Skip to main content
. 2023 Aug 16;15(17):1095–1107. doi: 10.4155/bio-2023-0078

Table 8. . Stability studies of LB-100 and endothall under various conditions.

Analyte Stability recoveries, % (mean ± SD)
    Auto-sampler (4°C, 24 h) Benchtop (RT, 24 h) Long-term (-20°C, 60 days) Freeze–thaw (-20°C after three cycles)
LB-100 Plasma LQC 102 ± 1 100 ± 2 106 ± 4 104 ± 2
Plasma HQC 97 ± 2 104 ± 4 92 ± 2 92 ± 4
Endothall Plasma LQC 96 ± 1 103 ± 5 102 ± 5 106 ± 6
Plasma HQC 91 ± 3 103 ± 3 99 ± 2 100 ± 5

HQC: High-concentration quality control; LQC: Low-concentration quality control; RT: Room temperature; SD: Standard deviation.