Table 6.

Comparison of chromium (Cr) concentrations of spiked rumen fluid samples containing high (0.50 g/L), medium-high (0.25 g/L), medium-low (0.10 g/L), or low (0.005 g/L) concentrations of Cr as measured using an external calibration curve or the internal calibration of an energy dispersive X-ray fluorescence analyzer. The percent difference of the resulting concentration as extrapolated from an external calibration curve or internal calibration to determine which calibration method was more accurate

Cr concentration, g/L Percent difference Recovery, % Intra-assay CV Inter-assay CV
External calibration curve Internal calibration SEM1 P-value External calibration curve Internal calibration
High, 0.50 g/L 0.474 0.492 0.0001 <0.001 3.59 1.08 94.8 ± 5.64 3.32 5.83
Medium-High, 0.25 g/L 0.233 0.246 0.0003 <0.001 4.75 1.08 114.9 ± 22.80 2.63 2.45
Medium-Low, 0.10 g/L 0.109 0.120 0.0001 <0.001 5.66 11.76 108.6 ± 18.08 4.69 15.30
Low, 0.005 g/L 0.003 0.013 0.0006 <0.001 36.36 51.61 64.6 ± 11.23 11.02 10.21

1Standard error of the mean (n = 60).