Table 4.

Comparison of the deviations observed in surface analysis between 30 mm mouth opening and 50 mm mouth opening conditions

Intra oral scanner Arch Mouth opening P SE CI t P

Lower bound Upper bound
Parallel confocal microscopy Upper 30 mm mouth opening 0.59±0.15 0.06 0.42 0.75 0.39 0.70
50 mm mouth opening 0.55±0.14 0.05 0.40 0.70
Lower 30 mm mouth opening 0.36±0.08 0.03 0.27 0.45 0.91 0.38
50 mm mouth opening 0.42±0.14 0.05 0.27 0.57
3D in motion video technology with triangulation Upper 30 mm mouth opening 0.41±0.10 0.04 0.31 0.52 0.10 0.91
50 mm mouth opening 0.41±0.13 0.05 0.26 0.55
Lower 30 mm mouth opening 0.21±0.06 0.02 0.14 0.28 0.56 0.58
50 mm mouth opening 0.23±0.07 0.02 0.16 0.31

P value derived from paired sample t-test. 3D: Three dimensional, SD: Standard deviation, CI: Confidence interval, SE: Standard error