Table 5.
k-factor (mSv.mGy−1cm−1) for liver, chest, cardiac and abdomen-pelvis imaging protocols for Single Energy CT (SECT) and the Dual-energy Computed Tomography (DECT) for male and female phantoms from the present study and the study by Christner et al., [ 6 ].
| Imaging protocol | DECT (Present study) | SECT (Christner et al. [ 6 ]) | DECT (Christner et al. [ 6 ]) | |||
|---|---|---|---|---|---|---|
| Male | Female | Male | Female | Male | Female | |
| Liver | 0.018 | 0.023 | 0.015 | 0.015 | 0.017 | 0.017 |
| Chest | 0.011 | 0.021 | 0.014 | 0.014 | 0.019 | 0.019 |
| Cardiac | 0.013 | 0.029 | 0.014 | 0.014 | 0.031 | 0.031 |
| Abdomen-pelvis | 0.014 | 0.017 | 0.015 | 0.015 | 0.016 | 0.016 |
DECT: Dual Energy Computed Tomography, SECT: Single Energy CT