Table 6.
Comparison of chromium (Cr) concentrations of spiked rumen fluid samples containing high (0.50 g/L), medium-high (0.25 g/L), medium-low (0.10 g/L), or low (0.005 g/L) concentrations of Cr as measured using an external calibration curve or the internal calibration of an energy dispersive X-ray fluorescence analyzer. The percent difference of the resulting concentration as extrapolated from an external calibration curve or internal calibration to determine which calibration method was more accurate
| Cr concentration, g/L | Percent difference | Recovery, % | Intra-assay CV | Inter-assay CV | |||||
|---|---|---|---|---|---|---|---|---|---|
| External calibration curve | Internal calibration | SEM1 | P-value | External calibration curve | Internal calibration | ||||
| High, 0.50 g/L | 0.474 | 0.492 | 0.0001 | <0.001 | 3.59 | 1.08 | 94.8 ± 5.64 | 3.32 | 5.83 |
| Medium-High, 0.25 g/L | 0.233 | 0.246 | 0.0003 | <0.001 | 4.75 | 1.08 | 114.9 ± 22.80 | 2.63 | 2.45 |
| Medium-Low, 0.10 g/L | 0.109 | 0.120 | 0.0001 | <0.001 | 5.66 | 11.76 | 108.6 ± 18.08 | 4.69 | 15.30 |
| Low, 0.005 g/L | 0.003 | 0.013 | 0.0006 | <0.001 | 36.36 | 51.61 | 64.6 ± 11.23 | 11.02 | 10.21 |
1Standard error of the mean (n = 60).