Table 6. . Inter-run accuracy and precision for LB-100 and endothall in pooled human plasma.
| Analyte | Nominal concentration (ng/ml) | Mean measured (ng/ml) | SD (ng/ml) | Precision (%CV) | Accuracy (%RE) |
|---|---|---|---|---|---|
| LB-100 | 2.50 | 2.41 | 0.14 | 6 | -4 |
| 8.00 | 8.41 | 0.23 | 3 | 7 | |
| 40.0 | 43.4 | 1.0 | 2 | 9 | |
| 400 | 402 | 10.0 | 2 | 0 | |
| Endothall | 2.50 | 2.34 | 0.09 | 4 | -6 |
| 8.00 | 7.99 | 0.21 | 3 | 4 | |
| 40.0 | 40.1 | 0.5 | 1 | 0 | |
| 400 | 413 | 9.0 | 2 | 3 |
Each datum point was calculated by five parallel measurements of five identical quality controls at each concentration over three validation batches.
CV: Coefficient of variation; RE: Relative error; SD: Standard deviation.