Skip to main content
. 2023 Jun 12;4(8):1095–1102. doi: 10.34067/KID.0000000000000185

Table 2.

Summary of dialysis characteristics for each study phase

Variables MCO-HD HFX-HD MCO-HDF HFX-HDF
Qb (ml/min) 367.1±23.9 371.7±27.1 371.6±15.9 365.7±25.5
Qd (ml/min) 538.3±52.0c 547.1±48.0 449.4±18.6b,d 448.8±29.9a
TMP (mm Hg) 29.6±4.5c 33.8±11.5 175.2±25.8b,d 168.7±22.3a
RBV minimum (%) 81.5±6.0 81.9±5.9 79.9±5.2 79.3±5.5
spKt/V 1.6±0.3c 1.7±0.5 1.8±0.4d 1.9±0.4
Intradialytic weight change (kg) 2.4±0.8 2.3±0.8 2.5±0.8 2.4±0.8

Data presented as mean±SD.

All comparisons executed by using the pairwise dependent sample t test. Study phases: MCO-HD, hemodialysis using a medium cutoff dialyzer membrane; HFX-HD, high-flux hemodialysis; MCO-HDF, hemodiafiltration using a medium cutoff dialyzer membrane; HFX-HDF, high-flux hemodiafiltration. Qb, blood flow; Qd, dialysate flow; TMP, transmembranous pressure; RBV, relative blood volume; UFV, ultrafiltration cutoff

Hypothesis testing:

a

P < 0.001 comparing HFX-HDF with HFX-HD.

b

P < 0.001 comparing MCO-HDF with HFX-HD.

c

P < 0.001 comparing MCO-HD with HFX-HDF.

d

P < 0.001 comparing MCO-HDF with MCO-HD.