Skip to main content
. 2023 Aug 18;102(33):e34927. doi: 10.1097/MD.0000000000034927

Table 2.

Performance of the GenBody coronavirus disease 2019 rapid antigen based on antigen operator.

Non-laboratory staff (95% CI) Laboratory professionals (95% CI) Total (95% CI)
Sensitivity 79.92% (77.26%–82.39%) 79.80% (74.78%–84.22%) 79.89% (77.58%–82.06%)
Specificity 99.23% (98.73%–99.57%) 99.99% (99.94%–100.00%) 99.85% (99.76%–99.91%)
PLR 103.25 (62.31–171.11) 6983.92 (983.03–49617.00) 533.77 (326.87–871.63)
NLR 0.2 (0.18–0.23) 0.2 (0.16–0.25) 0.2 (0.18–0.22)
Diagnostic odds ratio 510.18 (299.81–868.18) 34566.45 (4770.30–250474.46) 2650.26 (1592.81–4409.74)
Prevalence 33.49% (31.78%–35.24%) 3.28% (2.92%–3.67%) 10.64% (10.09%–11.21%)
PPV 98.11% (96.91%–98.85%) 99.58% (97.09%–99.94%) 98.45% (97.50%–99.05%)
NPV 90.75% (89.64%–91.75%) 99.32% (99.15%–99.46%) 97.66% (97.39%–97.90%)
Accuracy 92.76% (91.76%–93.67%) 99.33% (99.13%–99.48%) 97.73% (97.44%–97.99%)

CI = confidence interval, NLR = negative likelihood ratio, NPV = negative predictive value, PLR = positive likelihood ratio, PPV = positive predictive value.