. 1998 Oct;36(10):2964–2969. doi: 10.1128/jcm.36.10.2964-2969.1998

TABLE 1.

Limit of detection for the ultrasensitive and standard HIV-1 MONITOR Testsa

Actual input concn (copies/ml) No. of replicates Standard test
Ultrasensitive test
No. (%) positive Avg no. of copies/mlb No. (%) positive Avg no. of copies/mlb
1,000 3 3 (100) 1,213 3 (100) 1,190
500 3 3 (100) 697 3 (100) 612
100 6 2 (33) 265 6 (100) 134
50 7 1 (14) 390 7 (100) 67
40 7 2 (29) 271 6 (86) 90
30 7 0 (0) Not detected 7 (100) 61
20 7 0 (0) Not detected 5 (71) 55
0 6 0 (0) Not detected 0 (0) Not detected
a

The limit of detection was determined by using the ultra-low-level HIV RNA proficiency panel described in Materials and Methods. 

b

The viral RNA concentration in each replicate reaction yielding a positive signal was calculated as described in Materials and Methods. The values presented here are the averages of the individual values for the positive replicates. Reactions yielding negative results were not included in the average.