Figure 1.

Figure 1

Acquisition protocol for 5D relaxation–diffusion MRI. (a) Pulse sequence for acquiring images encoded for relaxation and diffusion in a 5D space defined by the echo time τE , and b -tensor trace b , normalized anisotropy bΔ , and orientation ( Θ , Φ ). An EPI image readout block acquires the spin echo produced by slice-selective 90 and 180 radio-frequency pulses. The 180 pulse is encased by a pair of gradient waveforms allowing for diffusion encoding according to principles from multidimensional solid-state NMR (Topgaard, 2017) (red, green, and blue lines). The signal is encoded for the transverse relaxation rate R2 by varying the value of τE . (b) Numerically optimized gradient waveforms (Sjölund et al., 2015) yielding four distinct b -tensor shapes ( bΔ=-0.5 , 0.0, 0.5, and 1) (Eriksson et al., 2015).