Main features provided by the most common detectors coupled online with FFF techniques.

Detector Features
UV-Vis/DAD
  • Spectroscopical Characterization

  • Concentration detector

  • Studies on evolving system (drug loading and releasing, conjugations…)

  • Versatile signal (a specific)

FLD
  • Spectroscopical Characterization

  • Studies on evolving system (drug loading and releasing, conjugations…)

  • Very selective signal (quite specific)

  • Stable signal less prone to drifts and interference

dRI
  • Concentration detector

  • Unstable signal significantly affected by pressure changes

  • Universal signal

MALS
  • Requires concentration signal

  • Accurate determination of molecular weight/molar mass, radius of gyration (Rg) and fractal radius (δ)

  • Information on the aggregation rate of nanosystems

  • Information of the shape of nanoparticles (conformation plot)

  • Working size range(1 nm–100 μm)

DLS
  • Determination of hydrodynamic radius (Rh)

  • Estimation of the shape factor (Rg/Rh)

  • Working size range(1 nm–100 μm)

ICP-MS
  • Studies on evolving system by monitoring elemental composition

  • Low detection limit (ng/L)

High-Resolution MS
  • Accurate information on molar mass of molecules and complex