Table 8. . Stability studies of LB-100 and endothall under various conditions.
| Analyte | Stability recoveries, % (mean ± SD) | ||||
|---|---|---|---|---|---|
| Auto-sampler (4°C, 24 h) | Benchtop (RT, 24 h) | Long-term (-20°C, 60 days) | Freeze–thaw (-20°C after three cycles) | ||
| LB-100 | Plasma LQC | 102 ± 1 | 100 ± 2 | 106 ± 4 | 104 ± 2 |
| Plasma HQC | 97 ± 2 | 104 ± 4 | 92 ± 2 | 92 ± 4 | |
| Endothall | Plasma LQC | 96 ± 1 | 103 ± 5 | 102 ± 5 | 106 ± 6 |
| Plasma HQC | 91 ± 3 | 103 ± 3 | 99 ± 2 | 100 ± 5 | |
HQC: High-concentration quality control; LQC: Low-concentration quality control; RT: Room temperature; SD: Standard deviation.