Table 1.
Radiation dose experienced during hip replacement and hip arthroscopy surgery
| Staff member | Distance from source [m] | Dose rate (mSv/h) | Dose rate with lead apron (mSv/h) | Hip replacement | Hip arthroscopy | ||
|---|---|---|---|---|---|---|---|
| Cases per year to reach 1 mSv* No lead apron |
Cases per year to reach 1 mSv* With lead apron |
Cases per year to reach 1 mSv* No lead apron |
Cases per year to reach 1 mSv* With lead apron |
||||
| Surgeon | 0.5 | 3.1227 | 0.0390 | 36 | 2870 | 64 | 5144 |
| Surgical assistant | 0.7 | 0.6846 | 0.0086 | 164 | > 10,000 | 293 | > 10,000 |
| Anaesthetist | 1.5 | 0.0453 | 0.0006 | 2472 | > 10,000 | 4429 | > 10,000 |
| Anaesthetic nurse | 1.2 | 0.1259 | 0.0016 | 890 | > 10,000 | 1595 | > 10,000 |
| Instrument nurse (Scrub) | 1.5 | 0.9080 | 0.0114 | 123 | 9872 | 221 | > 10,000 |
| Circulating nurse (Scout) | 1.7 | 0.6984 | 0.0087 | 160 | > 10,000 | 287 | > 10,000 |