Table 2. . Accuracy and precision of LB-100 and endothall at the lower limit in six individual lots of human plasma matrices.
| Analyte | Plasma matrix | Nominal concentration (ng/ml) | Mean measured (ng/ml) | SD (ng/ml) | Precision† (%CV) | Accuracy‡ (%RE) |
|---|---|---|---|---|---|---|
| LB-100 | Lot 1 | 2.50 | 2.36 | 0.19 | 8 | -6 |
| Lot 2 | 2.50 | 2.58 | 0.34 | 13 | 3 | |
| Lot 3 | 2.50 | 2.25 | 0.29 | 13 | -10 | |
| Lot 4 | 2.50 | 2.33 | 0.12 | 5 | -7 | |
| Lot 5 | 2.50 | 2.35 | 0.15 | 6 | -6 | |
| Lot 6 | 2.50 | 2.23 | 0.14 | 6 | -11 | |
| Endothall | Lot 1 | 2.50 | 2.53 | 0.14 | 6 | 1 |
| Lot 2 | 2.50 | 2.39 | 0.33 | 14 | -4 | |
| Lot 3 | 2.50 | 2.51 | 0.32 | 13 | 1 | |
| Lot 4 | 2.50 | 2.35 | 0.21 | 9 | -6 | |
| Lot 5 | 2.50 | 2.49 | 0.13 | 5 | -1 | |
| Lot 6 | 2.50 | 2.40 | 0.15 | 6 | -4 |
%CV = (SD/mean) × 100%
%RE = [(measured – nominal)/nominal] × 100%
CV: Coefficient of variation; RE: Relative error; SD: Standard deviation.