. 2023 Apr 1;33(7):3059–3065. doi: 10.1007/s00590-023-03538-6

Table 1.

Radiation dose experienced during hip replacement and hip arthroscopy surgery

Staff member Distance from source [m] Dose rate (mSv/h) Dose rate with lead apron (mSv/h) Hip replacement Hip arthroscopy
Cases per year to reach 1 mSv*
No lead apron
Cases per year to reach 1 mSv*
With lead apron
Cases per year to reach 1 mSv*
No lead apron
Cases per year to reach 1 mSv*
With lead apron
Surgeon 0.5 3.1227 0.0390 36 2870 64 5144
Surgical assistant 0.7 0.6846 0.0086 164  > 10,000 293  > 10,000
Anaesthetist 1.5 0.0453 0.0006 2472  > 10,000 4429  > 10,000
Anaesthetic nurse 1.2 0.1259 0.0016 890  > 10,000 1595  > 10,000
Instrument nurse (Scrub) 1.5 0.9080 0.0114 123 9872 221  > 10,000
Circulating nurse (Scout) 1.7 0.6984 0.0087 160  > 10,000 287  > 10,000