Table 4. Correction factor, n E th , to apply to the fluence threshold calculation ( Equation 11) when irradiating with truncated beam by a circular aperture.
The values are represented graphically on Figure 3(b).
| η E th | Aperture power transfer ( P T ) | |||||
|---|---|---|---|---|---|---|
| 90% | 75% | 60% | 40% | 25% | ||
| Maximum considered energy | 1.5·Eth | 1.002 | 1.002 | 1.002 | 1.002 | 1.002 |
| 2·Eth | 1.005 | 1.006 | 1.006 | 1.007 | 1.007 | |
| 2.5·Eth | 1.010 | 1.011 | 1.011 | 1.012 | 1.012 | |
| 3·Eth | 1.014 | 1.016 | 1.016 | 1.017 | 1.017 | |
| 3.5·Eth | 1.018 | 1.020 | 1.021 | 1.022 | 1.024 | |
| 4·Eth | 1.022 | 1.026 | 1.027 | 1.028 | 1.029 | |
| 4.5·Eth | 1.028 | 1.031 | 1.032 | 1.033 | 1.034 | |
| 5·Eth | 1.032 | 1.035 | 1.037 | 1.038 | 1.040 | |