TABLE 1.
Specimens with discordant results by RIBA HCV 2.0 SIA for RIBA Processor System versus manual assaya
Specimen | RIBA Processor System
|
RIBA manual assay
|
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Rating (score)
|
Interp | Rating (score)
|
Interp | |||||||||
5-1-1 | c100-3 | c33-c | c22-3 | SOD | 5-1-1 | c100-3 | c33-c | c22-3 | SOD | |||
B | ± (0.36) | − (0.05) | ± (0.66) | − (0.05) | − (0.00) | NEG | ± (0.78) | − (0.07) | 2+ (1.18) | − (0.07) | − (0.01) | IND |
A | − (0.01) | − (0.01) | − (0.05) | 1+ (0.85) | − (0.01) | IND | − (0.04) | − (0.05) | − (0.03) | ± (0.59) | − (0.06) | NEG |
Ratings: −, 0.00 to 0.09; ±, 0.10 to 0.84; 1+ 0.85 to 1.14; 2+, 1.15 to 2.69. Abbreviations: SOD, superoxide dismutase; Interp, interpretation; NEG, negative; IND, indeterminate.