Incision depth, vaporization area, coagulation area, and total laser area for Th:YAG versus Ho:YAG lasers upon porcine renal urothelium ex vivo.

Th:YAG Ho:YAG p
Power, mean ± SD, W 15.21 ± 8.61 14.65 ± 7.79 0.930
ID, mean ± SD, mm 0.346 ± 0.120 0.458 ± 0.194 0.024*
VA, mean ± SD, mm2 0.070 ± 0.45 0.066 ± 0.46 0.572
CA, mean ± SD, mm2 0.123 ± 0.020 0.066 ± 0.035 0.001*
TLA, mean ± SD, mm2 0.264 ± 0.146 0.125 ± 0.055 0.005*

CA = coagulation area; ID = incision depth; TLA = total laser area; VA = vaporization area.

*p < 0.05.