Skip to main content
. 2023 Aug 17;22(9):2950–2958. doi: 10.1021/acs.jproteome.3c00272

Table 4. Precision (%CV) and Accuracy (%RE) Results for Quality Control Samples (n = 6) of ppCCK(21–44), Prepared in Pooled Blank Human Plasma and Quantified against a Calibration Curve.

spiked concentration (pg/mL) 1 3 15 150
mean 0.9 2.5 13.9 159.9
standard deviation 0.1 0.3 0.7 5.1
%CV 10.5 10.5 5.0 3.2
%RE –9.5 –17.1 –7.6 6.7