Skip to main content
. 2023 Aug 29;23(17):7499. doi: 10.3390/s23177499

Table 24.

Comparison of results of IMU1, IMU2, and Propriometer in 90 degrees of shoulder abduction; Day 1 vs. Day 2. N = 90.

ABD 90° Day 1 [°] Day 2 [°] Difference (D1-D2) [°] LOA [°]
IMU1 91 ± 1 * 92 ± 2 * 1 ± 1 −4; 3
IMU2 98 ± 3 * 98 ± 3 * 0 ± 1 −6; 7
Propriometer 92 ± 2 * 91 ± 2 * 1 ± 0 −3; 5

* significant difference between goniometer and given device; LOA—Limits of Agreement.