Figure 8.

Figure 8.

Point-wise correlation of total deviation (TD) and pattern deviation (PD) sensitivities at each testing location between the Humphrey Field Analyzer (HFA) and Smart Visual Function Analyzer (SVFA). Highly correlated visual field locations are demonstrated in darker colors and less correlated locations are demonstrated in decreasing color saturation. (a) ICC point-by-point plot and boxplot of TD between HFA and SVFA. (b) ICC point-by-point plot and boxplot of PD between HFA and SVFA.