TABLE 3.

Summary of treatment delivery parameters evaluated for TrueBeam and Halcyon plans for APBI patients.

Treatment and QA Parameters TrueBeam VMAT Halcyon VMAT p
Total Monitor Units (MU) per fraction 921 ± 114 (701−1025) 1188 ± 210 (762−1495) 0.001
Modulation factor (MF) 1.77 ± 0.22 (1.35−1.97) 2.28 ± 0.40 (1.46−2.87) 0.001
Beam‐on time (min) 1.68 ± 0.21 (1.27−1.86) 1.49 ± 0.26 (0.95−1.87) 0.036
VMAT QA pass rate (%) 97.85 ± 2.63 (91.5−100) 99.6 ± 0.39 (98.8−100) 0.053
Monte Carlo 2nd check pass rate (%) 99.24 ± 0.43 (98.7−99.9) 98.6 ± 0.52 (97.9−99.5) 0.001

Note: Statistically significant values are highlighted in bold.