Table 25.
Comparison of results of IMU1, IMU2 and Propriometer in 45 degrees of shoulder internal rotation; Day 1 vs. Day 2. N = 90.
InR 45° | Day 1 [°] | Day 2 [°] | Difference (D1-D2) [°] | LOA [°] |
---|---|---|---|---|
IMU1 | 47 ± 2 * | 46 ± 2 * | 1 ± 1 | −4; 6 |
IMU2 | 39 ± 3 * | 39 ± 4 * | 0 ± 1 | −9; 7 |
Propriometer | 48 ± 2 * | 47 ± 2 * | 1 ± 0 | −3; 6 |
* significant difference between goniometer and given device; LOA—Limits of Agreement.