Table 6. . Inter-run accuracy and precision for LB-100 and endothall in pooled human plasma.

Analyte Nominal concentration (ng/ml) Mean measured (ng/ml) SD (ng/ml) Precision (%CV) Accuracy (%RE)
LB-100 2.50 2.41 0.14 6 -4
8.00 8.41 0.23 3 7
40.0 43.4 1.0 2 9
400 402 10.0 2 0
Endothall 2.50 2.34 0.09 4 -6
8.00 7.99 0.21 3 4
40.0 40.1 0.5 1 0
400 413 9.0 2 3

Each datum point was calculated by five parallel measurements of five identical quality controls at each concentration over three validation batches.

CV: Coefficient of variation; RE: Relative error; SD: Standard deviation.