Abstract
During the past 10 years, performing real-time molecular imaging with positron emission tomography (PET) in combination with computed tomography (CT) during interventional procedures has undergone rapid development. Keeping in mind the interest of the nuclear medicine readers, an update is provided of the current workflows using real-time PET/CT in percutaneous biopsies and tumor ablations. The clinical utility of PET/CT guided biopsies in cancer patients with lung, liver, lymphoma, and bone tumors are reviewed. Several technological developments, including the introduction of new PET tracers and robotic arms as well as opportunities provided through acquiring radioactive biopsy specimens are briefly reviewed.
Keywords: Positron emission tomography computed tomography, Biopsy, Ablation techniques
Introduction to positron emission tomography/computed tomography guided interventions
The evolution of invasive medical procedures goes hand in hand with the development of diagnostic imaging techniques. Success in image-guided intervention relies on the quality and accuracy of medical imaging.
One of the major revolutions in this intertwined process came in the second half of the 20th century, with the development of cross-sectional imaging methods such as computed tomography (CT), ultrasound (US), and magnetic resonance imaging (MRI).1 The possibility of delving deeper into the anatomy of the human body with high quality images allowed for a significant increase in the complexity of interventions. Such was the impact of this era that new medical specialties were born, including interventional radiology, cardiology, and neuroradiology. The development of hybrid imaging methods in the last 30 years pushed the boundaries of image-guided interventions one step further. Hybrid imaging opened the possibility of simultaneously evaluating anatomy and physiology, which means we can now look at different organs and structures and simultaneously analyze how they function.
Positron emission tomography (PET)/CT is the main representative of this new era of hybrid imaging, where two sets of cross-sectional images are created and later fused by a dedicated software (while still enabling analysis of each set of images separately).
Previously acquired PET images can also be merged or saved with intra-procedural CT or US images to integrate physiological information from PET with the detailed anatomy of cross-sectional imaging techniques.2 The utility of this multimodal fusion has been demonstrated for both image-guided biopsies and tumor ablations. However, this technique is limited due to the use of “old” PET images and registration accuracy.3 The challenges of registering previously acquired diagnostic images with procedural images include differences in patient position between preprocedure imaging and procedure imaging (e.g. procedures can be performed in the supine, lateral, or oblique positions), differences in breathing, arm positioning, and tissue or organ mobilization, or instrumentation during a procedure.4–7 This brief review contains a summary of what is already well established in the literature regarding the use of real-time PET/CT for guiding interventions, focusing on two main procedures: 1) biopsies; and 2) ablations.
Although several authors have previously published comprehensive descriptions of these procedures,2, 8–12 we provide herein updates on indications, techniques, clinical applications, advantages, and limitations, and discuss new technological advancements of each procedure in a review oriented towards nuclear medicine readers.
Indications for PET/CT guided interventions
Indications for PET/CT guided biopsies
Percutaneous biopsies performed under CT or US guidance are commonly used to obtain definitive tissue diagnoses.2, 13, 14 High diagnostic success rates have been reported for lesions that are visible on CT imaging, often greater than 90%. However, many lesions are not easily identified on the CT images used to guide the biopsy needle, even after careful examination of previously acquired CT, MR, or US images.
To overcome this issue, intra-procedural intravenous contrast injection may be administered to improve the visibility of the target.2, 13, 14 However, depending on the size and location of the lesion, the improvement by transient contrast may not be sufficient to ensure accurate targeting, especially for small lesions (less than 2 cm). Anatomical landmarks may also be used to guide the tissue collection, although confirmation of precise needle positioning may be impossible. More false negative biopsies are observed in these cases than in those of lesions with optimal visualization. Alternatively, navigation tools can be used to merge different imaging modalities, including US, MRI, or PET/CT, although breathing movement or differences in position can reduce the accuracy of target alignment. Using MRI requires MRI-safe needles and other specially designed auxiliary apparatuses and equipment.
For PET radiotracer avid lesions, the use of PET/CT improves targeting tumors that are not well defined by CT or other imaging methods.3–7 The advantage of this technique is that a radiotracer can help to distinguish benign masses from malignant, viable from nonviable, or biologically aggressive from non-aggressive regions within malignant masses. Another advantage of PET/CT is to demonstrate abnormalities that are invisible on anatomical images or before morphological or radiographic changes are present and can be detected. Different radiotracers maybe used for achieving higher specificity. 14
Indications for PET/CT guided ablations
PET/CT images may be used to visualize, and to place ablation applicators inside radiotracer avid lesions. 15 PET/CT allows an improved estimate of the viability of tumor cells within an ablation zone (AZ) compared with anatomical imaging methods such as conventional CT, MR, or US. In addition, PET/CT imaging also provides a metabolic evaluation of the tumor and the AZ before and immediately after treatment. Such evaluation is critical because accurate AZ assessment can reduce the number of ablation failures and the incidence of local tumor progression, thereby improving oncologic outcomes.16, 17 The immediate 2-deoxy-2-[18F]fluoro-D-glucose ([18F]FDG) PET/CT evaluation of the AZ using the split dose technique has been shown to be a surrogate image biomarker of local tumor progression.15, 16 The development of such intraprocedural markers for incomplete ablation could guide additional therapy, either by repeating or prolonging ablation, or by considering additional treatment options (locoregional, systemic, or both).
Techniques for PET/CT guided interventions
Common technical requirements and procedure
Equipment
The interventional PET/CT suite requirements were described in detail by Shyn,8 including facilities with enabled access for anesthesia and code team coverage, ample space for all associated equipment and procedures, and a shielded waiting room for the patient between radiopharmaceutical administration and start of the procedure. The equipment required for PET/CT guided interventions may vary depending on the particular interventions performed in each setting. The common need is the availability of a PET/CT scanner in the interventional suite for acquiring intraprocedural and registered PET and CT images.2 In-room monitors are necessary to guide the interventionalist. Due to the long bore of a PET/CT scanner, the anesthesia tubing and other leads need to be long enough to cover the entire range of patient bed travel.8 Because dedicating such equipment to interventions is expensive, in some countries and/or institutions, shared use of a diagnostic scanner may be necessary. However, regulations surrounding interventional procedures must be considered. Other helpful modifications include a direct view into the scanner bore from the control room or a video camera-based system for patient monitoring during scanning. A reliable communication system between the control and procedure rooms is needed.
A scanner with a larger bore (e.g. 70 cm or larger diameter) should be used to reduce the risk of the needles or catheters touching the scanner gantry during table motion and scanning. Also, to facilitate biopsy needle, or catheter insertion, in the target lesion, the PET/CT scanner would preferably be equipped with CT fluoroscopy capabilities with in-room display. PET scanners with time-of-flight and shorter timing resolution provide higher quality images for the same number of counts, which may allow reduction in the injected activities and therefore also of the personnel and patient doses.18 Newer generation PET/CT scanners, with longer axial length, have much higher sensitivity,19, 20 which may allow further reduction in personnel and patient doses and make the procedures more time efficient.
General procedure flow for PET/CT guided interventions
The patient preparation as well as the general procedure flow for interventional FDG PET/CT scans is described in detail by Shyn,8 who also points to the benefit of having any previous PET/CT images for comparison. Patient fasting times and preparation details depend on the procedure,8, 21 which begins with placing the patient in the optimal body position for the planned procedure, appropriate immobilization, and anesthesia. If the patient had a recent PET scan, an initial free breathing PET/CT is acquired for 1 or 2 bed positions, with 1 to 5 min per bed position when using time-of-flight PET/CT scanners with axial length of approximately 15 cm.8 Although initially full diagnostic [18F]FDG activities were injected,8, 9 to reduce personnel dose, the currently injected activity for guidance is typically lower and may vary between practices: 255±74 MBq10 or 222 MBq (6 mCi)2 have been used for biopsy guidance whereas 148 MBq (4 mCi)2 has been used for ablation guidance in the split-dose technique.15
Because the exact time of interventional procedures is difficult to predict, quantification accuracy may be compromised and uptake times may vary from 30 min up to approximately 4 h postinjection, which may still allow reasonable visualization for many lesions when 222 MBq [18F]FDG is administered. The interventional radiologist localizes the FDG-avid lesion relative to the anatomical structures on the fused PETpre/CTpre pre-needle/applicator insertion scans and uses that image to plan the insertion path.22 To verify needle position for each FDG-avid lesion, multiple short length CT scans (referred to as CTneedle) may be acquired and fused with PETpre obtained at the same position. When misregistration due to motion is suspected, an additional PET/CT scan with the needle in place can be obtained. If adequate anesthesia is available, a second PET scan may be performed with forced breathing suspension, and in such cases, 20 to 90 s accumulation per bed position after preoxygenating the patient to high oxygen concentration is used.2, 7, 9, 23 A postablation PET/CT scan as well as a three-phase CT may be obtained for ablation evaluation as described below.
Radiation dose to personnel
Ryan et al.9 have measured the radiation dose to the personnel from PET/CT guided interventions, using optically stimulated luminescence dosimeters placed on the chest under the protective lead apron from 12 cases (11 ablations, 1 biopsy) injected with median activity of 447.7 MBq [18F]FDG. The median effective dose received was 0.02 (range 0–0.13) mSv for the operator, 0.01 (range 0–0.05) mSv for the nurse anesthetist, and 0.02 (range 0–0.05) mSv for the radiology technologists. Thermoluminescence dosimeter finger rings indicated median extremity dose equivalent to the operator of 0.05 (range 0–0.62) mSv.9 Using in-room image guidance (e.g. fluoroscopy or US) for 6 of the 12 cases resulted in higher median extremity dose equivalent (0.32 mSv); correlation with the duration of in-room procedure guidance was observed, but statistical significance was not reached.9 In the above study, the patients were encouraged to void just prior to the initial PET/CT scan, which was between 36 min and 123 min after injection; the procedures (11 ablations, 1 biopsy) lasted between 121 and 386 min (median 276 min).9
In a radiofrequency ablation of liver metastases from neuroendocrine tumors using, on average, 386 MBq 6-L-[18F]fluorodihydroxyphenylalanine ([18F]FDOPA), split into two injections, for PET/CT guidance and for technical success evaluation, Cazzato et al.21 have reported maximum right hand dose of 0.149 mSv.
The above personnel doses are lower than those used in [18F]FDG guided surgeries, which were found to be within the occupational limits and were performed with similar or higher injected activities.24, 25 Because dose to the operator from fluoroscopy-guided procedures was estimated at approximately 0.025 mSv per case,26 it was concluded that “the operator dose from PET/CT guided procedures is not significantly different from fluoroscopically guided procedures.”9 Nevertheless, the operator dose can increase rapidly if the time spent close to the patient increases. Because lead aprons and glasses used for protection in fluoroscopy are ineffective in shielding from 511 keV annihilation photons,8, 27 personnel dose can be reduced by reduction of the preablation injected activity (148 MBq used by Ryan et al.15) as well as by extending the uptake time or by providing rolling lead shields.8 However, even if effective in shielding, the latter option may not be practical due to their weight and shape, which may interfere and even be prone to causing accidents in the often-tight interventional suites.
Slightly longer procedure times and slightly higher patient doses (up to 6.2±2.9 mSv) were observed in one institution28 in 117 PET/CT guided liver ablations when compared with 251 CT-only guided ablations. With the introduction of longer axial length PET scanners, which have higher sensitivity (2–5 times higher for imaging one organ19, 20), the injected activities and doses to personnel and patients can be reduced proportionately.
Techniques for PET/CT guided biopsies
To overcome the registration problems associated with using previously acquired images as discussed above, a percutaneous biopsy can be performed in the PET/CT suite.3–7, 13, 14 Biopsies performed in this way allow direct control of the tumor position and needle placement in real time, which can improve their accuracy. Patient positioning is based on factors such as the location of the target, the optimal needle path, and the shortest skin-to-target distance. A radiopaque grid or a custom needle locator is placed on the patient’s skin. Patients with recent full-body PET/CT are subjected to a limited length PET/CT scan (1 to 2 bed positions) around the biopsy region of interest. A CT is obtained in all cases for attenuation correction and anatomical coregistration. Immediately after the CT, PET images are acquired. The insertion and guidance of the needles are carried out using fluoroscopy, US, or CT. An additional one bed position PET/CT might be acquired to confirm the correct position of the needle. Specimens are collected. Usually, an 11–13-gauge biopsy system is used for all bone biopsies performed under local anesthesia whereas for soft tissue biopsies the needle diameter is usually 18–20 G. Using coaxial technique (the biopsy needle is inserted through another tube-like device with a slightly larger diameter which is inserted first) and fluoroscopy may help to reduce procedure time. Sedation of patients is also performed.
Techniques for PET/CT guided ablations
Percutaneous ablations may be performed under general anesthesia in a dedicated interventional PET/CT scan.15–17 The modality of ablation (microwave, radiofrequency, cryoablation or irreversible electroporation) is determined by the operator considering several factors, in particular the size of the tumor, the proximity of the tumor to the blood vessels, bile ducts, or other critical structures adjacent to the ablation target.
Guiding ablation using [18F]FDG PET/CT provides the advantage of persistent tumor visibility from beginning to end of the procedure, even after ablation.29, 30 Therefore, the operator does not lose sight of the tumor and can place probes more accurately, even when the tumor is obscured on the CT by metal instrument artifacts, atelectasis, hemorrhage, edema, etc. In addition, [18F]FDG PET/CT aids targeting for postablation biopsies if any are needed.
An intraprocedural split activity FDG-PET imaging protocol may be used to target and evaluate the ablation zone. At some of the practicing institutions each patient receives an initial FDG activity injection equivalent to one-third of the standard diagnostic FDG activity, followed by a second FDG injection equivalent to two-thirds of the standard FDG diagnostic activity injection at the end ablation.15 Patients remain on the table for the entire procedure. In each case, an initial, free-breathing set of PET/CT images is followed by a localized PET/CT scan without contrast and with a ventilator-assisted breathing PET. The placement of the ablation applicator is performed using conventional CT mode, CT fluoroscopy, and/or US guidance as needed. Serial CT images of the applicator are obtained and merged with the initial or later PET images to confirm precise targeting. The repositioning of the applicator and the overlapping ablations are carried out to obtain a minimum ablation margin of 5 mm around the target tumor.31 In all cases, the protocol recommended by the ablation device manufacturer is applied and completed. Following the ablations in addition to PET imaging, a contrast CT is often obtained to help assess the ablation area.
Clinical applications
Some of the clinical applications of PET/CT guided interventions have been addressed previously.10, 32–34 Here details are added for some of the applications by including updates and additional experiences.
Incorporation of intra-procedural PET/CT into biopsy guidance
Lung lesions
Improved diagnostic performance has been observed with PET/CT-guided lung biopsies in comparison with CT-guided biopsies, by directing the needle into the areas of highest FDG uptake as noted on PET.35, 36 Sampling these areas increases the probability of obtaining representative tumoral cells, which usually display high FDG uptake.37 This is particularly helpful in the investigation of large and heterogeneous pulmonary masses, in which false negative rates are higher when procedures are guided by conventional CT, due to frequent sampling of non-representative material such as necrosis or inflammatory tissue (Figure 1).38 Another advantage of PET/CT, if not the most important, is allowing access to lesions that appear solely as metabolic alterations on PET images, without corresponding anatomical changes on CT.39
Figure 1.
Example of a heterogeneous mass in the right lung biopsied with [18F]FDG PET/CT guidance. Notice the predominantly peripheral FDG uptake in the lesion. Through posterior access, the biopsy needle was directed to one of the regions with the highest metabolism; thus, improving the odds of yielding representative neoplastic material for histopathological analysis (in this case, the results revealed adenocarcinoma). A: PET/CT transaxial; B: CT transaxial; C: PET maximum intensity projection; red arrow: site of biopsy.
Liver
As observed in other sites, PET/CT for liver biopsy guidance is especially helpful for: 1) large and heterogeneous masses, in which directing the needle to areas of highest FDG uptake increases the probability of sampling representative tumor cells (Figure 2); and 2) FDG-positive lesions on PET without anatomical correspondence on CT,37, 40 which might enable earlier therapeutic interventions.
Figure 2.
Example of an [18F]FDG PET/CT guided liver biopsy of an extensive lesion occupying most of the right hepatic lobe. The needle was directed towards a region of high metabolism inside the heterogeneous lesion. Pathology studies revealed that this lesion was a hemangiosarcoma. A: PET/CT transaxial; B: CT transaxial; C: PET maximum intensity projection; red arrow: site of biopsy.
PET/CT information can also aid biopsy of liver lesions through other imaging methods, such as fusion-guided sonography, improving diagnostic accuracy in comparison with conventional B-mode US.41
Benign liver lesions (hemangiomas, focal nodular hyperplasia, hepatocellular adenomas) are not usually FDG avid, which hinders the applicability of PET/CT as a guidance method for the rare occasions in which sampling of these findings is required. In general, multiphasic CT and/or MRI studies tend to suffice for diagnostic purposes of such lesions. However, we highlight that liver abscesses may be a source of false-positive findings on [18F]FDG PET/CT.42
Lymphoma
In patients with suspected primary or relapsed lymphoma, image-guided core needle biopsy, including CT, US, and fluoroscopy-guided procedures, has been suggested as a valuable alternative to surgical intervention.43, 44 These procedures are less invasive, reduce the risk related to anesthesia, lower morbidity and mortality, cost less, and involve a negligible amount of risk for local recurrence along the core needle biopsy trajectory.45
The reported diagnostic accuracy for the diagnosis of lymphoma through CT- and US-guided core needle biopsies is about 80% for nodal locations. However, it is considerably lower for extranodal sites44, 46, 47 because most extranodal lymphoma sites (or in the case of lymphomatous involvement of normal-sized lymph nodes, bone marrow, or spleen) morphological changes are difficult to detect by CT alone. For instance, some of these lesions may have attenuation coefficients similar to that of normal muscle and may only be detectable because of their irregular contour.48, 49 After injection of iodine contrast, enhancement of soft tissue lymphomas is mostly bland,50 and on MRI, lymphomas merely present as large masses with signal intensities comparable to, or slightly lower than, the adjacent muscles on T1-weighted images.51
The functional evaluation given by [18F]FDG PET/CT can thus increase the sensitivity for identifying areas of active disease (FDG avidity for most types of lymphomas is 90% or greater).13, 37, 39, 52 The intensity of FDG uptake in a lymphoma may vary with histology, grade, tumor cell proliferation, local perfusion, and hypoxia. This is reflected in variations of the standard uptake value (SUV), which is typically lower in indolent lymphomas than in aggressive types.53
A particular advantage of FDG PET/CT imaging guidance is the ability to differentiate and selectively sample the most metabolically active component of a lesion, obtaining the most representative sample thus identifying and sampling foci of “transformation” from indolent to aggressive non-Hodgkin lymphoma (Figure 3). Such an identification can be beneficial for heterogeneous targets, e.g. soft tissue lymphoma, reducing the false-negative biopsy rate54 and improving disease stratification.55 A single asymmetric focus, or multiple foci, of abnormally increased FDG accumulation in a specific muscle should raise suspicion for lymphomatous involvement, even though the muscles can accumulate variable degrees of FDG in a physiological state.56 Reliable differentiation of soft tissue lymphoma from other tumors, such as rhabdomyosarcoma or Ewing sarcoma, and from metastatic carcinoma, is necessary for therapeutic purposes.57
Figure 3.
Increased [18F]FDG uptake at multiple locations in subcutaneous adipose tissue in an 83-year-old woman with indolent lymphoma in transformation. A: positron emission tomography/computed tomography (PET/CT) transaxial; B: CT transaxial; C: PET transaxial; D: PET maximum intensity projection; red arrow: site of biopsy.
Due to better visualization, PET/CT guided biopsy may reduce procedure times for lesions with equivocal imaging findings in other modalities and subsequently reduce radiation exposure to the patient and physicians.9 Potential imaging pitfalls include areas of non-malignant lesions such as areas of inflammation, brown fat FDG uptake, bone marrow and spleen activation, and recent therapeutic intervention, e.g., surgery, radiation therapy, or chemotherapy.58 Such a pattern of FDG distribution can mimic or obscure malignant lymph-adenopathies at PET: if there is no correlation with CT findings, or if the result is difficult to interpret, there is a risk of false-positive results.59
Bone
Image-guided biopsies of bone lesions are a minimally invasive technique that allow tissue samples for histological processing. For example, percutaneous CT-guided biopsies enable successful sampling in 70–90% of cases with morphologically evident lesions,60, 61 with a reduction in success rates for lesions without significant morphological alterations resulting in non-representative tissue sampling and false-negative biopsies. In this case, a multimodal image-guided intervention, using the metabolic and topographic information of PET/CT,54 is essential to directly select the hypermetabolic region (Figure 4, 5) of active isolated/multifocal bone lesions, irrespective of structural landmarks, considering that PET/CT can visualize malignancies even before morphological changes become apparent.61 Such an intervention allows to obtain representative tissue samples and to reduce futile biopsy attempts, with complication rates comparable to traditional CT-guided biopsies.13, 62
Figure 4.
An [18F]FDG PET/CT guided biopsy on the right side of the sacrum (standard uptake value = 15.2) after two lines of chemotherapy, with diagnosis of relapse, in a 71-year-old woman with progression of skeletal aggressive B lymphoma. A: PET/CT transaxial; B: CT transaxial; C: PET maximum intensity projection; red arrow: site of biopsy.
Figure 5.
A 49-year-old man diagnosed with Hodgkin lymphoma with increased [18F]FDG uptake in the bone, lymph nodes (above and below the diaphragm), and the pleura. PET/CT guided biopsy was performed on the left side of the sacrum (standard uptake value = 19). A: PET/CT transaxial; B: PET/CT sagittal; C: PET maximum intensity projection; red arrow: site of biopsy.
PET/CT-guided biopsy is a feasible, safe, and efficient approach for the diagnosis of bone tumors and tumor-like lesions, targeting the metabolically active area, or portion of a bone lesion, allowing to identify lesions in the bone marrow (invisible on CT) and to target the viable portion of the tumor.63 The procedure can be beneficial for heterogeneous targets, may reduce the false-negative biopsy rate,54 and improve disease stratification.55, 64–69 Due to better visualization, targeted biopsy with PET/CT imaging might reduce the time of procedures for lesions with equivocal imaging findings from other modalities, and subsequently reduce radiation exposure to the patient and physicians.9
As reported in the study of Guo et al.,70 PET/CT-guided bone biopsy has a high overall diagnostic performance of 96.1% and 100% for first and repeat biopsy, respectively. Its main clinical impact concerns patients with singular suspected bone metastases,71 patients with minimal residual FDG uptake at the end-of-treatment,72 patients with hypermetabolic lesions whose morphology does not correlate with the suspicion of metastases and when individualized treatment should be planned.71
Operators need to be aware of several reasons for false positive or non-specific FDG uptake in bones, for instance due to several benign conditions (e.g. Paget’s disease,73 sarcoidosis,74 giant cell tumors, eosinophilic granulomas, Langerhans cell histiocytosis) or sites of inflammation or infection (e.g. Brodie abscess).
Incorporation of intra-procedural PET/CT into ablation guidance
Image-guided techniques have been proposed for the treatment of primary and secondary tumors and have become increasingly common in clinical practice.2, 16, 75 Classification systems, such as the Barcelona Clinic Liver Classification for liver lesions, help to determine the use of procedures as well as to distinguish the selection of one procedure over another. Although the selection is not always made according to the directives, it often depends on the availability and experience of interventional oncologists as well as on access to different technologies. Radiopharmaceuticals, such as [18F]FDG or [15O]water, may be useful to measure tumor viability during a procedure. Preliminary results showed promising results for those lesions that were hypermetabolic at baseline examination.
Multimodality image fusion may aid interventional radiologists substantially, but patient breathing, patient positioning, organ shift, and even procedure-/instrument-related motion challenge image registration and fusion. Therefore, ideally both imaging data sets are obtained in the same imaging session.3–7 The FDG PET/CT “split dose technique”15 provides an adequate visualization of the target as well as real-time metabolic assessment of the AZ, including the margins, intra-operatively and before any postablation inflammatory changes or reactive hyperemia are in effect (with ambiguous SUV signal). Recent data has indicated that when using the split dose technique and the postablation biopsy was negative for residual tumor, the uptake from the second, postablation [18F]FDG injection was an independent predictor of local tumor progression after ablation of colorectal liver metastases.16
PET/CT-guided interventions using specific radiotracers also seem promising for detecting radiotracer avid lesions that are not well visualized with other imaging methods. For instance, the prostate-specific membrane antigen (PSMA) is expressed by virtually all prostate cancers. PSMA-based radiotracers have been used successfully to obtain targeted biopsies and to treat tumors on PET-positive sites, with high sensitivity and specificity.14, 76
Table I28, 77–80 provides a summary review of the advantages and limitations of using PET/CT guidance in interventional procedures.
Table I. —
Advantages and limitations of using positron emission tomography/computed tomography (PET/CT) guidance in interventional procedures.
| PET/CT guided | Advantages | Limitations |
| Biopsies | -Guiding biopsies towards the areas of higher FDG uptake thus improving the possibility of yielding representative material (especially in heterogeneous lesions). -Allowing sampling of lesions characterized exclusively by metabolic alterations, before the appearance of correspondent anatomical changes. -Evaluation of new areas presenting metabolic alterations in patients already submitted to initial treatment. -Development of new radiotracers, broadening the spectrum of possibilities for PET/CT guided biopsies in the near future. -Similar complication rates to biopsies guided by CT only. -Measuring the specimen activity provides additional information, which can potentially be used for checking biopsy adequacy and for correlating tracer uptake with pathology with high resolution.177–80 |
-[18F]FDG PET/CT may not detect disease sites due to small lesion size and subsequent partial volume effect, low or heterogeneous glucose metabolism. -Benign inflammatory conditions often present with FDG avidity, which may lead to biopsies with negative results for malignancy. -Especially in developing countries, PET/CT availability might be significantly lower in comparison with conventional CT. |
| Ablations | - Prolonged visualization of the lesion. - Verification of ablation and prediction of local recurrence. |
-Slightly longer procedure time and slightly higher patient doses.28 -PET readers need to be aware of therapy-induced inflammation with its characteristic pattern of [18F]FDG uptake. |
Methodological developments and prospects
In the preface of a recent special issue dedicated to the evolving role of PET in interventional oncology, Lam et al. explain that the recent advances in targeted therapies impose higher demands to functional imaging and lead to the expanding role of PET and other potentially quantifiable functional imaging approaches in locoregional therapies.34 This transition is essentially accompanied by expanding the spectrum of PET tracers used in the interventional setting and with multiple technological and procedural innovations.
Some cancers show only low FDG uptake or may not be detectable on PET due to their small lesion size and/or heterogeneous glucose metabolism. Therefore, various other radiotracers, with unique biological specificities, are being investigated, including small molecules, peptides, and antibodies. For instance, 89Zr trastuzumab and pertuzumab have been used for imaging of HER-2 positive breast and gastric cancers,81 and PSMA-specific radiotracers for imaging of prostate cancer. Of note, 89Zr has a physical half-life of 78.4 hours, permitting PET-guided biopsies even at later time points in the case of labelled antibodies. A workflow that allows diagnostic PET/CT followed by a guided PET/CT biopsy several days later and after a multidisciplinary review meeting has been established.14
With respect to real time PET/CT guidance of ablations, after the introduction of the [18F]FDG “split dose technique”15 the procedure was extended to non-FDG tracers. For example, [18F]FDOPA was used for PET/CT-guided radiofrequency ablation of liver metastases from neuroendocrine tumors.21 Gallium-68 labeled peptidic vector ([68Ga]DOTATOC) has been used to guide biopsy and cryoablation of tumor-induced osteomalacia.77 In another approach, [13N]ammonia has been investigated for postablation injection, using perfusion PET for the ablation margin assessment:82 the ablation margin is seen as photopenic compared with the higher normal liver activity and the higher activity trapped in the tumor from the preprocedural [18F]FDG injection.82
In addition to reduction in injected activities and use of non-FDG PET radiotracers, another recent development is the use of robotic arms for procedure assistance.83–87 As described by Kumar et al.85 this robotic arm system uses physician-defined target and entry points based on the intraprocedural PET/CT to define the access path. Once the access path is determined, the robot places a guidance tube along this access path through which the interventional radiologist inserts the biopsy needle. Then, the needle is advanced to the depth determined by the robot computer, and the position of the needle is confirmed by low-dose CT scan, which is fused with the PET image.85
In the research domain, the fact that biopsy samples contain radioactivity at the location of the radiotracer molecules, opens the possibility for extracting additional information from such biospecimens by measuring their activity. Such measurements can be performed by using a well counter or by performing autoradiography.78, 79 Autoradiography allows quantification of the tracer uptake along the biopsy specimen (Figure 6),78 which can be beneficial for a wide variety of applications.78, 88 The first potential application of the method of quantitative autoradiography of biopsy specimens (QABS) is measuring the metabolic activity in specimens, which may help identify those that are inadequate. This may help reduce toxicity by lowering the number of needle placements and also decrease the possibility of misdiagnosis.22, 78
Figure 6.
An illustration of the capabilities of the Quantitative Autoradiography (ARG) of Biopsy Specimens (QABS) method for quantification of [18F]FDG in terms of standard uptake value derived from ARG (SUVARG) along the biopsy specimen.78 The location of the specimen extraction site is shown on the left in the PET/CT image obtained during a liver biopsy with intra-procedural PET/CT guidance. The correlation of the ARG with the photograph of the specimen (marked with pathology dyes) and with the histopathology micrograph is shown on the right. SUVARG and SUVPET do not need to match due to the different resolutions of the two measurements.
In addition to diagnostic adequacy, the same measurement can potentially be used to evaluate the adequacy of the specimens for genomic profiling by providing an estimate of the number of viable tumor cells in them.79 Other potential applications of the high-resolution information about radiotracer uptake in biopsy specimens are: 1) in situ validation of new radiopharmaceuticals;77 2) indication of regions in the specimens with high tumor burden for histopathology analysis; and 3) in localized cancer therapies, for tumor border and heterogeneity characterization. An alternative potential application of the QABS method is in the field of radiomics because correlating tracer uptake with molecular pathology with a higher resolution than that of PET images provides additional information for understanding and guiding radiomic investigations. Even if the biospecimen activity is not measured, having molecular pathology results from the same lesion and date as the PET image, reduces chances for lesion mutation mislabeling which may be caused by non-concordance between the primary and metastatic lesions and which may have therapeutic implications.89
Conclusions
The use of PET/CT in image-guided interventions has been expanding and improving steadily. When used for image-guided biopsy, real-time PET guidance allows accurate depiction and targeting of otherwise invisible avid lesions and refines the regions to be sampled. In addition, real-time PET/CT guided biopsies can yield radioactive specimens that may provide useful information for evaluating the adequacy of the specimens and for high-resolution in-situ validation of new PET tracers. For ablations, besides improving targeting, PET/CT provides an intraprocedural method for assessing the ablation zone as a surrogate image biomarker of complete ablation or detecting residual avid disease.16 In addition, the combination of metabolic imaging with the portal venous phase of contrast enhanced CT obtained just after the ablation significantly improves the depiction of the ablation margins. The combination of metabolic and anatomic assessment of the ablation zone can be used to optimize outcomes of image-guided ablation of avid tumors and may allow for immediate retreatment of those with residual metabolically active disease.16, 90 An open trial that is currently enrolling patients uses metabolic imaging and biopsy in an effort to optimize thermal ablation for colorectal hepatic metastases.16, 90
Acknowledgements.—
The authors acknowledge the diligent support from the staff of our departments and institutions.
Funding.—
This work was funded in part by the Marie-Josée and Henry R. Kravis Center for Molecular Oncology at MSK and by the National Cancer Institute Cancer Center Core Grant N. P30-CA008748.
Footnotes
Conflicts of interest.—Stephen B. Solomon received funds from GE Healthcare, Johnson and Johnson, AngioDynamics, he has a consultant relationship with Johnson and Johnson, Endoways, XACT Robotics, he also has personal financial interests for Johnson and Johnson.
References
- 1.Hopper KD. Percutaneous, radiographically guided biopsy: a history. Radiology 1995;196:329–33. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=7617841&dopt=Abstract 10.1148/radiology.196.2.7617841 [DOI] [PubMed] [Google Scholar]
- 2.Solomon SB, Cornelis F. Interventional Molecular Imaging. J Nucl Med 2016;57:493–6. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=26912443&dopt=Abstract 10.2967/jnumed.115.161190 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Yokoyama K, Ikeda O, Kawanaka K, Nakasone Y, Tamura Y, Inoue S, et al. Comparison of CT-guided percutaneous biopsy with and without registration of prior PET/CT images to diagnose mediastinal tumors. Cardiovasc Intervent Radiol 2014;37:1306–11. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=24263776&dopt=Abstract 10.1007/s00270-013-0793-x [DOI] [PubMed] [Google Scholar]
- 4.Venkatesan AM, Kadoury S, Abi-Jaoudeh N, Levy EB, Maass-Moreno R, Krücker J, et al. Real-time FDG PET guidance during biopsies and radiofrequency ablation using multimodality fusion with electromagnetic navigation. Radiology 2011;260:848–56. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=21734159&dopt=Abstract 10.1148/radiol.11101985 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bajpai S, Kambadakone A, Guimaraes AR, Arellano RS, Gervais DA, Sahani D. Image-guided treatment in the hepatobiliary system: role of imaging in treatment planning and posttreatment evaluation. Radiographics 2015;35:1393–418. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=26252365&dopt=Abstract 10.1148/rg.2015140281 [DOI] [PubMed] [Google Scholar]
- 6.Tatli S, Gerbaudo VH, Mamede M, Tuncali K, Shyn PB, Silverman SG. Abdominal masses sampled at PET/CT-guided percutaneous biopsy: initial experience with registration of prior PET/CT images. Radiology 2010;256:305–11. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=20574103&dopt=Abstract 10.1148/radiol.10090931 [DOI] [PubMed] [Google Scholar]
- 7.Shyn PB, Tatli S, Sahni VA, Sadow CA, Forgione K, Mauri G, et al. PET/CT-guided percutaneous liver mass biopsies and ablations: targeting accuracy of a single 20 s breath-hold PET acquisition. Clin Radiol 2014;69:410–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=24411824&dopt=Abstract 10.1016/j.crad.2013.11.013 [DOI] [PubMed] [Google Scholar]
- 8.Shyn PB. Interventional positron emission tomography/computed tomography: state-of-the-art. Tech Vasc Interv Radiol 2013;16:182–90. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=23993081&dopt=Abstract 10.1053/j.tvir.2013.02.014 [DOI] [PubMed] [Google Scholar]
- 9.Ryan ER, Thornton R, Sofocleous CT, Erinjeri JP, Hsu M, Quinn B, et al. PET/CT-guided interventions: personnel radiation dose. Cardiovasc Intervent Radiol 2013;36:1063–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=23229891&dopt=Abstract 10.1007/s00270-012-0515-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Cazzato RL, Garnon J, Shaygi B, Koch G, Tsoumakidou G, Caudrelier J, et al. PET/CT-guided interventions: Indications, advantages, disadvantages and the state of the art. Minim Invasive Ther Allied Technol 2018;27:27–32. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29168421&dopt=Abstract 10.1080/13645706.2017.1399280 [DOI] [PubMed] [Google Scholar]
- 11.Deng CJ, Dai FQ, Qian K, Tan QY, Wang RW, Deng B, et al. Clinical updates of approaches for biopsy of pulmonary lesions based on systematic review. BMC Pulm Med 2018;18:146. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=30176840&dopt=Abstract 10.1186/s12890-018-0713-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Young M, Shapiro R. Percutaneous Lung Lesion Biopsy (PLLB). Treasure Island, FL: StatPearls; 2019. [PubMed] [Google Scholar]
- 13.Cornelis F, Silk M, Schoder H, Takaki H, Durack JC, Erinjeri JP, et al. Performance of intra-procedural 18-fluorodeoxyglucose PET/CT-guided biopsies for lesions suspected of malignancy but poorly visualized with other modalities. Eur J Nucl Med Mol Imaging 2014;41:2265–72. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=25106463&dopt=Abstract 10.1007/s00259-014-2852-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Cornelis FH, Durack JC, Pandit-Taskar N, Ulaner GA, Lewis JS, Morris MJ, et al. Long-Half-Life 89Zr-Labeled Radiotracers Can Guide Percutaneous Biopsy Within the PET/CT Suite Without Reinjection of Radiotracer. J Nucl Med 2018;59:399–402. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28818992&dopt=Abstract 10.2967/jnumed.117.194480 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Ryan ER, Sofocleous CT, Schöder H, Carrasquillo JA, Nehmeh S, Larson SM, et al. Split-dose technique for FDG PET/CT-guided percutaneous ablation: a method to facilitate lesion targeting and to provide immediate assessment of treatment effectiveness. Radiology 2013;268:288–95. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=23564714&dopt=Abstract 10.1148/radiol.13121462 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Cornelis FH, Petre EN, Vakiani E, Klimstra D, Durack JC, Gonen M, et al. Immediate Postablation 18F-FDG Injection and Corresponding SUV Are Surrogate Biomarkers of Local Tumor Progression After Thermal Ablation of Colorectal Carcinoma Liver Metastases. J Nucl Med 2018;59:1360–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29439012&dopt=Abstract 10.2967/jnumed.117.194506 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Cornelis F, Sotirchos V, Violari E, Sofocleous CT, Schoder H, Durack JC, et al. 18F-FDG PET/CT is an immediate imaging biomarker of treatment success after liver metastasis ablation. J Nucl Med 2016;57:1052–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=26912433&dopt=Abstract 10.2967/jnumed.115.171926 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Vandenberghe S, Mikhaylova E, D’Hoe E, Mollet P, Karp JS. Recent developments in time-of-flight PET. EJNMMI Phys 2016;3:3. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=26879863&dopt=Abstract 10.1186/s40658-016-0138-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Cherry SR, Jones T, Karp JS, Qi J, Moses WW, Badawi RD. Total-body PET: maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med 2018;59:3–12. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28935835&dopt=Abstract 10.2967/jnumed.116.184028 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Pan T, Einstein SA, Kappadath SC, Grogg KS, Lois Gomez C, Alessio AM, et al. Performance evaluation of the 5-Ring GE Discovery MI PET/CT system using the national electrical manufacturers association NU 2–2012 Standard. Med Phys 2019;46:3025–33. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=31069816&dopt=Abstract 10.1002/mp.13576 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Cazzato RL, Garnon J, Ramamurthy N, Tsoumakidou G, Imperiale A, Namer IJ, et al. 18F-FDOPA PET/CT-Guided Radiofrequency Ablation of Liver Metastases from Neuroendocrine Tumours: Technical Note on a Preliminary Experience. Cardiovasc Intervent Radiol 2016;39:1315–21. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27048487&dopt=Abstract 10.1007/s00270-016-1334-1 [DOI] [PubMed] [Google Scholar]
- 22.Fanchon LM, Apte A, Schmidtlein CR, Yorke E, Hu YC, Dogan S, et al. Evaluation of the tumor registration error in biopsy procedures performed under real-time PET/CT guidance. Med Phys 2017;44:5089–95. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28494089&dopt=Abstract 10.1002/mp.12334 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Shyn PB, Tremblay-Paquet S, Palmer K, Tatli S, Tuncali K, Olubiyi OI, et al. Breath-hold PET/CT-guided tumour ablation under general anaesthesia: accuracy of tumour image registration and projected ablation zone overlap. Clin Radiol 2017;72:223–9. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27889091&dopt=Abstract 10.1016/j.crad.2016.10.017 [DOI] [PubMed] [Google Scholar]
- 24.Andersen PA, Chakera AH, Klausen TL, Binderup T, Grossjohann HS, Friis E, et al. Radiation exposure to surgical staff during F-18-FDG-guided cancer surgery. Eur J Nucl Med Mol Imaging 2008;35:624–9. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17955240&dopt=Abstract 10.1007/s00259-007-0532-0 [DOI] [PubMed] [Google Scholar]
- 25.Povoski SP, Sarikaya I, White WC, Marsh SG, Hall NC, Hinkle GH, et al. Comprehensive evaluation of occupational radiation exposure to intraoperative and perioperative personnel from 18F-FDG radioguided surgical procedures. Eur J Nucl Med Mol Imaging 2008;35:2026–34. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=18618106&dopt=Abstract 10.1007/s00259-008-0880-4 [DOI] [PubMed] [Google Scholar]
- 26.Paulson EK, Sheafor DH, Enterline DS, McAdams HP, Yoshizumi TT. CT fluoroscopy—guided interventional procedures: techniques and radiation dose to radiologists. Radiology 2001;220:161–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=11425990&dopt=Abstract 10.1148/radiology.220.1.r01jl29161 [DOI] [PubMed] [Google Scholar]
- 27.Ahmed S, Zimmer A, McDonald N, Spies S. The effectiveness of lead aprons in reducing radiation exposures from specific radionuclides. J Nucl Med 2007;48:470. [Google Scholar]
- 28.Hu EY, Levesque VM, Bay CP, Seol JG, Shyn PB. Liver Tumor Ablation Procedure Duration and Estimated Patient Radiation Dose: Comparing Positron Emission Tomography/CT and CT Guidance. J Vasc Interv Radiol 2020;31:1052–9. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=32534979&dopt=Abstract 10.1016/j.jvir.2019.11.036 [DOI] [PubMed] [Google Scholar]
- 29.Sainani NI, Shyn PB, Tatli S, Morrison PR, Tuncali K, Silverman SG. PET/CT-guided radiofrequency and cryoablation: is tumor fluorine-18 fluorodeoxyglucose activity dissipated by thermal ablation? J Vasc Interv Radiol 2011;22:354–60. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=21353987&dopt=Abstract 10.1016/j.jvir.2010.11.027 [DOI] [PubMed] [Google Scholar]
- 30.Schoellnast H, Larson SM, Nehmeh SA, Carrasquillo JA, Thornton RH, Solomon SB. Radiofrequency ablation of non-small-cell carcinoma of the lung under real-time FDG PET CT guidance. Cardiovasc Intervent Radiol 2011;34:S182–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=20508937&dopt=Abstract 10.1007/s00270-010-9898-7 [DOI] [PubMed] [Google Scholar]
- 31.Kaye EA, Cornelis FH, Petre EN, Tyagi N, Shady W, Shi W, et al. Volumetric 3D assessment of ablation zones after thermal ablation of colorectal liver metastases to improve prediction of local tumor progression. Eur Radiol 2019;29:2698–705. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=30402706&dopt=Abstract 10.1007/s00330-018-5809-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Fei B, Schuster DM. PET Molecular Imaging-Directed Biopsy: A Review. Am J Roentgenol 2017;209:255–69. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28504563&dopt=Abstract 10.2214/AJR.17.18047 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Cerci J, Fanti S, Delbeke D. Oncological PET/CT with histological confirmation. Amsterdam: Springer; 2016. [Google Scholar]
- 34.Lam MG, Hunt SJ, El-Haddad GE, Alavi A. Evolving Role of PET in Interventional Radiology-Based Oncology Procedures. PET Clin 2019;14:xiii–xiv. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=31472747&dopt=Abstract 10.1016/j.cpet.2019.07.001 [DOI] [PubMed] [Google Scholar]
- 35.İntepe YS, Metin B, Şahin S, Kaya B, Okur A. Our transthoracic biopsy practices accompanied by the imaging process: the contribution of positron emission tomography usage to accurate diagnosis. Acta Clin Belg 2016;71:214–20. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27142092&dopt=Abstract 10.1080/17843286.2016.1155810 [DOI] [PubMed] [Google Scholar]
- 36.Guralnik L, Rozenberg R, Frenkel A, Israel O, Keidar Z. Metabolic PET/CT-guided lung lesion biopsies: impact on diagnostic accuracy and rate of sampling error. J Nucl Med 2015;56:518–22. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=25698780&dopt=Abstract 10.2967/jnumed.113.131466 [DOI] [PubMed] [Google Scholar]
- 37.Purandare NC, Kulkarni AV, Kulkarni SS, Roy D, Agrawal A, Shah S, et al. 18F-FDG PET/CT-directed biopsy: does it offer incremental benefit? Nucl Med Commun 2013;34:203–10. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=23353885&dopt=Abstract 10.1097/MNM.0b013e32835c5a57 [DOI] [PubMed] [Google Scholar]
- 38.El-Haddad G. PET-based percutaneous needle biopsy. PET Clin 2016;11:333–49. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27321036&dopt=Abstract 10.1016/j.cpet.2016.02.009 [DOI] [PubMed] [Google Scholar]
- 39.Cerci JJ, Tabacchi E, Bogoni M, Delbeke D, Pereira CC, Cerci RJ, et al. Comparison of CT and PET/CT for biopsy guidance in oncological patients. Eur J Nucl Med Mol Imaging 2017;44:1269–74. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28364161&dopt=Abstract 10.1007/s00259-017-3658-8 [DOI] [PubMed] [Google Scholar]
- 40.Cerci JJ, Huber FZ, Bogoni M. PET/CT-guided biopsy of liver lesions. Clin Transl Imaging 2014;2:157–63. 10.1007/s40336-014-0056-0 [DOI] [Google Scholar]
- 41.Ewertsen C, Henriksen BM, Torp-Pedersen S, Bachmann Nielsen M. Characterization by biopsy or CEUS of liver lesions guided by image fusion between ultrasonography and CT, PET/CT or MRI. Ultraschall Med 2011;32:191–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=21225564&dopt=Abstract 10.1055/s-0029-1245921 [DOI] [PubMed] [Google Scholar]
- 42.Sacks A, Peller PJ, Surasi DS, Chatburn L, Mercier G, Subramaniam RM. Value of PET/CT in the management of primary hepatobiliary tumors, part 2. Am J Roentgenol 2011;197:W260–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=21785051&dopt=Abstract 10.2214/AJR.11.6995 [DOI] [PubMed] [Google Scholar]
- 43.Balestreri L, Morassut S, Bernardi D, Tavio M, Talamini R, Gloghini A, et al. Efficacy of CT-guided percutaneous needle biopsy in the diagnosis of malignant lymphoma at first presentation. Clin Imaging 2005;29:123–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=15752968&dopt=Abstract 10.1016/j.clinimag.2004.10.002 [DOI] [PubMed] [Google Scholar]
- 44.Zinzani PL, Corneli G, Cancellieri A, Magagnoli M, Lacava N, Gherlinzoni F, et al. Core needle biopsy is effective in the initial diagnosis of mediastinal lymphoma. Haematologica 1999;84:600–3. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=10406900&dopt=Abstract [PubMed] [Google Scholar]
- 45.Sklair-Levy M, Polliack A, Shaham D, Applbaum YH, Gillis S, Ben-Yehuda D, et al. CT-guided core-needle biopsy in the diagnosis of mediastinal lymphoma. Eur Radiol 2000;10:714–8. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=10823620&dopt=Abstract 10.1007/s003300050991 [DOI] [PubMed] [Google Scholar]
- 46.Agid R, Sklair-Levy M, Bloom AI, Lieberman S, Polliack A, Ben-Yehuda D, et al. CT-guided biopsy with cutting-edge needle for the diagnosis of malignant lymphoma: experience of 267 biopsies. Clin Radiol 2003;58:143–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=12623044&dopt=Abstract 10.1053/crad.2002.1061 [DOI] [PubMed] [Google Scholar]
- 47.Gaudio F, Pedote P, Ferrante A, Perrone T, Ingravallo G, Ianora AA, et al. CT-guided needle biopsy performed with modified coaxial technique in patients with refractory or recurrent lymphomas. Ann Hematol 2014;93:1559–64. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=24760400&dopt=Abstract 10.1007/s00277-014-2078-0 [DOI] [PubMed] [Google Scholar]
- 48.Malloy PC, Fishman EK, Magid D. Lymphoma of bone, muscle, and skin: CT findings. Am J Roentgenol 1992;159:805–9. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=1529847&dopt=Abstract 10.2214/ajr.159.4.1529847 [DOI] [PubMed] [Google Scholar]
- 49.Panicek DM, Lautin JL, Schwartz LH, Castellino RA. Non-Hodgkin lymphoma in skeletal muscle manifesting as homogeneous masses with CT attenuation similar to muscle. Skeletal Radiol 1997;26:633–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=9428069&dopt=Abstract 10.1007/s002560050301 [DOI] [PubMed] [Google Scholar]
- 50.Grunshaw ND, Chalmers AG. Skeletal muscle lymphoma. Clin Radiol 1992;45:399–400. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=1606798&dopt=Abstract 10.1016/S0009-9260(05)80999-7 [DOI] [PubMed] [Google Scholar]
- 51.Metzler JP, Fleckenstein JL, Vuitch F, Frenkel EP. Skeletal muscle lymphoma: MRI evaluation. Magn Reson Imaging 1992;10:491–4. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=1406100&dopt=Abstract 10.1016/0730-725X(92)90522-2 [DOI] [PubMed] [Google Scholar]
- 52.Bitencourt AG, Tyng CJ, Pinto PN, Almeida MF, Meyrellis LC, Pinheiro RP, et al. Percutaneous biopsy based on PET/CT findings in cancer patients: technique, indications, and results. Clin Nucl Med 2012;37:e95–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=22475924&dopt=Abstract 10.1097/RLU.0b013e3182443b78 [DOI] [PubMed] [Google Scholar]
- 53.Schöder H, Noy A, Gönen M, Weng L, Green D, Erdi YE, et al. Intensity of 18fluorodeoxyglucose uptake in positron emission tomography distinguishes between indolent and aggressive non-Hodgkin’s lymphoma. J Clin Oncol 2005;23:4643–51. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=15837966&dopt=Abstract 10.1200/JCO.2005.12.072 [DOI] [PubMed] [Google Scholar]
- 54.Klaeser B, Mueller MD, Schmid RA, Guevara C, Krause T, Wiskirchen J. PET-CT-guided interventions in the management of FDG-positive lesions in patients suffering from solid malignancies: initial experiences. Eur Radiol 2009;19:1780–5. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=19238391&dopt=Abstract 10.1007/s00330-009-1338-1 [DOI] [PubMed] [Google Scholar]
- 55.Völker T, Denecke T, Steffen I, Misch D, Schönberger S, Plotkin M, et al. Positron emission tomography for staging of pediatric sarcoma patients: results of a prospective multicenter trial. J Clin Oncol 2007;25:5435–41. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=18048826&dopt=Abstract 10.1200/JCO.2007.12.2473 [DOI] [PubMed] [Google Scholar]
- 56.Samuel LM, White J, Lessells AM, Roddie H, Matheson LM. Primary non-Hodgkins lymphoma of muscle. Clin Oncol 1999;11:49–51. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=10194587&dopt=Abstract 10.1053/clon.1999.9009 [DOI] [PubMed] [Google Scholar]
- 57.Schwalke MA, Rodil JV, Vezeridis MP. Primary lymphoma arising in skeletal muscle. Eur J Surg Oncol 1990;16:70–3. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=2137791&dopt=Abstract [PubMed] [Google Scholar]
- 58.Rousseau C, Bourbouloux E, Campion L, Fleury N, Bridji B, Chatal JF, et al. Brown fat in breast cancer patients: analysis of serial (18)F-FDG PET/CT scans. Eur J Nucl Med Mol Imaging 2006;33:785–91. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=16596378&dopt=Abstract 10.1007/s00259-006-0066-x [DOI] [PubMed] [Google Scholar]
- 59.Christensen CR, Clark PB, Morton KA. Reversal of hypermetabolic brown adipose tissue in F-18 FDG PET imaging. Clin Nucl Med 2006;31:193–6. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=16550009&dopt=Abstract 10.1097/01.rlu.0000204199.33136.05 [DOI] [PubMed] [Google Scholar]
- 60.Huch K, Röderer G, Ulmar B, Reichel H. CT-guided interventions in orthopedics. Arch Orthop Trauma Surg 2007;127:677–83. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17653733&dopt=Abstract 10.1007/s00402-007-0410-z [DOI] [PubMed] [Google Scholar]
- 61.Lis E, Bilsky MH, Pisinski L, Boland P, Healey JH, O’malley B, et al. Percutaneous CT-guided biopsy of osseous lesion of the spine in patients with known or suspected malignancy. Am J Neuroradiol 2004;25:1583–8. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=15502142&dopt=Abstract [PMC free article] [PubMed] [Google Scholar]
- 62.Kostakoglu L, Hardoff R, Mirtcheva R, Goldsmith SJ. PET-CT fusion imaging in differentiating physiologic from pathologic FDG uptake. Radiographics 2004;24:1411–31. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=15371617&dopt=Abstract 10.1148/rg.245035725 [DOI] [PubMed] [Google Scholar]
- 63.Kobayashi K, Bhargava P, Raja S, Nasseri F, Al-Balas HA, Smith DD, et al. Image-guided biopsy: what the interventional radiologist needs to know about PET/CT. Radiographics 2012;32:1483–501. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=22977031&dopt=Abstract 10.1148/rg.325115159 [DOI] [PubMed] [Google Scholar]
- 64.Klaeser B, Wiederkehr O, Koeberle D, Mueller A, Bubeck B, Thuerlimann B. Therapeutic impact of 2-[fluorine-18]fluoro-2-deoxy-D-glucose positron emission tomography in the pre- and postoperative staging of patients with clinically intermediate or high-risk breast cancer. Ann Oncol 2007;18:1329–34. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17693646&dopt=Abstract 10.1093/annonc/mdm139 [DOI] [PubMed] [Google Scholar]
- 65.Herbertson RA, Lee ST, Tebbutt N, Scott AM. The expanding role of PET technology in the management of patients with colorectal cancer. Ann Oncol 2007;18:1774–81. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17434894&dopt=Abstract 10.1093/annonc/mdm108 [DOI] [PubMed] [Google Scholar]
- 66.Piperkova E, Raphael B, Altinyay ME, Castellon I, Libes R, Sandella N, et al. Impact of PET/CT in comparison with same day contrast enhanced CT in breast cancer management. Clin Nucl Med 2007;32:429–34. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17515747&dopt=Abstract 10.1097/RLU.0b013e31805375e0 [DOI] [PubMed] [Google Scholar]
- 67.Loft A, Berthelsen AK, Roed H, Ottosen C, Lundvall L, Knudsen J, et al. The diagnostic value of PET/CT scanning in patients with cervical cancer: a prospective study. Gynecol Oncol 2007;106:29–34. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=17482666&dopt=Abstract 10.1016/j.ygyno.2007.03.027 [DOI] [PubMed] [Google Scholar]
- 68.Kitajima K, Murakami K, Yamasaki E, Kaji Y, Fukasawa I, Inaba N, et al. Diagnostic accuracy of integrated FDG-PET/contrast-enhanced CT in staging ovarian cancer: comparison with enhanced CT. Eur J Nucl Med Mol Imaging 2008;35:1912–20. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=18682935&dopt=Abstract 10.1007/s00259-008-0890-2 [DOI] [PubMed] [Google Scholar]
- 69.Lardinois D, Weder W, Hany TF, Kamel EM, Korom S, Seifert B, et al. Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography. N Engl J Med 2003;348:2500–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=12815135&dopt=Abstract 10.1056/NEJMoa022136 [DOI] [PubMed] [Google Scholar]
- 70.Guo W, Hao B, Chen HJ, Zhao L, Luo ZM, Wu H, et al. PET/CT-guided percutaneous biopsy of FDG-avid metastatic bone lesions in patients with advanced lung cancer: a safe and effective technique. Eur J Nucl Med Mol Imaging 2017;44:25–32. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27444682&dopt=Abstract 10.1007/s00259-016-3455-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Klaeser B, Wiskirchen J, Wartenberg J, Weitzel T, Schmid RA, Mueller MD, et al. PET/CT-guided biopsies of metabolically active bone lesions: applications and clinical impact. Eur J Nucl Med Mol Imaging 2010;37:2027–36. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=20680270&dopt=Abstract 10.1007/s00259-010-1524-z [DOI] [PubMed] [Google Scholar]
- 72.Kumar R, Mittal BR, Bhattacharya A, Singh H, Bal A, Prakash G, et al. 18F-FDG PET/CT-Guided Real-Time Automated Robotic Arm-Assisted Needle Navigation for Percutaneous Biopsy of Hypermetabolic Bone Lesions: Diagnostic Performance and Clinical Impact. Am J Roentgenol 2019;212:W10–8. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=30383406&dopt=Abstract 10.2214/AJR.18.19698 [DOI] [PubMed] [Google Scholar]
- 73.Mahmood S, Martinez de Llano SR. Paget disease of the humerus mimicking metastatic disease in a patient with metastatic malignant mesothelioma on whole body F-18 FDG PET/CT. Clin Nucl Med 2008;33:510–2. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=18580246&dopt=Abstract 10.1097/RLU.0b013e318177928a [DOI] [PubMed] [Google Scholar]
- 74.Ludwig V, Fordice S, Lamar R, Martin WH, Delbeke D. Unsuspected skeletal sarcoidosis mimicking metastatic disease on FDG positron emission tomography and bone scintigraphy. Clin Nucl Med 2003;28:176–9. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=12592122&dopt=Abstract 10.1097/01.RLU.0000053528.35645.70 [DOI] [PubMed] [Google Scholar]
- 75.Cornelis FH, Solomon SB. Treatment of Primary Liver Tumors and Liver Metastases, Part 2: Non-Nuclear Medicine Techniques. J Nucl Med 2018;59:1801–8. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=30361378&dopt=Abstract 10.2967/jnumed.116.186379 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Cornelis FH, Durack JC, Morris MJ, Scher HI, Solomon SB. Effective Prostate-Specific Membrane Antigen-Based 18F-DCFPyL-Guided Cryoablation of a Single Positive Site in a Patient Believed to Be More Metastatic on 11C-Choline PET/CT. Clin Nucl Med 2017;42:e516–8. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29099733&dopt=Abstract 10.1097/RLU.0000000000001846 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Maybody M, Grewal RK, Healey JH, Antonescu CR, Fanchon L, Hwang S, et al. Ga-68 DOTATOC PET/CT-Guided Biopsy and Cryoablation with Autoradiography of Biopsy Specimen for Treatment of Tumor-Induced Osteomalacia. Cardiovasc Intervent Radiol 2016;39:1352–7. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=27150801&dopt=Abstract 10.1007/s00270-016-1350-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Fanchon LM, Dogan S, Moreira AL, Carlin SA, Schmidtlein CR, Yorke E, et al. Feasibility of in situ, high-resolution correlation of tracer uptake with histopathology by quantitative autoradiography of biopsy specimens obtained under 18F-FDG PET/CT guidance. J Nucl Med 2015;56:538–44. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=25722446&dopt=Abstract 10.2967/jnumed.114.148668 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Kirov AS, Fanchon LM, Seiter D, Czmielewski C, Russell J, Dogan S, et al. Technical Note: scintillation well counters and particle counting digital autoradiography devices can be used to detect activities associated with genomic profiling adequacy of biopsy specimens obtained after a low activity 18 F-FDG injection. Med Phys 2018;45:2179–85. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29480927&dopt=Abstract 10.1002/mp.12836 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Kirov AS, Fanchon LM. Pathology-validated PET image data sets and their role in PET segmentation. Clin Transl Imaging 2014;2:253–67. 10.1007/s40336-014-0068-9 [DOI] [Google Scholar]
- 81.Yoon JK, Park BN, Ryu EK, An YS, Lee SJ. Current Perspectives on 89Zr-PET Imaging. Int J Mol Sci 2020;21:4309. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=32560337&dopt=Abstract 10.3390/ijms21124309 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Shyn PB, Casadaban LC, Sainani NI, Sadow CA, Bunch PM, Levesque VM, et al. Intraprocedural ablation margin assessment by using ammonia perfusion PET during FDG PET/CT–guided liver tumor ablation: a pilot study. Radiology 2018;288:138–45. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29613843&dopt=Abstract 10.1148/radiol.2018172108 [DOI] [PubMed] [Google Scholar]
- 83.Radhakrishnan RK, Mittal BR, Basher RK, Prakash G, Malhotra P, Kalra N, et al. Post-therapy lesions in patients with non-Hodgkin’s lymphoma characterized by 18F-FDG PET/CT-guided biopsy using automated robotic biopsy arm. Nucl Med Commun 2018;39:74–82. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=29189443&dopt=Abstract 10.1097/MNM.0000000000000780 [DOI] [PubMed] [Google Scholar]
- 84.Radhakrishnan RK, Mittal BR, Gorla AK, Basher RK, Sood A, Bal A, et al. Real-time intraprocedural 18F-FDG PET/CT-guided biopsy using automated robopsy arm (ARA) in the diagnostic evaluation of thoracic lesions with prior inconclusive biopsy results: initial experience from a tertiary health care centre. Br J Radiol 2017;90:20170258. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=28937268&dopt=Abstract 10.1259/bjr.20170258 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Kumar R, Mittal BR, Bhattacharya A, Singh H, Bal A, Vadi SK, et al. Diagnostic performance of real-time robotic arm-assisted 18F-FDG PET/CT-guided percutaneous biopsy in metabolically active abdominal and pelvic lesions. Eur J Nucl Med Mol Imaging 2019;46:838–47. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=30167803&dopt=Abstract 10.1007/s00259-018-4133-x [DOI] [PubMed] [Google Scholar]
- 86.Kumar R, Mittal BR, Bhattacharya A, Vadi SK, Singh H, Bal A, et al. Positron emission tomography/computed tomography guided percutaneous biopsies of Ga-68 avid lesions using an automated robotic arm. Diagn Interv Imaging 2020;101:157–67. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=31722844&dopt=Abstract 10.1016/j.diii.2019.10.006 [DOI] [PubMed] [Google Scholar]
- 87.Nath A, Prashanth A, Lal H, Kumar S, Barai S, Gambhir S. Robotic-assisted computed tomography-guided 18F-FDG PET/computed tomography-directed biopsy for diagnosis of intra thoracic lesions: an experience from a tertiary care centre in North India. Nucl Med Commun 2020;41:246–51. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=31939902&dopt=Abstract 10.1097/MNM.0000000000001148 [DOI] [PubMed] [Google Scholar]
- 88.Seiter D, Miller B, Fung E, Cheal S, Maitrejean S, Zanzonico P, et al. Particle counting autoradiography of radioactive biopsy specimens. Proceedings of the 18th EANM Annual Meeting; 2018. Oct 13–17; Dusseldorf, Germany. [Google Scholar]
- 89.Popovic M, Talarico O, van den Hoff J, Kunin H, Zhang Z, Lafontaine D, et al. KRAS mutation effects on the 2-[18F]FDG PET uptake of colorectal adenocarcinoma metastases in the liver. EJNMMI Res 2020;10:142. https://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&list_uids=33226505&dopt=Abstract 10.1186/s13550-020-00707-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Sofocleous CT, Vakiani E. Tissue Analysis After Thermal Ablation for Colon Cancer Liver Metastases Leading to Immediate Retreatment. ClinicalTrials.gov; 2019. [Internet]. Available from: https://clinicaltrials.gov/ct2/show/NCT04143516 [cited 2021, Jan 19]. [Google Scholar]