Table 1. Photobiomodulation parameters according to the “analgesic, anti-inflammatory and healing” protocol of the ATP38® medical device .

Cold lights combination
Blue Green Amber Red Deep red Infrared
Wavelengths (nm) 470 525 590 620 680-760 820
Duration (s) 151/151/149 182/182/180 241/241/239 244/244/242 162/162/242 219/219/217
Fluency (J/cm2) 2/2/2 1/1/1 0.8/0.8 /0.8 2/2/2 4/4/4 4/4/4
Frequency (Hz) 70/5/0a
Spot dimension area 600 cmb
Distance from the target 4 cm
Total energy delivery per sessionc 124.2 J/cm2

Note:When necessary specific values are given for each step (1 to 3).

a A null frequency (0 Hz) means continuous irradiation.

b The spot dimension is calculated as the spot dimension of each panel (10 cm × 20 cm) multiplied by the number of used panel (3)