.[23,67,68] Second, the application of a DC bias for active electronics or normal charge-balanced constant current stimulation is known to hasten device failure, putatively by exacerbating fluid ingress through the polymer layers.[63,65] As of the writing of this paper, the long-term chronic viability of the Injectrode is also unproven; however, the Injectrode results presented here demonstrate that the metal particle-based strategy performs electrochemically under stimulation like a pure wire of the same material. This suggests that the chronic Injectrode behavior will resemble that of already clinically proven gold and platinum metal electrodes of similar geometric area.
By comparison to existing soft polymer electrode strategies, the Injectrode has the potential advantages of being very simple and inexpensive to manufacture. This is particularly important for viability in translation, as insurance payers determine the reimbursement price for implantable neuromodulation technology. Complexity increases the number of failure points, complicates the supply chain for manufacture, and ultimately increases the overall cost of manufacture. This is why simple implantable devices based off variants of the cardiac pacemaker but modified for the cervical vagus, deep brain, or spinal cord currently dominate the existing neuromodulation industry. Unlike the existing soft polymer strategies which have to be modified based on surgical approach and intended neural target, the Injectrode as a platform technology is extensible to injectable surgical deployment around any complex neural structure without modification to create highly conformal electrode interfaces. This also potentially enables unique uses such as injectable delivery of undoped Injectrode to insulate nerves implicated in therapy limiting off-target effects, or to “reconnect” existing clinical leads that, through scarring or migration, no longer interface efficiently with their intended neural targets. Consequently, we envision both the Injectrode and high-density “soft” thin film electrodes will likely each have unique uses and therefore their own “niche” in the quickly evolving neurotechnology clinical market.
3. Conclusion
The Injectrode was developed with the goal of addressing problems with invasiveness, complexity, and cost of the implantation procedure that hinder the adoption of neuromodulation therapies and reduce patient’s access to these otherwise promising treatments. In order to develop the Injectrode into a chronically implanted neural interface, optimization of material properties and surgical approaches will need to be addressed in future studies. Additionally, long-term electrochemical testing of the Injectrode and histological analysis showing chronic stability and biocompatibility of the implant will need to be performed. However, the electrochemical testing and proof-of-concept experiments presented in this paper demonstrate the feasibility of the injectable electrode approach. Electrochemical testing, performed based on common FDA preclinical benchtop tests, including scanning electron microscopy, cyclic voltammetry, electrochemical impedance spectroscopy, and voltage transient analysis showed that the Injectrode, which is conductive due to the incorporation of metal filler particles into a composite material, is electrochemically similar to metal wire of the same material (silver). Additionally, acute in vivo testing of an Injectrode, cured around the complex compound motor nerve branches of the brachial plexus, demonstrated stimulation-induced tetanic activation of the terminal muscles. Finally, in an animal model better matching the scale of human anatomy, proof-of-concept comparisons of Injectrode performance to a clinical LivaNova vagus nerve stimulation lead showed stimulation-induced heart rate changes in the swine. These experiments suggest the feasibility of the Injectrode for neural stimulation. Additionally, by virtue of being simpler than traditional electrode designs the Injectrode has the potential to increase the adoption of neuromodulation therapies and enable the treatment of new indications.
4. Experimental Section
Materials and Equipment:
All general supplies, chemicals, reagents, and buffers were sourced from Sigma (St. Louis, MO) and used as received. Mixtures of two-part Pt-curing silicone elastomer and metallic silver flakes were prepared according to instructions from Neuronoff, Inc. (Valencia, CA). The silicone elastomers were purchased from WPI Inc. (Kwik-Cast), and used as received. According to their MSDS sheet, this silicone is primarily composed of vinyl dimethyl polydimethylsiloxane, silicon dioxide, and polymethylhydrogen siloxane (base paste). The catalyst is a proprietary organo-platinum complex (catalyst paste). Readers are referred to other publications regarding the mechanisms of silicone crosslinking by addition reactions.[69] Blend percentages are defined according to a weight/volume (w/v) nomenclature with respect to silver/silicone, where 80% w/v corresponds to 800 mg silver flakes and 200 mg mixed silicone prepolymer. The silver flakes (25–50 μm aggregate size) were purchased from Inframat Advanced Materials and used as received. The flakes were additionally pretreated with (3-glycidyloxypropyl)-trimethoxysilane (GLYMO) to improve their incorporation into the silicone matrix according to manufacturer recommendations.
To control geometry for the benchtop characterization experiments, materials were mixed thoroughly and immediately transferred to cure in either a 96-well plate or a 1/16 in. inner diameter (ID) silicone tubing for incorporation into a working electrode, as shown in Figure 11. For the in vivo experiments, materials were mixed and immediately transferred to a syringe for injection (extrusion) around the nerve of interest.
Figure 11.

Working electrode fabrication schematic. A) A short segment of Injectrode material was filled into a 1/16 in. ID tubing and an uninsulated silver wire was inserted into one end. A larger ID tubing was placed over the smaller tubing to form an insulative outer sheath and filled with silicone. The constructed working electrode provides a consistent 1/16 in. diameter disk of Injectrode exposed at one end with a silver wire exposed at the other to provide a connection for performing electrochemical experiments. B) (Left) An image of a completed Injectrode sample fabricated into a working electrode. (Right) Cross section of working electrode tip showing the Injectrode sample with a controlled geometric surface area. C) Diagram of a three-electrode electrochemical setup used to perform electrochemical testing. D) Diagram of electrochemical setup used for acquiring voltage transient data, with a current source delivering the waveform between the working electrode (WE) and counter electrode (CE) as well as an oscilloscope for measuring applied voltage with respect to a silver/silver chloride (Ag/AgCl) reference electrode (REF).
Electrochemical Characterization:
Benchtop electrical characterization generally followed the techniques described in Stuart Cogan’s 2008 review of neural stimulation and recording electrodes.[17,28,29,70–74] To achieve consistent electrochemical testing, working electrodes with tightly controlled geometric surface area were fabricated. A prepolymer that was 80% (w/v) silver particles was mixed as previously described and immediately injected into a 1/16 in. ID tube (labeled Inner Tubing in Figure 11), which was then cut into 2 cm segments. While curing, a 0.25 mm diameter silver wire was embedded ≈1 cm deep into each Injectrode sample. After curing, the inner tubing piece was placed into a second piece of tubing (labeled Outer Tubing in Figure 11) chosen with an ID slightly larger than the outer diameter (OD) of the inner tubing. The electrode was then back-filled with silicone, which served to provide mechanical stability and insulate the silver wire. This yielded an electrode with a 1/16 in. diameter circular disc Injectrode material exposed at one end, which was used as a working electrode for electrochemical characterization experiments measuring cyclic voltammetry, electrochemical impedance spectroscopy, and voltage transients. A schematic of the fabricated working electrode is shown in Figure 11A and photographs showing both a working electrode sample and magnified cross section of the electrode surface are presented in Figure 11B.
For comparison of Injectrode samples to a pure silver electrode, silver working electrodes were fabricated by removing, under a microscope, insulation from 0.01 in. silver wire (A-M Systems, Carlsborg, WA) such that the total exposed surface area matched the geometric surface area of the Injectrode, which was 0.0197 cm2. These silver working electrodes will be referred to as “silver wire electrodes.” In order to compare properties of the Injectrode to a clinically relevant electrode, a LivaNova PerenniaFLEX stimulation electrode with platinum/iridium contacts was used and will be consistently referred to as the “LivaNova electrode.”
Cyclic voltammetry and electrochemical impedance spectroscopy were measured in 0.01 m PBS (NaCl 0.138 m; KCl 0.0027 m; pH 7.4; Sigma, St. Louis, MO) using a three-electrode cell consisting of a previously described Injectrode working electrode, a platinum sheet counter electrode (1 cm2 surface area; Metrohm, Herisau, Switzerland), and a single junction silver/silver chloride (Ag/AgCl) in 3 m KCl reference electrode (BASi, West Lafayette, IN). Measurements were performed with an AutolabPGSTAT128N potentiostat (Metrohm).
Cyclic voltammetry was performed with a staircase sweep at 50 mV s−1 with a step size of 2.44 mV. 15 voltammograms were performed on each Injectrode sample and the final three voltammograms on each sample were averaged to obtain a single measurement. A new working electrode was used for each measurement to minimize effects from previous measurements on any given sample. In order to compare the electrochemical characteristics of the Injectrode to that of a silver wire electrode fabricated as previously described, extended voltammograms from −1.9 to 3.0 V versus Ag/AgCl were also applied to both electrode types. In order to compare properties of the Injectrode to that of a clinically used LivaNova electrode, cyclic voltammograms from −0.62 to 0.9 V versus Ag/AgCl, a common range for platinum, were applied. Prior to the collection of voltammetry measurements, the LivaNova electrode was cleaned by sonicating for 30 min in isopropyl alcohol, then an additional 30 min in deionized (DI) water, followed by anodic etching for 2 min at an applied potential of 2 V versus Ag/AgCl in a 0.5 m sulfuric acid solution. From the collected voltammograms, the CSCC was calculated for each electrode as the time integral of the cathodic current.
To characterize the impedance properties of the Injectrode and compare it to silver wire electrodes and the LivaNova electrode, EIS was performed with the same electrodes and electrochemical cell as was used for cyclic voltammetry. Electrochemical impedance spectroscopy was performed following the application of 20 voltammograms from −0.62 to 0.9 V versus Ag/AgCl. Impedance spectroscopy was performed with 25 mV sinusoidal waveforms applied with equal spacing on a logarithmic scale from 0.1 Hz to 100 kHz.
Finally, VT measurements were taken, once again, following the application of 20 voltammograms from −0.62 to 0.9 V versus Ag/AgCl. A cathodic-leading charge-balanced biphasic waveform was used for the VT measurements with pulse widths of 500 μs, and a 10 μs interpulse delay. Pulses were applied at 50 Hz and VT measurements were made following at least 1000 pulses in order to allow for stabilization. A Keithley 6221 current source (Tektronics, Beaverton, OR) was used for the application of pulse trains used for VTs. Using a Tektronics TBS1154 Oscilloscope (Tektronics, Beaverton, OR), VTs were recorded with the application of increasing amplitude pulses until the maximum cathodic polarization (Emc) was greater than 0.6 V. To minimize the effects of noise, VT measurements were recorded as the average of 16 pulses using built-in averaging on the oscilloscope.
Electrode Imaging:
Light and electron microscopy of the electrode surface were performed on an unstimulated Injectrode sample from a working electrode that was fabricated as described in the previous section. In preparation for imaging, a cross-sectional slice (Figure 11B) of an Injectrode sample that had not undergone electrochemical testing or stimulation was cut with a scalpel.
Light microscopy was performed with an AmScope ZM-4TW3 stereo microscope, equipped with an 18 MP camera (AmScope, Irvine, CA). Images were converted to grayscale and contrast enhanced by histogram normalization in ImageJ (National Institutes of Health, Bethesda, MD). The slice was then platinum-coated in preparation for SEM. SEM was performed using a Zeiss LEO 1530 (Zeiss International, Oberkochen, Germany) with an accelerating voltage of 3.0 kV.
Mechanical Testing:
ASTM International has specified procedures for evaluating mechanical properties of thermoset plastics under Active Standard D412-16. Specified forms include flat specimens with uniform cross-sectional area, and dog bone shaped specimens. Molds were designed as closely as possible adhering to the guidelines for creating flat dog bone shaped specimens in SolidWorks. The molds were then cut from acrylic sheets using a laser cutter at ThinkBox (Case Western Reserve University). The mechanical properties of the cured electrodes were then tested using a pneumatically driven universal testing machine (Enduratec) located in the CWRU Department of Materials Science & Engineering Advanced Manufacturing and Mechanical Reliability Center (AMMRC). Specimens (n = 4, independently mixed) were placed into the grips of the testing machine pulled under tension at a rate of 0.5 mm s−1. Grip spacing was set to 2 cm. The first sample was tested to 140% strain and did not fail. Therefore, the maximum strain was increased for subsequent samples, which were pulled to a final extension of 6 cm (200% strain) or failure.
Acute Rat Brachial Plexus Stimulation:
All animal experiments were approved by the Mayo Clinic Institutional Animal Care and Use Committee (Rochester, MN). Rats were anesthetized with isoflurane, using 5% isoflurane in pure oxygen for induction and 1–2% for maintenance over the duration of the experiment. The compound motor nerve branches of the rat brachial plexus were exposed and freed from surrounding tissues via careful dissection through the pectoralis muscles. Components of the Injectrode composite were combined in a syringe and mixed for 30 s to make the flowable prepolymer, as previously described. The prepolymer mixture was injected within 1–2 min to create two circumferential “cuff” electrodes in a bipolar configuration with 5 mm spacing, around the exposed rat brachial plexus. Silver wires were inserted into the composite during curing to make direct wired connections to the cuffs. For the purposes of this proof-of-concept experiment testing, the Injectrode in an open surgical site, Kwik-Cast (World Precision Instruments, Sarasota, FL) was used to insulate between and around the two electrodes.
A biphasic charge balanced waveform at 30 Hz (200 μs pulse width, 5 mA amplitude) was applied using Master-8 stimulator (A.M.P.I., Jerusalem, Israel) with Iso-Flex stimulus isolators (A.M.P.I., Jerusalem, Israel). Effect of stimulation was observed visually as large muscle contractions of the forelimb. Approximate joint angles were calculated using post hoc video analysis. In brief, still images were captured from the video every five frames (30 fps), and analyzed in ImageJ. Two lines were drawn through the joint. The angle was measured using ImageJ’s built-in tools.
Acute Swine Vagus Nerve Stimulation:
Pigs (weight 35–40 kg range) were anesthetized using Telazol/Xylazine for induction followed by maintenance with 1–2% isoflurane for the duration of the experiment. The vagus nerve was carefully dissected free from surrounding tissue and a commercially available bipolar cuff (LivaNova) was placed on the nerve ≈0.5 to 1.0 cm inferior to the nodose ganglion and 1.0 to 1.5 cm from the carotid bifurcation. The cuffs were placed on the ventromedial aspect of the vagus nerve.
A stimulation current–dose–titration curve was generated, with a biphasic charge balanced waveform at 30 Hz (200 μs cathodic pulse width), ranging from 0 to 5 mA. Heart rate was monitored with a pulse oximeter and recorded.
To compare results obtained during stimulation with the Injectrode to those obtained using clinical LivaNova electrodes, cuff-like electrodes were fabricated using Injectrode material extruded onto a stainless-steel mesh with an insulated backing. The same stimulation current–dose–titration curve was generated and compared to the results obtained while stimulating with the LivaNova electrode.
Supplementary Material
Acknowledgements
The authors would like to acknowledge funding from The Defense Advanced Research Projects Agency (DARPA) Biological Technologies Office (BTO) Targeted Neuroplasticity Training Program under the auspices of Doug Weber and Tristan McClure-Begley through the Space and Naval Warfare Systems Command (SPAWAR) Systems Center with (SSC) Pacific grant no. N66001-17-2-4010. The authors would also like to acknowledge Erika Woodrum and the FES Center for their help in constructing Figures 1 and 2, and to Janet Gbur and John Lewandowski at the Advanced Manufacturing and Mechanical Reliability Characterization Center for their assistance in collecting mechanical testing data. Additionally, the authors gratefully acknowledge use of facilities and instrumentation supported by NSF through the University of Wisconsin Materials Research Science and Engineering Center (DMR-1720415). Funding from the National Institute of Health (1U18EB029251-01) and The Grainger Foundation are also acknowledged. Finally, the authors would like to thank Seth Hara from the Mayo Clinic for his feedback on this manuscript.
Footnotes
Supporting Information
Supporting Information is available from the Wiley Online Library or from the author.
Conflict of Interest
M.F. and A.J.S. are co-founders of Neuronoff Inc. and co-inventors on intellectual property relating to the Injectrode. K.A.L. is a consultant to and co-founder of Neuronoff Inc. J.W. and K.A.L. are scientific board members and have stock interests in NeuroOne Medical Inc. J.W. also has an equity interest in NeuroNexus technology Inc. Additionally, K.A.L. is a paid member of the scientific advisory board of Cala Health, Blackfynn, and Battelle as well as a paid consultant for Galvani.
Contributor Information
James K. Trevathan, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA
Ian W. Baumgart, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA
Evan N. Nicolai, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA
Brian A. Gosink, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Department of Neurosurgery, University of Wisconsin-Madison, Madison, WI 53706, USA
Anders J. Asp, Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA
Megan L. Settell, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA
Shyam R. Polaconda, Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH 44106, USA
Kevin D. Malerick, Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH 44106, USA
Sarah K. Brodnick, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA
Weifeng Zeng, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA.
Bruce E. Knudsen, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA
Andrea L. McConico, Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA
Zachary Sanger, Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA.
Jannifer H. Lee, Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA
Johnathon M. Aho, Division of General Thoracic Surgery, Mayo Clinic, Rochester, MN 55902, USA; Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN 55902, USA
Aaron J. Suminski, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA
Erika K. Ross, Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA
Jose L. Lujan, Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55902, USA; Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA; Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN 55902, USA
Douglas J. Weber, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15260, USA
Justin C. Williams, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA
Manfred Franke, Neuronoff Inc., Valencia, CA 91354, USA.
Kip A. Ludwig, Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI 53706, USA; Department of Neurosurgery, University of Wisconsin-Madison, Madison, WI 53706, USA; Department of Neurologic Surgery, Mayo Clinic, Rochester, MN 55902, USA; Neuronoff Inc., Valencia, CA 91354, USA
Andrew J. Shoffstall, Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH 44106, USA; Neuronoff Inc., Valencia, CA 91354, USA; Advanced Platform Technologies Center, Louis Stokes Cleveland Veterans Affairs Medical Center, Cleveland, OH 44106, USA
References
- [1].Birmingham K, Gradinaru V, Anikeeva P, Grill WM, Pikov V, McLaughlin B, Pasricha P, Weber D, Ludwig K, Famm K, Nat. Rev Drug Discovery 2014, 13, 399. [DOI] [PubMed] [Google Scholar]
- [2].Famm K, Litt B, Tracey KJ, Boyden ES, Slaoui M, Nature 2013, 496, 159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Reardon S, Nature 2014, 511, 18. [DOI] [PubMed] [Google Scholar]
- [4].Dones I, Levi V, Brain Sci. 2018, 8, 138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Heiduschka P, Thanos S, Prog. Neurobiol 1998, 55, 433. [DOI] [PubMed] [Google Scholar]
- [6].Güemes A, Georgiou P, Bioelectron. Med 2018, 4, 9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Bonaz B, Bioelectron. Med 2018, 4, 4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Pavlov VA, Tracey KJ, Nat. Rev. Endocrinol 2012, 8, 743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Koopman FA, Chavan SS, Miljko S, Grazio S, Sokolovic S, Schuurman PR, Mehta AD, Levine YA, Faltys M, Zitnik R, Tracey KJ, Tak PP, Proc. Natl. Acad. Sci. USA 2016, 113, 8284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Mosier EM, Wolfson M, Ross E, Harris J, Weber D, Ludwig KA, in Neuromodulation, 2nd ed. (Eds: Krames ES, Peckham PH, Rezai AR), Academic, San Diego, CA: 2018, pp. 55–68. [Google Scholar]
- [11].Tarn J, Legg S, Mitchell S, Simon B, Ng W-F, Neuromodulation: Technol. Neural Interface 2019, 22, 580. [DOI] [PubMed] [Google Scholar]
- [12].Silberstein SD, Mechtler LL, Kudrow DB, Calhoun AH, McClure C, Saper JR, Liebler EJ, Engel ER, Tepper SJ, Headache: J. Head Face Pain 2016, 56, 1317. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Hofstoetter US, Freundl B, Binder H, Minassian K, PLoS One 2018, 13, e0192013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Plonsey R, Barr RC, IEEE Trans. Biomed. Eng 1995, 42, 329. [DOI] [PubMed] [Google Scholar]
- [15].Tyler DJ, Neuroprosthetics: Theory and Practice, World Scientific, Singapore: 2017, pp. 300–347. [Google Scholar]
- [16].Kumar K, Bishop S, Neurosurg J. 2009, 10, 564. [DOI] [PubMed] [Google Scholar]
- [17].Cogan SF, Annu. Rev. Biomed. Eng 2008, 10, 275. [DOI] [PubMed] [Google Scholar]
- [18].Hudak EM, Mortimer JT, Martin HB, J. Neural Eng 2010, 7, 026005. [DOI] [PubMed] [Google Scholar]
- [19].Drake PL, Hazelwood KJ, Ann. Occup. Hyg 2005, 49, 575. [DOI] [PubMed] [Google Scholar]
- [20].Merrill DR, Stefan IC, Scherson DA, Mortimer JT, J. Electrochem. Soc 2005, 152, E212. [Google Scholar]
- [21].Choi Y-J, Luo T-JM, Int. J. Electrochem 2011, 2011, 404937. [Google Scholar]
- [22].Abd El Rehim SS, Hassan HH, Ibrahim MAM, Amin MA, Monatsh. Chem 1998, 129, 1103. [Google Scholar]
- [23].Wang Y-H, Gu H-Y, Microchim. Acta 2009, 164, 41. [Google Scholar]
- [24].Cui X, Martin DC, Sens. Actuators, B 2003, 89, 92. [Google Scholar]
- [25].Cui X, Hetke JF, Wiler JA, Anderson DJ, Martin DC, Sens. Actuators, A 2001, 93, 8. [DOI] [PubMed] [Google Scholar]
- [26].Ludwig KA, Uram JD, Yang J, Martin DC, Kipke DR, J. Neural Eng 2006, 3, 59. [DOI] [PubMed] [Google Scholar]
- [27].Weiland JD, Anderson DJ, IEEE Trans. Biomed. Eng 2000, 47, 911. [DOI] [PubMed] [Google Scholar]
- [28].Merrill DR, Bikson M, Jefferys JGR,Neurosci J. Methods 2005, 141, 171. [DOI] [PubMed] [Google Scholar]
- [29].Wilks SJ, Hara SA, Ross EK, Nicolai EN, Pignato PA, Cates AW, Ludwig KA, Front. Neurosci 2017, 11, 438. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Clark GM, Clark JCM, Furness JB, JAMA, J. Am. Med. Assoc 2013, 310, 1225. [DOI] [PubMed] [Google Scholar]
- [31].Clark GM, Clark J, Cardamone T, Clarke M, Nielsen P, Jones R, Arhatari B, Birbilis N, Curtain R, Xu J, Wagstaff S, Gibson P, O’Leary S, Furness J, Cochlear Implants Int. 2014, 15, S1. [DOI] [PubMed] [Google Scholar]
- [32].Nadol JB, O’Malley JT, Burgess BJ, Galler D, Hearing Res. 2014, 318, 11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [33].Spiers K, Cardamone T, Furness JB, Clark JCM, Patrick JF, Clark GM, Cochlear Implants Int. 2016, 17, 129. [DOI] [PubMed] [Google Scholar]
- [34].Cathodic leading pulses are traditionally used for voltage excursion tests, as cathodic leading biphasic pulses require lower thresholds for activation when an electrode is placed directly on a nerve fiber. When there is any separation between the electrode and the fibers, or a group of fibers are being stimulated that cover a span of distances from the electrode, there may be less of an advantage to the use of cathodic first pulsing strategies.
- [35].Tyler DJ, in Neuromodulation, 2nd ed. (Eds: Krames ES, Peckham PH, Rezai AR), Academic, San Diego, CA: 2018, pp. 239–274. [Google Scholar]
- [36].Engineering ToolBox, Young’s Modulus of Elasticity for Metals and Alloys, https://www.engineeringtoolbox.com/young-modulus-d_773.html (accessed: March 2018).
- [37].Giannessi E, Stornelli MR, Coli A, Sergi PN, Appl. Sci 2019, 9, 1115. [Google Scholar]
- [38].Premchand RK, Sharma K, Mittal S, Monteiro R, Dixit S, Libbus I, DiCarlo LA, Ardell JL, Rector TS, Amurthur B, KenKnight BH, Anand IS, J. Cardiac Failure 2014, 20, 808. [DOI] [PubMed] [Google Scholar]
- [39].Sridharan A, Nguyen JK, Capadona JR, Muthuswamy J, J. Neural Eng 2015, 12, 036002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [40].Nguyen JK, Park DJ, Skousen JL, Hess-Dunning AE, Tyler DJ, Rowan SJ, Weder C, Capadona JR, J. Neural Eng 2014, 11, 056014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [41].Cong H, Pan T, in 2009 4th IEEE Int. Conf. Nano/Micro Engineered and Molecular Systems, IEEE, Shenzhen, China 2009, pp. 731–734. [Google Scholar]
- [42].Araki T, Nogi M, Suganuma K, Kogure M, Kirihara O, IEEE Electron Device Lett. 2011, 32, 1424. [Google Scholar]
- [43].Niu XZ, Peng SL, Liu LY, Wen WJ, Sheng P, Adv. Mater 2007, 19, 2682. [Google Scholar]
- [44].Larmagnac A, Eggenberger S, Janossy H, Vörös J, Sci. Rep 2015, 4, 7254. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [45].Bosman W-MPF, van der Steenhoven TJ, Suárez DR, Valstar ER, de Vries AC, Brom HLF, Jacobs MJ, Hamming JF, J. Vasc. Surg 2010, 52, 152. [DOI] [PubMed] [Google Scholar]
- [46].Ignjatovic N, Jovanovic J, Suljovrujic E, Uskokovic D, Bio-Med. Mater. Eng 2003, 13, 401. [PubMed] [Google Scholar]
- [47].Burgalossi A, von Heimendahl M, Brecht M, Front. Neural Circuits 2014, 8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [48].Dryg ID, Ward MP, Qing KY, Mei H, Schaffer JE, Irazoqui PP, IEEE Trans. Neural Syst. Rehabil. Eng 2015, 23, 562. [DOI] [PubMed] [Google Scholar]
- [49].Kawada T, Li M, Kamiya A, Shimizu S, Uemura K, Yamamoto H, Sugimachi M, J. Physiol. Sci 2010, 60, 283. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [50].Harris JP, Capadona JR, Miller RH, Healy BC, Shanmuganathan K, Rowan SJ, Weder C, Tyler DJ, Neural Eng J. 2011, 8, 066011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Du ZJ, Kolarcik CL, Kozai TDY, Luebben SD, Sapp SA, Zheng XS, Nabity JA, Cui XT, Acta Biomater. 2017, 53, 46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].McCallum GA, Sui X, Qiu C, Marmerstein J, Zheng Y, Eggers TE, Hu C, Dai L, Durand DM, Sci. Rep 2017, 7, 11723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [53].Sohal HS, Jackson A, Jackson R, Clowry GJ, Vassilevski K, O’Neill A, Baker SN, Front. Neuroeng 2014, 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [54].Shoffstall AJ, Ecker M, Danda V, Joshi-Imre A, Stiller A, Yu M, Paiz JE, Mancuso E, Bedell HW, Voit WE, Pancrazio JJ, Capadona JR, Micromachines 2018, 9, 486. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [55].Luan L, Wei X, Zhao Z, Siegel JJ, Potnis O, Tuppen CA, Lin S, Kazmi S, Fowler RA, Holloway S, Dunn AK, Chitwood RA, Xie C, Sci. Adv 2017, 3, e1601966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [56].Stiller AM, Usoro J, Frewin CL, Danda VR, Ecker M, Joshi-Imre A, Musselman KC, Voit W, Modi R, Pancrazio JJ, Black BJ, Micromachines 2018, 9, 500. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [57].Minev IR, Musienko P, Hirsch A, Barraud Q, Wenger N, Moraud EM, Gandar J, Capogrosso M, Milekovic T, Asboth L, Torres RF, Vachicouras N, Liu Q, Pavlova N, Duis S, Larmagnac A, Vörös J, Micera S, Suo Z, Courtine G, Lacour SP, Science 2015, 347, 159. [DOI] [PubMed] [Google Scholar]
- [58].Chou N, Yoo S, Kim S, IEEE Trans. Neural Syst. Rehabil. Eng 2013, 21, 544. [DOI] [PubMed] [Google Scholar]
- [59].Cuttaz E, Goding J, Vallejo-Giraldo C, Aregueta-Robles U, Lovell N, Ghezzi D, Green RA, Biomater. Sci 2019, 7, 1372. [DOI] [PubMed] [Google Scholar]
- [60].Viventi J, Kim D-H, Vigeland L, Frechette ES, Blanco JA, Kim Y-S, Avrin AE, Tiruvadi VR, Hwang S-W, Vanleer AC, Wulsin DF, Davis K, Gelber CE, Palmer L, Van der Spiegel J, Wu J, Xiao J, Huang Y, Contreras D, Rogers JA, Litt B, Nat. Neurosci 2011, 14, 1599. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [61].Hong G, Viveros RD, Zwang TJ, Yang X, Lieber CM, Biochemistry 2018, 57, 3995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [62].González-González MA, Kanneganti A, Joshi-Imre A, Hernandez-Reynoso AG, Bendale G, Modi R, Ecker M, Khurram A, Cogan SF, Voit WE, Romero-Ortega MI, Sci. Rep 2018, 8, 16390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [63].Wellman SM, Eles JR, Ludwig KA, Seymour JP, Michelson NJ, McFadden WE, Vazquez AL, Kozai TDY, Adv. Funct. Mater 2018, 28, 1701269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [64].Dissado LA, Fothergill JC, Electrical Degradation and Breakdown in Polymers, IET Digital Library, 1992. [Google Scholar]
- [65].Xie X, Rieth L, Caldwell R, Negi S, Bhandari R, Sharma R, Tathireddy P, Solzbacher F, Biomed. Microdevices 2015, 77, 1. [DOI] [PubMed] [Google Scholar]
- [66].Hsu J-M, Tathireddy P, Rieth L, Normann AR, Solzbacher F, Thin Solid Films 2007, 576, 34. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [67].Ordonez JS, Rudmann L, Cvancara P, Bentler C, Stieglitz T, in 2015 37th Annual Int. Conf. IEEE Engineering in Medicine and Biology Society (EMBC), IEEE, Piscataway, NJ 2015, pp. 1045–1048. [DOI] [PubMed] [Google Scholar]
- [68].Pfau J, Stieglitz T, Ordonez J, in 2017 8th Int. IEEE/EMBS Conf Neural Engineering (NER), IEEE, Piscataway, NJ 2017, pp. 166–169. [Google Scholar]
- [69].Colas A, Curtis J, Biomater. Sci 2004, 2, 80. [Google Scholar]
- [70].McCreery DB, Agnew WF, Yuen TGH, Bullara L, IEEE Trans. Biomed. Eng 1990, 37, 996. [DOI] [PubMed] [Google Scholar]
- [71].McCreery DB, Agnew WF, Yuen TG, Bullara LA, Med. Biol. Eng. Comput 1995, 33, 426. [DOI] [PubMed] [Google Scholar]
- [72].McCreery DB, Yuen TGH, Bullara LA, Hearing Res. 2000, 749, 223. [DOI] [PubMed] [Google Scholar]
- [73].McCreery D, Pikov V, Troyk PR, J. Neural Eng 2010, 7, 036005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [74].Shannon RV, IEEE Trans. Biomed. Eng 1992, 39, 424. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.