Table 2. . Accuracy and precision of LB-100 and endothall at the lower limit in six individual lots of human plasma matrices.

Analyte Plasma matrix Nominal concentration (ng/ml) Mean measured (ng/ml) SD (ng/ml) Precision (%CV) Accuracy (%RE)
LB-100 Lot 1 2.50 2.36 0.19 8 -6
Lot 2 2.50 2.58 0.34 13 3
Lot 3 2.50 2.25 0.29 13 -10
Lot 4 2.50 2.33 0.12 5 -7
Lot 5 2.50 2.35 0.15 6 -6
Lot 6 2.50 2.23 0.14 6 -11
Endothall Lot 1 2.50 2.53 0.14 6 1
Lot 2 2.50 2.39 0.33 14 -4
Lot 3 2.50 2.51 0.32 13 1
Lot 4 2.50 2.35 0.21 9 -6
Lot 5 2.50 2.49 0.13 5 -1
Lot 6 2.50 2.40 0.15 6 -4

%CV = (SD/mean) × 100%

%RE = [(measured – nominal)/nominal] × 100%

CV: Coefficient of variation; RE: Relative error; SD: Standard deviation.