Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) operating parameters.

Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) operating parameters
RF power 1150 W
Nebulizer gas flow 12.5 L/min
Cool gas flow 12.5 L/min
Nebulizer gas pressure 0.2 L/min
Auxiliary gas flow 0.5 L/min
Pump speed 45 rpm
Metals Emission waveleghts (nm) Detection limits (mg/L) Quantification limits (mg/L)
Al 167.0 0.005 0.015
B 249.6 0.008 0.027
Ba 455.4 0.0006 0.002
Ca 315.8 1.629 5.432
Co 228.6 0.001 0.005
Cr 267.7 0.001 0.005
Cu 324.7 0.003 0.011
Fe 238.2 0.004 0.013
K 766.4 1.764 5.883
Li 670.7 0.013 0.031
Mg 383.8 1.580 5.268
Mn 257.6 0.0008 0.003
Mo 202.0 0.0016 0.005
Na 818.3 2.221 7.404
Ni 221.6 0.0009 0.003
Sr 407.7 0.003 0.011
V 292.4 0.0014 0.004
Zn 213.8 0.0027 0.009