Table 3:

Total effective dose rate at the points of interest and comparison to FLUKA calculations for single layer and multilayered collimator designs, occupancy factors were not included.

Position NCRP Report 151
Calculation
(mSv w−1)
FLUKA Calculation (Single Layer)
(mSv w−1)
FLUKA calculation (Multilayer)
(mSv w−1)
Control Room
(KM w/Attenuation Factor from FLUKA)
9.1 × 10−2 5.90 × 10−2 4.84 × 10−2
Control Room
(MKM w/Attenuation
Factor from FLUKA)
1.6 × 10−1
Control Room
(MKM w/ Eq 1)
2.0 × 10−1
Control Room
(MKM w/ Attenuation
Factor and ϕA from
FLUKA)
1.3 × 10−1
Mechanical Room 8.3 × 10−3 1.36 × 10−3 1.21 × 10−3
Public Sitting Area 1.0 × 10−4 4.04 × 10−4 2.03 × 10−4
Office Space 4.8 × 10−2 4.47 × 10−3 1.52 × 10−3
Office Space 1st Floor 4.6 × 10−2 1.50 × 10−3 9.54 × 10−4
Garden Area 1.1 × 10−4 1.96 × 10−5 1.02 × 10−5