RF pulse parameters to establish the safe exposure limits.

Frequency Duration TR
64 MHz 603 s 24.03 ms
Patient weight entered Patient weight entered
68.0 kg 13.6 kg
Flip Angle (°) B1 + rms [µT] Flip Angle (°) B1 + rms [µT]
360 4.80 360 4.82
330 4.40 330 4.41
300 4.00 295 3.95
270 3.60 285 3.81
240 3.20 278 3.72
210 2.80 270 3.61
180 2.40 265 3.55
170 2.27 245 3.28
150 2.00 235 3.15
225 3.01
205 2.74
195 2.61
185 2.48
175 2.34