Table 3:
Total effective dose rate at the points of interest and comparison to FLUKA calculations for single layer and multilayered collimator designs, occupancy factors were not included.
| Position | NCRP Report 151 Calculation (mSv w−1) |
FLUKA Calculation (Single
Layer) (mSv w−1) |
FLUKA calculation (Multilayer)
(mSv w−1) |
|---|---|---|---|
| Control Room (KM w/Attenuation Factor from FLUKA) |
9.1 × 10−2 | 5.90 × 10−2 | 4.84 × 10−2 |
| Control Room (MKM w/Attenuation Factor from FLUKA) |
1.6 × 10−1 | ||
| Control Room (MKM w/ Eq 1) |
2.0 × 10−1 | ||
| Control Room (MKM w/ Attenuation Factor and ϕA from FLUKA) |
1.3 × 10−1 | ||
| Mechanical Room | 8.3 × 10−3 | 1.36 × 10−3 | 1.21 × 10−3 |
| Public Sitting Area | 1.0 × 10−4 | 4.04 × 10−4 | 2.03 × 10−4 |
| Office Space | 4.8 × 10−2 | 4.47 × 10−3 | 1.52 × 10−3 |
| Office Space 1st Floor | 4.6 × 10−2 | 1.50 × 10−3 | 9.54 × 10−4 |
| Garden Area | 1.1 × 10−4 | 1.96 × 10−5 | 1.02 × 10−5 |