Skip to main content
. 2023 Jun 7;24(9):e14054. doi: 10.1002/acm2.14054

TABLE 2.

Optically stimulated luminescence dosimeter (OSLD) measurements of superficial dose

Bolus Average OSLD reading (cGy) a % of Rx
Plan A

No bolus

3 mm TEB

BMB

151.5

205.5

194

75.8

102.8

97.0

Plan B

No bolus

3 mm TEB

BMB

152

205

195

76.0

102.5

97.5

Plan C

No bolus

3 mm TEB

BMB

155

207.5

199.5

77.5

103.8

99.8

a

Overall uncertainty in OSLD measurement is approximated to be ±5.9% of the measured value.