Quality control testing of CHROMagar Orientation used for the evaluation

Test microorganism Mean colony count (CFU) ± SD with:
Pa
Reference mediumb CHROMagar Orientationc
E. coli ATCC 25922 174 ± 98 166 ± 47 0.511
K. pneumoniae ATCC 13883 210 ± 77 230 ± 99 0.747
P. aeruginosa ATCC 27853 138 ± 15 128 ± 11 0.084
P. mirabilis ATCC 4630 110 ± 34 128 ± 43 0.132
S. aureus ATCC 25923 129 ± 69 132 ± 68 0.219
E. faecalis ATCC 19433 168 ± 69 174 ± 75 0.468
a

No significant differences were found between CHROMagar Orientation and the reference medium (P > 0.05 [paired t test]). 

b

The reference medium was TSA plus 5% defibrinated sheep blood. The inoculum was diluted to an estimated concentration of 102 CFU/ml. Four batches were tested. 

c

The inoculum was diluted to a concentration of 102 CFU/ml. Seven batches were tested.