Skip to main content
. 2023 Mar 2;33(3):591–610. doi: 10.1007/s00062-023-01271-5

Table 3.

Dose reduction in other NCCT of the head

Author Year Subjects (n) Comparison CT System Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol
(mGy)
DLP
(mGy * cm)
E
(mSv)
Kaul [36] 2016

177

71SD

86LD1

20LD2

SDFBP vs. LD1ASIR20 vs. LD2ASIR30 64-MDCT (GE Lightspeed VCT, GE Healthcare, Milwaukee, WI, USA)

120 kV, 100–300 mAam, SD

120 kV, 100–300 mAam, LD

FBP

HIR (ASIR 20%, 30%)

41LD1

73LD2

51.6SD

30.2LD1

13.9LD2

768.0SD

455.0LD1

204.0LD2

1.6SD

1.1LD1

0.4LD2

Ozdoba [34] 2014

75

50SD

25LD

SDFBP vs. LDHIR

16/64-MDCT (Siemens SOMATOM Sensation, Siemens Healthineers, Erlangen, Germany)SD

128-MDCT (Siemens Somatom Definition Edge, Siemens Healthineers, Erlangen, Germany)LD

120 kV, 220/380 mAsSD

100 kV, 230 mAsLD

FBP

HIR (SAFIRE)

40

57.2SD

34.9LD

964.0SD

574.1LD

NA
Korn [33] 2013

60

30SD

30LD

SDFBP vs. LDFBP, HIR 128-MDCT (Siemens Somatom Definition Flash, Siemens Healthineers, Erlangen, Germany)

120 kV, 320 mAsSD

120 kV, 255 mAsLD

FBP

HIR (SAFIRE 3)

20

59.7SD

47.8LD

1042.0SD

829.0LD

2.2SD

1.7LD

Kilic [28] 2013

305

152SD

153LD

SDFBP vs. LDASIR 16-MDCT (GE Lightspeed VCT)

100/120 kV, 200–420 mASD

100/120 kV, 100–290 mALD

FBP

HIR (ASIR 30%)

30

38.8pf/29.0cer SD

27.7pf/20.2cer LD

466.5SD

329.2LD

2.2SD

1.5LD

Korn [57] 2012

90

30SD

30LD1

30LD2

SD vs. LD1 vs. LD2 128-MDCT (Siemens SOMATOM Definition Flash, Siemens Healthineers, Erlangen, Germany)

120 kV, 320 mAsSD

120 kV, 275 mAs LD1

120 kV, 225 mAs LD2

FBP

IRIS

15SD-LD1

30SD-LD2

60.1SD

51.8LD1

42.3LD2

1043.0SD

890.0LD1

733.0LD2

2.2SD

1.8LD1

1.5LD2

Rapalino [32] 2012

150

50SD

100LD

SDFBP vs. LDASIR 64-MDCT (GE Discovery CT750HD, GE Healthcare, Milwaukee, WI, USA)

120 kV, 250 mA, 175 mAsSD

120 kV, 200 mA, 140 mAsLD

FBP

HIR (ASIR 20–100%)

26

66.5SD

49.7LD

1270.3SD

932.3LD

2.7SD

2.0LD

Becker [31] 2012

150

50SD

50LD1

50LD2

SDFBP vs. LD1FBP vs. LD2IRIS MDCT (Siemens Somatom Definition Flash)

120 kV, 320 mAsSD

120 kV, 390 mAsam, LD1, LD2

FBP

IRIS

24CTDI

20DLP

60.0SD

46.0LD1

45.0LD2

887.0SD

722.0LD1

708.0LD2

NA
Kilic [30] 2011

149

51SD

98LD

SDFBP vs. LDASIR 16-MDCT (GE Brightspeed, GE Healthcare, Milwaukee, WI, USA)

140pf/120cer kV, 170pf/270cer mA SD

140pf/120cer kV, 125pf/190cer mA SD

FBP

HIR (ASIR 30%)

26pf, CTDI

35cer, CTDI

31DLP

93.5pf, SD

59.4cer, SD

69.1pf, LD

38.6cer, LD

1081.3SD

748.6LD

2.3SD

1.6LD

am automatic tube current modulation, ASIR adaptive statistical iterative reconstruction, ASIR20/30 adaptive statistical iterative reconstruction at level 20/30/40% (available options: 0–50% in 10% increments), cer cerebrum, CTDI based on CTDIvol, FBP filtered back projection, HIR hybrid iterative reconstruction, IRIS iterative reconstruction in image space, LD low-dose, LD1 low-dose protocol 1, LD2 low-dose protocol 2, pf posterior fossa, SAFIRE sinogram-affirmed iterative reconstruction, SAFIRE1–5 SAFIRE at level 1–5, SD standard-dose, SD-LD1 dose reduction between standard-dose protocol and low-dose protocol 1, SD-LD2 dose reduction between standard-dose protocol dose and low-dose protocol