Table 4. Robustness of Posaconazole Observed at 235, 240, and 245 nm Wavelengths along with 25, 30, and 35 °C.
| number of formulations | peak area at 235 nm | peak area at 240 nm | peak area at 245 nm | peak area at 25 °C | peak area at 30 °C | peak area at 35 °C |
|---|---|---|---|---|---|---|
| F1 | 248,796 | 289,173 | 359,327 | 281,411 | 289,173 | 281,835 |
| F2 | 245,594 | 287,919 | 358,356 | 280,341 | 287,919 | 280,770 |
| F3 | 244,510 | 286,574 | 356,679 | 279,579 | 286,574 | 279,730 |
| F4 | 244,657 | 286,666 | 353,810 | 275,421 | 286,666 | 279,239 |
| F5 | 244,500 | 286,449 | 355,912 | 280,575 | 286,449 | 279,368 |
| F6 | 244,525 | 286,499 | 356,354 | 281,132 | 286,499 | 279,539 |
| mean | 245,430.3 | 287,213.3 | 356,739.7 | 279,743.2 | 287,213.3 | 280,080.2 |
| SDa | 1702.04 | 1108.24 | 1935.31 | 2212.14 | 1108.24 | 1018.38 |
| RSDb | 0.69 | 0.38 | 0.54 | 0.79 | 0.38 | 0.36 |
Standard deviation.
RSD: relative standard deviation, F: formulation.